Upload
others
View
0
Download
0
Embed Size (px)
Citation preview
User Manual
biobas 10
For in-vitro diagnostic use only!
User Manual biobas 10
Revision History 2
Revision History:
Version Manual
(Art.-No.)
Date
(yyyy/mm/dd)
Analyser Software-Version - Release -
Comment
01.00
(105-96-510-00) 2019-07-31
V1.0, Feb. 8, 2019
Initial version
User Manual biobas 10
Content 3
Content
1 IMPORTANT INFORMATION 8
Definition OEM / PLD 8
Liability Disclaimer 8
Software Copyrights 9
Intellectual Property 9
Warranty 9
Exclusions of manufacturer’s warranty 10
Guidelines for getting service assistance 11
2 TERMS, ABBREVIATIONS AND SYMBOLS USED IN THIS MANUAL 13
3 SAFETY 16
Duty of care of the user 17
Intended use 17
Reagents 19
Safety instructions 19
Electrical safety 19
Mechanical safety 20
Reasonable foreseeable and remaining risks 20
Emergency and first aid 21
Fire fighting 21
4 INSTALLATION 22
Name Plate (Serial no.) 22
Component overview 23
Software 23
Mounting and connecting the analyzer 24
Thermal block and User interface 26
User Manual biobas 10
Content 4
External printer 28
Connection to an external Printer (Port A) 28
Activate and deactivate the external printer 30
CuvCARDs - Load a cuvette balance (optional) 31
ChipCARDs – Store or load methods 36
Connection to a HOST (Port B) 37
Connecting an external Barcode Scanner 38
5 ANALYZER AND METHOD PARAMETERS 41
Instrument parameters 41
Method memory positions and Default Method Parameters 42
6 OPERATION 44
Measuring principle 44
User qualification 45
Switch on the analyzer 45
STANDBY 47
Method selection 48
Measuring Mode (= “cuvette in”) 48
Measuring Mode (= “Patient ID”) 49
Carry out a PT Measurement 51
Carry out an aPTT Measurement 54
Carry out a Fibrinogen Measurement g/l 58
Parameter configuration - Access WITHOUT PIN 61
Calibration curve 61
Set calibration curve points for PT 62
Set calibration curve points for Fibrinogen 64
Set MNP value 66
Save parameters 67
Parameter configuration - Access WITH PIN 68
Menu <General> 69
6.6.1.1 Method name 69
User Manual biobas 10
Content 5
6.6.1.2 Load default 72
6.6.1.3 Copy method 74
6.6.1.4 Exchange method 76
6.6.1.5 Read from ChipCARD 78
6.6.1.6 Write to ChipCARD 80
Menu <1st conversion> 83
6.6.2.1 <None> 85
6.6.2.2 Set <Reference curve> 85
6.6.2.3 Set <Unit> 86
6.6.2.4 Set <Decimal place> 86
6.6.2.5 Set Min/Max value (Conversion limits) 87
6.6.2.6 Set calibration points 87
6.6.2.7 Set <Time Interpolation> 89
6.6.2.8 Set <Value Interpolation> 89
Menu <2nd conversion> 90
6.6.3.1 Select <INR> and set the ISI value 92
6.6.3.2 Select <RATIO> and set MNP 93
6.6.3.3 Set <none> 93
Menu <Measurement> 94
6.6.4.1 Set Start Reagent Volume/ Reagent Lot. Number
95
6.6.4.2 Set Incubation time 96
6.6.4.3 Set Start number for Print out 96
6.6.4.4 Set Mixer function 97
6.6.4.5 Set <min value> 98
6.6.4.6 Set <Learn-/Lag times> 98
6.6.4.7 Set Minimum and maximum time 99
6.6.4.8 Set measuring time 99
UTILITIES 101
SubMenu <Printer> 103
User Manual biobas 10
Content 6
SubMenu <Port A> 105
SubMenu <Port B> (virtual com port via USB) 106
SubMenu <Beeper> 110
SubMenu <Language> 111
SubMenu <Date/Time> 112
SubMenu <PIN Code> 114
SubMenu <Cuvette-test> 116
SubMenu <Parameter import> 118
SubMenu <Parameter export> 121
SubMenu <Info> 123
Sample print-outs PT and calibration 124
7 TROUBLESHOOTING 126
Application errors 126
Cancel incubation / measurement 127
Status messages CuvCARD (in alphabetical order) 128
Error messages (in alphabetical order) 130
Errors during operation 135
Warnings 136
8 MAINTENANCE AND HYGIENE 140
Safety 140
Operating and auxiliary materials 141
Disinfectant 141
Hand protection 141
Eye protection 141
Other 141
9 TRANSPORT 142
Prepare the analyzer for packaging 142
Pack the analyzer for shipment 143
10 SERVICE INFORMATION SHEET (CUSTOMER COMPLAINT) 145
11 DECOMMISSIONING AND STORAGE 146
User Manual biobas 10
Content 7
12 DISPOSAL OF THE ANALYZER 147
13 APPENDIX 148
Technical data 148
System related 148
Electrical data 149
Environmental requirements 150
Disposables 150
Materials supplied 151
Optional equipment 151
Supplier documentation 151
User Manual biobas 10
1 Important Information 8
1 IMPORTANT INFORMATION
Definition OEM / PLD
OEM (Original Equipment Manufacturer)
LABiTec®
LAbor BioMedical Technologies GmbH
An der Strusbek 6
D 22926 Ahrensburg
Tel.: +49-4102 47950
Fax.: +49-4102 479535
Email: [email protected]
Website www.labitec.de
PLD (Private Label Distributor)
SPINREACT, S.A.
Ctra Sta Coloma, 7
17176 St. Esteve d'en Bas (Girona) Spain
Liability Disclaimer
The OEM makes no express or implied warranty regarding this manual, its quality, performance, or appropriate use regarding any type of specific procedure. Furthermore, this manual may be modified by the OEM without notice and without implying any obligation or liability on the part of the company.
User Manual biobas 10
1 Important Information 9
Software Copyrights
All software used with the analyzer is the intellectual property of the OEM. Intellectual property rights shall remain with the OEM. You are entitled to use the software and the printed accompanying material at a place of work that cannot be transferred. Any violation of property rights or copyright or trademark or using conditions may be subject to legal action. The OEM reserves the rights to modify the software, documentation as well as this operator manual without prior written notice and without prior agreement from the PLD.
Intellectual Property
biobas 10 is a product name of the PLD. All rights reserved by OEM.
No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translation into any language (human or computer) in any form, or by any means whatsoever, without the prior written permission of the OEM.
Any reference to products, documentation, articles within this publication not expressly described as property of OEM shall be considered as intellectual property of the publishing company of this certain product, documentation and or article.
Warranty
The OEM, the PLD or his authorised sub-distributor warrant this analyzer against defects to the original purchaser in accordance with legal requirements or other mutual agreements after the date of supply in material and workmanship and defects arising from failure to conform to specifications applicable on the date of supply, provided they are properly installed, operated and used with accessories and consumables in accordance with the operating and service instructions.
The OEM, the PLD or his authorised sub-distributor further agree to correct, either by repair, or, at its election, by replacement, any such defect found on examination to have occurred, under normal use and service, during one-year period, provided The OEM, the PLD or his authorised sub-distributor is properly notified in writing upon discovery of such defect.
User Manual biobas 10
1 Important Information 10
Exclusions of manufacturer’s warranty
The use of none permitted peripheral devices, e.g. printer.
Improper use/operating faults and non-adherence to the user instructions.
Attempted repairs by the customer or third-parties without authorization by manufacturer.
Faulty maintenance by third parties without authorization by manufacturer.
Defects caused by power failure, like wrong main voltage, overvoltage, wrong frequency of main voltage or similar reasons.
Accidents, storms, lightening, fire, water/other liquids, other natural catastrophes, theft, riots, plundering, the effects of war or other instances of acts of god.
Defects caused by transportation/shipping.
The use of third party components, unauthorized consumables or accessories which do not comply to manufacturers specifications.
Defects caused by non-authorised analyzer and software changes.
Defects caused by bypass of safety functions, deletion of passwords etc.
Loss of customer data or software from repair and installation processes.
Defects caused by improper use of analyzer.
Defects caused by inappropriate environmental conditions.
Analyzer were information or analyzer parts have been removed or changed necessary for a clear identification
Parts of the analyzer which are subject to natural wear and tear.
Defects which are the result of breaks and scratches, dirtying on e.g. LCD displays, membrane keypads or dirtying of the measuring channel.
Defects which are caused by use of improper printer paper.
User Manual biobas 10
1 Important Information 11
Guidelines for getting service assistance
NOTE
Supporting our service department with the basic information as detailed as possible will help us to give you the needed assistance. Thank you in advance.
Customer identification
Customer
Name
Address
Country
Phone
Fax
Distributor / Enduser (in case of enduser, please state the distributor)
Analyzer identification
Type of analyzer
Serial-Number (SN)
Reference number (REF)
Software version
User Manual biobas 10
1 Important Information 12
Problem description
Describe the problem as detailed as possible. This may clarify the situation.
Display Messages
Printouts
Screenshots
Photos
Analytical data
In case of analytical problems further information will be needed.
Describe the problem as detailed as possible.
Test type: eg. PT, aPTT, Fibrinogen, ….
Sample (Patient sample, Control N, Control P, dilution, ….)
Reagent Manufacturer, Name of reagent, Order number, Lot Number, date of expire.
Leaflets
Actual measured results, raw value in seconds, converted values.
Conversion, Conversion table, ISI, …
User Manual biobas 10
2 Terms, abbreviations and symbols used in this manual 13
2 TERMS, ABBREVIATIONS AND SYMBOLS USED IN
THIS MANUAL
Table 2-1 Symbols used in this Manual, on the analyzer itself, and on available accessories and consumables
Symbol Meaning Symbol Meaning
Manufacturer
Date of manufacture
Batch Code
Serial Number
CE Conformity
Do not use if packaging is damaged
Refer to Operators Manual
Do Not Reuse
Protect From Sunlight.
Caution
Danger
Warning
Temperature Limitation.
Traceblility number according to customer.
Humidity Limitation.
Important Information.
Electric Shock Warning
Biological Risk.
User Manual biobas 10
2 Terms, abbreviations and symbols used in this manual 14
Symbol Meaning Symbol Meaning
Fuse
This side up.
Keep dry!
Separate collection, handling and disposal for waste electrical and electronic equipment and its components
Do not throw
Observe Precautions for handling electrostatic sensitive devices (ESD) / Warning of damages by electrostatic discharge
In vitro Diagnostica
User Manual biobas 10
2 Terms, abbreviations and symbols used in this manual 15
Table 2-2 Explanations of styles used in this manual
Symbol Meaning
● Instructions that have to be followed
Explanations to the given context
Things, that are required, have to be checked or fullfilled
Chapters
Steps that have to be carried out in the given order.
User Manual biobas 10
3 Safety 16
3 SAFETY
This analyzer conforms with the European Directives for in vitro diagnostica .
The analyzer type described in this manual bears an IVD and CE mark, which confirms the compliance with essential requirements of the Directive 98/79/EC of the European Parliament on in vitro diagnostic medical devices.
Table 3-1 Explanation of safety symbols and signal words according ISO 3864-1
Symbol Signal Word
Meaning
DANGER Indicates a direct hazard with high risk, which will have as consequence death or grievous bodily harm if it isn't avoided.
WARNING Indicates a possible hazard with medium risk, which will have as consequence death or (grievous) bodily harm if it isn't avoided.
CAUTION Indicates a hazardous situation, which, if not avoided, could result in minor or moderate injury.
NOTICE Indicates a Damage Message only device-related.
NOTE Indicates an important note in the manual. Duty of care of the operator
User Manual biobas 10
3 Safety 17
Duty of care of the user
This User manual describes the analyzer and is directed to all users of the device.
This User manual gives all information necessary for the implementation, maintenance and packing of the analyzer.
Careful observation of all information, especially of hazards and precautions will ensure the correct and safe operation of the analyzer. Therefore, it is absolutely necessary to read the User manual completely.
This User manual was prepared with greatest care. Should you have any questions or recommendations please refer to your distributor or the manufacturer.
Intended use
WARNING
For in-vitro diagnostic (IVD) use only!
The biobas 10 (Model CD12-1-17) named hereafter as analyzer, is a semi-automated 1-channel in-vitro diagnostic analyzer for coagulation of blood plasma. The type of the light source is an infrared LED.
The analyzer is capable to perform a wide range of coagulation tests such as Prothrombin time, activated and partial Thromin time, Fibrinogen.
The analyzer is intended for the use by trained operators in coagulation labs for in-vitro diagnostics.
Only use the analyzer according to the required ambient
conditions.
Do not use sharp objects to operate the buttons of the
anaylzer.
Protect the measuring channel from direct sunlight or other
light sources.
User Manual biobas 10
3 Safety 18
Only use pipettes which are being validated in regular
intervals.
Close the light protection cap prior to each measurement.
Make sure that pipetting does not cause any air bubbles.
Use a new tip after every pipetting operation to prevent
reagent/sample carry-over.
Only use pipette tips that fit the pipette (e.g. eppendorf
pipettes with eppendorf tips), otherwise secure handling is not
guaranteed.
