User ManualZirconia abutments
MIS Implants Technologies Ltd.P.O.Box 7, Bar Lev Industrial Park, 2015600, ISRAELWebsite: www.mis-implants.com
© MIS Corporation. All rights reserved. MP-UIZI1 Rev. 2, Apr. 2014
EC REP
MIS Implants Technologies GmbHSimeonscarre 2, 32423 Minden, GermanyTel: +49 571-972-7620 Email: [email protected]
Non-Cemented abutments after cementation or pre-cemented abutments should never be put into a ceramic furnace or exposed to temperatures, higher than those normaly found in the oral cavity due to the risk of cement failure ▪ All sharp edges resulting from adjustments must be rounded and polished ▪ Do not reduce wall thickness below 0.5mm ▪ Always use a new, smooth bur with capious amount of water for the adjustment of ZIrconia components.
New product▪ The device is not sterile ▪ Cleaning and sterilization are required prior to first use.
Cleaning, Disinfection and SterilizationAbutments should be cleaned, disinfected, and even sterilized for specific clinical procedures. Effective cleaning and disinfection is an obligatory requirement for effective sterilization.
Cleaning:▪ Clean under running water while brushing both outer and inner surfaces with the appropriate brushes ▪ Immerse in compatible cleaning solutions (e.g. Cidezyme/Enzol, Johnson & Johnson). The cleaner or disinfectant should not contain any of the following ingredients: Strong organic acids or mineral acids, strong lye, organic solvents (e.g. acetone, ether, hexane, benzene), oxidizing agents (e.g. peroxide), halogens (chlorine, iodine, bromine) or aromatic, halogenated hydrocarbons ▪ Post-rinse the implant abutment at least three times with water.
Disinfection: We recommend a high level disinfectant such as, Cidex OPA
(Johnson & Johnson) for disinfection of the implant abutments.▪ Soak the implant abutments in the disinfectant solution for the required time. Refer to the Cidex OPA instructions for use.▪ Remove the implant abutment from the disinfectant solution.▪ Rinse at least three times with highly purified water.
Heat Sterilization: Recommended autoclave sterilization protocol: 20 minutes at 121°C (250 °F)
Reusability: Abutments should never be reused.
Key to codes used
Attention, see instructions for use
Catalogue numberBatch Code
Manufacturer
Internal Hex. Connection
MK-SPZ4 (For internal hex. SP Premolar tooth) MK-WPZ4 (For internal hex. WP Premolar tooth) MK-NIZ6 (For internal hex. NP Incisor tooth) MK-SIZ6 (For internal hex. SP Incisor tooth)
Conical Connection
CK-SPZ4 (For conical connection SP Premolar tooth)CK-WPZ4 (For conical connection WP Premolar tooth)CK-SIZ6 (For conical connection SP Incisor tooth)
Internal Hex. Connection
MK-SPC4 (For internal hex. SP Premolar tooth) MK-WPC4 (For internal hex. WP Premolar tooth) MK-NIC6 (For internal hex. NP Incisor tooth) MK-SIC6 (For internal hex. SP Incisor tooth)
Conical Connection
CK-SPC4 (For conical connection SP Premolar tooth)CK-WPC4 (For conical connection WP Premolar tooth)CK-SIC6 (For conical connection SP Incisor tooth)
Non-Cemented Catalog number index
IncisorIncisor PremolarPremolar
Non-Cemented Pre-Cemented
Titanium-Zirconia prosthetic abutments are designed to satisfy high esthetic demands. The abutments are highly biocompatible, customizable, and feature superior strength characteristics. MIS’s Titanium-Zirconia abutments are available in two options:
Overview
Non-cemented Zirconia components can be modified by reduction or by addition of porcelain and/or by altering their shade. Please follow conventional techniques and procedures, and ensure that the screw hole is clean and clear. Follow cement manufacturer’s guidelines and instructions, including pre-cementation preparations.
Design & Cementation.
1
The Zirconia abutments are manufactured from an anatomically designed Zirconia component which is pre-cemented to a titanium base, allowing modification by reduction only.
Pre - cemented.
Non-CementedDesigned to be modified by the technician. White, Zirconia shade.
Pre-CementedReady to use. Pre-Cemented at the factory. A1 VITA ® shade.
Remove excess cement
Incisor set
Incisor set
Premolar set
Premolar set
Recommended Anatomic healing caps. For anatomic healing caps catalog no. please
refer to online catalog at: www.mis-implants.com
If needed, adjust the Zirconia component using high-speed smooth new diamond burs with copious amount of water-spray. Polish using appropriate rubber tips and wheels.
2
Final adjustment.
Use a torque ratchet and hex. driver to apply 30Ncm of force to secure abutment prior to delivery of final restorat ion.
3
30Ncm.
Titanium-Zirconia prosthetic abutments are designed to satisfy high esthetic demands. The abutments are highly biocompatible, customizable, and feature superior strength characteristics. MIS’s Titanium-Zirconia abutments are available in two options:
Overview
Non-cemented Zirconia components can be modified by reduction or by addition of porcelain and/or by altering their shade. Please follow conventional techniques and procedures, and ensure that the screw hole is clean and clear. Follow cement manufacturer’s guidelines and instructions, including pre-cementation preparations.
Design & Cementation.
1
The Zirconia abutments are manufactured from an anatomically designed Zirconia component which is pre-cemented to a titanium base, allowing modification by reduction only.
Pre - cemented.
Non-CementedDesigned to be modified by the technician. White, Zirconia shade.
Pre-CementedReady to use. Pre-Cemented at the factory. A1 VITA ® shade.
Remove excess cement
Incisor set
Incisor set
Premolar set
Premolar set
Recommended Anatomic healing caps. For anatomic healing caps catalog no. please
refer to online catalog at: www.mis-implants.com
If needed, adjust the Zirconia component using high-speed smooth new diamond burs with copious amount of water-spray. Polish using appropriate rubber tips and wheels.
2
Final adjustment.
Use a torque ratchet and hex. driver to apply 30Ncm of force to secure abutment prior to delivery of final restorat ion.
3
30Ncm.
Titanium-Zirconia prosthetic abutments are designed to satisfy high esthetic demands. The abutments are highly biocompatible, customizable, and feature superior strength characteristics. MIS’s Titanium-Zirconia abutments are available in two options:
Overview
Non-cemented Zirconia components can be modified by reduction or by addition of porcelain and/or by altering their shade. Please follow conventional techniques and procedures, and ensure that the screw hole is clean and clear. Follow cement manufacturer’s guidelines and instructions, including pre-cementation preparations.
Design & Cementation.
1
The Zirconia abutments are manufactured from an anatomically designed Zirconia component which is pre-cemented to a titanium base, allowing modification by reduction only.
Pre - cemented.
