Operating Manual
Serial No.
Micro Centrifuge
Purchase Date
Place of purchase
Micro 12
www.ihanil.com2/20
Copyright ⓒ 2017 Hanil Scientific Inc. All rights reserved.
Contact Us
If you have any questions, contact Hanil Scientific Inc. or place of purchase.
[email protected] / [email protected] 3
CONTENTS
2 . Product description --------------------------6 2.1 Appearance -6
3 . Installation -------------------------------------8 3.2 Installation -8
4 .
Operation
--------------------------------------12
4.1 Controller
-12 4.2
Setting RPM
-13
4.3
Setting run time -13
4.4
Starting/End the centrifugaion -13
4.5
Pulse function -13
5 .
Maintenance ----------------------------------14
6.
Troubleshooting ---------------------------15
7. Rotor
and
accessories ----------------------17
3.1 Packing inspection -8
3.3 Connect power -9
1. General Considerations -----------------------4
1.1 Safety -41.2 Transport & Storage -41.3 Safety labe -51.4 Electric safety information -5
2.2 Delivery package -6 2.3 Specifications -7
3.4 Opening/closing the lid -93.5 Loading and removal of the rotor -103.6 Loading tubes -11
5.1 Care instructions -14 5.2 Cleaning -14
6-1 Possible problems -15,16
4 www.ihanil.com
1. General Considerations
1. The centrifuge should be installed on flat surface to maintain level.
2. Check the voltage to be used, before connecting the centrifuge to the power source.
4. Do not exceed the maximum rated speed of the rotor or buckets or accessories in use.
6. If the centrifuge is contaminated by toxic or radioactive samples or infectious blood samples, remove contaminants completely
7. Substances that may generate volatile or explosive vapor can not be centrifuged.
8. The balancing work of samples shoud be done in advance before operation.
11. Only centrifuge with rotor and rotor lid firmly tightened.
12. Do not block vents.
Do not insert foreign objects into the holes of the device.
14. Repairs must be carried out by a technologist authorized by Hanil scientific Inc .
15. According to IEC61010-2-020 maintain a 30cm “clearance envelope” around the centrifuge while the rotor is spinning.
16. Please contact the place of purchase or Hanil Scientific Inc. ([email protected]/[email protected])
for
product
repairs.
1.2 Transport & Storage
1.1 Safety
Ambient temperature 5℃ ~ 35℃Maximum relative humidity 30% ~ 85%
Air pressure 500 ~ 1060 hpa
Follow precautions and all the safety requirements described on this user manual to prevent any damages and failure of equipment and loss of lives.
Abnormal vibration is caused when the device is installed on an inclined surface.
3. Only use rotors, parts, and accessories provided by Hanil Scientific Inc.
Hanil
Scientific
Inc.
is
not
responsible
for
damages of the device and accidents caused by using parts and accessories not recommended.
5. Make sure to prepare necessary safety measures before using samples that are toxic or radioactive samples or pathogenic or samples or infectious blood.
and take needful actions such as ventilation or isolation of centrifuge.
9. Before operation, rotor and chamber should be dry.
10. Do not attempt to slow or stop the spinning rotor by hand.
13.
Storage TransportAmbient temperature -10℃ ~ 40℃Maximum relative humidity 10% ~ 90%
Air pressure 500 ~ 1060 hpa
[email protected] / [email protected] 5
1. General Considerations
1.3 Safety label attached to products
1.4 Electric safety information
Electric shock
1. Use a power cord only provided with the device.
2. Use sockets with a protective earth conductor and suitable power cord.
3. Do not use an extension cord.
4. Do not place anything on the power cable
5. Do not block vents.
6. If you have the following emergencies, shut off the power supply and unplug the power cord from outlet
and
Do not open lid until rotor stops completely.
Rotation direction of the rotor.
contact your place of purchase or Hani l Scientific Inc.
