12
SURGICAL TECHNIQUE V2.1– POLYAMIDE TORNIER BLUEPRINT 3D Planning + PSI

TORNIER BLUEPRINT...BLUEPRINT 3 e Surgical Technique v2.1 – Polyamide 5 3D Planning Once BLUEPRINT 3D Planning Software is installed on a clinic, hospital, or personal computer and

  • Upload
    others

  • View
    4

  • Download
    0

Embed Size (px)

Citation preview

Page 1: TORNIER BLUEPRINT...BLUEPRINT 3 e Surgical Technique v2.1 – Polyamide 5 3D Planning Once BLUEPRINT 3D Planning Software is installed on a clinic, hospital, or personal computer and

S U R G I C A L T E C H N I Q U E

V 2 .1 – P O LYAM I D E

TORNIER

BLUEPRINT3D Planning + PSI

Page 2: TORNIER BLUEPRINT...BLUEPRINT 3 e Surgical Technique v2.1 – Polyamide 5 3D Planning Once BLUEPRINT 3D Planning Software is installed on a clinic, hospital, or personal computer and
Page 3: TORNIER BLUEPRINT...BLUEPRINT 3 e Surgical Technique v2.1 – Polyamide 5 3D Planning Once BLUEPRINT 3D Planning Software is installed on a clinic, hospital, or personal computer and

3BLUEPRINT 3D Planning Software Surgical Technique v2.1 – Polyamide

Contents

Patient-Specific Instrumentation Overview

3D Planning

PSI Guide Creation and Order

Use of PSI Guide Intra-Operatively

Patient Specific Instrumentation Breakdown

4

5

6

7

9

Page 4: TORNIER BLUEPRINT...BLUEPRINT 3 e Surgical Technique v2.1 – Polyamide 5 3D Planning Once BLUEPRINT 3D Planning Software is installed on a clinic, hospital, or personal computer and

4 BLUEPRINT 3D Planning Software Surgical Technique v2.1 – Polyamide

Surgical Technique

Patient-Specific Instrumentation Overview

If pre-operative planning and patient-specific instrumentation are desired, BLUEPRINT™ 3D Planning software may be used prior to the procedure to visualize the anatomy in 3-dimensional space (automated 3D reconstruction) and perform a virtual implantation. The software allows the surgeon to virtually position the various implants and understand the appropriate path of treatment based on the patient’s anatomy. The software creates a visual and quantified aid when making important treatment decisions, including:

» Identification of glenoid wear patterns

» Visualization of humeral head subluxation and migration

» Planning correction of glenoid version and inclination

» Ensuring full implant seating

» Analyzing bone removal required for various implant options

» Assuring implant containment within the glenoid vault (size and radius curvature)

When the planning is complete, the software can generate patient-specific instrumentation (PSI) that replicates intra-operatively the same implant positioning as previously planned.

Visualize

Virtual Implantation

Create Guide

Patient-Specific Guide

OPERATING ROOM3D PLANNING

Replicate Planning on Patient

Page 5: TORNIER BLUEPRINT...BLUEPRINT 3 e Surgical Technique v2.1 – Polyamide 5 3D Planning Once BLUEPRINT 3D Planning Software is installed on a clinic, hospital, or personal computer and

5BLUEPRINT 3D Planning Software Surgical Technique v2.1 – Polyamide

3D Planning

Once BLUEPRINT 3D Planning Software is installed on a clinic, hospital, or personal computer and compatible CT scan is obtained (refer to BLUEPRINT

Scan protocol), run the software and plan a new case by loading the CT scan (in DICOM format).

BLUEPRINT 3D Planning provides preoperative glenoid measurements in order to identify the glenoid wear patterns, humeral head subluxation and migration, and bone stock.

Select and position the glenoid component in order to get an appropriate position within the three displayed views of scapula.

Positioning is defined by adjusting the following parameters:

» Version

» Inclination

» Position (Antero-Posterior, Supero-Inferior, implant Rotation)

» Medialization or Lateralization

» Bone seating of implant

Note: Implant turning to red means the presence of bone perforation of the implant existing from the glenoid vault. When needed, visualize the perforation and correct the perforation by repositioning the implant.

Page 6: TORNIER BLUEPRINT...BLUEPRINT 3 e Surgical Technique v2.1 – Polyamide 5 3D Planning Once BLUEPRINT 3D Planning Software is installed on a clinic, hospital, or personal computer and

6 BLUEPRINT 3D Planning Software Surgical Technique v2.1 – Polyamide

PSI Guide Creation ad OrderWhen the implant positioning is set and a patient-specific guide is desired, select the blue “Continue” button in the lower right corner of the screen (Figure 1).

Choose four different points on the edge of the glenoid (Figure 2). These points will establish the position of the feet of the PSI guide. One point needs to be on the posterior part of the glenoid fossa and three points on the anterior part of the glenoid fossa.

Note: Select the points according to the position of retractors of desired surgical approach.

A 3D PSI guide will be generated automatically based on the planning previously performed and the four selected points. (Figure 3)

By saving the plan and specifying the date of surgery, the software will generate a planning report including a summary of parameters and the specifications of the created PSI guide.

Note: When a PSI guide needs to be ordered for a case, confirm the computer is connected to

internet and upload the planned case.

All planned cases and PSI orders can be reviewed and managed online on www.shoulderblueprint.com in your surgeon portal.

