1 X-MET 880 FIELD PORTABLE X-RAY FLOURESCENCE OPERATING PROCEDURES SOP#: 1707 DATE: 12/22/94 REV. #: 0.0 1.0 SCOPE AND APPLICATION The purpose of this standard operating procedure (SOP) is to serve as a guide to the start-up, check-out, operation, calibration, and routine use of the X-Met 880 instrument, for field use in screening of hazardous or potentially hazardous inorganics. It is not intended to replace or diminish the use of the X-MET 880 Operating Instructions. The Operating Instructions contain additional helpful information to assist in the optimum instrument utilization and which form the basis on which new and varying applications can later be based. The procedures contained herein are general operating procedures which may be changed as required, dependent on site conditions, equipment limitations, limitations imposed by the QA/QC procedure or other protocol limitations. In all instances, the ultimate procedures employed should be documented and associated with the final report. 1.1 Principles of Operation X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the chemical composition of samples. In a source excited XRF analysis, primary X-rays emitted from a sealed radioisotope source are utilized to irradiate samples. During interaction of the source X-rays with samples, they may either undergo scattering (dominating process) or absorption by sample atoms in a process known as the photoelectric effect (absorption coefficient). This most useful analytical phenomenon originates when incident radiation knocks out an electron from the innermost shell of an atom. The atom is excited and releases its surplus energy almost instantly by filling the vacancy created with an electron from one of the higher energy shells. This rearrangement of electrons is associated with emission of X-rays characteristi c (in terms of energy) of the given atom. This process is referred to as emission of fluorescent X-rays (fluorescent yield). The overall efficiency of the process described is referred to as excitation efficiency and is proportional to the product of absorption coefficient and fluorescent yield. Generally, the X-MET 880 utilizes characteristic X-ray lines originating from the innermost shells of the atoms, K, L, and occasionally M. The characteristic X-ray lines of the K series are the most energetic lines for any element and, therefore, are the preferred analytical lines. The K lines are always accompanied by the L and M lines of the same element. However, being of much lower energy than the K lines, they can usually be neglected for those elements for which the K lines are analytically useful. For heavy elements (such as Ce, atomic number (Z)=58, to U, Z=92), the L lines are the preferred lines for X-MET 880 analysis. The L and L lines have " $ almost equal intensities and the choice of one or the other depends on what interfering lines might be present. A source just energetic enough to excite the L lines will not excite the K lines of the same element. The M lines will appear together with the L lines. The X-MET 880 Operating Instructions contain tables that show the energies and relative intensities of the primary characteristic X-ray lines for all the applicable elements, see Section 1.3. An X-ray source can excite characteristic X-rays from an element only if the source energy is greater tha n the absorption edge energy for the particular line group (i.e., K absorption edge, L absorption edge, M absorption edge) of the element. The absorption edge energy is somewhat greater than the corresponding line energy. Actually, the K absorption edge energy is approximately the sum of the K, L and M line energies, and the L absorption edge energy is approximately the sum of the L and M line energies of the particular element. Energies of the characteristic, fluorescent X-rays are

X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

  • Upload
    others

  • View
    16

  • Download
    1

Embed Size (px)

Citation preview

Page 1: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

1

X-MET 880 FIELD PORTABLE™

X-RAY FLOURESCENCEOPERATING PROCEDURES

SOP#: 1707DATE: 12/22/94

REV. #: 0.0

1.0 SCOPE AND APPLICATION

The purpose of this standard operating procedure(SOP) is to serve as a guide to the start-up, check-out,operation, calibration, and routine use of the X-Met880 instrument, for field use in screening of hazardousor potentially hazardous inorganics. It is not intendedto replace or diminish the use of the X-MET 880Operating Instructions. The Operating Instructionscontain additional helpful information to assist in theoptimum instrument utilization and which form thebasis on which new and varying applications can laterbe based.

The procedures contained herein are general operatingprocedures which may be changed as required,dependent on site conditions, equipment limitations,limitations imposed by the QA/QC procedure or otherprotocol limitations. In all instances, the ultimateprocedures employed should be documented andassociated with the final report.

1.1 Principles of Operation

X-Ray Fluorescence Spectroscopy (XRF) is anondestructive qualitative and quantitative analyticaltechnique used to determine the chemical compositionof samples. In a source excited XRF analysis,primary X-rays emitted from a sealed radioisotopesource are utilized to irradiate samples. Duringinteraction of the source X-rays with samples, theymay either undergo scattering (dominating process) orabsorption by sample atoms in a process known as thephotoelectric effect (absorption coefficient). Thismost useful analytical phenomenon originates whenincident radiation knocks out an electron from theinnermost shell of an atom. The atom is excited andreleases its surplus energy almost instantly by fillingthe vacancy created with an electron from one of thehigher energy shells. This rearrangement of electronsis associated with emission of X-rays characteristic(in terms of energy) of the given atom. This processis referred to as emission of fluorescent X-rays

(fluorescent yield). The overall efficiency of theprocess described is referred to as excitationefficiency and is proportional to the product ofabsorption coefficient and fluorescent yield.

Generally, the X-MET 880 utilizes characteristicX-ray lines originating from the innermost shells ofthe atoms, K, L, and occasionally M. Thecharacteristic X-ray lines of the K series are the mostenergetic lines for any element and, therefore, are thepreferred analytical lines. The K lines are alwaysaccompanied by the L and M lines of the sameelement. However, being of much lower energy thanthe K lines, they can usually be neglected for thoseelements for which the K lines are analytically useful.For heavy elements (such as Ce, atomic number(Z)=58, to U, Z=92), the L lines are the preferred linesfor X-MET 880 analysis. The L and L lines have" $

almost equal intensities and the choice of one or theother depends on what interfering lines might bepresent. A source just energetic enough to excite theL lines will not excite the K lines of the sameelement. The M lines will appear together with theL lines.

The X-MET 880 Operating Instructions contain tablesthat show the energies and relative intensities of theprimary characteristic X-ray lines for all theapplicable elements, see Section 1.3.

An X-ray source can excite characteristic X-rays froman element only if the source energy is greater thanthe absorption edge energy for the particular linegroup (i.e., K absorption edge, L absorption edge, Mabsorption edge) of the element. The absorption edgeenergy is somewhat greater than the correspondingline energy. Actually, the K absorption edge energyis approximately the sum of the K, L and M lineenergies, and the L absorption edge energy isapproximately the sum of the L and M line energies ofthe particular element.

Energies of the characteristic, fluorescent X-rays are

Page 2: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

2

converted (within the detector) into a train of electricpulses, the amplitudes of which are linearlyproportional to the energy. An electronicmultichannel analyzer (electronic unit) measures thepulse amplitudes which, since they are proportional tooriginal energies of emitted characteristic X-rays, arethe basis of a qualitative X-ray analysis. The numberof equivalent counts at a given energy isrepresentative of element concentration in a samplebasis for quantitative analysis.

1.1.1 Scattered X-rays

The source radiation is scattered from the sample by 1.17.two physical reactions: coherent or elastic scattering(no energy loss) and Compton or inelastic scattering For measurement, the sample is positioned in front of(small energy loss). Thus, the backscatter the source-detector window and exposed to the(background signal) actually consists of two primary (source) X-rays by pulling a trigger on thecomponents with X-ray lines close together, the probe (or pushing the top of the probe unit back on thehigher energy line being equal to the source energy. sample type probe) which exposes the sample toSince the whole sample takes part in scattering, the radiation from the source. The sample fluorescent andscattered X-rays usually yield the most intense lines backscattered X-rays enter through the beryllium (Be)in the spectrum. It is also obvious from the detector window and are detected in the active volumeaforementioned that the scattered X-rays have the of a high-resolution, gas-filled proportional counter. highest energies in the spectrum and contribute themost part of the total measured intensity signal. Elemental count rates (number of net element pulses

1.2 Sample Types

Solid and liquid samples can be analyzed for elementsAl (aluminum) through U (uranium) with proper X-ray source selection. Typical environmentalapplications are:

C Heavy metals in soil (in-situ or samplescollected from the surface or from bore holedrillings, etc.), sludges, and liquids (i.e., Pbin gasoline)

C Light elements in liquids (i.e., P, S and Cl inorganic solutions)

C Heavy metals in industrial waste streameffluents

C PCB in transformer oil by Cl analysis

C Heavy metal air particulates collected onmembrane filters, either from personnelsamplers or from high volume samplers.

2.0 METHOD SUMMARY

The X-MET 880 Portable XRF Analyzer employsradioactive isotopes, such as Fe-55, Cm-244, Cd-109and Am-241 for the production of primary X-rays.Each source emits a specific energy range of primaryX-rays that cause a corresponding range of elementsin a sample to produce fluorescent X-rays. Whenmore than one source can excite the elements ofinterest, the appropriate source(s) is selectedaccording to its excitation efficiency for the elementsof interest. See X-MET 880 Operating Instructions fora chart of source type versus element range, Section

per second) are used in correlation with actual samplecompositions to generate calibration models forqualitative and quantitative measurements.

Analysis time is user selectable from 1 to 32767seconds. The shorter measurement times (30 - 100s)are generally used for initial screening and hot spotdelineation, while longer measurement times (100 -500s) are typically used for higher precision andaccuracy requirements.

3.0 SAMPLE PRESERVATION,CONTAINERS, HANDLING ANDSTORAGE

This SOP specifically describes equipment operatingprocedures for the X-MET 880; hence, this section isnot applicable to this SOP.

4.0 INTERFERENCES ANDPOTENTIAL PROBLEMS

The total error of XRF analysis is defined as thesquare root of the sum of squares of both instrumentprecision and user or application related error.

Page 3: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

3

Generally, the instrument precision is the leastsignificant source of error in XRF analysis. User orapplication related error is generally the moresignificant source of error and will vary with each siteand method used. The components of the user orapplication related error are the following:

4.1 Sample Placement

This is a potential source of error since the X-raysignal decreases as you increase the distance from theradioactive source. However, this error is minimizedby maintaining the same sample distance from thesource.

4.2 Sample Representivity

This can be a major source of error if the sampleand/or the site-specific or site-typical calibrationsamples (see Section 4.0) are not representative of thesite. Representivity is affected by the soil macro- andmicro-heterogencity. For example, a sitecontaminated with pieces of slag dumped about by asmelting operation will be less homogeneous than asite contaminated by liquid plating waste. This errorcan be minimized by either homogenizing a largevolume of sample prior to analyzing an aliquot, or byanalyzing several samples (in-situ) at each samplingpoint and averaging the results.

4.3 Reference Analysis

Soil chemical and physical matrix effects may becorrected by using Inductively-Coupled Plasma (ICP)or Atomic Absorption (AA) spectrometer analyzedsite-specific soil samples as calibration samples. Amajor source of error can result if the samplesanalyzed are not representative of the site and/or theanalytical error is large. With XRF calibrations basedon reference analyses results, the XRF analyticalresults can be reported in the same units as thecalibration samples reference analyses. Results, forexample, will be in Contract Laboratory Protocol(CLP) extractable metals if the CLP specifiedHNO /H O digestion is used. Results will be in total3 2 2

metals if total (HF) digestion or KOH fusion is used.

4.4 Chemical Matrix Effects (Effectsdue to the chemical composition ofthe sample)

Chemical matrix effects result from differences inconcentrations of interfering elements. These effectsappear as either spectral interferences (peak overlaps)or as X-ray absorption/enhancement phenomena.Both effects are common in soils contaminated withheavy metals. For example, Fe (iron) tends to absorbCu (copper) K shell X-rays, reducing the intensity ofCu measured by the detector. This effect can becorrected if the relationship between Fe absorptionand the Cu X-ray intensities can be modeledmathematically. Obviously, establishing all chemicalmatrix relationships during the time of instrumentcalibration is critical. These relationships aremodeled mathematically, with X-MET 880 internalsoftware, using ICP or AA analyzed site-specific soilsamples as the XRF calibration standards.Additionally, increasing the number of standards andthe range of the standard concentrations used maydecrease the error in the calibration mathematicalmodeling. Generally, as rule-of-thumb, a minimum offive calibration samples per element to be analyzedare used to generate reliable X-MET 880 calibrationmodels.

4.5 Physical Matrix Effects (Effects dueto sample morphology)

Physical matrix effects are the result of variations inthe physical character of the sample. They mayinclude such parameters as particle size, uniformity,homogeneity and surface condition. For example,consider a sample in which the analyte exists in theform of very fine particles within a matrix composedof much courser material. If two separate aliquots ofthe sample are ground in such a way that the matrixparticles in one are much larger than in the other, thenthe relative volume of analyte occupied by theanalyte-containing particles will be different in each.When measured, a larger amount of the analyte willbe exposed to the source X-rays in the samplecontaining finer matrix particles; this results in ahigher intensity reading for that sample and,consequently, in apparently higher measuredconcentration of that element.

Page 4: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

4

4.6 Modeling Error

The error in the calibration mathematical modeling isinsignificant (relative to the other sources of error) IFthe instrument's modeling operating instructions arefollowed correctly (see Section 14.0).

