48
1 Today’s Program Today’s Program Foam System Testing Foam System Testing And Calibration And Calibration

1 Todays Program Foam System Testing And Calibration

Embed Size (px)

Citation preview

Page 1: 1 Todays Program Foam System Testing And Calibration

1

Today’s ProgramToday’s Program

Foam System Testing Foam System Testing And CalibrationAnd Calibration

Page 2: 1 Todays Program Foam System Testing And Calibration

2

Your Presenters Are…Your Presenters Are…

• Jim Johannessen– Fire Equipment Services Group of Underwriters Laboratories

Inc. for the past 26 years– Member of the NFPA 1901 (Automotive Apparatus) and 1500

(Fire Department Occupational Safety and Health) Committees

• Bill Ballantyne– Vice President – FoamPro (16 years)– President Fire Apparatus Manufacturers Association (2004)– NFPA 1901 Committee Principal Member (FAMA)

– Chairman – Foam Technical Committee

Page 3: 1 Todays Program Foam System Testing And Calibration

3

Disclaimer…

• This is not an official NFPA presentation

• No committee member can speak for the NFPA

• Members can only give their opinion

Page 4: 1 Todays Program Foam System Testing And Calibration

4

NFPA StandardsNFPA Standards

• Compliance with standards is VoluntaryVoluntary

• They are MinimumMinimum standards

• They are ConsensusConsensus standards

• Committee work is voluntary – No PayNo Pay

Page 5: 1 Todays Program Foam System Testing And Calibration

5

NFPA Standards…NFPA Standards…

• Editions are reviewed and updated every 5 years

• Editions can be “short-cycled” to address significant developments or changes

• A TIA (Temporary Interim Amendment) can be implemented to address issues that may arise

• Over 250 standards in the NFPA system

Page 6: 1 Todays Program Foam System Testing And Calibration

6

NFPA Standards…NFPA Standards…

• Main body is mandatory…SHALLSHALL

• Annex material is recommended…SHOULDSHOULD

Page 7: 1 Todays Program Foam System Testing And Calibration

7

NFPA Apparatus CommitteeNFPA Apparatus Committee

• NFPA 1901 Committee is responsible for:– 1901:1901: FD Apparatus– 1906:1906: Wildland Apparatus– 1911:1911: In-Service Testing, Maintenance and

Replacement– 1912:1912: Refurbishment

Page 8: 1 Todays Program Foam System Testing And Calibration

8

NFPA Apparatus Committee…NFPA Apparatus Committee…

• Committee balance is mandatory

– 1/3 Fire ServiceFire Service (10)

– 1/3 ManufacturersManufacturers (10)

– 1/3 Experts, Specialists, TestingExperts, Specialists, Testing (10)

Page 9: 1 Todays Program Foam System Testing And Calibration

9

NFPA Apparatus Committee…NFPA Apparatus Committee…

• Almost all current members have been in the fire service

• Chairman is alwaysalways afire service officer

– Chief Don Frazeur, LAFDChief Don Frazeur, LAFD

Page 10: 1 Todays Program Foam System Testing And Calibration

10

One Last Word About The StandardOne Last Word About The Standard

• Everything in 1901 is there for a reason!

• Usually something undesirable has happened:– A firefighter has been injured or worse, killed

• Pay attention – they could save your life!

Page 11: 1 Todays Program Foam System Testing And Calibration

11

NFPA 1901 2009 EditionNFPA 1901 2009 Edition

• Now available. Order at:– 800-344-3555

– www.nfpa.org

• Applies to all orders signed after January 1, 2009

Page 12: 1 Todays Program Foam System Testing And Calibration

12

What WeWill Cover

NFPAAccuracy

Requirements

NFPAAcceptedTesting

Methods

NFPAManufacturer,

Installerand AHJTesting

Requirements

Testing AFoamProSystem

Page 13: 1 Todays Program Foam System Testing And Calibration

13

Proportioning RatesProportioning Rates

• Ratio of concentrate to water

• Expressed as a percentage– 1.0% = 1 unit of concentrate to 99 units of

water– 3.0% = 3 units of concentrate to 97 units of

water– 0.5% = 0.5 unit of concentrate to 99.5 units of

water

Page 14: 1 Todays Program Foam System Testing And Calibration

14

Common Proportioning RatesCommon Proportioning Rates

• Class A– 0.1% - 1.0%

• Class B– 1%, 3%, 6%– (1% x 3%), (3% x 6%), (3% x 3%)

