Vibration Control
Typ 663
English
Manual
¨
¨
¨
¨
Vibration Velocity (
2 Relay-Switching Outputs(Adjustable)
Analog Current Output: 4...20 mA
Frequency Range: 10 Hz...1000 Hz
mm/s, rms)
1 Hz...1000 Hz
Attention!Before Start-Up Procedure the Instruction Manual must be read and
understood!
HAUBER-Elektronik GmbH
Should any question arise, please contact:
HAUBER-Elektronik GmbHFabrikstraße 6D-72622 NuertingenGermanyPhone.: +49 (0) 7022 / 21750-0Fax: +49 (0) 7022 / [email protected]
Instruction Manual
Edition: 26.06.13
Vibration Control Typ 663Standard
Zone-1-21
Zone-2-22
1 Safety Informations .............................................................................4
2 Instruction Manual Scope...................................................................5
3 The Vibration Control Typ 663 ...........................................................5
4 Intended Use ........................................................................................5
5 Safety Level..........................................................................................5
6 Documents and Certificates ...............................................................5
7 Responsibility for the Safe Operation / Disclaimer ..........................6
8 Application Fields................................................................................6
9 Delivery Contents ................................................................................6
10 Electrical Data....................................................................................7
11 Mechanical Data ................................................................................9
12 Connections.....................................................................................10
13 Functional Description....................................................................11
14 Limit Value Adjustment...................................................................12
15 Self-Test ...........................................................................................12
16 Mounting and Demounting .............................................................13
16.1 Fastening at the Mounting Surface ...............................................................13
16.2 Zone-2-22 - Fastening Safety Clip / Protective Cover ...................................14
17 Installation and Start-Up .................................................................15
18 Maintenance and Repairs ...............................................................15
19 Grounding Concepts.......................................................................16
HAUBER-Elektronik GmbH
3
Content
Standard, Zone-1-21, Zone-2-22
HAUBER-Elektronik GmbH
1 Safety Informations
In General
The safety instructions serve the protection of persons and things from damage and danger that
arise from not intended use and further misuse of products especially in explosion endangered
areas. Therefore read the instruction manual carefully, before working with or starting-up the
product. To the operating personnel the instruction manual has to be accessable anytime.
Before the starting-up or miscellaneous works with the product please check, wether all the
documents are available completely. If not all the documents are committed completely or futher
copies are required, they can be obtained in different languages.
Our product is designed to the latest state of the art. Nevertheless there are a number of residual
risks. This means that each person in the operators firm, concerned with mounting and
dismounting, installation, start-up, operating or maintanance of the product, has to have read and
understood the instruction manual.This means furthermore that each person in the operators firm, concerned with mounting and
dismounting, installation, start-up, operating or maintanance of the product, has to be an
authorized expert, familiar with the safety instructions for handling electrical components. For
handling ATEX-certified products within explosion endangered surroundings the expert in addition
has to be familiar with the safety instructions relevant there.
This symbol indicates an explosion hazard.
Used Symbols
This symbol indicates a risk from electrical current.
This symbol indicates a (non-safety relevant) information.
4
HAUBER-Elektronik GmbH
5
2 Instruction Manual Scope
The present instruction manual of the Vibration Control Type 663 is applicable for the variants:
Standard, Zone-1-21 and Zone-2-22.
The functionality of the variants is identical. In addition the variants have certifications and
labellings, that allow operation in explosion endangered areas. (see chap. 7, Operation Areas)
3 Vibration Control Typ 663
5 Safety Level
The Type 663 is tagged by subsequent safety levels:
- Safety Integrity Level, SIL2, SIL3 / FMEDA, nach IEC 61508
- Performance Level, PL-d, nach DIN ISO 13849
4 Intended Use
The Type 663 serves as protection for machines and mechanical equipment against undue
mechanical vibrations. It exclusively serves for measurement of mechanical vibrations.
Industrial fans, ventilators, blowers, electric motors, pumps,
centrifuges, seperators, generators, turbines, and similar mechanical equipment.
Main areas of application:
Subsequent Type 663 Documents und Certifikates can be consulted on
www.hauber-elektronik.de:
EC-Conformity-Declaration
Certificate - Safety Integrity Level, SIL2, SIL3 / FMEDA
Certificate - Performance Level, PL-d
EC-Type-Examination-Certificate ATEX-Zone 1 und 21, no.: SNCH 09 ATEX 4380
Statement-of-Conformity ATEX-Zone 2 und 22, no.: LU 09 ATEX 0065X
?
