5020 hornpuff resuscitator

Embed Size (px)

Citation preview

Page 1: 5020 hornpuff resuscitator

nice NeOumltech Medical Systems Pvt Ltd

Presentation Flow

COMPANY OVERVIEW

CLINICAL ASPECTS OF nice 5020 RESUSCITATOR

PRODUCT OVERVIEW

TROUBLESHOOTING

Company INFO

Established in the year 1997 Group of highly skilled professionals 4000+ highly satisfied customers Deep penetration in NICU and respiratory domain User friendly range of products With deep skills and high end technology we always strive

for growth and delighted customers

Timeline-nice Neotech

2005bullFirst New Factory Unit 1 Opened

bullIncrease in production Line

2001bullRegistered as Pvt ltd

bullProduction started for different models

2013bullSecond New Factory Unit 2 Opened

1999bull Neotech Logo

Designed

bull Designed first incubator amp open

care prototype

2003bull Export to

Vietnam

2009-2011

bull CE certification for products

2015bull 125+skilled

employee strong firm

bull Multiple range of NICU products

1997bullStarted Neotech Medical Systems with 2 employees

bullFirst Infant Radiant Warmer

Why nice Neotech

User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction

What is neonatal resuscitation

Newborn resuscitation is a series of actions which are used to

assist newborn babies who have difficulty with making the

physiological lsquotransitionrsquo between the womb and outside world rsquo

Newborn resuscitation assists babies who fail to initiate or sustain

regular breathing at birth

It maintain FRC (functional Residual Capacity) and recruitment of

alveoli which prevent lung collapse

Why Infant Resuscitators are Used

Prevent neonatal death

Ensure baby is stabilized after birth

Infant Resuscitators

Contrsquodhellip

Substantial physiologic changes occur in the transition from fetal to extra uterine life includingbullThe role of the placenta in gas exchange is taken over by the lungsbullChanges from fluid-filled to air filled lungsbullDramatic increase in blood flow to the lungs with reversal then closure of intra and extra cardiac shunts

Assistance to breathe at birthPrevent neonatal deathPrevent Adverse longterm neurodevelopmental sequelae

Failure or disruption of these changes may result in further difficulties with resuscitation in the newborn infantThe need for resuscitation of the newborn infant at birth cannot always be anticipated or predicted

Benefit

A range of Baby friendly equipments Introducing nice NeOumltech 5020 HORNPUFF Resuscitator

Designed to perfection for your baby

nice

502

0 H

OR

NP

UF

F R

esus

cita

tor

bullSafe Controlled and Precise Peak Inspiratory Pressure (PIP)bullFlexibility in OperationbullPrecise Positive End Expiratory PressurebullPrecise Pressure DeliverybullFace Mask of Different sizesbullCompact and HandybullMounting Options

Features

How to operate

Gas Supply Flow rate Setting the Maximum PressureSetting the Peak Inspiratory Pressure (PIP) Setting the Positive End Expiratory Pressure (PEEP)Resuscitate with nice 5020 Infant Resuscitator

Gas Supply Flow rate

1048729 Set the Flow rate at 5 LPM in the Flow Meter

Setting the Maximum Pressure

1048729 Occlude PEEP knob and turn PIP control fully clockwise adjust maximum pressure control knob clockwise or counterclockwise to set desired maximum pressure

Setting the Peak Inspiratory Pressure (PIP)

1048729 Occlude the PEEP knob turn PIP control knob counter-clockwise until the desired peak inspiratorypressure is set

Setting the Positive End Expiratory Pressure (PEEP)

1048729 Adjust PEEP knob clockwise or anti-clockwise to the desired PEEP level

Resuscitate with nice 5020 Infant Resuscitator1048729 Adjust gas supply to the desired flow rate

1048729 Fit patient T-piece to neonatal resuscitation mask and place over the babyrsquos mouth andor nose Fitpatient T-piece to the endotracheal tube

1048729 Resuscitate by placing and removing thumb over the PEEP knob to allow inspiration and expiration

How nice Resuscitatorfare over competitors

nice Resuscitator

User-friendly easy to access all parameters

Low consumables price

Compact and handy

Infant Resuscitator ndash nice 5020 Accessories

Nasal Prong

Humidifier Air Oxygen Blender Mounting Bracket

Patient Circuit

The addition of respiratory humidifier and Air-Oxygen blender improvises the usage of Resuscitator to a greater extent Humidifier adjusting the level of humidity thus minimizing evaporation andAir-Oxygen blender mixing the right amount of gas

Contdhelliphellip

Humidifier

Air Oxygen Blender

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1048729 Clean external surfaces of the nice 5020 infant resuscitator and Gas Supply line using a damp cloth and mild soapy water or Isopropyl Alcohol

1048729 Dry all surfaces after cleaning with a clean soft cloth or paper towel

1048729 Clean the Bag Resuscitator and any similar equipment in strict accordance with the Manufacturers instructions

1048729 All other components may be disinfected using an approved chemical disinfectant solution

1048729 When all components are completely dry assemble the system again

Preventive Maintenance

1048729 The nice 5020 infant resuscitator should require minimal servicing or maintenance when used under normal conditions

1048729 At least once every six months the equipment should be examined by a qualified Service Engineer to ensure safety and operational integrity

1048729 If any part of the equipment is damaged or fails to operate correctly take it out of service immediately and contact a qualified Service Engineer to ensure operational safety

1048729 A Service Manual enabling qualified technical personnel to carry out routine service schedules fault finding and repairs is available from nice Neotech

1048729 Always carry out a functional test after cleaning or maintenance to ensure safety and operational integrity

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to

internal components Do not autoclave or gas sterilize the unit Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Use of nonstandard components Consult the manufacturer for repair and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 2: 5020 hornpuff resuscitator

Presentation Flow

COMPANY OVERVIEW

CLINICAL ASPECTS OF nice 5020 RESUSCITATOR

PRODUCT OVERVIEW

TROUBLESHOOTING

Company INFO

Established in the year 1997 Group of highly skilled professionals 4000+ highly satisfied customers Deep penetration in NICU and respiratory domain User friendly range of products With deep skills and high end technology we always strive

for growth and delighted customers

Timeline-nice Neotech

2005bullFirst New Factory Unit 1 Opened

bullIncrease in production Line

2001bullRegistered as Pvt ltd

bullProduction started for different models

2013bullSecond New Factory Unit 2 Opened

1999bull Neotech Logo

Designed

bull Designed first incubator amp open

care prototype

2003bull Export to

Vietnam

2009-2011

bull CE certification for products

2015bull 125+skilled

employee strong firm

bull Multiple range of NICU products

1997bullStarted Neotech Medical Systems with 2 employees

bullFirst Infant Radiant Warmer

Why nice Neotech

User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction

What is neonatal resuscitation

Newborn resuscitation is a series of actions which are used to

assist newborn babies who have difficulty with making the

physiological lsquotransitionrsquo between the womb and outside world rsquo

Newborn resuscitation assists babies who fail to initiate or sustain

regular breathing at birth

It maintain FRC (functional Residual Capacity) and recruitment of

alveoli which prevent lung collapse

Why Infant Resuscitators are Used

Prevent neonatal death

Ensure baby is stabilized after birth

Infant Resuscitators

Contrsquodhellip

Substantial physiologic changes occur in the transition from fetal to extra uterine life includingbullThe role of the placenta in gas exchange is taken over by the lungsbullChanges from fluid-filled to air filled lungsbullDramatic increase in blood flow to the lungs with reversal then closure of intra and extra cardiac shunts

Assistance to breathe at birthPrevent neonatal deathPrevent Adverse longterm neurodevelopmental sequelae

Failure or disruption of these changes may result in further difficulties with resuscitation in the newborn infantThe need for resuscitation of the newborn infant at birth cannot always be anticipated or predicted

Benefit

A range of Baby friendly equipments Introducing nice NeOumltech 5020 HORNPUFF Resuscitator

Designed to perfection for your baby

nice

502

0 H

OR

NP

UF

F R

esus

cita

tor

bullSafe Controlled and Precise Peak Inspiratory Pressure (PIP)bullFlexibility in OperationbullPrecise Positive End Expiratory PressurebullPrecise Pressure DeliverybullFace Mask of Different sizesbullCompact and HandybullMounting Options

