Transcript
Page 1: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

The Field-Flow Fractionation Platform

Field-Flow Fractionation of Macromolecules and Structures That Cannot be Characterized

by Conventional GPC/SEC Techniquesy

Trevor Havard, Evelin Moldenhaur, Soheyl Tadjiki (Postnova)

Sebastien Perrier *, Sylvain Catrouillet* (Warwick and Monash)

Ad d Ch t i ti f P t i Bi l d N ti l

www.postnova.com .

Advanced Characterization of Proteins, Biopolymers and Nanoparticles

Page 2: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

FFF Separation Range

Comparison Chromatography vs. Field-Flow Fractionation

www.postnova.com .www.postnova.com . 22.10.2015 2

Page 3: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

FFF Principle

Separation Mechanism

• Separation in a narrow ribbon-like channel• Laminar flow inside the channel• External field perpendicular to the solvent flow

www.postnova.com .www.postnova.com . 22.10.2015 3

• External field perpendicular to the solvent flow

Page 4: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

FFF Principle

Separation Mechanism

• Centrifugal Field• Flow Field• Thermal Field• Gravitational Field

www.postnova.com .www.postnova.com . 22.10.2015 4

Gravitational Field

Page 5: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

Asymmetric Flow FFF - Principle

Step 1• Sample is injected directly into the

flow stream• A second pump provides focus flow• Cross flow is achieved by a precise

syringe pump• Constant flow to the detectors, no valves

Step 2• Sample is focused to narrow band• Improved resolution and sample

washing• Enrichment of very low concentrations

Step 3 • Focus flow stops and main pump elutes

the sample• Variable cross flow • Detector flow stays constant

www.postnova.com . 22.10.2015 5

Page 6: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

Asymmetric Flow FFFSmart Stream Splitting

• Major portion of flow in the channel is containing no sample• Minor portions of the flow is carrying the sample above the membrane• Minor portions of the flow is carrying the sample above the membrane• Splitting of those two stream via 2 outlets at the channel end• Only the sample containing sub-stream is guided to the detector• Sample dilution is reduced at channel outlet

www.postnova.com .www.postnova.com . 22.10.2015 6

• Increase of sensitivity by factor 5 or more depending on conditions

Page 7: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

Smart Stream SplittingHDL Lipoprotein Preparation for Mass Spectroscopy

Enrichment of HDL by a factor of 2 using Smart Stream Splitting

www.postnova.com .www.postnova.com . 22.10.2015 7

Page 8: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

FFF versus SEC

Reversed elution order in FFF compared to SEC

Additional information fromFFF!

M Co-elution of large

SEC

Log

gmaterial due tointeraction withstationary phase

SEC

FFF

www.postnova.com . 22.10.2015 8

Elution Volume

Page 9: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

How Low Can a Postnova AF4 go!!

16000

18000 12,384 g/mole

12000

14000

16000

(A

U)

Aprotinin Cytochrome C Bromophenol Blue669.7 g/mole

8000

10000

bsor

banc

e (

6511 4 / l

4000

6000

UV

ab 6511.4 g/mole

0 2 4 6 8 100

2000

www.postnova.com . 10/22/2015 9

Retention time (min)

Page 10: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

Flexibility of the Cross-Flow Gradient“Tailor-made” separation in AF4 realized by cross-flow-adjustment

• Extension of Cross-Flow causes better separation

• Selective enhancement of separation

• In SEC not possible!In SEC not possible!→ Column change is time

consuming and expensive→ Calibration of column

determines separationdetermines separation

www.postnova.com . 22.10.2015 10

Mixture of narrow PS: 300, 1200 & 8000 kg/mol

Page 11: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

Flexibility of the Cross-Flow Gradient

1

2

ow [m

L/m

in]

• Gradient of any

Polydisperse PE

0 ,8 10 7

0 1 2 3 4 5 6 7 8 9 10 11 1 2 1 3 140

Cro

ss-F

loE lu tion Volum e [m L]

yshape can be used

• Adjustment of

0 ,5

0 ,6

0 ,7

0 ,8

10 5

10 6

10

V] ol]

jseparation according to special requirements of

0 ,2

0 ,3

0 ,4

10 3

10 4

IR-S

igna

l [m

Mol

ar M

ass

[g/m

oqthe sample

0 1 2 3 4 5 6 7 8 9 10 11 1 2 1 3 14-0 ,1

0 ,0

0 ,1

10 1

10 2

M

www.postnova.com . 11

Elution Volum e [m L]