(Hydrocarbons x Polar Solvent fuels)

• Emulsifiers / Gels– Varies by manufacturer / usage

Page 15: 1 Todays Program Foam System Testing And Calibration

15

• For proportioning rates equal or greater than 1%:– - 0% to + 30% or 1 percentage point, whichever is less

(20.10.3)• For proportioning rates less than 1%:

– - 0% to + 40% (20.10.2)

6%

30%30% 20%20% 10%10% 10%10% 20%20% 30%30%

3%

1%

- +40%40%

Lean Rich

Allowable Range:Allowable Range:

6%:6%: 6.0 – 7.0%6.0 – 7.0%

3%:3%: 3.0 - 3.9%3.0 - 3.9%

1%:1%: 1.0 – 1.3%1.0 – 1.3%

0.5%: 0.5 – 0.70%0.5%: 0.5 – 0.70%

0.2%: 0.2 – 0.28%0.2%: 0.2 – 0.28%

NFPA 1901 and 1906 Now Require…NFPA 1901 and 1906 Now Require…

Page 16: 1 Todays Program Foam System Testing And Calibration

16

Why Is Accuracy So Important?Why Is Accuracy So Important?

• Performance– Ability to make wet, fluid or dry foam (Class A

foam)– Ability to richen for heavy rain (Class B foam)– Ability to seal (Class B foam)

• Logistics

• Cost

• Environmental Concerns (Class B)

Page 17: 1 Todays Program Foam System Testing And Calibration

17

What WeWill Cover

NFPAAccuracy

Requirements

NFPAAcceptedTesting

Methods

NFPAManufacturer,

Installerand AHJTesting

Requirements

Testing AFoamProSystem

Page 18: 1 Todays Program Foam System Testing And Calibration

18

Manufacturer’s Responsibilities – Manufacturer’s Responsibilities – What to ExpectWhat to Expect

• Chapter Four of NFPA 1901 requires:– If the apparatus has a foam

proportioning system, the foam proportioning system manufacturer’s certification of accuracy (see 20.10.4.2) and the final installer’s certification the foam proportioning system meets this standard(see 20.11.2)

Page 19: 1 Todays Program Foam System Testing And Calibration

19

Labels and Instruction PlatesLabels and Instruction Plates

• An instruction plate shall be provided for the foam proportioning system that includes, at a minimum, a piping schematic of the system and basic operating instructions.

Page 20: 1 Todays Program Foam System Testing And Calibration

20

Labels and Instruction PlatesLabels and Instruction Plates

• Foam classification type

• Maximum and minimum proportioning rates (percent)

• Maximum and minimum waterflow [gpm (L/min)]

• Maximum and minimum operating pressures

• The statement “Use only concentrates that are compatible with this foam proportioning system. Refer to the foam proportioning system manufacturer’s operating manual.”

Page 21: 1 Todays Program Foam System Testing And Calibration

21

Foam Proportioning System Accuracy.

• The foam proportioning system shall be type tested and certified by the foam proportioning system manufacturer to be accurate throughout the foam proportioning system’s declared range of waterflow, water pressure, foam percentage (or foam proportioning system capacity), and concentrate viscosity.

Page 22: 1 Todays Program Foam System Testing And Calibration

22

The methods of testing a foam proportioning system for

calibration accuracy. • Substituting water for foam concentrate

• Measuring foam concentrate pump output directly

• Determining foam percentage by use of a refractometer

• Determining foam percentage by use of a conductivity meter

Page 23: 1 Todays Program Foam System Testing And Calibration

23

How Accurate Are These Test Methods?How Accurate Are These Test Methods?

• Volumetric is most accurate (and least expensive)

• Measuring equipment must be calibrated correctly

• Foam viscosities may effect your results if you are substituting water for foam

• Conditions may not lend themselves to using refractometer or conductivity

Page 24: 1 Todays Program Foam System Testing And Calibration

24

Substituting water for foam Substituting water for foam concentrateconcentrate

• The proportioning system is operated at the waterflow rates at which the system is to be tested.

• Using water as a foam substitute draw from a separate calibrated tank rather than the foam concentrate tank.

