14
Voluson P8 Product Description A Voluson* Performance Series system delivers hard-working capabilities that fit your practice, your patient base and your budget…with an upgrade pathway that will help keep your system current for years to come. Highlights • Lightweight and portable • High Resolution Flat Panel Display 15” • Real-Time 4D • 3D Multiplanar Display • Automatic Optimization (AO) • CrossXBeamCRI* (CRI) • SonoNT • SonoRender Start • Anatomical M-Mode • On Board Archive including Preview and Pre-selection All options are not available in all regions. Extraordinary vision: Performance Series health net connections making connections in healthcare

Voluson_P8_Datasheet

Embed Size (px)

DESCRIPTION

USG Manual

Citation preview

Page 1: Voluson_P8_Datasheet

Voluson P8Product DescriptionA Voluson* Performance Series system delivers hard-working capabilities that fit yourpractice, your patient base and your budget…with an upgrade pathway that will helpkeep your system current for years to come.

Highlights• Lightweight and portable• High Resolution Flat Panel Display 15”• Real-Time 4D• 3D Multiplanar Display• Automatic Optimization (AO)• CrossXBeamCRI* (CRI)• SonoNT• SonoRender Start• Anatomical M-Mode• On Board Archive including Preview and Pre-selection

All options are not available in all regions.

Extraordinary vision: Performance Series

healthnetconnectionsmaking connections in healthcare

Page 2: Voluson_P8_Datasheet

General specificationsDimensions and weight

Height 1450mm(57in)

Width 420mm(16.5in)

Depth 670mm(26.4in)

Weight(noPeripherals) 143lbs/65kg

Electrical power

Voltage 100-120VAC,220-240VAC

Frequency 50/60Hz

Console design

3ActiveProbePorts

IntegratedHD(500GB)

3USBPortsforExternalPeripherals

2USBPortsforOn-boardPeripherals

RJ45LANPort

1HDMIOutPort

1AUDIOOutPort

User interfaceOperator keyboard

Ergonomic hard key layout

Interactiveback-lighting

Programmableprint/store/exportkeysforprinting, archiving and exporting

Monitor

15”HighResolutionFlatPanelScreen

Resolution:1024x768

Imagesize:800x566

Highbrightnesswith220cd/m2 typical

Manualbrightnessandcontrastadjustment

System overviewApplications

Obstetrics

Gynecology

Abdominal

Small-Parts

Vascular

Pediatrics

Cardiology

Urology

Musculoskeletal(MSK)

Operating modes

B-Mode(2D)

M-Mode(conventionalM-Mode)

AMM(AnatomicalM-Mode)–notavailableinsomecountries

PWDopplerwithhighPRF(PW)

HighPRFDopplerMode

ColorFlowDopplerMode(CFM)

PowerDopplerMode(PD)

HD-Flow*DopplerMode(HD-Flow)

M-ModeFlowModes(M/CF,M/HD-Flow)

VolumeMode(3D/4D):•3DStatic•4DRealTime•VCI-OmniView/SingleView–notavailableinsomecountries

ExtendedView(XTDView)

Scanning methods

ElectronicConvex

ElectronicLinear

Electronic Sector

MechanicVolumeSweep

Transducer types

ConvexArray

LinearArray

Phased Array

Volumeprobes‘4D’:•ConvexArray•MicroconvexArray

System standard features(some features can be optional or unavailable in some countries)

3D/4DMode

Innovativeuserinterfacewithonscreenmenus

HD-Flow

AutomaticTissueOptimization

CodedHarmonicImagingwithPulseInversionTechnology

SRI(SpeckleReductionImaging)

CrossXBeamCRI(CompoundResolutionImaging)

BModeonly

B+PowerDopplerMode

B+CFMDopplerMode

B+HD-FlowMode

B+CRI

B+CRI+CFM

Page 3: Voluson_P8_Datasheet

System overview (cont.)

System standard features (cont.)

(some features can be optional or unavailable in some countries)

B+CRI+PD

B+CRI+HD-Flow

HighResolutionZoom

PanZoom

Steering

VirtualConvex

Wide Angle on endovaginal probes

Beta-View

Patient information database

ImageArchiveonharddrive

3D/4Ddatacompression(lossy/lossless)

Real-timeautomaticDopplercalcs

Measurement&CalculationsincludingWorksheets/Reportfor:•OB •GYN•Vascular •Cardio•Abdominal •Small-Parts•Urology •Pediatrics•Musculoskeletal •Neurology

System options(some features can be standard in some countries)

Static3DMode

3D/4DAdvanced–notavailableinsomecountries

SonoRenderStart

SonoNT–notavailableinsomecountries

4DViewPCSoftware

DICOM3

XTD

AnatomicalM-mode–notavailableinsomecountries

Peripheral options

Integratedprinters:•B/Wprinter•Colorprinter•DVDrecorder

BluetoothLinePrinter

ECGDigitalModule

FootSwitchwithprogrammablefunctionality

Display modes

Simultaneous capability: •B+PW•B+CFM,B+PD,B+HD-Flow•B+M,B+AMM •B+3D,B+4D•B+CRI •B+SRI•B+SRI+CRI •B+SRI/4D+SRI•B+CRI/3D+CRI •B+CRI+SRI/3D+CRI+SRI•B+CRI/4D+CRI •B+SRI/4D+SRI•B+CRI+SRI/4D+CRI+SRI •B/B+SRI•B/CFM+CRI •B/CFM+SRI•B/CFM+CRI+SRI •B/PD+CRI

Display modes (cont.)

