Output calibration using TRS 398

Sothing vashum

Page 2: Output  calibration  using trs 398

Table of contents

Beam calibration in Co-60 gamma ray beam - dosimetry equipment

- phantom and chamber sleeves

- reference conditions

- determination of absorbed dose under reference condition

- worksheet

Page 3: Output  calibration  using trs 398

Dosimetry equipment Recommended reference instrument for calibration - cylindrical and plane-parallel chamber

In cylindrical – chamber axis at the centre of the cavity volume

In plane-parallel – the inner surface of the entrance window at the centre of the window

THE REFERENCE POINT SHOULD BE POSITIONED AT THE REFERENCE DEPTH IN A WATER PHANTOM

Page 4: Output  calibration  using trs 398

Phantom and chamber sleeve

Extend 5cm – four sides –field size – at depth of measurement

Extend at least 5g/cm^2 beyond max depth of measurement.

Page 5: Output  calibration  using trs 398

Reference conditions

SSD/SCD – 80cm/100cm field size - 10cm * 10cm

Page 6: Output  calibration  using trs 398
Page 7: Output  calibration  using trs 398

Co-60

After the positioning of the chamber according to the reference conditions, the absorbed dose to water is given by

For a dosimeter in a reference beam of quality Qo, where Qo is Co-60, in this situation,

DW,Q = DW , KQ,QO = KQ ( which has a value of unity) and ND,W,QO = ND,W.

DW,Q = MQND,W,QOKQ,QO

Page 8: Output  calibration  using trs 398

then we have,

Here, MQ = corrected meter reading which includes KTP,

Kpol, Ksat where KTP, which is the correction factor for temperature and

pressure given by,

KTP = (273.2 + T) Po/ (273.2 + To)P

To = 20oC ; Po = 760 mmHg

DW = MQND,W.

Page 9: Output  calibration  using trs 398

Kpol, which is the factor to correct the response of an

ionization chamber for the effect of change in polarity of the polarizing voltage applied to the chamber given by

Kpol = |M+| + |M-|/2M

M+ = meter reading for +300V

M - = meter reading for -300V

Page 10: Output  calibration  using trs 398

Ksat, factor to correct the response of an ionization chamber for the lack of complete charge collection given by

Ksat = (V1/V2)^2 – 1/ (V1/V2)^2 – (M1/M2)

V1 = voltage (+300V) V2 = voltage (+150V)

M1 = meter reading for voltage V1(+300V)

M2 = meter reading for voltage V2(+150V)

Page 11: Output  calibration  using trs 398

The Ksat for a pulsed beam is given by,

Ksat = ao + a1(M1/M2) +a2(M1/M2)^2

Where ao, a1 and a2 are constants and the values are

2.303, -3.636 and 2.299 respectively.