Using Hydrophobic Air Filters as Bidirectional Protection between an Integrity Tester and a Filter Housing

turning science into solutions

Application Note

Page 2: Using Hydrophobic Air Filters as Bidirectional Protection ...microsite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · Using Hydrophobic Air Filters as Bidirectional

BackgroundIntegrity testing of sterilizing-grade filters is necessary in order to unambiguously rule out potential damage to these sterile barriers during the production of biophar-maceuticals. The principle of filter integrity testing is based on using an integrity test unit to pressurize the filter cartridge hous-ing and to measure the pressure drop that occurs when gas molecules diffuse through a wetted filter membrane. The pneumatic connection required for integrity testing between the test unit and the filter to be tested entails a certain risk that liquid, aerosols or even microorganisms can be drawn back into the test unit. Vice versa, sterile test gas is not used during integrity testing; i.e., the interior of the test unit is not sterile. This, in turn, poses the risk that particles or microbes might be carried into the filter cartridge housing during the test procedure. Although the membrane of the filter under test ultimately constitutes a sterile barrier, it is still advisable to work meticulously in order to prevent contami-nation from occurring on the upstream side.

This Application Note describes the use of hydrophobic air filters (Sartofluor®, Midisart®) as reliable sterilizing-grade filters and liquid barriers between the integrity tester and the filter to be tested. Theoretical considerations and test data show that even a small, Midisart-type air filter installed in the test line will not affect the measurement results in any way.

Dr. Jens MeyerProduct ManagerSartorius Stedim Biotech

Why use a hydrophobic air filter between the integrity tester and the filter cartridge housing? It is a fact that integrity testing of steriliz-ing-grade filters is not performed under sterile conditions, ultimately due to the pneumatic connection between the non-sterile integrity test unit and the filter car-tridge under test. The integrity test itself is performed on the unsterile, or upstream, side. As a result, the membrane to be test-ed, if intact, will retain its function as a sterile barrier for protection of a down-stream product. Nevertheless, in many cases, it might be desirable to perform integrity tests under virtually sterile condi-tions. As the internal pneumatic compo-nents of the integrity testers used cannot be sterilized, installation of sterilizing-grade air filters (pore size of 0.2 µm) in the test line is ideal for ensuring sterile conditions. These filters reliably retain any potential particulate and microbial con-taminants, stopping them from reaching the filter membrane of the cartridge to be tested. The air filters presented in this report have been validated as sterilizing-grade filters with respect to their microbial retention in compliance with the regulations generally valid throughout the industry.

Page 3: Using Hydrophobic Air Filters as Bidirectional Protection ...microsite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · Using Hydrophobic Air Filters as Bidirectional

Air filters feature hydrophobic filter membranes. This hydropho-bicity is required to prevent blockage of the membrane caused by moisture present in air. The polytetrafluoroethylene (PTFE) material used in Midisart® and Sartofluor® filters is extremely hydrophobic, which not only provides protection against microbial contamina-tion, but also simultaneously prevents the passage of water or aqueous solutions. Installed as bidirectional protection in the test line, these filters at the same time prevent liquid from the filter cartridge housing from flowing back into the integrity tester used. Only at the so-called water penetration pressure can water pass through a hydrophobic PTFE membrane. For the Midisart® mem-brane, this pressure is above 4.5 bar (65.3 psi), however, so the effectiveness of this filter as a water barrier is guaranteed up to this pressure.

Ultimately, it must be ensured that such a filter installed in a test line will not have any adverse effect on the integrity test performed. This is to be explained in the following:

Is the filter area of Midisart® sufficient?The Midisart-type air filter has an effective filter area of only 20 cm2 (~3.1 square inches). This raises the question as to whether this small filter area is sufficient and will not unacceptably limit the air flow rate during the integrity test. If so, this would then automatically have an unsuitable impact on the results of an integrity test performed.

