  • BasicOperationProcedure

    TurningONtheMachine1.)TurnOntheAVR2.)TurnOntheUPS3.)TurnOntheMachineNote:AttheleftBackportionthemachine(leftside,facinginfrontofthemachine),youcanfindboththekeyactivatedswitchandasmallseesawswitch.TurnontheSeesawswitchtoPowerontheMachine.Donotremoveoralterthekeypositiontopreventunnecessaryproblems.4.)TurnOntheComputer

  • BasicOperation,PerformingAnalysisanddatagathering

    1.) AftertheComputerfinishedbootinginWindows,opentheroutineDialoglocatedattheXLabProCommunicationServer>RoutineorlookfortheSPXEUserDlg.exeshortcutlocatedintheDesktop.

    2.) Doubleclicktheshortcutlinkoftheroutinedialogtoopen,theroutinedialogwindowwillpopup.

    XLabProLogInwillshow,useSupervisorasLoginNameandsaixlabproasthepassword

  • 3.) AfteraSuccessfulLogin,anoptionwindowwillshow.ClickYestoenterthemachineinStandbyMode.

    4.) OncetheMachineisinStandbyMode(TheXrayTubeisinStandbyONPosition),youarenow

    readytostartanalyzingyoursamples.Note:Seebasicfunctionsandfieldsbelow:

  • 5.)ToAnalyzeasample,followthestepsbelow: *KeyintheSampleNameintheNameFieldoftheRoutineDialogWindow *SelecttheSpecificMethodforAnalysisbyclickingthedropdownwindowofthemethodfield Note:AvailableMethods:

    1.) NiFeCo04292012=forNickel,IronandCobaltAnalysisOnly2.) Complete04292012=ForCompleteAnalysisofNiOres

    Note:Ni,Fe,Co,P,S,Al203,SiO2,MgO,CaO,MnOandCr2O3aretheelementsanalyzedinCompletemethod

    3.) ChromiteOres=forCr2O3andFeAnalysisonly4.) MCA=forMCARecalibration5.) Global=forGlobalRecalibration6.) Turboquantpellets=FundamentalParametersmethodusedforunknown

    samplesinpelletform.*JobFolderCreationundertheJobFieldintheroutineDialog,justkeyinafoldername/jobcodeforaspecificsamplecodes(likeFSandShipmentsamples)Note:leavingtheJobfieldempty,thesoftwarewillautomaticallyusethedefaultfolderbydate.Allanalysiswillsortedandgroupedinacertainfolderbydateofanalysis

    6.)Afterfillingoutthesampleparameterfields,clicktherunbuttonlocatedatthelowerleftcorneroftheroutineDialog.*AwarningwindowwillshowThesamplewillbemeasuredinVacuum,wouldyouliketocontinue,clickyestoproceed*seeRoutineDialogsampleplatecolorcodesbelow:

    Gray=vacant Red=withsampleandwaitingforAnalysis Yellow=AnalysisOngoing Green=Analysisdone Blue=positionwashighlighted

  • 7.)Aftertheanalysisofallsamples,youcanviewtheresultsin3differentways a.)Firstbyrightclickingthesamplenameandclickingtheshowresultoption

  • b.)byprintingtheresultsintoanexcelfile,youcanclickdirectlytheprintershortcutbuttonbellowthemainmenuoryoucanusetheCrtl+Pfunctiontocallouttheprintingoptionfunction

  • c.)YoucanviewtheresultsintheJobManager,toaccesstheJobManager,simplyclicktheJobManagerShortcutbelowthemainmenuorclickXLabPro>JobManagerorrightclicktheCommunicationServerlocatedinthetaskbarandclickJobs.

  • TurningOfftheMachineNote:TurningOfftheMachineisjustthesameasturningOnbutinOppositeOrder.SeecorrectturningOffproceduresbelow:1.) ClicktheStandbyOffbuttonlocatedinthelowerleftcorneroftheroutinedialogtoturnoff

    theXrayTube2.) ClosetheroutinedialogandShutdowntheComputer3.) TurnOFFthemachineusingtheseesawpowerswitchlocatedintherightbackcornerofthe

    machine.4.) TurnOfftheUPSbypressingthePowerOFFbuttonforfewseconds5.) TurnOFFtheAVR6.) LastlyturnOffthemainSwitch

    SafetyPrecautionsforlongermachinelifespan:

    1.) Donotturnoffthemachineasmuchaspossibleevenduringlongbreaks*Duringlongbreaksandweekends,youcanstandbyOfftheXrayTubeandTurnofftheComputerbutleavethemachineandanAirconditioningunitrunningallthetime.

    2.)MonitorbothElectronicNoiseandVacuumLevelduringanalysisdailyandlogthedata.3.)Duringlongpowerfailurewithoutbackuppowergenerator,standbyofftheXrayTubeand shutdownthecomputer.Themachineshouldbeleftrunningunderthebatterybackuppower oftheUPS.4.)Duringlongpowerfailure,backupgeneratorshouldbereadilyavailableandrunningafter10~15 minutes.5.)ReportimmediatelyanyunusualmachinedatalikeElectronicNoiseXrayTubesActualVoltage andCurrentvaluesonbothstandbyandrunningconditions.Shouldyouhaveanyconcernregardingthemachine,pleaseletusknow.Contactusat :027711248(ManilaOffice)lookforOdensePirante :0325207881(CebuOffice)lookforSherwinMinoza/JasonBajenting :+639263019023(myhandphone)[email protected],[email protected]@qesnet.comPreparedby:

    JasonA.Bajenting AssistantServiceEngineer QESManila,Inc.(CebuOffice)