Non-CementedDesigned to be modified by the technician. White, Zirconia shade.
Pre-CementedReady to use. Pre-Cemented at the factory. A1 VITA ® shade.
Remove excess cement
Incisor set
Incisor set
Premolar set
Premolar set
Recommended Anatomic healing caps. For anatomic healing caps catalog no. please
refer to online catalog at: www.mis-implants.com
If needed, adjust the Zirconia component using high-speed smooth new diamond burs with copious amount of water-spray. Polish using appropriate rubber tips and wheels.
2
Final adjustment.
Use a torque ratchet and hex. driver to apply 30Ncm of force to secure abutment prior to delivery of final restorat ion.
3
30Ncm.
Titanium-Zirconia prosthetic abutments are designed to satisfy high esthetic demands. The abutments are highly biocompatible, customizable, and feature superior strength characteristics. MIS’s Titanium-Zirconia abutments are available in two options:
Overview
Non-cemented Zirconia components can be modified by reduction or by addition of porcelain and/or by altering their shade. Please follow conventional techniques and procedures, and ensure that the screw hole is clean and clear. Follow cement manufacturer’s guidelines and instructions, including pre-cementation preparations.
Design & Cementation.
1
The Zirconia abutments are manufactured from an anatomically designed Zirconia component which is pre-cemented to a titanium base, allowing modification by reduction only.
Pre - cemented.
Non-CementedDesigned to be modified by the technician. White, Zirconia shade.
Pre-CementedReady to use. Pre-Cemented at the factory. A1 VITA ® shade.
Remove excess cement
Incisor set
Incisor set
Premolar set
Premolar set
Recommended Anatomic healing caps. For anatomic healing caps catalog no. please
refer to online catalog at: www.mis-implants.com
If needed, adjust the Zirconia component using high-speed smooth new diamond burs with copious amount of water-spray. Polish using appropriate rubber tips and wheels.
2
Final adjustment.
Use a torque ratchet and hex. driver to apply 30Ncm of force to secure abutment prior to delivery of final restorat ion.
3
30Ncm.
Titanium-Zirconia prosthetic abutments are designed to satisfy high esthetic demands. The abutments are highly biocompatible, customizable, and feature superior strength characteristics. MIS’s Titanium-Zirconia abutments are available in two options:
Overview
Non-cemented Zirconia components can be modified by reduction or by addition of porcelain and/or by altering their shade. Please follow conventional techniques and procedures, and ensure that the screw hole is clean and clear. Follow cement manufacturer’s guidelines and instructions, including pre-cementation preparations.
Design & Cementation.
1
The Zirconia abutments are manufactured from an anatomically designed Zirconia component which is pre-cemented to a titanium base, allowing modification by reduction only.
Pre - cemented.
Non-CementedDesigned to be modified by the technician. White, Zirconia shade.
Pre-CementedReady to use. Pre-Cemented at the factory. A1 VITA ® shade.
Remove excess cement
Incisor set
Incisor set
Premolar set
Premolar set
Recommended Anatomic healing caps. For anatomic healing caps catalog no. please
refer to online catalog at: www.mis-implants.com
If needed, adjust the Zirconia component using high-speed smooth new diamond burs with copious amount of water-spray. Polish using appropriate rubber tips and wheels.
2
Final adjustment.
Use a torque ratchet and hex. driver to apply 30Ncm of force to secure abutment prior to delivery of final restorat ion.
3
30Ncm.
Titanium-Zirconia prosthetic abutments are designed to satisfy high esthetic demands. The abutments are highly biocompatible, customizable, and feature superior strength characteristics. MIS’s Titanium-Zirconia abutments are available in two options:
Overview
Non-cemented Zirconia components can be modified by reduction or by addition of porcelain and/or by altering their shade. Please follow conventional techniques and procedures, and ensure that the screw hole is clean and clear. Follow cement manufacturer’s guidelines and instructions, including pre-cementation preparations.
Design & Cementation.
1
The Zirconia abutments are manufactured from an anatomically designed Zirconia component which is pre-cemented to a titanium base, allowing modification by reduction only.
Pre - cemented.
Non-CementedDesigned to be modified by the technician. White, Zirconia shade.
Pre-CementedReady to use. Pre-Cemented at the factory. A1 VITA ® shade.
Remove excess cement
Incisor set
Incisor set
Premolar set
Premolar set
Recommended Anatomic healing caps. For anatomic healing caps catalog no. please
refer to online catalog at: www.mis-implants.com
If needed, adjust the Zirconia component using high-speed smooth new diamond burs with copious amount of water-spray. Polish using appropriate rubber tips and wheels.
2
Final adjustment.
Use a torque ratchet and hex. driver to apply 30Ncm of force to secure abutment prior to delivery of final restorat ion.
3
30Ncm.
Titanium-Zirconia prosthetic abutments are designed to satisfy high esthetic demands. The abutments are highly biocompatible, customizable, and feature superior strength characteristics. MIS’s Titanium-Zirconia abutments are available in two options:
Overview
Non-cemented Zirconia components can be modified by reduction or by addition of porcelain and/or by altering their shade. Please follow conventional techniques and procedures, and ensure that the screw hole is clean and clear. Follow cement manufacturer’s guidelines and instructions, including pre-cementation preparations.
Design & Cementation.
1
The Zirconia abutments are manufactured from an anatomically designed Zirconia component which is pre-cemented to a titanium base, allowing modification by reduction only.
Pre - cemented.
Non-CementedDesigned to be modified by the technician. White, Zirconia shade.
Pre-CementedReady to use. Pre-Cemented at the factory. A1 VITA ® shade.
Remove excess cement
Incisor set
Incisor set
Premolar set
Premolar set
Recommended Anatomic healing caps. For anatomic healing caps catalog no. please
refer to online catalog at: www.mis-implants.com
If needed, adjust the Zirconia component using high-speed smooth new diamond burs with copious amount of water-spray. Polish using appropriate rubber tips and wheels.
2
Final adjustment.
Use a torque ratchet and hex. driver to apply 30Ncm of force to secure abutment prior to delivery of final restorat ion.
3
30Ncm.
Titanium-Zirconia prosthetic abutments are designed to satisfy high esthetic demands. The abutments are highly biocompatible, customizable, and feature superior strength characteristics. MIS’s Titanium-Zirconia abutments are available in two options:
Overview
Non-cemented Zirconia components can be modified by reduction or by addition of porcelain and/or by altering their shade. Please follow conventional techniques and procedures, and ensure that the screw hole is clean and clear. Follow cement manufacturer’s guidelines and instructions, including pre-cementation preparations.
Design & Cementation.
1
The Zirconia abutments are manufactured from an anatomically designed Zirconia component which is pre-cemented to a titanium base, allowing modification by reduction only.
Pre - cemented.
