7/26/2019 L14 Ultrasound(1)
1/27
BME 340: Bioimaging
Lecture 14: Introduction to Ultrasound
Characteristics of ultrasound and its interactions with matter
1
7/26/2019 L14 Ultrasound(1)
2/27
Ultrasound
Ultrasound imaging is the class of techniques that form images on thebasis of the reflection of sound waves from structure boundaries.
A pulse-echo acquisition (like radar).
Frequency range for medical imaging is 2-10 MHz.
An imaging modality that is genuinely wavelength limited with respect
to spatial resolution.
2
7/26/2019 L14 Ultrasound(1)
3/27
3
7/26/2019 L14 Ultrasound(1)
4/27
Sound Propagation
Propagation of mechanical energy through a continuous
elastic medium bycompressionand subsequent expansionofthe medium itself.
As with EM radiation, wavelength is given as:
= c / f
with c typically given as m/s.
Sound velocity varies widely among materials and is a
function of the bulk modulus (B, a measure of elasticity and
resistance to compression) and thedensity() of the
medium:
Units are kg/(ms2) for B and kg/m3for .
Bc =
4
7/26/2019 L14 Ultrasound(1)
5/27
Tissue Velocities
5
7/26/2019 L14 Ultrasound(1)
6/27
Sound Propagation
Typical wavelengths in tissue:
f = 2MHz in soft tissue, = 0.77mm.
f = 10MHz is soft tissue, = 0.15mm.
f = 5MHz is soft tissue, = 0.31mm.
f = 5MHz in fat, = 0.29mm (5.8% from
soft tissue).
Wavelength determines the achievable spatial
resolution along the direction of the beam.
Higher frequency has a shorter wavelength.
6
7/26/2019 L14 Ultrasound(1)
7/27
Sound Propagation
Penetrabilityof the sound energy varies with frequency.
In general, lower frequency beams will penetrate further.
Increasing frequency reduces penetrability.
For abdominal imaging, f = 3.5 5 MHz. For thyroid, breast, etc., f = 7.5 10 MHz.
Scanners generally supplied with a range of transducers to
cover several applications.
7
7/26/2019 L14 Ultrasound(1)
8/27
Sound Propagation Theconstructiveanddeconstructive interference phenomena are very
important in shaping and steering the ultrasound beam.
8
7/26/2019 L14 Ultrasound(1)
9/27
Pressure
Sound energy represents local variations inpressurewithin the propagating medium.
Pressure amplitude defined as the peak maximum or
minimum of the pressure waveform referenced to the
average value. SI unit is the Pascal (Pa) = 1 N/m2.
1 atmosphere = 100kPa.
Typical ultrasound beams in diagnostic imaging exceed 1
MPa in peak pressure.
9
7/26/2019 L14 Ultrasound(1)
10/27
Intensity
Intensity is defined as thepower(energy/time) per unit area
and is proportional to the square ofpressure amplitude:
I P2
Typical unit is mW / cm2.
Intensity varies with peak pressure, and operating mode. Relative values expressed as dB:
Relative Intensity (dB) = 10 log (I2 / I1)
Relative Pressure (dB) = 20 log (P2 / P1)
Typical return echoes can be 60dB reduced in intensity over thetransmitted pulse.
10
7/26/2019 L14 Ultrasound(1)
11/27
11
7/26/2019 L14 Ultrasound(1)
12/27
Example: Calculate the remaining intensity of a 100 mWultrasound pulse that loses 30 dB while traveling through
tissue.
12
7/26/2019 L14 Ultrasound(1)
13/27
Ultrasound Interactions in Matter
A propagating sound wave can undergo reflection, refraction,
absorption, and scatter. Reflectionoccurs at boundaries where there is a difference in
acoustic impedance with the reflected energy traveling back
towards the source. This is the basis for image formation.
Refractionis the change in direction resulting from non-perpendicular incidence.
Scattering results from both reflection and refraction from small
structures and produces the characteristic background texture seen
in ultrasound images. Absorption is the conversion of sound energy to heat.
Attenuationis defined as the loss of intensity due to scatter and
absorption.
13
7/26/2019 L14 Ultrasound(1)
14/27
Acoustic Impedance Acoustic impedance (Z) is defined as: Z = c
Where c is the speed of sound in m/s and is density in kg/m3.
The unit of Z is kg / (m2s), and is known as a rayl. Acoustic impedance
can be thought of as a measure of compressibility.
Sound waves will reflect from boundaries representing a mismatch in
acoustic impedance.
