IEC 60601-1-11 is a mandatory standard provision geared toward determining and documenting failure and hazard potential of medical electrical equipment intended for home use. By addressing the challenges associated with regulating the use of medical devices in the home, we've identified key performance requirements that must be considered when designing a medical device for use outside of the traditional clinical environment.
Citation preview
Home HealthcareIEC 60601-1-11
October 24, 2013
Michael Brousseau Safety ManagerMichael Brousseau, Safety Manager
www.intertek.com1
Part 1Background
www.intertek.com2
Background: Size & Scope
F t F tFast Facts12 Million Americans Receive Home Care
75 Million Americans by 203075 Million Americans by 2030$74.3 Billion Annual Expenditure 2011 (CMS)
3% Annual Health Care Expenditure 201175% Receive Skilled Nursing Care in part by a caretaker or self
Majority are >65 years44 Million are caregivers of someone >1844 Million are caregivers of someone >18
www.intertek.com3
Background: The Drivers
Overall SpendingOverall Spending
2011: $2.7 Trillion2011: $2.7 Trillion
(projected growth)
2020: $4.6 Trillion2020: $4.6 Trillion
www.intertek.com4
Background: The Drivers
Per Capita Expenditure Per PersonPer Capita Expenditure Per Person
2011: $8,6802011: $8,680
(projected growth)
2020: $13 7082020: $13 7082020: $13,7082020: $13,708
The US spends $2,800+ more per person than the second highest spender among developed countries
www.intertek.com5
Background: The Drivers
U.S. Spends More Money on Health Care Per Capita as a %GDP than Any Country in the World
Obesity 400M
Chronic Disease Non -
Aging 600M
860M Non
Compliance 300B
U.S. Healthcare System Ranks 37th out of 191 Countries
www.intertek.com6
Background: The Drivers
Chronic DiseaseAging Population Hospital R d i i$2.025 Trillion 2011
$3 out of every $4 health d ll
Aging Population65 & Older
Will increase from
Readmissions$26 B in 2011
($17 B Preventable)care dollars
1 out of every 2 Americans
Will increase from40 million in 2010
to88 million in 2050
Nearly 1 in 5 patients readmitted within 30 days
Medical Technology Remote Monitoring & Telemedicine
Seeking cost-effective delivery of care
www.intertek.com7
Background: FDA Definition
Device intended for users in a non-clinical or transitory environmentManaged partl or holl b the serManaged partly or wholly by the user
May require training for the user by a healthcare professional in order to be used safely and effectively.
I l d tl d t il i l t d d iIncludes permanently and temporarily implanted devices Any type of equipment that a person may use to recover and rehabilitate
Must be capable of safely accommodating the many needs of the person using it.
“Home Use Extends Beyond the Home”Home Use Extends Beyond the HomeEncompass all environments in which a person plans to
use his or her medical device in day-to-day life.Delivers Medical Care Anywhere and Everywhere
www.intertek.com8
Delivers Medical Care Anywhere and Everywhere
Background: US vs. EU
M j iMajor variances exist betweenUnited StatesUnited States and Europe
www.intertek.com9
Background: US vs. EU
S• U.S. Nursing Homes are considered professional environments and products are classified as Class 1USA
• Europe Nursing Homes are considered a EU p ghome environment and products are classified as Class 2
EU
Class One Rely on a Ground, Class Two Do not rely on a ground
www.intertek.com10
y g
Part 2Challenge & Opportunity
www.intertek.com11
From 1997 through 2009, FDA received over 19,000 reports of adverse events in which the reporter listed the
location where the incident took place as “home.”
www.intertek.com12
Hospital vs. Home
The Home environment is fundamentally different from the clinical environment. These unique challenges have the potential to impact patient safety:
• Caretaker Knowledge
• Device Usability
• Environmental Predictability
www.intertek.com13
Challenges: Environmental Issues
ChildrenLocation
Power Sources
SanitationEMI
Temperature & Humidity
Pets & Vermin
Public Emergencies Noise
Air Quality
Water
www.intertek.com14
Challenges: Environmental Issues
Presence of Ho sehold PetsCase Example: A patient was put on peritoneal dialysis because of her failing kidneys. She lived in a one bedroom apartment with
Presence of Household Pets:
g y pher cat. Although the patient kept the cat out of her bedroom while she was undergoing dialysis, there was cat fur and dander throughout the room. Cat fur got into the patient’s dialysis tubing and entered her peritoneum and she contracted peritonitisand entered her peritoneum, and she contracted peritonitis.
