Upload
le-huy
View
214
Download
0
Embed Size (px)
Citation preview
7/30/2019 v44_055
1/7
A CRITICAL EVALUATION OF LINE OVERLAP CORRECTIONS
IN X-RAY SPECTROMETRYLarry E. Creasy
Titanium Metals Corporation
900 Hemlock Road, Morgantown, PA 19543
ABSTRACT
While it is generally recognized that line overlaps in wavelength X-ray spectrometry may be
affected by factors such as variations in mass absorption, conventional line overlap correction
models define the correction function as a constant and assume that it does not vary down to zero
percent concentration of the interfering element. However, the overlap function is not a constant,
even within the scope of a normal calibration set, and it becomes zero before the concentration of
the interfering element reaches zero. By treating the overlap as a variable background and
measuring the background using the Lorentz model, the need to establish an overlap function is
removed and a more accurate calibration can be obtained.
INTRODUCTION
In X-ray spectrometry, the measured intensity (I) of the analyte peak is a composite of several
factors and may be represented by the following equation:
I = F[bkg] + F[conc.] + F[enh] F[abs] + F[ovr]
Where: F[bkg] is a function of background
F[conc.] is a function of concentration
F[enh] is a function of enhancement
F[abs] is a function of absorption, andF[ovr] is a function of line overlap
In the calibration equation, the background is assumed to be a constant or the background is
measured and subtracted from the measured peak intensity to give a net peak intensity. Intensity
due to concentration is a linear function of concentration. The absorption and enhancement
effects are not linear or constant. However, within narrow ranges, they may be approximated by
a single factor for each affecting element using such models as LaChance-Traill1, Rasberry-
Heinrich2, etc. These may be determined by regression or calculated using fundamental alpha
type equations3. In more complex calibrations, these factors may be calculated using
Fundamental Parameter models4. The line overlap function is always assumed to be constant
with a zero concentration/zero interference intercept. These constants are typically determined byone of a number of techniques. The three most common techniques are:
3
1. Measure the intensity at the analyte position on a sample of pure interfering element and ratiothat value to another peak of that element. For example, to determine the constant for the
overlap of Cr-K on the Mn-K peak, measure the intensity at the Mn-K location on a
sample of pure chromium and ratio this value to the intensity of the Cr-K or K peak. Thisrequires the use of very pure chromium.
Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 3
7/30/2019 v44_055
2/7
2. Using the same technique as above, but use a specimen that contains chromium, but nomanganese. It is not always easy to find such a specimen.
3. Determine the constant by regression in the calibration process.Procedures 1 and 2 assume that there are no major absorption edges between the analyte peak
and peak of the overlapping element that is used to make the correction. Even if there were nomajor absorption edges present, significant variations in the mass absorption coefficient could
affect this correction factor.
However, in dealing with the most common line overlaps where the analyte peak lies on the edge
of another overlapping peak, such as the overlap of the Cr-K on the Mn-K, the Ni-K on Cu-
K, or Fe-K on Co-K, etc., it will be shown that both of the assumptions, a zero/zero interceptand constancy, are not true.
In an effort to increase the accuracy of the line overlap correction, it proposed that the overlap be
treated as a variable background and measured for each sample, instead of assigning a single
constant that would be applied to all samples. The model for this measurement is the Lorentzmodel that is available in some software packages for background correction. With properly
selected locations around the analyte peak, the Lorentz equation models the shape of theshoulder of the overlapping peak very well. The equation for this model is as follows 5:
( )i
i
bxaa
aax
aI 54
3
2
2
1++
++
=
Where:Ib = Intensity of the background,xi = 2-Theta angle, anda1.a5 are coefficients.
Figure 1. Background Fit for Overlap Using Lorentz Equation
Intensity
62 62.4 62.8 63.2 63.6 64
Degrees 2-Theta
Mn-K
Cr-K
Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 3
7/30/2019 v44_055
3/7
An example of an overlap correction using the Lorentz equation to model the overlap
interference under the analyte is shown in Figure 1. A minimum of five points are selected andthe curve is generated, basically treating the overlap as a curved background. These five points
are measured for every specimen and the overlap is subtracted, resulting in a net peak intensity.