•B/PD+SRI •B/PD+CRI+SRI•B/HD-Flow+CRI •B/HD-Flow+SRI•B/HD-Flow+CRI+SRI

Real-timeTriplexmode: •B/CFM/PW •B/PD/PW•B/HD-Flow/PW •B/M/CFM•B/M/HD-Flow •B/AMM/CFM•B/AMM/HD-Flow

Selectable alternating modes: •B+PWorCW •B/CFM+PWorCW•B/PD+PWorCW •B/HD-Flow+PWorCW•B+CFMorPDorHD-FloworCW

Multi-image(split,quad):•Liveand/orfrozen•Split:B+B,B/CFM+B/CFMorB/PD+B/PDorB/HD-Flow+ B/HD-Flow•Split:B+PWorMorCW•Split:FrameReview/XTD-View•Quad:B+B+B+B,B/CFM+B/CFM+B/CFM+B/CFM orB/PDorB/HD-Flow•IndependentCineplayback•Quad:A+B+C+3Dor4D•TUI:1x1,1x2,2x2,3x2,3x3,3x4,4x4•Segmentation:quad(A/B/C/Segm.Object),single (Segm.Object)•Split:TUIOverview+1slice ZoomRead/Write(withorwithoutoverviewimage)

Colorizedimage:•ColorizedB •ColorizedM•ColorizedPW

Timelinedisplay:•IndependentDualB/PWDisplay•DisplayFormats•Top/Bottomselectableformat(Size:1/2:1/2;1/3:2/3;2/3:1/3) Left/Rightformat(Size:1/2:1/2)

Display annotation

Patient name:•Last:max32characters•First:max15characters•Middle:max15characters

ID:max32characters

SecondarypatientID(CitizenServiceNumber)

Accession#:max16characters

HospitalName:30characters

Sonographer(upto5charactersaredisplayeddependingonfontsize)

Gestationalage(OB)orLMP(Gyn)

Birthdate(selectable)

Date:3typesselectable•MM/DD/YYYY•DD.MM.YYYY•YYYY/MM/DD

Page 4: Voluson_P8_Datasheet

System overview (cont.)

Display annotation (cont.)

Time:2typesselectable•24hours •12hours

ProbeName

ApplicationName

GrayScalebar

DepthScale

FocalZonemarker

FrameRate

ZoomStart/Depth

B-Mode:•Userprogram •AcousticPower•ReceiverFrequency •Gain•DynamicContrast •GrayMap•EdgeEnhance •Persistence•SRI,CRI •FocalZoneMarkers•DepthScaleMarker •ProbeOrientation

M-Mode/AMM-Mode:•Gain •DynamicContrast •EdgeEnhance •Reject•M-Cursor,AMM-Cursor •TimeScale

DopplerMode:•AcousticPower •Gain•Angle •WallMotionFilter•VelocityorFrequencyScale •SpectrumInversion•TimeScale •PRF•HPRF •DopplerFrequency•SampleVolumeDepthandWidth

ColorFlowImagingMode(CFM,PD,HD-Flow):•AcousticPower •ColorGain•ColorBalance •ColorBalanceMarker•Quality •WallMotionFilter•PRF •ColorMap•ColorScale:kHz,cm/s,m/sPowerandSymmetrical VelocityImaging•SpectrumInversion

3D/4DMode:•3D/4DSubProgram •Threshold•Quality •VolumeBoxAngle•Mix •AcquisitionMode•Compression •OrientationMarkers

TGCCurve

CineFrameNumber

RecorderStatus

BodyPattern:130typesorganizedin10anatomicalgroups

MeasurementResults

DisplayedAcousticOutput:•TIS:ThermalIndexSoftTissue •TIC:ThermalIndexCranial(Bone)•TIB:ThermalIndexBone •MI:MechanicalIndex

PowerOutput

BiopsyGuideLine

ECGLine

Trackballfunction

Zoomoverviewimage(zoomboxposition)

GELogo

System parametersSystem setup

Pre-programmable categories date format

UserProgrammablePresetCapability,Userprogram,etc.

Languages:English,French,German,Spanish,Italian,Norwegian,Dutch,Finnish,Swedish,Danish,Chinese,Japanese,Russian

EUMLanguages:English,German,Spanish,Italian,French,Russian

Upto400ProgrammableAnnotationsorganizedin10 anatomical groups

ThreeprogrammablePxbuttonsfordocumentationpreferenceslikeSave,DICOMSend,Print,Check,Cinelengthetc.

Several user configurable functions:•ClinicName•Display(TGCcurve,ScreenLock,Screensaver,AutoScan Stop,Beeper,3D/4DScreenControls)•Trackballspeed•Dimfunction•Zoom:Overviewwindow•PatientInfodisplay•Titlebarsettings•StartExam&EndExamConfiguration

Measure setup

M&ASetupincludingAdd,Delete,EditandReorderof measure items

Application Setup including several parameters of Measurement,DopplerTraceandCalculationpresets

GlobalSetupincludingseveralparametersofMeasurement,CursorandResultwindowpresets

Biopsy setup

Userprogrammableneedleguidelines

Pre-processing

WriteZoomupto8x

B/M-Mode:•Gain •TGC•DynamicRange •AcousticOutput•TransmissionFocusPosition •TransmissionFocusNumber•TransmissionFrequency •EdgeEnhancement•PersistenceControl •LineDensityControl•Reject •SweepSpeed•M-Cursorposition

PW/CW-Mode:•Gain •DynamicRange•AcousticOutput •TransmissionFrequency•PRF •WallMotionFilter•SampleVolumeGate •Length,Depth,Pos•VelocityScale •SweepSpeed

ColorFlowMode(CFM,PD,HD-Flow):•AcousticOutput •PRF•WallMotionFilter •Linedensity •Ensemble •Smooth(RiseandFall) •Frequency •Balance •LineFilter •Quality •ArtifactSuppression •Gain •FlowResolution

Page 5: Voluson_P8_Datasheet

System parameters (cont.)