Always place a cuvette in the measuring channel before
pipetting. Ensure each cuvette is equipped with a mixer.
Pipetting reagent or sample into the measuring channel can
significantly contaminate the analyzer and could even render
it faulty, so that expensive cleaning or repair might become
necessary.
Only use the original manufacturer’s cuvettes and mixers as
they are subject to strict quality control. The use of 3rd party
products and any associated analyzer problems as a result,
will invalidate your guarantee.
Use cuvettes once only. Multiple use of cuvettes (e.g.
cleaned) may produce incorrect results; this in turn can
become an indirect hazard for the patient.
Perform regular quality control checks. Refer to the guidelines
of use provided by the reagent manufacturer and the local
law.
Before use make sure that the reagents, controls and
disposables are not yet past their expiration date.
We recommend NOT to set the analyzer to silent mode.
User Manual biobas 10
3 Safety 19
Reagents
For proper coagulation analysis we recommend to use reagents, controls and buffers supplied by PLD.
Always read the information leaflet in the package and observe the instructions given by the reagent manufacturer.
NOTE
Utilize reagents and controls only according to directions as provided by the reagent manufacturer to avoid incorrect measuring results or malfunction of the analyzer.
Safety instructions
Electrical safety
DANGER - ELECTRICAL SHOCK
Never remove protective covers or secured components.
The stated input voltage of the wide-range power supply has to comply with the local mains voltage.
Never place the analyzer on a moist surface (floor, worktable or countertop), nor place containers of liquids on top of the analyzer. Liquids leaked out into the analyzer may cause an electrical shock.
Even after a device has been switched off, components can be still under voltage as a result of electrical charge and represent a risk of an electrical shock.
User Manual biobas 10
3 Safety 20
Mechanical safety
WARNING - MECHANICAL SAFETY
Put the analyzer out of operation and secure it against inadvertent use if you come to the conclusion that it can no longer be operated safely.
This is to be assumed in the following cases:
The analyzer shows visible damage.
The analyzer does not longer work.
The analyzer has been stored or transported for
a prolonged period under unfavorable
conditions.
Reasonable foreseeable and remaining risks
WARNING – BIOLOGICAL RISKS
Use disposable nitrile rubber gloves to avoid any hand contact with potentially infectious material.
Wear goggles as splashes of reagent, plasma or disinfectant might enter the eye.
WARNING - LASER LIGHT
Be careful when using the external barcode scanner.
Avoid direct eye contact with the laser beam.
Follow the manufacturer’s instructions.
User Manual biobas 10
3 Safety 21
Emergency and first aid
CAUTION
In case of emergency immediately interrupt the operation of the analyzer and pull out the external power supply from the socket.
Relevant actions to rescue personnell are given in the operational rules determined by the company.
Following procedure is recommendatory, not binding.
1. If the risk of injury remains when staying where the accident
happenend, bring the person immediately out of the danger
zone.
2. Otherwise take care of the person.
3. Provide first aid.
4. Call for help (colleagues, manager, doctor, ambulance…).
5. Take care of the injured person until further help arrives.
Fire fighting
There are no special measurements necessary for fire fighting.
We recommend to use common CO2 fire extinguisher.
Please consider the instructions defined by your company regarding fire protection and fire fighting.
User Manual biobas 10
4 Installation 22
4 INSTALLATION
Name Plate (Serial no.)
The name plate can be found on the bottom of the analyzer.
Information (signs) given on the plate is explained in the following.
Manufacturer according to IVD
Manufacture date
Type: Analyzer name (according to customer)
Traceblility number according to customer
Serial number of analyzer
Refers to the User manual with further information
Separate collection, handling and disposal for waste electrical and electronic equipment and its components
CE Conformity
In vitro Diagnostica
Input: Information about voltage and current
Fuse characteristic 1.25 AT
User Manual biobas 10
4 Installation 23
Component overview
Thermal block +37.4°C +/- 0.4°C
4 x cuvette positions
1 position for reagent bottles (NO stirring implemented)
1 measuring channel with light protection cap
Display 2 lines, 20 characters each
Membrane keypad with keys (0 - 9, Mode, Enter, ESC, <,>, Start and Reset)
ChipCARD / CuvCARD reading unit
Secure Digital Memory Card (SD Card)
Power supply socket
1x 6-pin Mini DIN RS232-C
1x USB interface (Type B)
Software
The analyzer software is stored in a memory chip and activated as soon as the analyzer is connected to the plug-in power supply attached to the mains supply.
The software controls the analyzer via start functions for the analytic program. Visual communication between the analyzer and the user is accomplished via a liquid crystal display with two rows and a 20 character string each.
Via the display the user is lead through all measurement steps. The user will confirm these steps either by key strokes or by pipetting start reagent. Thus, correct handling of the system is guaranteed.
User Manual biobas 10
4 Installation 24
Mounting and connecting the analyzer
Upon receipt remove the analyzer from its packaging check
the content for damage and completeness and verify that the
accessory kit is complete.
Immediately inform your distributor in the event that the
shipment was damaged or incomplete. Otherwise it results in
a loss of claims.
Upon unpacking keep all package material (e.g. membrane
packaging and card boards) stored at a dry place. You might
need them later on to send back the analyzer to the
manufacturer for repair or maintenance purposes.
Proceed as follows to install the analyzer:
1. Prior to installation of the analyzer read chapter 3, page 16.
2. Compare the input range of the plug-in power supply with the local mains voltage. Make sure that the external power supply meets the local mains voltage.
3. Place the analyzer in a location that is not exposed to excess humidity, any explosive gases, magnetic influences and direct light, e.g. sun and other light sources.
4. Place the analyzer on a stable and level surface with at least 150 mm space for ventilation at all sides.
5. Make sure to meet the environmental requirements, see chapter 13.1.3, page 150.
Figure 4-1 External Power supply
User Manual biobas 10
4 Installation 25
NOTICE
The analyzer is delivered with an external power supply.
The analyzer has to be connected exclusively with the delivered external power supply. Make sure to have a protected ground at the main socket. Otherwise a safe operation can not be guaranteed.
We do not warrant against a defect of the analyzer in case another plug-in power supply than the delivered has been used.
6. Connect the cable of the power supply to the analyzer and the external power supply itself to a socket outlet without voltage surges produced by heavy power users, e.g. lifts and centrifuges.
7. Ensure easy accessibility of the power plug at the analyzer rear side and/or at the external power supply at the wall socket to disconnect the device in case of an emergency.
Interfaces on the rear side
1 USB Port B
2 Serial interface for Barcode-Scanner or external Printer
3 Power supply socket
4 SD card-Slot
Figure 4-2 Interfaces on the analyzer rear side
1 2 3 4
User Manual biobas 10
4 Installation 26
Thermal block and User interface
Figure 4-3 Thermal block (left area), User interface (right area) and ChipCARD / CuvCARD reader slot on the right side of the analyzer
The operation area of the analyzer consists of the heated Thermal block with
Reagent position
Cuvette positions
Measuring channel
… and of the User interface with
Number keys
Mode / Esc / Enter keys
Start / Reset key
Arrow keys (for navigation)
Display
The function of the keys is described in the following table.
User Manual biobas 10
4 Installation 27
Table 4-1 Key description of the interface
Arrow-key left, right
Left = 1. select display to the left
2. delete
Right = 1. select display to the right, 2. set decimal point
Esc-key
1. Switch from measuring to STANDBY
2. Exit a submenu
3. Quit parameter change
4. Exit a menu or save entered or modified data.
Enter-key
1. Confirm selection
2. Confirm parameter change
Mode-key
1. Calibration
2. Menu selection, analyzer settings and method parameterization
Number-keys = Enter parameters
0-Key = A print-out of the respective
parameters for the selected method is generated by pressing 0 during measuring.
Start-key: manual start of measuring
- Start sample incubation timer
- Sample adjustment
- Manual test start
- Manual test stop
Reset key: manual stop of measurement,
Break of run, adjustment of sample
+
Current measurement is cancelled if keys are pressed simultaneously. Display: "break"
User Manual biobas 10
4 Installation 28
External printer
NOTE
Please read the suppliers manual carefully.
Connection to an external Printer (Port A)
For software activation of the external printer, see chapter 6.7.1, page 103.
1 Serial interface RS-232-C) / Barcode-Scanner / external printer (Port A)
Figure 4-4 Serial interface for an external printer
Serial interface data for external printer:
9600 baud
8 bit
no parity
1 stop bit
no Handshake
1
User Manual biobas 10
4 Installation 29
NOTE
Never operate the printer without paper!
Without inserted paper or if the printer function is deactivated you may not be able to print results and parameters.
We recommend to purchase and use the printer exlusively with the thermal paper offered by the manufacturer.
Operating the printer with another paper might result in malfunction of the printer with the following effects:
No paper transport.
Paper abrasion.
No print out.
Damage of the printing unit.
These effects may have the following reasons:
The paper is too thick / thin
The paper is too wide or small.
No thermal paper is used.
Paper is inserted upside down.
NOTICE
We do not warrant against a defect of the printing unit if the use of printer paper other than recommended is the is the cause of the defect.
NOTE
Mind proper storage conditions for the thermal printer paper:
Dry environment.
Protect from light.
Protect from heat.
User Manual biobas 10
4 Installation 30
Activate and deactivate the external printer
The printer is connected to the analyzer as well as to a power supply with a seperate cable.
The printer function in the analyzer is preset to "Off" by the manufacturer.
In case no thermal printer paper is available or necessary you can deactivate the printer.
To set the printer to <Off> proceed as described.
1. Press or and select UTILITIES.
2. Press and enter the PIN (default = 11111).
3. Press and select Port A.
4. Press to enter Port A Submenu.
5. Press and select „Printer <On>“ or „Printer <Off>“.
6. Press to confirm the selection.
User Manual biobas 10
4 Installation 31
7. Press if you want to adjust more parameters.
8. Press to switch to measuring mode.
If you have changed the settings following display appears:
9. Press to save parameters or to leave the menu without saving.
If parameters are unchanged an according message appears
and the analyzer turns to STANDBY automatically.
CuvCARDs - Load a cuvette balance (optional)
Without a cuvette balance no measurements can be carried out.
In this chapter it is described how to load a cuvette credit balance of original cuvettes by means of a cuvette card, so-called CuvCARD.
The individually loaded cuvette credit balance enables the user to perform coagulation measurements in the same quantity of cuvettes as loaded to the analyzer.
NOTE
Upon delivery the analyzer is preloaded with a cuvette balance that corresponds to the quantity of cuvette as delivered in the accessory box.
NOTE
Disposal of the CuvCARD may be done in the domestic waste. Ensure in your own interest that no balance remains on the CuvCARD.
User Manual biobas 10
4 Installation 32
Each unit of original cuvettes, suitable for the analyzer, is also
equipped with a corresponding CuvCARD. This CuvCARD
carries the same quantity of cuvettes as you will find in the
box.
Using the CuvCARD you can select, whether the total cuvette
credit balance shall be loaded to the analyzer. Otherwise, if
more analyzers of the same type are available, you can also
load only the necessary number to each analyzer.
Once the cuvette credit balance is fully used, meaning, if
measurements of the same quantity as the loaded cuvette
balance have been performed, the analyzer automatically
requests to re-load (reminder to order new cuvettes if used
up).
NOTE
We recommend to load the total credit balance on either one analyzer or, if more analyzer are in the laboratory, split it and load the necessary amount.
REASON: If the CuvCARD gets lost or is damaged the remaining balance is lost.
If you do not want to load the total balance, store the CuvCARD at a dry and safe place.
1. Set the analyzer to STANDBY.
Only in this state the analyzer can recognize CuvCARDs.
User Manual biobas 10
4 Installation 33
Figure 4-5 Insertion of CuvCARD with memory chip face up
2. Insert the CuvCARD face up into the reader slot (memory chip in the direction of insertion). Leave the CuvCARD in the reader.
The following display appears.
Display Meaning
<bookable> Displays the quantity of cuvette balance still remaining on the CuvCARD
<debit> Quantity of cuvettes on the CuvCARD which shall be loaded to analyzer.
3. Press to to enter the desired quantity of cuvettes.
The following entries are possible:
Minimum quantity
= 100 cuvettes (if e.g. 99 is entered, a beep indicates the automatic input of a balance of 100)
Maximum quantity
= as displayed at <bookable>
User Manual biobas 10
4 Installation 34
Entries between 100 and the maximum available quantity may
be done in steps of 10 cuvettes, for example 100, 110, 120,
…200 etc.
The display shows the remaining balance of cuvettes still
available on the analyzer.
This menu section only appears as long as the cuvette credit
balance on the analyzer is larger than or equals 100 cuvettes.
4. Press or to select between the following options:
Selection Meaning
<keep> Remaining cuvette balance on the analyzer shall be kept, no loading from CuvCARD.
<overwrite> Remaining cuvette balance on the analyzer shall be overwritten by loading a cuvette balance from the CuvCARD.
NOTE
When loading a cuvette balance from the CuvCARD to the analyzer, the remaining balance eventually still on the analyzer will be overwritten. Make sure the balance on the analyzer is fully used prior to loading a new balance from the CuvCARD!