22.10.2015

Page 12: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

System Information

► Injector

► Separator

PN5300 Auto Injector

AF2000 MF Flow FFF System (AF4)► Separator

► Channel

AF2000 MF Flow FFF System (AF4)

AF2000 Analytical Channel (AF4) 350 µm Thickness; NovaRC 10kDa Membrane

PN3241 UV – Absorbance, 254 nm

► Detection,

PN3150 RI – Refractive IndexPN3621 MALS – Static Light Scattering, 532 nm

n.a.

► Fraction.

► Eluent

Sample UW001: 0.1 M NaCl + 0.2 g/L NaN3Eluent filtered by 0.1 µm filterSample UW002: THF

Sample UW001: 0 5 mL/min and 0 5 mL/min Smart Stream Splitting► Eluent

► Flow Rate

Sample UW001: 0.5 mL/min and 0.5 mL/min Smart Stream SplittingSample UW002: 0.5 mL/min

Postnova AF2000 Multi Flow FFF

www.postnova.com .www.postnova.com .

► Flow Rate

22.10.2015 - C O N F I D E N T I A L - 12

Page 13: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

Summary (AT THE BEGINNING)

The best fit of the shape (SAS ie ) doesn’t agree ith theThe best fit of the shape (SAS view) doesn’t agree with the fit of the intensity

Results of SLS agree with the SANS results in intensity

The hypothesis: Most of the system is composed of unimersThe hypothesis: Most of the system is composed of unimerswith a small amount of very long cylinders that explain both the results of SLS and SANS was confirmed by AF4 resultsresults

www.postnova.com . 13

Page 14: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

β-Sheet Forming Cyclic Peptide Nanotubes

D‐ L‐

www.postnova.com . 14Ghadiri, M. R.; Granja, et al., Nature, 1993, 366 (6453), 324.Horne, W. S., et al., Bioorg. & Med. Chem. 2005, 13 (17), 5145.

Page 15: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

SANS

Study of SL155

PEGA60 DP = 60

www.postnova.com . 15

60

Page 16: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

Nanotube Design Elements

Internal Diameter

External Functionality

.

www.postnova.com . 16Chapman, R., et al., Chem. Soc. Rev. 2012, 41, 6023.Couet, J.; Biesalski, M., Small 2008, 4 (7), 1008.

Page 17: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

Sample UW001: SL155Raw Data Fractogram of AF4 - MALS and UV

SystemSystem• PN5300 Auto Injector• AF2000 FFF System• PN1650 Smart Stream Splitter

PN3621 MALS D t t• PN3621 MALS Detector• PN3241 UV Detector• PN3150 RI Detector

Conditions• Injection Volume: 20 µL• Concentration: 10 mg/mL• LS 90°(red trace)S 90 ( ed t ace)• RI Detector (blue trace)• The 1st peak shows a double structure and was detected between 8 and

25 min by light scattering detection.• A 2nd peak was detected between 25 and 60 min by light scattering

d t ti

www.postnova.com .www.postnova.com . 22.10.2015 - C O N F I D E N T I A L - 17

detection.

Page 18: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

Sample UW001: SL155O l M l M d LS Si lOverlay: Molar Mass and LS Signal• The Molar Mass was calculated from MALS and RI data• given value for dn/dc = 0.12 mL/g

In the 1st Fraction the Sample contains Material with a Molar Mass of appr 5 4 x 104 g/mol (w• In the 1st Fraction the Sample contains Material with a Molar Mass of appr. 5.4 x 104 g/mol (w-Average), in the 2nd Fraction of 2.6 x 105 g/mol and in the 3rd Fraction of 2.3 x 106 g/mol.

