5
TOWARD OPTIMIZING ELECTRODE CONFIGURATIONS TO IMPROVE MYOELECTRIC PATTERN RECOGNITION Aaron Young 1,2 , Levi Hargrove 2,3 1. Dept. of Biomedical Engineering, Northwestern University, Evanston, IL, USA 2. Center for Bionic Medicine, Rehabilitation Institute of Chicago, Chicago, IL, USA 3. Dept. of Physical Medicine and Rehabilitation, Northwestern University, Chicago, IL, USA ABSTRACT Standard myoelectric control systems use carefully placed bipolar electrode pairs to provide independent myoelectric signals (MESs) for prosthesis control. Because myoelectric pattern recognition systems do not require isolated MESs, the two electrode poles used for each MES channel may be placed longitudinally along individual muscles or transversely across multiple muscles. In addition, each electrode pole can be combined with a number of additional poles to form multiple channels. However, practical issues limit the number of poles that can be used in clinical settings. In this study, we investigated classification error reduction and controllability improvements provided by a combination of transverse and longitudinal MES channels in two conditions: (1) a constant number of electrode poles, and (2) a constant number of MES channels. In both cases, we also investigated performance when the electrodes were slightly shifted from their original positions to evaluate sensitivity to electrode shift. We found that a combination of two transverse and two longitudinal electrode channels constructed from four poles significantly outperformed the individual performances of either two transverse or two longitudinal channels each constructed from four poles (p<0.01). Using eight poles, we found that the best channel subset was always comprised of a combination of transverse and longitudinal channels. These results are important because the number and arrangement of poles and channels is a practical consideration for successful clinical implementation of myoelectric pattern recognition control. INTRODUCTION Myoelectric pattern recognition systems show promise for intuitive control of prostheses with multiple degrees of freedom [1]. Despite two decades of extensive research, these systems have yet to be clinically implemented. Typically, studies of pattern recognition systems have focused on signal processing aspects including data windowing [2], feature extraction [3], classification [4], and post-processing [5]. However, a key component of the system is the placement and configuration of the electrode poles. This is true in two contexts: first, the information content of each MES channel is affected by the configuration of the two electrode poles, and second, use of more electrode channels increases computational and financial costs and use of more electrode poles poses technical difficulties in embedding electrodes into prosthetic sockets. The information content of an MES is determined by the electrode detection volume, which defines the selectivity of the electrode. The primary factor affecting selectivity is the interelectrode distance: a rough estimate of detection volume is given by a sphere with radius equal to the interelectrode distance [6]. Most experiments have been conducted with interelectrode distances of approximately 2 cm, resulting in selective recordings. However, in pattern recognition systems, nonselective MES recordings may provide a different set of information that is complementary to the selective information. In addition, nonselective MES recordings may be less sensitive to changes in the location of the recording electrodes [7], such as those that result from donning and doffing the prosthesis or socket shift during use. Electrode shift is a potential problem in clinical applications because data presented to the classifier when electrodes are shifted are different from training data [8]. The number of channels used for classification of MESs varies between studies based on the classification problem and available recording equipment. Four channels are often used in studies of myoelectric pattern recognition control by transradial amputees [9]. One study [4] showed that for a myoelectric control of 10 motion classes using forearm muscles, four channels provided a sufficient amount of information for classification, and additional channels did not increase classification accuracy. In fact, the study showed a small decrease in accuracy as the number of channels was increased to 16. Other studies have shown similar findings with a plateau effect in classification accuracy with increasing numbers of channels [10, 11]. The number of channels is an important property of the MES detection system, and this study examines this property in terms of robustness to electrode shift. We compared pattern recognition system performance when using combinations of selective and nonselective recordings by measuring classification error and controllability testing. Selective recordings were obtained from bipolar electrode pairs with small interelectrode distances aligned with the underlying muscle fibers. Nonselective recordings were obtained from bipolar electrode pairs aligned in a transverse orientation to underlying muscle fibers and spanning muscle groups: one electrode pole was placed on the wrist flexor muscle group and one electrode pole is placed on the wrist extensor From "MEC 11 Raising the Standard," Proceedings of the 2011 MyoElectric Controls/Powered Prosthetics Symposium Fredericton, New Brunswick, Canada: August 14-19, 2011. Copyright University of New Brunswick. Distributed under a Creative Commons Attribution-Noncommercial-No Derivative Works 3.0 United States License by UNB and the Institute of Biomedical Engineering, through a partnership with Duke University and the Open Prosthetics Project.