www.intertek.com15
Challenges: Environmental Issues
Electromagnetic InterferenceCase Example: An individual with an implantable cardioverterdefibrillator (ICD) was playing electric guitar at home. As he was
Electromagnetic Interference:
( ) p y g gunplugging the guitar from the amplifier, feedback from the amplifier caused the individual to feel a painful vibration at the site of his ICD. His left arm became immobile for almost two minutes. The painful sensation stopped when his wife unplugged theThe painful sensation stopped when his wife unplugged the amplifier.
www.intertek.com16
Challenges: Environmental Issues
Noise Le els
Case Example: A pediatric patient was receiving ventilator therapy in the home, and the ventilator tubing became
Noise Levels:
py , gdisconnected. The patient’s mother was asleep and did not hear the ventilator alarm. There were other ambient noises in the home (e.g., television), and they may have interfered with the mother’s ability to hear the alarm The patient diedability to hear the alarm. The patient died.
www.intertek.com17
Challenges: User Issues
Off-Label
LifestyleDurability
Use
Physical Capabilities
Education
Ease of Use Availability
Education Level
of Instructions Emotional
Stability
www.intertek.com18
Challenges: User Issues
Ph sical CapabilitiesCase Example: A patient doing home nocturnal dialysis was found by her husband, unresponsive. She was disconnected from
Physical Capabilities:
y , pthe dialysis machine and her tubing was in a closed circuit filled with saline. There was a syringe attached to one of her lines, but the other line was open, and the patient had lost a significant amount of blood It appeared that she had been unable to clampamount of blood. It appeared that she had been unable to clamp her catheter by herself. The patient died due to blood loss.
www.intertek.com19
Part 3The Standard
www.intertek.com20
The Standard: National Deviations
“Medical Electrical Equipment – Part 1: General Requirements forMedical Electrical Equipment Part 1: General Requirements for Basic Safety and Essential Performance – Collateral Standard:
Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment”Electrical Systems Used in the Home Healthcare Environment
ANSI/AAMI HA60601-1-11: 2011 – US Deviations
CAN/CSA-C22.2 No. 60601-1-11: 2011
CENELEC EN 60601-1-11: 2010
IEC 60601-1-11: 2010
www.intertek.com21
Clause 4.1 – Mains Supply for ME Equipment and ME Systems
Non-life-supporting Medical Electrical Equipment or Medical Electrical Systems
- Testing completed at 110% and 85% of rating voltage or voltage range
Life-supporting Medical Electrical Equipment or Medical Electrical Systems
- Testing completed at 110% and 80% of rating voltage or voltage rangeg p g g g g
www.intertek.com22
Clause 4.2.1 – Environmental Conditions of Transport and Storage between Uses
Unless otherwise indicated in the IFU or if the Medical Electrical Equipment is t ti th i t h ll l t th f ll i t t d tstationary, the equipment shall comply to the following transport and storage
conditions
-25°C without relative humidity control; and
70°C at a relative humidity up to 93% non-condensing
If the IFU states a more restricted range j stification is pro ided as part of theIf the IFU states a more restricted range, justification is provided as part of the Usability Engineering File which appears in the Risk Management File; and
The range is marked on the product and the product’s carrying case
www.intertek.com23
Clause 4.2.1 Test Conditions
a) Prepare equipment for transport or storage per the IFU
b) Expose equipment at lowest environmental conditions for 24 hrs or till thermal stability is reached (minimum 2 hrs)thermal stability is reached (minimum 2 hrs)
c) Expose equipment at highest environmental conditions for 24 hrs or till thermal stability is reached (minimum 2 hrs)
d) Allo eq ipment to stabili e back to operating conditions of normal sed) Allow equipment to stabilize back to operating conditions of normal use
