1000TR ORP% Instructions

1000TR % Eng · G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc - 3 - - Senast utskrivet 2006-10-27 11:41 2. INSTALLATION 2.1 CASING The casing is made of die-cast

  • Upload
    others

  • View
    6

  • Download
    0

Embed Size (px)

Citation preview

Page 1: 1000TR % Eng · G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc - 3 - - Senast utskrivet 2006-10-27 11:41 2. INSTALLATION 2.1 CASING The casing is made of die-cast

1000TR

ORP%

Instructions

Page 2: 1000TR % Eng · G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc - 3 - - Senast utskrivet 2006-10-27 11:41 2. INSTALLATION 2.1 CASING The casing is made of die-cast

G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc

- 1 -

- Senast utskrivet 2006-10-27 11:41

CONTENTS 1. INTRODUCTION ............................................................................................................................................. 2

1.1 COMMON INTRODUCTION...................................................................................................................... 2 1.2 PARTS & ACCESSORIES........................................................................................................................... 2

2. INSTALLATION .............................................................................................................................................. 3

2.1 CASING ........................................................................................................................................................ 3 2.2 MOUNTING ................................................................................................................................................. 3 2.3 ELECTRICAL INSTALLATION................................................................................................................. 3

2.3.1 Connection of the power supply ............................................................................................................. 4 2.3.2 Connection of the electrode.................................................................................................................... 4 2.3.3 Connection of a recorder........................................................................................................................ 4

2.4 CHECK THE CONNECTIONS.................................................................................................................... 4

3. FUNCTIONS ..................................................................................................................................................... 5

3.1 COMMON FUNCTIONS ............................................................................................................................. 5 3.2........................................................................................................................................................................ 5

3.2.1 MEASURE.............................................................................................................................................. 5 3.2.2 CAL. 1..................................................................................................................................................... 5 3.2.3 OUTPUT ................................................................................................................................................ 5 3.2.4 ALARM ................................................................................................................................................... 5

3.3 KEY FUNCTIONS ....................................................................................................................................... 6

4. PARAMETER SETTINGS .............................................................................................................................. 7

4.1 DEFAULT SETTINGS ................................................................................................................................. 7 4.2 PARAMETER SETTINGS ........................................................................................................................... 7

4.2.1 Set the output signal ............................................................................................................................... 7 4.3 CALIBRATION............................................................................................................................................ 7 4.4 OPERATION ................................................................................................................................................ 7 4.5 CHANGING THE PARAMETER SETTINGS ............................................................................................ 8

4.5.1 Changing the output signal..................................................................................................................... 8

5. MAINTENANCE .............................................................................................................................................. 9

5.1 HARDWARE CHECK ................................................................................................................................. 9 5.2 PRE-CALIBRATION ................................................................................................................................... 9

5.2.1 Pre-calibration ....................................................................................................................................... 9 5.3 RESET OF THE DEFAULT SETTINGS ................................................................................................... 10

6. TROUBLE SHOOTING................................................................................................................................. 11

7. SPECIFICATIONS ......................................................................................................................................... 11

APPENDIX

Calibration instruction

Page 3: 1000TR % Eng · G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc - 3 - - Senast utskrivet 2006-10-27 11:41 2. INSTALLATION 2.1 CASING The casing is made of die-cast

G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc

- 2 -

- Senast utskrivet 2006-10-27 11:41

1. INTRODUCTION

1.1 COMMON INTRODUCTION 1000TR is a new CE-approved series of transmitters from Elmacron AB. In the series, there are instruments for pH, Redox%, Redox mV and temperature measurement. 1000TR is easy to program, calibrate and use. The layout on the instrument is clear; a large display with four LED-segments, diode indication for the menu and four large function keys. All settings are made through the function keys on the front panel. When connecting an electrode to 1000TR-ORP% the display shows the actual value in the solution measured. 1000TR-ORP% is provided with an isolated output 0/4 - 20 mA, proportional to the measured value, for connection to, for example, a computer, printer or other recording equipment. When measuring is interrupted, the output stays frozen at the last measured value.

1.2 PARTS & ACCESSORIES 1000TR is delivered without connection cables and electrodes.

Article Function Code number Power supply cord, standard Power supply 20-R037-000 Low noise coax cable with BNC-connection ( xx* = 01, 02, 03, 05, 07 or 10 meter )

Connection of electrode/probe 60-T810-0xx*

Page 4: 1000TR % Eng · G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc - 3 - - Senast utskrivet 2006-10-27 11:41 2. INSTALLATION 2.1 CASING The casing is made of die-cast

G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc

- 3 -

- Senast utskrivet 2006-10-27 11:41

2. INSTALLATION

2.1 CASING The casing is made of die-cast polystyrol with a transparent front cover that is closed with a snaplock. The connection terminals are located in a separate partition at the lower part of the transmitter. At the bottom of the instrument there are four cable glands ( two 15,2 and two 18,6 ) for the electrical connections. The protection class is IP65.

2.2 MOUNTING The instrument is designed for wall mounting. Be sure that the instrument is mounted at a non-vibrating place. 1000TR is mounted vertically with two fastening screws (∅ 7) through the mounting brackets on the instrument.

