Technical Specification of Goniophotometer & Goniospectroradiometric System
for setting up of LED based Lighting products & Luminaries Test Laboratory for
NTH(ER), Salt Lake, Kolkata
Three (3) copies of Technical Bids are to be submitted
The luminous intensity distribution of a luminaire is determined by measuring its illuminance at
a given distance in all possible directions in space by Goniometer. Goniophotometer (mirror C
type) consists of luminaire holder, mirror and photometer. During measurement, the mirror reflects
the light emitted by the test object towards the photometer. The photometer head positioned at fixed
distances sufficiently away from the light source to be able to evaluate it as point source.
On the basis of this data - luminous flux, luminous efficacy, average luminance of the tested
luminaires can be calculated.
By equipping a Spectroradiometer with Goniometer, the spectral characteristics of the light
source at specific angles of emission can be evaluated. The color characteristics of the source, for
example the Color Rendering Indices (CRI), the co-related color temperature and the color
coordinates can be ascertained through mathematical calculations. Thus, a complete color distribution
& evaluation of the light source can be obtained, which is especially helpful for luminaire designers of
LED luminaires to produce lighting with uniform color appearance.
Sl No
Items Features/Parameters
Scope of work
1. The successful bidder has to set up the
Goniophotometer & Goniospectroradiometric System in
the room to be provided by NTH of approximate size 24m
(L) x 6.5m(W) x 6.2m(H) (lay out diagram in APPENDIX -
1). Necessary infrastructure for setting up the same has to
be carried out by the firm.
The bidder has to categorically state the work he
intends to carry out for the above job in Technical Bid,
failing which the offer will not be considered and no
correspondence will be made. However, break up cost
for the work to be carried out may be mentioned in the
Price Bid only.
2. The successful bidder has to provide appropriate
gadgets & devices for maintaining the required
environmental conditions of Temperature, Humidity etc. as
per the requirement of specification.
The bidder has to categorically state the work he
intends to carry out for the above job like Air
Conditioning, De-humidifier etc. in Technical Bid,
failing which the offer will not be considered and no
correspondence will be made. However, break up cost
for the work to be carried out may be mentioned in the
Price Bid only.
3. For setting up the entire system NTH will provide 3 phase
supply of 415V±5% at 50±2.5 Hz & single phase supply of
240V±10 V at 50±2.5 Hz. Bidder will have to provide
necessary gadgets to cope up with voltage & frequency
Page 2 of 7
Sl No
Items Features/Parameters
fluctuation for running the system, if required.
The bidder has to categorically state the work he
intends to carry out for the above job like providing
Constant Voltage Constant Frequency (CVCF) Source
etc., if required in Technical Bid, failing which the offer
will not be considered and no correspondence will be
made. However, break up cost for the work to be
carried out may be mentioned in the Price Bid only.
4. Short-circuit protection for the equipment is to be provided.
The bidder has to categorically state the protective
device that has been employed for the above.
1. Goniophotometer & Goniospectroradiometric System A Goniophotometer system and
Working Principle
C-type mirror Goniometric system fully automated with
computerized system & software.
The Goniometer shall be configured for operation with both
Photometer & Spectro-radiometer.
Mode selection should be fully automatic.
B Room size NTH will provide a room of approximate size
24m (L) x 6.5m(W) x 6.2m(H). The drawing of room may be
referred in APPENDIX - 1
Bidder to quote the best fit model and indicate it's model no. along
with proposed Laboratory utilisation plan.
The total room volume required for maintaining ambient
conditions such as temperature, humidity, air speed, air
pressure etc. as laid down in relevant specifications covering
NTH requirements are to be indicated.
C Goniophotometer in
accordance with
-EN 13032
-LM 75-01
-CIE Publication No. 70 (1987)
D Standards to be fulfilled by
complete system (limited to
sections applicable to
Goniophotometric and
Goniospectroradiometric
procedures only)
Indian Standard
IS 16106 : 2012 for Solid State Lighting Products
IS 16107 (Part-5/Sec-1): 2012 for LED Luminaire.