e) Evaluate equipment against its specifications, Basic Safety, and Essential Performance characteristics
www.intertek.com24
Clause 4.2.1 – Environmental Operating Conditions
Unless otherwise indicated in the IFU or if the Medical Electrical Equipment is t ti th i t h ll l t th f ll i t t d tstationary, the equipment shall comply to the following transport and storage
conditions
A t t f 5°C t 40°CA temperature range of 5°C to 40°C;
A relative humidity range of 15% to 93%, non-condensing; and
An atmospheric pressure range of 700 hPa to 1060 hPap p g
If the IFU states a more restricted range, justification is provided as part of the Usability Engineering File which appears in the Risk Management File; andy g g pp g
The range is marked on the product and the product’s carrying case
www.intertek.com25
Clause 4.2.2 – Test Conditions
a) Expose equipment to ambient conditions for 6 hrs or till thermal stability (minimum 2 hrs)
b) Evaluate equipment against its specifications, Basic Safety, and Essentialb) Evaluate equipment against its specifications, Basic Safety, and Essential Performance characteristics
c) Evaluate equipment against its specifications, Basic Safety, and Essential Performance characteristics at lowest specified atmospheric pressurep p p
d) Evaluate equipment against its specifications, Basic Safety, and Essential Performance characteristics at highest specified atmospheric pressure
www.intertek.com26
Clause 4.2.2 – Test Conditions, cont’d
e) Cool equipment to lowest specified environmental operating conditions for 6 hrs or till thermal stability (minimum 2 hrs)
f) Evaluate equipment against its specifications, Basic Safety, and Essentialf) Evaluate equipment against its specifications, Basic Safety, and Essential Performance characteristics
g) Warm equipment to highest specified environmental operating conditions for 6 hrs or thermal stability (minimum 2 hrs)y ( )
h) Evaluate equipment against its specifications, Basic Safety, and Essential Performance characteristics
www.intertek.com27
Clause 4.2.3 – Environmental Shock to Transit-Operable ME Equipment
a) Set up equipment for operation according to intended use
b) Expose equipment to lowest specified environmental operating conditions for 6 hrs or thermal stability (2 hrs)
c) Within 5 minutes, expose equipment to highest specified operatingc) Within 5 minutes, expose equipment to highest specified operating conditions
d) While maintaining step c) evaluate equipment against its specifications, Basic Safety, and Essential Performance characteristics until thermal y,stability or for 2 hrs
Repeat test new sample allowed by reversing steps b) and c)Repeat test, new sample allowed, by reversing steps b) and c)
www.intertek.com28
Clause 5 – General Requirements for Testing ME Equipment
In addition to the requirements of clause 5.9.2.1 of the general standard, accessibility also determined by:
Small Finger Probe – in a bent or straight position
a) For all positions of equipment when operated in normal use) p q p p
b) After opening access covers and removal of parts
i) Opened without use of tool
ii) IFU states Lay Operator is to open access cover
www.intertek.com29
Clause 6 – Classification of ME Equipment and ME Systems
Unless equipment in the Home Healthcare Environment is to be permanently installed, the equipment shall be:
Class II or Internally Powered;
Shall not have a Functional Earth Terminal; and
Applied Parts, shall be Type BF or Type CF
www.intertek.com30
Clause 7.1 – Usability of Accompanying Documents
Usability of the identification, marking, and accompanying documents intended f L O t L R ibl O i ti h ll b l t d i tfor Lay Operator or Lay Responsible Organization shall be evaluated against the Operator Profile assuming a maximum of eight years education.
Equipment in the Home Healthcare Environment should be designed to be simple to use and not require reference to complex Accompanying Documents.
Equipment in the Home Healthcare Environment should be designed to be q p gsimple to use and to minimize, reference to Accompanying Documents.