TOWARD OPTIMIZING ELECTRODE CONFIGURATIONS TO …

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: TOWARD OPTIMIZING ELECTRODE CONFIGURATIONS TO …

TOWARD OPTIMIZING ELECTRODE CONFIGURATIONS TO IMPROVE

MYOELECTRIC PATTERN RECOGNITION Aaron Young

1,2, Levi Hargrove

2,3

1. Dept. of Biomedical Engineering, Northwestern University, Evanston, IL, USA

2. Center for Bionic Medicine, Rehabilitation Institute of Chicago, Chicago, IL, USA

3. Dept. of Physical Medicine and Rehabilitation, Northwestern University, Chicago, IL, USA

ABSTRACT

Standard myoelectric control systems use carefully

placed bipolar electrode pairs to provide independent

myoelectric signals (MESs) for prosthesis control. Because

myoelectric pattern recognition systems do not require

isolated MESs, the two electrode poles used for each MES

channel may be placed longitudinally along individual

muscles or transversely across multiple muscles. In

addition, each electrode pole can be combined with a

number of additional poles to form multiple channels.

However, practical issues limit the number of poles that can

be used in clinical settings. In this study, we investigated

classification error reduction and controllability

improvements provided by a combination of transverse and

longitudinal MES channels in two conditions: (1) a constant

number of electrode poles, and (2) a constant number of

MES channels. In both cases, we also investigated

performance when the electrodes were slightly shifted from

their original positions to evaluate sensitivity to electrode

shift. We found that a combination of two transverse and

two longitudinal electrode channels constructed from four

poles significantly outperformed the individual

performances of either two transverse or two longitudinal

channels each constructed from four poles (p<0.01). Using

eight poles, we found that the best channel subset was

always comprised of a combination of transverse and

longitudinal channels. These results are important because

the number and arrangement of poles and channels is a

practical consideration for successful clinical

implementation of myoelectric pattern recognition control.

INTRODUCTION

Myoelectric pattern recognition systems show promise

for intuitive control of prostheses with multiple degrees of

freedom [1]. Despite two decades of extensive research,

these systems have yet to be clinically implemented.

Typically, studies of pattern recognition systems have

focused on signal processing aspects including data

windowing [2], feature extraction [3], classification [4], and

post-processing [5]. However, a key component of the

system is the placement and configuration of the electrode

poles. This is true in two contexts: first, the information

content of each MES channel is affected by the

configuration of the two electrode poles, and second, use of

more electrode channels increases computational and

financial costs and use of more electrode poles poses

technical difficulties in embedding electrodes into prosthetic

sockets.

The information content of an MES is determined by the

electrode detection volume, which defines the selectivity of

the electrode. The primary factor affecting selectivity is the

interelectrode distance: a rough estimate of detection

volume is given by a sphere with radius equal to the

interelectrode distance [6]. Most experiments have been

conducted with interelectrode distances of approximately 2

cm, resulting in selective recordings. However, in pattern

recognition systems, nonselective MES recordings may

provide a different set of information that is complementary

to the selective information. In addition, nonselective MES

recordings may be less sensitive to changes in the location

of the recording electrodes [7], such as those that result from

donning and doffing the prosthesis or socket shift during

use. Electrode shift is a potential problem in clinical

applications because data presented to the classifier when

electrodes are shifted are different from training data [8].

