Use USB_DAQ to Complete the Calibration of PGA308 EVM. Iven Xu Mar. 13 th , 2013 Thanks to Ian Williams. Agenda. USB_DAQ and PGA308 EVM overview Calibration principle of PGA308 EVM Three output mode of PGA308 EVM 4-wire voltage output mode 3-wire voltage output mode Current output mode - PowerPoint PPT Presentation
Citation preview
Use USB_DAQ to Complete the Calibration of PGA308 EVM
Iven Xu
Mar. 13th, 2013
Thanks to Ian Williams
1
Agenda
• USB_DAQ and PGA308 EVM overview
• Calibration principle of PGA308 EVM
• Three output mode of PGA308 EVM– 4-wire voltage output mode
– 3-wire voltage output mode
– Current output mode
• Multi-Calibration principle of PGA308
2
3
a) 4-wire voltage output mode
a) 4-wire voltage output mode:
• USB_DAQ Board Default Configuration:
4
a) 4-wire voltage output mode:
• PGA308 EVM Default Configuration:
• T1 and T2 of USB_DAQ need not
to connect anything.
• Notice only in current output mode,
we should connect the +-15V to T2
and 24V to T1.
5
a) 4-wire voltage output mode:• Firstly, we should make sure we have selected “Run out of RAM” in “Block Diagram”:
• Then, move to “Calibration”. Make sure the step 1 ~ step 3 as the same of below fig:
• Thirdly, click “select Cal Preset” in step 4 and select “pre_vout_4p096.csv” for voltage output.
• Then, click “Calibrate” in step 6 and follow the prompted window to change the output of sensor.
6
• About the OTP Program, note that you only have 4 times. So before your calibration is completed, do not use OTP program.
7
a) 4-wire voltage output mode:
a) 4-wire voltage output mode:
8
9
b) 3-wire voltage output mode
b) 3-wire voltage output mode:
• USB_DAQ Board Default Configuration:
10
b) 3-wire voltage output mode:
• PGA308 EVM Default Configuration:
11
b) 3-wire voltage output mode:
• Pre-cal file : pre_OW_to_vo_4p096.csv
12
13
c) Current output mode
c) Current output mode:
14
• USB DAQ Configuration
c) Current output mode:• PGA308 EVM Configuration
* When we want to use current output mode of XTR117: * we should make sure the configuration of
PGA308 is in 4-wire voltage output mode; * Because in 3-wire voltage output mode of
PGA308, we should turn on/off the Vs of PGA308. But in 3-wire mode, the one-wire
and Vout is connect in one junction, so we
cannot use one-wire to control the Vs (turn on/off). * conclusion:
is we should use 4-wire voltage output
for current output.
15
c) Current output mode:
16
c) Current output mode:• Firstly, we should make sure we have selected “Run out of RAM” in “Block Diagram”:
• Then, move to “Calibration”. Make sure the step 1 ~ step 3 as the same of below fig:
• Thirdly, click “select Cal Preset” in step 4 and select “pre_Iout_4p096.csv” for current output.
• Then, click “Calibrate” in step 6 and follow the prompted window to change the output of sensor.
17
c) Current output mode:
• We can use
the Graph to
monitor the
output.
Also, we can
use Multi-meter
to monitor the
output.
18
• About the OTP Program, note that you only have 4 times. So before your calibration is completed, do not use OTP program.
19
c) Current output mode:
Multi-Calibration principle of PGA308:
• Need to update later.
20
Thanks!
21
How to create the model file:• (1), click the “Load Preset From….”, select a exist file like
“pre_vout_4p096.csv”
22
How to create the model file:• (2), then you can change the value as you want, for example we change
the Vref (V) from 4.096 to 4.
23
How to create the model file:• (3), then click the “Save Preset to File”, write a name like “pre_cal.csv”,
then click “Save”.
24
How to create the model file:• (4), Then re-click the “Load Preset From…”, select the “pre_cal.csv” we
saved in the last step (3):
25
How to create the model file:
• (5), then the value should be the same as you just changed in step (2) (we changed Vref from 4.096 to 4).
• (6), then click the “OK”, now finished the preset file.
• (7), then you can do the calibration in “step 6” of
PGA308 software.