Training Manual: Training Manual: KI9600A Series Pocket Power MetersKI9600A Series Pocket Power Meters
Level 1Level 1
June 2011
2
Course ContentsCourse Contents1. General Features2. Models3. Detector Types4. Keypad5. Instrument Care6. Operation7. Operating Modes8. Log / Linear display9. Special Features
3
1. General Features1. General Features
Battery life – 300 hours Calibration at multiple wavelengths Ruggedised, water & dust resistant 3 year re-calibration interval 3 year warranty – extendable to 5 or
7 years Special Features: ‘Multi-fiber ID’ ,
SlowMode and TamperLock
Entry skill level Fits in shirt pocket Test tone detection Industry standard
interchangeable connectors – Including SFF
4
There are two model styles:
A. KI9600A: for popular connector types – the most popular.
B. KI9600-XL: fitted with a wide area detector - plastic fibre, MT-RJ, ribbon and specialty fibre applications up to 3 mm core diameter.
2. Models2. Models
5
3. Detector Types3. Detector Types
Several Detector types are available to suit your application. Choose a detector style to suit your power and wavelength requirements.
H series High power – CATV, DWDM• Power level configurations of up to +24 dBm• Filtered InGaAs
Si Wide area detector configuration - LAN at 850nm, POF
• Power levels of up to 0 dBm•Will not work at 1300 nm• Available XL series meters only
InGaAs Telco & LAN• Power Levels of up to +5 dBm•Most accurate detector type at Telco wavelengths
Ge LAN at 850 & 1300nm• Power levels of up to +10 dBm• Not as accurate as InGaAs.• Can be affected by temperature• Not recommended above 1550 nm
6
3.1 Typical Detector Spectral responses3.1 Typical Detector Spectral responses
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
1.1
1.2
600 700 800 900 1000 1100 1200 1300 1400 1500 1600 1700
Rela
tive R
esp
onsi
vity A
/W
Wavelength nm
Power Meter Wavelength Dependence
Si
Early InGaAs Meters
Ge 25 ºC
Ge 0 ºC
InGaAs B series
New InGaAs detectors work well at 850 nm.
7
4. Keypad4. Keypad
Power On/Off
Connector release
Wavelength toggle and
special features
DUAL FUNCTION
dBm, dBr & W+
SET REF
DUAL FUNCTION
Hold+
Min/Max
Connector housing cap
8
5. Instrument Care5. Instrument Care
Keep the instrument in a carry case during storage and transport • Use only high quality batteries.
• For prolonged storage remove batteries. • The instrument is resistant to normal dust and moisture, however it is not
waterproof. • If moisture gets into the instrument, remove batteries & dry it out carefully
before using it again.
• Where possible, keep instrument away from strong sunlight.
• Clean the instrument case using Iso-Propyl-Alcohol (IPA) or other non solvent cleaning agents.
• DO NOT use Acetone or other active solvents as damage may result.
9
6. Operation6. Operation
1. Fit batteries
2. Turn power meter on
3. Select / determine test cord configuration
4. Fit correct adaptor style to meter
5. Measure
10
6.1 Fit batteries6.1 Fit batteries
To Fit Batteries.• Press battery cover latch down and push away from case.• Insert 2 ‘AAA’ cells• Replace battery cover
Battery life.Alkaline ‘AAA’ batteries: 300 Hrs
Low Battery Display.Indicator shows when batteries are low.
Warning.Do not use lithium batteries, or other batteries with a nominal voltage greater
than 1.8 V as the instrument may become damaged.
dB/ dBm/mW
HOLD
Max MinSet Ref
Pocket Fiber Meter
1550-0.59 dB m
Hold
nm
dB/ dBm/mW
HOLD
Max MinSet Ref
Pocket Fiber Meter
onPerm
11
6.2 Power On6.2 Power On
To turn onPush the Aqua [On/Off] button
To Defeat auto time out. Press [On/Off] & [HOLD] for several seconds. •Instrument will beep twice.• ‘on’ will display. •‘Perm’ will display on the upper part of the LCD.