Post-processing

ReadZoom:0.8x-3.4xZoom(withHD-Zoomfunctionality upto22xZoom)

B-Mode:•2DGain •DynamicContrast•GrayMap •ColorizedB•SRIII(SpeckleReductionImaging)

M-Mode:•GrayMap •ColorizedM•DisplayFormat •SweepSpeed

PWMode:•GrayMap •BaseLineShift•AngleCorrection •ColorizedD•Scale(KHz,m/s,cm/s) •Trace•Invert •SweepSpeed

ColorFlowImagingModes(CFM,PD,HD-Flow):•ColorMap •DisplayThreshold•DisplayMode(V,V-T,T,P,P-T)(CFMonly)•Scale(CFMandHD-Flow) •Baseline

Image processing and presentation DigitalBeamformer

System processing channel counts up to 161,616 by configuration

MinimumDepthofField:0-1cm(Zoom,probedependent)

MaximumDepthofField:0-36cm(probedependent)

TransmissionFocus1-5FocusPointsselectable(probeand applicationdependent)

FocalZoneposition,upto7steps

ContinuousDynamicReceiveFocus/ContinuousDynamicRe-ceive Aperture

256shadesofgray

16.8MillionColors24bit

Upto255dBProcessingDynamicRange

ImageReverse:Right/Left,Rotation:0°,180°

Cine features Cinefeatures:•Dual/QuadImageCINEDisplay•CINEGaugeandCINEImagenumberdisplay•CINEReviewLoop•SelectableCINESequenceforCINEReview (byStartFrameandEndFrame)•SideChangeindualCINEMode•Measurements/Calculations&AnnotationsonCINE

Length:•2D:140MB:upto3minsandupto7,000frames(depending onB-imagesizeandFPS);typical:about1min./1000images (withcurvedarray:15cmdepth,angle81°,22FPS)•M-Mode:32MB:upto20minmotiontime(dependingon sweepspeedanddepth)•Dop.–Mode:32MB:upto10minmotiontime(dependingon sweepspeed)

Cine features (cont.)

Cineoperation:•Manual:imagebyimage•Autorun:speed:6to400%ofreal-timerate,playrepeat mode:forward–forward,forward-backward-forward

Image/volume storage

Imagedatastoredas:•RawDatafile(proprietaryformat)•DICOMfile(Single-orMulitframe)

Volume file stored as:•RawDatafile(proprietaryformat)•Size:typically:0.8-5MB (dependingonprobeandadjustedvolumesize)

Compression:•2D:JPEG,Lossless,high,mid,low•3D/4D:Lossyandlosslesscompressionavailable.Typical compressionratesare50%withlosslesscompression,15% withlossycompressionbutmaximumqualityand5%with lossycompressionandreducedquality(approximatevalues).

Review:ReviewofcurrentExamandarchiveddatasets(SingleImagesandCineClips).ViewFormat:Rawdata,DICOMdata.DisplayFormats:1x1,2x2,3x3

Reload:•Reloadofcurrent/archiveddatasets:2DRawData(incl. ColorDoppler,SpectralDopplerandM-Mode).3DRawData (SingleVolumeincl.Calc.Cines).4DRawData(VolumeCine)

Export as:•Bitmapfiles:BMP,TIFF,JPEG•Rawfiles:RAW(2D),VOL(Volumedata),4DV(RAW,VOLincl. Patientdata)•SequenceofBitmaps:BMP,AVI,MOV•DICOMFiles:DCM,DICOMFileswithDICOMDIR•3DRawData:conversiontoCartesianformatpossible

AVICodec:MPEG4,MSVideo1,FullFrames

Exportto:DVD+R(W),CD-R(W),Network,USBdevices

Export Anonymous function: yes, available for following image types:AVI,MOV,BMP,TIFF,JPEG

Backupfunctionto:DVD+R(W)/CD-R(W),Network,USBdevices

Reprofunction:Settingsrecall(e.g.Geometry,Gain,Colormap,etc.)fromastoredorreloadedpicture

ExamHistory:directaccesstoimagesfrompreviousexams;directaccesstoMeasureReportsimagesfrompreviousexams;Imagecomparewindowonscreentocompareimagesfromprevious exams with current exam image

HardDriveDataStoragesize:about430GB

Page 6: Voluson_P8_Datasheet

Image processing and presentation (cont.)

Connectivity

Ethernet network connection

WLANnetworkconnection

2USBPortsforharddisks/memorysticks

1USBPortforFootSwitch/Printer

DICOMsupport(option)•Verify•Print•Store•ModalityWorklist•StructuredReporting•StorageCommitment•MPPS(modalityperformedprocedurestep)•MediaExchange•Offnetwork/mobilestoragequeue•Query/Retrieve

Scanning ParametersB-Mode

BAcousticPower 1-100%

Scanangle changedin5º (probedependent)

Frequencyrange 1-18Mhz(Dependingonthe probe,3stepshigh,mid,low)

Framerate >740fps(dependingonprobe andapplication)

GAINrange +15to-15dB

Grayscalevalues 8bit

SRI 6steps(1-6)

CRI 8steps(1-8)

CRIfilter 4steps:off,low,mid,high

CE On/Off(Probedependent)

FFC On/Off(Probedependent)

Persistencefilter 8steps(pre)

Linefilter 3steps(pre)off, low(12,5/75/12,5%), high(25/50/25%)

