IRB Space Choice
Considerations
• Patient access– Most walking, some wheelchair– Access for ambulance stretcher– No line of sight to bench work (?)
• Equipment space– 11’ min room width– Room for tube to be completely out of reach of
patient during set-up
Patient access- 2nd floor
Patient access- 1st floor
Room Layout- 2nd floor
Patient/Tech Space
Scanner
Scanner
Patient/Tech Space
Scanner Electronics
Rack
Scanner Electronics
Rack
120V Double Duplex
Data Ports (4)D4
D2 Data Ports (2)
D2
D2 D4
Grounding Lug (to existing Cu strap)
208V, 1f, 30A; L6-30R
208V, 3f, 50A; Disconnect
6” Leaded Cable Pass-
Through
7’H Partition wall, ¼” Pb-lined
~3’x3’ window, ¼” Pb-equivalent
Built-in sitting-height counter
Room Layout- 1st floor