Md100 digital spo2 user's manual

Preview:

Citation preview

Advanced DSP algorithm diminishes the influence of motion artifact All digital-signal processing improves accuracy of low perfusion Small size, easy to be assembled

Meditech.OEM

[MD100 DIGITAL SPO2 ]

Meditech SpO2 Module MD200 is a Low-power oxygen monitor module; the SPO2 Module is used in the sampling inspection or

measurement of the patient's oxygen saturation and pulse rate. As to the average oxygen

saturation, low-power module can send instantaneous blood oxygen saturation.

It is very useful, especially in the sleep research. The SPO2 module is compatible with the entire

Page 2 of 4

Range and accuracyPulse rate: 30 ~ 250 bpmSpO2: 0 ~ 100%Weak perfusion 0.1%Resolution(1SD)Pulse rate:(30~250BPM)

2bpm Normal3bpm Motion or low perfusion

SpO2: (70~100%)

2% Normal3% Motion or low perfusionBelow 70% Not specified

Environmental Requirements

Operating temperature: 0~45 °CStorage temperature: -20~60 °C

Operating Humidity:30~95%Storage Humidity:10~100%

Operating Altitude:-500~5000mStorage Altitude:-500~10000m

Power Specifications

Par. Min. Typ. Max.

5V +4.75V +5V +5.25V

Ripple --- --- 100mVp-p

Current --- 20mA ---

Communication ProtocolMD200 Module communicates with the host via a asynchronous serial port at TTL levels, the dataformat is 4800baud, 8 data bits, odd parity, 1stop bit(4800,8-Odd-1).

Page 3 of 4

Byte Bit Description Byte Bit Description

0

0 Signal strength 0

3

0 Pulse rate 0

1 Signal strength 1 1 Pulse rate 1

2 Signal strength 2 2 Pulse rate 2

3 Signal strength 3 3 Pulse rate 3

4 1 = Searching too long 4 Pulse rate 4

5 1 = Sensor unplugged 5 Pulse rate 5

6 1 = Pulse beep 6 Pulse rate 6

7 1 (Syn. bit) 7 0 (Syn. bit)

1

0 Plethysmogram 0

4

0 SpO2 0

1 Plethysmogram 1 1 SpO2 1

2 Plethysmogram 2 2 SpO2 2

3 Plethysmogram 3 3 SpO2 3

4 Plethysmogram 4 4 SpO2 4

5 Plethysmogram 5 5 SpO2 5

6 Plethysmogram 6 6 SpO2 6

7 0 (Syn. bit) 7 0 (Syn. bit)

2

0 Bargraph 0

1 Bargraph 1

2 Bargraph 2

3 Bargraph 3

4 1 = Sensor error

5 1 = Searching for pulse

6 Pulse rate 7

7 0 (Syn. bit)

Page 4 of 4

Connections

P1-Probe connectorP1 Pin Description

1 Probe output2 3.3V3 RED+4 RED-5 GND6 PRO-DET

P2-Host connectorP3 Pin Description

1 TX Output2 NC3 GND4 +5V

Size

Unit:mm

SensorOnly digital sensor can be used for this module.

Recommended