INTRODUCTION/TABLEOFCONTENTS StepOne
TheLVU2700Series isageneralpurposeultrasonic leveltransmitterthatprovidesa looppowered420mAoutput.The420mAoutputcanbeusedtoprovidetheproportionallevelofliquidinanytankorvessel.Thesignalcanbeconnectedtoanydevicethatacceptslooppowered420mAsignals,suchasaPLC,SCADA,DCS,display,controller,etc.
NewFeatures Easilychangeoutputfromdistancetovolume Selectfrom6standardtankshapes 16pointstrappingtableforcustomer/oddsizetanks AllwithinthesimpleconfigurationPCSoftwareProgram(LVCN414SW)
TableofContentsSpecifications/Dimensions..............................................................................................................................4SafetyPrecautions............................................................................................................................................5Components ...................................................................................................................................................6GettingStarted..................................................................................................................................................7 USBFobInterface...............................................................................................................................7 LVCN414SWSoftware..........................................................................................................................8 1.Configuration........................................................................................................................9 2.TankShapeSelection..........................................................................................................11 DimensionalEntryVerticalCylinderExample.........................................................12 DimensionalEntryHorizontalCylinderwithEndcapsExample...............................14 3.LevelConfiguration.............................................................................................................16 4.WritetoUnit.......................................................................................................................16Wiring .................................................................................................................................................17 WiringDiagram&WireConnections.................................................................................................17 WiringtoDisplays,Controllers&PLCs............................................................................................19Installation .................................................................................................................................................20 MountingGuide..................................................................................................................................20 FittingSelection..................................................................................................................................21 TankAdapters&Risers:..........................................................................................................21 Flanges&SideMountFittings................................................................................................22 StandPipe...............................................................................................................................23AdvancedFeature...........................................................................................................................................24Appendix .................................................................................................................................................25 SampleProgram.................................................................................................................................25 UpdatingSoftware..............................................................................................................................26 UpdatingTransmitterFirmware.........................................................................................................27 StrappingTable...................................................................................................................................28 Linervs.NonLinear...........................................................................................................................29 FactoryDefaults..................................................................................................................................31 DemoPage.........................................................................................................................................32 Troubleshooting..................................................................................................................................33 UserSettings.......................................................................................................................................34
4
SPECIFICATIONS/DIMENSIONS StepTwo
Range:Accuracy: 0.2%ofrangeResolution: LVU2710:0.019(0.5mm) LVU2718:0.039(1mm) LVU2726:0.079(2mm) LVU2726:0.079(2mm)Deadband: LVU2710:4(10cm) LVU2718/26/32:8(20cm)Beamwidth: LVU2710:2(5cm) LVU2718/26/32:3(7.6cm)Configuration: LVCN414SW(software)PC WindowsUSB2.0Memory: NonvolatileSupplyvoltage: 24VDC(loop)Consumption: 0.5WLoopresistance: 500Ohms@24VDCSignaloutput: 420mA,twowireSignalinvert: 420mAor204mASignalfailsafe: 4mA,20mA,21mA, 22mAorholdlast
Processtemp.: F:4to140 C:20to60Temp.comp.: AutomaticAmbienttemp.: F:31to140 C:35to60Pressure: MWP=30PSIEnclosuretype: Type6Pencapsulated,corrosion
resistance&submersibleEncl.material: PolycarbonateTrans.material: PVDFCablejacketmat.:PolyurethaneCabletype: 4conductorshieldedCablelength: 10(3m)Processmount: LVU2710:1NPT(1G) LVU2718/26/32:2NPT(2G)Mount.gasket: FKMClassification: GeneralpurposeCompliance: CE,RoHS
Dimensions:
SideView/LVU2710Series
SideView/LVU2718,LVU2726andLVU2732Series
LVU2710:4"to8'
(10cmto3m)
LVU2718:8"to9.8'
(20cmto5.5m)
LVU2726:8"to26.2'
(20cmto8m)
LVU2732:8"to32.8'
(20cmto10m)
5
SAFETYPRECAUTIONS StepThree
AboutthisManual:PLEASEREADTHEENTIREQUICKSTARTPRIORTOINSTALLINGORUSINGTHISPRODUCT.This manual includes information on the LVU2700 Series Ultrasonic Level Transmitter from OMEGAENGINEERING. Please refer to the part number located on the switch label to verify the exact modelconfiguration,whichyouhavepurchased.