During a filter integrity test, the highest air flow rate must be measured during the phases of pressurization and venting. In these phases, the filter cartridge housing under test is filled up to the programmed test pressure and, at the end of the test, vented so that the pressure inside returns to the normal atmospheric level. A 10’’ filter cartridge housing has an upstream volume of approxi-mately one liter. At a differential pressure of 100 mbar (~1.5 psi), Midisart® 2000 delivers an air flow rate of around five liters per minute. This clearly shows that a Midisart® filter can be considered to have a negligible effect on the flow rate. When large housings with correspondingly higher volumes are to be integrity tested, it is naturally preferable to complete the filling process as fast as possible. In these cases, the inner diameter of the tubing used as the test line, the quick-connect couplings and the valves ultimately restrict the inlet flow rate so that Midisart® 2000 does not have any limiting effect either.

An integrity test is comprised of the phases of pressurization, stabilization, testing and venting. The pressurization and venting phases are necessary, but initially do not have any influence on the test results. For the test phase, however, it must be absolutely ensured that the Midisart® filter used does not have any impact on the test results. In this phase, only drops in pressure are measured, which are caused by diffusion of the test gas through a wetted membrane. In this case, tests can be performed to verify that the flow rates (diffusion rates) are very low and the Midisart® filter connected between the test unit and the filter housing has no restrictive effect on these rates. For instance, a 10’’ Sartopore® 2

cartridge (0.2 µm) has a diffusion limit of merely 18 ml/min. This clearly indicates that the air flow rate of Midisart® filters does not limit flow in any case during the test phase, not even in multi-cartridge housings.

The purpose of the following test procedure was to verify our theoretical assumptions outlined above:Sartopore® 2 sterilizing-grade filters (0.2 µm) with and without Midisart® installed in the test line were tested. The small Midisart® filter was used in the “worst case” test scenario. As larger Sartofluor® air filter capsules have considerably more filter area than does Midisart®, these can be considered to have a negligible impact on flow rate, if it can be demonstrated that Midisart® itself does not have any limiting effect. Following successful wetting of the Sartopore® 2 capsules to be tested, they were integrity tested using the Sartocheck® 4 plus tester set to standard test parameters.

Figure 1: The two different test setups for the investigation. For the standard test the Sartocheck® test tubing was directly connected to the MidiCaps. The protective Midisart® 2000 filter was positioned between the test tubing and the MidiCaps filter in order to investigate its potential impact on the test result.

Three diffusion tests and three bubble point tests were performed on each Sartopore® 2 MidiCaps, using different Midisart® filters in each case. The test results are given in Table 1:

Test No. Diffusion Test Results [ml/min]

Bubble Point Test Results[mbar]

Without Midisart®

With Midisart®

Without Midisart®

With Midisart®

1 3.6 3.7 3928 3934

2 3.4 3.5 3977 3933

3 3.4 3.4 3978 3983

Mean value 3.47 ± 0.12 3.53 ± 0.15 3961 ± 28 3950 ± 28

Table 1: Test results of filter integrity tests of a Sartopore® 2 MidiCaps filter (0.2 µm, size 9). The test with a Midisart® filter in the test line has no significant impact on the test result.

Standard Test Setup

Test Setup with Protective Filter

Page 4: Using Hydrophobic Air Filters as Bidirectional Protection ...microsite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · Using Hydrophobic Air Filters as Bidirectional

These results show that there are no significant differences in the outcomes of the tests performed with and without a Midisart® 2000 filter installed in the test line. Nearly identical results were obtained both for the diffusion tests and for the bubble point tests. Therefore, these results prove that the use of a Midisart® filter in the test line is safe and reliable.

Figure 2: Bubble Point Tests without (black curve) and with Midisart® filter (red curve) in the test line. Each curve represents the average curve of three bubble point tests. Both curves with and without protective filter are almost congruent, i.e. no effect of the additional Midisart® 2000 filter in the testline.

Generally, it must be noted that the flow rates given apply only to dry filters. Should a film of liquid partially or completely cover the hydrophobic PTFE membrane of the Midisart® filter, this will reduce the flow rate accordingly and may even lead to complete blockage of the membrane. In this respect, users need to ensure at all times that any liquid that collects inside the integrity test tubing will not affect Midisart®.

