Dr.
A
nto
nio
s L
on
tos
Slide 1/25
Dr. Antonios Lontos
Associate Professor
The project ΥΓΕΙΑ/ΔΥΓΕΙΑ/0311/27 (ΒΙΕ) was co-financed by the European Regional Development Fund and the
Republic of Cyprus through the Research Promotion Foundation.
Scientific Session: Track SS8
Optimizing the diagnostic value of Myocardial Perfusion Imaging
using a dynamic phantom assembly
Construction of inflatable lungs to simulate
respiratory motion in Myocardial Perfusion Imaging (MPI)
XIV MEDITERRANEAN CONFERENCE ON MEDICAL AND BIOLOGICAL ENGINEERING AND COMPUTING
MARCH 31ST - APRIL 2ND 2016, PAPHOS, CYPRUS
Dr.
A
nto
nio
s L
on
tos
Slide 2/25
1. Introduction
2. Image processing and 3D design
3. Design and construction of the moulds
4. Construction of inflatable lungs
5. Conclusions
Contents
Dr.
A
nto
nio
s L
on
tos
Slide 3/25 1. Introduction
• One of the most common heart disease is the Coronary Artery Disease (CAD)
• Coronary artery stenosis can be evaluated by Myocardial Perfusion Imaging (MPI)
• Myocardial Perfusion Images (MPI) is influenced not only by the cardiac motion but also by various other
parameters such as the respiratory motion
Computational models and patient data are useful.
But mechanically controlled phantoms (with inflatable lungs and heart beating) are very
important to optimize MPI studies.
Research scope
• Accurate design and Construction of inflatable lungs with lung-equivalent material.
• Validation of the inflatable lungs
• Development of a respiratory-diaphragm motion phantom, and implement it within the existing cardiac
phantom
• Help in validation of new protocols before implementation using phantom.
• Educational purposes to physicians, medical physicists, technicians and students
Dr.
A
nto
nio
s L
on
tos
Slide 4/25 2. Image processing and 3D design
Anthropomorphic phantom for testing
Thorax-CT flash-anatomy 86 planes
It is a fully tissue equivalent and represents an average male in size and shape. Both lungs and heart will be operating by special equipment for the actual simulation of the respiratory motion and heart beating
Using special software, lungs can be design. From each image, the outside contour can be collected. By attaching the one contour to other in specific height the real 3D design can be constructed. As it is shown the geometrical shape of the lungs is very complicated and difficult to construct
Dr.
A
nto
nio
s L
on
tos
Slide 5/25 2. Image processing and 3D design
Left and right lung shape The shape of both lungs was slightly changed in order to predict smoother geometry The total volume of the lung was the same as measured from CT scan images The left lung volume is 510 ml and the right lung volume is 565 ml
• The inside volume filled by tissue equivalent material • Upper openings help respiratory simulation and filling of the lung with the tissue
equivalent material • Transparent acrylic material was attached at the top of each lung • Elastic tubes were connected with the acrylic material and air pumps • Using these tubes and controlled air pump, respiratory motion can be simulated
Dr.
A
nto
nio
s L
on
tos
Slide 6/25
Left lung assembly Right lung assembly
2. Image processing and 3D design
Silicon membrane thickness = 2mm Main consideration: Anatomy, strength and elasticity of the silicon
Air tight fittings for air tubes
Dr.
A
nto
nio
s L
on
tos
Slide 7/25 3. Design and construction of the molds
Design of the different parts of the left lung mold
Design and construction considerations
(1) Better injection of the silicon rubber, (2) Better solidify of the material, (3) Silicon lung removal,
(4) Dimensioning and injection parameters, (5) Assemble and disassemble of the different mold parts
Dr.
A
nto
nio
s L
on
tos
Slide 8/25
Construction of aluminum moulds for heart and lungs using CNC machines
4-axis Computer Numerical Control machine
3. Design and construction of the molds
Dr.
A
nto
nio
s L
on
tos
Slide 9/25
Preparation of CNC manufacturing using special CAM software
3. Design and construction of the molds
Dr.
A
nto
nio
s L
on
tos
Slide 10/25
Construction of left lung mold
• The assembly and disassembly of lung molds is very important • Fit perfectly each other in order to leave smooth and uniform shape • Assembly and disassembly procedure was described and followed
3. Design and construction of the molds
• Material aluminum 6061 T6, which is very good material for injection molding • Easy to manufacture and can predict very good and shiny surfaces • Shiny surfaces are very important for the silicon lungs quality
Dr.
A
nto
nio
s L
on
tos
Slide 11/25
Molds for the heart - inner part
3. Design and construction of the molds
Previous Research work: Heart design
Dr.
A
nto
nio
s L
on
tos
Slide 12/25
Moulds for the heart – outer part
3. Design and construction of the molds
Previous Research work: Heart design
Dr.