- Unusual noises or smell from the equipment.- Damage or wear of a power cord.- Breakdown of circuit breaker, fuse or safety device.- If you spill liquid on the equipment.- If the equipment is exposed to water.- If Some of equipment has been damaged.
6 www.ihanil.com
2. Product description
2.1 Appearance
① Lid : Lid protects the inside of the chamber.②
Rotor
locking
screw:
Connect
the
rotor
to
the
motor shaft.
③
Rotor :
Connected
to
the
motor
shaft
for
centrifugal
separation
④Lid
sensor
:
Detect
whether
lid
is
cloesd
⑤
Lid
open
switch: Press
the
button
to
open
to
the
lid.
⑥Controller
:
Keys
and
knobs
for
operating
the
centrifuge.
⑦Power
socket
:
Connection
for
power
cord.
⑧Power
switch
:
Switch
for
switching
the
centrifuge
on/off.
⑨Absorbent
rubber
:
Fix
the
device
on
the
laboratory
bench.
①Standard offering
-
- Micro12 main body (Rotor included : A2.0-12)
- Power cable (300/500V, 6A)
- User manual
②Optional products [see 7. Rotor and accessories]
-
PCR
strip
rotor
(A0.2-32)
①
②
⑥ ⑤ ⑦
⑧ ⑨
③④
④
2.2 Delivery package
[email protected] / [email protected] 7
2. Product description
2.3 Specifcations
* RPM, RCF value is for A2.0-12.
* This device does not provide automatic recognition of the rotor. So do not exceed 6,000 rpm when using the PCR rotor.
Max. RPM 13,500 rpm
Max. RCF 12,032 xg
Time ≤ 15 min, Pulse
Acc. time to max. speed ≤18 sec
Dec. time from max. speed ≤8 sec
Tube Capacity 1.5/2.0ml x 12
Power supply AC110/220V, 50/60Hz
Dimension (W x D x H, mm) 190 x 202 x 129
Weight (kg) 1.5
※ RCF Table
* Rcf value was calculated based on the bottom of the rotor hole.
RPMA2.0-12
RPMA0.2-32
0.2ml 0.5ml 1.5/2.0ml 0.2 ml Inner line0.2 ml
Outer line
3,000 443 545 594 3,000 313 ~ 443 464 ~ 560 6,000 1,231 1,515 1,651 5,000 870 ~ 1,231 1,289 ~ 1,5568,000 3,151 3,877 4,226
6,000[ Max.]
1,253 ~ 1,773
1,857 ~ 2,241 10,000 4,924 6,058 6,603 12,000 7,090 8,724 9,508 13,500 8,973 11,042 12,032
8 www.ihanil.com
3. Installation
3.1 Packing inspection
3.2 Installation
▶ Check the delivery for completeness.
· Check packing conditions very carefully before unpacking. · Contact Hanil Scientific immediately if damages found.· Check the delivery for completeness.· You can get contact details on packing box or back of the manual.
▶ Check all parts for damage.
Installation on hard and flat ground.• Centrifuge should be installed on hard and flat place.• If the centrifuge is installed in an inclined place, the shaft may be bent due to the weight of the rotor.
Good ventailation.• For air circulation and safety, maintain a 30cm“clearance envelope”around the centrifuge while the rotor is spinning.
Constant temperature/humidity• Centrifuge equipped with the sensitive electronic software which is fragile with humidity and temperature.• Must avoid direct ray or heater and be put in the ambience of controlled temp. and air.
Avoid the corrosive gas• Install the centrifuge in a place where corrosive gas is not generated.• Sulfur dioxide gas and chlorine gas may cause corrosion.
Leveling• The shaft should be put exactly vertical on the hotizontaliy flat ground by the leveling tool.
3. Installation
3.3 Connect power
3.4 Opening/ Closing the lid
· Connect the device to voltage sources which correspond to the electrical requirements on
When closing the lid, close it completely until it clicks.
(O) (X)
· Use sockets with a protective earth conductor and provided power cord.
the label attached to the device.