Verified by inspection of the Usability Engineering Process/FileVerified by inspection of the Usability Engineering Process/File
www.intertek.com31
Clause 7.2 – Additional Requirements for Marking IP Classification
For the following types of products, an IP rating is required:
a) Body-worn b) Hand-held
If a carrying case provides the protection against ingress of liquids, then it shall be marked but
c) Mobile d) Portable
be marked, but…
www.intertek.com32
Clause 7.2 – Additional Requirements for Marking IP Classification
If the enclosure does not provide the minimum protection against ingress of t it h ll b k dwater, it shall be marked:
“KEEP DRY”KEEP DRY
Or
www.intertek.com33
Clause 7.3.1 – Contact Information
The IFU shall include the following:
) f i t if d d i tti i i t i i th d ta) for assistance, if needed, in setting up, using, or maintaining the product
b) to report unexpected operation or events
A postal address and either a phone number or web address to contact the Manufacturer or Manufacturer’s Representative
www.intertek.com34
Clause 7.3.2 – Lay Operator Briefing Information
IFU shall include information for the Healthcare Professional to brief Lay Operator or Lay Responsible Organization on any known contraindications to the equipment and precautions to be taken including the following:be taken including the following:
• Changes in performance of equipment
• Exposure to reasonably foreseeable environmental conditions
• Information on medicinal substances or human blood derivatives
• Measuring function accuracies
www.intertek.com35
Clause 7.4.1 – Additional Requirements for Warning and Safety Notices
For each warning and safety sign, the IFU shall:
D ib th t f th h d- Describe the nature of the hazard;
- Consequences that could occur from not following advice; and
- Precautions for reducing riskg
www.intertek.com36
Clause 7.4.1 – Additional Requirements for Warning and Safety Notices
The IFU shall, at least, address the following:
St l ti d t bl d h- Strangulation due to cables and hoses
- Small parts being inhaled or swallowed
- Allergic reactions to accessible materialsg
- Contact injuries
www.intertek.com37
Clause 7.4.1 – Additional Requirements for Warning and Safety Notices
The IFU shall include warnings that it is unsafe to:
U i t d ib d i th IFU- Use accessories not described in the IFU
- Interconnect equipment to other equipment not described in the IFU
- Modify the equipmenty q p
- Use equipment outside of the carrying case, if the case provides a degree of protection
www.intertek.com38
Clause 7.4.2 – Additional Requirements for an Electrical Power Source
If Basic Safety and Essential Performance is dependent on an Internal Electrical Power Source, the IFU shall describe the following:
- Typical operation time or number of procedures
- Typical service life
- For a Rechargeable Internal Electrical Power Source, the behavior of the equipment while charging
www.intertek.com39
Clause 7.4.3 – Additional Requirements for ME Equipment Description
The IFU shall include easily understood (effective):
• diagrams,
• illustrations, or
h t h• photographs
of a fully assembled, ready to operate equipment including:
• controls,
• visual information signals, and
• indicators• indicators
www.intertek.com40
Clause 7.4.4 – Additional Requirements for ME Equipment Start-up Procedure
The IFU shall include easily understood (effective):
• diagrams,diagrams,
• illustrations, or
• photographs
of the proper connection of the Patient to the equipment, accessories or other equipment.
www.intertek.com41
Clause 7.4.4 – Additional Requirements for ME Equipment Start-up Procedure
The IFU shall include the time from switching “ON” until the equipment is ready for Normal Use, if the time is greater than 15 seconds
www.intertek.com42
Clause 7.4.5 – Additional Requirements for Operating Instructions
The IFU shall include information on conditions in the Home Healthcare E i t th t ff t B i S f t d E ti l P f i l diEnvironment that can affect Basic Safety and Essential Performance including:
- Effects of lint, dust, or lightEffects of lint, dust, or light
- Devices or sources that could cause interference problems
- Effects of degraded sensors and electrodes
- Effects of pets, pests, and children
Information shall include steps the Lay Operator can take to identify andInformation shall include steps the Lay Operator can take to identify and resolves these issues.
www.intertek.com43
Clause 7.4.5 – Additional Requirements for Operating Instructions
The IFU shall explain the meaning of the IP rating that appears on either:
Th i t- The equipment, or
- The equipment’s carrying case
www.intertek.com44
Clause 7.4.6 – Additional Requirements for ME Equipment Messages
The IFU shall include a troubleshooting guide for malfunctions during start-up and operation.
The troubleshooting guide shall disclose the necessary steps to be taken for an alarm condition.