IS 16103 (Part-2) : 2012 for LED Modules for general lighting
IS 16102 (Part-2) : 2012 for Self-ballasted LED lamps
IS 10322 (Part-5/Sec-1) : 2012 for Fixed General purpose
luminaire
IS 10322 (Part-5/Sec-2) : 2012 for Recessed luminaire
IS 10322 ( Part-5/Sec-3) : 2012 for Road & Street light luminaire
IS 10322 (Part-5/Sec-4) for portable luminaire
IS 10322 (Part-5/Sec-5) for flood light luminaire
IS 13383 (Part-1) : 1992 : Method of photometric measurement of
luminaires for use in interior lighting
IS 13383 (Part-2) : 1992 : Method of photometric measurement of
luminaires for road & street lighting
Page 3 of 7
Sl No
Items Features/Parameters
I IS 13383 (Part-3) : 1992 : Method of photometric measurement of
luminaires for flood lighting
CIE Technical Reports and Standards
CIE technical specifications covering
1) Test Method for LED Lamps, LED Luminaires, and LED
Modules
2) Photometry of Floodlights
3) Methods of Characterizing Illuminance Meters and
Luminance Meters
4) The Measurement of Absolute Luminous Intensity
Distributions
5) The Measurement of Luminous Flux
6) The Photometry and Goniophotometry of Luminaires
7) Measurement of LEDs
8) Method of Measuring and Specifying Colour Rendering
Properties of Light Sources
European standard
European standard covering
Light and Lighting Measurement and presentation of photometric
data of lamps and luminaires for LED Lamps, Modules, and
Luminaires
1.1 Goniometer
A Method of alignment of Test
Object
Motorized height adjustment of test object. Built-in alignment laser
to be provided for verification.
B Power/sense connection
between test object, power
supply, power meter
Sufficient number of Electrical connection junctions for test object
are to be provided, with all cables integrated into goniometer
framework.
The number of junctions are to be specified with utility & drawing
layout .
C Angular encoder Optical type
D Size of test object ≥1600mm in diagonal length
E Weight of test object ≥50 Kg. Appropriate counter weights are to be provided which is
to be quoted.
F Photometric test distances
Instrument shall be capable of testing at multiple test distances
covering low intensity as well as high intensity sources.
G Method of switching from one
test distance to another.
Switching shall be automatic and software controlled only.
H Stray light correction Method of integrated stray-light correction compulsory
I Test object mounting jigs Universal jig to be provided for mounting ceiling, downlights, panel
light, street light, flood light, post top light etc.
J Remote control Hand held remote system to be provided for control of goniometer
assembly.
K Measuring lampholders 4 contact, gold plated, Kelvin type lamp-holders for line and
sense, contacts at lamp base for lamp types B22, G5, G13 with
ballast, E40, E 26/27, E12, GU10, GX5.3. for easy connection to
goniometer.
L Safety Features Software operated Limit Switches or equivalent devices are to be
provided to prevent all types of rotation for mirrors, accessories
etc. where there are specified limits of rotation.
Page 4 of 7
Sl No
Items Features/Parameters
Mirror
N Mirror base and mirror Uniformity, reflectance and flatness deviation: ≤1.5%
With manufacturer's certificate to be provided in support of above.
O Polarization effect of mirror ≤1% as per LM 79-08 sec. 9.3.5
Test/Calibration report from a laboratory accredited as per ISO
17025 or equivalent, to be provided in support of above.
P Mirror Sample A 10 cm x 10 cm sample of the mirror to be employed shall be
delivered with the system for future reference for re-calibration of
photometer head & verification of mirror characteristics.
1.2 Photometer for employment with mirror Goniometer for Goniophtometry
A Photometer Type V(λ) corrected Silicon photo detector
B Area of application Photometric measurements of test object in conjunction with the
Goniometer for evaluation of the photometric characteristics of the
object, angularly resolved.
C Lux measuring range 0.0001 lx – 80000 lx or better
Test/Calibration report from a laboratory accredited as per ISO
17025 or equivalent, to be provided in support of above.
D Spectral irradiance
responsivity
f1' ≤1.0% including spectral influence of the mirror.