The number of channels used for classification of MESs

varies between studies based on the classification problem

and available recording equipment. Four channels are often

used in studies of myoelectric pattern recognition control by

transradial amputees [9]. One study [4] showed that for a

myoelectric control of 10 motion classes using forearm

muscles, four channels provided a sufficient amount of

information for classification, and additional channels did

not increase classification accuracy. In fact, the study

showed a small decrease in accuracy as the number of

channels was increased to 16. Other studies have shown

similar findings with a plateau effect in classification

accuracy with increasing numbers of channels [10, 11]. The

number of channels is an important property of the MES

detection system, and this study examines this property in

terms of robustness to electrode shift.

We compared pattern recognition system performance

when using combinations of selective and nonselective

recordings by measuring classification error and

controllability testing. Selective recordings were obtained

from bipolar electrode pairs with small interelectrode

distances aligned with the underlying muscle fibers.

Nonselective recordings were obtained from bipolar

electrode pairs aligned in a transverse orientation to

underlying muscle fibers and spanning muscle groups: one

electrode pole was placed on the wrist flexor muscle group

and one electrode pole is placed on the wrist extensor

From "MEC 11 Raising the Standard," Proceedings of the 2011 MyoElectric Controls/Powered Prosthetics Symposium Fredericton, New Brunswick, Canada: August 14-19, 2011. Copyright University of New Brunswick.

Distributed under a Creative Commons Attribution-Noncommercial-No Derivative Works 3.0 United States License by UNB and the Institute of Biomedical Engineering, through a partnership with Duke University and the Open Prosthetics Project.

Page 2: TOWARD OPTIMIZING ELECTRODE CONFIGURATIONS TO …

muscle group. This configuration has a large interelectrode

distance and records a global signal from multiple muscles

[3].

In this paper, we seek to provide clinical

recommendations for electrode placement and number of

electrodes and recording channels based on offline

classification error, real-time controllability scores, and

robustness to electrode shift.

METHODS

Experiment 1: Effect of Pole Number and Placement

Seven able-bodied subjects participated in the study,

which was approved by the Northwestern University

Institutional Review Board. Two control sites on the

forearm were used: one on the flexor muscle group and one

on the extensor muscle group. At each control site, two

surface electrodes were placed longitudinal to the direction

of the underlying muscle fibers. A ground electrode was

placed on a bony region near the elbow away from the

muscles of interest. Four bipolar MES channels were

formed from these four electrode pole locations (

). Two channels were longitudinal (spanning each

individual control site), and two were transverse (one pole

on flexors and one on extensors).

For each subject, the classifier was trained and tested

offline with electrodes located at the nominal (or no-shift)

location. The electrodes were manually shifted 1 and 2 cm

from the nominal position in the direction parallel to the

underlying muscle fibers (distal to the subject) and 1 and 2

cm from the nominal position perpendicular to the

underlying muscles (clockwise from the subject’s

perspective). Testing data were recorded at each of these

four shift locations. Seven motion classes were recorded:

wrist flexion, wrist extension, forearm pronation, forearm

supination, hand close, hand open, and no movement.

A pattern recognition system similar to that used

previously [1] was used to discriminate motion classes.

MESs were sampled at 1 kHz and high-pass filtered at 20

Hz. Data were windowed in 250 ms intervals with 50 ms

overlap [2]. Time domain features [3] were classified using

linear discriminant analysis [1].

The performance of the classifier with three different

channel combinations using the same four electrode pole

locations were tested: (1) using two longitudinal channels,

(2) using two transverse channels, and (3) using a

combination of two longitudinal and two transverse

channels. We also compared our results to those achieved

when data collected from the shifted locations were

incorporated into the training data (referred to as

displacement training) in order to reduce classifier

sensitivity to electrode shift [12].

A Target Achievement Control (TAC) test [5] was used

to evaluate controllability. Subjects controlled a virtual

prosthesis to achieve target postures shown on a screen (see

[2, 5] for more details on TAC testing). Only one motion

class (e.g. wrist flexion and extension) was required per

trial. If mistakes were made, for example, activation of a

different motion class or overshooting the target, the subject

had to make a corrective activation. The subject had 17 s to

complete each trial. A test consisted of two trials of each

motion class. Performance was assessed by failure rate: the

percentage of trials that the subject did not complete during

a test. TAC tests were completed at the no-shift and 2 cm

shift (both parallel and perpendicular) locations. At each

location, one test was completed for two longitudinal

channels, and one test was completed using two longitudinal

and two transverse channels.

