® EC REP ® MIS Implants Technologies Ltd. P.O.Box 7 , Bar Lev Industrial Park 2015600, ISRAEL Website: www.mis-implants.com User Manual Complete Prosthetic Kit - CPK (Conical Connection) MP-UIC29 Rev.1 Apr. 2014 © MIS Corporation. All rights reserved. The plastic copings for using CPK abutments are fitted a snap mechanism which makes use of the elasticity of the plastic material. In order for the metal framework to fit on the analog/abutment, the snap mechanism should be removed. Use a MT-CS450 reamer to remove excess material prior to seating the framework on analogs or abutments. 7 Insert the abutment analog into the plastic cap (in the impression tray). * Verify that the analog fits properly. Laboratory procedure. Place the wax-up anti-rotation plastic cap on the abutment analog. * For single tooth restorations, ALWAYS use the red wax-up cap with the anti-rotation mechanism. For bridges, it is easier to use the white wax-up caps, which do not include an anti-rotation mechanism. * Cut the plastic cap according to analog height. 6 5 Take the impression. * Recommended impression material: silicone over 20-shore hardness. Temporary stage Place the temporary healing cap on top of the abutment, to allow soft tissue healing and prevent irritation by the abutment. * Use temporary cement. 3 4 Manufacturer Batch Code Catalog Number Single Use Attention, see instructions for use MIS Implants Technologies GmbH Simeonscarre 2, 32423 Minden, Germany Tel: +49 571-972-7620 Email: [email protected] Key to codes used: Procedures recommended by MIS cannot replace the judgment and professional experience of the surgeon. The CPK is not intended for use with crowns cast directly onto the abutment analog, such as a CAD/CAM or Captek™ process. When using such a process, work directly off the CPK abutment. Instrument maintenance. Sterilization: The device must be sterilized before use by autoclave, at a temperature of 134°C (273°F) at a pressure of 315 Kpa during 6 minutes. Do not exceed 134ºC during sterilization. * All parts of the Complete Prosthetic Kits are supplied non- sterile. The plastic components are for single use only.

User Manual Complete Prosthetic Kit - CPK - Mis ImplantsMP-UIC29)_Rev1… · EC REP ® MIS Implants Technologies Ltd. P.O.Box 7, Bar Lev Industrial Park 2015600, ISRAEL Website: User

Embed Size (px)

Citation preview

Page 1: User Manual Complete Prosthetic Kit - CPK - Mis ImplantsMP-UIC29)_Rev1… · EC REP ® MIS Implants Technologies Ltd. P.O.Box 7, Bar Lev Industrial Park 2015600, ISRAEL Website: User

®

EC REP

®

MIS Implants Technologies Ltd.P.O.Box 7, Bar Lev Industrial Park 2015600, ISRAELWebsite: www.mis-implants.com

User ManualComplete Prosthetic Kit - CPK(Conical Connection)

MP-UIC29 Rev.1 Apr. 2014© MIS Corporation. All rights reserved.

The plastic copings for using CPK abutments are fitted a snap mechanism which makes use of the elasticity of the plastic material.

In order for the metal framework to fit on the analog/abutment, the snap mechanism should be removed.

Use a MT-CS450 reamer to remove excess material prior to seating the framework on analogs or abutments.

7

Insert the abutment analog into the plastic cap (in the impression tray).

* Verify that the analog fits properly.

Laboratory procedure.

Place the wax-up anti-rotation plastic cap on the abutment analog.

* For single tooth restorations, ALWAYS use the red wax-up cap with the anti-rotation mechanism.

For bridges, it is easier to use the white wax-up caps, which do not include an anti-rotation mechanism.

* Cut the plastic cap according to analog height.

65

Take the impression.

* Recommended impression material: silicone over

20-shore hardness.

Temporary stage

Place the temporary healing cap on top of the abutment, to allow soft tissue healing and prevent irritation by the abutment.

* Use temporary cement.

3 4

Manufacturer

Batch Code

Catalog Number Single Use

Attention, see instructions for use

MIS Implants Technologies GmbHSimeonscarre 2, 32423 Minden, GermanyTel: +49 571-972-7620 Email: [email protected]

Key to codes used:

Procedures recommended by MIS cannot replace the judgment and professional experience of the surgeon.

The CPK is not intended for use with crowns cast directly onto the abutment analog, such as a CAD/CAM or Captek™ process. When using such a process, work directly off the CPK abutment.

Instrument maintenance.

Sterilization: The device must be sterilized before use by autoclave, at a temperature of 134°C (273°F) at a pressure of ≈315 Kpa during 6 minutes. Do not exceed 134ºC during sterilization.

* All parts of the Complete Prosthetic Kits are supplied non-sterile. The plastic components are for single use only.

Page 2: User Manual Complete Prosthetic Kit - CPK - Mis ImplantsMP-UIC29)_Rev1… · EC REP ® MIS Implants Technologies Ltd. P.O.Box 7, Bar Lev Industrial Park 2015600, ISRAEL Website: User

The Complete Prosthetic Kit (CPK) is a comprehensive set designed for cemented type restorations on single or multiple parallel implants.

■ Compatible with conventional crown and bridge procedures. ■ A selection of lengths, diameters and gingival heights. ■ Can be used immediately after implant placement.■ Offers improved tissue healing.

The CPK includes all required components for the fabrication of cement retained restorations: an abutment, impression cap, comfort cap, casting sleeves and an analog.

A complete solution

The kits are available for all MIS implant platforms: Narrow, Standard and Wide, in four vertical heights and with two crown heights.

Versatility

PrecisionThe CPK ensures simple and accurate impressions and casting procedures.

Kit contents.

Burnout plastic cap

Procedure Steps.

Abutment assembly

Assemble the CPK abutment on the implant.

1

* Recommended tightening moment: 25-30 Ncm.

Burnout anti-rotation plastic cap

Mount transfer plastic cap

Plastic healing cap

- Crown height for Narrow platform: 4, 6mm - Crown height for Standard platform: 4, 6, 8mm- Crown height for Wide platform: 6, 8mm

Anatomic transgingival abutment with screw- Gingival height: 1, 2, 3, 4mm- Crown height for Narrow platform: 4, 6mm - Crown height for Standard platform: 4, 6, 8mm- Crown height for wide platform: 6, 8mm

- Crown height for Narrow platform: 4, 6mm - Crown height for Standard platform: 4, 6, 8mm- Crown height for Wide platform: 6, 8mm

Abutment analog

Advantages at a glance.

Impression taking

Snap the impression coping plastic cap on to the CPK abutment.

* The notch on top of the plastic transfer indicates the location of the flat area

(anti-rotation).

2.

Compatible with conventional crown and bridge procedures.