Upload
others
View
3
Download
0
Embed Size (px)
Citation preview
TOF/QTOF Users Meeting
Jim Lau, Agilent Technologies.
Agilent Users Meeting
September 17th, 2008
Agenda • Goal – Share Best Practices, Develop Network of Agilent Users
Today’s Topics:
• Care and Feeding Agilent LCMS Systems
• Data Acquisition: Optimize TOF/QTOF for Best Mass Accuracy
• MassHunter Qual: MFG and MFE for Small and Large Molecule Applications
• Overview MassProfiler Software
• MassHunter Bioconfirm Software
• Questions and Answer Period
Agilent Users Meeting
September 17th, 2008
Care and Feeding of QTOF/TOF (Cont.)
• Defragment Hard Drives (Weekly) – Use Diskeeper to Defragment Drives – If Errors – Run Check Disk for Errors Program
• Check LC Components Firmware – If older version update LC Firmware
Agilent Users Meeting
September 17th, 2008
Data Acquisition Overview How to Improve Mass Accuracy
• Optimize Source Parameters: – Drying Gas Temperature and Flow; Fragmenter Voltage
• Optimize Ion Transmission in Quads – For Low Mass Analytes < 300 Change Quad Amu Gain to 100
• Review How to Calibrate/Tune Q(TOF) • Optimize Points Defining a Peak
– Reduce maximum mass from m/z 3200 to 1700 – Increase Acquisition Rate MS (2-4 spectra/s)
• Causes of High Mass Error (> 3ppm) – Saturated Ions – Unresolved Ions
Agilent Users Meeting
September 17th, 2008
Optimization of ESI Source Parameters
VCap – Entrance of the Capillary, 2000 to 4500V (3500V) Dependent on compound and MW Concentrated samples (µg/µL) lower VCap
Drying Gas - Dependent on LC flow rate 10-13 L/sec
Gas Temp - Start at high temperature work down 275-325oC
Skimmer - Voltage on skimmer and lens after capillary 65V
Fragmentor- Not Same as QQQ/Quad, Compound Dependent
Vary fragmentor/scan (30V steps) 120V to 240V FIA with Injection Program or MultiFragmentor acquis.
Agilent Users Meeting
September 17th, 2008
Vary Fragmentor with FIA Injection Program Multiple Methods or Scan Functions Auto Sampler Injection Program • REMOTE Startpulse • REPEAT 5 times • VALVE bypass • DRAW Def Amount from Sample, def.
speed, def offset • VALVE mainpass • WAIT 0.75 min • END REPEAT
Vary Fragmentor Voltage • Right Click in Expt Area and select “Add
Experiment” • Vary Fragmentor Voltage by 20-40V per
Experiment • Acquire Data using Injector Program • Determine Optimum Voltage from EIC as a
Function of Fragmentor Voltage
Agilent Users Meeting
September 17th, 2008
Extracted Ion Chromatograms: Advanced Tab “Fragmentor Setting”
Review How to Tune and Calibrate the Q(TOF)
Agilent Users Meeting
September 17th, 2008
Entering Calibration and Tuning
In Mass Hunter Workstation Data Acquisition, select the Tune Context.
September 17th, 2008
Agilent Users Meeting
Set Instrument State for Calibration
• Mass Range: Low (1700 m/z), Standard (3200 m/z) or
High (20000)
• Instrument Modes: • High Resolution (4GHz) • 4GHz (High Resolution disabled) • Extended Dynamic Range (2GHz) • Minimum Storage Size (1 GHz)
Advanced showing
Agilent 6520
September 17th, 2008
Agilent Users Meeting
Calibration
• Calibrate weekly. • Generates fifth order polynomial to accurately assign masses over entire mass range. •Takes 10 –15 seconds.
1. Turn on MS. 2. Select Instrument State. 3. Select Bottle B. 4. Press Calibrate. 5. Apply
September 17th, 2008
Agilent Users Meeting
Calibration Result - Mass Calibration Curve
• The Mass Calibration Curve is generated by selecting the Calibrate button.
• The calibration curve shows the time-to-mass fit of the polynomial correction across the full mass range.
• Apply accepts new TOF Mass Coefficients to polynomial mass correction function.
• Residual Error display often exhibits polynomial “wave” pattern.