Non-CementedDesigned to be modified by the technician. White, Zirconia shade.
Pre-CementedReady to use. Pre-Cemented at the factory. A1 VITA ® shade.
Remove excess cement
Incisor set
Incisor set
Premolar set
Premolar set
Recommended Anatomic healing caps. For anatomic healing caps catalog no. please
refer to online catalog at: www.mis-implants.com
If needed, adjust the Zirconia component using high-speed smooth new diamond burs with copious amount of water-spray. Polish using appropriate rubber tips and wheels.
2
Final adjustment.
Use a torque ratchet and hex. driver to apply 30Ncm of force to secure abutment prior to delivery of final restorat ion.
3
30Ncm.
Titanium-Zirconia prosthetic abutments are designed to satisfy high esthetic demands. The abutments are highly biocompatible, customizable, and feature superior strength characteristics. MIS’s Titanium-Zirconia abutments are available in two options:
Overview
Non-cemented Zirconia components can be modified by reduction or by addition of porcelain and/or by altering their shade. Please follow conventional techniques and procedures, and ensure that the screw hole is clean and clear. Follow cement manufacturer’s guidelines and instructions, including pre-cementation preparations.
Design & Cementation.
1
The Zirconia abutments are manufactured from an anatomically designed Zirconia component which is pre-cemented to a titanium base, allowing modification by reduction only.
Pre - cemented.
Non-CementedDesigned to be modified by the technician. White, Zirconia shade.
Pre-CementedReady to use. Pre-Cemented at the factory. A1 VITA ® shade.
Remove excess cement
Incisor set
Incisor set
Premolar set
Premolar set
Recommended Anatomic healing caps. For anatomic healing caps catalog no. please
refer to online catalog at: www.mis-implants.com
If needed, adjust the Zirconia component using high-speed smooth new diamond burs with copious amount of water-spray. Polish using appropriate rubber tips and wheels.
2
Final adjustment.
Use a torque ratchet and hex. driver to apply 30Ncm of force to secure abutment prior to delivery of final restorat ion.
3
30Ncm.
Titanium-Zirconia prosthetic abutments are designed to satisfy high esthetic demands. The abutments are highly biocompatible, customizable, and feature superior strength characteristics. MIS’s Titanium-Zirconia abutments are available in two options:
Overview
Non-cemented Zirconia components can be modified by reduction or by addition of porcelain and/or by altering their shade. Please follow conventional techniques and procedures, and ensure that the screw hole is clean and clear. Follow cement manufacturer’s guidelines and instructions, including pre-cementation preparations.
Design & Cementation.
1
The Zirconia abutments are manufactured from an anatomically designed Zirconia component which is pre-cemented to a titanium base, allowing modification by reduction only.
Pre - cemented.
Non-CementedDesigned to be modified by the technician. White, Zirconia shade.
Pre-CementedReady to use. Pre-Cemented at the factory. A1 VITA ® shade.
Remove excess cement
Incisor set
Incisor set
Premolar set
Premolar set
Recommended Anatomic healing caps. For anatomic healing caps catalog no. please
refer to online catalog at: www.mis-implants.com
If needed, adjust the Zirconia component using high-speed smooth new diamond burs with copious amount of water-spray. Polish using appropriate rubber tips and wheels.
2
Final adjustment.
Use a torque ratchet and hex. driver to apply 30Ncm of force to secure abutment prior to delivery of final restorat ion.
3
30Ncm.
Titanium-Zirconia prosthetic abutments are designed to satisfy high esthetic demands. The abutments are highly biocompatible, customizable, and feature superior strength characteristics. MIS’s Titanium-Zirconia abutments are available in two options:
Overview
Non-cemented Zirconia components can be modified by reduction or by addition of porcelain and/or by altering their shade. Please follow conventional techniques and procedures, and ensure that the screw hole is clean and clear. Follow cement manufacturer’s guidelines and instructions, including pre-cementation preparations.
Design & Cementation.
1
The Zirconia abutments are manufactured from an anatomically designed Zirconia component which is pre-cemented to a titanium base, allowing modification by reduction only.
Pre - cemented.
Non-CementedDesigned to be modified by the technician. White, Zirconia shade.
Pre-CementedReady to use. Pre-Cemented at the factory. A1 VITA ® shade.
Remove excess cement
Incisor set
Incisor set
Premolar set
Premolar set
Recommended Anatomic healing caps. For anatomic healing caps catalog no. please
refer to online catalog at: www.mis-implants.com
If needed, adjust the Zirconia component using high-speed smooth new diamond burs with copious amount of water-spray. Polish using appropriate rubber tips and wheels.
2
Final adjustment.
Use a torque ratchet and hex. driver to apply 30Ncm of force to secure abutment prior to delivery of final restorat ion.
3
30Ncm.
Titanium-Zirconia prosthetic abutments are designed to satisfy high esthetic demands. The abutments are highly biocompatible, customizable, and feature superior strength characteristics. MIS’s Titanium-Zirconia abutments are available in two options:
Overview
Non-cemented Zirconia components can be modified by reduction or by addition of porcelain and/or by altering their shade. Please follow conventional techniques and procedures, and ensure that the screw hole is clean and clear. Follow cement manufacturer’s guidelines and instructions, including pre-cementation preparations.
Design & Cementation.
1
The Zirconia abutments are manufactured from an anatomically designed Zirconia component which is pre-cemented to a titanium base, allowing modification by reduction only.
Pre - cemented.
Non-CementedDesigned to be modified by the technician. White, Zirconia shade.
Pre-CementedReady to use. Pre-Cemented at the factory. A1 VITA ® shade.
Remove excess cement
Incisor set
Incisor set
Premolar set
Premolar set
Recommended Anatomic healing caps. For anatomic healing caps catalog no. please
refer to online catalog at: www.mis-implants.com
If needed, adjust the Zirconia component using high-speed smooth new diamond burs with copious amount of water-spray. Polish using appropriate rubber tips and wheels.
2
Final adjustment.
Use a torque ratchet and hex. driver to apply 30Ncm of force to secure abutment prior to delivery of final restorat ion.
3
30Ncm.
Titanium-Zirconia prosthetic abutments are designed to satisfy high esthetic demands. The abutments are highly biocompatible, customizable, and feature superior strength characteristics. MIS’s Titanium-Zirconia abutments are available in two options:
Overview
Non-cemented Zirconia components can be modified by reduction or by addition of porcelain and/or by altering their shade. Please follow conventional techniques and procedures, and ensure that the screw hole is clean and clear. Follow cement manufacturer’s guidelines and instructions, including pre-cementation preparations.
Design & Cementation.
1
The Zirconia abutments are manufactured from an anatomically designed Zirconia component which is pre-cemented to a titanium base, allowing modification by reduction only.