4.7 Moisture Content

If measurement of soils or sludges is intended, thesample moisture content will affect the accuracy ofthe analysis. The overall error from moisture may bea secondary source of error when the moisture rangeis small (5-20%), or may be a major source of errorwhen measuring on the surface of soils that aresaturated with water.

4.8 Cases of Severe X-ray SpectrumOverlaps

Certain X-ray lines from different elements, whenpresent in the sample, can be very close in energy and,therefore, interfere by producing a severelyoverlapped spectrum.

The typical spectral overlaps are caused by the K line$

of element Z-1 (or as with heavier elements, Z-2 or Z-3) overlapping with the K line of element Z line."

This is the so-called K K interference. Since the"/ $

K :K intensity ratio for the given element usually" $

varies from 5:1 to 7:1, the interfering element, (Z-1),must be present in large concentrations in order todisturb the measurement of analyte Z. The presenceof large concentrations of titanium (Ti) could disturbthe measurement of chromium (Cr). The TiK and K" $

energies are 4.51 and 4.93 Kev, respectively. TheCrK energy is 5.41 Kev. The resolution of the"

detector is approximately 850 eV. Therefore, largeamounts of Ti in a sample will result in spectraloverlap of the TiK with the CrK peak.$ "

Other interferences are K/L, K/M and L/M. Whilethese are less common, the following are examples ofa severe overlap:

AsK PbL , SK PbM"/ " "/ "

In the arsenic/lead case, lead can be measured fromthe PbL line, and arsenic from either the AsK or the$ "

AsK line; this way the unwanted interference can be$

corrected. However, due to the severeness of theoverlap (energy of AsK is almost identical to that of"

PbL ), measurement sensitivity is reduced."

4.9 Inappropriate Pure ElementCalibration

It is of paramount importance that the pure elementcalibration, also called "instrument calibration" (seeSection 12.0), include all elements that can be presentat the site, (i.e., in the samples to be analyzed). Thismeans that even if the element is not a target element,as long as it is present in detectable amounts with thesource in use, it must be included in the pure elementcalibration in order for the X-MET 880 to correct forits potential spectral interference effect on the targetelement.

5.0 EQUIPMENT/APPARATUS

5.1 Description of the X-Met 880System

The X-MET 880 analyzer includes a compact, sealedradiation source contained that is in a measuringprobe. This probe is connected by cable to anenvironmentally sealed electronic module.

The analyzer utilizes the method of Energy DispersiveX-Ray Fluorescence (EDXRF) spectrometry todetermine the elemental composition of soils, sludges,aqueous solutions, oils, and other waste materials.

Each probe is equipped with a high resolution gas-filled proportional detector, or a high resolution solidstate lithium drifted silicon (Si/Li) detector.

The complete X-MET 880 system consists of fivealternate configurations.

1. The X-MET 880ES Extended Range X-MET880 Silicon Detector System includes theSilicon Detector Surface Probe System(SDPS) with a 60 millicurie (mCi) Curium-244* (Cm-244) excitation source and a 30mCi Americium-241** (Am-241) excitationsource.

2. The X-MET 880SH, Standard X-MET 880System comes with a SAPS probe containinga 60 mCi Cm-244* excitation source.

3. The X-MET 880ER Extended Range X-MET

Page 5: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

5

880 System comes with a DOPS probe that The electronic module includes a 256 channel multi-contains a 60 mCi Cm-244* excitation channel analyzer and a high speed, 16/32 bit,source and a 30 mCi Am-241** excitation Motorola 68000 microprocessor. Up to 32 multi-source. element analysis programs, called models, can be

4. The X-MET 880AS Alternate StandardRange X-MET 880 System comes with a The unit comes factory pre-calibrated based on theSAPS probe containing 10 mCi Cd-109*** Outokumpu Electronics synthetic soil calibration(in place of the 60 mCi Cm-244). suite, or based on customer supplied standards (see

5. The X-MET 880AE Alternate Extended calibration data specific to each X-MET 880).Range X-MET 880 System includes theDOPS with 10 mCi Cd-109*** plus 30 mCi Optional calibrations can be installed at the factory forAm-241**. other soil or waste material types on request.

* Cm-244 in SDPS probe (or Surface Analysis Additional models tailored to specific needs may beProbe Set (SAPS) and Double Source added by the user after attending the X-MET 880Surface Probe Set (DOPS) probes) will allow Calibration and Operators Training Course which isanalysis of all elements from atomic number conducted by Ouotkumpu Electronics at regular19 (potassium) to 35 (bromine) and from intervals. (Request course description and scheduleatomic number 56 (barium) to atomic number from Outokumpu Electronics, Langhorne, PA.)83 (bismuth).

** The Am-241 excitation source extends the 220-volt) wall outlet or its 12-volt, 10-hour capacityelemental range of the system to include such battery, or a standard 12-volt car or truck battery.important priority pollutants as Cd, Ag, andBa. Am-241 in SDPS, SAPS or DOPS will The X-MET 880 can be operated in a temperatureallow analysis of all elements from atomic range from 32 to 140 Fahrenheit (F), or may benumber 30 (zinc) to 60 (neodymium), and operated down to -13 F with the low temperatureatomic number 73 (tantalum) to 92 option. The freezing point of a discharged battery is(uranium). 14 F.

*** Replacing the Cm-244 with Cd-109 provides The probe and electronic unit may be exposed to asomewhat improved precision and accuracy light rain. However, additional protection is providedfor Pb when As is present. Cd-109 in SAPS if the system (electronic unit and probe) is containedor DOPS will allow analysis of elements in the optional water repellant carrying case.from atomic number 24 (chromium) to 42(molybdenum) and from 65 (terbium) to 92 The instrument can be calibrated for up to 10 elements(uranium). per model, six (6 target elements) of which can

Optional sample type probes (for laboratory or mobilelab use only) are available for use when all samples In the Assay Mode, up to 30 reference samples perwill be contained in X-ray cups. These probes can assay model can be used to generate the samplecontain any of the excitation sources described above. calibration curve in the X-MET 880.

The use of the special optional sample probes, LightElement sample Probe System (LEPS), or the SurfaceLight element Probe System (SLPS), with the Fe-55ring source, enables analysis of light elements rangingfrom atomic number 13 (aluminum), to atomic number24 (chromium), and heavy elements from 37(rubidium) to 56 (barium).

stored in its memory.

supplemental documentation on factory calibration for

The X-MET 880 can be operated from a 115-volt (or

0

0

0

provide a readout in the Assay Mode.

5.2 Equipment and Apparatus List

5.2.1 X-MET 880 Analyzer System

The complete X-MET 880 Analyzer System includes:

C 880 Electronics Module

Page 6: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

6

C Single source SAPS or DOPS or SDPS with 1. The probe should always be in contact withoptional LEPS or optional Heavy Element the surface of the material being analyzedSample Probe Set (HEPS) in place of, or in and the analyzed material should completelyaddition to, the SAPS and/or DOPS, each cover the probe opening (aperture), when thecontaining appropriate excitation source(s) probe shutter is open. The indicator flag on

C Pure element standards green when the shutter is closed and red

C Battery charger

C Battery pack pointed at the operator or surrounding

C X-MET 880 Operating Instructions, X-MET880 Operator's Manual and any applicable 3. Do not place any part of the operator's or co-X-MET 880 factory calibration worker's body in line of exposure when thedocumentation. shutter is opened and not fully covered.

5.2.2 Optional Items

C 31 millimeter (mm) diameter sample cups

C XRF polypropylene film, 0.2mm thickness

C Nylon reinforced, water-repellant backpack

C Metal reinforced shipping case with die-cutfoam inserts for X-MET 880 and accessories

C Peripheral devices such as theTerminal/Printer data logger, the DMS-1Data Management System, the ESP extendedsoftware package for use with IBMcompatible Personal Computer (PC).

C Surface probe shield assembly. Shieldassembly must be used when the SAPS orDOPS probes are inverted for measuringsample in XRF cups.

See Outokumpu Electronics X-MET 880 AccessoriesPrice List for additional Outokumpu Electronicsoptions.

For mobile lab or laboratory X-ray sample preparationaccessories, such as drying ovens, grinders, sieves,etc., consult general laboratory equipment suppliers .

5.2.3 Limits and Precautions

The probes should be handled in accordance with thefollowing radiological control practices:

the side of the DOPS and SAPS probes is

when it is open.

2. Under no circumstances should the probe be

personnel with the shutter open.

4. The SAPS or DOPS probe trigger must bekey-locked when not in use.

5. Notify Outokumpu Electronics immediatelyof any condition or concern relative to theprobe structural integrity, source shielding,shutter condition or operability.

6. Notify the appropriate state agency orNuclear Regulatory Commission (NRC)office (see factory supplied data onradiological safety) immediately of anydamage to the radioactive source, or any lossor theft of the device.

7. Labels or instructions on the probe(s) mustnot be altered or removed.

8. The user must not remove the probe coversor attempt to open the probe.

9. The source(s) in the probe must be leaktested every six (6) months as described inthe X-MET 880 Operating Instructions. TheLeak Test Certificates must be kept on fileand a copy must accompany the instrumentat all times.

10. The probe shield assembly must be usedwhen the SAPS or DOPS probe is invertedfor measuring samples contained in cups.

11. During operation, keep the probe at least 10feet from computer monitors and any othersource of radio frequency (RF). Some

Page 7: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

7

monitors have very poor RF shielding and If the receiving device, serial port, does not have a 25will affect measurement results. pin, male, RS-232 C standard connector, THEN it will

12. The X-MET 880 should not be dropped or to-female converter), OR, in the case of 9 pin deviceexposed to conditions of excessive shock or connectors, a 25 to 9 pin adapter (with or vibration.

13. Keep the force on the probe, with the trigger the connector at the receiving device). Allpulled, to less than four (4) pounds to avoid Outokumpu supplied peripherals are delivered withshutter binding. appropriate connections.

Additional precautions include:

1. Do not pull on the probe wire to unplug theprobe. Grasp the probe plug at the ribbedrubber connector cover and squeeze, thenpress firmly while plugging, and pull whileunplugging the connector.

2. Do not attempt to rotate the handle on theelectronic unit unless the release buttons oneach side of the handle are depressed.

3. The X-MET 880 should not be operated orstored at an ambient temperature below 32 F0

(-13 F with low temperature version) or0

above 140 F. 0

4. The battery charging unit should only beused indoors at dry conditions.

5.3 Peripheral Devices

The X-MET 880 may be used with a wide range ofperipheral devices for electronic data capture orprinted readout as long as they are equipped withinput compatible with the RS-232 serial protocol.Such devices include terminals, printers, electronicdata loggers, personal computers, etc. Any time aperipheral device is connected to the X-MET 880, alltext and commands shown on the X-MET 880 displaywill be automatically output to and copied by theperipheral.

5.3.1 Communication Cable Connection

Plug the round connector of the RS-232 cable, into the * 1 stop bit (MARK)X-MET 880 "IN/OUT" port (the connection justabove the probe connection on the electronic unit) ,and the rectangular (25 pin) connector of the cable tothe RS-232 port of the receiving device (serial port).

be necessary to obtain a "gender mender plug" (male-

without gender changer, depending on the gender of

5.3.2 Communication Speed

To communicate with the external device, the X-MET880 MUST be set at the same Baud Rate as thereceiving device. The X-MET 880 command forsetting or resetting the Baud Rate is CSI (ConfigureSerial Interface). The CSI command is a sub-command under the EMP (Enter MaintenanceParameters) command, which must therefore precedeit. Enter EMP followed by "CONT/YES", THENenter CSI on the keypad FOLLOWED by"CONT/YES". The X-MET 880 will display

BAUD RATE: XXXX NEW?

Press "CONT/YES" until the desired baud rate (datatransfer speed) is in the display, then press "END/NO"to accept the displayed reading. The baud rate canhave values 50, 75, 110, 134.5, 150, 200, 300, 600,1200, 1800, 2400, 4800, 9600 and 19200 baud. Selectthe baud rate of the peripheral device with which youare communicating.

5.3.3 Character Set

The RS-232 C interface supports all the standardASCII characters. Upper and lower case letters areequivalent in data transmission. This means that theX-MET 880 will execute any legal command typed inlower case on the peripheral keyboard. (See Section5.3.5 for some special keys.) The serial data formatis:

* 1 start bit (SPACE)* 8 data bits (ASCII from LSB to MSB)

5.3.4 Parity

Parity MUST be set to NO PARITY.

Page 8: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

8

5.3.5 Terminal Emulation 5.4.2 Further Information and

Although most keys on the standard typewriter stylekeypad (as used on most terminals and computers) arethe same as the X-MET 880 keypad, there are,however, some specific keys on the X-MET 880 thatrequire the operator to use an "equivalent" key on theterminal. The following lists correlates the unique X-MET 880 keys and their terminal equivalents:

X-MET 880 KEY TERMINAL KEY (or keycombination)

CONT/YES ENTER (or RETURN onsome devices)

END/NO ESCMODEL CTRL DMTIME CTRL TRECALC CTRL R< SHIFT <^ SHIFT ^ START CTRL A

5.3.6 User Software

Refer to your PC software manual for details onadditional settings that may be required for properinterfacing between X-MET 880 and your particularsoftware.

