1 / 14 Increased Throughput and Purity of Combinatorial Libraries Utilizing a Targeted Gradient Profile Based on Preliminary Analytical Screening Todd M. Anderson Shimadzu Scientific Instruments, Inc., Columbia, MD

HPLC Combinatorial Libraries

Embed Size (px)

DESCRIPTION

With increasing pressure of a higher sample throughput and fewer chemists, purification labs in medicinal chemistry groups need to be more productive now than ever before. This presentation will describe a technique that allows the analyst to obtain a higher purity and better resolution using information from the preliminary analytical screening of these samples prior to purification.

Citation preview

Page 1: HPLC Combinatorial Libraries

1 / 14

Increased Throughput and Purity of Combinatorial Libraries Utilizing a Targeted Gradient Profile Based on Preliminary Analytical ScreeningTodd M. AndersonShimadzu Scientific Instruments, Inc., Columbia, MD

Page 2: HPLC Combinatorial Libraries

2 / 142 / 14

Challenge

With increasing pressure of a higher sample throughput and fewer chemists, purification labs in medicinal chemistry groups need to be more productive now than ever before.

Many of these labs have historically utilized a steep low-to-high organic gradient, as the broad spectrum of compounds that separate tend to have a wide range of elution profiles. Typically, a 5 to 95% organic gradient profile is utilized.

Separating tightly resolved compounds and impurities can be somewhat difficult with these typical gradient profiles. The analyst is then forced to sacrifice either speed or purity, and ultimately requires multiple purifications.

Page 3: HPLC Combinatorial Libraries

3 / 143 / 14

Objective

This presentation will describe a technique that allows the analyst to obtain a higher purity and better resolution using information from the preliminary analytical screening of these samples prior to purification.

By utilizing the retention time of the compound from preliminary runs, an optimal set of conditions may be obtained that allows for a greater success rate of separation. Compared to the standard 5 to 95% elution gradient, a narrow, short gradient profile can be determined.

Along with a better separation, the chromatographic time can be shortened, saving the analyst both instrument time and solvent consumption, while allowing for purification with a single injection.

Page 4: HPLC Combinatorial Libraries

4 / 144 / 14

Methodology

HPLC: Shimadzu Prominence HPLCInjector SIL-20AC autosampler

Pumping System

2 X LC-20AD gradient pumps

Oven CTO-20A

Detector SPD-M20A PDA detector

Mobile PhaseA: H2O with 0.1% TFAB: Acetonitrile (LC grade)

Software LabSolutions V. 5.54

Column Shimadzu ODS (C-18) 5 um

Page 5: HPLC Combinatorial Libraries

5 / 145 / 14

Analytical Sample Data

A random selection of samples was obtained from the lab’s inventory of test compounds. Below is the compound list and their stock concentrations:

1) Saccharin 13.0 mg/ml

2) Caffeine 6.8 mg/ml

3) Papaverine 5.36 mg/ml

4) Verapamil 8.95 mg/ml

5) Butylparaben 14.31 mg/ml

6) Naphthalene 5.78 mg/ml

7) Anthracene 8.0 mg/ml

Individual retentions for these compounds were obtained by: 1) Running an initial screening of these compounds using a 5 to 95% ACN gradient paired with an aqueous phase of 0.1% TFA at 1.5 ml/min, and 2) A one-minute hold at 95% organic and a two-minute re-equilibration back at 5% ACN.

Page 6: HPLC Combinatorial Libraries

6 / 146 / 14

Chromatographic Profile

Datafile Name:7 compound mix_7 min gradient_1652 PM_002.lcdSample Name:7 compound mix

0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 min

0

100

200

300

400

500

600

700

800

900

mV

0.00

0.25

0.50

0.75

1.00

1.25

psi(x10,000)B.ConcAD2

Caffe

ine

Papp

aver

ine

Vera

pami

l

Butyl

Par

aben

Anthr

acen

e imp

urity

Napth

alene

Anthr

acen

e

0.0 1.0 2.0 3.0 4.0 5.0 6.0 min

200

225

250

275

300

325

350

375

400

nm

To obtain a single chromatogram with all components, a mixture of 100 ul of each solution was combined and run on the same gradient profile (Figure 1). Figure 2 shows the PDA contour plot.

Figure 1.

Figure 2.

Page 7: HPLC Combinatorial Libraries

7 / 147 / 14

Gradient Profile Optimization

After obtaining retention times for each of the individual compounds, two compounds were chosen, one with a late elution and one with an early elution, to plot the optimal organic percent needed to elute those compounds at the height of the gradient profile.

This should achieve the best resolution and purity for preparative conditions.

Page 8: HPLC Combinatorial Libraries

8 / 148 / 14

The highlighted compounds, Anthracene and Caffeine, were used for this purpose (Figure 3).