Pre - cemented.
Non-CementedDesigned to be modified by the technician. White, Zirconia shade.
Pre-CementedReady to use. Pre-Cemented at the factory. A1 VITA ® shade.
Remove excess cement
Incisor set
Incisor set
Premolar set
Premolar set
Recommended Anatomic healing caps. For anatomic healing caps catalog no. please
refer to online catalog at: www.mis-implants.com
If needed, adjust the Zirconia component using high-speed smooth new diamond burs with copious amount of water-spray. Polish using appropriate rubber tips and wheels.
2
Final adjustment.
Use a torque ratchet and hex. driver to apply 30Ncm of force to secure abutment prior to delivery of final restorat ion.
3
30Ncm.
Titanium-Zirconia prosthetic abutments are designed to satisfy high esthetic demands. The abutments are highly biocompatible, customizable, and feature superior strength characteristics. MIS’s Titanium-Zirconia abutments are available in two options:
Overview
Non-cemented Zirconia components can be modified by reduction or by addition of porcelain and/or by altering their shade. Please follow conventional techniques and procedures, and ensure that the screw hole is clean and clear. Follow cement manufacturer’s guidelines and instructions, including pre-cementation preparations.
Design & Cementation.
1
The Zirconia abutments are manufactured from an anatomically designed Zirconia component which is pre-cemented to a titanium base, allowing modification by reduction only.
Pre - cemented.
Non-CementedDesigned to be modified by the technician. White, Zirconia shade.
Pre-CementedReady to use. Pre-Cemented at the factory. A1 VITA ® shade.
Remove excess cement
Incisor set
Incisor set
Premolar set
Premolar set
Recommended Anatomic healing caps. For anatomic healing caps catalog no. please
refer to online catalog at: www.mis-implants.com
If needed, adjust the Zirconia component using high-speed smooth new diamond burs with copious amount of water-spray. Polish using appropriate rubber tips and wheels.
2
Final adjustment.
Use a torque ratchet and hex. driver to apply 30Ncm of force to secure abutment prior to delivery of final restorat ion.
3
30Ncm.
Titanium-Zirconia prosthetic abutments are designed to satisfy high esthetic demands. The abutments are highly biocompatible, customizable, and feature superior strength characteristics. MIS’s Titanium-Zirconia abutments are available in two options:
Overview
Non-cemented Zirconia components can be modified by reduction or by addition of porcelain and/or by altering their shade. Please follow conventional techniques and procedures, and ensure that the screw hole is clean and clear. Follow cement manufacturer’s guidelines and instructions, including pre-cementation preparations.
Design & Cementation.
1
The Zirconia abutments are manufactured from an anatomically designed Zirconia component which is pre-cemented to a titanium base, allowing modification by reduction only.
Pre - cemented.
Non-CementedDesigned to be modified by the technician. White, Zirconia shade.
Pre-CementedReady to use. Pre-Cemented at the factory. A1 VITA ® shade.
Remove excess cement
Incisor set
Incisor set
Premolar set
Premolar set
Recommended Anatomic healing caps. For anatomic healing caps catalog no. please
refer to online catalog at: www.mis-implants.com
If needed, adjust the Zirconia component using high-speed smooth new diamond burs with copious amount of water-spray. Polish using appropriate rubber tips and wheels.
2
Final adjustment.
Use a torque ratchet and hex. driver to apply 30Ncm of force to secure abutment prior to delivery of final restorat ion.
3
30Ncm.
®
®
User ManualZirconia abutments
MIS Implants Technologies Ltd.P.O.Box 7, Bar Lev Industrial Park, 2015600, ISRAELWebsite: www.mis-implants.com
© MIS Corporation. All rights reserved. MP-UIZI1 Rev. 2, Apr. 2014
EC REP
MIS Implants Technologies GmbHSimeonscarre 2, 32423 Minden, GermanyTel: +49 571-972-7620 Email: [email protected]
Non-Cemented abutments after cementation or pre-cemented abutments should never be put into a ceramic furnace or exposed to temperatures, higher than those normaly found in the oral cavity due to the risk of cement failure ▪ All sharp edges resulting from adjustments must be rounded and polished ▪ Do not reduce wall thickness below 0.5mm ▪ Always use a new, smooth bur with capious amount of water for the adjustment of ZIrconia components.
New product▪ The device is not sterile ▪ Cleaning and sterilization are required prior to first use.
Cleaning, Disinfection and SterilizationAbutments should be cleaned, disinfected, and even sterilized for specific clinical procedures. Effective cleaning and disinfection is an obligatory requirement for effective sterilization.
Cleaning:▪ Clean under running water while brushing both outer and inner surfaces with the appropriate brushes ▪ Immerse in compatible cleaning solutions (e.g. Cidezyme/Enzol, Johnson & Johnson). The cleaner or disinfectant should not contain any of the following ingredients: Strong organic acids or mineral acids, strong lye, organic solvents (e.g. acetone, ether, hexane, benzene), oxidizing agents (e.g. peroxide), halogens (chlorine, iodine, bromine) or aromatic, halogenated hydrocarbons ▪ Post-rinse the implant abutment at least three times with water.
Disinfection: We recommend a high level disinfectant such as, Cidex OPA
(Johnson & Johnson) for disinfection of the implant abutments.▪ Soak the implant abutments in the disinfectant solution for the required time. Refer to the Cidex OPA instructions for use.▪ Remove the implant abutment from the disinfectant solution.▪ Rinse at least three times with highly purified water.
Heat Sterilization: Recommended autoclave sterilization protocol: 20 minutes at 121°C (250 °F)
Reusability: Abutments should never be reused.
Key to codes used
Attention, see instructions for use
Catalogue numberBatch Code
Manufacturer
Internal Hex. Connection
MK-SPZ4 (For internal hex. SP Premolar tooth) MK-WPZ4 (For internal hex. WP Premolar tooth) MK-NIZ6 (For internal hex. NP Incisor tooth) MK-SIZ6 (For internal hex. SP Incisor tooth)
Conical Connection
CK-SPZ4 (For conical connection SP Premolar tooth)CK-WPZ4 (For conical connection WP Premolar tooth)CK-SIZ6 (For conical connection SP Incisor tooth)
Internal Hex. Connection
MK-SPC4 (For internal hex. SP Premolar tooth) MK-WPC4 (For internal hex. WP Premolar tooth) MK-NIC6 (For internal hex. NP Incisor tooth) MK-SIC6 (For internal hex. SP Incisor tooth)
Conical Connection
CK-SPC4 (For conical connection SP Premolar tooth)CK-WPC4 (For conical connection WP Premolar tooth)CK-SIC6 (For conical connection SP Incisor tooth)
Non-Cemented Catalog number index
IncisorIncisor PremolarPremolar
Non-Cemented Pre-Cemented
®
®
User ManualZirconia abutments
MIS Implants Technologies Ltd.P.O.Box 7, Bar Lev Industrial Park, 2015600, ISRAELWebsite: www.mis-implants.com
© MIS Corporation. All rights reserved. MP-UIZI1 Rev. 2, Apr. 2014
EC REP
MIS Implants Technologies GmbHSimeonscarre 2, 32423 Minden, GermanyTel: +49 571-972-7620 Email: [email protected]
Non-Cemented abutments after cementation or pre-cemented abutments should never be put into a ceramic furnace or exposed to temperatures, higher than those normaly found in the oral cavity due to the risk of cement failure ▪ All sharp edges resulting from adjustments must be rounded and polished ▪ Do not reduce wall thickness below 0.5mm ▪ Always use a new, smooth bur with capious amount of water for the adjustment of ZIrconia components.