www.intertek.com45
Clause 7.4.7 – Additional Requirements for Cleaning, Disinfection and Sterilization
The IFU shall indicate, for equipment, parts, or accessories, the frequency of:
- cleaning,
- cleaning and disinfection, or
l i d t ili ti- cleaning and sterilization
The IFU shall also include the following methods for:
- drying,y g,
- handling, and
- storage between uses
www.intertek.com46
Clause 7.4.7 – Additional Requirements for Cleaning, Disinfection and Sterilization
These instructions shall cover either:
- single patient,
lti l ti t- multiple patients, or
- indicate that professional hygienic maintenance prior to re-use is required (including contact information for the service)
www.intertek.com47
Clause 7.4.8 – Additional Requirements for Maintenance
The IFU shall include the following information:
- Expected Service Life of the Medical Electrical Equipment
E t d S i Lif f i hi d ith M di l El t i l- Expected Service Life of accessories, shipped with Medical Electrical Equipment
- If the Shelf Life is less than the Expected Service Life, the Shelf Life of accessories shipped ith the Medical Electrical Eq ipmentof accessories, shipped with the Medical Electrical Equipment
www.intertek.com48
Clause 7.4.9 – Additional Requirements for Environmental Protection
The IFU shall include information on:
di l f th M di l El t i l E i t t- proper disposal of the Medical Electrical Equipment, parts, or accessories and
- a statement that the Lay Responsible Organization shall contact local a thorities for the disposal of potentiall bio ha ardo s parts andauthorities for the disposal of potentially bio-hazardous parts and accessories
www.intertek.com49
Clause 7.4.10 – Additional Requirements for ME Equipment and ME Systems
For equipment that incorporates the use of a Distributed Alarm System, the IFU shall state the recommended placement of the remote part of the Distributed Alarm System to ensure the operator is notified at all times.
www.intertek.com50
Clause 7.5.1 – Permanently Installed Class I Equipment
For Permanently Installed Class I Equipment, the Technical Description shall i l dinclude:
- Warning that connection of Protection Earth is completed by aWarning that connection of Protection Earth is completed by a qualified Service Personnel
- Specifications for the Protective Earth Conductor
Warning to erif the integrit of the e ternal protecti e earthing- Warning to verify the integrity of the external protective earthing system
- Warning to connect and verify the Protective Earth Terminal is connected to the external protective earth systemconnected to the external protective earth system
www.intertek.com51
Clause 7.5.2 – Additional Requirements for Professional Hygienic Maintenance
If equipment requires profession hygienic maintenance prior to re-use
Th T h i l D i ti i l d l i d di i f ti l i dThe Technical Description include cleaning and disinfection or cleaning and sterilization:
- before and after any servicing
- when equipment is transferred to another patient
www.intertek.com52
Clause 7.5.2 – Additional Requirements for Cleaning & Disinfection of ME Equipment and ME Systems
Cleaning and disinfection procedures for equipment in the Home Healthcare E i t h ll b bl f b i f d b L O tEnvironment shall be capable of being preformed by a Lay Operator.
The Usability of these processes shall be documented in the UsabilityThe Usability of these processes shall be documented in the Usability Engineering File
www.intertek.com53
Clause 8.2 – Additional Requirements for Sterilization of ME Equipment and ME Systems
Sterilization procedures for equipment in the Home Healthcare Environment h ll b bl f b i f d b L O tshall be capable of being preformed by a Lay Operator.
The Usability of this process shall be documented in the Usability EngineeringThe Usability of this process shall be documented in the Usability Engineering File
www.intertek.com54
Clause 8.3.1 – Ingress of Water or Particulate Matter into ME Equipment
An IP Rating of IP22 shall be provided for the following equipment:
) T it bl H d h ld i ta) Transit-operable, Hand-held equipment
b) Body-worn equipment
All other Transportable Equipment shall be provided with an IP Rating of IP21
If carrying case provides the protection, then the case shall be marked.
www.intertek.com55
Clause 8.3.2 – Ingress of Water or Particulate Matter into ME Equipment
Enclosures of non-ME Equipment shall also provide a degree of protection against harmful ingress of water or particulate matter in accordance with there relevant end of product standards
www.intertek.com56
Clause 8.4 – Additional Requirements for Interruption of the Power Supply / Supply Mains to ME Equipment and ME Systems
Life-supporting ME Equipment or ME Systems shall maintain Essential P f f ffi i t ti b f d t ll f lt tPerformance for a sufficient time or number of procedures, to allow for alternate methods to be employed, after failure of the supply mains.
The following types of devices can be utilized for maintaining Essential Performance:
- Internal Electrical Power Source or
- An independent means
www.intertek.com57
Clause 8.4 – Additional Requirements for Interruption of the Power Supply / Supply Mains to ME Equipment and ME Systems
The IFU shall disclose the time or number of procedures available following the failure of electrical power system
The IFU shall describe the alternate means of life support employed
The Technical Description shall describe methods that can employed for a longer period of time
www.intertek.com58
Clause 8.4 – Additional Requirements for Interruption of the Power Supply / Supply Mains to ME Equipment and ME Systems
If a Internal Electrical Power Source is not used, the equipment shall be id d ith l t i l di di i it l fprovided with an alarm system including an medium priority alarm for power
supply failure.