Direction in which the air filter is installedAn additional question to answer is in which direction the air filter needs to be installed between the integrity test unit and the filter cartridge housing. Basically, air filters can be used in either direction. However, regarding their design, the filters are not sym-metrical. As a result, the maximum differential pressure for an air filter installed in the direction of filtration will differ from that for a filter installed in the direction opposite to filtration, for instance. Therefore, the direction in which a filter is installed would play a critical role only in situations in which this maximum pressure tolerance would be exceeded during integrity testing. The PTFE membrane used in Midisart® is highly permeable and features high flow rate performance, among other characteristics. Therefore, no substantial pressure differences will occur across the filter mem-brane during the pressurization and venting phases. In all integrity test phases, there are thus no limiting differences in pressure. In this respect, the direction in which the air filter installed would not matter.

Tests of bag systems: necessity of sterilization by gamma irradiationAs a result of the growing trend toward single-use products employed in the manufacture of biopharmaceuticals, pre-assem-bled filter capsules are increasingly being installed on single-use bag systems. To avoid compromising the sterility of the filter capsules connected to these systems even during integrity testing, a Midisart® BV filter (gamma-sterilized variant of Midisart® air filters) must be used, which has been presterilized by gamma irradiation together with the entire bag system. The tubing of the Sartocheck® unit is then connected by the appropriate coupling to the air filter, and the downstream filter capsule can then be tested.

Even in this test equipment setup, Midisart® BV will not affect the integrity test. Rather, it ensures that the required conditions for sterility are reliably met during the test. However, it must be con-sidered that the properties of plastic polymer materials are altered by the gamma irradiation procedure used. This will result in lower housing burst pressure values, which may be critical for integrity tests under specific circumstances. While a maximum pressure of 3 bar (~44 psi) is permitted for Midisart® 2000, only 1.5 bar (~22 psi) maximum is allowed for Midisart® BV after gamma irradiation. Depending on the type of filter to be integrity tested, the test pressure applied may exceed the maximum permissible pressure for the air filter. For example, Sartopore® 2 Gamma MidiCaps with a pore size of 0.1 µm must be tested at a pressure of 4 bar (58 psi) using the diffusion method. Taken at face value, the maximum burst pressure specification for Midisart® BV would preclude its use.

As mentioned above, the housing burst pressure limits the maxi-mum allowable operating pressure of an air filter. Yet this limita-tion can be overcome by using a suitable mechanical stabilizer for Midisart® BV, which restores the air filter’s resistance. As a result, test pressures higher than the maximum burst pressure can be used without compromising the integrity of Midisart® BV. The following test procedure is designed to demonstrate this:

Figure 3: Mechanical stabilizer for Midisart® filters. The clamp stabilizer can be easily mounted and reliably protects the irradadiated Midisart® filter from bursting during integrity testing of gamma capsules.

Page 5: Using Hydrophobic Air Filters as Bidirectional Protection ...microsite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · Using Hydrophobic Air Filters as Bidirectional

The housing burst pressure of 10 Midisart® BV filters in unsterilized condition and of 10 Midisart® BV filters in gamma irradiated condition (50 kGy) was tested with and without a mechanical stabilizer (max. test pressure 10 bar = 145 psi).

Midisart® BVHousing Burst Pressure [bar]

Without Tri-Clamp With Tri-Clamp

Unsterilized Irradiated with 50 kGy

Unsterilized Irradiated with 50 kGy

7.57±2.7 3.41±0.7 >10 >10

Table 2: Effect of the mechanical stabilizer on the housing burst pressure of Midisart® BV. The burst pressure can be increased to more than 10 bar even after gamma irradiation using this stabilizer. Therefore, this device can be used to enable the usage of Midisart® BV in the test line for testing gamma capsules.

