Upload
bile86
View
17
Download
1
Tags:
Embed Size (px)
DESCRIPTION
.
Citation preview
HPLC can be used better…Great tips for better
results! Shimadzu Corporation
Analytical & Measuring Instruments Division
JAIMA SHOW 2003 -
New Technology Briefing
( September 11, 2003 Makuhari Messe )
• Mobile phase– Buffer solution / solvent selection, gradient optimization, flow rate selection
– Degassing, flow rate stability, mobile phase preparation
• Column– High separation, secondary interaction, lot variation, temperature control
• Detection– Sensitivity, selectivity, ambient temperature fluctuation affect
• Sample– Extraction method selection, sample solvent selection, dissolved oxygen
• Instrument– Performance, stability, durability, ease-of-use
• Evaluation overall– Stability of analytical technique, robustness, quantitative accuracy
For better analysis in HPLC…
Mobile Phase Solvent Mixing Ratio Notation and Preparation (1)
• Is the composition of the following solutions the same?– water / ethanol = 1/1 (v/v)
– 50%(v/v) ethanol aqueous solution
* Beware of volume change when mixing organic solvents!
– The density of the solvent mixture is not a simple average of the original solvent densities.
– If 50mL of water and 50mL of ethanol are mixed, the final volume becomes about 96mL at about 25ºC.
• water/ethanol=1/1 (v/v)
– Measure 500mL each of water and ethanol, and then mix
– The total is not 1L.
• 50%(v/v) ethanol aqueous sol.– Transfer 500mL ethanol to 1L
volumetric flask, bring to volume with water.
– The water percentage is high.
1LWater Ethanol
EthanolWater
Mobile Phase Solvent Mixing Ratio Notation and Preparation (2)
Mobile phase A: 20mM phosphoric acid (Na) buffer solution <pH 2.5> / acetonitrile = 9 / 1 (v/v)
Mobile phase B : 10%(v/v) acetonitrile in 20mM phosphoric acid buffer solution <pH 2.5>
1
2
3
31
A
B
2Peaks 1: acetaminophen 2: dihydrocodeine 3: caffeine
0 2 4 6 8 10min
Column :Shim-Pack VP-ODS (150mmL. x 4.6mmI.D.)
Flow rate :1mL/min
Temp. :40ºC
Detection :UV-VIS (210nm)
Mobile Phase Solvent Mixing Ratio Notation and Preparation (3)
• Preparation: 20mM Phosphoric acid buffer solution (pH2.5)
Mobile Phase Buffer Solution Notation and Preparation (1)
• A : “20mM” interpreted as phosphoric acid concentration Mix 10mmol phosphoric acid and 10mmol sodium dihydrogen
phosphate, dissolve in 1L water
• B : “20mM” interpreted as sodium concentration Dissolve 20mmol sodium dihydrogen phosphate in 1L water,
add phosphoric acid to adjust pH to 2.5
• C : “20mM” interpreted as phosphoric acid and sodium concentrations, adjust pH with perchloric acid Dissolve 20mmol sodium dihydrogen phosphate in 1L water,
add perchloric acid to adjust pH to 2.5
Column :Shim-Pack VP-ODS (150mmL. x 4.6mmI.D.)
Flow rate :1mL/min
Temp. :40ºC
Detection :UV-VIS (210nm)
Mobile phase A: phosphoric acid interpreted as 20mMMobile phase B: sodium interpreted as 20mMMobile phase C: phosphoric acid, sodium both
interpreted as 20mM, pH adjusted with perchloric acid
Mobile Phase Buffer Solution Notation and Preparation (2)
min
1
2
3
31
A
B
2
Peaks 1: acetaminophen 2: dihydrocodeine 3: caffeine
0 2 4 6 8 10
C
3
1,2
If interpreted wrong…Dihyrocodeine elution position is different!!
If mobile phase buffer capacity is weak…Peak shapes of acids and bases with pKa near mobile phase pH deteriorate.
Column : STR ODS-II (150mmL. x 4.6mm I.D.)