Users Responsibility for Safety: OMEGA ENGINEERING manufactures a broad range of level sensingtechnologies.Whileeachof these sensors isdesigned tooperate inawidevarietyofapplications, it is theusersresponsibilitytoselectasensormodelthatisappropriatefortheapplication,installitproperly,performtests of the installed system, andmaintain all components. The failure to do so could result in propertydamageorseriousinjury.
ProperInstallationandHandling:Onlyprofessionalstaffshouldinstalland/orrepairthisproduct.Installthetransmitterwith the includedFKMgasketandneverovertighten thetransmitterwithin the fitting. Alwayscheckforleakspriortosystemstartup.
WiringandElectrical:Asupplyvoltageof12to28VDCisusedtopowertheLVU2700Series.Electricalwiringofthetransmittershouldbeperformedinaccordancewithallapplicablenational,state,andlocalcodes.
Material Compatibility: The enclosure is made of Polycarbonate (PC). The transducer is made ofPolyvinylideneFluoride(PVDF).Makesurethatthemodel,whichyouhaveselected,ischemicallycompatiblewiththeapplicationmedia.
Enclosure: Whilethetransmitterhousing isencapsulated,corrosionresistance&submersible,theLVU2700Series is not designed to be operationalwhen immersed. It should bemounted in such away that theenclosure and transducer do not come into contactwith the applicationmedia under normal operationalconditions.
Safety Installationshouldbedonebyproperlytrainedstaff Supplyvoltageshouldneverexceedamaximumof28VDC Makesurethesensorischemicallycompatiblewithyourapplication Designafailsafesystemthataccommodatesthepossibilityofsensorand/orpowerfailure. Thissensorshouldnotbeusedinclassifiedhazardousenvironments
MakeaFailSafeSystem: Designafailsafesystemthataccommodatesthepossibilityoftransmitterand/orpower failure. OMEGA ENGINEERING recommends the use of redundant backup systems and alarms inadditiontotheprimarysystem.
Flammable,ExplosiveorHazardousApplications:LVU2700Seriesshouldnotbeusedwithinclassifiedhazardousenvironments.
Warning: Alwaysuse theFKMgasketwhen installing theLVU2700Series,andmakesure thatallelectricalwiringoftheswitchisinaccordancewithapplicablecodes.
6
Components StepFour
LVU2700Seriesisofferedinfourdifferentmodels.Dependingonthemodelpurchased,youmayormaynothavebeenshippedallthecomponentsshownbelow. Youdohowever,needanLVU2700Series,USBFobandFKMgaskettoconfigure,installandoperateLVU2700Series.
LVU2700Series
PartNumber Range Enclosure MountingThread USBFob
LVU2710LVU2710B 9.8
(3m)
Type6Penclosure
1NPT IncludedNoFobLVU2710GLVU2710GB 1G
IncludedNoFob
LVU2718LVU2718B 18.0
(5.5m)
2NPT IncludedNoFobLVU2718GLVU2718GB 2G
IncludedNoFob
LVU2726LVU2726B 26.2
(8m)
2NPT IncludedNoFobLVU2726GLVU2726GB 2G
IncludedNoFob
LVU2732LVU2732B 32.8
(10m)
2NPT IncludedNoFobLVU2732GLVU2732GB 2G
IncludedNoFob
FKMGasketo Part#200128forLVU2710serieso Part#200129forLVU2718,LVU2726orLVU2732series
USBFobo Part#LVCN414USB
Manual
7
GETTINGSTARTED StepFive
LVU2700 Series is configured through a PC software program (LVCN414SW). The LVCN414SW is a freedownload fromOmega Engineeringswebsite. Youmust download an