Introduction
This guide is intended to improve an undergraduates understanding of how an MRI machine
functions. It contains the very basics of NMR imaging. The information herein is compiled from
various sources (listed below). It has not been reviewed by any person. It is not intended as a study
aide, merely to help a student get to grips with the basic concepts which are used in NMR imaging.
Every atomic nucleus possesses angular momentum - it spins around an axis which transects its
centre. Every atomic nucleus is magnetically dipolar, meaning that it possesses both a positive
magnetic pole at one end of its axis and a negative magnetic pole at the opposite end of its axis.
When an atom is placed in external magnetic field, such as is produced by the earth, the atom tends
to align with the direction of the magnetic field which transects it (like the needle of a compass
does). When the atoms axis is not exactly in line with the external magnetic field, it starts to rotatearound the axis (which is in line with the direction) of the magnetic field which transects it. This
phenomenon is called precession. The greater the deviation of the atoms axis from the direction of
the magnetic field, the greater the circumference of the imaginary circle created by the precession
is, reaching a maximum when the axis of the atom is perpendicular to the direction of the external
magnetic field. The precession is also invisible when the axis of the atom points in the opposite
direction of the magnetic field. These concepts are used to create MRI images. Protons (hydrogen
atoms) are used in MRI imaging because of
An MRI machine consists of a number of coils. Electrical currents are applied to these currents to
create magnetic fields. Each individual coil has a different configuration and orientation in the MRI
machine. Each of these coils serves a unique function.
The largest coil (and the most expensive one) is used to create an extremely homogeneous magnetic
field in the bore. The purpose of this coil is to create a very strong field which will cancel the effects
of other unwanted external magnetic fields which may disturb the (close to) perfect alignment of all
the protons which is fundamental to the functioning of NMR imaging. This homogeneous external
magnetic field is a vector, represented by the symbol B. This field causes all the protons to align
with the magnetic field, either in the direction of the field (termed spin up) or in the exact
opposite direction (termed spin down). Protons in the spin up orientation have a low level of
potential energy in comparison with protons in the spin down orientation which have a high level of
potential energy. At room temperature the proportion of protons in these two orientations are
almost equal, with a very slight preponderance to the spin up state. The ratio of spin up protons to
spin down protons is called the Boltzmann factor and is determined by the strength of B and
temperature. The proportion of excess spin up protons (termed spin excess) is very small, but the
absolute number is massive and this spin excess produces the net signal that is measured during
NMR imaging (the signals of the other protons cancel each other out). If the strength of the
magnetic field increases, the Boltzmann factor increases, causing a greater spin excess and
consequently a greater signal, increasing the signal-to-noise ratio or SNR (see below). An increase in
temperature decreases the Boltzmann factor (the protons absorb the energy and deviate toward thehigher energy state of down spin), reducing spin excess and consequently the SNR. The vector of the
8/3/2019 Introduction to MRI Physics for the Undergraduate Med Student by the Undergraduate Med Student
2/6
magnetic field produced by the spin excess (which corresponds to the direction/axis of the angular
momentum) is represented by the symbol M.
The precession frequency of an atom is also known as the Larmor frequency or resonance frequency
and is represented by the symbol . The precession frequency of an atom is a product of the
strength of the magnetic field and the gyromagnetic ratio () of the atom in question.
= B.
Every atom has a unique gyromagnetic ratio. A proton has a very high gyromagnetic ratio which is
an important reason why it is the atom measured in NMR. All protons in an equal B precess at an
equal frequency. If the B is inhomogeneous this means that the field is stronger in some areas than
in others, therefore some protons will precess at a higher frequency and other protons will precess
at a lower frequency. This is the reason why the large NMR coil is so critical, because any significant
external, unwanted, unmeasured interference cannot be entered into the mathematical processesinvolved in NMR imaging and will therefore give inaccurate results.
Another important coil is the oscillating transverse magnetic field coil (OTMF). When an
electromagnetic radiofrequency is applied at a frequency equal to an atoms Larmor frequency,
these frequencies are said to resonate. The proton absorbs the electromagnetic energy and
consequently the protons spin tilts (it shifts to an orientation with a higher energy state, in the
direction of spin down). The angle of the tilt is proportional to the strength and the duration of the
applied electromagnetic pulse. The OTMF is applied for the exact time that it is necessary to tilt the
spin 90 degrees towards the xy-plane (transverse plane; the orientation at which precession causes
the greatest signal or, in other words, is most visible). This pulse ofelectromagnetic energy is
termed a 90 degree pulse. When the pulse is turned off, the vector of the spin gradually returns to
B (the direction of the B; also called the z-axis). This is called T1 relaxation. As the proton relaxes, it
emanates the energy as electromagnetic waves, known as an electromagnetic signal, which is
detected by a detector coil and gives information about the spins. Some of the energy is also
released in the form of heat to its environment (this environment is called the lattice). T1 relaxation
is therefore also known as spin-lattice relaxation. T1 is time it takes for the M to return to B. T1 is
unique to every type of tissue.
