Upload
wisbiomed
View
352
Download
1
Embed Size (px)
Citation preview
Biosensing Instrument Inc.
Installation and Training Tutorial
BI 2000 SPR Instrument
Please read the manual before
viewing this quick tutorial
Tutorial Sections
• Installation
• Operation
• Experimental Tips
• Clean Up
• Example Experiments
Home Section
Installation
• Unpacking
• Instrument Setup
• Software Installation
Home Section
Unpacking
• Check that the instrument was not damaged during shipping.
• Do not obstruct the ventilation slots.
• Place the instrument away from breezeways such as windows, door, AC vents, and major traffic areas.
Home Section
Instrument Setup
• Plug the power cords into the back of the SPR instrument and syringe pump
• Connect the SPR instrument to the computer via the USB communication cable
Home Section
Software Installation
• If a computer system was not purchased from BI, you will need to install the software yourself.
• Located the Data Acquisition program on the BI Software disk and install it first.
• Next install the BI SPR software package located on the disk.
• Simply, follow the onscreen instructions to complete the software installation process.
Home Section
BI Software Package
• BI- Control
• BI- Data Analysis
• BI- Kinetics Analysis
For more details, please review the BI Manual
Home Section
BI Control - Plotting Features
Measuring Tool Double-click on axis to
adjust rangesTake notes
More plot options in
drop-down menu
Quickly zoom out
Baseline correction
tool
Start data collection
Valve information
Plot Labels
Double-click on plot
name to change plot
options
Select any area to
quickly zoom in the
data
Export to BI- Data
Analysis
Real-time label-free
data acquisiton
Home Section
BI Data AnalysisMore plot options in
drop-down menu
Make local or global
notes
Data Selection Tool
Referenced Data
Selection Tool
Add multiple date
files to workplace
Overlay multiple
plots
Remove plots
Select name to
isolate data
Export to BI-KA
Selected region for
analysis or export
Home Section
BI Kinetics Analysis
Calculate Kinetics
Overlay simulation
and experimental
data Residuals Calculate Affinity
Reference
subtraction
Crop injections
Extracted kinetic
constants
Align baselines
Align injections
Home Section
Operation
• Data Acquisition
• Calibration
• Flow Cell
• Valves and Syringes
• Preparation
Home Section
Data Acquisition
• Data acquisition property options are found in the setup drop down menu.
• Higher gain settings increase sensitivity, but decrease detection range. – The Gain can be set : 1, 10, 100, and 1000.
– 10 is the default value
• Higher sapling rates have faster time resolution, but noisier data and larger files.– 10 is the default value
Home Section
Calibration
• Calibration property options are found in the setup drop down menu.
• Accurate calibration of both channels critical in obtaining accurate data and high quality background and reference subtraction.
• System calibration should be checked every several months.
• A calibration example is found later in this tutorial.
The potential may be
inverted for EC-SPR
applications
A drifting signal is a good indicator of poor calibration.
Home Section
Flow Cell (FC)• Open Position
– Lift the FC holder, rotate CCW, then gently lower.
• Closed Position– Lift the FC holder, rotate back over the detection area, then gently
lower the FC holder.
– Gently press down on the FC holder to ensure contact.
Home Section
Valves
Mode Select
InjectionChannel Select
Home Section
Mode Select ValveUse this valve to select single or serial flow modes.
– In the single flow mode (CCW): the injected sample only passes through one channel.
– In the serial flow mode(CW): the injected sample passes through both channels serially.
Home Section
CH1
CH2
CH1
CH2
Single Flow Mode Serial Flow Mode
Channel Select Valve
Use this valve to select Channel 1 or Channel 2 operation.
– Turn the handle CCW to select Channel 1.
– Turn the handle CW to select Channel2.
Home Section
CH1
CH2
Channel 1 Channel 2
CH1
CH2
Injection Valve• Use this valve to Load and Inject
samples.1. Turn the handle CCW to load samples
through the injection port.
2. Turn the handle CW to release the sample to the sensor.
3. Turn the handle CCW to stop the sample release and reload another samples.
• The default loop size is 100uL, and may be changed.
• Use the Injection Timer to time the injections
Home Section
Loading Injection Syringes
• Load the injection syringe with an additional 20-25 uL of sample for operational waste. – At least 10 uL of waste-sample should precede the test-sample to
flush out prior residue
– At least 10 uL of waste-sample should remain in the injection syringe to avoid injecting air bubbles.
• Wipe dry the injection syringe needle before inserting it into the injection valve for loading.
Home Section
Preparation• Warm Up
• Carrier Solution
• Syringe Pump
• Carrier Pre-Flow
• Sensor Chips
• Pre-Experiment Check
• Experimental Tips
Home Section
Warm up
• Turn on the instrument and laser
• Allow at least 15 minutes for warm up
Home Section
Carrier Solution Preparation
Filter and degas the carrier solutions prior to use
Home Section
Syringe filters are a
convenient way to
filter solutions.