Test/Calibration report from a laboratory accredited as per ISO
17025 or equivalent, to be provided in support of above.
E f' spectral mismatch
correction factor
Shall be less than 1.0 %. Indicate f' for phosphor & RGB type
white LEDs including spectral influence of the mirror,
Test/Calibration report from a laboratory accredited as per ISO
17025 or equivalent, to be provided in support of above.
F Thermostatic stabilisation Built-in to head
G Signal amplification Built-in to head
H Control unit Photometer display and control unit, manual control with digital
display in LUX and CANDELA for each test distance, built-in
interface for software control.
I Calibration method Calibration against a Photometric standard lamp via the
mirror.
All calibration hardware and software are to be included in
the scope of supply
1.3 Spectroradiometer for employment with mirror Goniometer for
Goniospectroradiometry
A Detector CCD or upgraded version. The bidder is to specify the detector
and justify the upgraded version, if any.
B Area of application Spectral measurements of test object in conjunction with the
goniometer for evaluation of the colorimetric characteristics of the
object, angularly resolved.
C Wavelength range 380nm-780nm
D Spectral resolution ≤3 nm
E Wavelength Accuracy 0.15 nm or better
F Colour Accuracy ≤0.002 in x, y
Test/Calibration report from a laboratory accredited as per ISO
17025 or equivalent, to be provided in support of above.
G Field of view Adjustable to mirror distance and diameter
Page 5 of 7
Sl No
Items Features/Parameters
H Calibration method Calibration against a spectral standard lamp via the mirror.
All calibration hardware and software are to be included in
the scope of supply
I Software functions for
Colorimetrics with
Goniospectroradiometer
Software for computer controlled operation of spectroradiometer
in conjunction with the goniometer for computer based operation,
measurement, system calibration, and automatic test report
generation for particular measurement of the spectral values and
subsequent calculation of the colorimetric characteristics
1.4 Power Sources and Meters for Powering of Test Object
A AC power source Voltage range 0-300V, Current Range:0-16A AC at 50 Hz with
accuracy ≤ 0.2% for voltage & current.
Separate Source to be provided for test object of less than 30 W.
Input Current regulation: The input current shall be monitored
and regulated to within ± 0.5% of the rated rms value during
photometric measurements as per IS:16105-2012.
B DC power source 0-120V, 0-10A(min), with accuracy of ≤ 0.2%,the ripple voltage
shall not exceed 2% of the DC output voltage as per IS:16105-
2012.
C Power Analyzer Power Analyzer with Harmonics package with voltage/current
accuracy: ≤ ± 0.2%
Test/Calibration report from a laboratory accredited as per ISO
17025 or equivalent, to be provided in support of above.
D Software Software for computerised operations & data acquisition of power
source & wattmeter as well as all applicable electrical parameters
doing test along-with photometric/colorimetric data.
1.5 PC and Software
A Computer System Standard PC, Monitor, Printer inclusive of OS Windows 7 or latest
for operating the system (sense and feedback of the object under
test, performance of Goniophotometric & Goniospectroradiometric
testing, evaluation of test results, report generation etc.)
B Test routines Commonly employed test routines for general lighting applications
are pre-programmed
C Custom routines Software shall enable custom generation of test procedure by
user
D Report generation The software shall enable automatic test report generation,
documentation and visualization. Export of test data in EXCEL
E Colorimetric parameters
required to be calculated
from Goniospectroradiometric
measurements
-Chromaticity coordinates
-Correlated color temperature
-Color rendering indices R1-R16 & Ra of R1 to R8
-Δ u’ v’ as required by Energy Star requirement
-Weighted CRI
-Optional Calculation of CQS
-Flexible selection of n-step (n=1 to 7) MacAdam ellipse portrayal
around average/target color point
-3-D interactive portrayal of the measured colorimetric parameters
-Plot of spectral power distribution
-Plot of chromaticity diagrams along with Planckian Locus
-Built-in verification procedure for spectroradiometer: Calibration
Page 6 of 7
Sl No
Items Features/Parameters
against luminous intensity, traceable to the photometer of the
system expressed in W/(sr nm)
-Scotopic/Photopic ratio
F Software reporting examples Bidders shall attach sample software reports demonstrating
Compliance
1.6 System Calibration A Standard lamps for
Goniophotometer
4 (FOUR) Standard lamps with certificate traceable to NMI/ISO
17025 for at least 2 (TWO) lamps.