September 17th, 2008
Agilent Users Meeting
Calibration Result - Detailed Residual Plot
Detailed Residual Plot shows: • uncorrected mass error in ppm (blue diamonds). • corrected mass error in ppm (red squares) after applying fifth-order polynomial mass correction. • actual calibration curve – green line.
September 17th, 2008
Agilent Users Meeting
Optimization for Low Mass Operation
1. Calibrate using the entire mass range (High 3200 m/z). 2. Perform an AutoTune. 3. Instrument State – change mass range from 3200 to
1700. 4. Deselect higher masses not in new range from mass list. 5. Re-calibrate.
September 17th, 2008
Agilent Users Meeting
Mass Calibration Error Reduced
•The errors have been reduced. •Must have 6 points. •No gain with fewer points.
September 17th, 2008
Agilent Users Meeting
Automatic and Manual Tuning
Automated tuning for most users
Manual tuning for diagnostics and special applications
September 17th, 2008
Agilent Users Meeting
Order of Tuning Choices
(Check TOF Tune) • If necessary, do… (Quick TOF Tune) • If necessary, do… (TOF Tune) • If necessary, do…FIRST every few months Initial TOF Tune
(Check Quad Tune) • If necessary, do… (Quad Tune) • If necessary, do…AT PM time (Yearly) Initial Quad Tune
September 17th, 2008
Agilent Users Meeting
Autotunes – TOF and Quad Autotune
TOF Autotune steps are displayed while the system tunes.
Autotune Report
September 17th, 2008
Agilent Users Meeting
Save Tune File When Changing Context
Make certain to specify correct tune file in acquisition method.
Or, save and load tune files in the Instrument State tab.
September 17th, 2008
Agilent Users Meeting
Agilent Users Meeting
September 17th, 2008
Agilent Users Meeting
September 17th, 2008
New Makeup Procedure for Reference Solution
Changing from Acetonitrile to Methanol Solution • 2 mL of HP121 into 98 mL of 95/5 MeOH/Water
• 0.8 mL of HP921 into 99.2 mL of 95/5 MeOH/Water
• Using a 500 mL volumetric flask, mix 25 mL of HP121 and 921 into 450 mL of 95/5 MeOH/Water with 0.2% Acetic Acid.
• Internal Reference Mass Correction – Two Ions (1 < m/z 350, second 500 amu higher) – (m/z 121.050873, 149.02332, 322.048121, 922.009798, 2421.91339) – Neb Pressure at 5 psi
Better for Negative ion performance and for use with APCI and Multimode sources without changing reference solution.
Addition of Negative Ion Reference Ion
• From Acquisition • MS TOF Tab • General Tab
– Ion Polarity – Negative
• Ref Mass Tab – Right Click “Add” – Input Mass 966.000725 – Input Mass 980.016375
Reference Solution can also be introduced using a Binary or Isocratic with a 100:1 splitter designed by Agilent. Flow rate of LC set at between 0.5-1.0 mL/min and gives very stable flow.
Agilent Users Meeting
September 17th, 2008
My Mass Error is Above 3 ppm What should I do?
• Check for Interference from Chemical Background – Change Chromatography – Create Narrow (+/- 10ppm) Extracted Ion Chromatogram with subtraction
• Check Resolution of Peak in Spectrum Peak List • Check for Saturated Peak – Peak List or “*” in Mass Spectrum
Agilent Users Meeting
September 17th, 2008
Agilent Users Meeting
September 17th, 2008
Agilent Users Meeting
September 17th, 2008
Agilent Users Meeting
September 17th, 2008
Poor Mass Accuracy Example – Unresolved Peaks “Spectrum Peak List 1 Table – Resolution Column”
EIC m/z 337.09847
EIC m/z 338.17521
BPC
1.59 ppm error
Agilent Users Meeting
September 17th, 2008
Compare Manual versus Automated Analysis “Find Compounds by Feature Extraction”
Manual Spectral Subtraction
Automated Spectral Extraction
Agilent Users Meeting
September 17th, 2008
QUALITATIVE ANALYSIS
Find Compounds by Formula
Find Compounds in analysis by inputting: •Formulas separate by “,” •A database with Formulas •Worklist Formulas
Find Compounds by Formula - Options Input Comma Separated Formulas
Find Compounds by Formula: Accurate Mass Database Search
Modify default.csv file adding formula, RT and exact mass of compound
Results “Compound Found” in Table
Find Compounds by Formula
Compounds Found
Identify Compounds
Generate Formulas after a Find Compounds
Search a Database after Find Compounds
Example: Characterization of Synthetic Urine
Mixture of 72 Compounds MW 100-625 Daltons, Mixed Classes.