LineDensity 3steps(pre)low,norm,high

Reject 51steps(pre)from0to225

Enhance 6steps(pre)0,1,2,3,4,5

Graymaps 21(18basicmapsand 3User-definedmaps)

Tintmaps 15

Dynamic 12differentdynamiccurves C1–C12

DisplayModes B,XTD

ScreenFormats•2DImaging:Single(B),Dual(B+B),Quad(B+B+B+B)•XTDView:Single(XTD),Dual(B+XTD)

M-Mode

Workingmodes M(conventionalM-Mode), AMM(AnatomicalM-Mode)

Frequencyrange 1-18Mhz(Dependingonthe probe,3stepshigh,mid,low)

MAcousticPower 1-100%

MGain ±15dBrange,1dBsteps

Msweepspeeds:•900/450/300/225/150/100pixels/sec;•26.44/13.22/8.81/6.61/4.40/2.94cm/sinrelation to system monitor

Review(memorytimes) >60s(32MB)

SignalprocessingM:•Dynamicrange:1to12 •Reject:0to255•Enhance:0to5 •Graymaps:18•Tintmaps:15

Displaymodes:•M:2D+M,2D+M/CFM,2D+M/HD-Flow,2D+M/PD•AMM:2D+AMM,2D/CFM+AMM/CFM,2D/HD-Flow+AMM/ HD-Flow

Screenformats(windowarrangement):•2D+Mand2D+AMM:up/down(horizontal):threedifferentsub formats40/60,50/50,60/40%left/right(vertical):50/50%•2D+AMM+AMM:left/rt-up/rt-down:50/25/25%

M-Color Flow Mode

AcousticMCFMPower 1-100%

Frequencyrange 1-18Mhz(Dependingonthe probe,3stepshigh,mid,low)

MCFMColorMaps 8maps

CFMGain ±16dBrange,1dBsteps

CFMVelocityScaleRange PRF:150Hzto13kHz

WallMotionFilter 8-3000Hz

Ensemble(colorshotsperline)8-16,stepsize1

Gentlecolorfilter

Smooth filter:•Rise:12steps•Fall:12steps

CFMSpectrumInversion

CFMBaselineShift 17steps

Pre-settableandindependentlyadjustableB-,MandMCFMGain

CFMThreshold 1-255steps

Balance 25-225,stepsize5

Artifactsuppression On/Off

ColorDisplayMode:•V(Velocity)•V-T(Velocity+Turbulence)•V-P(Velocity+Power)•T(Turbulence)•P-T(Power+Turbulence)

Real-timeTriplexMode:•B+M+MCFMinanydepth

Page 7: Voluson_P8_Datasheet

Scanning parameters (cont.)

Spectral Doppler Mode (PW, CW)

Operatingmodes: •PW(PulsedWaveDoppler,SingleGate) •CW(ContinuousWaveDoppler)

Frequencyrange 1.75-16Mhz(Dependingonthe probe,3stepshigh,mid,low)

PulseRepetitionFrequency(PRF): •PW-Doppler:0.9...22.0kHz •CW-Doppler:1.3...41.7kHz

SampleVolume(DopplerGate):•Length:0.7,1,2,3,4,5,6,7,8,9,10,15mm•Position:5mmtoB-scanend•Anglecorrection:-85°...0°...+85°

Powercontrolrange 1-100%

GAINrange PW+15to-25dB CW+15to-15dB

WMF(wallmotionfilter) PW:30...500Hz CW:30...1000Hz

Zerolineshift ±PRF/2,±8 steps

SpectrumAnalyzer FFT(FastFourier Transformation),max.256 channels,255-256amplitude levels

PW sweep speeds:•Simplex(26,44/13.22/8.81/6.61/4.40/2.94cm/s)•Duplex/Triplex(8.81/6.61/4.40/2.94cm/s)

•Review(memorytimes) >60s(32MB)

Measurableflowvelocities:•PW:1cm/s-8m/s(a=0°,2.0MHz,max.zeroshift) 1cm/s-16m/s(a=60°,2.0MHz,max.zeroshift) •CW:1cm/s–15.40m/s(a=0°,2.0MHz,max.zeroshift) 1cm/s–30.80m/s(a=60°,2.0MHz,max.zeroshift)

Signal processing: •Dynamicrange:15steps(10to40) •Graymaps:18basiccurvesand3User-defined(pre,post) •Tintmaps:5

Scale display: •Vert.:kHz,cm/s,m/s(selectable)•Hor.:1smarker(big),½smarker(small)

Screen formats: •2D/D:up/down(horizontal):threedifferentsubformats40/60, 50/50,60/40%left/right(vertical):50/50%.D:pencilprobesonly

Displayformats:•2D/D(duplexupdate,simultaneous);2D+CFM/D,2D+HD-Flow/D, 2D+PD/D (triplexupdate,PWorCW).2D+CFM/PW, 2D+PD/PW,2D+HDFlow/PW(triplexsimultaneous, PWonly)

Audio-Modes Stereo(bothdirections separatelyinbothchannels)

AudioVolume Adjustable,controldigipots

Color Doppler Mode

Screenformats 2D+CFM(Single,Dual,Quad)

Frequencyrange 1-16Mhz(Dependingonthe probe,3stepshigh,mid,low)

Displaymodes:•Triplexmode:2D+CFM/PW,2D/M+MCFM

Colorcoding:•Steps:65536colorsteps

Displaymodes:•V-T(velocity+turbulence) •V(velocity)•V-P(velocity+power) •T(turbulence)•P-T(power+turbulence)

Depthrange:•Axial:0toBscanrange •Lateral:0toBscanrange

Baselineshift 17steps(independentfrom spectralDoppler)