1. Connect the power cable to the power socket on the right side of the product and plug the cord into the outlet.
2. Switch the centrifuge on using the power switch on the right side of the device.
▶ It is normal product to hear a beep when power is connected.
the lid-1. Press the 'lid open' switch.-Opening -Closing the lid-
1. Put the lid on and press the edge of the lid
If the lid is not completely closed, the device will not operate.
10 www.ihanil.com
3. Installation
3.5 Loading and removal of the rotor
Motor Hub(Motor Shaft)
· This device is equipped with a rotor .· Only use rotors, parts, and accessories provided by Hanil Scientific Inc . · Do not use a defective rotor .
1 .
Using
a
flat
head
screwdriver,
turn
the
rotor
locking
screw
counterclockwise
and
pull
it
completely
out
with
the
washer
under
the
screw.
2.
Lift
up
the
rotor
from
the
motor
shaft .
3.Load
the
rotor
into
the
motor
shaft .
4.
Align
the
washer
with
the
hole
at
the
top
of
the
rotor .
5.
Insert
the
rotor
lock
screw
into
the
screw
hole
in
the
center
of
the
rotor
and
turn
it
clockwise
using
a
flat
head
screwdriver
to
completely
lock
it
6.
Check
that
the
rotor
is
completely
locked .
Flat head screwdriver
Rotor lock screw
Washer
Rotor
Follow the steps below to remove or replace the rotor.
[email protected] / [email protected] 11
3. Installaion
3.6 Loading tubes
Caution
Wrong arrangement
· Only use tubes provided or approved by Hanil Scientific Inc . · Always use the same type of tube .· Tubes should be loaded symmetrically .· Do not exceed the maximum rated speed of the tube . · Same volume of sample should be put on tubes . · Check symmetric loading by balncing tubes with scales .
1. Check the inside of the rotor hole for moisture or foreign matter.
2. Tubes must be loaded symmetrically.
Correct arrangement
12 www.ihanil.com
4.1 Operation panel
4. Operation
버튼 설명
① RPM setting knob
② Time setting knobRotate clockwise or counterclockwise to set the run time.(To use pulse function, set the time to '0' and press the start key.)
③ Start key
④ Stop key
※ When using a A0.2-32(PCR strip rotor) do not exceed 6,000rpm.
Rotate clockwise or counterclockwise to set the rpm value.
Start key
Stop key
[email protected] / [email protected] 13
4.2 Setting RPM
4.3 Setting run time
4.4 Starting/End the centrifugation
4.5 Pulse function
4. Operation
Rotate the 'RPM setting knob' clockwise or counterclockwise to set the rpm value. ▶When using a A0.2-32 (PCR strip rotor) do not exceed 6,000 rpm.
Rotate 'Time setting knob' clockwise or counterclockwise to set the run time.
▶0~15 min.
1. Check the all parameters are correct and the lid is closed.2. Press the 'Start' key (Operation begins with a beep.)
▶ If the lid is not completely closed, it will not operate. ▶ When the set time is over, the device is decelerated and stops with beep sound.▶ Press 'Stop' key to abort centrifugation.▶ Set the 'Rpm setting knob' or 'Time setting knob' to '0' to abort centrifugation. ▶ When the lid is opened during operation, it decelerates and stops.
1. Set the 'Time setting knob' to '0'2. Press the Start key.3. The device decelerates and stops when the set rpm is reached.
4.2 Setting RPM
14 www.ihanil.com
5.Maintenance
5.1 Care instructions
• The following procedures should be performed regulary.1. Regularly inspect the rotor chamber for check the motor shaft is nomal.2. Rotate the shaft with your hand to make sure it turns smoothly.3. Use the stopwatch to check that the time setting is correct
5.2 Cleaning
• Oustside of the device
1. Clean the outside of equipment with a soft and dry cloth. ▶ If the device is contaminated, use a mild cleaning fluid to clean.2. Do not use aggressive chemicals on the device such as alchol, benzene, acetone or phenol. 3. Use the stopwatch to check that the time setting is correct.4. Make sure do not scratch the surface of equipment when cleaning it. ▶ Do not use a metal sponge. ▶ If the device is rusted, remove it with a mild detergent and wipe it with a dry cloth.