Experiment 2: Effect of Number of Channels

This experiment was similar to the first except four

control sites spaced equally around the circumference of the

forearm were used. Eight bipolar MES channels were

formed from eight electrode pole locations (Figure 2). Four

channels were longitudinal and four were transverse in an

arrangement similar to experiment 1.

Medial View Lateral View

Figure 1: Electrode placement for experiment 1.

Electrode poles 1 and 2 are located on wrist flexors and

3 and 4 on wrist extensor muscles. The two longitudinal

channels were the bipolar pairs of 1 & 2 and 3 & 4. The

transverse channels were 1 & 3 and 2 & 4.

From "MEC 11 Raising the Standard," Proceedings of the 2011 MyoElectric Controls/Powered Prosthetics Symposium Fredericton, New Brunswick, Canada: August 14-19, 2011. Copyright University of New Brunswick.

Distributed under a Creative Commons Attribution-Noncommercial-No Derivative Works 3.0 United States License by UNB and the Institute of Biomedical Engineering, through a partnership with Duke University and the Open Prosthetics Project.

Page 3: TOWARD OPTIMIZING ELECTRODE CONFIGURATIONS TO …

Medial View Lateral View

Figure 2: Electrode placement for experiment 2.

Longitudinal channels were formed between poles 1 & 2, 3

& 4, 5 & 6, and 7 & 8. Transverse channels were formed

between poles 1 & 3, 2 & 4, 5 & 7, and 6 & 8.

The same motions and classification techniques as

described for experiment 1 were used. The classifier was

trained and tested at the no-shift location. In this

experiment, electrodes were only shifted 1 and 2 cm

perpendicular to the no-shift location, because results from

experiment 1 showed that the classifier was more sensitive

to perpendicular shifts than to parallel shifts. Classifier

performance resulting from use of from one to eight

channels was evaluated and the best channel subset for each

number of electrodes was determined based on the error at

the no shift, 1 cm shift, and 2 cm shift locations using the

following weighted error formula:

Weighted Error = 2*Error(No Shift) + 1.5*Error(1 cm Shift)

+Error(2 cm shift).

Every combination of channels was tested and a

weighted error was assigned to each combination. The

optimal combination was that which had the lowest

weighted error compared to others with the same number of

channels.

RESULTS

Experiment 1: Effect of Pole Number and Placement

In the first experiment, we evaluated the best electrode

channel configuration for a constant number of electrode

poles. Two longitudinal and two transverse channels were

formed from the same four electrode poles in different

configurations. The combined selective and nonselective

information from both longitudinal and transverse channels

performed the best with and without shift (Figure 3)

compared to only transverse channels (p<0.01) and only

longitudinal channels (p<0.05).

0

10

20

30

40

50

60

No shift 1cm || 2cm || 1cm ⊥ 2cm ⊥

Cla

ssif

ica

tio

n E

rro

r

Trasverse Only

Longitudinal Only

Longitudinal Only with Displacement Training

Longitudinal and Transverse

Longitudinal and Transverse with Displacement Training

Figure 3: Classification error of electrode configurations

using four electrode poles. Displacement training results are

also displayed for two of the configurations. Error bars

show one standard error of the mean. || refers to parallel

shifts and ⊥ refers to perpendicular shifts.

Training with displacement data increased error at the

no-shift location in both cases, but also decreased sensitivity

to electrode shift (Figure 3). Displacement training did not

reduce sensitivity to shift for the longitudinal channels as

much as adding the transverse channel did. However, by

adding the transverse channels and incorporating

displacement training data, sensitivity to shift was greatly

reduced at all shift locations (p<0.05) compared to a

combination of longitudinal and transverse channels without

displacement training.