Inversionofcolordirection Yes

WallMotionFilter 8steps(low1,low2,mid1,mid2, high1,high2,max1,max2)

SmoothingFilter:•12stepsrisetime •12stepsfalltime

Gaincontrol +15dBto-15dB,0.2dBstep

LineDensity(colorlinedensity) 10steps

Ensemble(colorshotsperline):•CFM:7to31 •MCFM:8to16

FlowResolution 4steps(low,mid1,mid2,high)

Pulserepetitionfrequency:•CFM:150Hzto20.5kHz •MCFM:150Hzto20.5kHz

ColorMap 8differentcolorcodesfor each probe

Balance From25to225

Max.meas.velocity 4.23m/sec

Min.meas.velocity 0.3cm/sec

Scale kHz,cm/s,m/s

Automaticmovingtissue Yes suppression

Power Doppler Mode (PD)Screenformats 2D+PD(Single,Dual,Quad)

Frequencyrange 1-16Mhz(Dependingonthe probe,3stepshigh,mid,low)

DisplayModes:•Triplexmode:2D+PD/PW PDcoding 256colorsteps

PDwindowsize:•Lateral:maximumtominimumBmodescanangle•Axial:B-scanrange

Page 8: Voluson_P8_Datasheet

Scanning parameters (cont.)

Power Doppler Mode (PD) (cont.)

Displaymode P(power)

WallmotionFilter 8steps(low1,low2,mid1,mid2, high1,high2,max1,max2)

SmoothingFilter:•Riseedge:12steps •Falledge:12steps

Gaincontrol +15dBto-15dB,0.2dBsteps

PDEnsemble 7to31

PDLineDensity 10steps

Pulserepetitionfrequency 150Hzto20,5kHz

PDMap 8differentcolorcodesforeach probe

FlowResolution 4steps(low,mid1,mid2,high)

Balance From25to225in41steps

Artefactsuppression Yes

HD-Flow

Screenformats 2D+HDF(Single,Dual,Quad)

Frequencyrange 1-16Mhz(Dependingonthe probe,3stepshigh,mid,low)

Displaymodes:•Triplexmode:2D+HDF/PW;2D/M+MHDF

HD-FlowCodingSteps 256colorsteps

HD-Flowwindowsize:•Lateral:maximaltominimalBmodescanangle•Axial:B-scanrange

Displaymode P(power)

WallMotionFilter 8steps(low1,low2,mid1,mid2, high1,high2,max1,max2)

SmoothingFilter:•12stepsriseedge •12stepsfalledge

GainControl +15dBto-15dB,0.2dBsteps

HD-FlowEnsemble 7to31

HD-FlowLineDensity 10steps

PulseRepetitionFrequency 150Hzto20.5KHz

HD-FlowMap 8differentcolorcodesfor eachprobe)

FlowResolution 4steps(low,mid1,mid2,high)

Balance From25to225

Artifactsuppression Yes

Volume Scan Module

Vol.scansize:max.64MBforgrayvolumes;therequiredmemo-ryspacedependsonscanparameters(VOL-boxsizeandquality(low,mid1,mid2,high1,high2,max).Typical:0.8-5MB

Lines/2D-image:max.1024(typ.80to350)

2D-images/volume:Upto4096(AcquisitionModedependent)

VOL-Frames/sec.:max.40;theframeratedependsonscanparameters:VOL-Boxsize,qualityandprobe.

4DVolumeCine:upto128volumes

Displayofsectionalplaneimages:synchronouswithcontrolsetting, arbitrary movement in volume, monitored position in volume.

Rotation:360°,1°or3°increments(X-,Y-andZ-axis)

Magnification:adjustablefrom0.3toafactorof4.00

Acquisitionmodes:

3DStatic:•3D(2Dincl.CRI)

4D:•4DRealTime •SingleView

Visualizationmodes:•3DRendering(diversesurfaceandintensityprojectionmodes)•SonoRenderStart•SectionalPlanes•Multiplanar •SingleView(Option)•Niche•TUI •ThresholdVolume:measurevolumebelowandabovea threshold

Rendermodes:•Surfacetexture•SurfaceSmooth•SurfaceSkinandSmooth•Max-,min-andX-ray(averageintensityprojection)•Gradient•MixModeoftwoRenderModes

Displaygraphics:•Rotationaxis,centerpoint;•ROIbox,3DFrame;•Temporarydisplayofonscreencontrols(rotation,translation)

Graymaps:Slices:21(18basiccurvesand3User-defined(pre,post)3DImage:onegeneralmapadjustablewithbright(1-100)&contrast(1-100)

Tintmaps:Slices:15;3DImage:15

Depthrendermaps:3

Frequencyrange:1-18Mhz(Dependingontheprobe,3stepshigh,mid,low)

Page 9: Voluson_P8_Datasheet

Scanning featuresCoded Harmonic Imaging (CHI)

HarmonicImaging

Available on all probes

Compound Resolution Imaging (CRI)

Availableonallprobesexceptthe3Sc

Speckle Reduction Imaging

Available on all probes

Virtual Convex

Available on all linear probes

Wide Sector

Available on the following probes:•E8C-RS•RIC5-9W-RS

Measurements toolGeneric measurements

Distance:•Distance(PointtoPoint) •Distance(LinetoLine)•2DTrace(TraceLength) •2DTrace(PointLength)•Stenosis(%Dist)

Area/Circumference:•Ellipse •Trace(Line)•Trace(Point) •Stenosis(%Area)•Area(2Dist.)