Solutions may be degassed
with a low-pressure vacuum
for 10-15 min with gentle
intermittent shaking.
Loading Carrier Solution
• Load the filtered degassed carrier solution into the two provided carrier syringes
• Tap out any large air pockets that may have trapped in the syringes
• Connect the syringe fittings onto the syringes
Home Section
Watch for air pockets
Loading the Syringe Pump Loading the Pump
1. Press the Drive Button to slide the Drive Bar back
2. Lift and rotate the Syringe Retainer Bar away
3. Load the two syringes into the Syringe Cradle. Make sure the collars of the syringes are flush with the Syringe Cradle.
4. Secure the syringes with the Retainer Bar.
5. Press the Drive Button to slide the Drive Bar against the syringe plungers
Home Section
Programming the Pump
• Set the syringe diameter– Scroll through the table menu and select Beckman Plastic, 10cc
– A diameter of 14.48mm should automatically load
• Set a syringe volume for automatic stop, otherwise input zero.
• Flow rates should be less 150ul/min on BI-2000 models.
Home Section
Carrier Pre-FlowPre-Flow the carrier solution before loading the chip to remove residual air pockets and equilibrate the fluidic lines.
• Place the flow cell and holder in the closed position (either on a chip or the bare prism),
• Switch to single flow mode
• Run the pump at a high flow rate, ~150 uL/min, for 3-5 minutes.
• Stop the pump, and wait ~2 minutes for residual pressure to pass.
Home Section
Sensor Chips
• Prism Stage Cleaning
• Flow Cell Cleaning
• Removing Chip from Container
• Chip Loading
• Closing the Flow Cell
Home Section
Prism Stage Cleaning
• Lift and rotate the Flow Cell Holder away from the detection stage.
• Clean off the detection stage using lens paper moistened with ethanol.
• Make sure the detection stage is free of residue and debris.
Home Section
Flow Cell Cleaning• Remove the FC from the holder and wipe clean the
gasket with a paper towel moistened with ethanol.
• N2 blow dry the gasket, making sure the FC is free of residue and debris
• Remount the FC into the FC holder, make sure that the FC sits flush with the back of the holder
Home Section
Removing Chip from Container
• Remove the seal from the container of SPR chips.
• Remove the lid and retainer ring.
• Carefully remove a sensor chip with a clean pair of tweezers.
Home Section
Loading Chip• Place a 3ul drop of Matching Fluid on the center of the prism stage (~3mm wide).
• With tweezers, slowly place the chip (Gold side up) on the prism stage.
• Lower the chip at an angle to avoid trapping air bubbles.
• Use a soft tip to re-center the chip on the prism stage.
Note: Be careful not to scratch the prism surface . A tooth pick or pipette tip works well as a soft point
Home Section
Which Face is Gold Coated?Color Test:• Hold the chip with a pair of tweezers
• Look at the reflections of both sides.
• The gold coated side is more golden or shinny, whereas the glass side will show a duller yellow or bronze tint.
Scratch Test (if still not sure):• Gently scratch one of the chip’s corners.
• Scratch marks appear on the gold-coated side only
Home Section
Close the Flow Cell• Lift and rotate the FC holder over the sensor chip,
and gently lower the FC holder.
• Gently press down on the FC holder to confirm the contact.
Home Section
Pre-Experiment Checks
• Interface Quality
• Baseline Quality
• Resonance Quality
Home Section
Interface Quality• Turn on the liquid laser.
• Observe that the optical reflection in the viewing window has uniform intensity.
• If an intensity spot or dark streak is observed, an air bubble may be trapped at the sensor-prism interface. – Try pressing on one of the chips’ corners with a soft point.
– A tooth pick or pipette tip works well as a soft point
Home Section
Baseline Quality• In serial flow mode, start the pump at a high flow
rate, ~150ul/min.
• Let flow for ~3 min, then reduce to ~60um/min
• Wait for the baseline to stabilize, ~10 min depending upon previous use and clean up.
Home Section
Resonance Quality
• Examine the SPR reflection in the viewing window.
• Two nearly identical SPR dark lines ~2mm wide should be clearly visible and neighboring each other.– If they are not similar, you may have trapped an air bubble. (see
maintenance section)
• SPR experiments are ready to begin.
Home Section
Experimental Tips
Obtaining high quality binding information requirescareful consideration of the following factors :
• Instrument Calibration
• Buffer Chemistry
• Surface Coverage
• Sample Preparation
• Flow Rate and Sample Volume
Home Section
Quick Calibration Check
Accurate calibration of both channels is critical in obtaining accurate data and high quality background and reference subtraction. • Use a brand new bare gold chip.