?
?
?
?
6 Documents and Certificates
The Vibration Control Typ 663 ist applied for measurement and control of machines absolute
bearing vibration, referring to DIN ISO 10816. It offers the following features:
• Two limit values and two associated delays are seperately adjustable.
• An exceeding of the adjustable limit value is signalled via relay outputs.
This can be used to generate a pre- and a main alarm.
• Measurement parameter: The root mean square (rms) of the vibration velocity (mm/s).
• Analoge current output: Interference-free direct current from 4...20 mA, proportional to the
vibration amplitude.
• Cable break at the control cable detectable by a succeeding evaluation device:
Value of the direct current signal < 3,5 mA.
Variant Delivery Contents
Available Supplies:
• Allocable Mating Connector, M12, 8-pole
• Connection Cable, M12-Socket, 8-pole, 0,25 mm2,
L= 2 m, 5 m oder 10 m
• EMC-Adapter
• Vibration Control Typ 663
• Cylinder Bolt, M8 x 20 mm
• Spring Ring, M8
• Instruction Manual
Standard
• Vibration Control Typ 663
• Protective Cover for M12-Plug
• Safety Clip
• Cylinder Bolt, M8 x 20 mm
• Spring Ring, M8
• Instruction Manual
• Vibration Control Typ 663,
Integrated Cable, L= 2, 5, 10, 25 m or on request
• Cylinder Bolt, M8 x 20 mm
• Spring Ring, M8
• Instruction Manual
Zone-2-22
Zone-1-21
9 Delivery Contents
8 Application Fields
6
HAUBER-Elektronik GmbH
The correct layout of the electrical plant under conditions of explosion protection, as well as the
correct switch on procedure, is the sole responsibility of the user of the plant.
The current valid explosion protection rules and security regulations must be adhered to and
must be under given circumstances checked by a competent person. Should the plant on the
order of the user be erected by a subcontractor, the plant must only be switched on after the
subcontractor has submitted an installation certificate as prove of the correct nature of the
installation, according to the relevant valid regulations.
The primary switch on of explosion protected plants or part of plants, as well as the subsequent
switch on after major adjustments or maintenance work, must be reported to the relevant
authorities by the owner.
7 Responsibility for the Safe Operation / Disclaimer
Variante Application Fields
None explosion
endangered areasStandard
Explosion endangered
areas Zone 1 and 21
Explosion endangered
areas Zone 2 and 22
Zone-1-21
Zone-2-22
Labelling
II 2 G Ex db IIC T4
II 2 D Ex tb IIIC T120 °C IP6X
II 3 G Ex nCc IIC T4
II 3 D Ex tc IIIC T125 °C IP6X
HAUBER-Elektronik GmbH
7
10 Electrical Data
The housing cover may only be unscrewed, if the Type 663 is either seperated from themains or no explosion endangered atmosphere exists. Otherwise danger of explosionbecause of sparkling, when the ATEX-certified Type 663 is operating in explosionendangered areas!
8
16
32
64
128
256
10 Hz...1000 Hz
1 Hz ...1000 Hz
Before Starting-Up the mains must be secured with a microfuse ()!
Type 663, time delay,160 mA, breaking capacity C
Measuring range: 0...
Measuring accuracy: 5%
Transverse sensitivity: < 5 %
mm/s
0... mm/s
0... mm/s
0... mm/s
0... mm/s
0... mm/s
Frequency range: (Standard)
(optionally)
Output signals: 1 x 4...20 mA
2 x Relay contact (Pre- and Mainalarm)
Relay switching load: 1A / 30V DC
Voltage supply: 24V DC 10%
Power input (max.): 80 mA
Surroundings temperatur: -30°C...+60°C
Operating temperatur range: -30°C...+85°C (Meas.-head-temp. a.t. fastening)
Burden/load (max.): 500
Fusing: Microfuse
"
"
S
(time delay, 160 mA, breaking capacity C)
Automatic: As soon as the vibration values fall below the limit
values, the relays pick up again.automatically
• Each Type 663 hasof the listed
measuring ranges.one
• Further measuringranges on request.
• Please indicate themeasuring range inyour order.
