Upload
ali-rostami
View
216
Download
0
Embed Size (px)
Citation preview
7/30/2019 Viscosity Measuremant
1/8
Introduction
The purpose of this experiment is to provide an introduction to the use of the Cannon
Fenske viscometer. This viscometer requires calibration in order to relate the measured
values to the actual values. Thus, another purpose of this experiment is to introduce the
concept of calibration. The viscosity of liquids is quite sensitive to changes in temperature.
Data will be taken at different temperatures to investigate the effect of temperature on
viscosity.
The Cannon Fenske viscometer is a glass capillary viscometer. A diagram of the viscometer
is shown on Figure 1. At the beginning of the measurement, capillary tube R and bulb C
'are filled with the fluid to be tested. The liquid level in C is held slightly: above line E by
applying a slight vacuum to tube N and quickly placing your finger over the end of the
tube. When the vacuum is removed (remove finger from end of tube),
The fluid flows from bulb C through capillary tube R into bulb A. The time required for the
liquid to flow between marks E and F is related to the kinematics viscosity of the liquid by
Equation 1. The viscometer selected must have a minimum flow time of 200 seconds.
Where:
=Ct
=/ , kinematics viscosity, centistokes (ct)
C= viscometer constant
t= flow time, sec.
= viscosity, cp
= density, gm/cm3
7/30/2019 Viscosity Measuremant
2/8
The viscometer coefficient, C, is different for each viscometer and must be determined by
calibration. Calibration may be done with viscosity oil standards or using standard
viscometers.
Viscosity oil standards are oils which have known viscosities at specified temperatures. To
calibrate a viscometer, standard oil which has a flow time of at least 200 seconds is used.
Flow time is determined as described previously. Thus, from Equation (1) .
SSSStC /)/( =
(2)
A standard viscometer is one which has a known calibration C. This constant may have
been determined previously by using viscosity oil standards. Calibration of an uncelebrated
viscometer is done by determining the flow time for the same oil in the ca1ibrated and the
uncelebrated viscometers. .The viscometer constant for the uncelebrated viscometer can be
computed from Equation (3).
Cl = 1/)*( ttC
SS (3)
Where: C1 = constant of uncelebrated viscometer
t1 = efflux time at A fixed temperature for liquid in uncelebrated viscometer
SC
= Calibration constant of standard viscometer
ts = efflux time for same oil as used for t1 and at same temperature but in calibrated
viscometer.
S
= Viscosity, CP, of standard oil, centipoises S
= density of standard oil, gm/3
cm .
S
= density of standard oil, gm/3
cm .
Viscosity measurement
7/30/2019 Viscosity Measuremant
3/8
Cleaning the viscometer between successive determinations, clean the viscometer
thoroughly by several rinsing with toluene, followed by acetone. Dry the viscometer by
passing a slow stream of filtered dry air through the viscometer for 2 minutes or until the
last trace of solvent is removed.
Measurement of flow time
Charge the viscometer with the oil sample and mount in a constant temperature water bath
as described previously. When thermal equilibrium has been reached (15 min.). determine
the flow times for three trial runs.
Compute the kinematics viscosity from the viscometer constant and the three flow times.
Note: viscosities which differ by more than 0.35% from their mean are considered suspect.
Experimental Apparatus and Procedure
7/30/2019 Viscosity Measuremant
4/8
1. The viscometer should be cleaned with a suitable solvent and dry, clean, filtered air
should be blown through the viscometer to remove any remaining traces of solvent.
2. The instrument should be periodically cleaned 'with chromic acid to remove any possible
traces of organic deposits.
3. Jf a possibility of lint, dust, or other solid material is present In liquid sample, this may
be removed by filtering through a sintered glass filter or fine mesh screen. .
4. To introduce sample into the viscometer, Invert viscometer Immerse tube "A" into liquid
and apply suction to "J" which causes the sample to rise to etched line "E", Turn the
viscometer to normal position and wipe tube "A" clean.
5. Insert the viscometer into a holder and place in Constant. Temperature Bath, Allow 10
minutes for viscometer to reach equilation at 100 degrees F.(9 8.89C,) and 15 minutes at
210 degrees F. (98.89c), Or that time required to reach equilibrium.
6. Vertical alignment may be accompanied in bath by suspending a plumb bob In tube"J".
7. Apply suction to tube "A" and bring sample into bulb "D", a short distance above mark
"C".
8. The efflux time is measured by allowing the sample to flow freely through mark "C",
measuring the time for the meniscus to pass C" to E".