New product▪ The device is not sterile ▪ Cleaning and sterilization are required prior to first use.
Cleaning, Disinfection and SterilizationAbutments should be cleaned, disinfected, and even sterilized for specific clinical procedures. Effective cleaning and disinfection is an obligatory requirement for effective sterilization.
Cleaning:▪ Clean under running water while brushing both outer and inner surfaces with the appropriate brushes ▪ Immerse in compatible cleaning solutions (e.g. Cidezyme/Enzol, Johnson & Johnson). The cleaner or disinfectant should not contain any of the following ingredients: Strong organic acids or mineral acids, strong lye, organic solvents (e.g. acetone, ether, hexane, benzene), oxidizing agents (e.g. peroxide), halogens (chlorine, iodine, bromine) or aromatic, halogenated hydrocarbons ▪ Post-rinse the implant abutment at least three times with water.
Disinfection: We recommend a high level disinfectant such as, Cidex OPA
(Johnson & Johnson) for disinfection of the implant abutments.▪ Soak the implant abutments in the disinfectant solution for the required time. Refer to the Cidex OPA instructions for use.▪ Remove the implant abutment from the disinfectant solution.▪ Rinse at least three times with highly purified water.
Heat Sterilization: Recommended autoclave sterilization protocol: 20 minutes at 121°C (250 °F)
Reusability: Abutments should never be reused.
Key to codes used
Attention, see instructions for use
Catalogue numberBatch Code
Manufacturer
Internal Hex. Connection
MK-SPZ4 (For internal hex. SP Premolar tooth) MK-WPZ4 (For internal hex. WP Premolar tooth) MK-NIZ6 (For internal hex. NP Incisor tooth) MK-SIZ6 (For internal hex. SP Incisor tooth)
Conical Connection
CK-SPZ4 (For conical connection SP Premolar tooth)CK-WPZ4 (For conical connection WP Premolar tooth)CK-SIZ6 (For conical connection SP Incisor tooth)
Internal Hex. Connection
MK-SPC4 (For internal hex. SP Premolar tooth) MK-WPC4 (For internal hex. WP Premolar tooth) MK-NIC6 (For internal hex. NP Incisor tooth) MK-SIC6 (For internal hex. SP Incisor tooth)
Conical Connection
CK-SPC4 (For conical connection SP Premolar tooth)CK-WPC4 (For conical connection WP Premolar tooth)CK-SIC6 (For conical connection SP Incisor tooth)
Non-Cemented Catalog number index
IncisorIncisor PremolarPremolar
Non-Cemented Pre-Cemented
®
®
User ManualZirconia abutments
MIS Implants Technologies Ltd.P.O.Box 7, Bar Lev Industrial Park, 2015600, ISRAELWebsite: www.mis-implants.com
© MIS Corporation. All rights reserved. MP-UIZI1 Rev. 2, Apr. 2014
EC REP
MIS Implants Technologies GmbHSimeonscarre 2, 32423 Minden, GermanyTel: +49 571-972-7620 Email: [email protected]
Non-Cemented abutments after cementation or pre-cemented abutments should never be put into a ceramic furnace or exposed to temperatures, higher than those normaly found in the oral cavity due to the risk of cement failure ▪ All sharp edges resulting from adjustments must be rounded and polished ▪ Do not reduce wall thickness below 0.5mm ▪ Always use a new, smooth bur with capious amount of water for the adjustment of ZIrconia components.
New product▪ The device is not sterile ▪ Cleaning and sterilization are required prior to first use.
Cleaning, Disinfection and SterilizationAbutments should be cleaned, disinfected, and even sterilized for specific clinical procedures. Effective cleaning and disinfection is an obligatory requirement for effective sterilization.
Cleaning:▪ Clean under running water while brushing both outer and inner surfaces with the appropriate brushes ▪ Immerse in compatible cleaning solutions (e.g. Cidezyme/Enzol, Johnson & Johnson). The cleaner or disinfectant should not contain any of the following ingredients: Strong organic acids or mineral acids, strong lye, organic solvents (e.g. acetone, ether, hexane, benzene), oxidizing agents (e.g. peroxide), halogens (chlorine, iodine, bromine) or aromatic, halogenated hydrocarbons ▪ Post-rinse the implant abutment at least three times with water.
Disinfection: We recommend a high level disinfectant such as, Cidex OPA
(Johnson & Johnson) for disinfection of the implant abutments.▪ Soak the implant abutments in the disinfectant solution for the required time. Refer to the Cidex OPA instructions for use.▪ Remove the implant abutment from the disinfectant solution.▪ Rinse at least three times with highly purified water.
Heat Sterilization: Recommended autoclave sterilization protocol: 20 minutes at 121°C (250 °F)
Reusability: Abutments should never be reused.
Key to codes used
Attention, see instructions for use
Catalogue numberBatch Code
Manufacturer
Internal Hex. Connection
MK-SPZ4 (For internal hex. SP Premolar tooth) MK-WPZ4 (For internal hex. WP Premolar tooth) MK-NIZ6 (For internal hex. NP Incisor tooth) MK-SIZ6 (For internal hex. SP Incisor tooth)
Conical Connection
CK-SPZ4 (For conical connection SP Premolar tooth)CK-WPZ4 (For conical connection WP Premolar tooth)CK-SIZ6 (For conical connection SP Incisor tooth)
Internal Hex. Connection
MK-SPC4 (For internal hex. SP Premolar tooth) MK-WPC4 (For internal hex. WP Premolar tooth) MK-NIC6 (For internal hex. NP Incisor tooth) MK-SIC6 (For internal hex. SP Incisor tooth)
Conical Connection
CK-SPC4 (For conical connection SP Premolar tooth)CK-WPC4 (For conical connection WP Premolar tooth)CK-SIC6 (For conical connection SP Incisor tooth)
Non-Cemented Catalog number index
IncisorIncisor PremolarPremolar
Non-Cemented Pre-Cemented
®
®
User ManualZirconia abutments
MIS Implants Technologies Ltd.P.O.Box 7, Bar Lev Industrial Park, 2015600, ISRAELWebsite: www.mis-implants.com
© MIS Corporation. All rights reserved. MP-UIZI1 Rev. 2, Apr. 2014
EC REP
MIS Implants Technologies GmbHSimeonscarre 2, 32423 Minden, GermanyTel: +49 571-972-7620 Email: [email protected]
Non-Cemented abutments after cementation or pre-cemented abutments should never be put into a ceramic furnace or exposed to temperatures, higher than those normaly found in the oral cavity due to the risk of cement failure ▪ All sharp edges resulting from adjustments must be rounded and polished ▪ Do not reduce wall thickness below 0.5mm ▪ Always use a new, smooth bur with capious amount of water for the adjustment of ZIrconia components.