HAUBER-Elektronik GmbH
Arbeitsbereich interner Sensorbaustein
0
20
40
60
80
100
120
140
160
180
200
0 100 200 300 400 500 600 700 800 900 1000
Frequenz (Hz)
Sc
hw
ing
ge
sc
hw
ind
igk
eit
(m
m/s
)
Operating Range
Operating Range of Vibration Control Type 663
Reading example:
Frequency Max. measurable(Hz) Vibration Velocity
100 160400 40
1000 18
The operating range is independent of the measuring range. The diagram shows that withincreasing frequency the height of the measurable vibration velocity decreases.
Vib
rati
on
Velo
cit
y (
mm
/s)
Frequency (Hz)
Frequency Range 10 Hz...1000 Hz
-40
-35
-30
-25
-20
-15
-10
-5
0
5
1 10 100 1000 10000 Hz
dB
Frequency Range 1 Hz...1000 Hz
-40
-35
-30
-25
-20
-15
-10
-5
0
5
0,1 1,0 10,0 100,0 1000,0 10000,0 Hz
dB
8
Housing Material: Stainless Steel V2A; material no: 1.4305
M12-connector material: CuZn (brass), nickel plated
Cable gland: Stainless Steel V2A
Fastening: Inner hexagon screw, M8 x 20 mm,
Threat pitch: 1,25 mm
Weight: ca. 500 g
Protection Style: IP 67
Securing: The sensor must be earthed via the M8
fastening (see chapt.16).
HAUBER-Elektronik GmbH
9
11 Mechanical Data
Housing Dimensions and Direction of Measurement
Standard, Zone-2-22 Zone-1-21
Note: Direction of Measurement = Direction of Fastening
30
M12-Plug
Connection Cable
Direction ofmeasurement
Cable Gland
Direction ofmeasurement
Measures in mm.
Housing Dimensions and Direction ofMeasurement: Standard, Zone-2-22
Housing Dimensions and Direction ofMeasurement: Zone-1-21
HAUBER-Elektronik GmbH
10
Connection Plan for all 3 Variants
12 Connections
Note: Displayed in the Connection Plan is the Alarm Condition resp. Currentless Condition!The Relays 1 and 2 are dropped out. (Further informations on operating conditions, page 11.)
Connection Cable Socket(Supplies)
M12-Plug
2
3
4
7 8
56
1
M12-Plug, 8-pole.Pin allocation see Connection Plan.
Connection cable socket M12, 8-pole,0,25 mm². Pin allocation see Connection Plan.
Standard, Zone-2-22 Zone-1-21
Connection Cable
PUR-sheathed-cable, Ø: ca. 6,5 mm,8 -pole 0,25 mm².Pin allocation see Connection Plan.
Evaluation-Device
Type 663
Pin 1: white
Pin 2: brown
Pin 3: green
Pin 4: yellow
Pin 5: grey
Self-Checkswitch
Relay 2
Relay 1
Evaluation of the 4...20 mAcurrent signal via amperemeteror PLC-control
Pin 6: pink
Pin 7: blue
Pin 8: red
1
2
3
4
5
6
7
8
WH
BN
GN
YE
GY
PK
BU
RD
ALARMWARNING
OK
HAUBER-Elektronik GmbH
13 Functional Description
The Typ 663 consistes of two channels LIM1 and LIM2 independent from each other. Both
channels are constructed identically.
For both channels the Limit Value and the Delay Time can be adjusted seperately.
The exceeding of the adjustable Limit Value is signalled via relay outputs. This can
be used to generate a pre- and a main alarm.
In addition the Typ 663 has a analogue current output. This delivers a direct current
from 4...20 mA proportional to the vibration amplitude.
User Interface Channels LIM1 and LIM2 - Controls and LED-Indicators
Operating Conditions Channels LIM1 and LIM2
SET-Turn-SwitchLimit Value in 15 Steps(Table: Limit Value, page 12)
Time-PotentiometerDelay-Time -stepless, 0...30 sec.
- red- yellow- green
Operating Condition Measurement Relais LED-Anzeige
OK < Limit Value energized OK
WARNING> Limit Value,
Delay Time runningenergized WARNING+OK
ALARM> Limit Value,
Delay Time expireddropped out ALARM
The housing cover may only be unscrewed, if the Type 663 is either seperated from themains or no explosion endangered atmosphere exists. Otherwise danger of explosionbecause of sparkling, when the ATEX-certified Type 663 is operating in explosionendangered areas!
