System Fischerscope HM2000S Shimadzu DUH-211 (230V)Helmut-FischerGermany Japan
Measurement Theoretical Theoreticaltechnology based on calculation based on calculation
using IIT using IIT
Compliance ASTM E2546 ASTM E2546
Load range up to 2 N (203 gf)
Stage Manual
Microscope NoneNo image acquisition
PC Included with software
Unit price S$ 56,010.00
Demo test result
2014-10-29 (2.7 μm) 2458 HV with 0.3 μm maxpenetration at 40 mN
2014-11-3 (11 μm) 2645 HV with 0.4 μm maxpenetration at 80 mN
Remark Based on theoretical calculationin Instrumented Indentation TestASTM E2546.
Provides hardness vs load graphthat enables penetration point to
be determined.
Not based on indent diameter measurement, no image acquisition.
Need to adjust IIT hardness valueto optical method hardness value(using software).
Zwick Roell ZHVμ-SGermany
Opticalbased on the diagonaldiameter of indentation
ASTM E384
up to 19.6 N (2 kgf)
Manual
USB Camerawith 3 objective lens (x2.5, x10, x40)Included with software
S$ 49,510.00
Based on indent diameter measurement as of ASTM E384for Vickers micro hardness.
Provides more "actual" hardness value.
Sample details
Base material : Stainless steel
Coating film : iTAC / DLC
Coating thickness : 5 - 10 μm
HV : appx. 1700 - 2700 @ ≤100 mN
Instrumented Indentation Tester
Penetrationdepth (μm)
Possible "human error" if the indent mark is not captured properly.
Able to hold 2 indenters, in casewe need additional indenter in future.
Measuring technique
ASTM E2546 ASTM E384
Diagonal length of pyramid
Instrumented Indentation Tester
Load (mN)