5. Press to confirm the selection.
6. Wait until the message appears to remove the CuvCARD.
User Manual biobas 10
4 Installation 35
7. Remove the CuvCARD from the reading unit and the analyzer automatically turns to the measuring mode “cuvette in”.
An automatic printout follows if the printer is set to «On»
NOTE
To print a short overview of method relevand data e.g. Lot Number, remaining cuvette balance, select
the desired method, press to enter the
measuring mode and then press . The printout starts automatically.
As soon as the cuvette balance has been loaded, the analyzer automatically prints the following information.
CuvCARD balance
loaded to
biobas 10 Type of analyzer
Vxx.xx
SerNo. Axxxxxxx
Actual software version of the analyzer
Serial number of analyzer
Date/Time: Date and time of loading process
dd:mm:yyyy, hh:mm:ss
--CuvCARD Info--
Lot Number = xxxxxxxx Lot Number of loaded cuvette/mixer
Balance = xxx Loaded balance on analyzer
Remaining balance
on CuvCARD = xxx
Remaining balance on CuvCARD
User Manual biobas 10
4 Installation 36
ChipCARDs – Store or load methods
ChipCARDs are used to load or store method-specific
information in or from the analyzer.
Storage of the Methods on a ChipCARD is helpful, if the user
has changed the parameters inadvertendly and wants to
restore the default parameters.
The operator has the possibility to store method-specific data
and parameters from the analyzer on the ChipCARD in case
they are unitentionally changed and have to be reloaded. In
this case, the method stored on the ChipCARD can be
transferred/ written to another analyzer.
Unless a measurement is currently running, a ChipCARD can
be inserted into the reading unit to read a method from the
ChipCARD or to safe a method to a ChipCARD at any time.
If a ChipCARD has been read-in, no further adjustments need
to be carried out on the analyzer, because the analyzer
processes the data automatically.
The method-specific data on the analyzer will be overwritten
by the ChipCARD data.
Only one method can be stored at the time on the ChipCARD.
The operator has the possibility to set customized methods
not stored as default by the manufacturer and write them onto
a ChipCARD.
ChipCARDs can be ordered at any time. Please contact our
sales department.
For further information how to enter the menu, read chapter
6.6.1.5, page 78 and chapter 6.6.1.6, page 80.
For troubleshooting see chapter 7, page 126.
User Manual biobas 10
4 Installation 37
Connection to a HOST (Port B)
For software activation of the HOST connection, see chapter 6.7.3, page 106.
To connect the analyzer to the HOST a special USB cable with USB-A (for HOST) and USB-B (for analyzer) connection.
Figure 4-6 USB cable with USB-A (for HOST) and USB-B (for analyzer)
Figure 4-7 USB cable connected to the analzyer via USB-B
Figure 4-8 USB cable connected to HOST via USB-A
User Manual biobas 10
4 Installation 38
Connecting an external Barcode Scanner
1 USB-B Port (PORT B)
2 Serial interface RS 232-C / external Barcode-Scanner / Printer (PORT A)
Figure 4-9 Interface for External Barcode scanner, Printer and Host connection
If the analyzer is connected to a HOST via Port B, the software requests to scan a barcode (= patient-ID) or a manual input of the patient-ID number.
For software activation of the external Barcode Sanner see chapter 6.7.3 SubMenu <Port B> (virtual com port via USB), page 106.
The Barcode scanner can be connected via the RS 232-C interface (2).
Hardware adaption of the scanner cable might be required.
The power supply of the analyzer for the barcode is 5V DC
max. 150 mA.
The number input of the patient-ID can either be carried out by
entering the number manually (numerical data) or with a
barcode scanner (alpha numerical).
For functionality of the external BC Scanner the analyzer has
to be set to Port B = HOST to enable the communication.
1. Connect the analyzer to HOST via the USB-B Port (1)
2 1
User Manual biobas 10
4 Installation 39
2. Connect the Barcode scanner to serial interface RS 232-C (2).
3. Switch on the HOST and wait at least 30 sec.
4. Switch on the analyzer.
5. Select UTILITIES with or .
6. Press and enter the PIN (default = 11111).
7. Press or , select <Port B> and set Port B to HOST.
8. Press .
9. Press .
10. Press to save parameters or to leave the menu without saving.
The display shows “Patient ID” to indicate, that a number is
requested.
11. Press and manually enter the Patient-ID or scan the
barcode.
User Manual biobas 10
4 Installation 40
Automatically “cuvette in” appears to indicate, that the
measuring procedure can be started.
See chapter 6.4.3, page 49, to start e.g. a PT measurment.
User Manual biobas 10
5 Analyzer and Method parameters 41
5 ANALYZER AND METHOD PARAMETERS
Parameters are stored in the EEPROM memory. The parameters of the analyzer are preset by the manufacturer.
NOTE
Before routine tests can be performed, the user must change certain reagent- specific parameters such as Lot Number and calibration curve.
Otherwise results have no correct relation to the actual reagent.
Instrument parameters
The following default parameters are preset:
Program version: V x.xx Release mm.dd.yy
Printer Off
Port A Off
Port B Off
Beeper On
Language English
PIN Code 11111
User Manual biobas 10
5 Analyzer and Method parameters 42
Method memory positions and Default Method Parameters
The analyzers provide up to 15 method memory positions, whereas some are already preset with method parameters (factory defaults). The following overview in
Table 5-1 Default method parameter (for 1-channel), no parameters are deposited in Memory positions 8-15
on the next page, shows the preset method parameter and memory positions.
Before you start with your analysis the method parameters must be updated for the respective reagent as described in chapter 6.5, page 61 and chapter 6.6, page 68.
<empty>
The method memory positions <empty> are free positions where new methods can be entered manually or installed and modified by copying existing methods or using a ChipCARD, see chapter 6.5, page 61 and 6.6.1.5.
User Manual biobas 10
5 Analyzer and Method parameters 43
User Manual biobas 10
6 Operation 44
6 OPERATION
Measuring principle
The analyser operates according to the opto-mechanical measuring principle. This measuring principle is especially suited for lipemic and/or icteric colored samples as well as reagents with kaolin.
A light beam passes through the cuvette containing the test plasma onto a photo detector. Any change in the intensity of the transmitted light, that is light increase or decrease, is converted into an electric signal. Hence, even the most unstable clot can be detected.
The period from adding the start reagent until clot formation is measured. It then can be converted into the appropriate units (%, ratio, INR, mg/dl, g/l).
Once the start reagent has been added, the measuring channel is adjusted, that is the lamp intensity automatically adjusts up or down depending on the turbidity of the test sample. In this process the turbidity of the sample plasma and the reagent are adjusted.
A mixer is located in the cuvette. During the measuring process the mixer provides homogeneity of the reagent-plasma medium. At the same time a small whirl emerges through the mixer movement which assures that even the smallest fibrin clot is formed in front of the photo detector.
This stirring action combined with the optical measurement constitutes the basic features of the "turbodensitometric measuring principle".
User Manual biobas 10
6 Operation 45
User qualification
WARNING
It is absolutely necessary that only skilled personnel will access the parameter menu with a PIN Code as improper handling of the analyzer might cause inaccurate measuring results. (see chapter 6.7.7, page 114 <PIN Code>).
All settings in the parameter menu must be done in accordance with manufacturer requirements.
Upon alteration a parameter protocol must be printed in order to check all settings again.
Switch on the analyzer
NOTE
Communication with the analyzer is performed via the liquid crystal display. We assume that you are familiar with the function of the individual keys as described in chapter 4.5, page 26.
Contact your local distributor with regard to adjustment of display contrast if the display cannot be correctly read.
Connect the power supply plug to the analyzer and the socket,
respectively.
Following information appears on a “floating display”:
Analyzer name
Program version and date
Name of distributing company
User Manual biobas 10
6 Operation 46
Warm-up time
NOTE
The analyzer requires approximately 30 minutes to warm-up the thermal block to an operating temperature of 37.4°C.
During the warm-up no measurements are possible.
Only switching between menus is possible and settings can be adjusted.
The display shows the actual temperature of the thermal
block.
If the target temperature is reached, another 15 minutes serve
as temperature stabilizing time.
A timer counts down to 00:00 min. and displays the remaining
time until the analyzer is ready for measurements,
respectively.
Use the warm-up phase to load the analyzer with cuvettes and
reagents for testing. Ensure each cuvette is equipped with a
mixer.
Comply with the instructions of the reagent manufacturer.
Compare the method parameters with those stored in the analyzer.
For your own safety follow the instructions for hygiene.
User Manual biobas 10
6 Operation 47
STANDBY
When the analyzer has reached the required temperature the display turns automatically to STANDBY. The temperature of the thermal block as well as the last method used will be displayed.
PT Method is shown by default. However, always the last method set by the user will be displayed when switching on the analyzer.
1. Press .
2. Remove the cuvette from the measuring channel and close
the light protection cap.
NOTE
Make sure that the cuvette is removed out of the measuring channel prior to confirm the message, otherwise the automated cuvette detection and measuring channel adjustment may not work properly.
3. Press any key to confirm the message
The measuring channel is adjusted automatically.
4. The display shows “cuvette in” and the measurement can be
started.
User Manual biobas 10
6 Operation 48
Method selection
STANDBY is the level from where the user can switch between and select the necessary method or carry out settings in UTILITES menu
Press or to select a method or UTILITIES
After the required selection press and enter the PIN
(11111) to access the according menu.
You can also:
Press up to to quick select a test (method). Each
test is assigned to an according number (1 = PT; 2 = aPTT; 3
= Fibrinogen etc.).
Press to quick select UTILITIES.
To start the measuring see next chapter.
Measuring Mode (= “cuvette in”)
NOTE
Make sure the cuvette is removed from the measuring channel prior confirm the message and settings are correct.
Press to switch to the measuring mode of the according
method
The selected method has been initialized.
User Manual biobas 10
6 Operation 49
Only in this mode measurements can be carried out.
“cuvette in” requests the user to place the cuvette into the
measuring channel.
NOTE
For all tests you can do single determination measurement, with one result. A mean value calculation from two single values afterwards is not supported by the software.
Measuring Mode (= “Patient ID”)
If HOST is activated in the UTILITIES menu (chapter 6.7.3, page 106) the measuring is carried out the following.
1. From STANDBY press to enter the measuring mode (Patient ID).
2. Press .
3. Enter the patient ID manually (with number keys) or use the external barcode scanner.
4. Press within <30 sec.
User Manual biobas 10
6 Operation 50
NOTE
If you wait over 30 sec without entering the ID or with entering the ID but without pressing Enter to proceed, the following message appears:
<Read Pat-ID Time Out>
Then, the display automatically switches back to Patient ID display.
From this point (= “cuvette in”) the measurement can be
started as described in the next chapter.
When the measurement is finished (= result appears on the display) all information is also transferred to HOST via USB-B Port . See also chapter 4.9.1 Connecting an external Barcode Scanner, page 38.
User Manual biobas 10
6 Operation 51
Carry out a PT Measurement
NOTE
For the special preparation of Plasma and Reagents (reconstitution, prewarming) please see the according Plasma/Reagent leaflet.
Place the PT start reagent vial and a sufficient number of cuvettes in the thermal block at the same time as starting the analyzer (or at least 30 min before starting with the measurement). The temperature has to be constantly kept at 37,4°C in the reagent and cuvette positions.
Sample incubation
Sample incubation has to be carried out always in the measuring channel as the automatically started test-specific incubation time is necessary to get plausible and reproducible results.
1. Set the analyzer to STANDBY and select the PT method.
2. From STANDBY press to switch to the measuring
mode.
3. Pipette 50 µl plasma without air bubbles into a cuvette that
has been prewarmed in the thermal block to 37,4°C.
4. Open the light protection cap of the measuring channel, place
the cuvette into the measuring channel and close the light
protection cap.
The analyzer automatically recognizes the cuvette and starts
the timer for sample incubation (timer count down).
User Manual biobas 10
6 Operation 52
A 5 second period of intermitting beeps indicates, that the
measurement can be startet soon.
The display shows INCUBATION READY (incubat. ready) and
then AUTO ADJUST.
After the sample incubation the measuring channel will be
adjusted (zero adjustment) for the measurement (adjust).
Add start reagent
After the adjustment the following display appears for the PT-
Measurement and requests to add 100 µl start reagent.
5. Aspirate 100 µl pre-warmed start reagent and pipette it
vertically through the hole of the light protection cap of the
measuring channel into the cuvette.
The measurement starts automatically.
During the measurement the display shows the measuring
time until the clot is formed.
User Manual biobas 10
6 Operation 53
If the beeper is activated an acoustic signal indicates the
recognition of clotting in the cuvette and stops the timer.
The maximum measuring time is set for every method
individually in the parameter settings, thus if no clotting is
detected an acoustic signal indicates that the maximal
measuring time is reached without clotting detection.
If conversions are programmed the measured values will be
converted automatically and appear on the display.
Results
Results (time in sec, Conversion to % and INR) are displayed.
If a printer is connected and set to <Auto> the result is printed
automatically.
If the printer is set to <On> press to print the result.
PT
Patient _ _ _ _ _
dd:mm:yyyy, hh:mm:ss
No. = 1
Channel = 1
Time = 13.7 s
% = 74.4
INR = 1.18
Figure 6-1 Example print out of PT result
After the print out, “cuvette out” appears in the display.
User Manual biobas 10
6 Operation 54
Open the light protection cap, take out the cuvette, close the
cap and press
The display shows “cuvette in” again and the next
measurement can be started.