11
Measuring Range e.g.:
SET-Turn-Switch Pos.:
Limit Value:
HAUBER-Elektronik GmbH
12
15 Self-Test
Example: Limit Value Adjustment
14 Limit Value Adjustment
0...32 mm/s
16 mm/s
8
0 0 0 0 0 0 0
1 0,5 1 2 4 8 16
2 1 2 4 8 16 32
3 1,5 3 6 12 24 48
4 2 4 8 16 32 64
5 2,5 5 10 20 40 80
6 3 6 12 24 48 96
7 3,5 7 14 28 56 112
8 4 8 16 32 64 128
9 4,5 9 18 36 72 144
10 5 10 20 40 80 160
11 5,5 11 22 44 88 176
12 6 12 24 48 96 192
13 6,5 13 26 52 104 20814 7 14 28 56 112 22415 7,5 15 30 60 120 240
SET-
Turn-Switch
Position
Limit Value
Range
0…8 mm/s
Range
0…16 mm/s
Range
0…32 mm/s
Range
0…64 mm/s
Range
0…128 mm/s
Range
0…256 mm/s
The Self-Test is to verify the correct function of the vibration control. Therefore a Testsignal
is applied to pin 4 of the control. The Testsignal is simulating a vibration amplitude, that exceeds
the maximum limit value. So that the control will start the Alarming.
Direct current, 24 V DC or
Rectangular signal, 24 V DC / 0,5 Hz
Normal Operation
The Testsignal is applied to pin 4.
The Output-Currentsignal increases up to ~23 mA.
After < 35 s the Alarmrelays switch.
The Testsignal will be removed.
After further < 60 s the Output-Currentsignal has
decreased to 4 mA or to the actual value.
Normal Operation
approx. 95 s
Testsignals: a.
b.
Initial State:
Start Selftest: •
•
End Selftest.
Final State:
Duration:
•
•
may be used only from Productionnumber 75478 on.
à
à
à
Testsignal a
In normal operation Pin 4 shall be wired to GND (Risk of elektromagnetic influences!).
HAUBER-Elektronik GmbH
16 Mounting and Dismounting
Mounting and Dismounting works at and with the Control may only be executed by an authorized expert,familiar with the safety instructions for handling electrical components. For handling ATEX-certified controlswithin explosion endangered surroundings the expert in addition has to be familiar with the safetyinstructions relevant there.
Preconditions
?
?
Mounting surface clean and flat,i.e. free from paint, rust, etc.
Threaded hole at the Mountingsurface:Depth: 15 mmThread: M8
Unscrew housing cover from housing
base.
Allen wrench, SW8?
Fasten control friction-locked at
the mounting surface by use of
allen screw and snap ring.
Allen wrench, SW6?
Tighten housing cover on housing base
loosely by hand.
(Do not tilt the thread!)
Tighten housing cover with a tightening
torque = 5 Nm.
Torque wrench SW8?
Working Steps
1.
2.
3.
?
?
?
?
Allen wrench, SW6, SW8
Torque wrench SW8
Allen screw M8x20
Snapring for M8
Tools and Materials
Note: ex factoryTo avoid a possible galling of the housing cover with the housing base,the thread is treated with a mounting paste for high-grade steel connections .
Before mounting and dismounting works the control has to be seperated from the mains!Seperated plug and socket devices always have to be disconnected from the mains!Otherwise danger of explosion because of sparkling, when operating ATEX-certified Types663 in explosion endangered areas!
The control housing must be earthed via ist fastening - i. e. via machine earth or via aseperate earth wire (PE)!
16.1 Fastening at the Mounting Surface
Fastening at the mounting surface
Mounting Surface
zu 5 Nm
Tightening housing cover with a torque wrench
13
HAUBER-Elektronik GmbH
Half Shell Arrow and Eye
Sign
Fastened Safety Clip
Protective Cover Fastened Protective Cover
Safety Clip
Plug in the connection cable socket into the M12-plug completely.
(Pay attention to the code cam!).
16.2 Zone-2-22 - Fastening Safety Clip / Protective Cover
1. Disconnect mains / electric circuit.
2. Separate both shell halves of the fuse clip with a screw driver.
3. Fasten protective cover and skrew it tightly onto the sensor plug.
1.
2.
3.
Tighten firmly the lock-nut of the connection cable socket by hand.