The electromagnetic waves emitted by the precessing protons causes a potential difference in the
detector coil (according to the laws of electromagnetic induction) and produces the signal which is
measured in the NMR computer system. As the proton precesses, the direction of its magnetic field
(corresponding with the axis of its angular momentum) also rotates. This causes an oscillating
voltage in the detector coil. When M is aligned in the xy-plane, the oscillating voltage is at its
greatest. As relaxation takes places, this oscillating voltage/signal decreases. This decrease in signal
is termed free induction decay.
After a 90 degree pulse, a certain period of time (called half echo time) elapses in which the protons
relax toward the z-axis and also go out of phase in the xy-plane (due to the magnetic field gradient
and minor inhomogeneity). Before the vector reaches to the z-axis, another electromagnetic pulse isapplied the 180 degree pulse (which is twice as long as a 90 degree pulse). The axes of angular
8/3/2019 Introduction to MRI Physics for the Undergraduate Med Student by the Undergraduate Med Student
3/6
momentum of each proton is displaced to the exact opposite direction and the protons commence
relaxation towards the direction of B. The direction of precession is reversed and at echo time
(which is half echo time after the 180 degree pulse), the axes of angular momentum are once again
in phase and Mis in the transverse plane. The signal is at its greatest again. The protons relax
further until they align with Bagain. The form of the signal forms an echo and this process is
called spin echo.
When spin echoes are performed repeatedly the amplitudes of the echoes are seen to decrease
gradually. This is because the axes of angular momentum go out of phase in the transverse plane for
reasons other than differences in the external magnetic field. This loss of phase is irreversible and is
because of the effect of the magnetic fields caused by the atoms surrounding the protons. Because
the composition of tissues differs, these effects vary and cause differences in Larmor frequencies.
This effect is termed T2-relaxation. T2 is the time it takes for signal amplitude to decrease with36,8%.
The T2 is dependent on the chemical milieu of the protons is thus relatively characteristic of a certain
type of tissue - this permits the identification of tissue on the NMR image.
A third major coil, the magnetic field gradient coil, produces a magnetic field which varies linearly in
three orthogonal directions (along the x-, y- and z-axis).Two solenoids of equal strength are placed at
opposing points of the bore (a pair along each axis). They create equal but opposite magnetic fields
and this creates a gradient in the magnetic field. Because the Larmor frequency of the proton is
proportional to the magnitude of the external magnetic field, the gradient causes a linear variation
in precession frequencies, with the greatest frequency at and the smallest at a wanted and
predictable variation in magnetic field (as opposed to unwanted, unpredictable inhomogeneity). The
Larmor frequency of the proton is thus dependent on its position. This allows the computer to
calculate the position of signal in each of the 3 planes.
The magnetic field gradient coil also corrects minor distubances of B, primarily caused by the
Earths magnetic field. As mentioned earlier, if certain areas of the magnetic field are stronger than
others, the protons precess at different frequencies, causing the vectors of the individual protons
dipoles(axes of angular momentum) to point in different directions, causing the vector of M (the
total magnetization vector) to decrease and consequently a decrease in signal. The free induction
decay occurs more rapidly (because of lower initial signal amplitudes). The magnetic field gradient
coil decreases the effects of external unwanted disturbances this process is called shimming. The
field is more homogeneous, the strength of the signal greater and the free induction decay is longer.
All the signals received by the detector coil are organized into a virtual matrix, called the k-space.
This matrix represents the manner in which the signals are gathered spatially (only in two
dimensions in the axial plane or slices). The process of gathering the signals is called encoding
and three different types of encoding exist and these represent the three gradients which are
applied by the magnetic field gradient coil.
The first encoding gradient is the slice encoding gradient. A gradient is applied along the z-axis (i.e.
along the length of the bore) or B which is in the cephalic direction. Consequently the magnetic
field is greatest at the head of the patient (only slightly stronger than B) and gradually decreasesuntil it is at its weakest at the feet of the patient (only slightly weaker than B). When the
8/3/2019 Introduction to MRI Physics for the Undergraduate Med Student by the Undergraduate Med Student
4/6
electromagnetic pulses are applied, they only resonate with the protons precessing at their specific
frequency. Consequently only spins at that axial level will produce a signal. The matrix represents a
specific slice.
The phase encoding gradient is applied along the y-axis. Phase encoding is only done one row (of
the matrix or k-space) at a time. This gradient is applied for a small period and then switched off. The
result is that protons along the y-axis are out of phase in a predictable pattern. When this gradient is
switched off, the phase difference remains constant. Phase encoding produces the columns of the
k-space.