TAC test controllability results demonstrated a trend

similar to that of the classification error: a combination of

longitudinal and transverse channels outperformed

longitudinal channels alone (Figure 4), especially when

electrodes were shifted. The controllability test was not hard

enough to separate out the performance of the two

classifiers at the no shift location, as both had very low

failure rates.

From "MEC 11 Raising the Standard," Proceedings of the 2011 MyoElectric Controls/Powered Prosthetics Symposium Fredericton, New Brunswick, Canada: August 14-19, 2011. Copyright University of New Brunswick.

Distributed under a Creative Commons Attribution-Noncommercial-No Derivative Works 3.0 United States License by UNB and the Institute of Biomedical Engineering, through a partnership with Duke University and the Open Prosthetics Project.

Page 4: TOWARD OPTIMIZING ELECTRODE CONFIGURATIONS TO …

0

10

20

30

40

50

60

70

80

No Shift Parallel Shift Perpendicular Shift

Failu

re R

ate

(%)

Longitudinal and Transverse

Longitudinal Only

Figure 4: TAC test failure rates for two electrode

configurations with and without 2 cm shifts. Error bars

show one standard error of the mean.

The longitudinal channels alone performed well without

shift, but had high failure rate with shift in either direction.

By adding transverse channels, the no-shift failure rate

dropped slightly and the failure rate was substantially

reduced for both shift directions. In particular, using both

longitudinal and transverse channels and a 2 cm parallel

shift the failure rate was lower than the longitudinal

channels without shift.

Experiment 2: Effect of Number of Channels

First, the best subset of channels for each number of

channels (between one and eight) was determined based on

their weighted classification errors. Only the unweighted

classification errors are displayed here (Figure 5).

Interestingly, combinations with at least one longitudinal

and one transverse channel were always in the best subset

when more than one channel was used.

The use of four to six channels had the lowest

classification error across the three shift conditions. An

important result was that errors were below 15% at the

nonshifted and 1 cm shift locations when using more than

two recording channels. The 2 cm shift location had high

classification error regardless of the number of recording

channels.

0

5

10

15

20

25

30

35

40

45

50

1 2 3 4 5 6 7 8

Cla

ssif

icat

ion

Err

or

(%)

Number of Channels

no shift

1 cm shift

2cm shift

Figure 5: Effect of the number of recording channels on

classification error. Error bars show one standard error of

the mean.

DISCUSSION

The first goal of this study was to find the electrode

configuration that gave the highest performance given a

limited number of electrode pole locations. Performance

was measured in terms of classification error,

controllability, and robustness to electrode shift. We found

that a combined configuration that included both

longitudinal and transverse channels was the highest

performing configuration of those that were tested. This

configuration had lower error and better controllability

without shift and at every shift location tested compared to

using only longitudinal or transverse channels.

Previous investigators have considered including

displacement location data in the training data in order to

train a more robust pattern recognition system [12]. We

repeated this analysis for the configuration with longitudinal

channels and the configuration with combined longitudinal

and transverse channels. We found that displacement

training increased classification error at the no-shift

location, but helped to reduce sensitivity at shift locations.

The combination of displacement location with longitudinal

and transverse channels performed especially well with less

than 15% classification error at all tested locations. This is a

clinically useful result as with only four channels and four

pole locations, low classification errors were achieved with

high robustness across shift conditions.

The second goal of this study was to analyze the

number of recording channels necessary for sufficiently

high classification and to determine the optimal composition

of channel subsets. Based on weighted averages, it was

found that a combination of longitudinal and transverse

channels was always optimal regardless of the number of

channels. The combination of selective and nonselective

From "MEC 11 Raising the Standard," Proceedings of the 2011 MyoElectric Controls/Powered Prosthetics Symposium Fredericton, New Brunswick, Canada: August 14-19, 2011. Copyright University of New Brunswick.

Distributed under a Creative Commons Attribution-Noncommercial-No Derivative Works 3.0 United States License by UNB and the Institute of Biomedical Engineering, through a partnership with Duke University and the Open Prosthetics Project.