If the equipment incorporates an Internal Electrical Power Source automatic switch over shall occur in the case of power supply failure.
The equipment shall be provided with an alarm system including an low priority alarm for the automatic switchover.
www.intertek.com59
Clause 8.4 – Additional Requirements for Interruption of the Power Supply / Supply Mains to ME Equipment and ME Systems
The equipment shall be provided with an alarm system including a medium i it t h i l l i di ti th I t l El t i l P S i ipriority technical alarm indicating the Internal Electrical Power Source is nearing
insufficient remaining power to keep the equipment functional.
The technical alarm shall provide the Lay Operator sufficient time or number of procedures to act.
A low priority technical alarm shall remain on until the Internal Electrical Power Source is returned above the alarm limit.
www.intertek.com60
Clause 9 – Accuracy of Controls & Instruments and Protection against Hazardous Outputs
The following, at a minimum, shall be considered as part of the Usability E i i P d FilEngineering Process and File:
Changes of Controls
Confusion in Operation Modes
Unexpected Movement
Transfer of Energy /
Substances
Potential of Disconnection
Exposure to Biological Improper / Unsafe
OperationParts Inhaled /
SwallowedDisconnection Materials Operation Swallowed
www.intertek.com61
Clause 10.1.1 – General Requirements for Mechanical Strength – Table 1
• Push, Drop, Molding Stress Relief, Shock, and Vibration
Non-Transit-operable and Hand-Held
• Push, Impact, Drop, Molding Stress Relief, Shock, and Vibration
Non-Transit-operable and Body-worn
• Push, Impact, Drop, Molding Stress Relief, Shock, and Vibration
Non-Transit-operable and Portable
• Push, Impact, Drop, Rough Handling, Molding Stress Relief, Shock, and
Non-Transit-operable and Mobile
Vibration
Fixed or Stationary
www.intertek.com62
• Push, Impact, and Molding Stress Relief
Clause 10.1.1 – General Requirements for Mechanical Strength – Table 2
P h D M ldi S R li f Sh k d Vib i
Transit-operable and Hand-Held• Push, Drop, Molding Stress Relief, Shock, and Vibration
Transit-operable and Body-worn• Push, Impact, Drop, Molding Stress Relief, Shock, and Vibration
Transit-operable and Portable• Push, Molding Stress Relief, Free Fall, Shock, and Vibration
Transit-operable and Portable
• Push, Impact, Rough Handling, Molding Stress Relief, Free Fall, Shock and Vibration
Transit-operable and Mobile
www.intertek.com63
Shock, and Vibration
Clause 10.1.2 – Requirements for Mechanical Strength for Non-Transit-Operable ME Equipment
a) Shock Test in accordance IEC 60068-2-27: 2008
- Three shocks per direction per axis (18 total)
b) Broad-band Random Vibration Test in accordance with IEC 60068-2-64: 2008
- 30 minutes per perpendicular axis
c) Verify that Basic Safety and Essential Performance are maintainedc) Verify that Basic Safety and Essential Performance are maintained
www.intertek.com64
Clause 10.1.3 – Requirements for Mechanical Strength for Transit-Operable ME Equipment
For Body-worn Equipment
a) Shock Test in accordance with IEC 60068-2-27: 2008
- Either Type 1 or Type 2 three shocks per direction per axisEither Type 1 or Type 2, three shocks per direction per axis
b) Broad-band Random Vibration Test in accordance with IEC 60068-2-64: 20082008
- 30 min per perpendicular axis (three total)
c) Verify that Basic Safety and Essential Performance are maintained
www.intertek.com65
Clause 10.1.3 – Requirements for Mechanical Strength for Transit-Operable ME Equipment
For Hand-Held Equipment
a) Shock Test in accordance with IEC 60068-2-27: 2008
Eith T 1 T 2 th h k di ti i- Either Type 1 or Type 2, three shocks per direction per axis
b) Broad-band Random Vibration Test in accordance with IEC 60068-2-64: )2008
- 30 min per perpendicular axis (three total)
c) Verify that Basic Safety and Essential Performance are maintained
www.intertek.com66
Clause 10.1.3 – Requirements for Mechanical Strength for Transit-Operable ME Equipment
For Portable and Mobile Equipment
a) Shock Test in accordance with IEC 60068-2-27: 2008
- Either Type 1 or Type 2, three shocks per direction per axisEither Type 1 or Type 2, three shocks per direction per axis
b) Broad-band Random Vibration Test in accordance with IEC 60068-2-64: 2008
30 min per perpendic lar a is (three total)- 30 min per perpendicular axis (three total)
c) Free Fall Test in accordance with IEC 60068-2-31: 2008
- Procedure 1, two falls in each specified altitudep
d) Verify that Basic Safety and Essential Performance are maintained
www.intertek.com67
Clause 10.2 – Additional Requirements for Internal Electrical Power Sources
If a Internal Electrical Power Source is essential to maintain Basic Safety, E ti l P f t l i k f l f E ti l P f itEssential Performance, or control risks from loss of Essential Performance it shall be provided with one of the following means of indication:
- Number of procedures remaining
- Remaining operating time
- Percentage of remaining operating time or energy
a “f el” ga ge- a “fuel” gauge
This means may be indicated continuously or by operator action.