This study will deal only with the situations where the overlap is such that the analyte peak lies
on the tail of the overlapping element, such as the overlap of Cr-K on the Mn-K peak. It will
not address the situation where there is a complete overlap, such as the Mo-L1 on the P-K peak.
EXPERIMENTAL
Over 130 specimens were measured to determine overlap intensities. The measurements were
made on a Rigaku RIX-2000 scanning x-ray spectrometer. The matrices included aluminum,
iron, nickel, cobalt, copper (brasses), zirconium and titanium. The Mn-K, Cu-K, and Co-Klines were measured. The overlap intensities at the analyte locations were determined by direct
measurement of samples that had no analyte in them or by using a Lorentz fit. The Lorentz fit
was made using the standard software package for the instrument.
In addition, a calibration scheme was set up to compare the conventional treatment of line
overlap with a model that used the Lorentz equation to measure the overlap. In this test, about 25iron-based reference materials were used, generally of the type that would be considered
stainless steels. The accuracy of the Lorentz measured overlaps were compared to regression fits
using both the K and the K lines of the overlapping element.
RESULTS AND DISCUSSION
Below are example scans of the region surrounding the K- lines of Co and Cu withconcentrations of Fe and Ni at low levels.
Figure 2. Zero Overlap from Interfering Elements
In Figure 2, the overlap is shown to reach zero before the concentration or the intensity of the
overlapping element reaches zero. The intensity level at which the overlap reaches zero will
depend on the resolution of the spectrometer and the difference between the location of the twopeaks. The concentration level at which this occurs will also depend on the matrix, since mass
0
1
2
3
4
5
51.0 51.5 52.0 52.5 53.0
0
1
2
3
43.0 43.5 44.0 44.5 45.0 45.5 46.0
Intensity
(kcps)
Fe-K Ni-K
Cu-K Co-K
2-Theta 2-Theta
a b
Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 3
7/30/2019 v44_055
4/7
absorption affects the sensitivity of the peaks. In Figure 2a, the material is Ti with about 4.4%
Fe. In Figure 2b, the material is cast iron with about 2.3% Ni.
Having shown that the intercept of the overlap function is not zero-zero, the next question is Is
the overlap function a constant?. Over 130 specimens were measured as previously described.
The overlap ratios are shown plotted and described below.
Intensity of Cr-K (kcps) Intensity of Cr-K (kcps)
Figure 3. Overlap Ratio - Cr-K on the Mn-K peak
In the plot of all the data, Figure 3a, the ratios are fairly consistent as the intensity of the Cr-Kpeak decreases until the intensity reaches about 100 kcps, at which point, the ratio increases
significantly. Most of this rise is because the overlap values are at or near background and are
fairly constant, and these numbers are divided by increasingly smaller Cr-K peak intensities.When the plot of ratio values less than 0.01 are examined (Figure 3b where more detail is seen),
there is much scatter. Even in the region where the data is relatively flat, there is a range of ratios
from less than 0.001 to 0.004. If pure Cr were used to determine the overlap ratio, the ratiowould be found to be 0.0012, a value at the extreme low end.
Intensity of Fe-K (kcps) Intensity of Fe-K (kcps)
Figure 4. Overlap Ratio - Fe-K on the Co-K - Fe intensity less than 20 kcps (a) andratios less than 0.02 (b)
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0 200 400 600 800 1000 1200
OverlapRatio
a
0.000
0.002
0.004
0.006
0.008
0.010
0 200 400 600 800 1000 1200
0.0
0.4
0.8
1.2
1.6
2.0
0 4 8 12 16 20
0.000
0.004
0.008
0.012
0.016
0.020
0 50 100 150 200 250
OverlapRatio
b
ab
Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 3
7/30/2019 v44_055
5/7
The patterns for Fe-Co are similar to those for Cr-Mn. The range of the ratio values for Fe
intensities greater than 30kcps is 0.006 to 0.014. The values increase dramatically at the lowintensity levels. Again, the value for pure Fe is 0.0064, near the minimum.
Intensity of Ni-K (kcps) Intensity of Ni-K (kcps)
Figure 5. Overlap Ratio - Ni-K on Cu-K - Ni intensity less than 20 kcps (a) and ratiosless than 0.02 (b)
Exhibiting the same pattern as previously observed, the range of ratio values for Ni intensities
greater than 50 kcps is 0.002 to 0.009 and increases at the low intensity levels. In this case, thevalue for pure Ni is near the average at 0.0049.