5.4 Instrument Maintenance

5.4.1 Probe Window

Should the probe window become damaged orpunctured, it should be replaced as soon as possible toprevent dust and moisture from entering the probe.Replacement window assemblies can be ordered fromOutokumpu Electronics. Simply pry out the oldwindow aperture using a small, flat-blade screwdriver,or similar. PRIOR to reinstalling the new aperture,rub a very thin film of silicon grease (or liquid soap,IF silicon grease is not available) around the rubber"O" ring, inside the probe aperture opening, tofacilitate re-installation.

The removal of any loose dirt on the probe windowshould be done with a soft brush or cloth and thenblown with air. To remove adhering dirt, a solventsuch as methanol or ethanol may be used.

Troubleshooting

Refer to the X-MET 880 Operating Instructions (oroptional Maintenance Manual) for additional detailedoperational and/or maintenance and troubleshootinginstructions. IF no solution is not found in eithermanual, THEN contact Outokumpu Electronics forassistance.

It is recommended that an Instrument Log bemaintained for documenting specific correctiveactions taken to alleviate any instrumental problems,or to record any service that has been performed.

6.0 REAGENTS

6.1 Site-Specific Calibration Standards(SSCS)

The SSCS must be representative of the matrix to beanalyzed by the XRF. They are employed in the"sample calibration" stage of programming theinstrument (see Section 7.5.6). They are alsoemployed in the subsequent calibration checks (seeSection 7.2 and 9.0) and any re-calibrations that maybe performed.

The concentration of the target elements in the SSCSshould be determined by independent AA or ICPanalysis that meets an acceptable quality for refereedata.

Additionally, the concentration in the SSCS ofelements adjacent (+/-2 atomic numbers Z) to the Z ofthe target element should be determined byindependent AA or ICP analysis if: 1) they areexcited by the source used, and/or 2) theirconcentrations are unknown or suspected to be greaterthan ten percent of the target element concentrationand/or 3) it is unknown or suspected that theirconcentration variance is greater than twenty percentin the site matrix, or if this variance (if greater thantwenty percent) has a non-linear relationship to thevariance of the target element concentration.

For example, the requested target elements are Cd andSb for a site. Review of the site history indicates thatSn and Ag may be present. The SSCS should beanalyzed for Sn and Ag in addition to Cd and Sb, todetermine their concentrations and the relationship

Page 9: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

9

(linear or non-linear) to the Cd and Sb concentrations with a greater content of non-representative materialin the SSCS samples. such as rocks and/or organic debris. Standard glass

6.1.1 SSCS Sampling

Review Section 4.2 on sample representivity. TheSSCS samples must be representative of the matrix to The SSCS samples should be dried either by airbe analyzed by XRF. It does not make sense to drying overnight, or oven drying at less than 105EC.collect the SSCS samples in the site containment area Aluminum drying pans or large plastic weighing boatsif you are interested in investigating off-site for air drying may be used. After drying, remove allcontaminant migration. The matrices may be different large organic debris and non-representative materialand could affect the accuracy of the XRF results. I f (twigs, leaves, roots, insects, asphalt, rocks, etc.).there are two different matrices on-site, collect twosets of SSCS samples. The sample should be sieved through a 20-mesh

A full range of target element concentrations is should be broken up against the sieve using a stainlessneeded to provide a representative calibration curve. steel spoon. Pebbles and organic matter remaining inMixing high and low concentration soils to provide a the sieve should be discarded. The under-sievefull range of target element concentrations is not fraction of the material constitutes a sample. recommended due to homogenization problems. Thehighest and lowest SSCS samples will determine the Although a maximum final particle size of 20-mesh islinear calibration range. Unlike liquid samples, solid normally recommended, a smaller particle size may besamples cannot be diluted and re-analyzed. desired (see Section 4.5). The sample should be

The number of SSCS samples needed for calibrating and physically mixing opposite quarters with a cleanan assay model depends on: stainless steel spoon. Re-composite and then repeat

1. The number of target elements (analyte). For Place the sieved sample in a clean glass sample jareach additional target element, increase the and label it using both the site name and samplenumber of SSCS samples by five (up to a identification information.maximum of 30).

2. The number of elements adjacent to the using soap and water and dried between samples.target elements. For each additional adjacentelement known or found to be present in the One or more plastic XRF sample cups should be filledsamples, you should increase the number of with the sieved soil for each SSCS sample. A piece ofSSCS samples by five (up to a maximum of .2mm polypropylene film is cut and tensioned,30) to insure that the calibration model wrinkle-free, over the top of the x-ray sample cup andproperly corrects for X-ray then sealed using the plastic film securing ring. The

interferences and spectral cup should be labeled using both the site name andoverlaps. specimen identification information.

Additionally, collect several SSCS samples in the Either the XRF sample cup or the balance of theconcentration range of interest. If the action level of prepared sample is submitted to the approvedthe site is 500 mg/kg, providing several SSCS samples laboratory for analysis of the requested element(s) byin this concentration range will tend to improve the AA or ICP.XRF analytical accuracy in this concentration range.

Generally, a minimum of 10 and a maximum of 30appropriate SSCS samples should be taken. Aminimum of a 4 oz. sample is required. A larger sizesample should be provided to compensate for samples

sampling jars should be used.

6.1.2 SSCS Preparation

stainless steel sieve. Clumps of soil and sludge

homogenized by dividing the sieved soil into quarters

the quartering and mixing procedure three times.

The stainless steel sieves should be decontaminated

6.2 Site Typical Calibration Standards(STCS)

When the goal of the analysis with X-MET 880 issemi-quantitative measurements, such as hot spot

Page 10: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

10

delineation or determination of sampling points for a GAIN CONTROL: COUNT RATE TOO LOW SSCS, then use of a STCS may be the mostappropriate method. STCS are SSCS from a different will flash intermittently (along with a "beep") on thesite that have the identical target elements in a similar display. This is normal and will stop approximatelyrange and matrix as the site that is to be analyzed. It 30 seconds after power is on.should be noted that the STCS are not from the site tobe analyzed and may generate false positive and Verify that a prompt sign (>) appears in the lower leftnegative results. corner of the display after the gain control message

For example, samples are going to be taken at leadbattery manufacturing site for a SSCS. There is no Verify that the upper right corner of the display readsinformation in the site history on the location or the number of seconds (0 - 32,000s) on the top lineconcentration of the Pb contamination. A model and the model number (1 - 32) on the bottom line.calibrated for Pb with a SSCS from another batterybreakage site could be used as a STCS to screen this If a "battery low" message appears recharge thesite and locate low, mid and high Pb contamination battery before proceeding or operate using linepoints for the SSCS sampling. voltage.

7.0 PROCEDURE

7.1 Prerequisites

If the X-MET 880 will be used in a location whereAC power outlets are conveniently accessible, connectthe battery charger to the battery and plug the chargercord into the outlet. The cable probe must beconnected before the power is switched on. Pluggingand unplugging this cable with the power on candamage the detector.

Verify that the probe shutter is closed by checking themechanical flag color on the side of the SAPS orDOPS. When the flag is green the shutter is closedand open when it is red.

Connect the probe cable to the connector labeled"PROBE" on the electronic module. Make certain theplug has been firmly pushed in all the way (you willfeel and hear a slight "click" as the probe connectorlocks into position).

Apply power to the X-MET 880 by pressing the "ON"button.

Verify that the display briefly reads:

X-MET 880 VX.X.X (software version) and DATE

SELF TEST COMPLETED, followed by (if the X-MET 880 has been off, more than a few minutes) themessage:

has stopped flashing.

Allow the X-MET 880 to warm up for approximately60 minutes.

If the X-MET 880 is being used in a location wherethe temperature of the environment has changed bymore than 5 F, then allow the X-MET 880 to stabilizeo

at the new ambient temperature. Approximately 1minute stabilization time for each 1 F change ino

ambient temperature should be allowed.

7.1.1 Gain Control

Allow 5 minutes after temperature stabilization for theX-MET 880 to perform a complete cycle of automaticelectronic gain control. The trigger must be releasedon the probe to activate the gain control operation.Additionally, the prompt sign (>) must be in thedisplay.

The X-ray spectrum contains a number of incident X-ray quantum energies of which there arecorresponding channels in the multichannel analyzer.These channels are limited by certain changes inconditions such as the resolution of the detector andtemperature coefficients. This means that aproportionally large error from measurement would beobtained, if no compensation was made for thesevariations.

The elimination of such errors is made possible bymonitoring the state of the probe spectrum andcompensating for any spectral drift as described inSection 7.1.2.

Page 11: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

11

The state of the detection region is maintained by a maximum (FWHM) of the peak and the counts will befeedback gain control system, which operates all the displayed. Record the PKCH and the FWHM in a logtime the probe is in a non-operative position (gain book.control position, DOPS, SAPS, SLPS placed on theirsides, HEPS, LEPS with the lid in the forward The peak channel (N) should not vary more than N±1position) and the instrument is not under any during all of the operations listed above.command (i.e., prompt displayed (> )).

The gain control routine is accomplished by allowing instrument to gain control for another 5.0 minutes. Ifthe X-MET 880 to measure a reference material the peak channel continues to drift after allowing it to(usually copper) mounted on the shutter and then gain control several times, contact an Outokumpumaintaining a track on the peak spectral position. representative. DO NOT continue to perform any

The initial peak position is set up during theinitialization of the probe (INI) and from this point theinstrument will adjust accordingly, during gaincontrol.

7.1.2 Gain Control Monitoring

During initialization, the initial peak channel is target element (i.e., typical detection limits for leadestablished for the gain control. During gain control, are 100-200 mg/kg. Selection of a 300 or 400 mg/kgadjustments are made to return the maximum peak SSCS would be appropriate). Select a mid SSCS at or(usually copper), after the backscatter peak, to this near the action level for each target element or aninitial peak channel location. The peak channel is SSCS about 5 or 10 times the low SSCS concentrationmonitored and recorded to verify if the gain control (i.e., typical action levels for lead are 500-2000mechanism is working and returning the peak to the mg/kg. Selection of a 1000 mg/kg SSCS would becorrect channel. Failure of the gain control appropriate). Select a high SSCS at or near the end ofmechanism will result in spectral drift and calculation the linear calibration range of each target element.of incorrect intensities in the element windows orincorrect pure element window calibration. The gain These SSCS should be measured using the samecontrol peak channel should be measured and recorded measuring that will be applied to the sample analysis.a the beginning, end, and every 25 to 40 minutes However, a minimum of 60 seconds should be usedduring the following operations: for Model verification and preoperational checks if the

1. Pure element calibration screening analysis.2. SSCS Measurements3. All sample analysis The low SSCS should be measured ten times, using4. All preoperational check sample the anticipated site measuring time, after the model

measurements. calibration has been completed. These results will be

The gain control peak channel measurement is and quantitation limit (QL) as described in Sectionperformed with the probe in the same position as it 15.0. A control range may be calculated using thewas during gain control (shutter is closed for the average of the results plus or minus the detection limitDOPS or SAPS probes and the sample chamber is (i.e., a low SSCS has a mean of 200 mg/kg and a DLpulled forward for the HEPS probe). The instrument of 120 mg/kg; the control range would be 80 to 320should be set for a minimum measuring time of 60 mg/kg). The low SSCS will have the largest relativeseconds. The enter maintenance program command percent variance due to its proximity to the DL.(EMP) is entered followed by "CONT/YES"command. Then the test measurement (TSM) The mid and high SSCS can be measured as describedcommand is entered followed by the "CONT/YES". above and a range calculated using the average plus orThe peak channel (PKCH), the full width half- minus three times the standard deviation of the results.

If the PKCH varies more than one channel, allow the

analysis until the problem has been corrected.

7.2 Preoperational Checks

Select a minimum of one low, mid and high SSCS(used in the model to be checked, not detected) for alltarget elements for every model to be checked. Selecta low SSCS above the typical detection limit for each

instrument is going to be using 15 to 30 second

used to calculate a preliminary detection limit (DL)

Page 12: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

12

Generally, three measurements of the standards is Where Y is the currently selected model number.sufficient. A control range may be calculated usingthe average of the three measurements plus or minus Enter the desired model number using the numbertwenty-five percent of the average. keys and press the "CONT/YES" key. The analyzer

These SSCS should be measured at least once calibration - spectral data only), INTENSITY (purewhenever the instrument is transported. Additionally, element intensities only - if the pure element routinethey should be analyzed at the beginning of each has been completed), LIBRARY (identificationanalysis day and at the end of the analytical period. calibration completed), or ASSAY (chemistry orAll results should be logged in the operator's log book composition calibration completed), along with theor saved on a computer disk in a report format. model's name (if assigned a name). Note that the

These SSCS may be used for verifying the model as programmed measuring time) has been registered indescribed in Section 7.5.10 and for Quality Assurance the lower right corner of the display.and Control as described in Section 9.0.

Results outside the described range indicate that there "CONT/YES" or "END/NO" key is pressed, theis an instrument problem. model number remains unchanged.

7.3 Normalization and Standardization

Normalization and standardization should never beperformed. These procedures have never been neededat ERT/REAC and have never been performed. It isrecommended that the operator recalibrate the modelif the calibration is older than four months for a Cd-109 source; six months for an Fe-SS source; and threeyears for a Cm-244 source. Recalibration is neverrequired for the Am-241 source.