Features

How to operate

Gas Supply Flow rate Setting the Maximum PressureSetting the Peak Inspiratory Pressure (PIP) Setting the Positive End Expiratory Pressure (PEEP)Resuscitate with nice 5020 Infant Resuscitator

Gas Supply Flow rate

1048729 Set the Flow rate at 5 LPM in the Flow Meter

Setting the Maximum Pressure

1048729 Occlude PEEP knob and turn PIP control fully clockwise adjust maximum pressure control knob clockwise or counterclockwise to set desired maximum pressure

Setting the Peak Inspiratory Pressure (PIP)

1048729 Occlude the PEEP knob turn PIP control knob counter-clockwise until the desired peak inspiratorypressure is set

Setting the Positive End Expiratory Pressure (PEEP)

1048729 Adjust PEEP knob clockwise or anti-clockwise to the desired PEEP level

Resuscitate with nice 5020 Infant Resuscitator1048729 Adjust gas supply to the desired flow rate

1048729 Fit patient T-piece to neonatal resuscitation mask and place over the babyrsquos mouth andor nose Fitpatient T-piece to the endotracheal tube

1048729 Resuscitate by placing and removing thumb over the PEEP knob to allow inspiration and expiration

How nice Resuscitatorfare over competitors

nice Resuscitator

User-friendly easy to access all parameters

Low consumables price

Compact and handy

Infant Resuscitator ndash nice 5020 Accessories

Nasal Prong

Humidifier Air Oxygen Blender Mounting Bracket

Patient Circuit

The addition of respiratory humidifier and Air-Oxygen blender improvises the usage of Resuscitator to a greater extent Humidifier adjusting the level of humidity thus minimizing evaporation andAir-Oxygen blender mixing the right amount of gas

Contdhelliphellip

Humidifier

Air Oxygen Blender

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1048729 Clean external surfaces of the nice 5020 infant resuscitator and Gas Supply line using a damp cloth and mild soapy water or Isopropyl Alcohol

1048729 Dry all surfaces after cleaning with a clean soft cloth or paper towel

1048729 Clean the Bag Resuscitator and any similar equipment in strict accordance with the Manufacturers instructions

1048729 All other components may be disinfected using an approved chemical disinfectant solution

1048729 When all components are completely dry assemble the system again

Preventive Maintenance

1048729 The nice 5020 infant resuscitator should require minimal servicing or maintenance when used under normal conditions

1048729 At least once every six months the equipment should be examined by a qualified Service Engineer to ensure safety and operational integrity

1048729 If any part of the equipment is damaged or fails to operate correctly take it out of service immediately and contact a qualified Service Engineer to ensure operational safety

1048729 A Service Manual enabling qualified technical personnel to carry out routine service schedules fault finding and repairs is available from nice Neotech

1048729 Always carry out a functional test after cleaning or maintenance to ensure safety and operational integrity

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to

internal components Do not autoclave or gas sterilize the unit Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Use of nonstandard components Consult the manufacturer for repair and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 3: 5020 hornpuff resuscitator

Company INFO

Established in the year 1997 Group of highly skilled professionals 4000+ highly satisfied customers Deep penetration in NICU and respiratory domain User friendly range of products With deep skills and high end technology we always strive

for growth and delighted customers

Timeline-nice Neotech

2005bullFirst New Factory Unit 1 Opened

bullIncrease in production Line

2001bullRegistered as Pvt ltd

bullProduction started for different models

2013bullSecond New Factory Unit 2 Opened

1999bull Neotech Logo

Designed

bull Designed first incubator amp open

care prototype

2003bull Export to

Vietnam

2009-2011

bull CE certification for products

2015bull 125+skilled

employee strong firm

bull Multiple range of NICU products

1997bullStarted Neotech Medical Systems with 2 employees

bullFirst Infant Radiant Warmer

Why nice Neotech

User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction

What is neonatal resuscitation

Newborn resuscitation is a series of actions which are used to

assist newborn babies who have difficulty with making the

physiological lsquotransitionrsquo between the womb and outside world rsquo

Newborn resuscitation assists babies who fail to initiate or sustain

regular breathing at birth

It maintain FRC (functional Residual Capacity) and recruitment of

alveoli which prevent lung collapse

Why Infant Resuscitators are Used

Prevent neonatal death

Ensure baby is stabilized after birth

Infant Resuscitators

Contrsquodhellip

Substantial physiologic changes occur in the transition from fetal to extra uterine life includingbullThe role of the placenta in gas exchange is taken over by the lungsbullChanges from fluid-filled to air filled lungsbullDramatic increase in blood flow to the lungs with reversal then closure of intra and extra cardiac shunts

Assistance to breathe at birthPrevent neonatal deathPrevent Adverse longterm neurodevelopmental sequelae

Failure or disruption of these changes may result in further difficulties with resuscitation in the newborn infantThe need for resuscitation of the newborn infant at birth cannot always be anticipated or predicted

Benefit

A range of Baby friendly equipments Introducing nice NeOumltech 5020 HORNPUFF Resuscitator

Designed to perfection for your baby

nice

502

0 H

OR

NP

UF

F R

esus

cita

tor

bullSafe Controlled and Precise Peak Inspiratory Pressure (PIP)bullFlexibility in OperationbullPrecise Positive End Expiratory PressurebullPrecise Pressure DeliverybullFace Mask of Different sizesbullCompact and HandybullMounting Options

Features

How to operate

Gas Supply Flow rate Setting the Maximum PressureSetting the Peak Inspiratory Pressure (PIP) Setting the Positive End Expiratory Pressure (PEEP)Resuscitate with nice 5020 Infant Resuscitator

Gas Supply Flow rate

1048729 Set the Flow rate at 5 LPM in the Flow Meter

Setting the Maximum Pressure

1048729 Occlude PEEP knob and turn PIP control fully clockwise adjust maximum pressure control knob clockwise or counterclockwise to set desired maximum pressure

Setting the Peak Inspiratory Pressure (PIP)

1048729 Occlude the PEEP knob turn PIP control knob counter-clockwise until the desired peak inspiratorypressure is set

Setting the Positive End Expiratory Pressure (PEEP)

1048729 Adjust PEEP knob clockwise or anti-clockwise to the desired PEEP level

Resuscitate with nice 5020 Infant Resuscitator1048729 Adjust gas supply to the desired flow rate

1048729 Fit patient T-piece to neonatal resuscitation mask and place over the babyrsquos mouth andor nose Fitpatient T-piece to the endotracheal tube

1048729 Resuscitate by placing and removing thumb over the PEEP knob to allow inspiration and expiration

How nice Resuscitatorfare over competitors

nice Resuscitator

User-friendly easy to access all parameters

Low consumables price

Compact and handy

Infant Resuscitator ndash nice 5020 Accessories

Nasal Prong

Humidifier Air Oxygen Blender Mounting Bracket

Patient Circuit

The addition of respiratory humidifier and Air-Oxygen blender improvises the usage of Resuscitator to a greater extent Humidifier adjusting the level of humidity thus minimizing evaporation andAir-Oxygen blender mixing the right amount of gas

Contdhelliphellip

Humidifier

Air Oxygen Blender

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1048729 Clean external surfaces of the nice 5020 infant resuscitator and Gas Supply line using a damp cloth and mild soapy water or Isopropyl Alcohol

1048729 Dry all surfaces after cleaning with a clean soft cloth or paper towel

1048729 Clean the Bag Resuscitator and any similar equipment in strict accordance with the Manufacturers instructions

1048729 All other components may be disinfected using an approved chemical disinfectant solution

1048729 When all components are completely dry assemble the system again

Preventive Maintenance

1048729 The nice 5020 infant resuscitator should require minimal servicing or maintenance when used under normal conditions

1048729 At least once every six months the equipment should be examined by a qualified Service Engineer to ensure safety and operational integrity

1048729 If any part of the equipment is damaged or fails to operate correctly take it out of service immediately and contact a qualified Service Engineer to ensure operational safety

1048729 A Service Manual enabling qualified technical personnel to carry out routine service schedules fault finding and repairs is available from nice Neotech