Acquire ESI Positive and Negative ion MS and MSMS .
First Step: Find Compounds by Molecular Feature.
Second Step: Search Compound Table by Metlin & Exact Mass Database.
Compare Accurate Mass, Coverage, Dynamic Range.
Database Searching
Search Mode
Automated Compound Search Metlin Database
Creating a Database
Modify existing database file and save to unique name.
MassHunter\Databases\default.csv.
Mass Hunter Qual…Molecular Formula Generation
Agilent Users Meeting
September 17th, 2008
Molecular Formula Generator
September 17th, 2008
Agilent Users Meeting
Molecular Formula Generation Parameters
• Pull Down Menu.. – Compounds or Spectrum Peaks
• Adduct Selection
• Electron State: Even Electron
– Allows “none”
• Add Natural Isotopes with “[ ]”
• Limits for Matches using MS, MSMS Mass Error, DBE.
• DBE can be Negative for Sulfate and Phosphate Compounds.
September 17th, 2008
Agilent Users Meeting
Molecular Formula Generator
MS Formula Results
MS/MS Formula Results
September 17th, 2008
Agilent Users Meeting
Molecular Formula Generation
Agilent Users Meeting
September 17th, 2008
Allowed Species, Limits and Charge State User Inputs
Agilent Users Meeting
September 17th, 2008
Charge State and Isotope Distribution (B.02.00)
Agilent Users Meeting
September 17th, 2008
New in B.02.00
Isotope Model:
Common Organic Molecules Peptides Unbiased (unique isotope patterns)
Charge State: 1 or 2 for small molecule 10-15 for large molecule
Isotope Distribution (Experimental/Theoretical)
Agilent Users Meeting
September 17th, 2008
Cross Score (using MFG with MFE)
Agilent Users Meeting
September 17th, 2008
MFG Overview
Agilent Users Meeting
September 17th, 2008
Agilent Users Meeting
September 17th, 2008
Molecular Feature Extractor (MFE) – How it works
• Map signals in the 3-dim. space in time and mass at the MS level. IMPORTANT: High resolution in mass and time!
• Remove areas which only contain noise, and NO signals. • Identify all mass signals with a common RT (narrow time window) • Combine mass signals with common RT and chemical relation (isotope,
adduct, dimer, higher charge state) => Molecular feature or Compound. • Create Extracted Compound Chromatograms (ECC) and Compound Mass
Spectra from ions associated to a molecular feature (compound)
RT m/z Abund … … …
… … … 2.11 195.1745 21000 2.11 257.2566 33550
2.34 224.2134 11784
… … …
… … …
Compound list or Molecular feature list
Find by Molecular Feature
September 17th, 2008
Agilent Users Meeting
Method Editor – MFE (cont.)
Extraction Parameters •Target Data Type Small Molecule Large Molecule Infusion
•Peak Filters S/N Profile only Peak Height both
September 17th, 2008
Agilent Users Meeting
Method Editor – MFE (cont.)
Agilent Users Meeting
September 17th, 2008
Method Editor – MFE (cont.)
Agilent Users Meeting
September 17th, 2008
Method Editor – MFE (cont.)
Agilent Users Meeting
September 17th, 2008
Method Editor – MFE (cont.)
Agilent Users Meeting
September 17th, 2008
Executed
September 17th, 2008
Agilent Users Meeting
Find Compounds by Molecular Feature 6x10
0
0
0
0
1
1
1
1
1
2
2
2
2
2
Cpd 279:+ ECC Scan CoffeeSpiked12.d
Counts vs. Acquisition Time (min)
0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 9
Over 280 Compounds Found in Coffee Spiked with Pesticide
September 17th, 2008
Agilent Users Meeting
Compound List
Use found compounds for further analysis: •Use Molecular Formula Generator on list. •Search a Database and match findings •Upload the MFE results to Mass Profiler or GeneSpring for analysis.
September 17th, 2008
Agilent Users Meeting
Large Molecule Applications Proteins, Oligonucleotides, Peptides
Agilent Users Meeting
September 17th, 2008
API-Electrospray: Multiply-Charged Data
Multiply charged compounds acquire a range of charge states yielding a series of mass peaks for a single compound.
Each mass peak represents a different charge state.