7.4 General Keys and Commands

7.4.1 The "Shift" Key

Prior to any operations using the keypad, determinewhich keypad function is to be used:

ALPHA = SH showing in displayNUMERIC = SH not showing in display

Then press the "SHIFT" key to change keypadfunctions. For selecting the model as outlined inSection 7.4.2 below, the SH must not be showing inthe display.

7.4.2 The "Model" (Selection) Key

If it is desired to change models, then depress the"MODEL" key. The analyzer will show the currentmodel in the lower right corner of the display. Theanalyzer active display will read:

MODEL Y?

will display the model type, UNCALIBRATED (no

change in model number (and the associated pre-

If no new model number is entered and only the

Example of model selection:

>MODELMODEL 1? CONT/YES 15sUNCALIBRATED MODEL 1>

or

>MODELMODEL 1? 2 CONT/YES 10sOLD LIBRARY MODEL PB LEVEL 2>

or

>MODELMODEL 1? 32 CONT/YES 100sOLD ASSAY MODEL SITE X 32

The factory pre-programmed measurement time foreach model can be changed using the MTM,Measurement Time by Model, command. The changewill not be executed until the model is exited and re-selected.

7.4.3 The "MTIME" (Measurement TimeSelection) Key

If the measurement time needs to be changed, depressthe "MTIME" key. The analyzer will show themeasuring time in seconds followed by a lower case"s" in the upper right corner of the display. The activedisplay (bottom line) will read:

Page 13: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

13

MEASURING TIME XXX? only the result for the Last Source Measurement can

Depress the number keys to enter the desired ment has to be carried out.measurement time (15 to 240 seconds are typicalmeasuring time for hazardous waste application - 1 to32767 is the total range) and depress the"CONT/YES" key. Note the corresponding change inmeasuring time in the upper right corner of thedisplay.

The measurement time remains unchanged if the"CONT/YES" or "END/NO" keys alone are pressed.The new measurement time replaces the old value inthe upper right hand corner of the display.

All uncalibrated models are defaulted to a measuringtime of 15 seconds. All calibrated models aredefaulted to the pre-programmed (under the MTMcommand) measurement time.

Example of selection of measurement time:

>MTIME CONT/YESMEASURING TIME 15 ? 60 CONT/YES

The MTIME command provides a temporary changein the measurement time. The measurement time willreturn to the model pre-programmed (MTMcommand) measurement time when a new model isselected.

7.4.4 AMS Command: Data AveragingMode of the instrument's precision. The statistical error is

Use the AMS command (Average MeasurementService) if an average of several measurements isdesired. First, select the desired model (see Section7.4.2). Then enter the command AMS and depress"CONT/YES". The analyzer will display:

MEASURE

Measure the sample(s). Depress the "END/NO" keyto terminate the AMS command and display theaverage of the measurements.

7.4.5 The "RECALC" Key

Results (spectrum) from the last measurement can berecalculated using another model. This is done byswitching to a new model, and pressing the"RECALC" key. If the new model is for two sources,

be calculated; if this is not sufficient, a new measure-

7.4.6 The Function Keys "F1 - F5"

The five function keys can be programmed to containany of the three letter command acronyms. Up to five(5) pre-programmed expressions can be stored in theanalyzer's memory and are retained as long as acharged battery is connected. The programming isinitiated with the FNC command.

If the keyboard is locked (LOC command), only the"ON, OFF, CONT/YES, END/NO, RECALC,START, and F1- F5" keys can be used.

7.4.7 STD Command: Standard Deviations

The STD command computes the statistical standarddeviations, the error due to counting statistics, fromthe last measured sample spectrum for the analyzed(target) elements.

The display format is similar to the concentrationoutput:

STDEVS: FE 1.04 CR .361 CU .142 PB .006

This does not reflect the total error of the measure-ment (accuracy), but only the part due to countingstatistics (precision). Generally, it is a good estimate

reduced fifty percent for each quadrupling(multiplying by four) of the measurement time.

7.4.8 Other Commands

Many other commands are available on the X-MET880 and are confirmed and executed with the"CONT/YES key. Refer to the X-MET 880 OperatingInstructions for further information.

7.5 Instrument Calibration

7.5.1 DEL Command: Deleting a Model

If all 32 models have (an) old library(s) or assaymodel(s) in them, then a model must be deleted beforeproceeding with a new calibration. Enter DELfollowed by the model number (1-32) and confirm the

Page 14: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

14

action by pressing the "CONT/YES" key. The X- general parameters sub-section of the overall modelMET 880 will display the selected model number and parameters) will be displayed. This sub-section isit's name and ask if this model is to be deleted. accepted by pressing the "CONT/YES" key. If theAnother "CONT/YES" key response deletes the old model has already been assigned a name, the name ofmodel clearing the space for a new model. Therefore, the model number entered will appear next and thedeleting a model with the DEL command means that model can be re-named if desired, if not, press thea new pure element and sample calibration will be "CONT/YES" key. The model type will appear at therequired. prompt. Enter "CONT/YES".Example:

To delete model 6: of model to one of the three choices: Identification

>DEL CONT/YESWHICH MODEL TO DELETE (1-32) ? 6

CONT/YESDELETE MODEL 6 CRFECU CHEM ?

CONT/YESDELETED>

An "END/NO" key response to the above will resultin the following display:

NOTHING CHANGED

and the model will not be deleted.

7.5.2 Changing a Model's "MaximumCount Rate" and Default MeasuringTime

All uncalibrated models have a default maximumcount rate limit of 6 Khz (6000 counts per second) anda default measuring time of 15 seconds. Prior to pureelement calibration, the maximum count-rate limitshould be increased to 15 Khz. Simultaneously, themodel default measuring time can be changed to theanticipated calibration time (and after calibration, tothe anticipated field or sample analysis time). Thesechanges are performed with the following procedure.

Unlock the maintenance program by typing in theEMP command, followed by the "CONT/YES" key.This enables the maintenance program. Enter themodel parameters section of the maintenance programby typing in the PRM command, (Parameterscommand) followed by the "CONT/YES" key. Enterthe number of the model to be accessed (1-32), oraccept the number offered by pressing the"CONT/YES" key. The words "UNCALIBRATEDMODEL" will be displayed. Press the "CONT/YES"key. The words "GENERAL PARAMETERS" (the

Entering "END/NO" at this query will change the type

(IDENT), Assay (ASSAY), or Undefined (UNDEF).The wrong computational algorithm will be applied ifthe incorrect type of model is entered. Answer"END/NO" to the incorrect model type and"CONT/YES" to the correct model type.

The measuring time will be displayed next. Key inthe new default measuring time (performs the samefunction as using the MTM command) or accept theoffered time by pressing the "CONT/YES" key. Thedisplay will show "Number of channels: XX".Continue past this with the "CONT/YES" key. Thetitle "Flow Check Channel" will appear at the prompt.Press the "CONT/YES" key. The title "Check CountRate" will appear at the prompt. Press the"CONT/YES" key. The title "Max Count Rate" willappear at the prompt followed by "6 Khz ?". Press the"END/NO" key. This will scroll the display to thenext count rate default of 10 Khz. Press the"END/NO" key and the display will show "15 Khz?".To accept the 15 Khz maximum count rate limit press"CONT/YES". The display will show "Safety Limit".Continue past this with the "CONT/YES" key. Thetitle "CHANNEL PARAMETERS?", which is thenext sub-section of the PRM command, will appear atthe prompt. Enter the "END/NO" key. The title "G-MATRIX?", the next sub-section of the PRMcommand, will appear at the prompt. Enter the"END/NO" key. The title "ASSAY/IDENTPARAMETERS?", which is the last sub-section of thePRM command, will appear at the prompt. Enter the"END/NO" key. You have exited the PRM file,which is evidenced by the reappearance of the promptsign ">" on the display.

To re-lock the parameters file type in the XMPcommand, followed by the "CONT/YES" key. Thisinhibits access to the maintenance program.

Example of changing a model's Maximum Count Ratelimit and Default Measuring Time:

Page 15: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

15

>EMP CONT/YES>PRM CONT/YESMODEL 5 ? CONT/YESUNCALIBRATED MODEL CONT/YESGENERAL PARAMETERS CONT/YESName: ?CONT/YESType: UNDEF ? CONT/YESMeas. Time: 15? 200 CONT/YESNo. of channels: 0 ? CONT/YESFlow check channel: 0 ? CONT/YESCheck count rate: 0.00 ? CONT/YESMax count rate: 6 Khz ? END/NOMax count rate: 10 Khz ? END/NOMax count rate: 15 Khz ? CONT/YESSafety limit %: 0.00 ? CONT/YESCHANNEL PARAMETERS ? END/NOG MATRIX ? END/NOASSAY/IDENT PARAMETERS ? END/NO

7.5.3 PUR Command: Pure ElementCalibration

7.5.3.1 Introduction element calibration in several models without

During pure element calibration, the analyzerdetermines the locations of the channels which willbracket the energy of the element(s) to be measured,including the backscatter peak. The analyzer alsosimultaneously calculates the element spectral overlapfactors, or the values to be used in the "G-Matrix"table to correct each channel for the influence ofadjacent element peaks (in the form of spectraloverlap).

The pure element samples for the up to 10 elementschosen for calibration (1 to 9 elements plusbackscatter) may be measured in any order, however,it is recommended that the elements be measured inorder of their respective X-ray energies to facilitateoperator review of the channel limit (LIM command)settings (LIM review follows the PUR command).

The elements selected for an assay model shouldinclude all elements to be analyzed plus any elementsthat might interfere, either by spectral overlap or bymatrix interference.

The analyzer stores the spectra measured in a mainspectra table and computes the correct channel limitsfor each element, that are assigned to the specificmodel. Pulses falling between these limits areincluded in the channel or "window" of the respectiveelement. The analyzer calculates spectral overlapfactors and stores them in the G-MATRIX. This is theinformation from the spectra that will be required forthe deconvolution calculations. Each spectrum andthe respective channel limits are automatically labeledwith the element, probe number and source used.

The instrument must be stabilized (powered-up forelectrical stabilization and kept in a constanttemperature environment for thermal stabilization) for60 minutes, and allowed to gain control for 5.0minutes prior to the pure element calibration. Theinstrument must be allowed to gain control for 5.0minutes every 25 minutes during pure elementcalibration. The area in which the instrument is beingcalibrated in must be thermally stable to within ±3EF.A minimum measurement time of 240 seconds isnormally used for pure element calibration.

Pure element spectra are stored in a main spectrastorage table to facilitate copying to another model.This gives the operator the option of performing pure

measuring the pure elements each time. Pure elementspectra stored in the main spectrum table should notbe used unless they have been acquired and storedwithin the previous eight hours by the same operatorthat performed the initial pure element calibration.

The model channel limits can be reviewed and re-setmanually after pure element calibration using the LIMcommand. They can be output to a peripheral deviceby using P command (print) followed by the"CONT/YES" key while in the LIM Table.

7.5.3.2 Operation

Gain control monitoring must be performed asdescribed in Section 7.1.2. Pure element calibrationis started by entering the PUR command followed bythe "CONT/YES" key. The instrument asks if youwish to "Delete the old spectra?". This is answered bythe "CONT/YES" or "END/NO" key. Answeringwith the "CONT/YES" key erases all spectra in thelibrary, making copying impossible until new spectrahave been generated from pure elementmeasurements.

Page 16: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

16

The X-MET 880 asks for a new model name if the to Cd. The minimum PUR command calibrationmodel has not been previously calibrated. It is elements for source A would includerecommended that the operator include a statement inthe model name's title concerning the concentration of Cd, Sn, Fe, and BS (backscatter peak). Fe is alwaysthe XRF readings (e.g., if the assay model is going to included in the PUR calibration of all sources becausemeasure 0 -20,000 ppm Pb (mg/kg), the assay values it is always present in soil. The minimum PURwill have to be entered as parts per million by weight calibration for source B would include Fe, Ni, and Bs.(ppmw) or mg/kg/10 since the instrument only has 4 Note that the Fe and BS must be measured twice, onceplaces before and after the decimal point for data for each source used. The channel limits for anentry). To ensure all operators are aware of this, it is element cannot be interchanged between sources.recommended that in addition to noting this fact in theXRF logbook (see Section 19.0), the model name For DOPS or SAPS Probesreflects this also (i.e., title: BBATY PPM =XRFX10). Place the pure element sample in position and start the

Next, the instrument prompts for the first element and type probe) or by pulling the trigger on the surfaceasks for it to be measured. The elements can be probe and holding it open until the measurement isnamed either by their one or two letter element completed. On DOPS probe, make sure thesymbol or by their atomic number. The scattering appropriate source is selected by observing the sourcesample has to be named by the symbol BS (or atomic window display on the side of the DOPS probe andnumber 0). the display panel on the X-MET 880 electronic unit .