Provision of Halogen incandescent lamp, FEL type, for calibration
of the Goniophotometer via the mirror with certificate issued for
lamp luminous flux, voltage, current, issued by a National
Metrology Institute / ISO 17025 accredited laboratory
Provision of suitable lampholder for connection to goniophotometer enabling electrical connection of line and sense leads to power source/wattmeter in system rack for system calibration purposes.
B Standard lamps for
Goniospectroradiometer
4 (FOUR) Standard lamps with certificate traceable to NMI/ISO
17025 for at least 2 (TWO) lamps.
Provision of Halogen incandescent lamp, FEL type, for calibration
of the Goniospectroradiometer via the mirror with certificate
issued for lamp CCT, CRI, x,y; voltage, current, issued by a
National Metrology Institute / ISO 17025 accredited laboratory
Provision of suitable lampholder for connection to Goniospectroradiometer enabling electrical connection of line and sense leads to power source/wattmeter in system rack for system calibration purposes
C Calibration of the relative
spectral sensitivity of the
photometer head
- Calibration of the relative spectral irradiance responsivity of the
photometer head(s) and mirror assembly
- Verification of the quality index of the photometer head and
mirror assembly for the V(λ) correction according to EN 13032-1
- Verification of the spectral mismatch indices of the photometer
head and mirror assembly.
Certificate to be issued in accordance with NMI/ISO 17025
D Verification of system
alignment
Verification of alignment of optical path is to be demonstrated by
means of a distance laser with calibration certificate traceable to
NMI/ISO 17025 along-with Manufacturer’s Certificate
1.7 Infrastructural Requirements
A Setting up of Laboratory
Infrastructure
NTH will provide the room for Laboratory purpose (lay out
diagram in APPENDIX-1) with wooden partition & painted matt
finish black. For housing the Goniometer assembly, the bidding
firm has to prepare necessary structures, as deemed fit at their
own cost and for which break up price is to be provided in the
price bid only, which may be in Indian Rupees.
The bidder has to categorically state the work he intends to
carry out for the above job.
B Creation of Environmental
Conditions
For maintaining the ambient conditions in the Laboratory like
Air Conditioning , Air Pressure, Relative Humidity etc. the
Page 7 of 7
Sl No
Items Features/Parameters
bidding firm has to arrange for that, as deemed fit at their own
cost and for which break up price is to be provided in the price
bid only, which may be in Indian Rupees.
The bidder has to categorically state the work he intends to
carry out for the above job.
The total volume of the room is to be indicated where
controlled ambient conditions of temperature, humidity, air
speed, air pressure etc. will be maintained.
1.8 Training Requirement
A Training at Manufacturer site 1. Training of at least two scientists / four scientists at the
manufacturing works are to be provided. Number of days of
training for two scientists / four scientists are to be indicated
with break up cost in the price bid which include air fare,
boarding, lodging etc.
2. Training duration for two / four scientists at the manufacturing
works are to be mentioned only in the technical bid.
B Training at place of
installation
At least four scientists from NTH(ER), Kolkata should be trained
free of cost for all types of instrument related works viz.
operation, servicing, maintenance etc. for at least 10 working
days at the place of installation.
1.9 Other terms & conditions for Compliance by the Bidder :
(i) Two copies of Operation & Service Manual with electrical / electronics circuit diagrams should be
provided along with the Machine.
(ii) Original Printed Literature of the Machine quoted along with accessories etc. should be included
in the offer along with all application notes.
(iii) Accessories are to be quoted with break up cost.
(iv) The Instrument is to be installed, commissioned and demonstrated satisfactorily at NTH(ER),
Kolkata,at free of cost.