New product▪ The device is not sterile ▪ Cleaning and sterilization are required prior to first use.
Cleaning, Disinfection and SterilizationAbutments should be cleaned, disinfected, and even sterilized for specific clinical procedures. Effective cleaning and disinfection is an obligatory requirement for effective sterilization.
Cleaning:▪ Clean under running water while brushing both outer and inner surfaces with the appropriate brushes ▪ Immerse in compatible cleaning solutions (e.g. Cidezyme/Enzol, Johnson & Johnson). The cleaner or disinfectant should not contain any of the following ingredients: Strong organic acids or mineral acids, strong lye, organic solvents (e.g. acetone, ether, hexane, benzene), oxidizing agents (e.g. peroxide), halogens (chlorine, iodine, bromine) or aromatic, halogenated hydrocarbons ▪ Post-rinse the implant abutment at least three times with water.
Disinfection: We recommend a high level disinfectant such as, Cidex OPA
(Johnson & Johnson) for disinfection of the implant abutments.▪ Soak the implant abutments in the disinfectant solution for the required time. Refer to the Cidex OPA instructions for use.▪ Remove the implant abutment from the disinfectant solution.▪ Rinse at least three times with highly purified water.
Heat Sterilization: Recommended autoclave sterilization protocol: 20 minutes at 121°C (250 °F)
Reusability: Abutments should never be reused.
Key to codes used
Attention, see instructions for use
Catalogue numberBatch Code
Manufacturer
Internal Hex. Connection
MK-SPZ4 (For internal hex. SP Premolar tooth) MK-WPZ4 (For internal hex. WP Premolar tooth) MK-NIZ6 (For internal hex. NP Incisor tooth) MK-SIZ6 (For internal hex. SP Incisor tooth)
Conical Connection
CK-SPZ4 (For conical connection SP Premolar tooth)CK-WPZ4 (For conical connection WP Premolar tooth)CK-SIZ6 (For conical connection SP Incisor tooth)
Internal Hex. Connection
MK-SPC4 (For internal hex. SP Premolar tooth) MK-WPC4 (For internal hex. WP Premolar tooth) MK-NIC6 (For internal hex. NP Incisor tooth) MK-SIC6 (For internal hex. SP Incisor tooth)
Conical Connection
CK-SPC4 (For conical connection SP Premolar tooth)CK-WPC4 (For conical connection WP Premolar tooth)CK-SIC6 (For conical connection SP Incisor tooth)
Non-Cemented Catalog number index
IncisorIncisor PremolarPremolar
Non-Cemented Pre-Cemented
®
®
User ManualZirconia abutments
MIS Implants Technologies Ltd.P.O.Box 7, Bar Lev Industrial Park, 2015600, ISRAELWebsite: www.mis-implants.com
© MIS Corporation. All rights reserved. MP-UIZI1 Rev. 2, Apr. 2014
EC REP
MIS Implants Technologies GmbHSimeonscarre 2, 32423 Minden, GermanyTel: +49 571-972-7620 Email: [email protected]
Non-Cemented abutments after cementation or pre-cemented abutments should never be put into a ceramic furnace or exposed to temperatures, higher than those normaly found in the oral cavity due to the risk of cement failure ▪ All sharp edges resulting from adjustments must be rounded and polished ▪ Do not reduce wall thickness below 0.5mm ▪ Always use a new, smooth bur with capious amount of water for the adjustment of ZIrconia components.
New product▪ The device is not sterile ▪ Cleaning and sterilization are required prior to first use.
Cleaning, Disinfection and SterilizationAbutments should be cleaned, disinfected, and even sterilized for specific clinical procedures. Effective cleaning and disinfection is an obligatory requirement for effective sterilization.
Cleaning:▪ Clean under running water while brushing both outer and inner surfaces with the appropriate brushes ▪ Immerse in compatible cleaning solutions (e.g. Cidezyme/Enzol, Johnson & Johnson). The cleaner or disinfectant should not contain any of the following ingredients: Strong organic acids or mineral acids, strong lye, organic solvents (e.g. acetone, ether, hexane, benzene), oxidizing agents (e.g. peroxide), halogens (chlorine, iodine, bromine) or aromatic, halogenated hydrocarbons ▪ Post-rinse the implant abutment at least three times with water.
Disinfection: We recommend a high level disinfectant such as, Cidex OPA
(Johnson & Johnson) for disinfection of the implant abutments.▪ Soak the implant abutments in the disinfectant solution for the required time. Refer to the Cidex OPA instructions for use.▪ Remove the implant abutment from the disinfectant solution.▪ Rinse at least three times with highly purified water.
Heat Sterilization: Recommended autoclave sterilization protocol: 20 minutes at 121°C (250 °F)
Reusability: Abutments should never be reused.
Key to codes used
Attention, see instructions for use
Catalogue numberBatch Code
Manufacturer
Internal Hex. Connection
MK-SPZ4 (For internal hex. SP Premolar tooth) MK-WPZ4 (For internal hex. WP Premolar tooth) MK-NIZ6 (For internal hex. NP Incisor tooth) MK-SIZ6 (For internal hex. SP Incisor tooth)
Conical Connection
CK-SPZ4 (For conical connection SP Premolar tooth)CK-WPZ4 (For conical connection WP Premolar tooth)CK-SIZ6 (For conical connection SP Incisor tooth)
Internal Hex. Connection
MK-SPC4 (For internal hex. SP Premolar tooth) MK-WPC4 (For internal hex. WP Premolar tooth) MK-NIC6 (For internal hex. NP Incisor tooth) MK-SIC6 (For internal hex. SP Incisor tooth)
Conical Connection
CK-SPC4 (For conical connection SP Premolar tooth)CK-WPC4 (For conical connection WP Premolar tooth)CK-SIC6 (For conical connection SP Incisor tooth)
Non-Cemented Catalog number index
IncisorIncisor PremolarPremolar
Non-Cemented Pre-Cemented
®
®
User ManualZirconia abutments
MIS Implants Technologies Ltd.P.O.Box 7, Bar Lev Industrial Park, 2015600, ISRAELWebsite: www.mis-implants.com
© MIS Corporation. All rights reserved. MP-UIZI1 Rev. 2, Apr. 2014
EC REP
MIS Implants Technologies GmbHSimeonscarre 2, 32423 Minden, GermanyTel: +49 571-972-7620 Email: [email protected]
Non-Cemented abutments after cementation or pre-cemented abutments should never be put into a ceramic furnace or exposed to temperatures, higher than those normaly found in the oral cavity due to the risk of cement failure ▪ All sharp edges resulting from adjustments must be rounded and polished ▪ Do not reduce wall thickness below 0.5mm ▪ Always use a new, smooth bur with capious amount of water for the adjustment of ZIrconia components.