• The volume of water drawn from the calibrated tank divided by the volume of water pumped over the same time period multiplied by 100 represents the percentage of foam the proportioning system is producing.

Page 25: 1 Todays Program Foam System Testing And Calibration

25

Measuring foam concentrate pump output directly

• With the foam proportioning system operating at a given waterflow rate and either foam concentrate or water used as a substitute for foam concentrate, the output of the foam concentrate pump is measured by diverting that output into a calibrated container for direct measurement over a given period of time. An alternative is to measure the foam concentrate flow or water substitute with a calibrated meter.

Page 26: 1 Todays Program Foam System Testing And Calibration

26

Determining Foam Percentage by Use of a Refractometer.

• This test determines the accuracy of a foam system’s proportioning equipment by measuring the percentage concentration of a foam concentrate in the foam solution.

Page 27: 1 Todays Program Foam System Testing And Calibration

27

Determining Foam Percentage by Use of a Refractometer.

• The amount of foam concentrate in the solution is measured with a refractometer.

• A base calibration curve is first drawn using the samples of foam concentrate and water taken from the foam system.

• Samples of the actual foam solution produced by the system are then tested and plotted on the graph.

• The results must fall within the parameters set by NFPA

Page 28: 1 Todays Program Foam System Testing And Calibration

28

Determining Foam Determining Foam Percentage by Use of a Percentage by Use of a

Conductivity Meter.Conductivity Meter.• The conductivity test method is

based on changes in electrical conductivity as foam concentrate is added to water.

• The accuracy of this type of testing will deteriorate as the conductivity of the water and concentrate approach each other.

Page 29: 1 Todays Program Foam System Testing And Calibration

29

Measuring Foam Output Using The Measuring Foam Output Using The Volumetric MethodVolumetric Method

Page 30: 1 Todays Program Foam System Testing And Calibration

30

InstrumentationInstrumentation

• The instruments on the right of the ruler are conductivity meters and the instrument on the left is a refractometer

Page 31: 1 Todays Program Foam System Testing And Calibration

31

NFPA 1911NFPA 1911Standard for the Inspection, Standard for the Inspection,

Maintenance, Testing, and Retirement Maintenance, Testing, and Retirement of In-Service Automotive Fire of In-Service Automotive Fire

ApparatusApparatus

2007 Edition

Page 32: 1 Todays Program Foam System Testing And Calibration

32

Performance Testing Of Foam Performance Testing Of Foam Proportioning Systems (1911)Proportioning Systems (1911)

• 20.1 Test to determine if capable of delivering foam for the agents used

• 20.1.1 Tested at least annually• 20.1.2 Inspect all components• 20.2 Operated at % specified by AHJ at flows and

pressures for the agent(s) used• 20.2.1 Measure calibration accuracy• 20.2.2 Use same test points for every test• 20.3 Acceptable Testing Methods• 20.4 Test all concentrates carried• 20.5 Meet accuracy level in effect during install

Page 33: 1 Todays Program Foam System Testing And Calibration

33

NFPA 11 Inspection and Test FormsNFPA 11 Inspection and Test Forms

Page 34: 1 Todays Program Foam System Testing And Calibration

34

What WeWill Cover

NFPAAccuracy

Requirements

NFPAAcceptedTesting

Methods

NFPAManufacturer,

Installerand AHJTesting

Requirements

Testing AFoamProSystem

Page 35: 1 Todays Program Foam System Testing And Calibration

35

NFPA 1901 Purchaser RequirementsNFPA 1901 Purchaser Requirements

20.3.2 The purchaser shall specify the following:

(1) Range of waterflows and pressures

(2) Proportioning rates

(3) Types of concentrate(s) (Class A, Class B, etc)

(4) Brand and viscosity of concentrate

Page 36: 1 Todays Program Foam System Testing And Calibration

36

20.10 Foam Proportioning System 20.10 Foam Proportioning System Accuracy (1901 and 1906)Accuracy (1901 and 1906)

20.10.1 The foam proportioning system shall be type tested and certified by the foam proportioning system manufacturer to be accurate throughout the foam proportioning system’s declared range of water flow, water pressure, foam percentage (or foam proportioning system capacity), and concentrate viscosity.