1048729 Always carry out a functional test after cleaning or maintenance to ensure safety and operational integrity

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to

internal components Do not autoclave or gas sterilize the unit Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Use of nonstandard components Consult the manufacturer for repair and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 4: 5020 hornpuff resuscitator

Timeline-nice Neotech

2005bullFirst New Factory Unit 1 Opened

bullIncrease in production Line

2001bullRegistered as Pvt ltd

bullProduction started for different models

2013bullSecond New Factory Unit 2 Opened

1999bull Neotech Logo

Designed

bull Designed first incubator amp open

care prototype

2003bull Export to

Vietnam

2009-2011

bull CE certification for products

2015bull 125+skilled

employee strong firm

bull Multiple range of NICU products

1997bullStarted Neotech Medical Systems with 2 employees

bullFirst Infant Radiant Warmer

Why nice Neotech

User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction

What is neonatal resuscitation

Newborn resuscitation is a series of actions which are used to

assist newborn babies who have difficulty with making the

physiological lsquotransitionrsquo between the womb and outside world rsquo

Newborn resuscitation assists babies who fail to initiate or sustain

regular breathing at birth

It maintain FRC (functional Residual Capacity) and recruitment of

alveoli which prevent lung collapse

Why Infant Resuscitators are Used

Prevent neonatal death

Ensure baby is stabilized after birth

Infant Resuscitators

Contrsquodhellip

Substantial physiologic changes occur in the transition from fetal to extra uterine life includingbullThe role of the placenta in gas exchange is taken over by the lungsbullChanges from fluid-filled to air filled lungsbullDramatic increase in blood flow to the lungs with reversal then closure of intra and extra cardiac shunts

Assistance to breathe at birthPrevent neonatal deathPrevent Adverse longterm neurodevelopmental sequelae

Failure or disruption of these changes may result in further difficulties with resuscitation in the newborn infantThe need for resuscitation of the newborn infant at birth cannot always be anticipated or predicted

Benefit

A range of Baby friendly equipments Introducing nice NeOumltech 5020 HORNPUFF Resuscitator

Designed to perfection for your baby

nice

502

0 H

OR

NP

UF

F R

esus

cita

tor

bullSafe Controlled and Precise Peak Inspiratory Pressure (PIP)bullFlexibility in OperationbullPrecise Positive End Expiratory PressurebullPrecise Pressure DeliverybullFace Mask of Different sizesbullCompact and HandybullMounting Options

Features

How to operate

Gas Supply Flow rate Setting the Maximum PressureSetting the Peak Inspiratory Pressure (PIP) Setting the Positive End Expiratory Pressure (PEEP)Resuscitate with nice 5020 Infant Resuscitator

Gas Supply Flow rate

1048729 Set the Flow rate at 5 LPM in the Flow Meter

Setting the Maximum Pressure

1048729 Occlude PEEP knob and turn PIP control fully clockwise adjust maximum pressure control knob clockwise or counterclockwise to set desired maximum pressure

Setting the Peak Inspiratory Pressure (PIP)

1048729 Occlude the PEEP knob turn PIP control knob counter-clockwise until the desired peak inspiratorypressure is set

Setting the Positive End Expiratory Pressure (PEEP)

1048729 Adjust PEEP knob clockwise or anti-clockwise to the desired PEEP level

Resuscitate with nice 5020 Infant Resuscitator1048729 Adjust gas supply to the desired flow rate

1048729 Fit patient T-piece to neonatal resuscitation mask and place over the babyrsquos mouth andor nose Fitpatient T-piece to the endotracheal tube

1048729 Resuscitate by placing and removing thumb over the PEEP knob to allow inspiration and expiration

How nice Resuscitatorfare over competitors

nice Resuscitator

User-friendly easy to access all parameters

Low consumables price

Compact and handy

Infant Resuscitator ndash nice 5020 Accessories

Nasal Prong

Humidifier Air Oxygen Blender Mounting Bracket

Patient Circuit

The addition of respiratory humidifier and Air-Oxygen blender improvises the usage of Resuscitator to a greater extent Humidifier adjusting the level of humidity thus minimizing evaporation andAir-Oxygen blender mixing the right amount of gas

Contdhelliphellip

Humidifier

Air Oxygen Blender

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1048729 Clean external surfaces of the nice 5020 infant resuscitator and Gas Supply line using a damp cloth and mild soapy water or Isopropyl Alcohol

1048729 Dry all surfaces after cleaning with a clean soft cloth or paper towel

1048729 Clean the Bag Resuscitator and any similar equipment in strict accordance with the Manufacturers instructions

1048729 All other components may be disinfected using an approved chemical disinfectant solution

1048729 When all components are completely dry assemble the system again

Preventive Maintenance

1048729 The nice 5020 infant resuscitator should require minimal servicing or maintenance when used under normal conditions

1048729 At least once every six months the equipment should be examined by a qualified Service Engineer to ensure safety and operational integrity

1048729 If any part of the equipment is damaged or fails to operate correctly take it out of service immediately and contact a qualified Service Engineer to ensure operational safety

1048729 A Service Manual enabling qualified technical personnel to carry out routine service schedules fault finding and repairs is available from nice Neotech

1048729 Always carry out a functional test after cleaning or maintenance to ensure safety and operational integrity

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to

internal components Do not autoclave or gas sterilize the unit Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Use of nonstandard components Consult the manufacturer for repair and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 5: 5020 hornpuff resuscitator

Why nice Neotech

User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction

What is neonatal resuscitation

Newborn resuscitation is a series of actions which are used to

assist newborn babies who have difficulty with making the

physiological lsquotransitionrsquo between the womb and outside world rsquo

Newborn resuscitation assists babies who fail to initiate or sustain

regular breathing at birth

It maintain FRC (functional Residual Capacity) and recruitment of

alveoli which prevent lung collapse

Why Infant Resuscitators are Used

Prevent neonatal death

Ensure baby is stabilized after birth

Infant Resuscitators

Contrsquodhellip

Substantial physiologic changes occur in the transition from fetal to extra uterine life includingbullThe role of the placenta in gas exchange is taken over by the lungsbullChanges from fluid-filled to air filled lungsbullDramatic increase in blood flow to the lungs with reversal then closure of intra and extra cardiac shunts

Assistance to breathe at birthPrevent neonatal deathPrevent Adverse longterm neurodevelopmental sequelae

Failure or disruption of these changes may result in further difficulties with resuscitation in the newborn infantThe need for resuscitation of the newborn infant at birth cannot always be anticipated or predicted

Benefit

A range of Baby friendly equipments Introducing nice NeOumltech 5020 HORNPUFF Resuscitator

Designed to perfection for your baby

nice

502

0 H

OR

NP

UF

F R

esus

cita

tor

bullSafe Controlled and Precise Peak Inspiratory Pressure (PIP)bullFlexibility in OperationbullPrecise Positive End Expiratory PressurebullPrecise Pressure DeliverybullFace Mask of Different sizesbullCompact and HandybullMounting Options

Features

How to operate

Gas Supply Flow rate Setting the Maximum PressureSetting the Peak Inspiratory Pressure (PIP) Setting the Positive End Expiratory Pressure (PEEP)Resuscitate with nice 5020 Infant Resuscitator

Gas Supply Flow rate

1048729 Set the Flow rate at 5 LPM in the Flow Meter

Setting the Maximum Pressure

1048729 Occlude PEEP knob and turn PIP control fully clockwise adjust maximum pressure control knob clockwise or counterclockwise to set desired maximum pressure

Setting the Peak Inspiratory Pressure (PIP)

1048729 Occlude the PEEP knob turn PIP control knob counter-clockwise until the desired peak inspiratorypressure is set

Setting the Positive End Expiratory Pressure (PEEP)

1048729 Adjust PEEP knob clockwise or anti-clockwise to the desired PEEP level

Resuscitate with nice 5020 Infant Resuscitator1048729 Adjust gas supply to the desired flow rate

1048729 Fit patient T-piece to neonatal resuscitation mask and place over the babyrsquos mouth andor nose Fitpatient T-piece to the endotracheal tube