For DOPS Probe only

If the DOPS probe is being used, the instrument source by releasing the trigger and pressing theprompts for the source to be used (either source A or "END/NO" key. Then re-enter the program with theB). The operator should determine the appropriate PUR command and respond to the X-MET 880source for the most efficient excitation of the desired prompts until you return to the desired channeltarget element(s) (see source selection chart in X- number. Type in the desired element symbol and re-MET 880 Operating Instructions) and select the start the measurement.appropriate source(s) for the target element(s). SourceA is the Am-241 source and source B is the Cm-244 It is recommended that the operator examine eachsource (or the Cd-109 source in the 880EA system). pure element sample prior to use to ensure that theBoth sources can be used in a model to analyze pure element plastic or metallic disc is still fixed indifferent target elements. If a dual source method is place and has not fallen out.used, a total maximum of 10 pure element calibrationsmay be used in the model for both sources (i.e., an It is recommended that the elements be measured ineven division of pure elements between two sources order of their respective X-ray energies to facilitatecould result in 4 elements and a backscatter for Source operator review of the channel limit (LIM command)A and 4 elements and a backscatter for Source B for settings (LIM review follows the PUR command).a total of 10 pure elements in the model). Rememberto include all elements that might interfere, either by All of the calibration samples will have to be re-spectral overlap or by matrix effect in each source measured, if any pure elements are added, deleted, orpure element calibration. re-measured with the PUR command after the

For example, a site requires the soil analysis of CAL command.cadmium (Cd) and nickel (Ni). Ni analysis requiresthe use of the Cm-244 source (B). Cd analysis While the measurement is in progress, the remainingrequires the use of the Am-241 source (A). The measurement time is continuously displayed. Whenoperator notes that there is also tin (Sn) present in the the measurement is completed, the peak channel andsoil and this may cause spectral overlap in the Cd the Full-Width at Half-Maximum (FWHM) of thewindow due to the fact that Sn is an adjacent element pure element spectrum is displayed and the instrument

measurement by pressing the "START" key (sample

Interrupt any measurements that start with an incorrect

calibration samples have been measured using the

Page 17: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

17

requests the next pure element. The PUR calibrationcan be terminated with the "END/NO" key.

Example of pure element calibration of a new modelusing a DOPS probe (6), with an Am-241 source (A)and a Cm-244 source (B):

>PUR CONT/YESDELETE OLD SPECTRA ? END/NO OR

CONT/YES*

*(CONT/YES deletes all existing spectra in mainspectral library; maximum = 20)

CALIBRATING A NEW MODELNAME? PLATE PPM=XRFX10 CONT/YES1. PURE SAMPLE:CR CONT/YESSOURCE A ? END/NOSOURCE B ? CONT/YESMEASURE CR6B PULL TRIGGERMEASURING: PROBE 6 TYPE DOPS (B) 200SECONDS - measurement, time is counted down; atconclusion display reads:PEAK CHANNEL: 69 FWHM: 12 (peak channel &channel width)2. PURE SAMPLE: MN CONT/YES should not be used unless they have been acquired andSOURCE A ? END/NOSOURCE B ? CONT/YES

MEASURE MN6B PULL TRIGGERMEASURING: PROBE 6 TYPE DOPS (B) 200SECONDSPEAK CHANNEL: 75 FWHM: 123. PURE SAMPLE: FE CONT/YESSOURCE A ? END/NOSOURCE B ? CONT/YESMEASURE FE6B PULL TRIGGERMEASURING: PROBE 6 TYPE DOPS (B) 200SECONDS PEAK CHANNEL: 81 FWHM: 124. PURE SAMPLE:BS CONT/YES(scattering sample)SOURCE A ? END/NOSOURCE B ? CONT/YESMEASURE BS6B PULL TRIGGERMEASURING: PROBE 6 TYPE DOPS (B) 200SECONDS PEAK CHANNEL: 255 FWHM: 15. PURE SAMPLE: END/NOCALIBRATION FINISHED>

7.5.3.3 Multiple Models

Pure element calibration can be performed using thepure element spectra stored in the buffer memorycalled the spectrum table, which holds 20 completepure element spectra. The pure element spectra storedin the spectrum table can be copied into any model(s)using the PUR command. Each model can accept upto 10 such spectra. These pure element spectra areaccessed during PUR calibration for transfer of theappropriate channel limits data as well as forcalculating the appropriate G-Matrix overlap factorcorrections. Thereafter, the model no longer needs torefer to the spectrum table. Therefore, the 20 elementlimitation of the spectrum table does not limit the X-MET 880 to the measurement of only 20 elements. Ifmore than 20 spectra are added, the 21st one willoverwrite the first one in the spectra table, and so on,but previously calibrated models will not be affected.If calibrating several models with more than 20 totalelements using PUR, it is advisable to begin withthose elements that occur only in the first model(s) tobe calibrated as they will be overwritten in thespectrum table, but not in the calibrated model(s).Pure element spectra stored in the main spectrum table

stored within the previous eight hours by the sameoperator that performed the initial pure elementcalibration in the spectrum table.

7.5.3.4 LIM Command:Examining andVerifying the ChannelLimits

The channel limits are printed out and examined usingthe LIM command once the pure element calibrationhas been completed. The LIM command first displaysthe number of channels and asks if the operator wantsto examine the individual channel information. Thebottom line (active line) of the X-MET 880 displayreads: "EXAMINE ?". Pressing the "END/NO" keyexits the LIM command. The operator may use"CONT/YES" key to review the data for the firstchannel; element symbol, probe and sourceidentification, pure element gross count rate and thenormalization coefficient. On the second line, the X-MET 880 displays the channel limits. Elementchannels are advanced in the forward direction withthe "CONT/YES" key and backwards with the "^"(up arrow) key. The P command followed by the

Page 18: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

18

"CONT/YES" key gives an output in table form to a FE 6B 450.06 1.000 78 86printer or a terminal (see Section 5.3). BS 6B 185.18 1.000 255 255

During scanning, new channel limits can be manuallyentered in place of the old ones. The count-rate forthe new channel limits will then change accordingly.Manual channel limit setting is required in certainsituations, such as when setting a channel at an L-betaline, (i.e., when the L-alpha line has an overlap withanother element channel (such as the case of As(arsenic) K-alpha overlapped with Pb (lead) L-alpha).The pure element spectra must be in the spectrumtable in order to manually change the channel limits.

Before proceeding to the next step, the pure elementchannel limits must be examined and verified for eachsource. None of the pure element channel limitsshould overlap or coincide. If any do, the overlappingpure element spectra should be deleted andremeasured. Additionally, the order of the channellimits must be identical to the corresponding X-rayenergy order of the elements measured (low to highmeasurement order facilitates this review). Note inthe example below, that the order of the Cr(chromium), Mn (manganese) and Fe (Iron) channellimits corresponds to the atomic number order and theX-ray energy order of the elements. If the channellimits do not follow this pattern, the anomalous pureelement spectra should be deleted and remeasured.

Stepping through channel limits is discontinued withthe "END/NO" key. After termination of thecommand, the instrument pure element calibrationdata (spectral overlap factors called G-Matrix) is re-calculated on the basis of the new channel limits.

Example of examining the channel limits:

> LIM CONT/YES4 channelsEXAMINE ? CONT/YES

CR6B 820.17 1.000 63 71 NEW ? P CONT/YES

P command initiates formatted printout, if printer isconnected):(MODEL 5: PLATE PPM=XRFX10)

ELEM PROBE COUNT NORM. LIMITS RATE FACTOR CR 6B 820.17 1.000 63 71MN 6B 1194.64 1.000 72 77

7.5.4 SPE Command: Examining Spectra

The pure element and sample spectra can be output bythe SPE command. The last sample spectrum isoutput by answering "CONT/YES" to the query"LATEST?" (latest measurement). Pure elementspectrum are output by answering "END/NO" to thequery "LATEST?". If "END/NO" is selected, thepure element spectrum can be specified the followingthree ways:

1. The position number in the spectrum table(1-20).

2. The pure element symbol.3. The atomic number of the pure element (but

only if it is larger than 20; otherwise it isinterpreted by the instrument as the positionnumber in table).

If the same pure element has been measured withmore than one probe, the program will also prompt forthe probe number (and for source A or B if DOPSprobe is being used) after entry of the symbol orelement number. When the pure element has beenspecified with its position number or symbol, then theprobe number and measurement time will bedisplayed. Display of the channel numbers and countsof two channels at a time can be obtained by scanningbackwards or forwards. Forward scanning is donewith the "CONT/YES" key and backward scanningwith the "^" (up arrow) key. Keying in the micro-channel number n (0 to 255) displays the counts ofmicro-channel n. Scanning is discontinued with the"END/NO" key and the instrument will then ask for anew pure element spectrum. Another "END/NO" atthis point will terminate the SPE function.

7.5.5 SPL Command: Spectrum Plot Usinga Peripheral Device

A spectrum can be plotted with a printer or on acomputer terminal by using the command SPL. Thespectrum is chosen as in the SPE command.Thereafter, X-MET 880 will prompt for the first andlast channels, the output window settings and thelower and upper limits for the scaling of the countsaxis. The output "window" means the number ofmicro-channels which are integrated and printed asone channel of the plot. If the automatic lower and

Page 19: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

19

upper limits are not changed, the plot will be scaled calibration sample, or every 25 minutes, during theaccording to the highest peak of the spectrum. CAL sample measurements. Gain control monitoring

7.5.6 CAL Command: Assay ModelSample Calibration

7.5.6.1 Introduction

When the pure element calibration (PUR command) ofthe analyzer has been carried out, the X-MET 880 iscapable of computing the net counts (intensities) inthe element channels. In order to proceed from netintensities to concentrations (assay), samplecalibration is required. This entails measuring knownsamples using the CAL calibrate command (assayreadout).

Assay models are used for hazardous wasteapplication models since they provide results in aconcentration which can be QA/QC'd. Identificationmodels produce identification names of the sampletypes by the quality of spectral matchings (i.e., LOWPB LEVEL or HIGH PB LEVEL). Identificationmodels are never used for hazardous wasteapplications because matching low resolution spectracan result in false qualitative and quantitative results.Additionally, identification models will not beaddressed in this operating procedure.

For calibration of an assay model, the samples aremeasured using the CAL command, and the assaymodel is calculated using X-MET 880 internalsoftware by multi-variable regression analysis. Thecalculation can be performed either internally, usingthe analyzer software (with a maximum of 30calibration samples), or, externally in which case thereis no restriction on the number of samples used. Thisoperating procedure will only address internalcalculations since ERT/REAC does not have externalsoftware.

7.5.6.2 Measurement ofCalibration Samples

The instrument must be stabilized (turned-on andwarmed- up for 60 minutes) and allowed to gaincontrol for 5.0 minutes prior to the measurement ofcalibration samples. The area in which the instrumentis being calibrated should be thermally stable(maintain ±3EF). A measurement time of 200 secondsor more should be used for measurement of calibrationsamples. The instrument must be allowed to gaincontrol for 5.0 minutes after measuring every 6th

must be performed as described in Section 7.1.2.

In the sample calibration of an assay model, the X-MET 880 stores the intensities of the calibrationsamples in a calibration intensities table, from whichthey can be output using the CIN command, or useddirectly for the generation of a regression model bythe internal regression program of the X-MET 880(MOD calibration modeling command). Thecalibration intensities table is an electronic scratch-pad that is erased after each model is completed andthen moved to the next model to be calibrated. Thistable can be used for the calculation of only one modelat a time. The maximum size of the calibrationintensities table is 30 samples, hence, if a greaternumber of samples is required, the intensities must beoutput for external calculation of the regressionmodel. The calibration samples can be measured inan arbitrary order, AS LONG AS THE ORDER ISDOCUMENTED. Therefore, be sure to make a noteof the order used and document it in the operator's logbook. The calibration intensities table stores thesamples by cardinal number (1-30) only, in the orderthey were measured. The function is started with theCAL command. The X-MET 880 begins bydisplaying the status of the calibration table:

- NO OLD DATA- OLD DATA: 30 SAMPLES (TABLE FULL) or- OLD DATA: n SAMPLES, MODEL XX OVERWRITE?

If the OVERWRITE? question appears, it isreminding the operator to finish the modelingcalculations using the calibration intensities tablemeasured in an earlier model. If the earlier model isalready completed (modeling is finished using theMOD command) and the calibration coefficients arealready calculated, then the "CONT/YES" key shouldbe pressed, to erase the old model scratchpad andmove it to the current model. If the earlier model isnot completed, then the "END/NO" key should bepressed and calibration calculations finished in theearlier model. NOTE: if "CONT/YES" is pressedonly new samples can be measured and the oldcalibration intensities table is erased.

The X-MET 880 asks for one sample at a time:

MEASURE SAMPLE n

This can be answered in any of the following ways:

Page 20: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

20

Measure the sample PULL TRIGGERProceed to Sample m Enter sample number m, followed by: CONT/YESExit from Sample query END/NO

Repeat measurement of a particular sample can beaccomplished by means of the sample number. Afterpressing "END/NO", the last sample to be stored isverified.

LAST SAMPLE n ?

This can be answered with a smaller number if it isdesired to omit some of the samples. If the tablebecomes full, the message TABLE FULL is displayedand the command is terminated.