2.3 ELECTRICAL INSTALLATION It is recommended that each instrument is provided with a separate power switch. The electrode coax cable must be protected by a screen and may not be installed near power cables. Avoid extension of cables.

The connection terminals are located in a separate space at the lower part of the transmitter.

Page 5: 1000TR % Eng · G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc - 3 - - Senast utskrivet 2006-10-27 11:41 2. INSTALLATION 2.1 CASING The casing is made of die-cast

G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc

- 4 -

- Senast utskrivet 2006-10-27 11:41

2.3.1 Connection of the power supply Connect the power supply to terminal 1 (earth), terminal 2 (zero) and terminal 3 (live).

2.3.2 Connection of the electrode Electrode cabling is a critical part of the whole system.

• Use a low noise coax cable between the electrode and the input terminal on the controller. • Low noise cables has, in general, a black semi-conductive sheath between the centrewire

insulation and the shield. Remove this sheath. • Keep the cable away from power wires. • Max recommended length of cable is 10 meters. • Connect the shield ( reference ) of the coax cable to terminal 4 ( EL - ). • Connect the centre ( metall ) of the coax cable to terminal 5 ( EL + ). NOTE! When using electrode PtPK/G: • connect the shield ( reference ) of the coax cable to terminal 5 ( ELECTR+ ) • connect the centre ( metal pin ) of the coax cable to terminal 4 ( ELECTR- )

2.3.3 Connection of a recorder Connect the recorder to terminal 10 ( REC + ) and terminal 11 ( REC - ).

2.4 CHECK THE CONNECTIONS Before the power supply is turned on; check that the connections are mechanical and electrical correct.

Page 6: 1000TR % Eng · G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc - 3 - - Senast utskrivet 2006-10-27 11:41 2. INSTALLATION 2.1 CASING The casing is made of die-cast

G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc

- 5 -

- Senast utskrivet 2006-10-27 11:41

3. FUNCTIONS

3.1 COMMON FUNCTIONS All settings are made through the function keys on the front panel. The keys + and – are accelerating in three steps when they are kept pressed. Chosen function is indicated by a green diode in the menu, at alarm a red diode is activated at the same time as the diode at the actual function activates. When pressing M you step down in the menu. When calibration or setting mode is entered the output signal stays "frozen" at the latest value, until = is pressed.

3.2

MEASURE

CAL. 1

OUTPUT

ALARM

Offset

0/4 - 20 mA

3.2.1 MEASURE When power supply is connected the measuring starts..

3.2.2 CAL. 1 Offset calibration. Pre-set value: 5%.

3.2.3 OUTPUT Can be set to 0 - 20 mA or 4 - 20 mA.

3.2.4 ALARM When an error occurs in the measuring mode, the alarm diode activates at the same time as the error code is shown on the display. At serious errors, the error code is flashing on the display.

Page 7: 1000TR % Eng · G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc - 3 - - Senast utskrivet 2006-10-27 11:41 2. INSTALLATION 2.1 CASING The casing is made of die-cast

G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc

- 6 -

- Senast utskrivet 2006-10-27 11:41

3.3 KEY FUNCTIONS

KEY FUNCTION 1 ( at parameter setting mode ) FUNCTION 2 ( in measuring mode )

Steps down the menu

Accelerating key. Increase the value in setting mode. Used when making a choice according to the display.

Interrupt measuring together with

Accelerating key. Decrease the value in setting mode.

Interrupt measuring together with

Confirm performed settings and choices.

Starts measuring.

Page 8: 1000TR % Eng · G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc - 3 - - Senast utskrivet 2006-10-27 11:41 2. INSTALLATION 2.1 CASING The casing is made of die-cast

G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc

- 7 -

- Senast utskrivet 2006-10-27 11:41

4. PARAMETER SETTINGS

4.1 DEFAULT SETTINGS When delivered, 1000TR have the following default settings. Output 0 - 20 mA Offset 5% Slope 100% ( 20 °)

4.2 PARAMETER SETTINGS All settings has to be confirmed with = to be saved. To leave the setting mode without saving the settings: press M. Before starting up the transmitter the following parameters has to be set: Temperature compensation and output range. Turn the power on.

4.2.1 Set the output signal 1. Interrupt the measuring by pressing + and – at the same time until the measuring is

interrupted and the diode at CAL.1 is activated. 2. Press M until the diode at OUTPUT. is activated. At the display: 0 - 20 . 3. Choose between 0 - 20 and 4 - 20 mA with +. 4. Press = to confirm. The diode at MEASURE is activated and the display shows Hold. 5. Press = to start the measurement.

4.3 CALIBRATION It is recommended that the instrument is calibrated regularly, to get the highest possible accuracy in the measuring. The instrument is preset for calibration at 5%. During the calibration, the latest output signal stays "frozen". See attachment "Calibration instructions" for calibration.