9. To repeat efflux time. repeat steps 7 and 8.
10. The Viscosity is calculated by multiplying the efflux time by the viscometer constant.
1. Constant temperature water baths (30 C, 40 C, 50 C)
2. Oil samples
7/30/2019 Viscosity Measuremant
5/8
3. Cannon Fenske viscometers and holding clamps
4. Electronic timer or stopwatch capable of measuring time to the nearest 0.1 sec.
5. Rubber bulb to supply vacuum and rubber hose
6. Thermometer
7. Standard oil for viscosity calibration.
8. Cleaning liquids such as toluene and acetone.
Procedure
You will measure the viscosity of two different oils at three different temperatures.
Temperature variations will be minimized by conducting the measurements in constant
temperature baths. Three constants temperature baths (30C,40C,and 50C) will be set up
and operating in the laboratory.
Select a viscometer which has a minimum flow time of 200 seconds at the calibration
temperature. Each viscometer has an identification number. Record this identification
number. Table 1 contains "approximate" viscometer constants for standard ASTM
viscometers. Note: we have the sizes identified with an asterisk. Clean the viscometer by
rinsing it with toluene followed by acetone. See cleaning procedure described later in this
section.
Calibration of viscometer
Bulb to the large tube (L) on the Cannon Fenske viscometer. Invert the viscometer and
apply suction to tube L with tube N immersed in the liquid sample. Draw the sample to
7/30/2019 Viscosity Measuremant
6/8
timing mark F and quickly turn the viscometer so that it is upright. Mount the viscometer in
the constant temperature bath by attaching a small water bath clamp to tube L. Tube L
should be mounted in a vertical position.
Allow 15 minutes for the sample to attain bath temperature. Record the bath temperature.
Attach the hose .on the rubber bulb to tube N and draw the oil through bulb C to a position
about 5 mm above the first timing mark (E). The liquid level can be temporarily held at this
point by holding your finger over the Choose standard oil. Supplied by the laboratory
assistant. The viscosities of the standard oils are on the labels of the bottles. Distilled water
may be used as a standard when its flow time is 200 seconds or more (Sizes 25, 50).
The quantity of standard oil is limited. It will be necessary for other students to use the
same oil in their calibration experiments, so return the oil to the bottle after use. Be careful
not to contaminate the standard oil with water or any other solvent. In normal laboratory
operations, the oil would be discarded after removal from the viscometer.
Charge the viscometer with the standard oil as follows. Connect a rubber end of tube N.
Remove the suction on tube N and allow the sample to flow freely through the viscometer.
Measure to within 0.2 sec. the time required for the meniscus to pass from the first timing
mark to the second. Measure the flow time for three independent runs and use the average
for the flow time for the sample.
Compute the viscometer constant C from the known kinematics viscosity and flow time
using Equation 2. This will be the constant for this viscometer when determining
kinematics viscosities for all oils which have a minimum flow time of 200 seconds.
There are corrections which can be applied to account for kinetic energy changes, surface
tension effects and thermal expansion of the viscometer due to use at a temperature above
the fill temperature.
7/30/2019 Viscosity Measuremant
7/8
The first correction is usually small. The surface tension correction is not needed when the
calibrating fluid is a standard" oil and the fluid to be tested is oil. You can investigate the
temperature variation of the viscometer constant by transferring the viscometer containing
the standard oil to one of the other temperature baths and detonating the flow time of the
standard oil as described previously. Do this for one viscometer and compare the
viscometer constants.
Return the standard oil to the appropriate bottle when calibration runs are finished.
Discussion and Recommendation
1. Try to use the correct viscometer size. Measuring ranges overlap, but the best
results will be obtained if you are operating near the center of the viscometer range.
If your sample has a viscosity of about 100 cSt, for example, the size 300 (range 50-
250 cSt) viscometer will give more accurate results than either size 200 (range 20-
100 cSt) or size 350 (range 100-500 cSt).
2. Some time is required to allow the sample to equilibrate at the temperature of the
water bath and to allow possibly occurring air bubbles to segregate. Therefore, after
loading the sample, wait at least 10 minutes before making measurements.
3. Avoid overloading the viscometer. Using either too much or too little sample will
produce inaccurate values of viscosity. The volume that fills the tube from E to the
inlet of the A tube will approximately half fill the bulb marked H.
7/30/2019 Viscosity Measuremant
8/8
References:
1. Amyx J.W., M.D Bass, and R.L. Whting, Petroleum Reservoir Engineering Mc
Graw Hill Book Co., (1960) p.4 and 7.
2. Product Safety Bureau, Reference Manual, Book 5, Laboratory Policies and
Procedures, p.C27-2.