Carry out an aPTT Measurement
NOTE
For the special preparation of Plasma and Reagents (reconstitution, prewarming) we refer to the according Plasma/Reagent leaflet.
Place the reagent vial (CaCl2) in the thermal block at the same time as starting the analyzer (or at least 30 min before starting with the measurement). The temperature has to be constantly kept at 37,4°C.
Sample incubation
Sample incubation has to be carried out always in the measuring channel as the automatically started test-specific incubation time is necessary to get plausible and reproducible results!
1. Set the analyzer to STANDBY and select the aPTT method.
2. Press to switch to measuring mode.
User Manual biobas 10
6 Operation 55
3. Pipette 50 µl of plasma sample without air bubbles into a pre-warmed cuvette.
4. Add 50 µl aPTT reagent without air bubbles into the same
cuvette onto the plasma.
5. Open the light protection cap of the measuring channel and
place the sample cuvette into the measuring channel.
6. Close the cap of the measuring channel.
The analyzer automatically recognizes the cuvette and starts
the timer for sample incubation (120 sec. timer count down).
A 5 seconds period of intermitting beeps indicates the
remaining incubation time and the display shows
INCUBATION READY and then AUTO ADJUST.
After sample incubation the measuring channel will be
adjusted (zero adjustment) for the measurement.
Add start reagent
After the adjustment the following display appears:
User Manual biobas 10
6 Operation 56
7. Pipette 50 µl of prewarmed start reagent (CaCl2) vertically
through the hole of the light protection cap of the measuring
channel into the sample cuvette.
The measurement starts automatically.
During measurement the display shows the measuring time
until the clot is formed.
If the beeper is activated an acoustic signal indicates the
recognition of clotting in the cuvette and stops the timer.
The maximum measuring time is set for every method
individually in the parameter settings, thus if no clotting is
detected an acoustic signal indicates that the measuring time
is reached without clotting detection.
If conversions are programmed the measured values will be
converted automatically.
Results
Results (time; Ratio) are displayed.
If a printer is connected and set to <Auto> the result is printed
out automatically.
User Manual biobas 10
6 Operation 57
If the printer is set to <On> press to print the result.
aPTT
Patient _ _ _ _
dd:mm:yyyy, hh:mm:ss
No. = 2
Channel = 1
Time = 31.5 s
Ratio = 1.13
Figure 6-2 Example print out of aPTT result
After the print out, “cuvette out” appears in the display.
Open the light protection cap, take out the cuvette, close the
cap and press .
The display shows “cuvette in” again and the next
measurement can be started.
User Manual biobas 10
6 Operation 58
Carry out a Fibrinogen Measurement g/l
NOTE
For the special preparation of Plasma and Reagents (reconstitution, prewarming) we refer to the according Plasma/Reagent leaflet.
Sample incubation
Sample incubation has to be carried out always in the measuring channel as the automatically started test-specific incubation time is necessary to get plausible and reproducible results.
1. Set the analyzer to STANDBY and select the Fibrinogen method.
2. Then press to switch to measuring mode.
3. Pipette 100 µl of a pre-diluted (1:10) plasma sample without air bubbles in a prewarmed cuvette in the thermal block to 37,4°C.
4. Open the light protection cap of the measuring channel.
5. Immediately insert the cuvette into the measuring channel.
6. Close the light protection cap of the measuring channel.
The analyzer automatically recognizes the cuvette and starts
the timer for sample incubation (60 sec. timer count down).
User Manual biobas 10
6 Operation 59
A 5 seconds period of intermitting beeps indicates the
remaining incubation time and the display shows
INCUBATION READY and then AUTO ADJUST.
After sample incubation the measuring channel will be
adjusted (zero adjustment) for the measurement.
Add start reagent
After the adjustment the following display appears:
7. Pipette 50 µl start reagent vertically through the hole of the light protection cap of the measuring channel into the sample cuvette.
The measurement starts automatically.
During measurement the display shows the measuring time.
If the beeper is activated an acoustic signal indicates the
recognition of clotting in the cuvette and stops the timer.
The maximum measuring time is set for every method
individually in the parameter settings, thus if no clotting is
detected an acoustic signal indicates that the maximum
measuring time is reached without clotting detection.
If conversions are programmed the measured values will be
converted automatically.
User Manual biobas 10
6 Operation 60
Results
Results (time; g/l) are displayed.
If a printer is connected and set to <Auto> the result is printed
automatically.
If the printer is set to <On> press to print the result.
Fib. g/l
Patient _ _ _ _ _
dd:mm:yyyy, hh:mm:ss
No. = 3
Channel = 1
Time = 5.7 s
g/l = 3.87
Figure 6-3 Example print out of Fibrinogen result
After the print out, “cuvette out” appears in the display.
Open the light protection cap, take out the cuvette, close the
cap and press .
The display shows “cuvette in” again and the next
measurement can be started.
User Manual biobas 10
6 Operation 61
Parameter configuration - Access WITHOUT PIN
The following parameters can be accessed from the measuring mode (cuvette in) directly without entering the PIN Code.
1st conversion with calibration points
ISI
Start Reagent Lot Number
MNP value (depending on method)
2nd conversion (MNP value)
This chapter is a Quick Guide for adjustments of the calibration points according to the selected method and the Start Reagent Lot.No.
Calibration curve
Although method parameters of the analyzer do include preset values for the calibration curve (of PT and Fibrinogen; %, g/l, mg/dl and time in sec) it is necessary to adjust the parameters any time you change the Lot Number of the reagents.
The reason is, that the reagent has lot-specific properties, which could lead to wrong results when the settings do not match the reagents results.
Calibration curves are carried out with special calibration plasma and the new reagent.
NOTE
To set calibration curves /points all relevant methods have to be set to “ON” under column “Reference curve” in Table 5-1, page 42.
A dilution series is prepared to later on set the respective
measured values as calibration points in the analyzer.
We recommend to carry out a double measurement for each
dilution and calculate the mean.
User Manual biobas 10
6 Operation 62
Each mean is entered in the analyzer as calibration point as
described in the following chapters.
NOTE
Points on the reference curve that are not to be used must have the entries 0.0% or 0.0 s.
A minimum of 2 points must be defined. A maximum of 9 points can be entered.
We recommend defining 4 points (100%, 50%, 25%, 12.5%).
Set calibration curve points for PT
1. Select the required Method in STANDBY, here PT.
2. Press to set the analyzer to measuring mode.
3. Press from "cuvette in" directly without entering the
PIN Code.
The analyzer switches directly to the first calibration point.
(Point of highest activity and lowest coagulation time,
previously determined by the calibration curve measurement).
4. Use number keys to enter the activity of e.g. 100.0% and to set the decimal point.
5. Press to confirm and enter the next line.
User Manual biobas 10
6 Operation 63
6. Use number keys to enter the clotting time and to set the
decimal point for the respective activity.
7. Press to confirm and set values of the second
calibration points.
8. Continue the same way to set subsequent calibration points (50%, 25%, 12,5%)
9. Press to confirm and switch to the ISI.
10. Enter the ISI-value of the respective reagent as indicated in
the according reagent leaflet in the table of values with
number keys and to set the decimal point.
11. Press to confirm the entry.
12. Press to confirm and set the Lot Number with number
keys of the according Start reagent.
13. Press
User Manual biobas 10
6 Operation 64
14. Press to save parameters or to leave the menu
without saving.
The analyzer switches automatically back to the measuring
mode.
Set calibration curve points for Fibrinogen
1. Select required Method in STANDBY, here Fibrinogen g/l.
2. Press to set the analyzer to measuring mode.
3. Press from " cuvette in" directly without entering the
PIN Code.
The analyzer switches directly to the first calibration point. (the
point with the highest concentration, thus the lowest time).
4. Use number keys to enter the activity of e.g. 6 g/l
and to set the decimal point.
5. Press to confirm and enter the next line.
User Manual biobas 10
6 Operation 65
6. Use number keys to enter the clotting time and
to set the decimal point for the respective activity
7. Press to confirm and set values of the second
calibration points.
Continue the same way to set subsequent calibration points.
8. Press to confirm and switch to set Lot.No. of the according Start reagent leaflet.
9. Use number keys to set numbers, press to
switch to the next entry position, and press or to enter characters or decimals).
10. Then press .
11. Press to save parameters or to leave the menu
without saving.
The analyzer switches automatically back to the measuring
mode.
User Manual biobas 10
6 Operation 66
Set MNP value
NOTE
MNP = Mean of Normal Plasma
1. Select required method in STANDBY, here aPTT.
2. Press to set the analyzer to measuring mode.
3. Press from " cuvette in" directly without entering the
PIN Code.
4. Press and set the MNP value with number keys
and to set the decimal point.
5. Press to confirm and switch to set Lot.No. of the according Start reagent.
6. Use number keys to set numbers, press or
. to enter characters or decimals and press to switch to the next entry position and continue until the following message appears.
User Manual biobas 10
6 Operation 67
Press to save parameters or to leave the menu
without saving.
The analyzer switches automatically back to the measuring
mode (cuvette in).
Save parameters
Generally, after finishing parameterization or when leaving the method menus the analyzer checks the parameters and the following sequence appears.
If no parameters were changed:
If parameters were changed:
Press to save parameters or to leave the menu
without saving.
User Manual biobas 10
6 Operation 68
Parameter configuration - Access WITH PIN
Detailed adjustments can only be accessed and set with a PIN code, respectively.
Method Parameter <General>
Method Parameter <1st conversion>
Method Parameter <2nd conversion>
Method Parameter <Measurement>
Adjustment of the settings are similar among the menus (PT, aPTT, Fibrinogen, etc.).
1. Set the analyzer to STANDBY.
2. Press or to select the required method (here PT is
shown).
3. Press to enter respective method.
4. Enter the PIN Code (Default no. 11111) to access the method
parameter menu. See also chapter 6.7.7 SubMenu <PIN
Code>, page 114.
If the wrong number was entered STANDBY will be displayed again. As soon as the correct number has been entered, access to PT Method parameter settings is possible.
5. Press to enter the respective menu and its submenus.
User Manual biobas 10
6 Operation 69
Menu <General>
Press to access the Menu <General>.
The following submenus are available:
Selection Meaning
<Method name> Enter or change a method name
<Load default> Load a default method from memory
<Copy method> Copy a method
<Exchange method>
Exchange method memory positions
<Read from ChipCARD>
Read a method from a ChipCARD to the analyzer
<Write to ChipCARD>
Safe a method to a ChipCARD
6.6.1.1 Method name
The method name, as displayed in STANDBY, can be edited with
the character generator (with arrow keys or ).
The character generator comprises the following characters:
! " § $ % & /= ? + * # < > , . : _
A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
a b c d e f g h i j k l m n o p q r s t u v w x y z.
To enter or modify a method name follow the description as shown below.
User Manual biobas 10
6 Operation 70
1. Select the necessary method, press and enter the PIN
with number keys to access the menu. The first
display “Method-Parameter <General> appears.
2. Press to confirm.
3. Press or to select <Method name>
4. Press to confirm selection.
The next display appears:
The cursor blinks on the first character position.
5. Press or to set the desired character.
6. Press to confirm the character position and to move to
the next position.
7. Repeat this process to enter more characters.
8. Once the last character is entered press until the final
position is reached.
User Manual biobas 10
6 Operation 71
9. Press to finally confirm your entry.
NOTE
If you keep the arrow-key pressed the characters will be displayed faster and you can speed up your selection.
If you keep the Enter-key pressed the cursor will move faster.
Delete character
Set the cursor with behind the character you want to
delete.
Press to delete the character to the left.
Every stroke means one deletion.
User Manual biobas 10
6 Operation 72
6.6.1.2 Load default
Some method memory positions of the analyzer have been preset by the manufacturer. In Table 5-1, page 42, you will find an overview of the default methods. These parameters are saved in the preset ROM of the analyzer and can be reloaded.
Either to restore adjusted parameters of e.g. <1 PT> or to load a default method to an empty memory position e.g. <14 empty> follow the instructions below.
This is an example to restore original PT settings to memory position <1 PT>.
1. Select <1 PT> in STANDBY.
2. Press and enter the PIN Code to access the parameter
menu.
3. Menu <General> appears, press to enter the menu.
4. Press or and select <Load default>
5. Press to confirm.
6. Press or to select the desired method, here
<1 PT>.
The following display appears:
User Manual biobas 10
6 Operation 73
7. Press to confirm.
Automatically the next display will be displayed.
8. Press to overwrite the actual method or to leave
the menu without overwriting the parameters.
Automatically the next submenu appears.
Either you enter more settings or return to STANDBY.
To return to STANDBY proceed with the following:
9. Press several times until the following display appears:
If parameters have been changed, the following display
appears:
10. Press to save the changes or press to leave the
menu without saving.
The analyzer automatically switches to STANDBY and the
user can proceed with the operation.
User Manual biobas 10
6 Operation 74
6.6.1.3 Copy method
If you want to load a method which is rather similar to an existing method you can copy it to the desired method memory position and modify the parameters where necessary.
Follow the instructions below to copy an existing method to another memory position. For example, method <1 PT> shall be loaded to an empty memory position.