Compounds in excess of 100,000 Da may be analyzed.
Picomole to femtomole full scan spectra.
Useful for obtaining molecular weights
Find Molecular Weight of Intact Protein
1000 1500
100000
120000
20000
40000
60000
80000
706
.85
726
.25
746
.15
767
.55
789
.95
839
.45
866
.45
895
.45
926
.15
959
.15
994
.85
103
2.95
1
074.
15
111
8.85
1
167.
55
122
0.45
127
8.75
134
2.65
141
3.25
157
9.65
Green Fluorescent Protein MW 27 kDa - Multiple Charged Envelope
Determine Charge State: • Assume that a pair of ions at m1 and m2 are two charge
states of the same ion (n1 and n2), where n2 = n1 - 1 • Assume an adduct ion weight, x.
(H+ in this example, mass = 1) • Calculate charge state for the second ion n2. If these peaks
are related, it must be an integer. n2 = (m1-x)/(m2-m1) = (694.0-1)/(867.3-694) = 4 n2 = 4
550 600 650 700 750 800 850 900
578.5
694.0 100
50
0
867.3
m1
m2
Determining Charge State of a Multiply Charged Ion
Calculate Molecular Weight, M • Use X = hydrogen = 1.0 • M = n2(m2-x) 4(867.3-1) = 3465.2
Repeat for next pair of peaks in the series • n2 = (578.5-1)/(694.0-578.5) = 5 • M = 5(694.0-1) = 3465.0
The next pair(s) form a coherent set if:
• They calculate out to an integer charge state in the proper series with earlier pairs
• The calculated MW is close enough to the MW derived from earlier pairs
550 600 650 700 750 800 850 900
578.5
694.0 100
50
0
867.3
5+ +4
6+
Determining Molecular Weight of a Multiply Charged Ion
Deconvolute Myoglobin
•Open the data file. •Integrate the data. •Obtain background subtracted peak spectrum. •Fill in Deconvolute Method parameters.
Deconvolution Parameters
Mass Range – Expected Mass Range of the Protein. Sets the x-axis of the deconvolution result display. Mass Step – Resolution of the deconvoluted protein mass range. Use limited m/z range – mass range examined in algorithm. Adduct – expected adduct, necessary for proper calculation. Proton,ammonium, potassium, sodium, proton loss. Average mass %peak – The % of the peak used for the calculation. Isotope width – Automatic, old automatic, specified
Highlight spectrum then Execute
Deconvolution Result
Mass intact protein if it were singly charged.
5x10
0
0.05
0.1
0.15
0.2
0.25
0.3
0.35
0.4
0.45
0.5
0.55
0.6
0.65
0.7
0.75
0.8
0.85
0.9
0.95
1
1.05
Abundance vs. Mass-to-Charge (m/z)
500 600 700 800 900 1000 1100 1200 1300 1400 1500 1600 1700 1800 1900
+ Scan (7.949-8.052 min, 6 scans) 4000-200myo.d
893.1872
1211.8108
1304.9478
1413.6178
1696.1247
MS Spectrum and Deconvolved Spectrum of Horse Myoglobin 6x10
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
1.1
1.2
1.3
1.4
1.5
1.6
1.7
1.8
1.9
2
2.1
2.2
Abundance vs. Mass (amu)
16750 16800 16850 16900 16950 17000 17050 17100 17150 17200 17250 17300 17350
+ Scan (474.4-483.1 sec, 8 scans) 4000-200myo.d Deconvoluted (1)
16951.6584
17050.3704
17218 125116781 0598
Theoretical: 16951.44 Observed: 16951.66 Error: -0.22 Da -12.9 ppm
MS
Deconvolved Spectrum
Export Data
Resolved Charge State Deconvolution
Agilent Users Meeting
September 17th, 2008
Converts Isotopically Resolved Charged States to Singly Charged Spectrum
Applications:
Surfactants Peptides Oligonucleotide HRES Mode Resolves to +13
Large Molecule Molecular Feature Extraction
Agilent Users Meeting
September 17th, 2008
Agilent Users Meeting
September 17th, 2008
Mass Profiling – A Powerful Workflow
Molecular Feature: • a discrete molecular entity defined by the combination of
RT, mass and response in an LC/MS or GC/MS analysis • typically represents a known or unknown compound
LC/MS or GC/MS Analysis
Identification Find molecular features
Compare sample sets
Validation
Mass Profiling: • is used to find differences between sample sets • originally developed for proteomics/metabolomics • can be applied to many analytical problems where finding
differences between samples is important • is done using statistical analysis
Mass Profiler GeneSpring MS
MFE (LC/MS) AMDIS (GC/MS)
Spectrum Mill Molecular Formulas DB Search (METLIN) Library Search