• 10 minutes auto off• Permanent operation• Low battery indicator
12
6.3 Determine type of test cord6.3 Determine type of test cord
Identify fibre type required for test cord and configuration required.
Connector / boot colours may identify the fibre type:-OM1 (62.5 µm) BeigeOM2 (50 µm) BlackOM3 (50 µm) AquaSMF Blue or Green
Cordage colours may identify the fibre type:-SMF: YellowMMF Blue, Grey/Slate, Aqua or Orange
Unlike sources, power meters accept both PC and APC connectors.•EXCEPTION – wavelength selective meters are end face specific!
13
6.4 Fit connector6.4 Fit connector
Once the required test cordage configuration is identified, a matching adaptor for the power meter can be selected.
Fitting the adaptor depends upon meter style
14
Standard & Wavelength Selective InstrumentsStandard & Wavelength Selective Instruments
Supplied with :- SC connector as standard
Other connector styles include:-ST, FC, LC, E2000, 2.5mm Universal , SMA , MU, D4, DIN
* Universal not recommended for sources
15
To fit / change an adaptor / connectorTo fit / change an adaptor / connector
• Locate quick release button on front of instrument at base of connector housing
• Push and hold button in• Pull out existing adaptor• Fit new adaptor
Slot to outside
16
XL Series ConnectorsXL Series Connectors
• Industry standard 7/8”-28 TPI adaptors• Select required adaptor style• Screw on adaptor
17
7. Operating Modes7. Operating Modes
1. Measure Power:If receiving light from a ‘simple’ source.
– Meter will display power. – dBm, dBr or W– The operator must select correct
wavelength.
2. Modulation Frequency:Automatic display of modulation frequency.
– Modulation 200 to 2,500 Hz
3. Multi-fibre Identification number:
Automatic display of 12 fibre identification numbers.
Connect power meter to DUT.In the presence of light the meter will operate in one of three ways:
dB/ dBm/mW
HOLD
Max MinSet Ref
Pocket Fiber Meter
1550-0.59 dB m
Hold
nm
18
7.1 Measure Power7.1 Measure Power
Power meter receives light from the source.User must select the correct wavelength.
To select wavelength.•Toggle [λ] or [λ] to select λ
– Wavelength toggling is circular.
[<λ] [>λ]
19
7.2 Test Tone7.2 Test Tone
When receiving modulated light. •Instrument displays modulated frequency & beeps.
When receiving Test Tone.•270, 1,000 or 2,000 Hz.•Instrument displays modulated frequency & beeps.
Tone measurement can be disabled in Slow Mode.
850270Hz
Hold
nm
dB/ dBm/mW
HOLD
Max MinSet Ref
Pocket Fiber Meter
7.3 Fibre ID Number7.3 Fibre ID Number
When receiving a Multi-fibre identification signal:•Displays Fibre identification number
• Applies to KI9600 Firmware version r0.05 and above.• Legacy instruments will display a modulation
frequency.
20
Some Light sources have the ability to send a ‘multi-fibre identification’ number in the range of 1 to 12.
dB/ dBm/mW
HOLD
Max MinSet Ref
Pocket Fiber Meter
Perm 1550
I d04Hold
nm
LEVELMOD
Pocket Fiber Source
1550dB m
nmSource
0.00 I d04
21
8. Log/linear Display8. Log/linear Display
1. Absolute ModeMeasure actual power level at a particular location – dBm
– Typically used during installation & maintenance activities.
2. Relative ModeMeasure power level ‘relative’ to a particular location – dBr
– Typically used for loss measurement.
3. LinearMeasures power in watts – nW, µW, W
– Typically a laboratory measurement.
The meter defaults to last setting and wavelength at power on.
The power meter can display power in logarithmic (dB) or Linear (W) units.
HOLD
Max Min
Pocket Fiber Meter
1625
dB/ dBm/mWSet Ref
3.83 u W
Hold
nm
dB/ dBm/mW
HOLD
Max MinSet Ref
Pocket Fiber Meter
1625
-0.56 dB R
Hold
nm-24.1
dB/ dBm/mW
HOLD
Max MinSet Ref
Pocket Fiber Meter
1625
-24.17 dB m
Hold
nm
22
Toggle - Absolute/Relative/LinearToggle - Absolute/Relative/Linear
To toggle between Absolute / Relative and Linear mode. •Briefly, Press soft button [dB/dBm/mW] to toggle between the three modes.