Fasten the safety clip against accidental disconnection of the plug connection:
1. Put both shell halves of the safety clip around the plug connection.
2. Press together by hand both shell halves of the safety clip until the
catch lock snaps in.
3. Put the arrow connected to one shell halve around the cable, then stick
it through the eye on the other end, so that the notice sign is readable
alongside the cable.
Fastening Protective Cover
Fastening Safety Clip
Disassemble the safety clip and mount the protective cover:
The operation of variant Zone-2-22 is not permitted without the safety clip, to avoidaccidently disconnecting the plug-in connection! Otherwise danger of explosion because ofsparkling, when operating in explosion endangered areas!
After disconnecting the plug connection the protective cover has to be mounted!
14
HAUBER-Elektronik GmbH
15
17 Installation and Start-Up
18 Maintenance and Repair
Note: The Type 663 and ist variants are maintenance free!
Errortable
Installing and starting-up the control may only be executed by an authorized expert, familiar with thesafety instructions for handling electrical components. For handling ATEX-certified controls withinexplosion endangered surroundings the expert in addition has to be familiar with the safetyinstructions relevant there.
Starting-up may only be executed with correctly tightened housing cover (tightening torque= 5 Nm)! Otherwise danger of explosion because of sparkling, when operating ATEX-certified Types 663 in explosion endangered areas!
Prior to starting-up the Type 663, the mains must be secured with a microfuse ()!
time delay,160 mA, breaking capacity C
The connection cable and possible extension cables must be protected against electricalinfluenzes and mechanical damages. Here local regulations and commissions absolutelyhave to be considered.
Repairing the control may only be executed by an authorized expert, familiar with the safetyinstructions for handling electrical components. For handling ATEX-certified controls withinexplosion endangered surroundings the expert in addition has to be familiar with the safetyinstructions relevant there.
Prior to repair and cleaning works the Type 663 has to be seperated from the mains!Seperated plug and socket devices always have to stay disconnected from the mains!Otherwise danger of explosion because of sparkling, when operating ATEX-certified Types663 in explosion endangered areas!
Defective connection cables immediately have to be replaced! Otherwise danger ofexplosion because of sparkling, when operating ATEX-certified Types 663 in explosionendangered areas!
A defective control has to be changed completely!
Error Cause Activity
No measured value (4-20 mA) No power supply Check power supply and/or connection cable
Connection cable interrupted Replace connection cable
Fuse defective Replace fuse
Wrong connection cable polarity Provide correct connection cable polarity
Control defective Replace Control
Relay does not switch Limit Value adjustement wrong Adjust Limit Value correctly
No power supply Check power supply and/or connection cable
Connection cable interrupted Replace connection cable
Fuse defective Replace fuse
Polarity of connection cable wrong Polarize connection cable correctly
Control defective Replace Control
Measured value wrong Control mounting not friction-locked Mount Control friction-locked
Control mounting at wrong position Mount Control at correct position
EMC-Problems see Cap. 19 Grounding Concepts
HAUBER-Elektronik GmbH
16
19 Grounding Concepts to avoid Ground Loops
Ground loops are among the most frequent problems for measurement setups with sensitive
sensor technology. They arise through unwanted potential differences inside the current circuit
between sensor and evaluation device. As countermeasure we recommend our Standard-
Grounding Concept or, according to application, our Alternative-Grounding Concept.
Alternativ-Grounding Concept
In the Alternative-Grounding Concept the shield of the sensor cable has connection to the
sensor-hausing (dotted circle). The sensor hausing is decoupled from the machine ground via
EMC-adapter (black).
Machine Evaluation Device
+24V / 4...20mA
Sensor Cable
Machine
+24V / 4...20mA
Sensor Cable
Evaluation Device:e.g. Amperemeter,PLC-Control
Standard-Grounding Concept
In the Standard-Grounding Concept the shield of the sensor cable has no connection to the
sensor-hausing (dotted circle). The sensor housing lies on the same potential as the machine
ground.
Maschine ground
Maschine Ground
Sensor-Housing
EMC-Adapter
Please indicate in your order, if you decide to use the .We will then offer to you the according Sensor Cable and EMC-Adapter.
Alternative-Grounding Concept
Evaluation Device
Evaluation DeviceAmperemeter,
:e.g.PLC-Control
Shield
ShieldSensor-Housing