1048729 Resuscitate by placing and removing thumb over the PEEP knob to allow inspiration and expiration

How nice Resuscitatorfare over competitors

nice Resuscitator

User-friendly easy to access all parameters

Low consumables price

Compact and handy

Infant Resuscitator ndash nice 5020 Accessories

Nasal Prong

Humidifier Air Oxygen Blender Mounting Bracket

Patient Circuit

The addition of respiratory humidifier and Air-Oxygen blender improvises the usage of Resuscitator to a greater extent Humidifier adjusting the level of humidity thus minimizing evaporation andAir-Oxygen blender mixing the right amount of gas

Contdhelliphellip

Humidifier

Air Oxygen Blender

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1048729 Clean external surfaces of the nice 5020 infant resuscitator and Gas Supply line using a damp cloth and mild soapy water or Isopropyl Alcohol

1048729 Dry all surfaces after cleaning with a clean soft cloth or paper towel

1048729 Clean the Bag Resuscitator and any similar equipment in strict accordance with the Manufacturers instructions

1048729 All other components may be disinfected using an approved chemical disinfectant solution

1048729 When all components are completely dry assemble the system again

Preventive Maintenance

1048729 The nice 5020 infant resuscitator should require minimal servicing or maintenance when used under normal conditions

1048729 At least once every six months the equipment should be examined by a qualified Service Engineer to ensure safety and operational integrity

1048729 If any part of the equipment is damaged or fails to operate correctly take it out of service immediately and contact a qualified Service Engineer to ensure operational safety

1048729 A Service Manual enabling qualified technical personnel to carry out routine service schedules fault finding and repairs is available from nice Neotech

1048729 Always carry out a functional test after cleaning or maintenance to ensure safety and operational integrity

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to

internal components Do not autoclave or gas sterilize the unit Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Use of nonstandard components Consult the manufacturer for repair and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 6: 5020 hornpuff resuscitator

What is neonatal resuscitation

Newborn resuscitation is a series of actions which are used to

assist newborn babies who have difficulty with making the

physiological lsquotransitionrsquo between the womb and outside world rsquo

Newborn resuscitation assists babies who fail to initiate or sustain

regular breathing at birth

It maintain FRC (functional Residual Capacity) and recruitment of

alveoli which prevent lung collapse

Why Infant Resuscitators are Used

Prevent neonatal death

Ensure baby is stabilized after birth

Infant Resuscitators

Contrsquodhellip

Substantial physiologic changes occur in the transition from fetal to extra uterine life includingbullThe role of the placenta in gas exchange is taken over by the lungsbullChanges from fluid-filled to air filled lungsbullDramatic increase in blood flow to the lungs with reversal then closure of intra and extra cardiac shunts

Assistance to breathe at birthPrevent neonatal deathPrevent Adverse longterm neurodevelopmental sequelae

Failure or disruption of these changes may result in further difficulties with resuscitation in the newborn infantThe need for resuscitation of the newborn infant at birth cannot always be anticipated or predicted

Benefit

A range of Baby friendly equipments Introducing nice NeOumltech 5020 HORNPUFF Resuscitator

Designed to perfection for your baby

nice

502

0 H

OR

NP

UF

F R

esus

cita

tor

bullSafe Controlled and Precise Peak Inspiratory Pressure (PIP)bullFlexibility in OperationbullPrecise Positive End Expiratory PressurebullPrecise Pressure DeliverybullFace Mask of Different sizesbullCompact and HandybullMounting Options

Features

How to operate

Gas Supply Flow rate Setting the Maximum PressureSetting the Peak Inspiratory Pressure (PIP) Setting the Positive End Expiratory Pressure (PEEP)Resuscitate with nice 5020 Infant Resuscitator

Gas Supply Flow rate

1048729 Set the Flow rate at 5 LPM in the Flow Meter

Setting the Maximum Pressure

1048729 Occlude PEEP knob and turn PIP control fully clockwise adjust maximum pressure control knob clockwise or counterclockwise to set desired maximum pressure

Setting the Peak Inspiratory Pressure (PIP)

1048729 Occlude the PEEP knob turn PIP control knob counter-clockwise until the desired peak inspiratorypressure is set

Setting the Positive End Expiratory Pressure (PEEP)

1048729 Adjust PEEP knob clockwise or anti-clockwise to the desired PEEP level

Resuscitate with nice 5020 Infant Resuscitator1048729 Adjust gas supply to the desired flow rate

1048729 Fit patient T-piece to neonatal resuscitation mask and place over the babyrsquos mouth andor nose Fitpatient T-piece to the endotracheal tube

1048729 Resuscitate by placing and removing thumb over the PEEP knob to allow inspiration and expiration

How nice Resuscitatorfare over competitors

nice Resuscitator

User-friendly easy to access all parameters

Low consumables price

Compact and handy

Infant Resuscitator ndash nice 5020 Accessories

Nasal Prong

Humidifier Air Oxygen Blender Mounting Bracket

Patient Circuit

The addition of respiratory humidifier and Air-Oxygen blender improvises the usage of Resuscitator to a greater extent Humidifier adjusting the level of humidity thus minimizing evaporation andAir-Oxygen blender mixing the right amount of gas

Contdhelliphellip

Humidifier

Air Oxygen Blender

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1048729 Clean external surfaces of the nice 5020 infant resuscitator and Gas Supply line using a damp cloth and mild soapy water or Isopropyl Alcohol

1048729 Dry all surfaces after cleaning with a clean soft cloth or paper towel

1048729 Clean the Bag Resuscitator and any similar equipment in strict accordance with the Manufacturers instructions

1048729 All other components may be disinfected using an approved chemical disinfectant solution

1048729 When all components are completely dry assemble the system again

Preventive Maintenance

1048729 The nice 5020 infant resuscitator should require minimal servicing or maintenance when used under normal conditions

1048729 At least once every six months the equipment should be examined by a qualified Service Engineer to ensure safety and operational integrity

1048729 If any part of the equipment is damaged or fails to operate correctly take it out of service immediately and contact a qualified Service Engineer to ensure operational safety

1048729 A Service Manual enabling qualified technical personnel to carry out routine service schedules fault finding and repairs is available from nice Neotech

1048729 Always carry out a functional test after cleaning or maintenance to ensure safety and operational integrity

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to

internal components Do not autoclave or gas sterilize the unit Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Use of nonstandard components Consult the manufacturer for repair and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 7: 5020 hornpuff resuscitator

Why Infant Resuscitators are Used

Prevent neonatal death

Ensure baby is stabilized after birth

Infant Resuscitators

Contrsquodhellip

Substantial physiologic changes occur in the transition from fetal to extra uterine life includingbullThe role of the placenta in gas exchange is taken over by the lungsbullChanges from fluid-filled to air filled lungsbullDramatic increase in blood flow to the lungs with reversal then closure of intra and extra cardiac shunts

Assistance to breathe at birthPrevent neonatal deathPrevent Adverse longterm neurodevelopmental sequelae

Failure or disruption of these changes may result in further difficulties with resuscitation in the newborn infantThe need for resuscitation of the newborn infant at birth cannot always be anticipated or predicted

Benefit

A range of Baby friendly equipments Introducing nice NeOumltech 5020 HORNPUFF Resuscitator

Designed to perfection for your baby

nice

502

0 H

OR

NP

UF

F R

esus

cita

tor

bullSafe Controlled and Precise Peak Inspiratory Pressure (PIP)bullFlexibility in OperationbullPrecise Positive End Expiratory PressurebullPrecise Pressure DeliverybullFace Mask of Different sizesbullCompact and HandybullMounting Options

Features

How to operate

Gas Supply Flow rate Setting the Maximum PressureSetting the Peak Inspiratory Pressure (PIP) Setting the Positive End Expiratory Pressure (PEEP)Resuscitate with nice 5020 Infant Resuscitator

Gas Supply Flow rate

1048729 Set the Flow rate at 5 LPM in the Flow Meter

Setting the Maximum Pressure

1048729 Occlude PEEP knob and turn PIP control fully clockwise adjust maximum pressure control knob clockwise or counterclockwise to set desired maximum pressure

Setting the Peak Inspiratory Pressure (PIP)

1048729 Occlude the PEEP knob turn PIP control knob counter-clockwise until the desired peak inspiratorypressure is set

Setting the Positive End Expiratory Pressure (PEEP)

1048729 Adjust PEEP knob clockwise or anti-clockwise to the desired PEEP level

Resuscitate with nice 5020 Infant Resuscitator1048729 Adjust gas supply to the desired flow rate