GC/MS LC/MS
QQQ MRM GeneSpring MS
Agilent Users Meeting
September 17th, 2008
MassProfiler Dependent on MFE Feature Extraction
Don’t extract all the compounds only Important ones
How to Mine for Gold… Generated a Base Peak Chromatogram Input Mass Range of Compounds Exclude Reference Ions Input m/z to 4 decimal places Set window to 20 ppm Run Standards to Determine RT Information
Agilent Users Meeting
September 17th, 2008
Good Quality *.MHD Gives Good Quality Profiler
1. Centroid versus Profile data 2. S/N versus Total Counts Typically use 100 counts Too high may not detect C13 peak Profile S/N 10 First Pass
3. Focused MHD generation Limit Retention Time Limit m/z Range
4. Ion Species (NH4 If buffer) 5. Check Box if High Salt 6. Check if Free Radicals/Quats
Agilent Users Meeting
September 17th, 2008
Good Quality *.MHD Gives Good Quality Profiler
7. Charge State = 1 to detect only singly charged species
8. Last Tabs only for Display in MassHunter QUAL
9. *.ProPos from Profile
10. *.CentPos from Centroid
Agilent Users Meeting
September 17th, 2008
Find Compounds by Molecular Feature
Large Molecule Applications
• Protein Characterization Software
– Resolved Charge State Deconvolution – Maximum Entropy Software – Large Molecule Molecular Feature Extraction – Bioconfirm B.02.00 – Peptides, Intact Proteins, Oligonucleotides
Agilent Users Meeting
September 17th, 2008
Agilent Users Meeting
September 17th, 2008
Mass Profiler B.01.00 – Highlights
• Impurity Profiling/Differential Analysis Software • Graphical Displays
– Mass vs Retention Time – Log2 Ratio (Group1/Group2) vs Retention Time – Log2 Abundance Group 2 vs Log2 Abundance Group 1 – Unique to Group 1 vs RT and Unique to Group 2 vs RT (New)
• Feature Identification (NEW) – Molecular Formula Generation and AMRT database – Single Feature or Batch Processing (summary report) – Web Internet Database Searching
Agilent Users Meeting
September 17th, 2008
Mass Profiling Software
Processed TIC Processed TIC Processed TIC Group A (Impure)
Processed TIC Processed TIC Processed TIC Group B (Control)
Aligns Data RT, Mass,
Abundance
All Data RT/Mass
What’s Changed? Mass/RT >2 Fold Change
What’s Unique!! Impurity Control
Agilent Users Meeting
September 17th, 2008
Graphical Displays
Agilent Users Meeting
September 17th, 2008
Mass Profiler – Reproducibility of Results
Spread in ChromatographyRT
Reproducible Retention Time Stability
Spread of Peak in Time
Agilent Users Meeting
September 17th, 2008
Mass Profiler – Reproducibility of Results
Outlier Sample or Feature
Agilent Users Meeting
September 17th, 2008
Blue Unique Group 2 and Red Unique Group 1
Agilent Users Meeting
September 17th, 2008
Mass Profiler – The Basics…
Double Click Mass Profiler Icon
“1” for Single Group Comparison “2” for Differential Analysis
Agilent Users Meeting
September 17th, 2008
Mass Profiler – Load MHD Files
*.MHD Data File Location
Select “Add” then add in the *.MHD files in Groups
Holding Down ctrl key select all *.MHD to add
Then Select “OK”
Agilent Users Meeting
September 17th, 2008
Set Up Mass Profiler “Method”
*.mpm is Extension of Mass Profiler Project
Agilent Users Meeting
September 17th, 2008
Mass Profiler – Edit Method
1. Limit Mass Range and RT
2. For Small Molecule Apps Multiply Charge Forbidden
3. For Biomolecules Apps Multiply Charged Required
4. Abundance to 0.1 to 0.01%
5. Number of Ions >=2
6. Isotope Pattern Matching
Agilent Users Meeting
September 17th, 2008
Mass Profiler – How to Import Exclusion List
Sort Table By Height/Volume
Right Click on Table Select “EXPORT”
From EXCEL Copy Masses From Table
Agilent Users Meeting
September 17th, 2008
Mass Profiler – Exclusion List of Masses
Paste List of Masses From Blank Into Exclude Masses
Agilent Users Meeting
September 17th, 2008
Mass Profiler – Alignment
Agilent Users Meeting
September 17th, 2008
Mass Profiler – Results Filters
Agilent Users Meeting
September 17th, 2008
Mass Profiler – Special Case – No Replicates