• Run DI water as the carrier solution.
• In Serial mode, inject 1% ethanol solution (diluted in DI water).
• Adjust the calibration factors so that a 60.0 mDeg response is observed in both channels.
System calibration should be checked every several months.
Home Section
Buffer Chemistry
• Buffer pH’s much greater or lower than the sample’s isoelectric point, may result in strong charge attraction or repulsion of the sample with the surface.
• Chose a buffer concentration that is high enough to sustain protein activity but low enough to prevent saturation of detectors.
• It may be helpful to include low concentrations of buffer additives/surfactants to reduce non-specific absorption.
• Degas and filter the buffer before use
Home Section
Surface Coverage• Surface Coverage densities too low will result in too small of
an analyte binding signal and densities too high which result in too demanding of a surface leading to mass transport limited kinetics.
• Mass transport limited kinetics means that the surface is demanding more analyte molecules than is currently being supplied by the flow.
• The BI-KA Software can compensate for mass limited transport effects, but a sufficient amount of kinetic curvature should still be present for a more accurate analysis.
Home Section
Sample Preparation
• To minimize the effects of bulk refractive index change, dilute samples with the carrier solution.
• Degas samples and let them warm to room temperature before use
• Start with low concentrations near that of the expected affinity constant.
Home Section
Flow Rate and Sample Volume
• Flow rates too low result in mass limited transport kinetics.
• Flow rates too high do not allow enough exposure time for binding information to be extracted.
• Note that higher flow rates consume more carrier solution and sample volume per time, thus larger sample volumes are required at higher flow rates for equal exposure times.
Home Section
Clean Up
Cleaning up after experiments is of critical importance in maintaining the high performance of the instrument.
• Valves and Tubing
• Flow Cell
• Prism Stage
Home Section
Clean Up – Valves and Tubing
1. Clean the carrier and injection syringes with DI water
2. While in inject-single position, load the carrier syringes with DI water and let flow through the system at high flow rate (~150uL/min) for a several minutes.
3. Toggle the valves several times to flush away possible internal valve debris.
4. Loosely insert injection syringe into injection port, and inject DI water so that solution escapes from port opening.
5. Load the carrier syringes with air and manually pushing the air through the system to flush out all the liquid
Home Section
Clean Up – Flow Cell
1. Lift the cell holder and rotating it such that it remains elevated.
2. Slide the Flow Cell out of the Flow Cell Holder and wipe it clean with an ethanol moistened paper towel.
3. N2 blow dry the FC, and return it to the Flow Cell Holder.
Home Section
Clean Up - Prism Stage
1. Use a soft point (cotton swab or pipette tip) to nudge most of the chip off the prism edge.
2. Use a pair of tweezers to gently lift away the chip.
3. Wipe the remaining RIM liquid off the prism with lens paper.
4. Moisten lens paper with ethanol for final wipes.
5. Gently return FC to bare Prism stage.
Home Section
Example Experiments
• Reference Subtraction
• Calibration
• BSA/ anti-BSA Experiment
Home Section
Reference Subtraction Example
Home Section
Reference Selection tool
Solid outline
Dashed outline
Align BaselineAlign injection timeReference Subtracted Injection
Calibration Example• Turn on instrument and laser, allow ~15min for instrument to warm up.
• Load filtered-degassed DI water into carrier syringes
• Allow several minutes for DI water pre-flow.
• Open the FC and wipe clean the FC
• Load a brand new chip bare gold chip
• Close FC, start pump at high flow rate
• Wait ~5min for baseline to stabilize with fluid
• In serial flow mode, iject 1% ethanol into the system
• Look for 60mDeg response change in both channels.
• Adjust the calibration values for both channels so that a 1% ethanol injection results in 60mDeg change.
• Repeat until the system is accurately calibrated.
Home Section
BSA Experiment Example• Turn on instrument and laser, allow ~15min for instrument to warm up.
• Load filtered-degassed buffer into carrier syringes
• Allow several minutes for DI water pre-flow.
• Open the FC and wipe clean the FC
• Load a modified chip onto the prism stage
• Close, start pump at high flow rate, wait ~5min for baseline to stabilize with buffer
• Begin acquiring data
• In serial flow ~30uL/min, inject NHS/EDC
• Switch to Ch1 single flow mode and inject BSA
• Switch to serial flow mode and inject EA blocker
• Inject regeneration solution to further clean the chip surface
• Increase flow rate to 100 uL/min and set the injection timer
• Inject Antibody.
• Inject regeneration solution twice
• Repeat antibody Injection or inject a higher concentration of antibody
• Export to BI-DA and BI-KA for analysis
Home Section