Volume: following methods:•1Distance •1Ellipse•1Dist.+Ellipse •3Distance

Angle: •Angle(3Point) •Angle(2Line)

M-Mode:•Distance(PointtoPoint) •Time•Slope •HR •Stenosis(%Dist)

DopplerMode:

•Auto&ManualTrace: -PS(PeakSystole) -ED(EndDiastole) -MD(Min.Diastole) -PS/ED(Ratio) -PI(PulsatilityIndex) -RI(ResistanceIndex)•TAmax(Timeavg.max.Velocity)•TAmean(Timeavg.meanVelocity)•VTI(VelocityTimeIntegral)•Ductusvenosus:S,D,a,PI,PLI,PVIV•HeartRate•SingleMeasurements: - Velocity - Acceleration -RI -PI -PS/ED -Time -HR

CalculationsAbdomen calculations

Liver

Gallbladder

Pancreas

Spleen

Kidney(right/left)

RenalArtery(right/left)

Aorta(Proximal,Mid,Distal)

Portal Vein

Vessel

BladderVolume

SummaryReports

Small Part default calculations

Thyroid(right/left)

Testicle(right/left)

Vessel

SummaryReports

Small Part Breast calculations

Lesion1-5(right/left)

SummaryReports

Obstetrics calculations

FetalBiometry

EarlyGestation

FetalLongBones

NBL(NasalBoneLength)

FetalCranium

NTMethod:SonoNT/Manual

AFI

Uterus

Ovary

FetalDopplermeasurements(DuctusArt.,DuctusVen.,Ao,Carotid,MCA,CeliacArtery,SuperiorMesentericArtery,UmbilicalArt.,UmbilicalVein,UterineArt.,UmbilicalVein,FHR,AtrialFHR)

GestationalAgeCalculation

GestationalGrowthCalculation

FetalWeight(FW)Estimation

FetalTrendGraph

Multi-GestationalCalculation&FetalCompare

CalculationandRatios

FetalQualitativeDescription(Anatomicalsurvey)

FetalEnvironmentalDescription(Biophysicalprofile)

SummaryReports

Page 10: Voluson_P8_Datasheet

Calculations (cont.)

Obstetrics Fetal Echo

4-Chamber-view

Thorax

OutflowTract,Aorticarch

Venous

Tricuspidvalve

MitralValve

Aortic Valve

MainPulmonaryArtery

Pulmonary Valve

Aorta,DuctusArt.

UmbilicalVein,DuctusVen.

FHR

AtrialFHR

LVOT

RVOT

Pulmonary Veins

Carotid

SummaryReports

Cardiology

2DMode:•LVSimpson(Single&Bi-Plane)•Volume(AreaLength)•LV-Mass(Epi&EndoArea,LVLength)•LV(RVD,IVS,LVD,LVPW)•LVOTDiameter•RVOTDiameter•MV(DistA,DistB,Area)•TV(Diameter)•AV/LA(AorticValve/LeftAtrium)•PV(Diameter)

M-Mode:•LV(IVS,LVD,LVPW,RVD)•AV/LA(AoRootDiam,LADiam,AVCuspSep.,AoRootAmpl)•MV(D-E,E-FSlope,A-CInterval,EPSS)•HR(HeartRate),AtrialHR

D-Mode:•MV(MitralValve)•AV(AorticValve),TV(TricuspidValve)•PV(PulmonaryValve)•LVOT&RVOTDoppler(Left&RightVentricleOutflowTract)•PulmonicVeins•PAP(PulmonaryArteryPressuremeasurement)•HR(HeartRate)

C-Mode:•PISA•Tei-Index

Cardiology (cont.)

Others:•Diast.Vol(Bi) •Syst.Vol.(Bi)•StrokeVolume •VolumeFlow•CardiacOutput •EjectionFraction•FractionalShortening •MyocardialThickness•LA/AoRatio •E/APeak•PeakGradientAcceleration •MeanGradient•MeanGradientAcceleration •VTI•TVA •PG•PHT •MVA•AVA •ERO•CVP(CardioVascularProfile)Score

SummaryReports

Urology

Bladder

Prostate

Left/RightTesticle

Left/RightKidney

Left/RightRenalArtery

Left/RightDorsalPenileArtery

Vessel

SummaryReportsincl.PSAD,PPSA(1),PPSA(2)calculation

Vascular

Left/RightCCA(CommonCarotidArtery)

Left/RightICA(InternalCarotidArtery)

Left/RightECA(ExternalCarotidArtery)

Left/RightVertebralArtery

Left/RightSubclav.

Left/RightBulb

Vessels

SummaryReports

Gynecology

Uterus

Right/LeftOvary

Right/LeftFollicle

Fibroid

Endometrialthickness(Dist,DoubleDist.)

CervixLength

Left/RightOvarianArtery

Left/RightUterineArtery

Vessels

PelvicFloor

FHR

SummaryReports

Page 11: Voluson_P8_Datasheet

Calculations (cont.)