Note: No patient data is requested by Wright Medical and no patient data will be sent to

Wright Medical.

Figure 1

Figure 2

Figure 3

Page 7: TORNIER BLUEPRINT...BLUEPRINT 3 e Surgical Technique v2.1 – Polyamide 5 3D Planning Once BLUEPRINT 3D Planning Software is installed on a clinic, hospital, or personal computer and

7BLUEPRINT 3D Planning Software Surgical Technique v2.1 – Polyamide

Use of PSI Guide Intra-Operatively

PSI guide assembly

The PSI guide is shipped with an identification (ID) tag where the anonymous patient-specific tracking information, part number, and lot number reside. This tag should be removed prior to use (Figure 4). Caution should be taken after the tag has been removed due to the potential for glove tears on a possible sharp edge created by the tag removal.

The removal of the ID tag should be performed on the back table. Using a pair of sterile scissors, cut along the outside of the central hole on the PSI guide (Figure 5).

With the ID tag removed, insert the pin guide into the central hole on the PSI Guide and give a quarter turn to secure the PSI guide on the pin guide.

cut pin guide

insertion

Note: The pin guide features a morse taper style press-fit design which fits in the central

hole on the PSI guide.

Figure 4

Figure 5

Page 8: TORNIER BLUEPRINT...BLUEPRINT 3 e Surgical Technique v2.1 – Polyamide 5 3D Planning Once BLUEPRINT 3D Planning Software is installed on a clinic, hospital, or personal computer and

8 BLUEPRINT 3D Planning Software Surgical Technique v2.1 – Polyamide

Use of the Glenoid 3D model and the PSI guide

The BLUEPRINT PSI guide references the patients’ operative glenoid bone. Osteophytes must be kept in order to ensure a good fit of the guide. Additionally, the operative glenoid must be cleaned of soft tissues such as cartilage and labrum prior to use of the PSI guide on the patient’s glenoid bone.

The surgeon may utilize the PSI guide with the 3D glenoid model and compare the fit to the glenoid (Figure 6 and 7).

The feet of the PSI guide are designed to match the landmarks chosen previously by the surgeon during the final guide design process.

Once the PSI guide is in place on the operative glenoid, insert the final guide pin that will guide the subsequent cannulated drills (refer to the surgical technique of the planned implant) and reamers. By placing the guide pin through the PSI guide in place, the surgeon replicates the plan on patient.

Note: Osteophytes must not be removed prior to the use of the BLUEPRINT PSI to ensure

guide fit.

Figure 6

Figure 7

Page 9: TORNIER BLUEPRINT...BLUEPRINT 3 e Surgical Technique v2.1 – Polyamide 5 3D Planning Once BLUEPRINT 3D Planning Software is installed on a clinic, hospital, or personal computer and

9BLUEPRINT 3D Planning Software Surgical Technique v2.1 – Polyamide

Patient-Specific Instrumentation Breakdown

MWJ001

MWJ020

MWJ010

BLUEPRINT Patient Specific Instrumentation Description Reference Associated Implants

BLUEPRINT Glenoid Guide MWJ001

All ImplantsBLUEPRINT Glenoid Bone Model MWJ010

BLUEPRINT Reusable Pin Guide MWJ020

Page 10: TORNIER BLUEPRINT...BLUEPRINT 3 e Surgical Technique v2.1 – Polyamide 5 3D Planning Once BLUEPRINT 3D Planning Software is installed on a clinic, hospital, or personal computer and

10 BLUEPRINT 3D Planning Software Surgical Technique v2.1 – Polyamide

Notes

Page 11: TORNIER BLUEPRINT...BLUEPRINT 3 e Surgical Technique v2.1 – Polyamide 5 3D Planning Once BLUEPRINT 3D Planning Software is installed on a clinic, hospital, or personal computer and

11BLUEPRINT 3D Planning Software Surgical Technique v2.1 – Polyamide

Notes

Page 12: TORNIER BLUEPRINT...BLUEPRINT 3 e Surgical Technique v2.1 – Polyamide 5 3D Planning Once BLUEPRINT 3D Planning Software is installed on a clinic, hospital, or personal computer and

CAW-8609 Rev A DCR 170421 Feb-2017

Proper surgical procedures and techniques are the responsibility of the medical professional. This material is furnished for information purposes only. Each surgeon must evaluate the appropriateness of the material based on his or her personal medical training and experience. Prior to use of any Tornier implant system, the surgeon should refer to the product package insert for complete warnings, precautions, indications, contraindications, and adverse effects. Package inserts are also available by contacting Wright. Contact information can be found in this document and the package insert. The BLUEPRINT™ Glenoid Guides are intended to be used as surgical instruments to assist in the intraoperative positioning of glenoid components used with total anatomic or reversed should arthroplasty procedures.

™ and ® denote Trademarks and Registered Trademarks of Wright Medical Group N.V. or its affiliates.©2017 Wright Medical Group N.V. or its affiliates. All Rights Reserved.

BLUEPRINT 3D PLANNING SOFTWARE HELPLINEEmail: [email protected]

Phone: 855.378.1459 Managed 8:00am–5:00pm CST

10801 Nesbitt Avenue SouthBloomington, MN 55437888 867 6437952 426 7600www.wright.com

161 Rue Lavoisier38330 Montbonnot Saint MartinFrance+33 (0)4 76 61 35 00