Page 5: TOWARD OPTIMIZING ELECTRODE CONFIGURATIONS TO …

information decreased error with and without shift

compared to configurations with only one type of electrode

configuration. Experiment 2 showed that there was little or

no improvement in terms of classification error when using

more than six channels. For the transradial case, this study

demonstrates that four to six channels are sufficient to

obtain classification errors of less than 15% both without

electrode shift and with shifts up to 1 cm.

ACKNOWLEDGEMENTS

The authors would like to acknowledge Ann Barlow and

Aimee Schultz for helping to revise and edit the manuscript,

Annie Simon for providing valuable contributions, and

Todd Kuiken for providing overall project guidance for this

study.

REFERENCES

[1] T. A. Kuiken, G. Li, B. A. Lock, R. D. Lipschutz, L. A. Miller,

K. A. Stubblefield, and K. B. Englehart, "Targeted muscle reinnervation for

real-time myoelectric control of multifunction artificial arms," JAMA, vol. 301, pp. 619-28, 2009.

[2] L. H. Smith, L. Hargrove, B. A. Lock, and T. Kuiken,

"Determining the optimal window length for pattern recognition-based myoelectric control: balancing the competing effects of classification error

and controller delay," IEEE Transactions on Neural Systems and

Rehabilitation Engineering, vol. In Press, 2010. [3] B. Hudgins, P. Parker, and R. N. Scott, "A new strategy for

multifunction myoelectric control," IEEE Transactions on Biomedical

Engineering, vol. 40, pp. 82-94, 1993. [4] L. Hargrove, K. Englehart, and B. Hudgins, "A Comparison of

Surface and Intramuscular Myoelectric Signal Classification," IEEE

Transactions on Biomedical Engineering, 2007.

[5] A. Simon, L. Hargrove, B. A. Lock, and T. Kuiken, "A

decision-based velocity ramp for minimizing the effect of

misclassifications during real-time pattern recognition control," IEEE Transactions on Neural Systems and Rehabilitation Engineering, vol.

Accepted, 2010.

[6] G. Kamen and D. Gabriel, Essentials of Electromyography. Champaign: Human Kinetics Publishers, 2009.

[7] A. Young, L. Hargrove, and T. Kuiken, "The Effects of

Electrode Size and Orientation on the Sensitivity of Myoelectric Pattern Recognition Systems to Electrode Shift," IEEE Transactions on

Biomedical Engineering, under minor revisions. [8] L. Hargrove, K. Englehart, and B. Hudgins, "The effect of

electrode displacements on pattern recognition based myoelectric control,"

in Proceedings of the 28th IEEE EMBS Annual International Conference, New York City, 2006, pp. 2203-2206.

[9] K. Englehart and B. Hudgins, "A robust, real-time control

scheme for multifunction myoelectric control," IEEE Transactions on Biomedical Engineering, vol. 50, pp. 848-54, 2003.

[10] H. Huang, P. Zhou, G. Li, and T. A. Kuiken, "An Analysis of

EMG Electrode Configuration for Targeted Muscle Reinnervation Based Neural Machine Interface," IEEE Transactions on Neural Systems and

Rehabilitation Engineering, vol. 16, pp. 37-45, 2008.

[11] G. Li, A. E. Schultz, and T. A. Kuiken, "Quantifying pattern recognition-based myoelectric control of multifunctional transradial

prostheses," IEEE Transactions on Neural Systems and Rehabilitation

Engineering, vol. 18, pp. 185-92, 2010. [12] L. Hargrove, K. Englehart, and B. Hudgins, "A training strategy

to reduce classification degradation due to electrode displacements in

pattern recognition based myoelectric control," Biomedical Signal Processing and Control, vol. 3, pp. 175-180, 2008.

From "MEC 11 Raising the Standard," Proceedings of the 2011 MyoElectric Controls/Powered Prosthetics Symposium Fredericton, New Brunswick, Canada: August 14-19, 2011. Copyright University of New Brunswick.

Distributed under a Creative Commons Attribution-Noncommercial-No Derivative Works 3.0 United States License by UNB and the Institute of Biomedical Engineering, through a partnership with Duke University and the Open Prosthetics Project.