7.5.6.3 Sample Calibration of aModel Using Both Sources ina DOPS Probe

A given analysis may require the use of both sourcesin a DOPS probe to cover all the elements to bemeasured. A single model employing both sources inthe DOPS probe can be calibrated as follows:

Every element to be measured by each source must 8 SAMPLESundergo independent pure element calibration for each >source. For example:

If BS is going to be used in the calibration of both sample # 4 was incorrect, then measuring the correctsources A & B, then the BS sample must be measured sample will automatically overwrite the previous # 4by both sources A & B during the pure element sample data in the table.calibration (PUR) .

If the model's pure element calibration channel limitscontain limits for both sources, then during samplecalibration the DOPS probe will automatically switchbetween sources, as required, to measure the spectrafor both sources, and store the intensities in therespective source's element channels. Be sure tocontinue to hold the trigger open after the 1st sourcemeasurement. The X-MET 880 will indicate thecompletion of the first of the two measurements witha "beep-beep-beep". This is immediately followed bya clicking noise that is the sound of the source changerswitching sources. Do not release the trigger until the2nd source has completed its measurement.

During measurements, always verify that theappropriate source is selected by observing the displayon the side of the DOPS probe. Interrupt and end

("END/NO") any measurements started with anincorrect source.

7.5.6.4 Deletion of theCalibration IntensitiesTable

The calibration intensities table can only be deleted bygoing to a different model to obtain the"OVERWRITE" message. The operator can theneither go back to the previous model or stay with thenew model to start a new calibration intensities table.

Example:

>MODEL CONT/YESMODEL 4 ? CONT/YES>CAL CONT/YESOLD DATA: 11 SAMPLES, MODEL 2 (meansprevious calibration is in Model 2)OVERWRITE? CONT/YESMEASURE SAMPLE 1 PULL TRIGGERMEASURING: PROBE 6 TYPE DOPS (B)200 SECONDSMEASURE SAMPLE 9 END/NOLAST SAMPLE 8? CONT/YES

If it is discovered that after measuring eight samples,

>CAL CONT/YESOLD DATA: 8 SAMPLESMEASURE SAMPLE 9 4 CONT/YESMEASURE SAMPLE 4 PULL TRIGGERMEASURING: PROBE 6 TYPE DOPS (B)200 SECONDSMEASURE SAMPLE 5 END/NOLAST SAMPLE 8 ? CONT/YES8 SAMPLES>7.5.7 ASY Command: Input of Calibration

Sample Assay Values

The chemical metal analysis (assay) results of theSSCS are entered using the ASY command. The X-MET 880 starts by prompting for the target elementsfor which concentrations are required (these are

Page 21: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

21

CAUTION:Do not enter a P "CONT/YES" commandwhile in the first section of the ASYcommand, the entry of DEPENDENTS, asin this section a P will be interpreted as theelement phosphorus. Should this be doneaccidently, the instrument will ask for re-verification that the element P(phosphorus) is desired in place of thepreviously selected element(s). At thispoint the operator would simply re-enterthe correct element symbols prior topressing "CONT/YES" which accepts theoffered element (P). IF the symbol P isinadvertently accepted as the element, byacknowledging it with "CONT/YES"twice, it will then be necessary to re-enterASY and the correct elements symbols andre-install all the assay values.

NOTE:Only assay values with less than foursignificant figures in front of the decimalpoint can be entered into this table. Therefore, if any samples contain assayvalues with greater than 4 significantfigures in front of the decimal point ALLassay values for all elements in the suiteshould be divided by some power of 10(such as 10, 100, 1000, or 10,000) prior toentry into the assay table. For example,one sample in the suite is assayed at33,999 mg/kg. The decimal fractionresulting from the division of 10 wouldconvert this value to 3,399.9. Therefore,the value 3,399.9 would be entered into theassay table and all other element assays inthe calibration suite would likewise bedivided by 10 prior to entry into the assaytable.

known as the DEPENDENTS). This is answered byentering the element symbols (XX) of all the desiredelements for readout (up to six) using the "SPACE"key in between the symbols, but not following the lastentry. To enter the element symbols typed onto thedisplay press "CONT/YES". The instrument will thenstart asking for each sample assay value to be entered.Type in each assay value for each sample followed by"CONT/YES", until all the calibration samples assayvalues have been entered. The concentrations can bechanged afterwards if required by stepping backwardwith the "^" (up-arrow). If concentration values fornewly measured samples, or additional elements, aredesired to be added later, then re-enter the ASYcommand and press "CONT/YES" until the first assayvalue is on the lower line of the display. Next, type a(/) followed by the sample number you desire to jumpto. The use of a (/) mark in front of the samplenumber advances the assay table directly to thesample number entered. From any location in theassay table, P followed by "CONT/YES" gives outputin tabular form to a printer or a terminal.

Page 22: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

22

Example 1: Entering Cr and Cu assays, for eight 2. SAMPLE CR: 34.2300calibration samples that have been measured using the ....CAL command.

>_ASY CONT/YESOLD DEPENDENTS:NO OLD DEPENDENTS

CONT/YESNEW: CR CU CONT/YESNEW DEPENDENTS?: CR CU

CONT/YES1. SAMPLE CR: 0.0000 ? 10.97

CONT/YES1. SAMPLE CR: 10.97001. SAMPLE CU: 0.0000 ? 34.23

CONT/YES1. SAMPLE CU: 34.2300

2. SAMPLE CR: 0.0000 ? 8.56CONT/YES

2. SAMPLE CR: 8.56002. SAMPLE CU: 0.0000 ? 42.29

CONT/YES....8. SAMPLE CR: 0.0000 22.39

CONT/YES8. SAMPLE CR: 22.39008. SAMPLE CU: 0.0000 ? 15.00

CONT/YES(Terminates assay command automatically)8*2 ASSAYS>

Example 2: Changing dependents from the old onesCr, Cu, to new: Cr, Cu and Fe. NOTE: This will noterase the previous assay table data, but will open upspace to add Fe data.

>ASY CONT/YESOLD DEPENDENTS: CR CUNEW ? CR CU FE CONT/YES1. SAMPLE CR: 10.9700 ? CONT/YES

1. SAMPLE CR: 10.9700

1. SAMPLE CU: 34.2300 ? CONT/YES1. SAMPLE CU: 34.2300

1. SAMPLE FE: 0.0000?40.2 CONT/YES1. SAMPLE FE: 40.20002. SAMPLE CR: 34.2300? CONT/YES

8. SAMPLE CR: 22.3900 CONT/YES8. SAMPLE CU: 15.0000? CONT/YES8. SAMPLE FE: 0.0000?51.00 CONT/YES(Terminates assay command automatically)

8*3 ASSAYS>

7.5.8 MOD Command: Generating theRegression Model

The MOD command of the X-MET 880 initiates theprogram for calculation of multi-variable regressionmodel coefficients. The program offers regressioncalculations for the DEPENDENTS (assay values)given in the ASY command (i.e., those elementswhose symbols were entered as DEPENDENTS).

7.5.8.1 Setting the Dependents

>MOD CONT/YESREGRESSION FOR CR ? END/NOREGRESSION FOR FE ? CONT/YES

Upon an affirmative answer ("CONT/YES"), theINDEPENDENTS (elements' intensities) for theelement(s) in question are requested: i.e., thoseelements that may affect the slope of the dependentelement are entered with the element symbol, to beincluded in the regression calculation.

7.5.8.2 S e t t i n g t h eIndependents

DEFINE INDEPENDENTS:Stop indep input by END-key1. indep: ? FE CONT/YES2. indep: ? CR CONT/YES3. indep: ? END/NO

At this stage it is possible to scan forward with the"CONT/YES" command and backwards with the "^"(up arrow) command: "-" deletes an independent."END/NO" terminates input, computes the regressionmodel and displays the quality-of-fit of the regressionline.

The simplest independent expressions arerecommended at first. The element can be specified

Page 23: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

23

either by its symbol or its element number. The other The X-MET 880 then asks if it is desired to examinepermissible forms of expression are as follows using the residuals (i.e., the comparison of concentrationsiron, chromium and backscatter as examples: calculated by the model with the assay values

FE otherwise "CONT/YES" leads to scanning of theFE/ or FE/BS BS = scatter intensity residuals. The notation is as follows:FE*CRFE*CR/ means FE*CR/(BS*BS) ASSAY = concentration enteredBS/ means 1/BS (from the AA, ICP orFE*BS/ means FE/(BS*BS) referee analysis)BS*BS means 1/(BS*BS) ESTIM. = concentration calculated with the modelIf independents have been entered incorrectly, i.e. , RESID. = ASSAY - ESTIMusing elements that are not included in the instrument ST.RES = RESID/standard deviation Scalibration, the X-MET 880 responds by beeping andthen listing the possible elements for inclusion as Example:independents (only those that were included in thepure element calibration) and then displays a NO ASSAY ESTIM R E S I Drepetition of the query: ST.RES

"beep" 1 30.300 29.875 0.425 0.274USE CR FE CU1. indep: Forward scanning is done with the "CONT/YES"

7.5.8.3 Reviewing theRegression FitParameters

After defining the independents, "END/NO"terminates the independent input and calculates theinternal regression fit. The figure-of-merit parametersare then displayed:

R = correlation coefficientS = standard deviation around

regression lineF (M,N) = F-test value of regression

with degrees of freedomM,N

M = number of independentsN = number of samples M-1

Example:

R = 0.975 S = 1.55 F = (2.5)= 4.25RESIDUALS ?

7.5.8.4 Examining theResiduals Table

entered). If these are not required, reply "END/NO";

command, backward scanning with the "^" (up-arrow)command; entering the number of a calibration samplefollowed by "CONT/YES" leads directly to thatsample in a table. P "CONT/YES" sends the output,in table form, to the printer or other peripheral device,D "CONT/YES" sends a calibration Data plot diagramto the printer or other peripheral device.

"END/NO" terminates the scanning.

7.5.8.5 Deleting Points

The scanning of residuals is followed by the question:

DELETE POINTS:

If it is required to delete samples from the regressionmodel calculation, enter the numbers of such samples,i.e., 3 SPACE 5 and press "CONT/YES". This wil lcalculate a new regression model and display the newfigure-of- merit data for this new regression model.This display will be followed by the RESIDUALS ?query. A "CONT/YES" response leads to scanningthe new residual table (using the new regressionmodel calculated without the deleted points). Thistable may be scanned also. "END/NO" terminates thescanning and returns to the DELETE POINTS:prompt. Additional points may be deleted at this time.

Page 24: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

24

NOTE:If the new residual table is scannedforward using the "CONT/YES"command and backward using the "^"(up arrow) command, the deletedpoint(s) will simply be missing in theresidual table.

If the new residual table is output in tabular formusing the P "CONT/YES" command, then the The slope and t-value for slope 1 correspond to thecalibration table knows the samples by cardinal first independent (element). The slope and t-value fornumber (1-30) only. The residual table's sample slope 2 correspond to the second independent, etc.numbers will change for points with cardinal numberslarger than those which are deleted. For example:Using a calibration table with 20 samples, an operatordeletes point number 5 and returns to the residualtable. The new residual table will contain points 1-19.Points 1-4 in the "output" new residual table willcorrespond to the cardinal numbers 1-4 in thecalibration table. Points 5-19 correspond to theoriginal calibration cardinal numbers 6-20. Therefore,if the operator wants to delete sample number 9 in the"output" residual table, sample (calibration cardinal)number 10 must be entered to the DELETEPOINTS:prompt after terminating the residual scanning.

The next new residual table will contain residuals forpoints 1-18 now. Points 1-4 in this new residual tablecorrespond to the calibration cardinal numbers 1-4.Points 5-8 in the new residual table correspond tocalibration cardinal numbers 6-9. Points 10-18 in thenew residual table correspond to calibration cardinalnumbers 11-20.

This loop will continue until the reply to the DELETEPOINTS: prompt is "CONT/YES" or "END/NO".The X-MET 880 response to an incorrect input is"GIVE POINT NUMBERS".

If the residuals are not examined, or the response tothe "DELETE POINTS" prompt is "CONT/YES" or"END/NO", the next question is COEFFICIENTSAND T VALUES?.

7.5.8.6 Examining Coefficients andT-Values

To the question: dependent regression), reiterating the model with newCOEFFICIENTS AND T-VALUES ? independents, computing a new regression, possibly

a response of "CONT/YES" will allow you to scan thecoefficients and T-Values for each independent usedin the regression model. Forward scanning is donewith the "CONT/YES" command, backward scanningwith the "^" (up-arrow) command.

INTERCEPT = 1.23 E-4SLOPE 1 = 3.827E-2 T = 15.60

CONT/YESSLOPE 2 =-1.00837 T = -9.86

"END/NO" terminates the scanning.P "CONT/YES" gives output in table form:

COEFFICIENTS AND T-VALUES ?CONT/YESINTERCEPT = -1.1433235E+1SLOPE 1 = 7.6707910E+2 T = 83.40P CONT/YES

COEFFICIENTS AND T-VALUES FOR *NI*ITC NI/ NI*FE

SL -11.433235 767.079104 -0.000234T- VALUE 83.401736 -7.457542

7.5.8.7 Reiterating the Modelor Exiting

After terminating the scanning with the "END/NO"command, the query CHANGE INDEPENDENTS? isdisplayed. If the model created for the element doesnot require changing, reply "END/NO". This wil lterminate the regression for the dependent (element)you have been working on. You will be asked if youwant to enter the next dependent regression("CONT/YES" if yes) or you will pass it and go to thenext element (END/NO).