(v) Warranty 60 (sixty) months from the date of satisfactory installation & commissioning of
the equipment.
Guarantee that they will supply spare parts, as and when required for a period of at least five (5)
years after expiry of warranty period of SIXTY (60) months
(vi) Instrument down time should be attended within 48 hours and the same is to be confirmed during
submission of the bid.
(vii) Terms & conditions for servicing the machine (both CMC and AMC) after expiry of warranty period
of 60 (sixty) months is to be quoted.
(viii) The manufacturer/supplier should submit the name of institute / organization preferably from
Govt. Sector, with “Performance Certificate” where machine/ instrument of similar type &
capacity has been supplied
LMT LICH T ME SSTECHNIK GMBH BERLIN · HELMHOLTZSTRASSE 2-9 · 10587 BERLIN GER MANYP HON E : +49-30-393 40 28 · FA X: +49-30-391 80 01 · E-MAIL: [email protected] · WWW.L MT.DE
Goniophotometerwith Mirror ArrangementLMT GO-DS
GO-DS DAT B404 Data subject to change without notice
Goniophotometer LMT GO-DS Technical Data
Marking LMT GO-DS Goniophotometer
Field of application Measurement and evaluation of light technical characteristics of all types of lamps and luminaries in their dedicated burning position with respect to gravity
Versions and test object dimensions
· GO-DS 1600 for test objects up to 1600 mm length (diagonal) · GO-DS 2000 for test objects up to 2000 mm length (diagonal) · Max. test object load: 50 kg
System conformity
· EN-13032-1 · CIE 121, 1996 · CIE 70, 1987 · IES LM 79-08 · Related internationally recognized standards
Tests performed
· Luminous intensity distribution in C-planes or on conical surfaces · Absolute and zonal luminous flux · Floodlight testing according to B-β system, conversion to C-γ · CIE color values, with optional colorimetric system
Goniometer type · Goniometer with centrically rotating mirror, mirror on optical axis for static and narrow light cone
Test distances · Fixed test distances to be placed at selectable positions between 5 m and 50 m · Numerous test distances possible
Straylight screening · Fixed baffles with adjustable apertures according to test distance and size of object
Mirror · Spectrally constant reflecting, non-polarizing mirror mounted in tempered aluminum base for
rigidity, flatness, and long time stability · Max. deviation in uniformity in accordance with EN 13032-1: < 1.5%
C-axis
· Test-object mounting on vertically oriented C-axis · Uninterrupted test-object rotation by means of 8-pole slip ring contact · Adjustable speed control · Motor controlled re-orientation of test object in pendant or upright position
γ-axis · Drive shaft with counterweight and forged cogged wheel for vibration free movement and reproducible positioning under load
Angular measurement · Optical angular encoders fixed to goniometer axes with guaranteed accuracy for encoder lifetime,
resolution 0.1° · Correlation of C-axis with γ-axis under rated loads at all positions: < 0.2°
Photometric system
· Characteristics in fulfillment of EN 13032-1 Table 3 · Total characteristics ≤ 3% (according to EN 13032-1, Table 3, Footnote C) · V(λ) adaptation f1 ≤ 1% by means of LMT Mosaic Filtering® with compensation for the spectral
characteristics of the mirror · Max. spectral mismatch correction factor fz against typical LED spectra for general lighting
(white LEDs 2.500 – 7.000 K): < 1% · Measurement range 0.0001 lx – 80000 lx
Colorimetric system (option)
· LMT array type spectrometric system in dedicated design for measurement of CIE color values in conjunction with mirror type goniometer
· Automatic positioning in and out of optical axis · Software routines for goniocolorimetrics according to LM 79-08 Chap 12
Electrical measurement · System integration and control of power supplies and wattmeters for supply of test objects and corresponding measurement values
Software · Complete software for system control, creation and execution of test programs, data administration,
evaluation, and export to light planning softwares · Presentation of data in graphic and text form · File import and export in IES, Eulumdat and related light-technical file formats
Building requirements · Narrow light cone of centrically rotating mirror enables building design with com pact, temperature controlled goniometer cabin and separate light tunnel for sensors