1. When in STANDBY select the method memory position e.g. <11 empty>.
2. Press and enter the PIN Code to access the parameter menu.
3. Select the menu <General> and press or to select the submenu <Copy method>.
4. Press to confirm the selection. The following display
appears:
5. Press to select the desired method <1 PT>.
6. Press to confirm the selection. The following display
appears:
User Manual biobas 10
6 Operation 75
7. Press to overwrite the desired method or to leave
the menu without saving.
Automatically the next submenu appears.
Either you enter more settings or return to STANDBY.
To return to STANDBY proceed with the following:
11. Press several times until the following display appears:
If parameters have been changed, the following display
appears:
12. Press to save the changes or press to leave the
menu without saving.
The analyzer automatically switches to STANDBY and the
user can proceed with the operation.
User Manual biobas 10
6 Operation 76
6.6.1.4 Exchange method
Sometimes it might be useful to exchange method memory positions. Follow the instructions below to exchange two method memory positions. For example, the method on position <11 empty> shall be exchanged with method <1 PT>.
1. When in STANDBY select the method memory position
<11 empty>.
2. Press and enter the PIN Code to access the parameter
menu.
3. In the Menu <General> press or to select the
submenu <Exchange method>.
4. Press to confirm the selection. The following display
appears:
5. Press or to select the desired method in this
example e.g <1 PT> and then press to confirm and the
next display appears:
6. Press to exchange or to keep the settings.
User Manual biobas 10
6 Operation 77
When exchanging, all parameters will be exchanged.
To return to STANDBY proceed with the following:
7. Press several times until the following display appears:
If parameters have been changed, the following display
appears:
8. Press to save the changes or press to leave the
menu without saving.
The analyzer automatically switches to STANDBY and the
user can proceed with the operation.
User Manual biobas 10
6 Operation 78
6.6.1.5 Read from ChipCARD
Background information for the ChipCARD in general is given in chapter 4.8 ChipCARDs – Store or load methods, page 36.
To load method-specific data from a ChipCARD to the analyzer follow the next steps described.
1. When in STANDBY press or to select the memory
position you would like to overwrite with the ChipCARD data,
for example <11 empty>.
2. Press and enter the PIN Code to access the selected
memory position.
3. Press or and select <General> , press .
4. Press or and select <Read from ChipCARD> with
.
5. Insert the ChipCARD into the reader slot with the memory chip facing up. Leave the ChipCARD in the reader, see Figure 4-5, page 33 for correct insertion and chapter 4.8 ChipCARDs – Store or load methods, page 36 for more information.
If no ChipCARD is inserted or if the ChipCARD is not correct a
message “INVALID ChipCARD” as floating text appears:
6. Remove the incorrect ChipCARD and insert a correct one.
7. Leave the ChipCARD in the reader.
8. Press to confirm.
User Manual biobas 10
6 Operation 79
9. Press or to select between the following options:
10. Select <Yes> to start the loading process.
11. Press to confirm.
The following display appears:
With “Reading parameters from ChipCARD “ the selected
method on the analyzer will be replaced by the new
parameters coming from the ChipCARD.
The following display with new method name and corresponding Lot.-No. appears. This is an example name of the method.
12. Remove the ChipCARD from the ChipCARD reader.
NOTE
Before starting measurements with the new method we recommend to set the new method name and enter the correct Lot.-No. of the reagent.
The next submenu <Write to ChipCARD> will be displayed automatically.
To return to STANDBY:
User Manual biobas 10
6 Operation 80
13. Press until the message "checking parameters please
wait" appears.
14. If there was no change in the parameters the analyzer
switches automatically to STANDBY.
15. If parameters have been changed the following message
appears:
16. Press to save parameters or to leave the menu
without saving.
6.6.1.6 Write to ChipCARD
Apart from the possibility to load method-specific data from a ChipCARD, such data may also be saved from the analyzer onto a ChipCARD for later use.
The steps are the same between all methods.
Remember however, that only one method can be saved on one ChipCARD at a time.
A CuvCARD also a used CuvCARD – that means – a CuvCARD with no credit left cannot be used as a ChipCard to store method parameters.
Measurement results and measurement curves cannot be saved.
To save method-specific data on a ChipCARD follow the steps as decribed.
1. When in STANDBY, select the method memory position you would like to save on the ChipCARD, for example <1 PT>.
2. Press and enter the PIN to access the selected memory position.
User Manual biobas 10
6 Operation 81
3. Press or and select <General> , press .
4. Press or and select <Write to ChipCARD> with
.
5. Insert the ChipCARD into the reader slot with the memory chip facing up. Leave the ChipCARD in the reader, see Figure 4-5, page 33 for correct insertion and chapter 4.8 ChipCARDs – Store or load methods, page 36 for more information.
6. Press to confirm the selection.
7. Press or to select among the following options:
8. Select <Yes> and and press to confirm.
The following display appears:
then
With “Writing parameters to ChipCARD “ the parameters on
the ChipCARD will be replaced by the new parameters coming
from the analyzer.
User Manual biobas 10
6 Operation 82
9. Remove the ChipCARD from the ChipCARD reader.
Automatically, the next menu appears:
10. Press until “checking parameters please wait" appears,
then
11. Press to save parameters or to leave the menu
without saving the changes.
12. Automatically STANDBY or the measuring mode is reached.
User Manual biobas 10
6 Operation 83
Menu <1st conversion>
NOTE
Evaluating the measurement results with the set 1st conversion parameters means that the results are calculated based on previously determined calibration points or calibration values given in the lot-specific PT reagent leaflet.
You have two options to change the <1st conversion> settings.
1. WITHOUT a PIN Code as described in chapter 6.5, Parameter configuration - Access WITHOUT PIN, page 61.
2. WITH a PIN Code as described in the following.
Method parameters are preset by the manufacturer. Before
you start with the analysis, certain parameters (Lot.-No.,
calibration points) must be updated for the respective reagent.
Table 6-1 Overview 1st conversion parameters
Selection Meaning
<Reference curve> <rising/falling>
Rising or falling reference curve
<Unit> Unit in % of reference curve to be entered by the text generator
<Decimal place> No. of digits after decimal point (format xxx.x)
<Min./Max. value> Ranges of activity/concentration
<Points> Entering a 9-point reference curve
<Time interpol.> Linear / Reciprocal / Logarithmic
<Value interpolation>
Linear / Reciprocal / Logarithmic
<None> No conversion
User Manual biobas 10
6 Operation 84
1. Set the analyzer to STANDBY <1 PT>.
2. Press enter the PIN Code (factory default: 11111).
See also chapter 6.7.7 SubMenu <PIN Code>, page 114.
If the number was incorrect STANDBY display appears again.
If the correct PIN Code was entered, the following display
appears:
3. Press or to switch to 1st conversion
4. Press .
5. Press to enter reference curve options
User Manual biobas 10
6 Operation 85
6.6.2.1 <None>
Press or and set 1st conversion to <None>.
Then press to confirm.
If <None> is selected instead of <Reference curve> under
<1st conversion> no conversion will be calculated.
6.6.2.2 Set <Reference curve>
1. Press the following display appears:
2. Press or to choose between falling and rising.
3. Press to confirm the selection.
NOTE
Reference curves must be entered as follows:
Rising curves (inhibitors) start with the lowest value.
Falling curves (coagulation factors) start with the shortest measuring times (sec) and the highest converted value (%, g/l, mg/dl).
The software will support this procedure only!
User Manual biobas 10
6 Operation 86
6.6.2.3 Set <Unit>
Character format, e.g. mg/dl.
max.: 4 digits.
The fraction bar is fix.
Select the characters for the unit and press to confirm.
6.6.2.4 Set <Decimal place>
NOTE
Regardless of the language used a dot is always used as decimal instead e.g. a comma.
The display to enter the digits (decimal place) appears.
There are 4 alternatives:
1. <Format xxxx.>
2. <Format xxx.x>
3. <Format xx.xx> or
4. <Format x.xxx>
Press or to select the required format and press
to confirm.
User Manual biobas 10
6 Operation 87
6.6.2.5 Set Min/Max value (Conversion limits)
You can enter the minimum and maximum values for the calibration curve and conversion.
Set minimum and maximum value with number keys
and to set the decimal point and press to
confirm.
An error message (Min Value reached – no more points)
appears if the values are outside the ranges or only the time
but no converted result will be displayed.
6.6.2.6 Set calibration points
In the following sequence a 9-point reference curve can be entered.
A minimum of 2 points must be defined. We recommend
defining at least 4 points.
Points in the reference curve that are not used need entries of
0.0 % and 0.0 s. As soon as e.g. the 5th point has the entry 0.0
the next point (here the 6th) will not appear.
Before you save the adjustments and return to the measuring
mode, check all calibration points for proper entries.
Press to enter the first reference curve point.
For a “PT” the point of highest activity and lowest coagulation
time in the calibration is for example the following:
User Manual biobas 10
6 Operation 88
1. Press to confirm the activity of 100.0%, or
2. Press number keys to overwrite the actual values and to
set the decimal point.
3. Press to confirm the entry. The cursor changes to a time setting.
4. Enter the clotting time for the respective activity of 100.0%.
5. Press to confirm the entry, the field for the next point on
the reference curve appears automatically.
1. Verify or adjust the second calibration point as previously
described.
2. Proceed as described above to set the remaining calibration
points.
User Manual biobas 10
6 Operation 89
6.6.2.7 Set <Time Interpolation>
As soon as the last point on the reference curve has been confirmed the following display will appear:
In the following display you can select Time interpolation modes of the time axis the following options can be selected:
<linear>, <reciprocal>, <logarithmic>.
Press or to select e.g. <linear> and press to
confirm.
6.6.2.8 Set <Value Interpolation>
In the following display you can select Value interpolation modes for the activity/concentration: <linear>, <reciprocal>, <logarithmic>.
Press or to select e.g. <reciprocal> and press
to confirm.
The following display appears:
Press to access measuring or to set or verify
more parameters.
User Manual biobas 10
6 Operation 90
Menu <2nd conversion>
NOTE
Results are calculated based on set MNP (= Mean Normal Plasma).
ISI values are given in the leaflet of the lot-specific PT-reagent.
RATIO = patient sample in sec is divided by MNP in sec.
INR = International Normalized Ratio
ISI = International Sensitivity Index according to package insert
1. Enter the settings WITHOUT a PIN Code as described in chapter 6.5, page 61.
2. Enter the settings WITH a PIN Code as described in the following.
NOTE
Only if <None> has been selected for <1st conversion> an input field to enter the MNP (Mean Normal Plasma) value will automatically be displayed for the INR/RATIO conversion under <2nd conversion>.
In case of the aPTT-Method, the 1st conversion is none and the MNP has to be entered.
If the <Reference curve> is selected under <1st conversion> and the reference curve consists of valid points, the MNP value is taken from the reference curve. Thus, there is no option to set the MNP.
The following sub-menus can be selected if <reference curve> or <none> was selected under 1st conversion.
User Manual biobas 10
6 Operation 91
Table 6-2 Overview 2nd conversion submenus
Selection Meaning
<INR> -→ MNP-→ ISI For calculation of the INR-value enter the ISI according the given in the PT reagent leaflet
<RATIO> -→ MNP To enter the MNP (only if <1st conversion> is set to <None>. e.g. Mean value of the normal plasma aPTT range.
<None> No conversion
Press to access the submenu <2nd conversion>.
User Manual biobas 10
6 Operation 92
6.6.3.1 Select <INR> and set the ISI value
1. Press and the following display appears:
2. Press to confirm and enter the next display.
3. Use number keys to enter the ISI-value and to set the decimal.
4. Press to confirm the entry.
5. Press to access measuring or to set or verify
more parameters.
User Manual biobas 10
6 Operation 93
6.6.3.2 Select <RATIO> and set MNP
If for <1st conversion> <none> is selected, the following display appears to enter the MNP time value.
1. Select 2nd conversion and press to confirm. The
following display appears:
2. Use number keys to enter the MNP, use to set the decimal point.
3. Press to confirm the entry of normal time value.
NOTE
If the 1st conversion is active, the MNP value is not accessible in the RATIO submenu.
6.6.3.3 Set <none>
If <none> is selected no INR and no Ratio will be calculated. The result is only displayed as the clotting time in seconds.
User Manual biobas 10
6 Operation 94
Menu <Measurement>
CAUTION
Any modifications within this menu can cause different measuring results!
To avoid incorrect measuring results only authorized personnel should have access to the method parameterization!
After each entry a parameter protocol must be printed in order to check the entry again!
Press to access menu <Measurement>.
The following selections are available.
Table 6-3 Overview Measurement parameter
Selection Meaning
<Start reagent> To enter the start reagent volume and reagent Lot Number
<Incubation> To enter the 1st and 2nd incubation time
<Printout No.> To enter a Start-No. for print-outs
<Mixer> Mixer starting speed - period of reduction until final speed is reached - mixer final speed
<learn, lag> To enter learn and lag times
<Meas.time> Total measuring time
User Manual biobas 10
6 Operation 95
6.6.4.1 Set Start Reagent Volume/ Reagent Lot. Number
Example for Method <PT> reagent and Lot Number.
NOTE: Entry range: 1-250 µl.
The Lot. No. can have an maximum of 12 characters.
1. Enter the volume for the start reagent with the number keys
, for example 100 µl.