4.4 OPERATION Measuring is started when the power supply is turned on. After an interruption off the measuring it can be restarted by pressing = when the MEASURE diode is activated. The display shows the actual value in the solution measured at the same time as the corresponding mA-signal is sent to the output. During the operation, the diode at MEASURE is activated.

Page 9: 1000TR % Eng · G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc - 3 - - Senast utskrivet 2006-10-27 11:41 2. INSTALLATION 2.1 CASING The casing is made of die-cast

G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc

- 8 -

- Senast utskrivet 2006-10-27 11:41

4.5 CHANGING THE PARAMETER SETTINGS All settings has to be confirmed with = to be saved. To leave the setting mode without saving the settings: press M.

4.5.1 Changing the output signal 1. Interrupt the measuring by pressing + and – at the same time until the measuring is

interrupted and the diode at CAL.1 is activated. 2. Press M until the diode at OUTPUT. is activated. 3. Choose between 0 - 20 and 4 - 20 mA with +. 4. Press = to confirm. The diode at MEASURE is activated and the display shows Hold. 5. Press = to start the measurement.

Page 10: 1000TR % Eng · G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc - 3 - - Senast utskrivet 2006-10-27 11:41 2. INSTALLATION 2.1 CASING The casing is made of die-cast

G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc

- 9 -

- Senast utskrivet 2006-10-27 11:41

5. MAINTENANCE

5.1 HARDWARE CHECK A check of the hardware is necessary only if you suspect that there is something wrong with your controller.

1. Disconnect the power supply. 2. Short-circuit the electrode input. 3. Connect a mA-meter to the REC-output. 4. Press M and connect the power supply, keep M pressed during 5 seconds until the four

upper diodes are activated. 5. The transmitter performs a self-test according to table 5.1.1 5.1.1 Hardware check

CH01 Test of LED on display All 4 LED-segments are activated

CH02 Test of function keys Press the keys one by one and the corresponding sign is shown on the display

CH03 Test of output signal Press =. 20.00 is shown on the display at the same time as the corresponding signal is sent to the output. Press =. 04.00 is shown on the display at the same time as the corresponding signal is sent to the output. Press =.

CH04 test of diodes All the menu diodes are activated

The version number is shown, then the instrument starts to measure.

5.2 PRE-CALIBRATION The transmitter is always pre-calibrated when delivered. Pre-calibration is only necessary when the instrument has been shut of for a longer period. It is recommended that the pre-calibration is carried out by means of the calibrator 1000CAL. Connect the calibrator to the electrode input.

5.2.1 Pre-calibration 1. Turn the power off. 2. Connect the calibrator to the electrode input. 2. Press – and turn on the power supply. Keep the key pressed until the display shows

CAL1.CAL1.CAL1.CAL1. (10 seconds). Every second diode is activated. 3. Set the calibrator at 5%. Press = to continue. 4. On the display is E05.0E05.0E05.0E05.0 shown during 5 s. 5. On the display CAL2.CAL2.CAL2.CAL2.

Page 11: 1000TR % Eng · G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc - 3 - - Senast utskrivet 2006-10-27 11:41 2. INSTALLATION 2.1 CASING The casing is made of die-cast

G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc

- 10 -

- Senast utskrivet 2006-10-27 11:41

6. Set the calibrator at 85%. Press = to continue. 7. On the display is 085.0085.0085.0085.0 shown during 5 s. 8. The instrument returns to measuring mode, the measuring starts automatically.

5.3 RESET OF THE DEFAULT SETTINGS The settings are reset according to chapter 4.1.

1. Turn off the power supply. 2. Press + and turn on the power supply. Keep the key pressed until the display shows CCCC

llll rrrr OOOO (10 seconds). The four lower diodes are activated. 3. Press = to continue. 4. The display shows the version number and then the measuring starts up.

Page 12: 1000TR % Eng · G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc - 3 - - Senast utskrivet 2006-10-27 11:41 2. INSTALLATION 2.1 CASING The casing is made of die-cast

G:\Instruktioner\Instruktioner i Word\1000TR\Engelska\1000TR % Eng.doc

- 11 -

- Senast utskrivet 2006-10-27 11:41

6. TROUBLE SHOOTING CODE INDICATES PROBABLE CAUSE CORRECTION E-01 The deviation is > +30% from

the preset value. *

Faulty buffer solution Old/bad electrode

Check the buffer solution Replace the electrode

E-02 The deviation is < -30% from the preset value*

Faulty buffer solution Old/bad electrode

Check the buffer solution Replace the electrode

* The deviation is calculated in relation to the offsetvalue ( the zero point ). The accepted range for calibration at 5% is -25% to 35%.

7. SPECIFICATIONS Instrument Version Dimensions 186x131x103 Weight ca 1,0 kg Max length, electrode cable 10 meter Display Four 7-segment LED Keyboard 4 push buttons Connection Terminals Power supply 230 VAC, 50 Hz Backup > 10 years Range, % 0 - 100%

Accuracy, % ± 1 %

Resolution, % ± 1 %

Offset Range ± 30% Calibration Offset, 5%

Input ORP-inp z = 1012 Ω Outputs Output 0 - 20 mA / 4 - 20 mA Max load, output 270 Ω