Pediatrics

Left/RightHipJoint

Pericallosal Artery

SummaryReport

Neurology

Left/RightACA(AnteriorCerebralArtery)

Left/RightMCA(MiddleCerebralArtery)

Left/RightPCA(PosteriorCerebralArtery)

BasilarArtery

A-Com.A(AnteriorCom.Artery)

P-Com.A(PosteriorCom.Artery)

Left/RightCCA(CommonCarotidArtery)

Left/RightICA(InternalCarotidArtery)

Left/RightVertebralArtery

Vessels

SummaryReports

OB tables

Age tables:•AC:ASUM,CFEF,Hadlock_82,Hadlock_84,Hansmann, Hobbins,Jeanty,JSUM,Kurmanavicius,Merz,Nicolaides, Shinozuka,Siriraj,Tokyo•AD:Persson•APAD:Merz•APTD:Hansmann•APTDxTTD:Shinozuka,Tokyo•BOD:Jeanty•BPD:ASUM,ASUM(old),Campbell,CFEF,Chitty(outer-outer) (outer-inner),Chitty,Hadlock_82,Hadlock_84,Hansmann, Hobbins,Jeanty,Johnsen,JSUM,Kurmanavicius,Kurtz, Persson,Merz,Nicolaides,OSAKA,Rempen,Sabbagha, Shinozuka,Siriraj,Tokyo,Verburg(outer-outer•CLAV:YARKONI•CRL:ASUM,ASUM(old),DAYA,Hadlock,Hansmann,JSUM, Persson,Nelson,OSAKA,Rempen,Robinson,Shinozuka,Tokyo, Verburg•EFW:Hadlock,JSUM2001,Osaka,Shinozuka,Tokyo•FL:ASUM,ASUM_OLD,CFEF,Chitty,Hadlock_82,Hadlock_84, Hansmann,Hobbins,Hohler,Jeanty,JSUM,Kurmanavicius, Persson,Merz,Nicolaides,O´Brien,OSAKA,Shinozuka,Siriraj, Tokyo,WARDA•FTA:OSAKA•FIB:Jeanty•GS:Hansmann,Hellman,Holländer,Rempen,Tokyo•HC:ASUM,CFEF,Chitty,Hadlock_82,Hadlock_84,Hansmann,Jeanty,Johnsen,Kurmanavicius,Merz,Nicolaides,Siriraj•HL:ASUM,Hobbins,Jeanty,Merz,OSAKA•LV:Tokyo•MAD:EIK-NES,Kurmanavicius•OFD:ASUM,Chitty,Hansmann,Jeanty,Kurmanavicius,Merz, Nicolaides•RAD:Jeanty,Merz

•TIB:Jeanty,Merz•TAD:CFEF,Merz,Chitty,Goldstein,HILL,Hobbins,Nicolaides, Hansmann•ULNA:Jeanty,Merz

Growthtables:•AC:ASUM,CFEF,Chitty,Hadlock,Hansmann,Jacot-Uillarmod, Jeanty,JSUM,Kurmanavicius,Lessoway,Merz,Nicolaides, Shinozuka,Siriraj,Tokyo,Verburg•AD:Persson•AFI:Moore•AortaVmax:Rizzo•APTDxTTD:Shinozuka,Tokyo•BOD:Jeanty•BPD:ASUM,Campbell,CFEF,Chitty,Hadlock,Hansmann,Jacot- Uillarmod,Jeanty,JSUM,Kurmanavicius,Lessoway,Persson, Merz,Nicolaides,OSAKA,Sabbagha,Shinozuka,Siriraj,Tokyo, Verburg•CLAV:YARKONI•CM:Nicolaides•CRL:ASUM,Hadlock,Hansmann,JSUM,Persson,OSAKA, Robinson,Shinozuka,Tokyo

•DVPI,DVPLI,DVPVIV,DVS/a:Baschat•EFW:Brenner,Doubilet,Hadlock,Hansmann,Hansmann(86), Hobbins/Persutte,JSUM2001,Persson,Osaka,,Shinozuka, Tokyo,Williams,Yarkoni(Twins),Ananth(Twins,Monochorionic), Ananth(TwinsDichorionic)•FL:ASUM,CFEF,Chitty,Chitty,Hadlock,Hansmann,Jacot- Uillarmod,Jeanty,JSUM,Kurmanavicius,Lessoway,Persson, Merz,Nicolaides,O´Brien,OSAKA,Shinozuka,Siriraj,Tokyo, Verburg,WARDA•FTA:OSAKA•FIB:Chitty,Jeanty,Siriraj•Foot:Chitty•GS:Hellman,Rempen,Tokyo•HC:ASUM,CFEF,Chitty,Hadlock,Hansmann,Jacot-Uillarmod, Jeanty,Kurmanavicius,Lessoway,Merz,Nicolaides,Siriraj, Verburg•HL:ASUM,Chitty,Jeanty,Merz,OSAKA,Siriraj•LV:Tokyo•MCAPI,RI:JSUM,Bahlman•MCAPV:Mari•MAD:EIK-NES,Kurmanavicius•MVE/A:HARADA•NBL:BUNDUKI,SONEK•OFD:ASUM,Chitty,Hansmann,Jeanty,Kurmanavicius, Merz,Nicolaides•MainPAVmax:Rizzo•RAD:Chitty,Jeanty,Merz,Siriraj•TAD:CFEF,JACOT-GUILLARMOD,Merz,•TCD:Goldstein,HILL,JACOT-GUILLARMOD,Nicolaides,Verburg•TIB:Chitty,Jeanty,Merz,Siriraj•TTD:Hansmann•TVE/A:HARADA•ULNA:Chitty,Jeanty,Merz,Siriraj•UmbArtPI:JSUM,Merz•UmbArtRI:JSUM,Merz,Kurmanavicius•FractionalLimbAvol/Tvol:Lee

Page 12: Voluson_P8_Datasheet

Calculations (cont.)

OB tables (cont.)