2. Press to confirm the entry.
3. Enter the reagent Lot Number manually the same way.
4. Press to confirm the entry.
NOTE
Alternatively, if a barcode scanner is connected:
Scan the barcode label of the reagent vial (a
maximum of 12 alphanumeric digits is imported
as string). The Software automatically displays
the next menu after the string was imported and
displayed for 3 seconds.
User Manual biobas 10
6 Operation 96
6.6.4.2 Set Incubation time
NOTE
For the PT method, the first incubation time is set to 0 by definition. The according value is assigned to the second incubation. If another parameterization is set, the analyzer might change the measuring procedure which could lead to missleading results and interpretation.
See according reagent leaflet for correct incubation times.
Example for PT-Method Incubation time of Start reagent.
NOTE: 0 s = no incubation, Max.: 600 s.
1. Enter the method-dependend 1st sample incubation time for
1st = 0 s and 2nd =xxx s with the number keys .
2. Press to confirm the entry.
6.6.4.3 Set Start number for Print out
The next display appears.
NOTE: Max.: 3-digits (1-999).
Enter the start number for print out by using the number keys
.
Press to confirm the entry.
User Manual biobas 10
6 Operation 97
6.6.4.4 Set Mixer function
NOTE
The mixer function is implemented to gently and properly mix the sample (and reagent).
Information to method-specific settings are given in chapter 5, page 41, Table 5-1.
Before modification consult your local distributor.
Example for Method Fibrinogen mg/dl
1. Enter the time period in sec = time from rpm1 to rpm2.
2. Enter the starting speed (0, 200 - 800 rpm).
3. Enter the end speed (0, 200 - 800 rpm).
In this example the measurement will start with 500 rpm (rpm1) and reaches 200 rpm (rpm2) after 40 seconds.
Figure 6-4 Exemplary function of mixer speed reduction
User Manual biobas 10
6 Operation 98
6.6.4.5 Set <min value>
Here you can enter the (individual evaluated) mininum value
(as digits) for the light that has to pass to the sample.
In case the sample is very turbid due to e.g. lipemic samples
the passing light goes below this minimum value.
As a result a "Break Dark" message appears on the display.
‘0’ means off, the detection of a mininum value is deactivated.
The curser is blinking and the value can be entered with
number keys.
Press to continue (to the next display).
6.6.4.6 Set <Learn-/Lag times>
Example for PT-Method
Here you can enter parameters for the optical detection of the
sample.
Learn 1…60 sec phase: Thresholds definition phase
Lag 0…60 sec phase: Time phase without measuring curve
analysis.
NOTE
Before any modification consult your local distributor.
User Manual biobas 10
6 Operation 99
1. Press to confirm the entry.
2. Press to access measuring or to set or verify more parameters.
6.6.4.7 Set Minimum and maximum time
Here the minimum time after which a clot can be detected and the upper limit of clotting detection can be set.
1. The cursor blinks on the first entry position for “min. time”.
2. Enter the “min. time” with number keys or confirm and proceed.
3. Press to confirm the entry and jump to “max. time”.
4. Enter the “max. time” with number keys or confirm and proceed.
5. Press to confirm the entry.
You jump automatically to the “meas. time” display.
6.6.4.8 Set measuring time
Here the total measuring time, depending on the test method (here aPTT), can be set.
User Manual biobas 10
6 Operation 100
1. Enter the measuring time with number keys or confirm and proceed.
2. Press to confirm the entry.
3. Automatically the next display appears.
4. Press until “checking parameters please wait" appears,
then:
5. Press to save parameters or to leave the menu
without saving the changes.
Automatically STANDBY or the measuring mode is reached.
User Manual biobas 10
6 Operation 101
UTILITIES
Parameter access WITH PIN Code only
The menu UTILITIES includes a group of submenus in which analyzer settings can be adjusted.
A "PIN Code" (default PIN 11111) is necessary to enter. It can be changed user-specific later on in UTILLTIES submenu <PIN Code>.
Table 6-4 Overview UTILITIES Submenus
Selection Meaning
<Printer> Select: AUTO / ON / OFF and print parameter protocol
<Port A> Activates functions at the RS232-C.
<Port B> Sends measuring results to HOST if “ON”.
Enables BC-Reader at Port A
If “OFF” the measuring results will not be sent to the HOST
<Beeper> Select: ON / OFF / CLICK
<Language> Select analyzer language
<Date/Time> Entry of Date and Time
<PIN Code> Enter a personal PIN Code (secret number)
Entry range is from <00001> to <59999>
00000 = Off/ disable
<CUVETTE-Test> Verification of automatic cuvette detection.
< Param.im/export > Service: Load parameter from a SD-Card, Save parameter onto SD-Card
Info Shows name of analyzer, software version with corresponding date.
User Manual biobas 10
6 Operation 102
1. Switch on the analyzer.
2. In STANDBY press or until <UTILITIES> appears.
3. Press and enter the PIN Code (factory default: 11111).
NOTE
If the number was incorrect STANDBY display appears again. The following display appears, provided the correct PIN Code was entered.
Generally, after a SubMenu has been selected and adjusted,
respectively, you are asked to adjust more parameter or
change to the measuring mode.
According display will appear:
4. Press to access the measuring mode (cuvette in) or
press to set or verify more parameters.
User Manual biobas 10
6 Operation 103
SubMenu <Printer>
In this SubMenu you can select the print-out of different data, e.g. test results, method- or analyzer parameters as well as error messages via the external printer.
Table 6-5 Printer selections
Selection Meaning
<On> Printer is activated, Press for print-out.
<Auto> Print of results automatically, as soon as the test result is available.
<Parameter protocol>
Print of Method- and global analyzer parameter
<Actual Method>
Print of method parameter of a selected method
<All Parameter>
Print of all analyzer- and method parameter
<Global> Print of global analyzer parameter
<Off> Printer is deactivated
Printout e.g. <Parameter protocol> Actual method
Very often it is of interest, to have a look at the actual method parameters in order to verify the measuring results.
The easiest way is to make a print-out as described in the following.
1. Enter UTILITIES and press or to switch to <Printer>.
User Manual biobas 10
6 Operation 104
2. Press to enter the <Printer> menu.
3. Press or to select <Parameter-Protocol>.
4. Press to enter the <Parameter-Protocol> menu.
5. Press or to select <Actual Method>.
6. Press to print out the actual method.
NOTE
To select and activate the actual method for the print out, the warm-up time of the analyzer after “switch on” must be finished. Then, select the
required method, press and change via UTILITIES to print out the actual method.
User Manual biobas 10
6 Operation 105
SubMenu <Port A>
The analyzer model is equipped with a 6-pin RS 232-C interface at the backside of the analyzer (= Port A). This interface can be used for external data communication, to connect an external printer or barcode scanner.
Table 6-6 Port A - Selections
Selection Meaning
<Off> RS 232-C interface (Port A) is deactivated
<ext. printer> Serial output of data to external printer
1. Enter UTILITIES and press or to switch to
<Port A>.
2. Press to enter the Port A submenus.
3. Press or to select among following options:
4. Press to confirm and activate respective option.
5. Press to return to the measuring mode or press
to set or verify more parameters.
User Manual biobas 10
6 Operation 106
SubMenu <Port B> (virtual com port via USB)
To link the analyzer to a HOST see chapter 4.9, page 37 to connect the HOST to the analyzer.
NOTE
If not already made available by the HOST operating system a USB driver has to be installed previously on the HOST in order to use the USB port.
The analyzer offers an USB interface (USB Type B connector
/ Port B) at the backside. This interface will be acknowledged
as virtual COM-Port at the HOST side (see also chapter 4.9
Connection to a HOST (Port B), page 37.
The USB interface can only be used to connect the analyzer
to a HOST. It is not possible to connect an external printer.
Table 6-7 Port B - Selections
Selection Meaning
<OFF> Interface is switched off
<HOST>
Serial Port B is activated for HOST communication.
The analyzer changes into a state that a barcodescanner can be connected to the RS232 interface.
If a HOST is activated/connected to the analyzer the software requests a patient barcode number. The number input can either be carried out by entering the number manually (numerical data) or by a barcode scanner (alpha numerical).
User Manual biobas 10
6 Operation 107
If a LIS application is installed on your HOST, the following order is necessary.
1. Switch on the HOST and wait until it is booted.
2. Switch on the analyzer.
Wait approx. 30 sec. until the HOST has detected the USB
connection. If available and activated a sound indicates the
identification of the analyzer.
3. Then open the LIS or other software application at HOST start the user software.
If the analyzer has been switched off in the meantime and has to be switched on again, follow the described order.
1. Stop the user software (close the program).
2. Switch on the analyzer.
3. Wait until the computer has identified the analyzer (beeps if sound are “on”).
4. Then start the user software.
Options in the analyzer settings
1. Off
2. HOST – measured results to be send in a formatted string to HOST.
Output example general:
<STX> Analyzer; S/N; Date; Time; Patient ID; Method_Name; Channel No; Raw Value1; R_Unit1; Raw Value2; R_Unit2; 1st Conversion; 1_Unit; 2nd Conversion; 2_Unit; 3rd Conversion; 3_Unit; Flags; CRC;<CR><LF>
Output example for PT method with normal Plasma:
<STX>biobas 10; G1560001; dd:mm:yyyy, hh:mm:ss; 147; PT; 1; 12,3;s; 89,7;%; 1,06;Ratio; void;void; OK; CB90;<CR><LF>
Output example for aPTT method with pathological Plasma
<STX>biobas 10; G1560001; dd:mm:yyyy, hh:mm:ss; 1895; aPTT; 1; 88,2;s; void;void; 3,15;Ratio; void;void; OK; D334;<CR><LF>.
User Manual biobas 10
6 Operation 108
HOST output Flags
o M = Mixer Motor (measurement stops, critical error)
o T = Temp (warning)
o OK = no flag
Set respective interface parameters in the User software (see “Options to set the analyzer settings”, point 3) as described in the following:
1. Enter UTILITIES and press or to switch to
<Port B>.
2. Press to Port B submenus.
3. Press or to select among following options:
4. In this case select HOST and press to confirm.
Automatically the next display appears:
5. Press to set more parameters or to return to STANDBY and measuring mode.
User Manual biobas 10
6 Operation 109
The analyzer checks the parameters.
If parameters had been changed and saved, the following
message appears:
If parameters are unchanged the following message appears:
If parameters are different to the previous ones:
6. Press to save new parameters and continue with the measurments.
7. Press to return to the measuring mode without saving the changes.
Automatically the next display appears
At this point the measuring can be started.
User Manual biobas 10
6 Operation 110
SubMenu <Beeper>
The beeper acoustically confirms:
Key functions upon confirmation.
An incorrect process.
The finished sample incubation.
The coagulation recognition.
1. Enter UTILITIES and press or to switch to <Beeper>
2. Press to enter the Beeper submenus.
Table 6-8 Beeper selctions
Selection Meaning
<On> Activates the beeper.
Each action is confirmed by the beeper
<Off> Deactivates the beeper.
<Click> Only keystrokes will be confirmed by the beeper
User Manual biobas 10
6 Operation 111
SubMenu <Language>
This menu allows you to select different analyzer software languages.
The Default language is English.
Table 6-9 Language selections
Selection Meaning
<English> Set analyzer language to English
<German> Set analyzer language to German
<Español> Set analyzer language to Spanish
<French> Set analyzer language to French
<Portuguese> Set analyzer language to Portuguese
<Russian> Set analyzer language to Russian
1. Enter UTILITIES and press or to switch to <language>.
2. Press
3. Press or to select the required language.
User Manual biobas 10
6 Operation 112
4. Press to confirm the selected language.
5. Press to save the selection or to return without
changes.
After that:
6. Press to return to the measuring mode.
7. Press to set more parameters.
SubMenu <Date/Time>
In this menu you can set the actual date and time of the analyzer.
1. Enter UTILITIES and press or to switch to
<Date/Time>
2. Press
User Manual biobas 10
6 Operation 113
The cursor blinks on the first character, the format is fix
(dd.mm.yy, and hh:mm:ss)
3. Press respective number keys to enter date and time (from left
to right).
4. Press to jump to the next number/ character.
The cursor automatically blinks on next number/ character.
There is no need to enter a decimal point.
5. Proceed with the adjustments until the last entry is reached.
6. Press
8. Press to return to the measuring mode, the analyzer
checks the parameters .
9. Press to save or to return to the measuring mode.
The analyzer switches to the measuring mode of the last
method set.
User Manual biobas 10
6 Operation 114
SubMenu <PIN Code>
A PIN Code can be entered to prevent inadvertent parameter changes of non-authorized personnel.
Without entering a PIN Code it is possible to:
Perform all tests.
Select the test method.
Set / adjust the conversion parameter and reagent Lot Number.
With a PIN Code it is possible to:
Perform all tests.
Select the test method.
Have full access to all method parameters.
Set / adjust the conversion parameter and reagent Lot Number.
Have full access to all parameter in the menu <UTILITIES>.
This menu allows you to select a PIN Code (secret number) of
5 digits to access the parameter and UTILITIES menu.
Entering <00000> will disable the PIN request.
Entry range is from <00001> to <59999>.
After the PIN Code has been entered, the following display
appears:
User Manual biobas 10
6 Operation 115
The default PIN set by the manufacturer (11111) can be overwritten at this point.