Mobile : 20mM buffer solution (pH 4.5)phase / acetonitrile (3/1, v/v)
Flow rate : 1mL/min
Temp. : 40ºC
Detection : UV-VIS (240nm)
Left figure: citric acid buffer solution (pH 4.5) used as mobile phaseRight figure: phosphoric acid buffer solution (pH 4.5) used as mobile phase
Mobile Phase pH Buffer Capacity and Peak Shape
• Absorbance HPLC grade acetonitrile has low absorbance at short UV wave lengths
• HPLC-grade acetonitrile good for hi-sens analysis at short UV wavelengths!
• Column pressure For Acetonitrile, lower pressure is sufficient. • At same flow rate, extra pressure need not be applied to column.
• Elution strength Acetonitrile generally stronger. • Some compounds strongly eluted with methanol. (carotene, cholesterol)
• Selectivity Both show differences.
• Peak shape Both show differences. • With polymer-type columns, acetonitrile is better.
• Mobile phase degassing efficiency Caution when mixing with water •Methanol…… Exothermic degassing •Acetonitrile… Endothermic gas dissolution
Mobile Phase Differences between Acetonitrile and Methanol
• Retention time change• Peak area change Quantitation Error!!
– Detector that responds absolute quantity of compound Peak area is unaffected by flow rate.
– Detector that responds to concentration of compound Peak area is affected by flow rate.
*Almost all LC detectors are concentration-response type
Mobile Phase Effect of Flow Rate Change (1)
1.009729145912886547n-butylparaben
1.009826520872626362n-propylparaben
1.009627710872744605ethylparaben
1.009539382133901107methylparaben
B/AB (0.99mL/min)A (1.00mL/min)
Area RatioPeak Area
Column : Shim-Pack VP-ODS (150mm L. x 4.6mm I.D.)
Mobile phase : methanol / water = 6/4 (v/v)
Flow rate : A ; 1.00mL/min B ; 0.99mL/min
Temperature : 40ºC
Detection : UV-VIS (260nm)
Mobile Phase Effect of Flow Rate Change (2)
When mixing is performed with 2 pumps…
A
B
A : Pre-mixed solutions (2) delivered using one pumpB : Each solvent delivered with a separate pump and mixed
Peaks 1: methylparaben 2: ethylparaben 3: propylparaben1
2
3
Mobile Phase Flow Rate Change Influence in High Pressure Gradient Instrument
(1)
• Retention time is different from when mobile phase is pre-mixed in a bottle.
1.01330752593035538n-butylparaben
1.01327971802761054n-propylparaben
1.01329225562885602ethylparaben
1.01341518764100127methylparaben
B/ABA
Area RatioPeak Area
Column : Shim-Pack VP-ODS (150mmL. x 4.6mm I.D.)
Mobile phase : methanol / water (6/4, vol/vol)
Flow rate : 1.00mL/min
Temperature : 40ºC
Detection : UV-VIS (260nm)
•A : Pre-mixed solutions (2) delivered using one pump
•B : Each solvent delivered with a separate pump and mixed
Mobile Phase Flow Rate Change Influence in High Pressure Gradient Instrument (1)
• Causes of bubble occurrence– Occurs when dissolved air exceeds air dissolution saturation point. Causes: Solution warming, decreasing pressure, solvent mixing, etc.
• Possible Troubles caused by bubbles– Mobile phase, pump
• Retention time fluctuation, unstable solvent delivery
– Column• Peak deformation, column degradation
– Detector• Noise generation
• Countermeasure: Offline and online degassing However, the problems caused by air does not just come from
bubbles, but from dissolved air as well.
Mobile Phase Troubles caused by Bubbles Formed in Mobile Phase
• Fluorescence detection– Quenching due to oxygen
→ Changes in peak area
• Absorption detection– Changes in background absorbance
→ Baseline fluctuation (especially with gradient)
– Appearance of peaks originating from oxygen
• Refractive index detection– Baseline fluctuation
Mobile Phase Influence of Dissolved Air in Mobile Phase on Detection
min0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0
mAU
0
100
200
300
400BP
A
Degassed
Not degassed
* Analysis of bisphenol A by fluorescence detection
Degassing : Helium purging
Mobile Phase Degassing Influence in Fluorescence Detection (1)
• Mobile phase degassing may affect fluorescence intensity.
• Gas-liquid membrane separation degasser takes time to get degassing result.