FetalweightExtimation(EFW):•Campbell(AC)•Hadlock(AC,BPD)•Hadlock1(AC,FL)•EFW•Hadlock2(BPD,AC,FL)•Hadlock3(HC,AC,FL)•Hadlock4(BPD,HC,AC,FL)•Hansmann(BPD,TTD)•Merz(AC,BPD)•Osaka(BPD,FTA,FL)•Persson(BPD,MAD,FL)•Persson2•Schild(HC,AC,FL)•Shepard(AC,BPD)•Shinozuka1(BPD,APTD,TTD,FL)•Shinozuka2(BPD,FL,AC)•Shinozuka3(BPD,APTD,TTD,LV)•Tokyo(BPD,APTD,TTD,FL)

Fetal ratios

CI(BPD/OFD)(Hadlock)

FL/AC(Hadlock)

FL/BPD(Hohler)

FL/HC(Hadlock)

HC/AC(Campbell)

Va/Hem(Nicolaides)

Va/Hem(Hansmann)

Vp/Hem(Nicolaides)

ProbesRAB2-5-RS

Applications Abdomen,Obstetrics, Gynecology

MaximumBandWidth(-20dB) 1-5MHz

NumberofElements 192

ConvexRadius 46.0mm

VolumeSweepRadius 22,6mm

FOV 80°(B),85°x80°(Volumescan)

Depth Max30cm

FootPrint 63.6x38.9mm

CenterFrequency 3.2MHz

DopplerTx.Frequency 3.22,2.95,2.38MHz

Harm.Tx.Frequency 2.13,2.0MHz

BiopsyGuideAvailable: •PEC74,Single-Angle,reusable •PEC78,Single-Angle,disposalwithreusablebracket

RIC5-9W-RS

Applications Obstetrics,Gynecology, Urology

BandWidth(-20dB) 4-9MHz

NumberofElements 192

ConvexRadius 11.6mm

VolumeSweepRadius 11.6mm

FOV 146°(B),146°x120° (Volumescan)

Wide 179°(B),179°x120° (Volumescan)

Depth Max16cm

FootPrint 22.4x22.6mm

CenterFrequency 6.6MHz

DopplerTx.Frequency 6.25,5.56,5.00MHz

Harm.Tx.Frequency 4.17,4.17,3.70MHz

BiopsyGuideAvailable PEC63,Single-Angle,reusable; disposable biopsy guide

4C-RS

Applications Abdomen,Obstetrics, Gynecology

MaximumBandWidth(-20dB) 2-5MHz

NumberofElements 128

ConvexRadius 60.0mm

VolumeSweepRadius n/a

FOV 58°

Depth Max30cm

FootPrint 68.7x18.3mm

Centerfrequency 3.1MHz

DopplerTx.Frequency 3.23,2.70,2.27MHz

Harm.Tx.Frequency 2.38,2.27,2.08MHz

BiopsyGuideAvailable Multi-Angle,disposable with reusable bracket

Page 13: Voluson_P8_Datasheet

Probes (cont.)E8C-RS

Applications Obstetrics,Gynecology, Urology

MaximumBandWidth(-20dB) 4-10MHz

NumberofElements 128

Convexradius 11.4mm

VolumeSweepRadius n/a

FOV 123°

Wide 161°

Depth Max16cm

FootPrint 22.1x10.7mm

CenterFrequency 6.5MHz

DopplerTx.Frequency. 6.67,5.88,4.76MHz

Harm.Tx.Frequency 4.55,4.17,3.70MHz

BiopsyGuideAvailable Single-Angle,Reusable andDisposable

12L-RS

Applications SmallParts,Obstetrics, Peripheral Vascular, Pediatrics, MSK

MaximumBandWidth(-20dB) 4-12MHz

NumberofElements 192

VolumeSweepRadius n/a

FOV(Width) 37mm

Depth Max11cm

FootPrint 47.2x13.8mm

CenterFrequency 7.7MHz

DopplerTx.Frequency 7.14,6.25,5.26MHz

Harm.Tx.Frequency 5.56,5.00,4.55MHz

BiopsyGuideAvailable Multi-Angle,disposable with reusable bracket

3Sc-RS

Applications Abdominal,Cardiology, Obstetrics,Pediatrics,Neurology

MaximumBandWidth(-20dB) 1-4MHz

NumberofElements 64

VolumeSweepRadius n/a

FOV(Width) 90°

Depth Max24cm

FootPrint 23.7x18.4mm

CenterFrequency 2.8MHz

DopplerTx.Frequency 2.5,2.17,1.85MHz

Harm.Tx.Frequency 2.0,1.92MHz

BiopsyGuideAvailable Multi-Angle,Reusable

External inputs and outputs Connectivity

HDMIOut

Network(RJ45)

ExternalAudioOut

USB(2xinfront,1xinrear)

ACPowerInput

Probe connector

Safety conformance

TheVolusonP8is:

ListedtoUL60601-1byaNationallyRecognizedTestLab

CertifiedtoCSA22.2,60601.1byanSCCaccreditedTestLab

CB-TestreportbyNationalCertificationBody

CEMarkedtoCouncilDirective93/42/EEConMedicalDevices

Conformstothefollowingstandardsforsafety:

IEC60601-1Electricalmedicalequipment

IEC60601-1-1Electricalmedicalequipment

IEC60601-1-2Electromagneticcompatibility

IEC60601-1-4Programmablemedicalsystems

IEC61157Declarationofacousticoutput

ISO10993Biologicalevaluationofmedicaldevices

NEMAUD3,UD2Acousticoutputdisplay(MI,TIS,TIB,TIC)

IEC60601-2-37Particularrequirementsforthesafetyof ultrasonicmedicaldiagnosticandmonitoringequipment

IEC60601-1-6Usability

Page 14: Voluson_P8_Datasheet

healthnetconnectionsmaking connections in healthcare

For more information visit hncl.net or e mail [email protected]

HealthNetConnections Ltd, Cavendish House, 36-40 Goldsworth Road, Woking, Surrey, GU21 6JT Tel: 01483 815120 Fax: 01483 815121