The frequency encoding gradient is applied along the x-axis. This gradient is applied after the
phase encoding gradient is halted. This causes a further deviation in phasealignment along the x-
axis which is also predictable. Frequency encoding produces the rows of the k-space.
By consolidating the data gathered during the various encodings, the NMR computer is able tocalculate the position of a signal. Fourier transformation (a mathematical process) is applied to the
data which are arranged in the k-space and a two-dimensional image is formed.
Different sequences
T1 weighted: better to view disruption in anatomy
T2 weighted: better to view pathological changes in tissue
STIR (fat-suppression): decreases intensity of adipose tissue, to differentiate it from other
hyperintense tissue, especially in the case of a history of cancer and trauma.
Field of view:
Field of view is the specific volume which is to be detected by the receiver coil. A greater field of
view means more electromagnetic waves to be detected, i.e. a greater signal, but a decrease in
resolution (if the matrix size is constant, because slices are thicker).
SNR
The signal to noise ratio is the ratio between the strength of the signal from the area of interest and
the signal from background electromagnetic activity (termed noise). A low SNR causes images to
appear grainy. The SNR is proportional to the size of the voxel. An increase in field of view, slice
thickness, phase steps and a decrease in matrix size cause an increase in SNR. The inverse applies for
a decrease in SNR. The SNR is proportional to the acquisition time.
Resolution
Resolution determines the sharpness or pixelated appearance of an NMR image. The resolution is
inversely proportional to the voxel. A decrease in the field of view, slice thickness and an increase in
matrix size cause an increase in resolution.
Voxel size
The size of the voxel is equal to the field of view divided by the matrix size. The depth of the voxel is
always the largest dimension, therefore the resolution perpendicular to the image plane is alwaysthe poorest.
8/3/2019 Introduction to MRI Physics for the Undergraduate Med Student by the Undergraduate Med Student
5/6
Acquisition time
Acquisition time is equal to repetition time (TR) multiplied by the number of phase encoding steps
multiplied by the number of averages (NSA or NEX).
Performing an MRI
Initially only an overview image is taken. This image is of poor resolution, but is used to plan the
definitive imaging study. In the imaging application, frames are placed and cropped to fit over the
areas of interest in the sagittal and axial plane. A number of parameters (e.g. field of view, number
of slices, etc.) are entered and the definitive imaging process is commenced. The imaging is repeated
for every sequence requested by the requesting physician. When the image collection is complete,
the radiographer annotates the images and digitally sends them to the radiologists job list as well as
the archive.
The duration of an MRI study is dependent on the anatomical area of interest, the number of
sequences requested, whether contrast is used and whether a dynamic study is performed.
Depending on the anatomical area of interest, different receiver coils are used during NMR imaging.
There are a number of special coils: brain coil, neck coil, torso coil, cardiac coil, breast coil, knee coil,
wrist coil and spinal coil. Some of these coils are place over the area of interest, e.g. the brain coil;
and some rest beneath the area of interest, e.g. the torso coil.
Special anaesthetic equipment, ECG machines, pulse oximeters, capnography machines,
defibrillators etc. are used in the MRI room. This is necessary because standard equipment will be
strongly attracted by the bore, causing objects to move rapidly and adhere to the bore. This causes a
great hazard for persons in the MRI room and may damage the bore.
Switching the large electromagnetic coil on is not a simple, quick or cheap procedure. It make take
days from initiation for the MRI machine to be fully functional. Therefore the MRI machine is not
switched of for any folly. In the event of an emergency necessitating the deactivation of the bore
(such as a large magnetic object which cannot otherwise be removed or an object impinging a
person), a procedure called quenching is performed.
8/3/2019 Introduction to MRI Physics for the Undergraduate Med Student by the Undergraduate Med Student
6/6
Sources
1. The knowledge and experience of the employees of Schnetler, Corbett and Partners2. Introductory NMR and MRI with Paul Callaghan (video series)3. www.mritutor.org4. Introduction to MRI Physics (www.simplyphysics.com)5. MRI Physics (www.hsc.csu.edu)6. MRI Physics (www.mri-physics.net)7. The Basic Principles of MRI (www.spinwarp.ucsd.edu)8. Basic Principles of MRI by Wm. Faulkner (www.e-radiography)
http://www.mritutor.org/http://www.mritutor.org/http://www.simplyphysics.com/http://www.simplyphysics.com/http://www.simplyphysics.com/http://www.hsc.csu.edu/http://www.hsc.csu.edu/http://www.hsc.csu.edu/http://www.mri-physics.net/http://www.mri-physics.net/http://www.mri-physics.net/http://www.spinwarp.ucsd.edu/http://www.spinwarp.ucsd.edu/http://www.spinwarp.ucsd.edu/http://www.spinwarp.ucsd.edu/http://www.mri-physics.net/http://www.hsc.csu.edu/http://www.simplyphysics.com/http://www.mritutor.org/