1. Enter the PIN Code with number keys.
2. Press to confirm the entry, parameters are checked.
3. Press to confirm the entry or to return to the measuring mode.
If changes are saved a print out will follow if a printer is
connected and the printer-mode is set to <Auto>.
-----------------------------
<Analyzer name>
Ser. No. B1770711 Example print-out!
PIN No. =xxxxx x = Number
-----------------------------
NOTE
If 5 times a 0 “zero” is entered and saved, the PIN Code is not required to enter the parameter or UTILITIES menu.
In any case, the PIN Code must have 5 characters.
The analyzer switches automatically to STANDBY of the last
method set.
User Manual biobas 10
6 Operation 116
SubMenu <Cuvette-test>
This menu checks the function of the automatic cuvette recognition.
The thermal block has to be warmed up to operating
temperature.
1. Enter UTILITIES and press or to switch to <Cuvette-test>.
2. Press to select <Cuvette-test> and the following display appears:
3. Insert a cuvette into the measuring channel
The transmission threshold value is used to decide whether a
cuvette is inserted or the light path is free of a cuvette.
If a cuvette is inserted into the measuring channel, the display
shows “cuvette” in case of cuvette detection.
When removing and inserting the cuvette alternately the
display must show the appropriate status “cuvette” and “- - -“.
In case the function does not operate as described above:
1. Remove all cuvette from the measuring channel.
User Manual biobas 10
6 Operation 117
2. Switch OFF the analyzer.
3. Wait about 1 minute.
4. Switch ON the analyzer.
5. Repeat all the steps described above a second time.
If the CUVETTE-test is carried out but the thermal block has not reached its final temperature the following info appears:
Wait until the temperature of 37.4°C is reached and the
adjustment of the measuring channel is finished. Then carry
out the step again.
User Manual biobas 10
6 Operation 118
SubMenu <Parameter import>
It is possible to import reagent and method data by use of a SD-card.
An already defined method with the same name is overwritten automatically.
At the end of this procedure modifications can be saved or skipped.
All parameters can be imported from a file named PARIMP.TXT to the analyzer.
To execute this function a fat32 formatted SD-Card must be inserted into the analyzer SD-Card slot.
Formatting of the SD-Card is only possible with a special software because windows does not support the formatting process properly.
Please visit: www.sdcard.org/downloads
NOTE
Make sure the SD-Card does not carry any additional adhesive stickers this could harm the card reading unit.
The write protection of the SC-Card has to be unlocked otherwise no data can be exported.
IF the SD-Card is protected the analyzer shows a P-FLAG. In this case remove the SD-Card, unlock it and insert it again
User Manual biobas 10
6 Operation 119
Insert the SD-card
1. Insert a SD-Card into the card reading unit located on the rear
side of the analyzer.
2. Ensure that the SD-Card latches in the card reader.
3. Enter UTILITIES and press or to switch to
<Param. im/export>.
4. Press .
5. Select < Import <-SD-Card>.
6. Press .
The parameter import starts automatically with according
message.
The parameter file of the SD card is imported automatically.
Subsequently, the display shows the next submenu.
7. Press to return to STANDBY of the respective method.
The analyzer checks the parameters.
User Manual biobas 10
6 Operation 120
If parameters are unchanged the following message appears:
If parameters are different to the previous ones the display
shows.
8. Press to save new parameters and continue with the measurments.
9. Press to return to STANDBY without saving the changes.
Remove the SD-card
Slightly press the outer visible face of the SD-Card into the
analyzer.
The pressure causes the internal ejector mechanism to
release the SD-Card.
User Manual biobas 10
6 Operation 121
SubMenu <Parameter export>
All parameters can be exported to a file named PAREXP.TXT onto SD-Card.
To execute this function a fat32 formatted SD-Card must be inserted into the analyzer SD-Card slot.
Formatting is only possible with a special software because windows does not support the formatting process properly.
Please visit: www.sdcard.org/downloads
Another possibility is to use a camera to format the SD-Card. See corresponding User manual to carry out the formatting.
NOTE
Make sure the SD-Card does not carry any additional adhesive stickers this could harm the card reading unit.
The write protection of the SC-Card has to be unlocked otherwise no data can be exported.
If the SD-Card is protected the analyzer shows a P-FLAG. In this case remove the SD-Card, unlock it and insert it again
1. Insert a SD-Card into the card reading slot located on the back
side of the analyzer.
2. Ensure that the SD-Card is inserted and pushed into the card
reader as far as possible.
3. Enter UTILITIES and press or to switch to <Param. im/export>
4. Press .
5. Select < Export - > SD-Card>.
User Manual biobas 10
6 Operation 122
6. Press .
The parameter export starts automatically with according
message.
Export of parameter from the analyzer to the SD-Card.
The parameter file of the SD card is exported automatically.
Subsequently, the display shows the next submenu.
7. Press to return to STANDBY of the respective method.
The analyzer checks the parameters.
If parameters are unchanged the following message appears:
If parameters are different to the previous ones the display shows.
8. Press to save new parameters and continue with the measurments.
User Manual biobas 10
6 Operation 123
9. Press to return to STANDBY without saving the changes.
Remove the SD-card
Slightly press the outer visible face of the SD-Card into the
analyzer.
The pressure causes the internal ejector mechanism to
release the SD-Card.
SubMenu <Info>
This submenu has the purpose to show the user
Name of the analyzer
Current software version
Date of release of software version
Serial number of the analyzer
Parameter ID number
Number of NTC-T
Press or to switch between lines.
User Manual biobas 10
6 Operation 124
Sample print-outs PT and calibration
General print-outs
Once a method has been selected the programmed calibration curve parameters will be printed out followed by the results.
The print-out is done automatically as soon as a result has been obtained by the measuring channel.
Each method has its own counter. As soon as the analyzer has been switched on, the counter starts with "1".
This feature can be set in menu <Method-Parameter>, <Measurement>, <Printout No.>.
Print-out of all parameters
A print-out of all programmed test parameters can be generated as described in chapter 4.3 UTILITIES.
Print-out of method parameters
To generate a parameter print-out for the selected method, press “0” when the analyzer is in the measuring mode.
However, the printer must be set to <On> or <Auto> in the UTILITIES menu <Printer>.
User Manual biobas 10
6 Operation 125
Print-out of method PT (Example):
Conversions via a 4-point calibration curve in % and INR.
- method store 1 - `PT´ Date dd.mm.yyyy Meas time = 420 s Gain_idx = 0 Cuv in ON
Reagens_sens ON
Automated cuvette detection
Start Reagent: Lot = 1 Reagent-Lot. No. Volume = 100 ul Reagent Start Volume incubat. = 60 s Incubation time 1st convers
INTERPOLAT.
1. conversion, calib. curve / interpolation
97.0%= 11.5 s Calibration curve points 43.0%= 20.1 s 23.0%= 31.9 s 12.0%= 57.0 s 2nd convers INR 2. conversion INR ISI = 1.05 ISI-constant
Results: PT Method Patient
___________
Patient name
dd:mm:yyyy,
hh:mm:ss
Date, Time
No. = 1 Print.-out no. Channel = 1 Channel No. 1 Time = 12.0s Measuring time % = 90.4 Conversion to PT % INR = 1.06 Conversion to INR
User Manual biobas 10
7 Troubleshooting 126
7 TROUBLESHOOTING
Faults may caused by actions during use and/or by the system itself. The analyzer displays error messages and warnings in the display.If an external printer is attached and activated the information are printed out.
Application errors
Application errors may cause error messages. Possible causes are:
Air bubbles were created during pipetting.
Pipetting was performed directly into the measuring channel
without cuvette.
The wrong pipette tips were used.
The pipetted volume is incorrect (e.g. variable Pipets).
The pipetting process was too slow or the pipetting angle
incorrect.
The temperature of the start reagent deviates from 37,4°C.
Cuvettes are not prewarmed / too cold for plausible
measuring.
The sample, control or reagent is too old.
No mixer has been placed into the cuvette.
Reagent carry over (e.g. PT reagent).
A reagent with the wrong Lot Number has been used.
The reagent has not been used according to the package
insert.
The reagent used does not correspond to the method
selected.
Start reagent has not been mixed properly.
User Manual biobas 10
7 Troubleshooting 127
No or an incorrect calibration curve is available.
Errors appeared during the sample collection or centrifugation.
Method parameters relevant for measuring are incorrect.
Should any of these errors occur and they are recognized in time, solve them immediately.
Cancel incubation / measurement
If you realize wrong handling and want to interrupt the measurement stop the measurement with the following key combination.
Press and at the same time to stop the
measuring process on the respective channel.
User Manual biobas 10
7 Troubleshooting 128
Status messages CuvCARD (in alphabetical order)
Status message (display/print)
Cause Action
Balance and reserve fully used.
The cuvette balance and reserve quantity is now fully used, no further measurement possible
Insert new CuvCARD and load new balance to analyzer, once the cuvette balance and reserve quantity is fully used.
Balance fully used!
The cuvette balance on the analyzer is now fully used, no further measurements possible.
Insert new CuvCARD and load new balance to analyzer, once the cuvette balance is fully used. Make sure always having enough cuvettes on stock.
CuvCARD devalued!
Cuvette balance on CuvCARD is fully used, CuvCARD is empty!
Use new CuvCARD delivered with cuvettes boxes.
Make sure enough cuvettes are available.
Only xx measurements remaining.
Message shows the remaining quantity of measurements left to utilize.
Perform remaining number of measurements, when used, insert new CuvCARD and load new balance.
Make sure enough cuvettes are available.
User Manual biobas 10
7 Troubleshooting 129
Only xx reserve measurements remain possible.
Message provides information, that only a certain reserve quantity is left to utilize.
Insert new CuvCARD and load new balance to analyzer, once the reserve quantity is fully used.
Make sure enough cuvettes are available.
Only a few measurements remain possible.
Message provides information, that only a small reserve quantity is left to perform.
Insert new CuvCARD and load new balance to analyzer, once the cuvette balance is fully used.
Make sure enough cuvettes are available.
Please order new cuvettes.
Note; balance and reserve will run short soon.
Timely order new cuvettes to be able continuing measurements.
User Manual biobas 10
7 Troubleshooting 130
Error messages (in alphabetical order)
Error message (Display)
Cause Action
break time Possibly no clotting;
The maximum measuring time has been exceeded.
Optical test for clots; repeat test
break top Exceeded measuring range (too high) possibly caused by air bubbles
Repeat test
break bott(om) Exceeded measuring range (too low) preparation becomes turbid
Repeat test
break moto(r) Mixer motor error occurred
Contact technical service
break nois(e) Signal too noisy after sample adjustment
Check for air bubbles or other particles, e.g. mixer fallen out of the cuvette into the channel.
break drft (drift) Measured curve drifted after reagent has been added or start pipette error.
Check sample for air bubbles or if start pipette is activated in soft-ware but not linked to analyzer
break adj(ust) Light value is too low during adjustment phase (too little light
Check sample and try again
User Manual biobas 10
7 Troubleshooting 131
Error message (Display)
Cause Action
passes the channel)
break Measurement cancelled with
and
Intentionally done by user!
Bright value failed
Channel is not free, e.g. remaining cuvette, dust or other debris disturbs the light path
Remove the cuvette
Clean the channel
Then press ENTER to retry or ESC to go to the next channel
If necessary call the technical service.
Channel
locked!
A measuring channel is locked due to no remaining cuvette balance available.
Error during adjustment
Load new cuvette balance to analyzer.
Switch OFF the analyzer and ON again to repeat the adjustment.
CuvCard or ChipCARD
Checksum error!
Parameter checksum on inserted card type is incorrect
Contact technical service
CuvCard or ChipCARD
Read error!
Inserted card type may be defect.
Re-insert card and repeat process, otherwise contact technical service
User Manual biobas 10
7 Troubleshooting 132
Error message (Display)
Cause Action
illegal card! please remove it.
Utilized CuvCARD may not carry cuvette credit or is a wrong card.
Try another CuvCARD or contact technical service.
Invalid customer ID
No booking done!
Customer ID number on CuvCARD does not correspond to cuvette type used.
Contact distributor
Empty/invalid card
Inserted ChipCARD could not be read due to possible incorrect or no data being on the card.
Repeat process or use other ChipCARD. Contact distributor.
card write error. NOT booked!
An error occurred while writing data to card.
Contact technical service
Wrong cuvette type
No booking done!
Cuvette type on cuvette card does not correspond to cuvette type on analyzer.
Contact distributor
Warning: ChipCARD Error!
Read Error
ChipCARD is inserted upside down
Wrong blank ChipCARD has been inserted
Wrong Method ChipCARD has been inserted
Invalid version of ChipCARD has been inserted
Insert the ChipCARD with the correct direction
Insert correct blanko ChipCARD
Insert correct Method ChipCARD
User Manual biobas 10
7 Troubleshooting 133
Error message (Display)
Cause Action
WARNING: Please remove ChipCARD!
A card still remains in the card reading unit
Remove card and press ENTER to continue
WARNING: incompatible
method
Wrong ChipCARD with incompatible method data was inserted into the analyzer.
Use ChipCARD with method data compatible to the analyzer software.
SD-CARD is write-protected
The SD-Card is locked
Unlock the write-protection of the SC-Card
CD12 BOOTLOADER V1.0
Inserted SD-Card is locked after connecting the analyzer to the power = starting.