14
7/26/2019 L14 Ultrasound(1)
15/27
Reflection and Refraction
15
7/26/2019 L14 Ultrasound(1)
16/27
Reflection Reflection of sound energy from a boundary is a function of the
impedance mismatch at the boundary separating two materials:
Where RPis thereflection pressure coefficient, Prand Piare the
reflected and incident pressures respectively, Z1and Z2are theacoustic impedances.
The intensity reflection coefficient is given as:
Following conservation of energy, the transmitted intensity is
given as: TI= 1 RI
12
12
ZZ
ZZ
P
PR
i
r
P
+
==
2
12
12
+
== ZZ
ZZ
I
I
Ri
r
I
16
7/26/2019 L14 Ultrasound(1)
17/27
17
7/26/2019 L14 Ultrasound(1)
18/27
Reflection
For perpendicular incidence on a boundary, the reflectedenergy returns to the transducer as an echo (time delay).
Where is much longer than structural variations in the
boundary, a smooth boundary is rendered (assumed in all
previous discussions). Reducing increases scatter off variations in the boundary
and a smaller return to the source.
For non-perpendicular incidence, reflection takes place at the
same angle away from the transducer resulting in loss ofsignal.
18
7/26/2019 L14 Ultrasound(1)
19/27
Where c2>c1,total reflection can occur when the angle of incidence
is greater than the critical angle, defined as
Refraction This is the change in direction of transmitted sound energy upon a non-
perpendicular encounter with a boundary. Angles of incidence, reflection, and transmission are measured relative
to the normal to the boundary.
The index of refraction is determined by the difference in the speed of
sound across the boundary.
192
1
arcsin c
c
c =
7/26/2019 L14 Ultrasound(1)
20/27
Scattering
Acoustic scattering results from objects on the order of a
wavelength in size (representing a rough surface).
Some organs have a characteristic surface structure that gives
rise to a characteristic scatter signature (useful diagnostically).
Rough reflectors reflect sound over a range of angles, thus
causing loss in amplitude of received echoes.
20
7/26/2019 L14 Ultrasound(1)
21/27
Scattering Intensity of return signal from rough reflectors has less of a
dependence beam direction than that from smooth reflectors.
Along a rough boundary, differences in scatter amplitude show
up as differences in image intensity.
Factors determining echo amplitude:
Number of scatter structures/volume
Size of scatter structures
Boundary Z difference
Signal frequency
All else being the same, scatter increases with frequency. In US lingo, Hyperechoic= high scatter amplitude,
Hypoechoic= low scatter amplitude, relative to average
background signal, useful in characterizing anatomy.
21
7/26/2019 L14 Ultrasound(1)
22/27
Attenuation
Attenuation results from loss of acoustic energy (converted to
heat) and scattering.
Attenuation coefficient () in units of dB/cm.
is approximately linear with frequency (i.e. doubling the
frequency doubles the attenuation).
General value for soft tissue is 0.5dB/cm/MHz.
Tissue half-value thickness (HVT) corresponds to thickness
for 3dB drop in intensity (or 6dB drop in pressure amplitude).
22
7/26/2019 L14 Ultrasound(1)
23/27
Attenuation
23
7/26/2019 L14 Ultrasound(1)
24/27
24
7/26/2019 L14 Ultrasound(1)
25/27
Attenuation
For medical imaging purposes, ultrasound detectors musthave a dynamic range on the order of 120-140dB in
pressure amplitude (factor of 107).
Deeper penetration requires lower frequencies.
Frequency dependence of transmission results in depth-dependent (low pass) filtering of broadband pulses.
25
7/26/2019 L14 Ultrasound(1)
26/27
Attenuation
26
Example: Calculate the approximate intensity HVT in soft
tissue for ultrasound beams of 2 and 10 MHz. Determine the
number of HVTs the incident beam and the echo travel at a 6-
cm depth.
7/26/2019 L14 Ultrasound(1)
27/27
27
Review
What is the typical range of ultrasound frequency used formedical imaging?
Does longer wavelength imply better ultrasound penetrability?
The wavelength of a 2 MHz ultrasound beam is ________ mm
in soft tissue. (In soft tissue c = 1540 m/sec).
What is the constructive/deconstructive interference?
Calculate the remaining intensity of a 10 mW ultrasound pulse
that loses 60 dB while traveling through tissue.
What intensity fraction of an ultrasound beam is reflected from
an interface between two media with Z (acoustic impedance)values of 1.65 and 1.55?