The results presented here show that the use of a ring-shaped stabilizer, a Tri-Clamp, on Midisart® BV enables burst pressure values of more than 10 bar to be measured. This means that Midisart® BV mounted inside a Tri-Clamp stabilizer is suitable as a sterile barrier for integrity testing of gamma-irradiated filter capsules. Even the high test pressure required for a 0.1 µm capsule is no longer a problem for this air filter. As a result, the limits specified for the maximum operating pressure of Midisart® BV do not apply.

SummaryThe tests described here show that Midisart® 2000 and Midisart® BV are suitable for use as protective filters between the test unit and the filter during integrity testing. The results of the filter integrity test are unaffected by the additional Midisart® filter. Furthermore, it could be demonstrated that when clamped inside a mechanical stabilizer, even gamma-irradiated air filters of the Midisart® BV type are appropriate for use as protective sterilizing-grade air filters during integrity testing. Their mechanical stability reduced by gamma irradiation is thus restored by the Tri-Clamp attached in place. The resulting maximum burst pressure values of these air filters are then much higher any of the test pressures relevant for the filters to be integrity tested.

The currently valid regulations do not require that such a pro- tective filter be installed in the test line. In this regard, it is up to the user to decide whether the benefits of preventing upstream contamination will outweigh the risks and cost of filters that may otherwise become contaminated by integrity testing.

Page 6: Using Hydrophobic Air Filters as Bidirectional Protection ...microsite.sartorius.com/fileadmin/newsletter/germany_biotechnology/... · Using Hydrophobic Air Filters as Bidirectional

Sales and Service ContactsFor further contacts, visit www.sartorius-stedim.com

EuropeGermanySartorius Stedim Biotech GmbHAugust-Spindler-Strasse 1137079 Goettingen

Phone +49.551.308.0Fax +49.551.308.3289

www.sartorius-stedim.com

Sartorius Stedim Systems GmbHSchwarzenberger Weg 73–7934212 Melsungen

Phone +49.5661.71.3400Fax +49.5661.71.3702

www.sartorius-stedim.com

FranceSartorius Stedim Biotech S.A.ZI Les PaludsAvenue de Jouques – BP 105113781 Aubagne Cedex

Phone +33.442.845600Fax +33.442.845619

Sartorius Stedim France SASZI Les PaludsAvenue de Jouques – CS 7105813781 Aubagne Cedex

Phone +33.442.845600Fax +33.442.846545

AustriaSartorius Stedim Austria GmbHFranzosengraben 12A-1030 Vienna

Phone +43.1.7965763.18Fax +43.1.796576344

BelgiumSartorius Stedim Belgium N.V.Leuvensesteenweg, 248/B1800 Vilvoorde

Phone +32.2.756.06.80Fax +32.2.756.06.81

DenmarkSartorius Stedim Nordic A/SHoerskaetten 6D, 1.DK-2630 Taastrup

Phone +45.7023.4400Fax +45.4630.4030

HungarySartorius Stedim Hungária KftKagyló u. 52092 Budakeszi

Phone +36.23.457.227Fax +36.23.457.147

ItalySartorius Stedim Italy S.p.A.Via dell’Antella, 76/A50012 Antella-Bagno a Ripoli (FI)

Phone +39.055.63.40.41Fax +39.055.63.40.526

NetherlandsSartorius Stedim Netherlands B.V.Edisonbaan 243439 MN Nieuwegein

Phone +31.30.6025080Fax +31.30.6025099

PolandSartorius Stedim Poland Sp. z o.o.ul. Wrzesinska 7062-025 Kostrzyn

Phone +48.61.647.38.40Fax +48.61.879.25.04

SpainSartorius Stedim Spain SAC/Isabel Colbrand 10, Oficina 70Polígono Industrial de Fuencarral28050 Madrid

Phone +34.90.2110935Fax +34.91.3589623

SwitzerlandSartorius Stedim Switzerland AGRingstr. 24 a8317 Tagelswangen

Phone +41.52.354.36.36Fax +41.52.354.36.46

U.K.Sartorius Stedim UK LimitedLongmead Business ParkBlenheim Road, EpsomSurrey KT19 9 QQ