Gradient Profile Optimization (continued)

Compound RT Opt. Org. % Final Peak RT

Saccharin 0.90 5 4.1

Caffeine 1.40 10 4.27

Papaverine 2.20 30 4.1

Verapamil 2.78 40 4.37

Butylparaben 3.10 45 4.33

Napthalene 3.60 55 4.28

Anthracene 4.20 67 4.27

Arbitrary RT Calculated Org. %

1 2.36

2 22.71

3 43.07

4 63.43

4.5 73.61

Values used to calculate slope of optimal elution percent.

Figure 3.

Page 9: HPLC Combinatorial Libraries

9 / 149 / 14

Gradient Profile Optimization (continued)

1 1.5 2 2.5 3 3.5 4 4.5 50

10

20

30

40

50

60

70

80

observed values

Calculated retention

A slope was then calculated, and all the compounds were run based on their predicted optimal elution gradient. Figure 4a compares the calculated and observed values with little deviation, and Figure 4b shows an overlay of the optimized chromatograms using LabSolutions software.

Figure 4a.

Page 10: HPLC Combinatorial Libraries

10 / 1410 / 14

0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 min

0

25000

50000

75000

100000

125000

150000

175000

200000

uV

Data7:Anthracene_47 to 67_1613 PM_001.lcd AD2Data6:Napthalene_35 to 55_1523 PM_003.lcd AD2Data5:Butyl Paraben_25 to 45_1523 PM_002.lcd AD2Data4:Verapamil_20 to 40_1709 PM_005.lcd AD2Data3:papaverine_10 to 30_1709 PM_003.lcd AD2Data2:caffeine_2 to 10_1728 PM_004.lcd AD2Data1:Sacharin_1 to 5_1709 PM_002.lcd AD2

Based on the chromatograms, it was determined that a mixture of Butylparaben, Naphthalene, and Anthracene (with its impurity peak – see below) would make the best example to test purification conditions.

Optimized Preparative Chromatography

Figure 4b.

Anthracene impurity

Data 1: Saccharin_1 to 5_1709 PM_002.Icd AD2Data 2: Caffeine_2 to 10_1728 PM_004.Icd AD2Data 3: Papaverine_10 to 30_1709 PM_003.Icd AD2Data 4: Verapamil_20 to 40_1709 PM_005.Icd AD2Data 5: Butylparaben_25 to 45_1523 PM_002.Icd AD2Data 6: Napthalene_35 to 55_1523 PM_003.Icd AD2 Data 7: Antharacene_47 to 67_1613 PM_001.Icd AD2

Page 11: HPLC Combinatorial Libraries

11 / 1411 / 14

Optimized Preparative Chromatography

A sample from the original stock solutions was created and run at three gradient profiles: one below (Figure 5), one at (Figure 6), and one above optimal conditions for Naphthalene (Figure 7).

Figure 5. A 25 to 45% organic gradient profile, optimized for Butylparaben.

Datafile Name:BP Nap Anth_25 to 45_1236 PM_003.lcdSample Name:BP Nap AnthSample ID:0 to 9

0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 min

0

50

100

150

200

250

300

350

400

450mV

0

10

20

30

40

50

60

70

80

90

%B.ConcAD2

Buty

lpar

aben

Nap

thal

ene

Anth

race

ne

Page 12: HPLC Combinatorial Libraries

12 / 1412 / 14

Datafile Name:BP Nap Anth_35 to 55 prep_1236 PM_004.lcdSample Name:BP Nap AnthSample ID:0 to 10

0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 min

0

50

100

150

200

250

300

350

400

450mV

0

10

20

30

40

50

60

70

80

90

%B.ConcAD2

Buty

lpar

aben

Nap

htha

lene

Anth

race

ne

Datafile Name:BP Nap Anth_45 to 65_1236 PM_006.lcdSample Name:BP Nap AnthSample ID:0 to 12

0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 min-50

0

50

100

150

200

250

300

350

400

450mV

0

10

20

30

40

50

60

70

80

90

%B.ConcAD2

Bu

tylp

ara

be

n

Na

ph

tha

len

e

An

thra

cen

e

Figure 6. A 35 to 55 % gradient profile optimized for Naphthalene.

Figure 7. A 45 to 65 % gradient profile optimized for Anthracene.

Optimized Preparative Chromatography (continued)

Page 13: HPLC Combinatorial Libraries

13 / 1413 / 14

Conclusion

After completion of all analysis, preparative separation was greatly enhanced by the shallow gradient determination.

With the results of the analytical screening, a table could be generated to indicate optimal gradient ranges for given windows of analytical retention time. This premise would need to be optimized for differences in analytical to preparative column performance, as well as dwell time differences from instrument to instrument. However, it would allow one analytical instrument to provide data for an entire lab running multiple preparative instruments, drastically increasing throughput and purity, while creating a more efficient workflow in the high-throughput purification lab.

Page 14: HPLC Combinatorial Libraries

14 / 1414 / 14

Thank you for viewing this presentation. Should you have any questions or require additional information about our research, products or services, please visit our Support page: www.ssi.shimadzu.com/support/