A
nto
nio
s L
on
tos
Slide 13/25
TYPICAL GENERAL CHARACTERISTICS
ELASTOSIL M 4645 A/B Silastic 3481
Color Transparent White
Mixed viscosity, mPa.s 50,000 25,000
Elongation at break, % 330 560
Hardness, (Shore A) 40 21
Tensile strength, MPa 5 4.9
Tear strength, N/mm 28 26
Base and Curing Agent mixture (by weight)
10:1 100:5
Curing time 12 hours 7 days
Typical properties of the silicon materials
Two different silicon rubbers were used: 1. Elastosil M4645 A/B, RTV-2 2. Silastic 3481, Hardener Silastic 81
• Tissue equivalent materials • Transparent for visibility during the experimental set up and procedure • Excellent mold release, fast and non-shrink cure at room temperature
4. Construction of inflatable lungs
Dr.
A
nto
nio
s L
on
tos
Slide 14/25
Construction of left lung mold and equipment
Special equipment: • Weight scale: High accuracy weight scale was use during the mixing of silicon and hardener. • Vacuum pump: Vacuum pump was used for removing air bubbles after mixing and stirring of the two
components. • Injection molding equipment: A special cylinder was used for the injection of silicon from the bottom
opening of the mold.
4. Construction of inflatable lungs
The mold cavity was filled from the bottom to the top in order to release air and avoid trapping air bubbles.
Dr.
A
nto
nio
s L
on
tos
Slide 15/25
Silicon left and right lungs
Left lung
Right lung from white silicon rubber
Right lung
White silicon rubber White and transparent silicon rubber
4. Construction of inflatable lungs
Dr.
A
nto
nio
s L
on
tos
Slide 16/25
Silicon lungs
• Both materials have the same size and shape with slight different in properties and color. • Special thermoplastic boundaries for the limitation of the lungs expansion during inhalation (air
pumping)
Thermoplastic boundaries
4. Construction of inflatable lungs
Dr.
A
nto
nio
s L
on
tos
Slide 17/25
Silicon heart final construction, transparent silicon
Inner part Outer part
Inner part Outer part
4. Construction of inflatable lungs and heart
Previous Research work: Heart design
Dr.
A
nto
nio
s L
on
tos
Slide 18/25
Inner part testing
Outer part testing
NO pressure With pressure
NO pressure With pressure
4. Construction of inflatable lungs and heart
Previous Research work: Heart design
Dr.
A
nto
nio
s L
on
tos
Slide 19/25
Silicon heart final assembly and testing
NO pressure With pressure
4. Construction of inflatable lungs and heart
Previous Research work: Heart design
Dr.
A
nto
nio
s L
on
tos
Slide 20/25
• A tissue equivalent material was inserted in the lungs cavity
• This material has similar characteristic mass and porosity as lung tissue
• The inserted material help in simulating the real movement of the lungs during respiratory
• Supports were placed on the top and the bottom of lungs
• All the materials that used in the phantom can be characterized as tissue equivalent
• Two compact size pumps with suction and blowing ports were used as vacuum pumps and small compressor (maximum flow of a 17,5 lt/min, pressure of 1 bar)
• Two electro-proportional valves control the respiratory motion (air pressure with respect to time)
• A special water tank was use to validate lungs volume expansion
o Expiration 510 ml and 565 ml, Inhalation up to 700 ml and 775 ml (Test for every 50 ml, tidal volumes 350-600 ml ), 2% Deviation
o Respiratory cycles 10-14 breaths per minute (± 20 ms)
Construction of the respiratory motion simulator for the phantom
4. Construction of inflatable lungs
Dr.
A
nto
nio
s L
on
tos
Slide 21/25
Cage construction for better simulation of the respiratory motion
Thermoplastic cage Thermoplastic cage with silicon lungs
4. Construction of inflatable lungs
Dr.
A
nto
nio
s L
on
tos
Slide 22/25 4. Construction of inflatable lungs
Dr.
A
nto
nio
s L
on
tos
Slide 23/25 4. Construction of inflatable lungs
Dr.
A
nto
nio
s L
on
tos
Slide 24/25 5. Conclusions
• Human-sized inflatable lungs were constructed to simulate the respiratory motion in MPI
studies.
• CT slices were used to design the lungs.
• Left and Right lungs with initial volumes 510 and 565 mL.
• Each lung was covered with a 1-mm-thick thermoplastic structure to properly inflate in
all directions in a physiologically correct manner.
• The tidal volume and the breathing cycle of the inflatable lungs were mechanically
validated.
• The tidal volume of the lungs was also validated from CT acquisitions using the OsiriX
imaging software.
• Mechanically controlled phantoms are useful to investigate Myocardial Perfusion
Imaging.
CONCLUSIONS
Dr.
A
nto
nio
s L
on
tos
Slide 25/25
The project ΥΓΕΙΑ/ΔΥΓΕΙΑ/0311/27 (ΒΙΕ) is co-financed by the European Regional Development Fund and the Republic of Cyprus through the Research Promotion Foundation.
The author would like to thank: - Yiannis Parpottas, Ph.D. in Nuclear Physics - Isabelle Chrysanthou, Ph.D. in Medical Physics - Antonis Antoniou, Ph.D. in Mechanical Engineering - Panayi Panayiotis (CNC manufacturing , Moulds Construction) “Machinery Panagiotis J. Panagi”
Thank you for your attention
Recommended