Conditions• Molar Mass (red dots)• LS Signal (blue trace)• LS Signal (blue trace)• Fitting by Random Coil Model

Mw [g/mol]

1 Fractionn-Average 4.2 x 1041 2 3

1. Fraction8.5 – 16.5 min

w-Average 5.4 x 104

z-Average 7.2 x 104

2. Fraction n-Average 2.5 x 105

A 2 6 10516.5 – 20.8 min

w-Average 2.6 x 105

z-Average 2.7 x 105

3. Fraction27 5 47 5

n-Average 7.1 x 105

w-Average 2.3 x 106

www.postnova.com .www.postnova.com . 22.10.2015 - C O N F I D E N T I A L - 18

27.5 – 47.5 min

w Average 2.3 x 10

z-Average 1.9 x 107

Page 19: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

Sample UW001: SL155

RadiusMolar Mass

Molar Mass Distribution Particle Size Distribution Radius

al

0.0550.0500.0450.0400 035

C

1.0

o a ass

al

1.0 C

1.0

Diff

eren

tia 0.0350.0300.0250.0200.0150.010

Cum

ulativeDiff

eren

tia

Cum

ulative

Radius [nm]140120100806040200

0.0050.000 0.0

Molar Mass [g/mol]1e+3 1e+4 1e+5 1e+6 1e+7 1e+8 1e+9

0.0 0.0

Differential Molar Mass/Particle Size Distribution (blue trace) Cumulative Molar Mass/Particle Size Distribution (red trace)Differential Molar Mass/Particle Size Distribution (blue trace), Cumulative Molar Mass/Particle Size Distribution (red trace)

The sample shows a multimodal distribution. For the 1st fraction (8.5 – 16.5 min) the distribution is in therange of 1.5 x 104 – 1.7 x 105 g/mol, for the 2nd fraction (16.5 – 20.8 min) in the range of 1.7 x 105 – 3.4 x 105

g/mol and for the 3rd fraction (27.5 – 47.5 min) in the range of 3.4 x 105 – 1.2 x 108 g/mol.

www.postnova.com .www.postnova.com . 22.10.2015 - C O N F I D E N T I A L - 19

g/mol and for the 3 fraction (27.5 47.5 min) in the range of 3.4 x 10 1.2 x 10 g/mol.

Page 20: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

Sample UW001: SL155Overlay: Radius of Gyration and LS Signal• The Radius of Gyration was calculated from MALS angular data• The 3rd Fraction shows a Radius of Gyration of 65 nm (z-average)

Conditions• Radius of Gyration (red dots)• LS Signal (blue trace)• Fitting by Random Coil Model2

Rg [nm]

3. Fraction27.5 – 47.5 min

n-Average 29

w-Average 35

A 65min z-Average 65

www.postnova.com .www.postnova.com . 22.10.2015 - C O N F I D E N T I A L - 20

Page 21: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

Study of SL155 in D2O 10g/L

Using the softare SAS view it’s possible to fit the shapeof the data to have an idea of the shape of the objects in solution

1

10

SANS SL155_D2O Fit Gaussian Chain

1

10

SANS SL155_D2O

0,1I 0,1I

0 01 0 1 1

1E-3

0,01

0 01 0 1 1

1E-3

0,01

0,01 0,1 1

q (Å-1 )0,01 0,1 1

q (Å-1 )

The tail at high q is well fitted with a Gaussian chain form factor

www.postnova.com . 21

The tail at high q is well fitted with a Gaussian chain form factor

Page 22: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

Study of SL155 in D2O 10g/L

Using the software SAS view it’s possible to fit the shape of the data to have an idea of the shape of the object in solution

1

10

SANS SL155_D2O Fit Rod

0,1I

0 01 0 1 1

1E-3

0,01

0,01 0,1 1

q (Å-1 )

The rollover is well fitted by a Rod form factor

www.postnova.com . 22

The rollover is well fitted by a Rod form factor

Page 23: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

Study of SL155 in D2O 10g/L

Using the software SASit’s possible to fit the shape of the data to have an idea of the shape of the objects in solution

1

10 SANS SL155_D2O Fit Rod Fit Gaussian Chain Sum Rod + GC

0,1I

L = 160Å

0 01 0 1 1

1E-3

0,01

R = 20Å 0,01 0,1 1

q (Å-1 )