Page 37: 1 Todays Program Foam System Testing And Calibration

37

20.10.1.1 Test Points for 20.10.1.1 Test Points for (Manufacturer’s)(Manufacturer’s) Certification of Foam Proportioning System Certification of Foam Proportioning System

Performance (1901 and 1906)Performance (1901 and 1906)

WaterflowWater

Pressure

Foam Percentage or Foam Proportioning

System Capacity

Minimum Minimum Minimum

Maximum Maximum Maximum

Minimum Maximum Minimum

Maximum Minimum Maximum

Midrange Midrange Midrange

Page 38: 1 Todays Program Foam System Testing And Calibration

38

3-D Description Of Terms3-D Description Of Terms

Water Flow

Wat

er P

ress

ure

Foam O

utput

Minimum / Minimum / Minimum

Maximum / Maximum / Maximum

Midrange

Page 39: 1 Todays Program Foam System Testing And Calibration

39

20.11 Testing and Documentation20.11 Testing and Documentation

20.11.1 The final installer shall test and certify the following:

(1) The foam proportioning system, as installed, complies with the foam manufacturer’s installation recommendations.

(2) The foam proportioning system has been calibrated and tested to meet the foam equipment manufacturer’s and the purchaser’s specifications.

(3) At a minimum, the foam proportioning system has been tested at the points defined in Table 20.11.1 for each foam system injection point.

Page 40: 1 Todays Program Foam System Testing And Calibration

40

20.11.1 Test Points for Installation Testing of 20.11.1 Test Points for Installation Testing of Foam Proportioning System PerformanceFoam Proportioning System Performance

(1901 and 1906)(1901 and 1906)

Waterflow Water PressureFoam Percentage or Foam Proportioning

System Capacity

Midrange Midrange Midrange

Minimum Minimum Minimum

Maximum Maximum Maximum

Page 41: 1 Todays Program Foam System Testing And Calibration

41

What WeWill Cover

NFPAAccuracy

Requirements

NFPAAcceptedTesting

Methods

NFPAManufacturer,

Installerand AHJTesting

Requirements

Testing AFoamProSystem

Page 42: 1 Todays Program Foam System Testing And Calibration

42

Performance Testing A FoamProPerformance Testing A FoamPro

• Most accurate test method is volumetric.– Measure water by either flow or volume

– Flow using a pitot gauge– Volume using a calibrated container

– Measure foam by volume or flow– Volume using a calibrated container– Flow using a calibrated flow meter

Page 43: 1 Todays Program Foam System Testing And Calibration

43

Tools RequiredTools Required

• Pitot gauge for measuring water flow

• Calibrated container for collecting foam (water)

• Pressure gauge with load valve to set “back pressure”

• Stop watch

Page 44: 1 Todays Program Foam System Testing And Calibration

44

Measuring Water Flow – Pitot GaugeMeasuring Water Flow – Pitot Gauge

Page 45: 1 Todays Program Foam System Testing And Calibration

45

Pressure Gauge / Load Valve AssemblyPressure Gauge / Load Valve AssemblyTo Simulate Back Pressure In Foam LineTo Simulate Back Pressure In Foam Line

1. Plumb to calibrate side of Cal / Inject Valve

2. Use Load Valve to adjust Back Pressure

Page 46: 1 Todays Program Foam System Testing And Calibration

46

Installer Testing A FoamPro Model 2001 Installer Testing A FoamPro Model 2001 And 2-1/2” FlowmeterAnd 2-1/2” Flowmeter

• Cross reference the proportioner model and the flowmeter size on the testing matrix chart

WaterFlow:

Use

Simulated

Flow Rate:

Use

Foam Percentage:

Water

Pressure

Set At:

Foam System Output

Should Be:

250 GPM 100 0.8% 400 PSI 0.8 Gallons

750 GPM 100 2.6% 150 PSI 2.6 Gallons

20 GPM 20 0.1% 150 PSI 0.02 Gallons

Page 47: 1 Todays Program Foam System Testing And Calibration

47

Questions / CommentsQuestions / Comments

Bill Ballantyne

[email protected]

651-766-6308

Download this presentation at:

www.foampro.com

Jim Johannessen

[email protected]

717-459-3391

Page 48: 1 Todays Program Foam System Testing And Calibration

48

And Finally, Remember To Always:And Finally, Remember To Always:

Buckle Up!

Use Foam!

Be Safe!

Thank You!