Advanced DSP algorithm diminishes the influence of motion artifact All digital-signal processing improves accuracy of low perfusion Small size, easy to be assembled
Meditech.OEM
[MD100 DIGITAL SPO2 ]
Meditech SpO2 Module MD200 is a Low-power oxygen monitor module; the SPO2 Module is used in the sampling inspection or
measurement of the patient's oxygen saturation and pulse rate. As to the average oxygen
saturation, low-power module can send instantaneous blood oxygen saturation.
It is very useful, especially in the sleep research. The SPO2 module is compatible with the entire
Page 2 of 4
Range and accuracyPulse rate: 30 ~ 250 bpmSpO2: 0 ~ 100%Weak perfusion 0.1%Resolution(1SD)Pulse rate:(30~250BPM)
2bpm Normal3bpm Motion or low perfusion
SpO2: (70~100%)
2% Normal3% Motion or low perfusionBelow 70% Not specified
Environmental Requirements
Operating temperature: 0~45 °CStorage temperature: -20~60 °C
Operating Humidity:30~95%Storage Humidity:10~100%
Operating Altitude:-500~5000mStorage Altitude:-500~10000m
Power Specifications
Par. Min. Typ. Max.
5V +4.75V +5V +5.25V
Ripple --- --- 100mVp-p
Current --- 20mA ---
Communication ProtocolMD200 Module communicates with the host via a asynchronous serial port at TTL levels, the dataformat is 4800baud, 8 data bits, odd parity, 1stop bit(4800,8-Odd-1).
Page 3 of 4
Byte Bit Description Byte Bit Description
0
0 Signal strength 0
3
0 Pulse rate 0
1 Signal strength 1 1 Pulse rate 1
2 Signal strength 2 2 Pulse rate 2
3 Signal strength 3 3 Pulse rate 3
4 1 = Searching too long 4 Pulse rate 4
5 1 = Sensor unplugged 5 Pulse rate 5
6 1 = Pulse beep 6 Pulse rate 6
7 1 (Syn. bit) 7 0 (Syn. bit)
1
0 Plethysmogram 0
4
0 SpO2 0
1 Plethysmogram 1 1 SpO2 1
2 Plethysmogram 2 2 SpO2 2
3 Plethysmogram 3 3 SpO2 3
4 Plethysmogram 4 4 SpO2 4
5 Plethysmogram 5 5 SpO2 5
6 Plethysmogram 6 6 SpO2 6
7 0 (Syn. bit) 7 0 (Syn. bit)
2
0 Bargraph 0
1 Bargraph 1
2 Bargraph 2
3 Bargraph 3
4 1 = Sensor error
5 1 = Searching for pulse
6 Pulse rate 7
7 0 (Syn. bit)
Page 4 of 4
Connections
P1-Probe connectorP1 Pin Description
1 Probe output2 3.3V3 RED+4 RED-5 GND6 PRO-DET
P2-Host connectorP3 Pin Description
1 TX Output2 NC3 GND4 +5V
Size
Unit:mm
SensorOnly digital sensor can be used for this module.