0
1
2
3
4
5
0 10 20 30 40 50 60Time from start of degassing (min)
Pea
k A
rea
Val
ue
(×
106 )
Degasser OFF
Degasser ONColumn : Shim-Pack VP-ODS
(150mm L x 4.6mm I.D.)Mobile Phase: acetonitrile/H2O = 4/1 (v/v)Flow rate : 1mL/minTemperature : 40ºC
Analyte : Pyrene
Detection : Fluorescence (Ex 310nm, Em 390nm)
Degassing:Gas-liq. separation membrane
Mobile Phase Degassing Influence in Fluorescence Detection (2)
Methanol Absorption Spectrum
• If dissolved oxygen concentration becomes high, organic solvent absorbance becomes high.
• This effect is pronounced in short wavelength region.
Mobile Phase Background Changes in UV Detection
Air saturation
Degassed
• Peak appears with mobile phase solution injection!?
Mobile Phase Dissolved Air-derived Peaks in UV Detection
Detection: UV- 210nm
The difference in dissolved air quantities in mobile phase and sample solvent solutions can be expressed as peaks.
Degassed mobile phase
Mobile phase not degassed
Oxygen bubbling
Air saturation
He bubbling
Oxygen bubbling
Air saturation
He bubbling
min5 10 15 20 25 30
mRIU
0
5
10
15
min5 10 15 20 25 30
mRIU
- 0.4
- 0.2
0.0
Magnified
Mobile Phase Dissolved Air-derived Peaks with RID Detector
Column : SCR-101C
Mobile phase : Water
Flow rate : 0.5mL/min
Temperature : 85ºC
Detection : RID detector
Degasser : Gas-liq. Membrane Separ. (DGU-14A)
Sample : 0.5% mannitol• The difference in dissolved air quantities in
mobile phase and sample solvent solutions can be expressed as peaks.
• In general, spectrum shape is affected by ambient temperature.
Measurement wavelength
Spectrum at high temperature
Absorbance difference due to temperature change
Spectrum at low temperature
Abs
orba
nce
Detection Cell Temperature Influence in UV Detection (1)
• Comparison of peak area values without temperature control (34C)
0.9680.985p-toluic acid
0.9860.991Benzoic acid
0.9910.994Phenol
50ºC40ºC
Peak Area Ratios Compared with Areas at 34ºC
Column : Shim-Pack VP-ODS (150mm L. x 4.6mm I.D.)
Mobile phase : 10mM acetic acid (sodium) buffer solution (pH 4.5) / Methanol = 7/3, (v/v)
Flow rate : 1mL/min
Temperature : 40ºC
Detection : UV-VIS (260nm)
Detection Cell Temperature Influence in UV Detection (2)
With temperature control (RF-10AXLsuper)
Areas 20→25C13.5% decrease
Areas at 20→25C2.2% decrease
Detection Cell Temperature Influence in Fluorescence Detection (1)
• Normally, the higher the temperature, the lower the fluorescence intensity.
Analysis result of acridine (Temp. effect on fluorescence intensity)
Without temperature control (RF-10AXL)
0.779110171185514LEU
0.7711266831457340ILE
0.777900131026839MET
0.7812260421565440VAL
0.78138426177078CYS
0.7712035291558560ALA
0.7919249002440772GLY
0.86436522505373PRO
0.7710177681315462GLU
0.7712042731570743SER
0.7711439251478447THR
0.7811657571500427ASP
40ºC20ºC
Area RatioA40/A20
Peak Area Area RatioA40/A20
Peak Area
40ºC20ºC
0.768391831105530ARG
0.69336837487433LYS
0.7914316651819675HIS
0.769958321315479PHE
0.748509761156850TYR
Instrument : Shimadzu Amino Acid Analysis System (Post-column derivitization using o-phthaladehyde)
Detection : RF-10AXL Super (Ex 350nm, Em 450nm) (Cell temperature ; 20ºC, 40ºC)
Detection Cell Temperature Influence in Fluorescence Detection (2)
Fluorescence Spectrum of Bisphenol A in Each Solvent (Excitation wavelength 270nm, with background correction)
Detection Solvent Influence on Fluorescence Intensity
Methanol
Acetonitrile
Wavelength(nm)
750000
800000
850000
900000
1 2 3 4 5 6 7 8
No. of Analyses
Pea
k Are
a
冷却あり冷却なし
• Without cooling, ascorbic acid gradually decomposes.