1048729 Fit patient T-piece to neonatal resuscitation mask and place over the babyrsquos mouth andor nose Fitpatient T-piece to the endotracheal tube

1048729 Resuscitate by placing and removing thumb over the PEEP knob to allow inspiration and expiration

How nice Resuscitatorfare over competitors

nice Resuscitator

User-friendly easy to access all parameters

Low consumables price

Compact and handy

Infant Resuscitator ndash nice 5020 Accessories

Nasal Prong

Humidifier Air Oxygen Blender Mounting Bracket

Patient Circuit

The addition of respiratory humidifier and Air-Oxygen blender improvises the usage of Resuscitator to a greater extent Humidifier adjusting the level of humidity thus minimizing evaporation andAir-Oxygen blender mixing the right amount of gas

Contdhelliphellip

Humidifier

Air Oxygen Blender

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1048729 Clean external surfaces of the nice 5020 infant resuscitator and Gas Supply line using a damp cloth and mild soapy water or Isopropyl Alcohol

1048729 Dry all surfaces after cleaning with a clean soft cloth or paper towel

1048729 Clean the Bag Resuscitator and any similar equipment in strict accordance with the Manufacturers instructions

1048729 All other components may be disinfected using an approved chemical disinfectant solution

1048729 When all components are completely dry assemble the system again

Preventive Maintenance

1048729 The nice 5020 infant resuscitator should require minimal servicing or maintenance when used under normal conditions

1048729 At least once every six months the equipment should be examined by a qualified Service Engineer to ensure safety and operational integrity

1048729 If any part of the equipment is damaged or fails to operate correctly take it out of service immediately and contact a qualified Service Engineer to ensure operational safety

1048729 A Service Manual enabling qualified technical personnel to carry out routine service schedules fault finding and repairs is available from nice Neotech

1048729 Always carry out a functional test after cleaning or maintenance to ensure safety and operational integrity

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to

internal components Do not autoclave or gas sterilize the unit Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Use of nonstandard components Consult the manufacturer for repair and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 8: 5020 hornpuff resuscitator

Contrsquodhellip

Substantial physiologic changes occur in the transition from fetal to extra uterine life includingbullThe role of the placenta in gas exchange is taken over by the lungsbullChanges from fluid-filled to air filled lungsbullDramatic increase in blood flow to the lungs with reversal then closure of intra and extra cardiac shunts

Assistance to breathe at birthPrevent neonatal deathPrevent Adverse longterm neurodevelopmental sequelae

Failure or disruption of these changes may result in further difficulties with resuscitation in the newborn infantThe need for resuscitation of the newborn infant at birth cannot always be anticipated or predicted

Benefit

A range of Baby friendly equipments Introducing nice NeOumltech 5020 HORNPUFF Resuscitator

Designed to perfection for your baby

nice

502

0 H

OR

NP

UF

F R

esus

cita

tor

bullSafe Controlled and Precise Peak Inspiratory Pressure (PIP)bullFlexibility in OperationbullPrecise Positive End Expiratory PressurebullPrecise Pressure DeliverybullFace Mask of Different sizesbullCompact and HandybullMounting Options

Features

How to operate

Gas Supply Flow rate Setting the Maximum PressureSetting the Peak Inspiratory Pressure (PIP) Setting the Positive End Expiratory Pressure (PEEP)Resuscitate with nice 5020 Infant Resuscitator

Gas Supply Flow rate

1048729 Set the Flow rate at 5 LPM in the Flow Meter

Setting the Maximum Pressure

1048729 Occlude PEEP knob and turn PIP control fully clockwise adjust maximum pressure control knob clockwise or counterclockwise to set desired maximum pressure

Setting the Peak Inspiratory Pressure (PIP)

1048729 Occlude the PEEP knob turn PIP control knob counter-clockwise until the desired peak inspiratorypressure is set

Setting the Positive End Expiratory Pressure (PEEP)

1048729 Adjust PEEP knob clockwise or anti-clockwise to the desired PEEP level

Resuscitate with nice 5020 Infant Resuscitator1048729 Adjust gas supply to the desired flow rate

1048729 Fit patient T-piece to neonatal resuscitation mask and place over the babyrsquos mouth andor nose Fitpatient T-piece to the endotracheal tube

1048729 Resuscitate by placing and removing thumb over the PEEP knob to allow inspiration and expiration

How nice Resuscitatorfare over competitors

nice Resuscitator

User-friendly easy to access all parameters

Low consumables price

Compact and handy

Infant Resuscitator ndash nice 5020 Accessories

Nasal Prong

Humidifier Air Oxygen Blender Mounting Bracket

Patient Circuit

The addition of respiratory humidifier and Air-Oxygen blender improvises the usage of Resuscitator to a greater extent Humidifier adjusting the level of humidity thus minimizing evaporation andAir-Oxygen blender mixing the right amount of gas

Contdhelliphellip

Humidifier

Air Oxygen Blender

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1048729 Clean external surfaces of the nice 5020 infant resuscitator and Gas Supply line using a damp cloth and mild soapy water or Isopropyl Alcohol

1048729 Dry all surfaces after cleaning with a clean soft cloth or paper towel

1048729 Clean the Bag Resuscitator and any similar equipment in strict accordance with the Manufacturers instructions

1048729 All other components may be disinfected using an approved chemical disinfectant solution

1048729 When all components are completely dry assemble the system again

Preventive Maintenance

1048729 The nice 5020 infant resuscitator should require minimal servicing or maintenance when used under normal conditions

1048729 At least once every six months the equipment should be examined by a qualified Service Engineer to ensure safety and operational integrity

1048729 If any part of the equipment is damaged or fails to operate correctly take it out of service immediately and contact a qualified Service Engineer to ensure operational safety

1048729 A Service Manual enabling qualified technical personnel to carry out routine service schedules fault finding and repairs is available from nice Neotech

1048729 Always carry out a functional test after cleaning or maintenance to ensure safety and operational integrity

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to

internal components Do not autoclave or gas sterilize the unit Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Use of nonstandard components Consult the manufacturer for repair and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 9: 5020 hornpuff resuscitator

A range of Baby friendly equipments Introducing nice NeOumltech 5020 HORNPUFF Resuscitator

Designed to perfection for your baby

nice

502

0 H

OR

NP

UF

F R

esus

cita

tor

bullSafe Controlled and Precise Peak Inspiratory Pressure (PIP)bullFlexibility in OperationbullPrecise Positive End Expiratory PressurebullPrecise Pressure DeliverybullFace Mask of Different sizesbullCompact and HandybullMounting Options

Features

How to operate

Gas Supply Flow rate Setting the Maximum PressureSetting the Peak Inspiratory Pressure (PIP) Setting the Positive End Expiratory Pressure (PEEP)Resuscitate with nice 5020 Infant Resuscitator

Gas Supply Flow rate

1048729 Set the Flow rate at 5 LPM in the Flow Meter

Setting the Maximum Pressure

1048729 Occlude PEEP knob and turn PIP control fully clockwise adjust maximum pressure control knob clockwise or counterclockwise to set desired maximum pressure

Setting the Peak Inspiratory Pressure (PIP)

1048729 Occlude the PEEP knob turn PIP control knob counter-clockwise until the desired peak inspiratorypressure is set

Setting the Positive End Expiratory Pressure (PEEP)

1048729 Adjust PEEP knob clockwise or anti-clockwise to the desired PEEP level

Resuscitate with nice 5020 Infant Resuscitator1048729 Adjust gas supply to the desired flow rate

1048729 Fit patient T-piece to neonatal resuscitation mask and place over the babyrsquos mouth andor nose Fitpatient T-piece to the endotracheal tube

1048729 Resuscitate by placing and removing thumb over the PEEP knob to allow inspiration and expiration

How nice Resuscitatorfare over competitors

nice Resuscitator

User-friendly easy to access all parameters

Low consumables price

Compact and handy

Infant Resuscitator ndash nice 5020 Accessories

Nasal Prong

Humidifier Air Oxygen Blender Mounting Bracket

Patient Circuit

The addition of respiratory humidifier and Air-Oxygen blender improvises the usage of Resuscitator to a greater extent Humidifier adjusting the level of humidity thus minimizing evaporation andAir-Oxygen blender mixing the right amount of gas