New product▪ The device is not sterile ▪ Cleaning and sterilization are required prior to first use.
Cleaning, Disinfection and SterilizationAbutments should be cleaned, disinfected, and even sterilized for specific clinical procedures. Effective cleaning and disinfection is an obligatory requirement for effective sterilization.
Cleaning:▪ Clean under running water while brushing both outer and inner surfaces with the appropriate brushes ▪ Immerse in compatible cleaning solutions (e.g. Cidezyme/Enzol, Johnson & Johnson). The cleaner or disinfectant should not contain any of the following ingredients: Strong organic acids or mineral acids, strong lye, organic solvents (e.g. acetone, ether, hexane, benzene), oxidizing agents (e.g. peroxide), halogens (chlorine, iodine, bromine) or aromatic, halogenated hydrocarbons ▪ Post-rinse the implant abutment at least three times with water.
Disinfection: We recommend a high level disinfectant such as, Cidex OPA
(Johnson & Johnson) for disinfection of the implant abutments.▪ Soak the implant abutments in the disinfectant solution for the required time. Refer to the Cidex OPA instructions for use.▪ Remove the implant abutment from the disinfectant solution.▪ Rinse at least three times with highly purified water.
Heat Sterilization: Recommended autoclave sterilization protocol: 20 minutes at 121°C (250 °F)
Reusability: Abutments should never be reused.
Key to codes used
Attention, see instructions for use
Catalogue numberBatch Code
Manufacturer
Internal Hex. Connection
MK-SPZ4 (For internal hex. SP Premolar tooth) MK-WPZ4 (For internal hex. WP Premolar tooth) MK-NIZ6 (For internal hex. NP Incisor tooth) MK-SIZ6 (For internal hex. SP Incisor tooth)
Conical Connection
CK-SPZ4 (For conical connection SP Premolar tooth)CK-WPZ4 (For conical connection WP Premolar tooth)CK-SIZ6 (For conical connection SP Incisor tooth)
Internal Hex. Connection
MK-SPC4 (For internal hex. SP Premolar tooth) MK-WPC4 (For internal hex. WP Premolar tooth) MK-NIC6 (For internal hex. NP Incisor tooth) MK-SIC6 (For internal hex. SP Incisor tooth)
Conical Connection
CK-SPC4 (For conical connection SP Premolar tooth)CK-WPC4 (For conical connection WP Premolar tooth)CK-SIC6 (For conical connection SP Incisor tooth)
Non-Cemented Catalog number index
IncisorIncisor PremolarPremolar
Non-Cemented Pre-Cemented
®
®
User ManualZirconia abutments
MIS Implants Technologies Ltd.P.O.Box 7, Bar Lev Industrial Park, 2015600, ISRAELWebsite: www.mis-implants.com
© MIS Corporation. All rights reserved. MP-UIZI1 Rev. 2, Apr. 2014
EC REP
MIS Implants Technologies GmbHSimeonscarre 2, 32423 Minden, GermanyTel: +49 571-972-7620 Email: [email protected]
Non-Cemented abutments after cementation or pre-cemented abutments should never be put into a ceramic furnace or exposed to temperatures, higher than those normaly found in the oral cavity due to the risk of cement failure ▪ All sharp edges resulting from adjustments must be rounded and polished ▪ Do not reduce wall thickness below 0.5mm ▪ Always use a new, smooth bur with capious amount of water for the adjustment of ZIrconia components.
New product▪ The device is not sterile ▪ Cleaning and sterilization are required prior to first use.
Cleaning, Disinfection and SterilizationAbutments should be cleaned, disinfected, and even sterilized for specific clinical procedures. Effective cleaning and disinfection is an obligatory requirement for effective sterilization.
Cleaning:▪ Clean under running water while brushing both outer and inner surfaces with the appropriate brushes ▪ Immerse in compatible cleaning solutions (e.g. Cidezyme/Enzol, Johnson & Johnson). The cleaner or disinfectant should not contain any of the following ingredients: Strong organic acids or mineral acids, strong lye, organic solvents (e.g. acetone, ether, hexane, benzene), oxidizing agents (e.g. peroxide), halogens (chlorine, iodine, bromine) or aromatic, halogenated hydrocarbons ▪ Post-rinse the implant abutment at least three times with water.
Disinfection: We recommend a high level disinfectant such as, Cidex OPA
(Johnson & Johnson) for disinfection of the implant abutments.▪ Soak the implant abutments in the disinfectant solution for the required time. Refer to the Cidex OPA instructions for use.▪ Remove the implant abutment from the disinfectant solution.▪ Rinse at least three times with highly purified water.
Heat Sterilization: Recommended autoclave sterilization protocol: 20 minutes at 121°C (250 °F)
Reusability: Abutments should never be reused.
Key to codes used
Attention, see instructions for use
Catalogue numberBatch Code
Manufacturer
Internal Hex. Connection
MK-SPZ4 (For internal hex. SP Premolar tooth) MK-WPZ4 (For internal hex. WP Premolar tooth) MK-NIZ6 (For internal hex. NP Incisor tooth) MK-SIZ6 (For internal hex. SP Incisor tooth)
Conical Connection
CK-SPZ4 (For conical connection SP Premolar tooth)CK-WPZ4 (For conical connection WP Premolar tooth)CK-SIZ6 (For conical connection SP Incisor tooth)
Internal Hex. Connection
MK-SPC4 (For internal hex. SP Premolar tooth) MK-WPC4 (For internal hex. WP Premolar tooth) MK-NIC6 (For internal hex. NP Incisor tooth) MK-SIC6 (For internal hex. SP Incisor tooth)
Conical Connection
CK-SPC4 (For conical connection SP Premolar tooth)CK-WPC4 (For conical connection WP Premolar tooth)CK-SIC6 (For conical connection SP Incisor tooth)
Non-Cemented Catalog number index
IncisorIncisor PremolarPremolar
Non-Cemented Pre-Cemented
®
®
User ManualZirconia abutments
MIS Implants Technologies Ltd.P.O.Box 7, Bar Lev Industrial Park, 2015600, ISRAELWebsite: www.mis-implants.com
© MIS Corporation. All rights reserved. MP-UIZI1 Rev. 2, Apr. 2014
EC REP
MIS Implants Technologies GmbHSimeonscarre 2, 32423 Minden, GermanyTel: +49 571-972-7620 Email: [email protected]
Non-Cemented abutments after cementation or pre-cemented abutments should never be put into a ceramic furnace or exposed to temperatures, higher than those normaly found in the oral cavity due to the risk of cement failure ▪ All sharp edges resulting from adjustments must be rounded and polished ▪ Do not reduce wall thickness below 0.5mm ▪ Always use a new, smooth bur with capious amount of water for the adjustment of ZIrconia components.