Column :SCR-101N (250mmL. x 7.9mm I.D.)
Mobile :10mM oxalic acid (sodium) buffer
phase solution (pH 3.8) including 1mM EDTA•2Na
Flow rate : 1mL/min
Temp. : 40ºC
Detection : UV-VIS (245nm)
Change in Ascorbic Acid Peak Area
Sample Curbing Decomposition of Components by Cooling
CoolingNo cooling
Sample Influence of Sample Solvent on Peak Shape (1)
Liquid flow
Methanol solvent
Analyte
Liquid flow
Aqueous solvent
Analyte
*Reversed phase chromatography with water / methanol mobile phase mixture
• Theoretical plate number changes with different sample solvent.
Aqueous solvent
Methanol / H2O
Methanol solvent
Sample: 10ul caffeine
Column:
Shim-pack CLC-ODS
(150mmL. x 6mm I.D.)
Mobile phase:
Methanol / H2O =3/7
Flow rate: 1mL/min
Temp. : Room temp.
Detection: 270nm
Theoretical
Plate
Number
Injection volume
Left Fig. : Sample solvent - water / acetonitrile = 3/1 (v/v)Right Fig.: Sample solvent - acetonitrile
Column : STR ODS-II (150mm L. x 4.6mm I.D.)
Mobile : 20mM citric acid (sodium) buffer
Phase : solution (pH 4.5) / acetonitile = 3/1 ( v/v)
Flow rate : 1mL/min
Temp. : 40ºC
Detection : UV-VIS (240nm)
Sample Influence of Sample Solvent on Peak Shape (2)
If sample solvent elution is stronger than that of mobile phase, peak shape deteriorates.
• Improves analytical accuracy– Difficult to perform offline processing operations
completely uniformly, possibly causing poor accuracy
• Improves ease of operation– Offline processing may require special skills
• Improves processing efficiency– Automated operation can be performed during
night, greatly improving processing efficiency
Sample Advantages of Automated Sample Pretreatment
Sample Column Switching HPLC System
for Direct Injection of Blood Plasma (Co-Sense for BA)
Pump
Analytical column
Autosampler
Inner Surface Reversed Phase Pretreatment Column
Shim-Pack MAYI-ODS
Pump
Detector
Mobile phase
Mobile phase
● ●
Dilution bypass
Polymer like protein
Drug
Coating film
Image of drug and inner surface
0 8.5 17Time (min)
proteinprotein IsopropylantipyrineIsopropylantipyrine
Injection volume : 100μLMobile phase (Smpl. Inj. side) : 0.1% phosporic acid / acetonitrile = 95/5 (v/v)Detection : Isopropylantipyrine ; 275nm
Plasma matrix ; 280nm
Sample Automated Pretreatment Example using Co-Sense for BA
•Analysis of Plasma Spiked with Isopropylantipyrine
• Recovery rate comparison after each of 2mg/mL ketoprofen and naproxen are added to plasma (50mLinj.)
Manual Pretreatment(Acetonitrile Added)
Automatic Pretreatment Using Co-Sense for BA
Ketoprofen Naproxen Ketoprofen Naproxenn=1 91.1 92.0 96.0 96.7n=2 96.6 96.3 95.7 97.0n=3 93.9 97.0 97.2 96.9n=4 92.4 94.1 97.3 97.0n=5 90.2 91.7 95.6 98.2
Average 92.8 94.2 96.4 97.2S.D. 2.5 2.4 0.8 0.6
C.V.(%) 2.7 2.6 0.9 0.6
Sample Comparison of Recovery by Pretreatment Methods
• Cause: Adsorption to metallic materials due to ionic interaction or coordination isomerism interaction (ionic compounds or basic compounds) – Remove adsorbed components by needle washing– Control adsorption by making needle inert
• Cause: Adsorption by hydrophobic interaction with resinous materials (fat soluble compounds)– Wash rotor seal groove with organic solvent– Minimize adsorption by changing rotor seal material
SampleCauses and Control of Carryover
Causes and Means of Control in Autosampler
Washing this surface is important.
NeedleWashing stateWashing stateReady stateReady state
Sample loop
Pump
Column
Wash portNeedle
Needle seal
Sample Needle Wash Mechanism (Total Volume Injection)
• To control adsorption of substances due to ionic interaction or coordination isomerism (ionic compounds or basic compounds)…– Acidify.– Make ion pair with large radius counter ion.