Contdhelliphellip

Humidifier

Air Oxygen Blender

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1048729 Clean external surfaces of the nice 5020 infant resuscitator and Gas Supply line using a damp cloth and mild soapy water or Isopropyl Alcohol

1048729 Dry all surfaces after cleaning with a clean soft cloth or paper towel

1048729 Clean the Bag Resuscitator and any similar equipment in strict accordance with the Manufacturers instructions

1048729 All other components may be disinfected using an approved chemical disinfectant solution

1048729 When all components are completely dry assemble the system again

Preventive Maintenance

1048729 The nice 5020 infant resuscitator should require minimal servicing or maintenance when used under normal conditions

1048729 At least once every six months the equipment should be examined by a qualified Service Engineer to ensure safety and operational integrity

1048729 If any part of the equipment is damaged or fails to operate correctly take it out of service immediately and contact a qualified Service Engineer to ensure operational safety

1048729 A Service Manual enabling qualified technical personnel to carry out routine service schedules fault finding and repairs is available from nice Neotech

1048729 Always carry out a functional test after cleaning or maintenance to ensure safety and operational integrity

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to

internal components Do not autoclave or gas sterilize the unit Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Use of nonstandard components Consult the manufacturer for repair and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 10: 5020 hornpuff resuscitator

nice

502

0 H

OR

NP

UF

F R

esus

cita

tor

bullSafe Controlled and Precise Peak Inspiratory Pressure (PIP)bullFlexibility in OperationbullPrecise Positive End Expiratory PressurebullPrecise Pressure DeliverybullFace Mask of Different sizesbullCompact and HandybullMounting Options

Features

How to operate

Gas Supply Flow rate Setting the Maximum PressureSetting the Peak Inspiratory Pressure (PIP) Setting the Positive End Expiratory Pressure (PEEP)Resuscitate with nice 5020 Infant Resuscitator

Gas Supply Flow rate

1048729 Set the Flow rate at 5 LPM in the Flow Meter

Setting the Maximum Pressure

1048729 Occlude PEEP knob and turn PIP control fully clockwise adjust maximum pressure control knob clockwise or counterclockwise to set desired maximum pressure

Setting the Peak Inspiratory Pressure (PIP)

1048729 Occlude the PEEP knob turn PIP control knob counter-clockwise until the desired peak inspiratorypressure is set

Setting the Positive End Expiratory Pressure (PEEP)

1048729 Adjust PEEP knob clockwise or anti-clockwise to the desired PEEP level

Resuscitate with nice 5020 Infant Resuscitator1048729 Adjust gas supply to the desired flow rate

1048729 Fit patient T-piece to neonatal resuscitation mask and place over the babyrsquos mouth andor nose Fitpatient T-piece to the endotracheal tube

1048729 Resuscitate by placing and removing thumb over the PEEP knob to allow inspiration and expiration

How nice Resuscitatorfare over competitors

nice Resuscitator

User-friendly easy to access all parameters

Low consumables price

Compact and handy

Infant Resuscitator ndash nice 5020 Accessories

Nasal Prong

Humidifier Air Oxygen Blender Mounting Bracket

Patient Circuit

The addition of respiratory humidifier and Air-Oxygen blender improvises the usage of Resuscitator to a greater extent Humidifier adjusting the level of humidity thus minimizing evaporation andAir-Oxygen blender mixing the right amount of gas

Contdhelliphellip

Humidifier

Air Oxygen Blender

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1048729 Clean external surfaces of the nice 5020 infant resuscitator and Gas Supply line using a damp cloth and mild soapy water or Isopropyl Alcohol

1048729 Dry all surfaces after cleaning with a clean soft cloth or paper towel

1048729 Clean the Bag Resuscitator and any similar equipment in strict accordance with the Manufacturers instructions

1048729 All other components may be disinfected using an approved chemical disinfectant solution

1048729 When all components are completely dry assemble the system again

Preventive Maintenance

1048729 The nice 5020 infant resuscitator should require minimal servicing or maintenance when used under normal conditions

1048729 At least once every six months the equipment should be examined by a qualified Service Engineer to ensure safety and operational integrity

1048729 If any part of the equipment is damaged or fails to operate correctly take it out of service immediately and contact a qualified Service Engineer to ensure operational safety

1048729 A Service Manual enabling qualified technical personnel to carry out routine service schedules fault finding and repairs is available from nice Neotech

1048729 Always carry out a functional test after cleaning or maintenance to ensure safety and operational integrity

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to

internal components Do not autoclave or gas sterilize the unit Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Use of nonstandard components Consult the manufacturer for repair and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 11: 5020 hornpuff resuscitator

bullSafe Controlled and Precise Peak Inspiratory Pressure (PIP)bullFlexibility in OperationbullPrecise Positive End Expiratory PressurebullPrecise Pressure DeliverybullFace Mask of Different sizesbullCompact and HandybullMounting Options

Features

How to operate

Gas Supply Flow rate Setting the Maximum PressureSetting the Peak Inspiratory Pressure (PIP) Setting the Positive End Expiratory Pressure (PEEP)Resuscitate with nice 5020 Infant Resuscitator

Gas Supply Flow rate

1048729 Set the Flow rate at 5 LPM in the Flow Meter

Setting the Maximum Pressure

1048729 Occlude PEEP knob and turn PIP control fully clockwise adjust maximum pressure control knob clockwise or counterclockwise to set desired maximum pressure

Setting the Peak Inspiratory Pressure (PIP)

1048729 Occlude the PEEP knob turn PIP control knob counter-clockwise until the desired peak inspiratorypressure is set

Setting the Positive End Expiratory Pressure (PEEP)

1048729 Adjust PEEP knob clockwise or anti-clockwise to the desired PEEP level

Resuscitate with nice 5020 Infant Resuscitator1048729 Adjust gas supply to the desired flow rate

1048729 Fit patient T-piece to neonatal resuscitation mask and place over the babyrsquos mouth andor nose Fitpatient T-piece to the endotracheal tube

1048729 Resuscitate by placing and removing thumb over the PEEP knob to allow inspiration and expiration

How nice Resuscitatorfare over competitors

nice Resuscitator

User-friendly easy to access all parameters

Low consumables price

Compact and handy

Infant Resuscitator ndash nice 5020 Accessories

Nasal Prong

Humidifier Air Oxygen Blender Mounting Bracket

Patient Circuit

The addition of respiratory humidifier and Air-Oxygen blender improvises the usage of Resuscitator to a greater extent Humidifier adjusting the level of humidity thus minimizing evaporation andAir-Oxygen blender mixing the right amount of gas

Contdhelliphellip

Humidifier

Air Oxygen Blender

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1048729 Clean external surfaces of the nice 5020 infant resuscitator and Gas Supply line using a damp cloth and mild soapy water or Isopropyl Alcohol

1048729 Dry all surfaces after cleaning with a clean soft cloth or paper towel

1048729 Clean the Bag Resuscitator and any similar equipment in strict accordance with the Manufacturers instructions

1048729 All other components may be disinfected using an approved chemical disinfectant solution

1048729 When all components are completely dry assemble the system again

Preventive Maintenance

1048729 The nice 5020 infant resuscitator should require minimal servicing or maintenance when used under normal conditions

1048729 At least once every six months the equipment should be examined by a qualified Service Engineer to ensure safety and operational integrity

1048729 If any part of the equipment is damaged or fails to operate correctly take it out of service immediately and contact a qualified Service Engineer to ensure operational safety

1048729 A Service Manual enabling qualified technical personnel to carry out routine service schedules fault finding and repairs is available from nice Neotech

1048729 Always carry out a functional test after cleaning or maintenance to ensure safety and operational integrity

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to

internal components Do not autoclave or gas sterilize the unit Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Use of nonstandard components Consult the manufacturer for repair and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 12: 5020 hornpuff resuscitator

How to operate

Gas Supply Flow rate Setting the Maximum PressureSetting the Peak Inspiratory Pressure (PIP) Setting the Positive End Expiratory Pressure (PEEP)Resuscitate with nice 5020 Infant Resuscitator