A sum of Rod and Gaussian Chain form factors fit nicely all the q rangeL = 160Å => Nagg = 32 (Number of Aggregates)

www.postnova.com . 23

L = 160Å => Nagg = 32 (Number of Aggregates)

Page 24: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

AF4 analysis of SL 155 in water

• Main populationMw = 54000g/molMw = 54000g/mol

(unimers)

• Second population

• Minor populationMw = 2.3x106g/mol

Nagg = 38.7p pMw = 2.6x105g/mol

Nagg = 4

AF4 results agree with SLS and SANS results: unimers are the main population

An additional population of Nagg = 4 that was not observed with SLS or SANS isobserved

The minor population (Nagg = 38.7) explain the result in shape of SANS

www.postnova.com . 24

p p ( gg ) p p

Page 25: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

C fCentrifugal FFF

www.postnova.com .www.postnova.com . 22.10.2015 25

Page 26: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

Centrifugal FFF – Principle

Separation PrincipleContent• FFF Principle• Systems• Applications• SummarySummary

• Gravity Separation Field up to 2.500 g• Size Separation Range: Particles 5 nm – 100 µm

Separation based on Si e and Densit

www.postnova.com . 22.10.2015 26

• Separation based on Size and Density

Page 27: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

Polyelectrolyte Encapsulated AroundNanoParticleShown are the raw data signals of the Light Scattering detector at 90° (red line) and the UV detector at 254 nm using Centrifugal Field Flow Fractionation CF3

Using Centrifugal FFF (CF3) we can separate the free Polyelectrolyte from the cross-linked encapsulated

www.postnova.com . 22.10.2015 27

nanoparticle.

Page 28: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

Encapsulated Nanoparticle with Cross-linked Polyelectrolyte Particle Distribution using CF3

In the following figures the particle size distribution is shown:Differential Radius Distribution (blue line) and Cumulative Radius Distribution

Radius0.0280.026 1.0

Differential Radius Distribution (blue line) and Cumulative Radius Distribution (red line) of the second peak of sample.

Rg [nm]

A 30entia

l

0.0240.0220.02

0.0180.0160 014

Cum

n-Average 30

w-Average 41

Diff

ere 0.014

0.0120.01

0.0080.0060.004

mulative

z-Average 55Radius [nm]

140120100806040200

0.0020

-0.0020.0

www.postnova.com . 22.10.2015 28

Page 29: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

Thermal FFF

www.postnova.com .www.postnova.com . 22.10.2015 29

Page 30: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

Thermal Field-Flow Fractionation

TF3 involves a hot and cold plate to generate a temperature gradient perpendicular to the separation channel. Thermal Diffusion occurs and this us often a property of the polymers morphology.morphology.

• Thermal gradient up to ∆120°C• Separation kDa up to several MDa• Separation kDa up to several MDa• Analysis time, 10 – 120 min (no upper limit)• Separation depends on D and DT

→ Separation according to size and chemical composition

www.postnova.com . 22.10.2015 30

composition

Page 31: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

SEC versus Thermal FFF

Analysis of PS and PMMA by SEC and Thermal FFF PS, PMMA and a mixture of both standards in THF. Taking advantage of the separation by chemical composition in TF3.

www.postnova.com . 22.10.2015 31

Page 32: Field-Flow Fractionation of Macromolecules and Structures ... Event/2015/Related Techniques...AF2000 MF Flow FFF System (AF4) Channel AF2000 MF Flow FFF System (AF4) AF2000 Analytical

Conclusion

• Field Flow Fractionation Complements SEC/GPC in standard separations

• FFF solves problems when the mass or particle size of the macromolecule exceeds the pore size of the column

• FFF solves the problem when the interaction between theFFF solves the problem when the interaction between the macromolecule and the packing material limits the solvent and pH choice ( GPC Symposium 2014 Presentation)

• FFF solves the problem when the macromolecule is reactive and unstable and the removal of packing and solvent control enables sensible elution characteristics

• FFF can be tailored to the application

• FFF is Great Fun to do in the lab ;) !!

www.postnova.com . 22.10.2015 32