Remove the power supply cable, then remove the SD-Card, unlock the SD-Card, insert the SD-Card and attach the power supply cable to start the analyzer again.
Invalid method param
Method parameters entered are incorrect.
Check method parameters
New name cancelled
Method name was entered incorrectly.
Enter method name again.
No card! There is no ChipCARD or CuvCARD inserted in the card reader.
Insert correct Card to card reader.
No SD-Card detected.
SD-card could not be read.
Check SD-Card, perhaps use another card instead.
User Manual biobas 10
7 Troubleshooting 134
Error message (Display)
Cause Action
No SD-Card found.
While trying to access the SD-Card card was not found.
Insert SD-Card to analyzer.
Blinking P in the right upper corner of the display.
Write protection of the SD-Card is activated.
Remove the SD-Card Remove the SC-Card, unlock the write portection and insert the SC-Card again.
Parameter error!
set to defaults
While loading parameters an error occurred. System default parameters loaded instead.
Check parameters and re-enter if necessary.
SD-Card write-protected!
The write protection on the Memory Card has been activated.
Deactivate write protect on the Memory Card and repeat the operation.
SD-Card: Init. error!
Initialization error while writing data to a card.
Repeat process, perhaps use other card instead.
SD-Card: Root directory error!
SD-Card defective or incorrectly formatted.
Use other card or try to initialize.
Write Error! Data cannot be written to SD-card.
Card defect, wrong card used or not correctly inserted. Retry. Contact technical Service if
User Manual biobas 10
7 Troubleshooting 135
Error message (Display)
Cause Action
problem persists.
WARNING: HOST not connected
PC is not connected.
USB driver is not installed.
PC and analyzer are switched on in the wrong order.
1. Switch on the HOST / PC.
2. Switch on the analyzer.
3. Set the analyzer to STANDBY.
4. Install a USB driver.
Errors during operation
Error Cause Action
Cannot start analyzer
Power supply failure.
Loose power cable.
Mains power failure.
Check the function of the power supply.
If the power supply is ok but the analyzer does not work, contact your local distributor or the technical service.
Analyzer fails during operation
Power supply failure.
Loose power cable?
Mains power failure.
Check the function of the power supply.
If the power supply is ok but the analyzer does not work, contact your local distributor or the technical service
User Manual biobas 10
7 Troubleshooting 136
of the manufacturer.
Measuring cells contaminated with liquids
Additional pipetting of plasma or reagent into the measuring cell without cuvette.
Remove liquid with pipette, clean with appropriate absorbent cloth and disinfectant.
Warnings
Warning Meaning Action
Cool down This message will appear when it is determined while no measurements are running that the thermal block is too warm. No other measurements can be started during cool down.
Might occur due to direct sunlight, place the analyzer to another place with lower light input.
Contact technical service.
WARN: Temp. instabile
(TEMP WARN)
If during a measurement the temperature of the thermal block deviates significantly from the set value, the measurement is not cancelled. Instead this warning appears in the display and is also printed out via the active printer.
Might occur due to direct sunlight.
Place the analyzer to another place with lower light input.
Contact technical service
ERR over
(calculation overflow)
This message appears if a calculated measuring value cannot be
Check conversion
If the err over message is displayed check the
User Manual biobas 10
7 Troubleshooting 137
Warning Meaning Action
(in display and printout)
displayed (calculation overflow) because of size overflow.
The err over message can appear during:
Extrapolation of the
100 % calibration
curve value
Interpolation of a
measuring value
based on the
calibration curve
Calculation of the
coefficient of
variation.
parameterization of the calibration curve parameters
ERR div0 Measurement calculation detetced a negative value.
Check adjustments on Time Interpolation and Value interpolation.
Repeat measurement and check reference curve.
ERR no clot No clotting detected during measurement.
Check sample and method parameter settings, repeat measurement.
ERR log0 Measurement calculation detetced a negative value.
Check adjustments on Time Interpolation and Value interpolation
Repeat measurement and
User Manual biobas 10
7 Troubleshooting 138
Warning Meaning Action
check reference curve.
ERR bright-calib ch%u:
Bright calib failed!
An error has been detected during bright calibration.
Remove all cuvettes from measuring channel and repeat adjustment. Might occur due to direct sunlight, place the analyzer to another place with lower light input.
Close light protection cap.
Contact technical service if problem persists.
ERR dark-calib ch%u:
Dark calib failed!
An error has been detected during dark calibration.
Remove all cuvettes from measuring channel and repeat adjustment. Might occur due to direct sunlight, place the analyzer to another place with lower light input.
Close light protection cap.
Contact technical service is problem persists.
min-value reached
- no more points
or
max-value reached
During the input of the calibration curve, the range of values (%, g/l, mg/dl) is limited for each method due to factory settings. In addition the points must increase or
Check the Range for the calibration and enter the values again.
User Manual biobas 10
7 Troubleshooting 139
Warning Meaning Action
- no more points
decrease depending on the presetting. If the largest or smallest permitted value for a point has been entered, no additional points can be entered. Input will be blocked when this message appears.
User Manual biobas 10
8 Maintenance and Hygiene 140
8 MAINTENANCE AND HYGIENE
Safety
CAUTION
Do not use organic acids to clean the analyzer.
Do not spray or pour on any liquids, as they can
affect the correct function of the analyzer or
damage it.
Keep the analyzer free from dust and liquid
spillages. When not used over a longer period
of time, place a dust cover over the analyzer or
place it into a cabinet.
If liquid has been spilled on the analyzer.
Remove the contamination with a clean,
absorbent non-woven cloth considering all
applicable hygienic requirements.
If liquid has accidentally entered, or been pipetted into a measuring channel:
Remove the liquid with a pipette, then clean the
measuring channel with a lint-free cloth.
Consider all applicable hygienic requirements!
Contact our technical service if subsequent control measurements do not produce the expected result.
NOTE
The analyzer is fitted with a lithium battery type Li-Mn CR 2430 (life approx. 5 years). It should be replaced by an authorised service operator after 5 years at the latest. Otherwise a faultless operation can not be guaranteed.
BIOLOGICAL RISK
The light protection cap is considered to be potentially contaminated. Therefore, the manufacturer recommends to replace it once a year. Contact your local distributor for further information.
User Manual biobas 10
8 Maintenance and Hygiene 141
With the application of different reagents, and here especially reagents containing thrombin, there is a risk of reagent carry-over.
If liquids or dried-on remnants can be seen on the rim of the opening of the light protection cap, remove them with laboratory disinfectant solution and lint-free cotton buds.
Operating and auxiliary materials
Disinfectant
96% Ethanol, denatured.
To clean and disinfect the analyzer (Thermal block, working
area)
Hand protection
Disposable nitrile rubber gloves.
Avoid any hand contact with potentially infectious material.
Eye protection
Safety eye glasses.
Splashes are likely to occur, there is a risk of infection.
Other
Lint-free cloth.
User Manual biobas 10
9 Transport 142
9 TRANSPORT
If you should have thrown away the packaging material, please contact the manufacturer, see chapter 1.1, page 8.
See also chapter 4.3 for further information.
Prepare the analyzer for packaging
NOTE
Please clean and disinfect the analyzer at first.
Take care when moving the analyzer.
Secure the analyzer against slipping and tilting.
There are no special signs regarding the transport of the analyzer.
1. Remove cuvettes incl. mixer.
2. Remove all vials and adapters (if inserted).
NOTE
Do not leave any loose parts inside the analyzer!
User Manual biobas 10
9 Transport 143
Pack the analyzer for shipment
1. Fold the flat card box, insert the analyzer and secure it as shown in the following pictures.
User Manual biobas 10
9 Transport 144
2. Open the box 3. Insert film-card
secured analyzer onto the botton of the box.
4. Take power supply plug and the accessory box and put them into the box.
5. Turn down the side flaps.
6. Close the cover of the box and seal it with packaging tape.
User Manual biobas 10
10 Service Information Sheet (Customer Complaint) 145
10 SERVICE INFORMATION SHEET (CUSTOMER
COMPLAINT)
NOTE
Please request a Return Authorization Number prior to any return shipment.
For proper handling of a complaint please make sure the following information is available.
AnalyzerType (Name Plate)
Serial Number of the analyzer (Name Plate)
Model (Name Plate)
Problem Description
If available a picture of the defective part
Furthermore, we recommend to keep a service log book to document maintenance work with information about date, reason and operator/ responsible person.
This makes the life cycle of the analyzer clearly visible and traceable.
User Manual biobas 10
11 Decommissioning and storage 146
11 DECOMMISSIONING AND STORAGE
If you intend not to use the analyzer over a longer period of time follow the instructions as described.
CAUTION
1. Remove the external power supply from the socket.
2. Remove the power supply from the analyzer.
3. Remove all loose items as cuvettes or vials from the analyzer.
4. Dispose vials with reagents and/or plasma according to given disposal specifications.
5. Clean / desinfect the analyzer surface properly with Ethanol (96%, denatured).
6. Let the analyzer dry for a certain time to get rid of any possible remaining liquid.
7. Insert red cuvettes (the same when delivered) into the measuring channel as dust protection.
8. Put the analyzer into a film bag.
9. Put the external power supply and power cable into a film bag.
10. Store the analyzer and additional items at a safe, dry place.
User Manual biobas 10
12 Disposal of the Analyzer 147
12 DISPOSAL OF THE ANALYZER
CAUTION
For the safety of the operating personnel make sure the analyzer has been disinfected before disposal.
All parts parts must be disposed of in accordance with national
directives for disposal.
Do not dismount and recylcle the analyzer by yourself
For disposal send the analyzer to a registered waste disposal
authority or contact your local distributor.
1. Remove the power supply from the socket and remove the power supply cable from the analyzer.
2. Remove all loose items, e.g. cuvettes or vials from the analyzer.
3. Clean / disinfect the analyzer properly with Ethanol (96%, denatured).
4. Let the analyzer dry for a certain time to get rid of any possible remaining liquid.
5. Put the analyzer into a film bag.
6. Put the power supply plug and power supply cable into a film bag.
7. Prepare the analyzer for transport as described in Chapter 9 Transport, page 142.
Do not dismount and recylcle the analyzer by yourself.
All parts parts must be disposed of in accordance with
national directives for disposal
For disposal send the analyzer to a registered waste
disposal authority.
User Manual biobas 10
13 Appendix 148
13 APPENDIX
Technical data
System related
Characteristics Description
Type of analyzer Semi-automated 1-channel in-vitro diagnostic analyzer for coagulation
Application Clotting tests
Operation Described in this User Manual
Measuring principle Turbodensitometric
Sensitivity ~ 1 g/l Fibrinogen
Test throughput ~30/h PT, ~15/h aPTT
Cuvette volume min. 150 µl, max. 200 µl (suspension)
Calibration Manual input of calibration points, method dependent
Software Stored in memory
Programmed methods PT, in sec, %, Ratio, INR (combinations)
aPTT, in sec, and Ratio
Fibrinogen, in sec, g/l and mg/dl
Thrombin Time, in sec
Intr. Factor, in %, Extr. Factor, in %
Light source LED approx. 875 nm
Display Liquid crystal display, 2 lines with 20 characters each
Thermal block Temperature controlled at 37.4°C +/- 0.4°C
Measuring channel 1
User Manual biobas 10
13 Appendix 149
Light protection caps: 1 (designed for yellow pipette tips by Eppendorf)
Reagent positions 1 á 22 mm diameter
Cuvette positions 4
Measuring time max. 420 sec
Interfaces RS232-C, USB-B, SD-Card, Card reader
System time Real time clock for time and date
Dimensions Approx. 160 x 210 x 80 mm (LxWxH)
Weight Approx. 0.80 kg
Electrical data
External power supply characteristics
Specification
Input voltage 100 – 240 VAC
Rated input current Approx. 0.4 A
Rated input frequency 50-60 Hz +/- 3Hz
Output voltage +12 VDC
Output current Max. 1.25 A
Power consumption Max. 15 VA
Analyzer characteristics Specification
Input voltage 12 VDC
Max. input current Max 0.8 A
Overvoltage category I
User Manual biobas 10
13 Appendix 150
Pollution degree 2
Means of protection Class III
Lithium battery type Li-Mn CR 2430
+3V
Environmental requirements
Parameter Specification
Installation site Inner rooms only, dry environment
Altitude 2.000 meters
Mains voltage fluctuation
Max. 10% of rated voltage
Temperature During operation; +15°C - +30°C
During transport and storage; +2°C - +50°C
Humidity 10% - < 85%, non-condensing
Disposables
Type Quantity Description
Cuvettes
5 x 100 Micro Cuvettes / Dispo-System with Mixer 1,0x4,0 mm
1x CuvCARD
1 x 500 Micro Cuvettes / Mixer 1,0 x 4,0 mm in plastic bag
1x CuvCARD
User Manual biobas 10
13 Appendix 151
Materials supplied
Quantity Description
1 x biobas 10
1 x 10 single cuvettes with mixer in plastic bag
1 x User manual on USB stick
1 x External power supply.
Optional equipment
Barcode scanner
External printer
USB driver (if not already available at HOST)
Cable for printer and scanner
Supplier documentation
Barcode scanner instruction
External printer instruction
USB driver information