Gas Supply Flow rate

1048729 Set the Flow rate at 5 LPM in the Flow Meter

Setting the Maximum Pressure

1048729 Occlude PEEP knob and turn PIP control fully clockwise adjust maximum pressure control knob clockwise or counterclockwise to set desired maximum pressure

Setting the Peak Inspiratory Pressure (PIP)

1048729 Occlude the PEEP knob turn PIP control knob counter-clockwise until the desired peak inspiratorypressure is set

Setting the Positive End Expiratory Pressure (PEEP)

1048729 Adjust PEEP knob clockwise or anti-clockwise to the desired PEEP level

Resuscitate with nice 5020 Infant Resuscitator1048729 Adjust gas supply to the desired flow rate

1048729 Fit patient T-piece to neonatal resuscitation mask and place over the babyrsquos mouth andor nose Fitpatient T-piece to the endotracheal tube

1048729 Resuscitate by placing and removing thumb over the PEEP knob to allow inspiration and expiration

How nice Resuscitatorfare over competitors

nice Resuscitator

User-friendly easy to access all parameters

Low consumables price

Compact and handy

Infant Resuscitator ndash nice 5020 Accessories

Nasal Prong

Humidifier Air Oxygen Blender Mounting Bracket

Patient Circuit

The addition of respiratory humidifier and Air-Oxygen blender improvises the usage of Resuscitator to a greater extent Humidifier adjusting the level of humidity thus minimizing evaporation andAir-Oxygen blender mixing the right amount of gas

Contdhelliphellip

Humidifier

Air Oxygen Blender

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1048729 Clean external surfaces of the nice 5020 infant resuscitator and Gas Supply line using a damp cloth and mild soapy water or Isopropyl Alcohol

1048729 Dry all surfaces after cleaning with a clean soft cloth or paper towel

1048729 Clean the Bag Resuscitator and any similar equipment in strict accordance with the Manufacturers instructions

1048729 All other components may be disinfected using an approved chemical disinfectant solution

1048729 When all components are completely dry assemble the system again

Preventive Maintenance

1048729 The nice 5020 infant resuscitator should require minimal servicing or maintenance when used under normal conditions

1048729 At least once every six months the equipment should be examined by a qualified Service Engineer to ensure safety and operational integrity

1048729 If any part of the equipment is damaged or fails to operate correctly take it out of service immediately and contact a qualified Service Engineer to ensure operational safety

1048729 A Service Manual enabling qualified technical personnel to carry out routine service schedules fault finding and repairs is available from nice Neotech

1048729 Always carry out a functional test after cleaning or maintenance to ensure safety and operational integrity

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to

internal components Do not autoclave or gas sterilize the unit Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Use of nonstandard components Consult the manufacturer for repair and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 13: 5020 hornpuff resuscitator

Gas Supply Flow rate

1048729 Set the Flow rate at 5 LPM in the Flow Meter

Setting the Maximum Pressure

1048729 Occlude PEEP knob and turn PIP control fully clockwise adjust maximum pressure control knob clockwise or counterclockwise to set desired maximum pressure

Setting the Peak Inspiratory Pressure (PIP)

1048729 Occlude the PEEP knob turn PIP control knob counter-clockwise until the desired peak inspiratorypressure is set

Setting the Positive End Expiratory Pressure (PEEP)

1048729 Adjust PEEP knob clockwise or anti-clockwise to the desired PEEP level

Resuscitate with nice 5020 Infant Resuscitator1048729 Adjust gas supply to the desired flow rate

1048729 Fit patient T-piece to neonatal resuscitation mask and place over the babyrsquos mouth andor nose Fitpatient T-piece to the endotracheal tube

1048729 Resuscitate by placing and removing thumb over the PEEP knob to allow inspiration and expiration

How nice Resuscitatorfare over competitors

nice Resuscitator

User-friendly easy to access all parameters

Low consumables price

Compact and handy

Infant Resuscitator ndash nice 5020 Accessories

Nasal Prong

Humidifier Air Oxygen Blender Mounting Bracket

Patient Circuit

The addition of respiratory humidifier and Air-Oxygen blender improvises the usage of Resuscitator to a greater extent Humidifier adjusting the level of humidity thus minimizing evaporation andAir-Oxygen blender mixing the right amount of gas

Contdhelliphellip

Humidifier

Air Oxygen Blender

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1048729 Clean external surfaces of the nice 5020 infant resuscitator and Gas Supply line using a damp cloth and mild soapy water or Isopropyl Alcohol

1048729 Dry all surfaces after cleaning with a clean soft cloth or paper towel

1048729 Clean the Bag Resuscitator and any similar equipment in strict accordance with the Manufacturers instructions

1048729 All other components may be disinfected using an approved chemical disinfectant solution

1048729 When all components are completely dry assemble the system again

Preventive Maintenance

1048729 The nice 5020 infant resuscitator should require minimal servicing or maintenance when used under normal conditions

1048729 At least once every six months the equipment should be examined by a qualified Service Engineer to ensure safety and operational integrity

1048729 If any part of the equipment is damaged or fails to operate correctly take it out of service immediately and contact a qualified Service Engineer to ensure operational safety

1048729 A Service Manual enabling qualified technical personnel to carry out routine service schedules fault finding and repairs is available from nice Neotech

1048729 Always carry out a functional test after cleaning or maintenance to ensure safety and operational integrity

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to

internal components Do not autoclave or gas sterilize the unit Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Use of nonstandard components Consult the manufacturer for repair and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 14: 5020 hornpuff resuscitator

Setting the Positive End Expiratory Pressure (PEEP)

1048729 Adjust PEEP knob clockwise or anti-clockwise to the desired PEEP level

Resuscitate with nice 5020 Infant Resuscitator1048729 Adjust gas supply to the desired flow rate

1048729 Fit patient T-piece to neonatal resuscitation mask and place over the babyrsquos mouth andor nose Fitpatient T-piece to the endotracheal tube

1048729 Resuscitate by placing and removing thumb over the PEEP knob to allow inspiration and expiration

How nice Resuscitatorfare over competitors

nice Resuscitator

User-friendly easy to access all parameters

Low consumables price

Compact and handy

Infant Resuscitator ndash nice 5020 Accessories

Nasal Prong

Humidifier Air Oxygen Blender Mounting Bracket

Patient Circuit

The addition of respiratory humidifier and Air-Oxygen blender improvises the usage of Resuscitator to a greater extent Humidifier adjusting the level of humidity thus minimizing evaporation andAir-Oxygen blender mixing the right amount of gas

Contdhelliphellip

Humidifier

Air Oxygen Blender

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1048729 Clean external surfaces of the nice 5020 infant resuscitator and Gas Supply line using a damp cloth and mild soapy water or Isopropyl Alcohol

1048729 Dry all surfaces after cleaning with a clean soft cloth or paper towel

1048729 Clean the Bag Resuscitator and any similar equipment in strict accordance with the Manufacturers instructions

1048729 All other components may be disinfected using an approved chemical disinfectant solution

1048729 When all components are completely dry assemble the system again

Preventive Maintenance

1048729 The nice 5020 infant resuscitator should require minimal servicing or maintenance when used under normal conditions

1048729 At least once every six months the equipment should be examined by a qualified Service Engineer to ensure safety and operational integrity

1048729 If any part of the equipment is damaged or fails to operate correctly take it out of service immediately and contact a qualified Service Engineer to ensure operational safety

1048729 A Service Manual enabling qualified technical personnel to carry out routine service schedules fault finding and repairs is available from nice Neotech

1048729 Always carry out a functional test after cleaning or maintenance to ensure safety and operational integrity

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to

internal components Do not autoclave or gas sterilize the unit Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Use of nonstandard components Consult the manufacturer for repair and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 15: 5020 hornpuff resuscitator

How nice Resuscitatorfare over competitors

nice Resuscitator

User-friendly easy to access all parameters

Low consumables price

Compact and handy

Infant Resuscitator ndash nice 5020 Accessories

Nasal Prong

Humidifier Air Oxygen Blender Mounting Bracket

Patient Circuit

The addition of respiratory humidifier and Air-Oxygen blender improvises the usage of Resuscitator to a greater extent Humidifier adjusting the level of humidity thus minimizing evaporation andAir-Oxygen blender mixing the right amount of gas

