System Fischerscope HM2000S Shimadzu Helmut-Fischer Germany Japan Measurement Theoretical Theoretic technology based on calculation based on using IIT using IIT Compliance ASTM E2546 ASTM E254 Load range up to 2 N (203 gf) Stage Manual Microscope None No image acquisition PC Included with software Unit price S$ 56,010.00 Demo test result 2014-10-29 (2.7 μm2458 HV with 0.3 μm max penetration at 40 mN 2014-11-3 (11 μm) 2645 HV with 0.4 μm max penetration at 80 mN Remark Based on theoretical calculation in Instrumented Indentation Test ASTM E2546. Provides hardness vs load graph that enables penetration point to

Vickers Hardness

  • Upload
    damien

  • View
    14

  • Download
    5

Embed Size (px)

DESCRIPTION

This is just a piece of rubbish to everybody

Citation preview

Page 1: Vickers Hardness

System Fischerscope HM2000S Shimadzu DUH-211 (230V)Helmut-FischerGermany Japan

Measurement Theoretical Theoreticaltechnology based on calculation based on calculation

using IIT using IIT

Compliance ASTM E2546 ASTM E2546

Load range up to 2 N (203 gf)

Stage Manual

Microscope NoneNo image acquisition

PC Included with software

Unit price S$ 56,010.00

Demo test result

2014-10-29 (2.7 μm) 2458 HV with 0.3 μm maxpenetration at 40 mN

2014-11-3 (11 μm) 2645 HV with 0.4 μm maxpenetration at 80 mN

Remark Based on theoretical calculationin Instrumented Indentation TestASTM E2546.

Provides hardness vs load graphthat enables penetration point to

Page 2: Vickers Hardness

be determined.

Not based on indent diameter measurement, no image acquisition.

Need to adjust IIT hardness valueto optical method hardness value(using software).

Page 3: Vickers Hardness

Zwick Roell ZHVμ-SGermany

Opticalbased on the diagonaldiameter of indentation

ASTM E384

up to 19.6 N (2 kgf)

Manual

USB Camerawith 3 objective lens (x2.5, x10, x40)Included with software

S$ 49,510.00

Based on indent diameter measurement as of ASTM E384for Vickers micro hardness.

Provides more "actual" hardness value.

Sample details

Base material : Stainless steel

Coating film : iTAC / DLC

Coating thickness : 5 - 10 μm

HV : appx. 1700 - 2700 @ ≤100 mN

Instrumented Indentation Tester

Penetrationdepth (μm)

Page 4: Vickers Hardness

Possible "human error" if the indent mark is not captured properly.

Able to hold 2 indenters, in casewe need additional indenter in future.

Page 5: Vickers Hardness

Measuring technique

ASTM E2546 ASTM E384

Diagonal length of pyramid

Instrumented Indentation Tester

Load (mN)