It is apparent that the overlap ratio is not a precise constant and the function does not have azero-zero intercept. Having shown that the single factor function for line overlap is inadequate,
the question arises What is the practical implication? To determine this, a calibration was set
up for Mn, Co, and Cu in stainless steels. The Fe concentrations varied from 51 to 88%, the Cr
from 2 to 22% and the Ni from 0.2 to 28%. The three elements in question were calibrated using
regression to determine an overlap factor for the K line and the K line of the overlappingelements and using the Lorentz function to subtract the overlap. The sigma values for each fit
were examined. The data for each calibration is shown in the Appendix. A summary of the sigmavalues is shown in Table 1.
Table 1. Summary of Calibration Sigma Values
Mn Co Cu
K K Lorentz K K Lorentz K K Lorentz
Sigma 0.036 0.041 0.021 0.024 0.032 0.011 0.012 0.014 0.012
Using the Lorentz model, there was significant improvement over the conventional regression
fits for Co and Mn. The sigma value of the calibration using the Lorentz model to subtract the
overlap for Co improved to 0.011, as compared to 0.024 using the regression fit against the Fe-
0.0
0.5
1.0
1.5
2.0
2.5
0 4 8 12 16 20
0.000
0.004
0.008
0.012
0.016
0.020
0 100 200 300 400
OverlapRatio
a b
Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 3
7/30/2019 v44_055
6/7
K and 0.032 against the Fe-K. The improvements for Mn were 0.021 for the Lorentz model ascompared to 0.036 and 0.041 for the regression fits. There was no significant improvement in the
Cu calibrations, probably because the separation of Cu- K from the Ni- K is greater than forthe other two pairs. Therefore, the overlap interference is less.
CONCLUSION
It has been shown that the line overlap for the situation where the analyte peak lies on the tail of
another element is not best described by single factor function. The overlap does become zerobefore the concentration of the overlapping element becomes zero. There is considerable scatter
in the overlap ratios and the function is non-linear, especially near the low end. It was not the
intent to resolve the line overlap function, that is, to determine the best model to define the lineoverlap, but rather, to show that there is a better way to handle the overlap in these situations.
The Lorentz model, commonly used for curved background correction, is a very effective tool to
subtract this overlap and to calibrate using a net analyte intensity. This results in a more accurate
calibration, in some cases by a factor of more than two.
REFERENCES
1. LaChance, G.R. and Traill, R.J., A Practical Solution to the Matrix Problem in X-ray
Analysis, Can. Spectry., Vol. 11, 1966, p43.
2.Rasberry, S.D. and Heinrich, K.F.J., Calibration for Interelement Effects in X-RayFluorescence Analysis,Analytical Chemistry, Vol. 46, 1974, p81.
3.Standard Guide for Correction of Interelement Effects in X-ray Spectrometric Analysis
(E1361),Annual Book of ASTM Standards Vol. 0306.
4.A number of these are described in Jenkins, R, et al. Quantitative Spectrometry, SecondEdition, Marcel Dekker, Inc., NY 1995, Ch. 10.
5. Private discussion with Katsu Toda, Rigaku/USA
Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 3
7/30/2019 v44_055
7/7
APPENDIX CALIBRATION DATA FOR STAINLESS STEELS
Calibration Calculations for Mn Calibration Calculations for CuK
Correction
KCorrection
Lorentz KCorrection
KCorrection
Lorentz
Sample Cert.
Value
Calc Diff. Calc Diff. Calc Diff. Cert.
Value
Calc. Diff. Calc. Diff. Calc. Diff.