Phone +44.1372.737159Fax +44.1372.726171

AmericaUSASartorius Stedim North America Inc.5 Orville DriveBohemia, NY 11716

Toll-Free +1.800.368.7178Fax +1.631.254.4253

Sartorius Stedim SUS Inc.1910 Mark CourtConcord, CA 94520

Phone +1.925.689.6650Toll Free +1.800.914.6644Fax +1.925.689.6988

Sartorius Stedim Systems Inc. 201 South Ingram Mill RoadSpringfield, MO 65802

Phone +1.417.873.9636Fax +1.417.873.9275

ArgentinaSartorius Argentina S.A. Int. A. Avalos 4251 B1605ECS Munro Buenos Aires

Phone +54.11.4721.0505 Fax +54.11.4762.2333

BrazilSartorius do Brasil LtdaAv. Dom Pedro I, 241Bairro Vila Pires Santo André São PauloCep 09110-001

Phone +55.11.4451.6226Fax +55.11.4451.4369

MexicoSartorius de México S.A. de C.V.Circuito Circunvalación Poniente No. 149Ciudad Satélite53100 Naucalpan, Estado de México

Phone +52.5555.62.1102Fax +52.5555.62.2942

Asia|PacificAustraliaSartorius Stedim Australia Pty. Ltd.Unit 5, 7-11 Rodeo DriveDandenong South Vic 3175

Phone +61.3.8762.1800Fax +61.3.8762.1828

ChinaSartorius Stedim BeijingRepresentative OfficeNo. 33, Yu’an Road,Airport Industrial Zone B, Shunyi DistrictBeijing 101300

Phone +86.10.80426516Fax +86.10.80426580

Sartorius Stedim ShanghaiRepresantative OfficeRoom 618, Tower 1, German Centre,Shanghai, PRC., 201203

Phone +86.21.28986393Fax +86.21.28986392.11

Sartorius Stedim Guangzhou OfficeRoom 704, Broadway Plaza,No. 233–234 Dong Feng West RoadGuangzhou 510180

Phone +86.20.8351.7921Fax +86.20.8351.7931

IndiaSartorius Stedim India Pvt. Ltd.#69/2-69/3, JakkasandraKunigal Road, Nelamangala TqBangalore – 562 123

Phone +91.80.4350.5361Fax +91.80.4350.5253

JapanSartorius Stedim Japan K.K.KY Building, 8–11 Kita Shinagawa 1-chomeShinagawa-kuTokyo 140-0001

Phone +81.3.3740.5407Fax +81.3.3740.5406

MalaysiaSartorius Stedim Malaysia Sdn. Bhd.Lot L3-E-3B, Enterprise 4Technology Park MalaysiaBukit Jalil57000 Kuala Lumpur

Phone +60.3.8996.0622Fax +60.3.8996.0755

SingaporeSartorius Stedim Singapore Pte. Ltd.1 Science Park Road, The Capricorn, #05-08A,Singapore Science Park 2Singapore 117528

Phone +65.6872.3966Fax +65.6778.2494 Sp

ecif

icat

ions

sub

ject

to c

hang

e w

itho

ut n

otic

e. P

rinte

d in

Ger

man

y on

pap

er th

at h

as b

een

blea

ched

wit

hout

any

use

of c

hlor

ine.

| W

Publ

icat

ion

No.

: XX-

0000

-e07

051

· Ord

er N

o.: 0

0000

-000

-00

· Ver

. MM

| YY

YY

www.sartorius-stedim.com

Spec

ific

atio

ns s

ubje

ct to

cha

nge

wit

hout

not

ice.

Prin

ted

in G

erm

any

on p

aper

that

has

bee

n bl

each

ed w

itho

ut a

ny u

se o

f chl

orin

e. |

W

Publ

icat

ion

No.

: SPI

4011

-e11

121

· Ord

er N

o.: 8

5032

-541

-61

· Ver

. 12

| 201

1