high blood pressure 8 To get more 2 decades, noninvasive hypertension (NIBP) monitors have already been widely used in operating rooms and critical care units to closely monitor blood pressure levels in patients of every age group. In spite of the widespread usage of automated blood pressure levels monitors, clinicians continue to deliberate over the accuracy and reliability of automated NIBP devices in comparison to other types of blood pressure levels determination. The next strategies to commonly asked questions about the usage of NIBP monitoring are derived from clinical research. Q: How can blood pressure measurements obtained with automatic noninvasive blood pressure levels (NIBP) devices compare with direct arterial measurement of blood pressure level? Clinical scientific studies have demonstrated that if blood pressures diastolic and systolic, and mean arterial) based on NIBP monitors from various manufacturers are compared to direct arterial pressures, the 2 values are, on average, within 5 mm Hg for each other. (1-9) Factors including the anatomical location of measurement play a role in the differences that can be found between direct and indirect methods. For example, in comparing brachial arterial pressure obtained with the NIBP monitor to radial arterial pressure obtained by direct arterial cannulation, radial arterial pressure is usually higher because the radial artery is a smaller vessel and fosters greater effectiveness against flow, which in turn results in a higher blood pressure levels reading than that determined using the brachial artery.

high blood pressure 8

Embed Size (px)

Citation preview

Page 1: high blood pressure 8

high blood pressure 8

To get more 2 decades, noninvasive hypertension (NIBP) monitors have already been widely used inoperating rooms and critical care units to closely monitor blood pressure levels in patients of everyage group. In spite of the widespread usage of automated blood pressure levels monitors, clinicianscontinue to deliberate over the accuracy and reliability of automated NIBP devices in comparison toother types of blood pressure levels determination. The next strategies to commonly asked questionsabout the usage of NIBP monitoring are derived from clinical research.

Q: How can blood pressure measurements obtainedwith automatic noninvasive blood pressure levels (NIBP)devices compare with direct arterial measurement ofblood pressure level?

Clinical scientific studies have demonstrated that if blood pressures diastolic and systolic, and meanarterial) based on NIBP monitors from various manufacturers are compared to direct arterialpressures, the 2 values are, on average, within 5 mm Hg for each other. (1-9) Factors including theanatomical location of measurement play a role in the differences that can be found between directand indirect methods. For example, in comparing brachial arterial pressure obtained with the NIBPmonitor to radial arterial pressure obtained by direct arterial cannulation, radial arterial pressure isusually higher because the radial artery is a smaller vessel and fosters greater effectiveness againstflow, which in turn results in a higher blood pressure levels reading than that determined using thebrachial artery.

Page 2: high blood pressure 8

Thesestudiesalsoindicatethatoccas

ionally, somebody NIBP blood pressure level determination value can vary by up to 37 mm Hg fromthe direct arterial value. (1-9) This large discrepancy signifies that treatment should not be madebased on a single NIBP determination without comparison to a auscultatory blood pressure leveldetermination or several consecutive measurements high blood pressure symptoms in men done byan NIBP monitor.

Q: Is it vital that you take advantage of the correct cuff size when working with automatic NIBPdevices?

Yes. Employing a cuff which is 1oo small will cause falsely high readings, and making use of a cuffthat may be too large will result in falsely low readings. The cuff width selected should equal 40% ofyour arm circumference (see Figure). The American Heart Association strategies for appropriate cuffsizes depending on upper-arm circumference should also be followed when working with NIBPmonitors (see Table). (10)

Page 3: high blood pressure 8

Q: Any kind of patient-related complications associated with using NIBP devices?

Skin and tissue compression from NIBP monitors, which can cause skin irritation and bruising, areprobably the most commonly occurring complications. Prolonged use and frequent blood pressurelevel determinations can cause venous pooling and congestion. Excessive venous pressures canbring about tissue ischemia and nerve damage. (11,12)

Q: What factors can hinder obtaining accurate NIBP measurements?

Page 4: high blood pressure 8

Several circumstances http://www.nhlbi.nih.gov/health/health-topics/topics/hbp/ can preventaccurate determination of blood pressure level with NIBP devices. Highly irregular or rapid cardiacrhythms ensure it is challenging to accurately determine blood pressure level using NIBP devicesdue to great beat to overcome variability. Most NIBP devices employ oscillometric technology that iscertainly influenced by fairly regular cardiac rhythms to find out blood pressure level. Excessivepatient movement such as shivering, restlessness, or external movement for example that from ahelicopter, ambulance transport, or even a rapid-cycling ventilator can obstruct detection of cardiacoscillations through the NIBP monitor. This can lead to erroneous hypertension measurements.