If the reply "CONT/YES" is entered, all points deletedat previous stages will be reinstated, and thedependent regression returns to the independentsquery:Example: 1. indep: FE ?

In this way, it is possible to stay in a loop (in a

deleting points and computing a new regression again,

Page 25: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

25

C = I *(K + SUM(K *I )) + B +i i i ij j i

SUM(B *I )ij j

j=1 j /=i

Terms: 1 2 3 4

NOTE: The R coefficient alone,however good, cannot be the only basisof determining the quality or goodnessof the model. Often in the case of toofew data points and/or over-complicating the calibration equation(too many independents), one can get a"perfect" R = 1.000, which may bemeaningless.

until the best possible regression model is found (see The fourth term reflects the change in backgroundSection 14.2). intensity under the analyte peak. This term is

When all elements in the model are finished, exit the the matrix scattering and its power to change theMOD program with "END/NO" which returns the general shape of the background, or more commonly,display to the prompt (>_). if the matrix element has a spectral peak that overlaps

7.5.9 Model Optimization Methodology

7.5.9.1 Basic Theory

The X-MET 880 uses the so-called empiricalcalibration approach, that is, the calibration equationsare developed based on intensities measured onknown calibration samples. The general calibrationformula is patterned after the intensity correctionmodel of Lucas-Tooth and Price. The generalized R - COEFFICIENT OF CORRELATION. Thiscalibration equation for the analyte i in a sample can calculation tells how well the given calibrationbe written as follows: equation explains the variation of data. Its value

The model assumes that the analyte concentration isa function of the intensities from elements in thesample. While this model is limited to relativelynarrow concentration ranges, it has an importantadvantage that only the analyte need be chemicallyassayed in order to develop it.

If no matrix effects are present in the sample, thecalibration equation would have only part 1 and 3, andwould be a simple straight line equation.

The second term modifies the slope of the calibration S - STANDARD ERROR OF THE FIT. Also knownline, according to the amounts of other elements under the term "Sum of Squares for Error" (SSE), orpresent in the sample, as determined by their "Root Mean Square Error" (RMS). This is oneintensities. This correction is needed when the matrix standard deviation of the spread of data points aroundelement is close in energy to the analyte, so that i t the fitted calibration line. The greater its valuestrongly absorbs or excites its radiation (matrix (error), the more scattered the data points. If two linesinterference). This slope modification is also useful are drawn parallel to the calibration line at plus S andwhen the matrix element varies over such a large minus S distance from it, the band created around theconcentration range that it significantly changes the calibration line will contain about 68% of all the data.effective matrix absorption of the incident or emitted The value of S is expressed in the units of thex-rays. The square term, I *I , would be the correction dependent variable being fit (generally concentrationi i

for self-absorption. units as in the case - mg/kg of soil). During the

significant if matrix element(s) vary enough to alter

the analyte peak. The overlapping peak may also bean escape peak. This situation often arises in alloyanalysis where strong peaks of iron, chromium ornickel can produce escape peaks (2.9 keV), which areto the low energy side of their fluorescent peaks(caused by argon filling).

7.5.9.2 Parameters Used inRegression Modeling

varies between 0 and 1. The closer its value is to1.000 (i.e., a one-to- one correlation between thecalculated element assay values and the correspondingelement intensity values), the better the calibrationmodel is. It should have a value greater than .65 to bemeaningful. The R value is a function of the numberof independents, the number of data points, andusually increases as those numbers increase.

Page 26: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

26

modeling one should try to minimize this parameter. mg/kg assay values are divided by 10 or 100 prior toIt should be noted that if the dependent mg/kg assay entry into the assay (ASY) table, the STDEV valuevalues are divided by 10 or 100 before being entered displayed must be multiplied by the same factor to getinto the assay (ASY) table, the S value displayed must mg/kg values.be multiplied by the same factor to get mg/kg values.

F - TEST VALUE. This is a statistic which isexpressed as the ratio of the sum of squares explainedby the given regression model, to the sum of squaresnot explained by the model (our familiar parameter S).If its value is low, say 5 to 10, then the model is notreliable, and may not be stable if a different data set isused for its calibration. Alternatively, we might saythat a low F value indicates the inability of a proposedregression model to predict the correlation betweenconcentrations and intensities of the element being fit.Assuming that an average calibration set has 10samples, the value of the F statistics should not besmaller than 10 for us to say with 95% confidence thatthe given model is a valid one. F is a function of thenumber of data points, the number of independentsand S. It usually decreases if the number of variablesincreases. It is a dimensionless number.

T - TEST VALUE. This statistic is a test ofsignificance for the calculated slope. It is expressedas the ratio of the slope value to the estimate of errorof that slope. For example, a T-value equal to 10,means that the error on the estimate of the slope is onetenth (or 10%) of the slope itself, which is significant.For an average case of ten calibration samples theabsolute value of T should not be smaller than 2.5 forus to say with 95% confidence that the given slope isstatistically valid. Therefore, the rule of thumb wouldbe to eliminate from the model all independentsfeaturing a T value less than |2.5| (±2.5). The T test isa dimensionless number and has the same sign as itsslope.

STDEV - STANDARD DEVIATION DUE TOCOUNTING STATISTICS. This is the error causedby the random nature of radioactive decay andcounting statistics. It is, by definition, equal to thesquare root of the total number of counts accumulatedwithin a given period of time. Alternatively, standarddeviation of intensity I is equal to the square root ofintensity I, divided by the square root of the time theintensity was measured for. It is seen that byextending the measurement time by a factor of 4, onecan reduce the standard deviation of counting (error)by a factor of 2. This error is the only one easilycontrolled. It should be noted that if the dependent

7.5.9.3 Iterative Process ofBuilding the Model

STEP 1

Always start with the intensity of the element ofinterest as the first independent variable. Next,depending on the nature of the sample matrix, try theintensity of the analyte ratioed to the intensity of thebackscattered (BS) radiation (XX/BS). This usuallyworks for sulfur in oil, chlorine in oil, metals in soil,or any element in a light (or highly scattering) matrix.If the analyte range is relatively wide (such as severalpercent), try the square term of the analyte intensity(XX*XX)-(example: Pb in soil in the range of 0 to10%). For each regression note the values of R, S,and F. The focus should be to MINIMIZE S as thisparameter is closely related to the accuracy of themethod. It can be shown mathematically that R and Fare functions of S.

Therefore, by concentrating on S we usually drive Rand F in the desired direction, that is high. Save theregression which yields the smaller S. However, keepin mind that two values of S, for example one of .12and another of .11, ARE FOR ALL PRACTICALPURPOSES THE SAME. The difference between theS's should be at least about 20% to be a criterion forselection. IF the S's are similar then select the modelwith the higher F and/or the higher R.

Examine the table of residuals. Make note of thepoints which are clearly off. The last column of thetable tells how many S's below or above (+/-) thecurve the given point is. About 68% of the data pointsshould have a STRES (last column) value smaller thanthe absolute 1, 95% smaller than the absolute 2, and99.9% (for practical reasons all points) smaller thanthe absolute 3.

Here is a worked example:

Total number of data points is 17 (17 samples).Therefore, about 12 points (.68*17 = 11.56 or 12points) should have a STRES (the last column) valuebetween -1.0 and +1.0.

Page 27: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

27

About 16 points (.95*17 = 16.15 or 16 points) should form of the product of the analyte and interferinghave a STRES value between -2.0 and +2.0. This also element intensities (XX*YY) may prove to be helpful.means that four points may fall between -2.0 and -1.0 Check the S, R, and F values again. If the S does notand between +1.0 and +2.0. improve, the independent should be rejected and

At the most, only one data point should have an S parameter, it is convenient to use the ratios of S'sabsolute value between 2.0 and 3.0. and F's from the two models. If one model is better

The number of points in each band may vary at leastby one from the predicted number due to randomness. F /F > (S /S ) ,

If any point is distant by more than ±(3S) from the where "better" refers to the smaller S value.line, it becomes a candidate for deletion. However ,the reason for deviation should be established first . Examine the table of residuals and the plot of the data.Often it is the incorrect entry of the referee assay Examine the t test values of the slopes. Reject(ASY command) for this point. Sometimes it just independent(s) which show T smaller than +/-2.5.may be a bad measurement which can be easilycorrected by remeasuring the sample(s). It may also Exit the modeling and enter the STD command. Thebe a case of the (outlying) sample being incompatible X-MET 880 should still have in its buffer memory the(i.e., different matrix) with the rest of the set. If al l spectrum of the last measured calibration sample.available information indicates that none of the above Therefore, it should output on the display the values ofapplies, then it is probably best to delete the point the standard deviations due to the counting statisticsfrom the regression. If deleting the "outlier" results in for all of the already modeled elements. Check andsignificant improvement of the model, then usually it make a note of those STD's. THEY CANNOT BEis justified. LARGER THAN THE S ERROR(S) OBTAINED

Often all data points will be within the ±(3S) band, model is over-corrected and should be changed bybut more of them than expected will exhibit large rejecting the variable (independent) with the lowest tdeviations from the line. In most cases this is caused value. The S error, by definition, includes theby incompleteness of the model. Examining the standard deviation, and cannot (be smaller than any ofcomposition of those samples which deviate too much its components.may reveal that those particular samples containsignificant amounts of some other element(s) which Continue the iterative process of modeling bywas not yet accounted for in the model. This situation repeating STEP 2 until no further improvement seemsprovides a hint as to which independent to include in feasible.the next round of the modeling process. Examine thetable of residuals and the plot of the data. Is anypoint, or group of points, indicating curvature of thecalibration line which could be corrected by includingin the model a square term (XX•XX)?

STEP 2

Now is the time to add to our first independent termthe next one. It may already be obvious what to dofrom the examination of the table of residuals asmentioned above in STEP 1. However, there aresome guidelines to keep in mind. The matrix elementwith the highest concentration should be examined, byincluding its intensity in the model as the nextvariable. If that element happens to be close in energyto the analyte peak, then the non-linear term in the

another one tried. To assess progress in reducing the

than the other one then:

better worse worse better2

DURING MODELING. If they are larger, then the

7.5.9.4 Concluding Remarks

Examination of the table of residuals during theiterative process is very important. During theexamination of the table of residuals, keep thefollowing in mind:

1. The results displayed in the table are in thesame concentration as the values entered intothe assay (ASY) table (i.e., values entered inthe assay table were mg/kg/10. Therefore,the values in the table of residuals must bemultiplied by 10 to get mg/kg.

2. A calibration based on several lowconcentration SSCS and one or two high

Page 28: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

28

concentration SSCS will report a very low analytical region. Additionally, SSCS in theS value. This low S value may be critical analytical region should be deletedmeaningless with respect to the analytical only if there is a large number of SSCS in theaccuracy of samples with concentrations region and the dilution of the point providesbetween the low and high SSCS. The significant improvement (greater than 20possibility of large analytical errors for percent) in the estimated values of the SSCSsamples in this concentration range should be in the region.brought to the attention of the task leader orproject manager prior to mobilization if Select as the final model the one with theSSCS cannot be obtained for this best (smallest) S and the best (most accurate)concentration range. estimated values for SSCS in the critical

3. Development of a model with SSCS similar S, select the one with greater value ofconcentrations greater than one or two F. If by including in the model one morepercent will generally result in increased independent(s), only a small improvement ofanalytical error in the region of the detection S is obtained, consider keeping the modellimit and the site action level (500 to 2000 with the larger S but with smaller STD value.mg/kg). Development of a low (DL to 1 or 2 In most cases, the smaller the number ofpercent concentration) and a high (one to ten independents in the model, the betterpercent concentration) concentration models (smaller error) the STD value.usually solves this problem. Therecalculation function (see "RECALC" Key Use common sense. Remember that simplerSection) can be used to recalculate is better. It is always easier to correct orconcentrations from spectra generated in the knowingly change the model which has theother model. smaller number of parameters. Do not over-

4. Pay particular attention to the SSCS with is too small, it is just not possible to developassay values in the critical analytical region. a good model. The rule of thumb is that ifThe critical analytical region is located you fit the equation with k independents, thebetween the DL and four or five times the number of samples in the model should be atsite action level. The goal of the operator least twice as much, that is N>=2•K. If theshould be to bring the estimated values off calibration set has only 6 samples and you fitthe SSCS in the region as close to their assay the calibration equation with 5 termsvalues as possible. Ideally, half of these (independents), then you will get a "perfectSSCS should have estimated values higher fit" with R=1.000, S=0.000 and F=****.then their assay values and half should have However, this "forced fit" will yield aestimated values lower then their assay calibration equation which will not work invalues. Select the model with the majority of practice, as it is not mathematicallythe SSCS estimated values above their assay representing the real-life scenario. This is avalues if an even distribution is impossible. case of over-defining the model on theThe goal is to provide the most accurate grounds of statistical analysis. analysis in the critical analytical regionbecause this is where managers are going to As a rule, do not bother to include asmake removal/remedial/health and independents those intensities whose elementsafety/sampling decisions. Additionally, concentrations in samples are very smallhigh-biased or false positive data is typically compared to the concentration range of thepreferred over low-biased or false negative analyte.data.