• Chamber
1. If the rotor chamber is not dry, wipe moisture from the chamber with a dry cloth 2. If the rotor chamber is dirty, wipe moisture from the chamber with a dry cloth
• Rotor
1. To prevent corrosion, take out the rotor from the rotor chamber.2. If any sample is spilt inside the rotor, wash and dry the rotor well
• Disposal
In case of product is to be disposed of, the local wastes laws and regulations ard to be observed.
[email protected] / [email protected] 15
증상 확인 사항
전원이 들어오지 않아요. [3.3 전원연결]을 참조하여 전원플러그가 빠져 있는지 확인해주십시오.
동작되지 않아요.Lid가 닫혀지지 않을 경우 동작되지 않습니다. [3.4 Lid 열기]를 참조하여 lid상태를 확인하고 lid를 잘 닫아주십시오.
Lid가 열리지 않아요.Lid 개방 스위치를 눌러 lid를 열 수 있습니다. Lid 개방 스위치가 파손되었다면 라벡스 기술지원팀으로 문의하여 주십시오.
Lid가 닫히지 않아요.Lid 걸쇠 부위가 파손되었는 지 확인하십시오. 파손되었다면 라벡스 기술지원팀으로 문의해 주십시오.
작동 중에 진동과 소음이 나타나요.
본체 설치 위치가 불안정하면 본체의 수평과 고정여부를 확인하고, 평평한 곳에 수평을 맞춰 다시 설치하십시오.
로터 장착상태가 불량한 경우 로터 탈착 후 로터 외관을 확인하고 파손된 부위가 있다면 로터 사용을 즉시 중지하십시오.또한 장착 방법이 잘못되었다면 [3.5 로터 장착과 분리]를 참조하여 로터를 정확하게 장착하십시오.
튜브 삽입이 비대칭이거나 무게가 맞지 않는 경우에는 [3.6 튜브 장착]을 참조하여 튜브 무게를 확인하고 대칭적으로 삽입하십시오.
6.Troubleshooting
6.1 Possible problems
Please check the following information, if the device is faulty.
16 www.ihanil.com
Lid sensor replacement may be required.
Need to repair the dial or knob.
Main body or lid repair may be required.
6. Troubleshooting
6.2 Possible problems
1. The device does not operate.2. The device stops during opertaion.3. There is no beep sound when power is connected.
Need to repair the controller.Please contact your dealer.
1. It operates when the lid is open.2. When press the 'start' key, only beeps .
1. 'Rpm/Time knob' dose not rotate.2. 'Start/stop' key does not press.
1. The lid does not close.2. The lid does not open.
Problem Recommended Action
[email protected] / [email protected] 17
7. Rotor and accessories
12 x 1.5/2.0 ml 45o
Hole diameter (mm) : 11.5Max. height for tube fit (mm) : 48
Fixed Angle Rotor, A2.0-12
Tube
Tube capacity 0.2 ml 0.5 ml 1.5/2.0 ml
Adaptor None
Cat. No. -Adaptor bore (Фx L, mm) 6 x 21 8 x 37 -
Radius (mm) 44 59 59Max. RPM 13,500Max. RCF 8,965 12,032
Tube
Tube capacity (ml) 0.2 8-tube PCR strip
Radius (mm) Inner line : 31.13 ~ 44.06Outer line : 46.13 ~ 55.68Max. RPM 6,000Max. RCF (g-force) Inner line : 1,773 / Outer line : 2,241
4 x 8-tube PCR strips, 32 x 0.2 ml 45o
Hole diameter (mm) : 6.5
PCR Rotor, A0.2-32
TR0.2 TR0.5
12,032
Hanil Scientific Inc.16 Arayukro, Gimpo 10136, Rep. of KOREAT. [email protected]