• (Example) 100mM perchloric acid aqueous solution / methanol (or acetonitrile)
• Adsorption of compound due to hydrophobic interaction (resinous compounds)– Wash with organic solvent.
* (Example) 100% methanol, acetonitrile, THF, etc.
Sample Selection of Needle Wash Solution
• Analyte : Chlorohexidin (basic compound)
• Inject 2L of standard solution (mobile phase with 1.2mg/ml dissolved compound), continuously inject 2L of mobile phase, compare areas.
Column : Shim-pack VP-ODS (150mmL.x 2.0mm I.D.)Mobile phase : 10mM phosphoric acid buffer solution (pH 2.6) containing
100mMNaClO4 / acetonitrile = 55 / 45 (v/v)Flow rate : 0.2mL/minTemperature : 40ºCDetection : UV (260nm , using semi micro cell)
Sample Minimizing Contamination due to Basic Substances (1)
0.04250.0004251W/O Wash
0.07300.0007301W/ Wash
BlankStandard Solution
Carryover [%]
Relative Peak Area Needle material : SUSWash solution : Mobile phaseNeedle wash process :
1.Immerse in wash port 3 sec.2.Draw sample solution.3.Immerse in wash port 3 sec.4.Inject.
• Comparison using variously processed SUS needles to make them inert
0.00230.0000231Teflon coating
0.04250.0004251SUS
0.00090.0000091Special metallic coating
0.00210.0000211PEEK coating
BlankStandard Solution
Carryover [%]
Relative Peak Area
Immersed in wash solution 3 sec. Other conditions same as previous page.
Sample Minimizing Contamination due to Basic Substances (2)
Chromatogram with injection of 2µL mobile phase
Sample Minimizing Contamination Due to Basic Substances (3)
• In analysis of chlorohexidin (basic compound), carryover decreases from 0.07% 0.001% using needle immersion washing and a special metallic coating on the needle surface!!
(Varies with component type, concentration and injection volume)
Chromatogram with injection of 2µL standard solution
* Ordinate full scale differs in the chromatograms.
chlorohexidin
chlorohexidin
• Adsorption of hydrophobic compounds is thought to occur mainly in rotor seal groove through which samples pass.
• It is believed that because the adsorptive substance is washed with mobile phase, it elutes gradually.
Sample loop
Pump
Column
Needle
Rotor seal
Sample Minimizing Contamination of Hydrophobic Substances (1)
• Analyte : Vitamin A Acetate • Inject 10µL of standard solution ( 10μg/mL: methanol dissolution ) ,
then inject 10µL of blank solvent (methanol), and compare peak areas.
Column : Shim-Pack FC-ODS (75mmL.x 4.6mmI.D. )
Mobile phase : (A) ; Water , (B) ; Methanol 0-2min, B solution 75%, hold 2-5min, B solution 75-100%, linear gradient 5-10min, B solution 100%, hold
Flow rate : 0.5mL/minTemperature : 50ºCDetection : UV (325nm)
Sample Minimizing Contamination of Hydrophobic Substances (2)
Relative Peak Area
Standard Blank
10 minutes aging 1 0.000051 0.0051
20 minutes aging 1 0.000136 0.0136
Carryover[%]
• By changing to a rotor seal made of PEEK, even after 20 minutes of aging, almost no carryover was observed.
Sample Minimizing Contamination of Hydrophobic Substances (3)
min0 2 4 6 8 10 12 14
V
0.0
0.5
1.0
1.594
1217
8
Chromatogram with injection of 10µL of blankChromatogram with injection of 10µL standard solution
* Ordinate full scale differs in the chromatograms
Sample High-Throughput Autosampler SIL-HT
• Reduced carryover
• Optimized materials for needle, rotor seal, etc.
• Equipped with rinse mode
• Multi-sample processing
• 2 sample trays / 4 micro-titer plates
• Maximum sample number: 350 samples with 1.0mL vials
1536 samples with four 384-well micro-titer plates • High-speed sample injection
• 15 seconds required for sample injection (with 10 L injection)!
• Equipped with high-performance measurement device
• Total volume injection mode
• Ideal LC autosampler for MS front end