A higher S value and an independent with tried to spell out above. One of the mosta|T| value <2.5 can be tolerated if the model significant exceptions is the calibration forprovides improved accuracy in the critical an element which is very strongly affected by

analytical region. If two models have a very

correct. If the number of calibration samples

There are some exceptions from the rules we

Page 29: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

29

the presence of all matrix elements. A (see Section 5.3).classical example is the analysis ofphosphorus in phosphate rock. Phosphorusx-rays are absorbed by calcium, silica,potassium, sulfur, all of which may bepresent in the sample in quantities that callfor each to be included in the calibration.Usually, for that calibration the t values of allof the slopes are quite marginal. However,none of the variables can be rejected withoutsignificant degradation of the modelparameters.

When testing the final model for accuracy,keep in mind when comparing the measuredvalues to the given ones (X-MET 880computed), that if the difference seems toolarge, perhaps the measured sample was alsooff during the calibration and modeling.THEREFORE, COMPARE THEMEASURED RESULT WITH THE VALUEOBTAINED DURING MODELING (in thetable of residuals) RATHER THAN WITHTHE GIVEN ONE. If the differencebetween the S value and the STD value (forthe same measurement time) is significant,then the S error should be used as thecriterion for judging any discrepanciesbetween the measured and the given values.This applies if S is about twice thecorresponding STD.

7.5.10 Verifying the Accuracy of theRegression Model

The Regression Model should be verified wheneverpossible, prior to sample analysis, to ensure thateverything was done properly. Ideally, all SSCS used(not deleted) in the model should be measured by themodel to verify its accuracy. As a minimum, one low,mid and high SSCS should be used to verify the modelas described in Section 7.2. Read Section 9.0 onrecommended Quality Assurance and Controlmethods.

7.5.11 Documenting the RegressionModel

After the regression model has been completed for allthe dependents the results should be captured to aperipheral device such as a printer or a PC floppy diskusing ProComm+, or other communication software

8.0 CALCULATIONS

The X-MET 880 is a direct readout instrument. Theelement concentrations displayed in the readout areidentical to the assay value concentration enteredduring instrument calibration (see Section 7.5.7).

9.0 QUALITY ASSURANCE/QUALITY CONTROL

9.1 Precision

The precision of the method is monitored by readingthe low SSCS selected as described in Section 7.2 atthe start and end of sample analysis and afterapproximately every tenth sample. The lowconcentration sample is analyzed by the instrument forthe normal field analysis time, and the results arerecorded in a log book. The standard deviation foreach dependent element is calculated (using the N-1formula).

9.1.1 Preliminary Detection Limit (DL) andQuantitation Limit (QL)

A preliminary DL and QL is needed to give theoperator an indication of the instruments capabilityout in the field. A low SSCS sample is selected asdescribed in Section 7.2. Models with multipledependent elements may require the use of more thanone standard to obtain low concentration values foreach element.

The sample is measured ten times without moving it,using the anticipated field analysis measuring time.The standard deviation of the mean for eachdependent element is calculated (using the N-1formula).

If the standard deviation has a fractional component,round up to the next whole number prior to calculatingthe DL and QL.

The definition of the DL is three times the calculatedstandard deviation value.

The definition of the QL is 10 times the calculatedstandard deviation value.

Page 30: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

30

9.1.2 Field Detection Limit (FDL) andField Limit of Quantitation (FLOQ)

The precision of the method is monitored in the fieldby reading the low SSCS selected as described inSection 7.2 at the start and end of sample analysis andafter approximately every tenth sample. The lowconcentration sample is analyzed by the instrument forthe normal field analysis time, and the results arerecorded in a log book. The standard deviation foreach dependent element is calculated (using the N-1formula) using the measurements from the entiresampling period.

If the standard deviation has fractional component ,round up to the next whole number prior to calculatingthe FDL and FLOQ.

The definition of the FDL is three times the calculatedstandard deviation value.

The definition of the FLOQ is 10 times the calculatedstandard deviation value.

9.2 Reporting Results

All raw XRF data should be reported including theindividual results of multiple analyses of samples andsampling points. The average and standard deviation(using the N-1 formula) of each multiple analysisshould also be reported.

A "reported" value for each analysis or average ofmultiple analyses should be messaged in the followingmanner.

1. First round the value to the same degree ofsignificance contained in the SSCS sampleassay values (usually 2).

2. All values less than or equal to the FDL arereported not detected (ND).

3. All values greater than the FDL and less thanor equal to the FLOQ are flagged (usuallywith a "J" next to the reported value) andnoted as such.

4. All values above the FLOQ and within thelinear calibration range are reported as is. Other types of QA/QC verification should include

5. All values above the linear calibration range appropriate for the specific site to be assessed. This

(greater than the highest SSCS used in themodel) are flagged (usually with a "*" nextto the reported value) and noted as such.

9.3 Accuracy

The accuracy of the method is monitored and verifiedby sending an XRF analyzed sample or sample cupout for AA or ICP analysis at an independentlaboratory.

Although AA and ICP are generally recognized ashaving good accuracy and precision over theconcentration range typical of metals contamination insoil, it is most important to recognize the possibilityof real differences in the composition of samples sentin for comparative analysis, due to heterogeneity ofthe soil. It is recommended that the prepared samplecups be sent in for analysis if the samples wereprepared in this manner.

Another very important source of potential differencebetween XRF and AA or ICP results is incompletedigestion of the leaching technique. Since XRF is atotal elemental technique, any comparison withreferee results must account for the possibility ofvariable extraction depending upon the extractionmethod used and its ability to dissolve the mineralform in question.

9.3.1 Additional QA/QC

Additional QA/QC plans may call for monitoring forpotential instrumental drift by measuring a mid-rangecalibration sample at regular intervals (such as every20th sample) in order to validate the previousmeasurements. The use of such measures insures thatthe instrument is continuing to provide the same levelof accuracy throughout the entire series of samples ata given site. Should, for example a cold front blowthrough during a series of measurements, it is requiredthat the X-MET 880 be allowed a gain control periodof five minutes to compensate for temperature effectson the detector gain. The need for such a gain controlcycle would become apparent based on anyinaccuracies noted during the QA/QC sample tests.

9.3.2 Matrix Considerations

verification that the instrument calibration is

Page 31: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

31

should include verification of the potentially multiple COEFFICIENTS AND T VALUES.soil matrix types that may exist at a site. Matrixvariations that affect the XRF include large variations 5. Enter "CONT/YES" to enter the Coefficientsin calcium content, such as may be encountered when and T- values table and then P "CONT/YES"going from siliceous to calcareous soils, as well as to capture the table.variations in iron content.

10.0 DATA VALIDATION

10.1 Active Calibration ModelDocumentation Method

Immediately following a calibration, the CalibrationIntensities Table and the Assay Table are both active;all model information can be captured at this point .Once a new model is started these tables areoverwritten and can no longer be documented. In thiscase, go directly to the Alternate DocumentationMethod described in Section 10.2.

If using a PC, first set the program to capture to disk.Enter the LIM command. Entering P "CONT/YES"will output the channel limits in table form to theperipheral device for capture.

Enter the "CIN" command next. Entering P"CONT/YES" will output the calibration intensitiesfrom the sample calibration in table form to theperipheral device for capture.

Enter the "ASY" command next. Entering P"CONT/YES" after the first sample will output theassay values entered, in table form, to the peripheraldevice for capture.

Enter the "MOD" command next. For each regressionmodeled:

1. Scroll down through all independents usedwith "CONT/YES" and then exit with the"END/NO" command.

2. Press "CONT/YES" to enter the ResidualsTable.

3. Re-delete the points deleted and print out theresidual table by entering P "CONT/YES".Next enter D "CONT/YES" to capture thecalibration plot.

4. Press "END/NO" to exit the Residuals Table,and "END/NO" again to pass the "DeletePoints:" query and step to the

6. Enter "END/NO" to exit the coefficients andT-values table and display the questionCHANGE INDEPENDENTS?.

7. Enter "END/NO" to go to the nextDEPENDENT and repeat steps a - g.

After capturing the above calibration information,print a copy (if captured to floppy disk) and make aphotocopy (if captured directly to a thermal printer,the thermal paper will fade after a few months) andstaple the copy into the XRF log book for futurereference. On the next page, staple or tape a copy ofthe Assay Table results for future reference.Additionally, hand write in the margin of the AssayTable, the ID name or number of the calibrationsamples corresponding to the X-MET 880 order,which is the order in which the calibration sampleswere measured using the CAL command (next to thesample sequence # on the Assay Table form).

10.2 Alternate Calibration ModelDocumentation Method

If using a PC, first set the program to capture to disk.For models that no longer contain an activeCalibration Intensities Table or Assay Table, thebalance of the model calibration data can be capturedwith one command. The Enter Maintenance Programcommand, EMP, followed by "CONT/YES" and thenthe Parameters command, PRM, followed by"CONT/YES", will result in the X-MET 880 display:

MODEL XX?

Enter the model number to be documented and press"CONT/YES". The X-MET 880 will display:

GENERAL PARAMETERS?

Enter P "CONT/YES" and all required data for themodel will be output to the data capture device.

After capturing the above calibration information,print a copy (if captured to floppy disk), make aphotocopy (if captured directly to a thermal printer,the thermal paper will fade after a few months) and

Page 32: X-MET 880 FIELD PORTABLE X-RAY FLUORESCENCE … · X-Ray Fluorescence Spectroscopy (XRF) is a nondestructive qualitative and quantitative analytical technique used to determine the

32

NOTE: Alternatively electronic datacapture may be used. See Section 5.0.

staple the copy into the XRF log book for future 4. "X-ray Fluorescence Analysis ofreference. Environmental Samples", Dzubay, T., Ed,

10.3 Confirmation Samples

Confirmation samples are recommended at aminimum rate of 10%. Confirmation samples arerequired if QA2 data objectives have been establishedfor site activities. Ideally, the sample cup that wasanalyzed by XRF should be the same sample that issent for AA/ICP analysis. When confirming an in-situanalysis, collect a sample from a six-inch by six-incharea for both an XRF measurement and confirmationanalysis.

The XRF and metals results are analyzed with aregression analysis using either SAS orTM

Statgraphics software with the intercept forcedTM

through zero. The correlation factor between XRFand AA/ICP data should be 0.7 or greater.

10.4 Recording Results

Record all results and monitoring activities in alaboratory or field notebook.

11.0 HEALTH AND SAFETY

When working with potentially hazardous materials ,follow USEPA, OSHA, corporate and/or any otherapplicable health and safety practices.

12.0 REFERENCES

1. Outokumpu X-MET 880 Portable XRFAnalyzer Operating Instructions, Revision A,October, 1989.

2. Outokumpu Notes and General Informationon the Regulatory Requirements for ownersof the utokumpu X-MET 880.

3. Outokumpu Supplemental Documentation onthe factory calibration of X-MET 880 (ifcalibrated at Outokumpu prior to delivery),including Table of Check samples andNormalizing Samples for Assay Models.

Ann Arbor Science, 1977, p. 310.

5. "Advancements in Portable XRFTechnologies for On-Site Hazardous WasteScreening", Pasmore, J., Piorek S., andMcLaughlin, J.

6. "Portable X-ray Fluorescence as a ScreeningTool for Analysis of Heavy Metals in Soilsand Mine Wastes", Chappell, R., Davis, A.,Olsen, R. Proceedings ConferenceManagement of Uncontrolled HazardousWaste Sites, Washington, D.C., 1986, p 115.

7. "A New Calibration Technique for X-rayAnalyzers Used in Hazardous WasteScreening, Piorek, S., Rhodes, J.,Proceedings 5th National RCRA/SuperfundConference, April 1988, Las Vegas, NV.

8. "Data Quality Objectives for RemedialResponse Activities", EPA\540\G-87\004,March 1987.

9. "Portable X-ray Survey Meters for In-SituTrace Element Monitoring of AirParticulates", Rhodes, J., Stout, J., Schlinder,J., and Piorek, S., American Society forTesting and Materials, Special TechnicalPublication 786, 1982, pp. 70 - 82.

10. "In-Situ Analysis of Waste Water UsingPortable Pre-concentration Techniques and aPortable XRF Analyzer", Piorek, S., Rhodes,J., Presented at the Electron Microscopy andX-ray Applications to Environmental andOccupational Health Analysis Symposium,Penn. State Univ., Oct. 14 - 17, 1980.

11. "Hazardous Waste Screening Using aPortable X-ray Analyzer", Piorek, S.,Rhodes, J., Presented at the Symposium onWaste Minimization and EnvironmentalPrograms within D.O.D., American DefensePreparedness Assoc., Long Beach, CA.,April 1987.

12. "Field-Portable X-Ray Fluorescence", U.S.EPA/ERT Quality Assurance TechnicalInformation Bulletin, Vol. 1, No. 4, May1991.