    PROCESS REFRACTOMETER

    UR24

    Technical Datasheetn 90034H3741

    Page 1 of 2

    TECHNI CAL SPECIF ICATIONS AND STANDARDS

    Ambient characteristics

    Temperature Limits:

    Ambient:

    -10...+45 C (14113 F).

    Storage: -20...+70 C (-4+158 F).

    Humidity Limits:Ambient: 5%...95% (R.H. non-condensing).Storage: 5%...95% (R.H. non-condensing).

    Altitude Limits:

  • 8/12/2019 ur24-3741-tds-gb

    2/2

    PROCESS REFRACTOMETER

    UR24

    Technical Datasheetn 90034H3741

    Page 2 of 2

    InterfacesAnalog: 0...20mA or 4...20mA into 470

    Digital: -RS422/485 for connection to Repeater, PC orAutomation Systems.

    -N1 PROFIBUS by means of 1 Module (optional)for Profibus-DP network with DC 24V powersupply.

    Inputs: N1 "Line Stop" input (HOLD).

    Outputs: N1 relay output for alarm signal or control ofCleaning System, with contact rating of DC/AC24V/500mA.

    All the interfaces are optically decoupled from the power supply(VDE0160) and completely configurable from the keypad. Allconnections to be made by means of round 16 pin metallicconnector.

    CONSTRUCTION FEATURES

    Execution: One-piece enclosure with 316 AISI stainlesssteel cover, Thermal isolation flange inPEEK, 316 AISI stainless steel prism

    holder with 3" Tri-Clamp BS 4825 ASME-

    BPE or Varivent type N connection forinstallation on process line. Tank mountingversion (LP) features attachment with 70 mmoffset.

    Measure. Section: - Measurement prism in Synthetic Sapphireor Optical Glass (Crown N-K5, BariumLight Flint BaLF4, Borosilicate Crown N-BK10).

    - Electronically compensated LED lightsource.

    - CCD sensor element.- Pt100 temperature sensor to install in line

    or internally to unit for tank applications.

    The optical section of the unit is dehumidified by means of aMolecular Sieve desiccant cartridge.

    Electronic Section:- Microprocessor CPU.- Measurement readings, program menus and

    error messages presented on a 128x64 pointbacklit graphic LCD display with LCDSaving function.

    - Moulded keypad in scratchproof polyesterwith dome keys.

    - Choice of 5 interface languages (English,German, French, Italian, Spanish) for thedisplay of menus and messages.

    - Temperature readings in C or F.

    Product contact materials:- Pt100 and Prism Holder in 316 AISI stainless

    steel (Hastelloy, Incoalloy on request).- O-rings and Seals in Kalrez 6230 + Viton or

    EPDM (Silicone, Kalrez 6375 or 6380 on

    request).- Measurement prism in Synthetic Sapphire

    or Optical Glass (Crown N-K5, BariumLight Flint BaLF4, Borosilicate Crown N-BK10).

    Dimensions and weight:

    Standard Version 176 (w) x 192.5 (h) x 132.5 (d), 3.3 kg

    LP and SA Version 176 (w) x 192.5 (h) x 214.5 (d), 5.0 kg

    ACCESSORIES

    - Mounting fittings for in-line or by-pass installation, made of316 AISI stainless steel (other materials on request) with

    Tri-Clamp connections, weldable, flanged, to DINstandards and pipelines from 1" to 4".

    - Adapter flanges in 316 AISI stainless steel (other materials onrequest) for mounting on tanks or on angle fittings.

    UR24 Standard Accuracy version UR24 High Accuracy (HA) version UR24 Super Accuracy (SA) version

    Scale Range

    Prism in Sapphire: 0...95 Brix

    Prism in Bal f : 080 Brix(0 Brix with 40 C max. -80 Brix with 20 C min.)

    Prism in Sapph ire HR: 30100 Brix

    Prism in Crown: 050 Brix

    Prism in Sapphire: 0...95 Brix

    Prism in Bal f : 080 Brix(0 Brix with 40 C max. -80 Brix with 20 C min.)

    Prism in Crown: 050 Brix

    Prism in NB-K10: 025 Brix

    Scale RangeSpan

    Prism in Sapphire :min. 30 - max. 95 Brix

    Prism in Bal f :min. 30 - max. 80 Brix

    Prism in Sapph ire HR:min. 30 - max. 70 Brix

    Prism in Crown:min. 30 - max. 50 Brix

    Prism in Sapphire:min. 15 - max. 95 Brix

    Prism in Bal f :min. 15 max. 80 Brix

    Prism in Crown:min. 15 max. 50 Brix

    Prism in NB-K10: 25 Brix

    Accuracy(the valuesprovided refer tostandard sucrosesolutions)

    0.5% of Scale Range / max. 0.15 Brix 0.3% of Scale Range / max. 0.05 Brix

    0.1% of Scale Range / max. 0.02 Brixvalid for maximum product and/orambient temperature variations

    of 10 C (18 F)

    Temperature -5...+105 C (23...221 F)

    LP version for temperature up to140 C (284 F)

    equal to the Standard Accuracy version -5+95 C (23203 F)

    Maximuntemperatureduring sanitization

    125 C (257 F) x 30 minutes

    145 C (293 F) x 30 minutesin the LP version

    equal to the Standard Accuracy version 145 C (293 F) x 30 minutes

    Pressure max. 10 bar (145 psi) at 20 C (68 F)

    max. 8 bar (116 psi) at 100 C (212 F)equal to the Standard Accuracy version equal to the Standard Accuracy version