New product▪ The device is not sterile ▪ Cleaning and sterilization are required prior to first use.
Cleaning, Disinfection and SterilizationAbutments should be cleaned, disinfected, and even sterilized for specific clinical procedures. Effective cleaning and disinfection is an obligatory requirement for effective sterilization.
Cleaning:▪ Clean under running water while brushing both outer and inner surfaces with the appropriate brushes ▪ Immerse in compatible cleaning solutions (e.g. Cidezyme/Enzol, Johnson & Johnson). The cleaner or disinfectant should not contain any of the following ingredients: Strong organic acids or mineral acids, strong lye, organic solvents (e.g. acetone, ether, hexane, benzene), oxidizing agents (e.g. peroxide), halogens (chlorine, iodine, bromine) or aromatic, halogenated hydrocarbons ▪ Post-rinse the implant abutment at least three times with water.
Disinfection: We recommend a high level disinfectant such as, Cidex OPA
(Johnson & Johnson) for disinfection of the implant abutments.▪ Soak the implant abutments in the disinfectant solution for the required time. Refer to the Cidex OPA instructions for use.▪ Remove the implant abutment from the disinfectant solution.▪ Rinse at least three times with highly purified water.
Heat Sterilization: Recommended autoclave sterilization protocol: 20 minutes at 121°C (250 °F)
Reusability: Abutments should never be reused.
Key to codes used
Attention, see instructions for use
Catalogue numberBatch Code
Manufacturer
Internal Hex. Connection
MK-SPZ4 (For internal hex. SP Premolar tooth) MK-WPZ4 (For internal hex. WP Premolar tooth) MK-NIZ6 (For internal hex. NP Incisor tooth) MK-SIZ6 (For internal hex. SP Incisor tooth)
Conical Connection
CK-SPZ4 (For conical connection SP Premolar tooth)CK-WPZ4 (For conical connection WP Premolar tooth)CK-SIZ6 (For conical connection SP Incisor tooth)
Internal Hex. Connection
MK-SPC4 (For internal hex. SP Premolar tooth) MK-WPC4 (For internal hex. WP Premolar tooth) MK-NIC6 (For internal hex. NP Incisor tooth) MK-SIC6 (For internal hex. SP Incisor tooth)
Conical Connection
CK-SPC4 (For conical connection SP Premolar tooth)CK-WPC4 (For conical connection WP Premolar tooth)CK-SIC6 (For conical connection SP Incisor tooth)
Non-Cemented Catalog number index
IncisorIncisor PremolarPremolar
Non-Cemented Pre-Cemented
®
®
User ManualZirconia abutments
MIS Implants Technologies Ltd.P.O.Box 7, Bar Lev Industrial Park, 2015600, ISRAELWebsite: www.mis-implants.com
© MIS Corporation. All rights reserved. MP-UIZI1 Rev. 2, Apr. 2014
EC REP
MIS Implants Technologies GmbHSimeonscarre 2, 32423 Minden, GermanyTel: +49 571-972-7620 Email: [email protected]
Non-Cemented abutments after cementation or pre-cemented abutments should never be put into a ceramic furnace or exposed to temperatures, higher than those normaly found in the oral cavity due to the risk of cement failure ▪ All sharp edges resulting from adjustments must be rounded and polished ▪ Do not reduce wall thickness below 0.5mm ▪ Always use a new, smooth bur with capious amount of water for the adjustment of ZIrconia components.
New product▪ The device is not sterile ▪ Cleaning and sterilization are required prior to first use.
Cleaning, Disinfection and SterilizationAbutments should be cleaned, disinfected, and even sterilized for specific clinical procedures. Effective cleaning and disinfection is an obligatory requirement for effective sterilization.
Cleaning:▪ Clean under running water while brushing both outer and inner surfaces with the appropriate brushes ▪ Immerse in compatible cleaning solutions (e.g. Cidezyme/Enzol, Johnson & Johnson). The cleaner or disinfectant should not contain any of the following ingredients: Strong organic acids or mineral acids, strong lye, organic solvents (e.g. acetone, ether, hexane, benzene), oxidizing agents (e.g. peroxide), halogens (chlorine, iodine, bromine) or aromatic, halogenated hydrocarbons ▪ Post-rinse the implant abutment at least three times with water.
Disinfection: We recommend a high level disinfectant such as, Cidex OPA
(Johnson & Johnson) for disinfection of the implant abutments.▪ Soak the implant abutments in the disinfectant solution for the required time. Refer to the Cidex OPA instructions for use.▪ Remove the implant abutment from the disinfectant solution.▪ Rinse at least three times with highly purified water.
Heat Sterilization: Recommended autoclave sterilization protocol: 20 minutes at 121°C (250 °F)
Reusability: Abutments should never be reused.
Key to codes used
Attention, see instructions for use
Catalogue numberBatch Code
Manufacturer
Internal Hex. Connection
MK-SPZ4 (For internal hex. SP Premolar tooth) MK-WPZ4 (For internal hex. WP Premolar tooth) MK-NIZ6 (For internal hex. NP Incisor tooth) MK-SIZ6 (For internal hex. SP Incisor tooth)
Conical Connection
CK-SPZ4 (For conical connection SP Premolar tooth)CK-WPZ4 (For conical connection WP Premolar tooth)CK-SIZ6 (For conical connection SP Incisor tooth)
Internal Hex. Connection
MK-SPC4 (For internal hex. SP Premolar tooth) MK-WPC4 (For internal hex. WP Premolar tooth) MK-NIC6 (For internal hex. NP Incisor tooth) MK-SIC6 (For internal hex. SP Incisor tooth)
Conical Connection
CK-SPC4 (For conical connection SP Premolar tooth)CK-WPC4 (For conical connection WP Premolar tooth)CK-SIC6 (For conical connection SP Incisor tooth)
Non-Cemented Catalog number index
IncisorIncisor PremolarPremolar
Non-Cemented Pre-Cemented