HXMA WIGGLER MAINTENANCE REVIEW
Linda LinJune 2010
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
HXMA Wiggler Overview Installed 2004 Started operation January 2005 Has been operated for about 6.5 years
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Maintenance scheduled1. Top cold heads (re-condenser) exchange - once a year2. Bottom cold heads (shield cooler) exchange - depending on
temperature rise3. Compressor - replace absorber after about 18,000 hours of
operation4. Charge helium gas to compressor as needed5. Check and clean water cooling system as needed6. Service cold head valve as needed7. Refill wiggler with liquid helium about every 1.5 month (two
runs).
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Maintenance occurred (2005)Time Maintenance item Reason Scheduled Emergency
repair
2005-10 Replaced two top cold heads.
Replaced compressor ( top right)
Replaced & purged associated flex lines
First maintenance. Oil leak on
compressor (top right)
Yes
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Maintenance occurred (2006)Time Maintenance item Reason Scheduled Emergency
repair
2006-10 Replaced two top cold heads.
Changed absorbers
Purged flex lines Purged
compressors Changed motor
drive speed from 120 rpm to 90 rpm.
Routine maintenance
Reduce wear
yes
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Maintenance occurred (2007)Time Maintenance item Reason Scheduled Emergency
repair
2007-10 Replaced top right cold head with top mid cold head.
Started to use only four cold heads for HXMA wiggler cooling
Routine maintenance
Reduce annual maintenance cost
yes
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Maintenance occurred (2008)Time Maintenance item Reason Scheduled Emergency
repair
2008-04 Replaced two top cold heads.
Replaced two absorbers
Purged flex lines Replaced power
module Moved HXMA
wiggler to down stream Position
Routine maintenance
Improve Injection Efficiency
Yes(Schedule changed)
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Maintenance occurred (2008)Time Maintenance item Reason Scheduled Emergency
repair
2008-05 Replaced compressor (bottom right)
Compressor stopped working, used the spare one
yes
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Maintenance occurred (2009)Time Maintenance item Reason Scheduled Emergency
repair
2009-04 Replaced two top cold heads.
Replaced two compressors (top)
Replaced flex line (top left)
Purged flex lines Purged bottom
compressors
Routine maintenance
Oil contamination on compressors and flex lines
yes
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Maintenance occurred (2009*)Time Maintenance item Reason Scheduled Emergency
repair
2009-09 Replaced top left cold head
Replaced top left compressor
Replaced flex line
Top left cold head stop working. Tried spare compressor to run, still no good.
Oil contamination in compressor & flex lines
yes
*Wiggler stopped running in July.
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Maintenance occurred (2010)Time Maintenance item Reason Scheduled Emergency
repair
2010-04 Replaced two top cold heads.
Replaced compressor (top left)
Replaced three absorbers
Replaced flex line (top left)
Purged all flex lines
Purged bottom compressors
Routine maintenance
Oil contamination on compressor (top left) and associated flex lines
yes
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Summary of replacement:
Items/service Top left Top right Bottom left
Bottom right
Middle (spare)
Cold head 6 6 1 1 0
Compressor 3 2 0 1 2
Transfer lines 3 2 0 0 0
Absorber 2 2 2 2 1
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Observations: Cold heads worn in specific area
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Cold head wear on one side – 2009-09
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Cold head wear on one side – 2009-09
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Cold head was fine on the other side -2009-09
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Cold head wear on one side – 2010-04
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Cold head wear on one side – 2010-04
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Wiggler stands with two top cold heads housing sit in 30 degree angle
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Cold heads are mounted with 30 degree angle to wiggler vertical position
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Observations (Continued): Oil contamination on cold heads, compressors and
transfer lines
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Oil appeared in the surface of the cold head – 2009-09
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Observations (continued): Compressor’s internal settings were configured for
automatic re-start. The top right compressor was only able to be started at
120 rpm – would not start at a lower rpm. Ion pump is difficult to be changed Gaskets may become stiff during liquid Helium refill - heat
tape used.
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
There is no isolation valve between ion pump and wiggler
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
We have to plug the heat tapes to prevent leak from frost of rubber gasket
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Conclusion The wiggler runs at 1.9 Tesla successfully with the annual
maintenance/emergency repair and regular refill. The total times of cold heads’ replacement were met with
the schedule in 6.5 years. Compressors and transfer lines had lot more
replacements than we expected. The 30 degree angle seems to be a problem. Extra work on heat tape – each refill and maintenance.
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Conclusion (continued): Average maintenance cost is more than 3 times the
maintenance cost of the BMIT Wiggler* Frequency of maintenance is approximately 3 times that
of the BMIT Wiggler* Liquid Helium consumption is much higher than the BMIT
Wiggler*
*BMIT Wiggler has seen limited operation
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Acknowledgements: Xiaofeng, Shen
Terry Johnson
Ning Chen
Mark Besse
Harley Berndt
Larry Wiebe
Alex Sitnikov
www.lightsource.ca
HXMA WIGGLER MAINTENANCE REVIEW
Questions?