The IFU shall state how to determine the state of the Internal Electrical Power Source
www.intertek.com68
Clause 10.3 – Additional Requirements for Actuating Parts of ME Equipment
Control of Medical Electrical Equipment that can affect Basic Safety or Essential P f h ll b t t d f id t l th i d dj t tPerformance shall be protected from accidental or unauthorized adjustment
- Access controlled by tool
- Access controlled by Responsible Organization passwordAccess controlled by Responsible Organization password
- Access controlled by Operator password
- Access controlled by fingerprints
Operator-adjustable controls for calibration shall include means to prevent unintentional changesg
www.intertek.com69
Clause 11 – Protection Against Strangulation or Asphyxiation
Means shall be provided to control the risk of strangulation or asphyxiation of the ti t d thpatient and others
- By routing of wirings/tubing
- By retention devicesBy retention devices
- Multiple length accessories
- Not providing removable small pieces
www.intertek.com70
Clause 12.1 – Emissions Classification
Medical Electrical Equipment and Medical Electrical Systems in the Home Healthcare Environment shall be classified as Class B according to CISPR 11: 2009
www.intertek.com71
Clause 12.2 – Protection of the Public Mains Network
For Home Healthcare Equipment the Public Mains Network requirements of Clause 6.1.3 of IEC 60601-1-2: 2007 apply
www.intertek.com72
Clause 12.3 – Additional Technical Description Requirements Applicable to ME Equipment & ME Systems
The IFU shall have a statement that the equipment is to be installed and put into i di t th i f ti i th i d tservice according to the information in the accompanying documents
The IFU shall have a statement that wireless communication equipment canThe IFU shall have a statement that wireless communication equipment can affect the equipment and should be kept at least a distance of “d” away from the equipment.
“d” is calculated by the manufacturer from 800 MHz to 2.5 GHz using Tables 5 y gand 6 of IEC 60601-1-2: 2007
www.intertek.com73
Clause 12.5 – Additional Requirements for Electrostatic Discharge (ESD) Tests
The standard test finger of the general standard (Figure 6) is used to determine ibl t d i th ESD t taccessible parts during the ESD test
www.intertek.com74
Clause 13.1 – Additional Requirements for Generation of Alarm Signals
Each High and Medium Priority Alarm provided in the Home Healthcare E i t h ll th ti f di lEquipment shall cause the generation of an audio alarm
Exception – Unless connected to a Distributed Alarm System that includesException Unless connected to a Distributed Alarm System that includes the generation of auditory alarm signals
www.intertek.com75
Clause 13.2 – Additional Requirements for of Alarm Signal Volume
If the Home Healthcare Equipment is capable of setting the volume below audio l l th i t h lllevels the equipment shall:
- Activate the indication of “ALARM OFF” or “AUDIO OFF” per IECActivate the indication of ALARM OFF or AUDIO OFF per IEC 60601-1-8
- For Life-supporting Medical Electrical Equipment not be possible unless connected to a Distributed Alarm System capable of generating y p g gAuditory Alarm Signals
www.intertek.com76
Part 4Why do products fail?
www.intertek.com77
Why do products fail?
Manuals
MarkingsMarkings
www.intertek.com78
Want More?