Contdhelliphellip

Humidifier

Air Oxygen Blender

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1048729 Clean external surfaces of the nice 5020 infant resuscitator and Gas Supply line using a damp cloth and mild soapy water or Isopropyl Alcohol

1048729 Dry all surfaces after cleaning with a clean soft cloth or paper towel

1048729 Clean the Bag Resuscitator and any similar equipment in strict accordance with the Manufacturers instructions

1048729 All other components may be disinfected using an approved chemical disinfectant solution

1048729 When all components are completely dry assemble the system again

Preventive Maintenance

1048729 The nice 5020 infant resuscitator should require minimal servicing or maintenance when used under normal conditions

1048729 At least once every six months the equipment should be examined by a qualified Service Engineer to ensure safety and operational integrity

1048729 If any part of the equipment is damaged or fails to operate correctly take it out of service immediately and contact a qualified Service Engineer to ensure operational safety

1048729 A Service Manual enabling qualified technical personnel to carry out routine service schedules fault finding and repairs is available from nice Neotech

1048729 Always carry out a functional test after cleaning or maintenance to ensure safety and operational integrity

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to

internal components Do not autoclave or gas sterilize the unit Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Use of nonstandard components Consult the manufacturer for repair and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 16: 5020 hornpuff resuscitator

Infant Resuscitator ndash nice 5020 Accessories

Nasal Prong

Humidifier Air Oxygen Blender Mounting Bracket

Patient Circuit

The addition of respiratory humidifier and Air-Oxygen blender improvises the usage of Resuscitator to a greater extent Humidifier adjusting the level of humidity thus minimizing evaporation andAir-Oxygen blender mixing the right amount of gas

Contdhelliphellip

Humidifier

Air Oxygen Blender

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1048729 Clean external surfaces of the nice 5020 infant resuscitator and Gas Supply line using a damp cloth and mild soapy water or Isopropyl Alcohol

1048729 Dry all surfaces after cleaning with a clean soft cloth or paper towel

1048729 Clean the Bag Resuscitator and any similar equipment in strict accordance with the Manufacturers instructions

1048729 All other components may be disinfected using an approved chemical disinfectant solution

1048729 When all components are completely dry assemble the system again

Preventive Maintenance

1048729 The nice 5020 infant resuscitator should require minimal servicing or maintenance when used under normal conditions

1048729 At least once every six months the equipment should be examined by a qualified Service Engineer to ensure safety and operational integrity

1048729 If any part of the equipment is damaged or fails to operate correctly take it out of service immediately and contact a qualified Service Engineer to ensure operational safety

1048729 A Service Manual enabling qualified technical personnel to carry out routine service schedules fault finding and repairs is available from nice Neotech

1048729 Always carry out a functional test after cleaning or maintenance to ensure safety and operational integrity

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to

internal components Do not autoclave or gas sterilize the unit Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Use of nonstandard components Consult the manufacturer for repair and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 17: 5020 hornpuff resuscitator

The addition of respiratory humidifier and Air-Oxygen blender improvises the usage of Resuscitator to a greater extent Humidifier adjusting the level of humidity thus minimizing evaporation andAir-Oxygen blender mixing the right amount of gas

Contdhelliphellip

Humidifier

Air Oxygen Blender

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1048729 Clean external surfaces of the nice 5020 infant resuscitator and Gas Supply line using a damp cloth and mild soapy water or Isopropyl Alcohol

1048729 Dry all surfaces after cleaning with a clean soft cloth or paper towel

1048729 Clean the Bag Resuscitator and any similar equipment in strict accordance with the Manufacturers instructions

1048729 All other components may be disinfected using an approved chemical disinfectant solution

1048729 When all components are completely dry assemble the system again

Preventive Maintenance

1048729 The nice 5020 infant resuscitator should require minimal servicing or maintenance when used under normal conditions

1048729 At least once every six months the equipment should be examined by a qualified Service Engineer to ensure safety and operational integrity

1048729 If any part of the equipment is damaged or fails to operate correctly take it out of service immediately and contact a qualified Service Engineer to ensure operational safety

1048729 A Service Manual enabling qualified technical personnel to carry out routine service schedules fault finding and repairs is available from nice Neotech

1048729 Always carry out a functional test after cleaning or maintenance to ensure safety and operational integrity

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to

internal components Do not autoclave or gas sterilize the unit Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Use of nonstandard components Consult the manufacturer for repair and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 18: 5020 hornpuff resuscitator

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1048729 Clean external surfaces of the nice 5020 infant resuscitator and Gas Supply line using a damp cloth and mild soapy water or Isopropyl Alcohol

1048729 Dry all surfaces after cleaning with a clean soft cloth or paper towel

1048729 Clean the Bag Resuscitator and any similar equipment in strict accordance with the Manufacturers instructions

1048729 All other components may be disinfected using an approved chemical disinfectant solution

1048729 When all components are completely dry assemble the system again

Preventive Maintenance

1048729 The nice 5020 infant resuscitator should require minimal servicing or maintenance when used under normal conditions

1048729 At least once every six months the equipment should be examined by a qualified Service Engineer to ensure safety and operational integrity

1048729 If any part of the equipment is damaged or fails to operate correctly take it out of service immediately and contact a qualified Service Engineer to ensure operational safety

1048729 A Service Manual enabling qualified technical personnel to carry out routine service schedules fault finding and repairs is available from nice Neotech

1048729 Always carry out a functional test after cleaning or maintenance to ensure safety and operational integrity

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to

internal components Do not autoclave or gas sterilize the unit Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Use of nonstandard components Consult the manufacturer for repair and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 19: 5020 hornpuff resuscitator

Cleaning Procedure

1048729 Clean external surfaces of the nice 5020 infant resuscitator and Gas Supply line using a damp cloth and mild soapy water or Isopropyl Alcohol

1048729 Dry all surfaces after cleaning with a clean soft cloth or paper towel

1048729 Clean the Bag Resuscitator and any similar equipment in strict accordance with the Manufacturers instructions

1048729 All other components may be disinfected using an approved chemical disinfectant solution

1048729 When all components are completely dry assemble the system again

Preventive Maintenance

1048729 The nice 5020 infant resuscitator should require minimal servicing or maintenance when used under normal conditions

1048729 At least once every six months the equipment should be examined by a qualified Service Engineer to ensure safety and operational integrity

1048729 If any part of the equipment is damaged or fails to operate correctly take it out of service immediately and contact a qualified Service Engineer to ensure operational safety

1048729 A Service Manual enabling qualified technical personnel to carry out routine service schedules fault finding and repairs is available from nice Neotech

1048729 Always carry out a functional test after cleaning or maintenance to ensure safety and operational integrity

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to

internal components Do not autoclave or gas sterilize the unit Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Use of nonstandard components Consult the manufacturer for repair and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 20: 5020 hornpuff resuscitator

Preventive Maintenance

1048729 The nice 5020 infant resuscitator should require minimal servicing or maintenance when used under normal conditions

1048729 At least once every six months the equipment should be examined by a qualified Service Engineer to ensure safety and operational integrity

1048729 If any part of the equipment is damaged or fails to operate correctly take it out of service immediately and contact a qualified Service Engineer to ensure operational safety

1048729 A Service Manual enabling qualified technical personnel to carry out routine service schedules fault finding and repairs is available from nice Neotech

1048729 Always carry out a functional test after cleaning or maintenance to ensure safety and operational integrity

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to

internal components Do not autoclave or gas sterilize the unit Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Use of nonstandard components Consult the manufacturer for repair and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 21: 5020 hornpuff resuscitator

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to

internal components Do not autoclave or gas sterilize the unit Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Use of nonstandard components Consult the manufacturer for repair and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 22: 5020 hornpuff resuscitator

Certifications and Approvals

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us
Page 23: 5020 hornpuff resuscitator

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Slide 6
  • Why Infant Resuscitators are Used
  • Contrsquodhellip
  • Slide 9
  • Slide 10
  • Slide 11
  • How to operate
  • Slide 13
  • Slide 14
  • How nice Resuscitator fare over competitors
  • Infant Resuscitator ndash nice 5020 Accessories
  • Slide 17
  • Slide 18
  • Cleaning Procedure
  • Preventive Maintenance
  • Slide 21
  • Certifications and Approvals
  • Contact Us