[dB/dBm/mW]
dBm dBr W
[dB/dBm/mW]
[dB/dBm/mW]
dB/ dBm/mW
HOLD
Max MinSet Ref
Pocket Fiber Meter
1625
-0.56 dB R
Hold
nm-24.1
23
Set ReferenceSet Reference
Must be in Relative mode (dBr) not Absolute mode (dBm).
• Press and hold [Set Ref] for several seconds.– Meter will beep 2 times and zero at indicated
wavelength.– Reference value remembered at power off.
[Set Ref]
dBr
dB/ dBm/mW
HOLD
Max MinSet Ref
Pocket Fiber Meter
1625
-0.00 dB R
Hold
nm-24.7
9. Special Features9. Special Features
24
1.HoldThis feature is useful where you want to show a measurement to a third party.
2. Min & Max recordingThis feature is useful where a level needs to be monitored over a time period.
3. SlowMode• Tone detection disabled, enables measurement power in the presence of audio level modulation.• Improved low power measurement accuracy.
4. TamperLock Mode• Display settings are locked.• Useful for low skilled areas or where only 1 wavelength is in use.
dB/ dBm/mW
HOLD
Max MinSet Ref
Pocket Fiber Meter
1550-0.59 dB m
Hold
nm
9.1 Hold Function9.1 Hold Function
To activate the Hold function:•Briefly press [HOLD]
• The instrument will beep.• The word ‘Hold’ will flash in the lower RHS of the
display.
To turn Hold off.•Briefly press [HOLD]
25
[Hold]
dB/ dBm/mW
HOLD
Max MinSet Ref
Pocket Fiber Meter
1550-12.03dB m
Hold
nm
max
dB/ dBm/mW
HOLD
Max MinSet Ref
Pocket Fiber Meter
1550-16.55dB m
Hold
nm
min
9.2 Min / Max9.2 Min / Max
To retrieve the Maximum and Minimum value:•Press and hold [Max Min]
• Instrument will beep twice and
• the word ‘min’ will display on the upper RHS of the LCD.
•Repeat above to toggle to ‘max’ •Repeat above to turn function ‘Off’.
26
Min
Max
[Max Min]
[Max Min]
To reset Max/Min:•Toggle to another wavelength and back or•Turn power Off and then back On.
9.3 SlowMode9.3 SlowMode
To Activate at Power On• Press and Hold [dB/dBm/mW] while turning meter On. “- tonE” will be displayed.
To Activate when meter powered On• Press and Hold [λ], then press [dB/dBm/mW] “- tonE” will be displayed.
To Activate when test tone detection is active• Press [dB/dBm/mW] .
27
When SlowMode is On, alternating “” and “B” symbols will be displayed.
To exit SlowMode • Press and hold [λ], then press [dB/dBm/mW] or turn power Off and then back On.
HOLD
Max Min
Pocket Fiber Meter
dB/ dBm/mWSet Ref
1625
-24.17 dB m
Hold
nm
B
9.4 Tamper Lock Mode9.4 Tamper Lock Mode
To activate• Set meter to required λ & measurement mode, e.g. 850 and dBm
• Press and hold [λ], then press and hold [HOLD]• ‘codE’ will display• Press any 6 keys, other than [Power]
• record keystroke sequence for future use.
• Instrument will beep 3 times and display ‘Lout’
When active• ‘<<<‘ and ‘>>>’ displayed
To clear• Press and hold [λ] & [HOLD] whilst turning power On• ‘codE’ will display , enter 6 key code sequence• If successful ‘PASS’ will display.• Alternatively, TamperLock will be cleared after 3 unsuccessful attempts.
28
dB/ dBm/mW
HOLD
Max MinSet Ref
Pocket Fiber Meter
850
-24.17 dB m
Hold
nm
>>><<<
Application NotesApplication Notes
Comprehensive selection available at
www.kingfisher.com.au/ApplicationNotes.htm
29