160 1.65 1.586 -0.064 1.647 -0.003 1.633 -0.017 0.172 0.168 -0.004 0.177 0.005 0.169 -0.003
1151 2.17 2.167 -0.003 2.141 -0.030 2.179 0.009 0.25 0.253 0.003 0.245 -0.005 0.250 0.000
1152 1.19 1.208 0.018 1.212 0.022 1.160 -0.030 0.50 0.516 0.016 0.504 0.004 0.490 -0.010
1154 1.74 1.731 -0.009 1.715 -0.025 1.748 0.008 0.56 0.560 0.000 0.544 -0.016 0.544 -0.016
1155 1.63 1.625 -0.005 1.607 -0.023 1.651 0.021 0.169 0.170 0.001 0.173 0.004 0.168 -0.001
1171 1.80 1.788 -0.012 1.875 0.075 1.795 -0.005 0.121 0.115 -0.006 0.121 0.000 0.121 0.000
1172 1.76 1.776 0.016 1.759 -0.001 1.764 0.004 0.105 0.111 0.006 0.110 0.005 0.114 0.009
1173 0.174 0.179 0.005 0.145 -0.029 0.175 0.001 0.204 0.204 0.000 0.199 -0.005 0.204 0.000
1184 1.04 0.997 -0.043 1.023 -0.017 1.022 -0.018 N.C. ------- ------- ------- ------- ------- -------
1185 1.22 1.243 0.023 1.224 0.004 1.236 0.016 0.067 0.063 -0.004 0.065 -0.002 0.065 -0.002
1193 0.65 0.615 -0.035 0.630 -0.020 0.653 0.003 0.103 0.139 0.036 0.153 0.050 0.141 0.038
1194 0.667 0.665 -0.002 0.652 -0.015 0.656 -0.011 0.047 0.061 0.014 0.064 0.017 0.052 0.0051223 1.08 1.170 0.090 1.153 0.073 1.140 0.060 0.081 0.067 -0.014 0.067 -0.014 0.083 0.002
1230 0.64 0.624 -0.016 0.639 -0.001 0.635 -0.005 0.14 0.138 -0.002 0.147 0.007 0.134 -0.006
1287 1.66 1.712 0.052 1.675 0.015 1.663 0.003 0.58 0.585 0.005 0.575 -0.005 0.571 -0.009
B474 1.70 1.660 -0.041 1.620 -0.080 1.702 0.002 0.35 0.355 0.005 0.340 -0.010 0.337 -0.013
C1154 1.42 1.450 0.030 1.499 0.079 1.433 0.013 0.40 0.394 -0.006 0.399 -0.002 0.388 -0.012
D845 0.77 0.782 0.012 0.762 -0.008 0.779 0.009 0.065 0.055 -0.010 0.056 -0.009 0.070 0.005
D846 0.53 0.511 -0.019 0.529 -0.001 0.505 -0.026 0.19 0.191 0.001 0.198 0.008 0.195 0.005
D848 2.13 2.083 -0.047 2.061 -0.069 2.096 -0.034 0.16 0.154 -0.006 0.148 -0.012 0.177 0.017
D849 1.63 1.672 0.042 1.664 0.034 1.650 0.020 0.21 0.210 0.000 0.203 -0.007 0.213 0.003
D850 N.C. ------- ------- ------- ------- ------- ------- 0.36 0.336 -0.024 0.353 -0.007 0.359 -0.001
Sigma 0.036 0.041 0.021 0.012 0.014 0.012
N.C. = Not Certified
Calibration Calculations for CoK
Correction
KCorrection
Lorentz
Sample Cert.
Value
Calc. Diff. Calc. Diff. Calc. Diff.
B474 0.019 0.017 -0.002 0.020 0.001 0.002 -0.017
1173 0.064 0.085 0.021 0.072 0.008 0.065 0.001
1171 0.10 0.084 -0.016 0.092 -0.008 0.088 -0.012
160 0.101 0.092 -0.009 0.098 -0.003 0.097 -0.004
1155 0.101 0.096 -0.005 0.098 -0.003 0.099 -0.002
1154 0.12 0.102 -0.018 0.100 -0.020 0.113 -0.0071172 0.12 0.111 -0.009 0.108 -0.012 0.117 -0.003
1230 0.15 0.147 -0.003 0.158 0.008 0.134 -0.016
1287 0.31 0.274 -0.036 0.227 -0.083 0.295 -0.015
C1154 0.38 0.326 -0.054 0.329 -0.051 0.382 0.002
1194 2.77 2.780 0.010 2.779 0.009 2.778 0.008
Sigma 0.024 0.032 0.011
Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 3