For Example without Replicates Min Differential Score MUST be set to ZERO before loading *.MHD Files
Agilent Users Meeting
September 17th, 2008
Feature Identification – Set Parameters for MFG
Agilent Users Meeting
September 17th, 2008
Identify Composition of Features
Right Click any feature in Graph
Right Click any feature in Table
Agilent Users Meeting
September 17th, 2008
Individual Feature Results # Compositions and Database Hits
Agilent Users Meeting
September 17th, 2008
Identification Summary Report
Agilent Users Meeting
September 17th, 2008
Feature Details
Agilent Users Meeting
September 17th, 2008
Exact Mass/Formula Database Searching
Agilent Users Meeting
September 17th, 2008
Export Feature Information
Agilent Users Meeting
September 17th, 2008
Questions and Answers
Agilent Users Meeting
September 17th, 2008
METLIN Personal Metabolite Database Software
Agilent Users Meeting
September 17th, 2008
METLIN Personal Database G6825AA
Metabolite-specific database for metabolomics research
• Database installed on a PC • Contains ~15000 compounds • Manual and batch searches
– Query based on monoisotopic mass
• Customizable – Add compounds – Assign chromatographic retention
times to metabolites – Create subset databases
• Works with other Agilent software • Cost: $3759
Agilent Users Meeting
September 17th, 2008
METLIN Personal Database Search Capabilities
• Manual searches – Search on neutral or charged
monoisotopic mass • User-specified search tolerance • Optionally restrict by retention time • Molecular formula
– Compound name – Retention time (user defined)
• Batch and manual mass searches from Agilent products – GeneSpring MS – MassHunter Qual – Mass Profiler
Agilent Users Meeting
September 17th, 2008
METLIN Personal Database Initial View
User search mode
Agilent Users Meeting
September 17th, 2008
METLIN Manual Single Mass Search
Metabolite search results
Active metabolite structure Single mass search settings
Hyperlinks to Metabolite websites
Agilent Users Meeting
September 17th, 2008
METLIN Batch Search
Load file for batch search
Active mass
Active metabolite structure
Active mass search results
Agilent Users Meeting
September 17th, 2008
METLIN Batch Summary
Active metabolite structure
Batch summary – best metabolite match
Exportable list of best metabolite search results
Agilent Users Meeting
September 17th, 2008
Creating a New METLIN Database
Create a new personal database from METLIN
METLIN is read only and can not be changed
Agilent Users Meeting
September 17th, 2008
Edit Metabolite Database – Update Entry
Add retention time to existing metabolite
Agilent Users Meeting
September 17th, 2008
METLIN Mass Search With Retention Time Adds Specificity
Adding retention time eliminates isobaric metabolites
Only 1 metabolite instead of 3 !!!!
Agilent Users Meeting
September 17th, 2008
Edit Metabolite Database – New Entry
Select add new then type new metabolite information
Agilent Users Meeting
September 17th, 2008
METLIN Search from Mass Hunter Qual
Identify compounds: search database
• Specify the METLIN database • Set search parameters
Manual search options • Select ion in spectra • Select MFE compound(s)
Batch search • Automate by Find and Identify
– Find compounds by MFE – Search database – Generate formula (optional)
Name and CAS number Database and MFG results
Set Parameters
View detailed results
Agilent Users Meeting
September 17th, 2008
METLIN Searching From Mass Profiler
Mass Profiler METLIN searching is virtually the same as METLIN • Mass Profiler can optionally add MFG
• MFG is used to assist METLIN search
• METLIN results are exportable
• Database customization is done only in METLIN
– Retention time changes
– New compounds
– New database
METLIN search integrated into Mass Profiler
Agilent Users Meeting
September 17th, 2008
Summary
• Care and Feeding of QTOF/TOF – Remember to Defrag Hard Drives Weekly – Software updates available
• Mass Profiler B.01.00 $3,150 – Differential Analysis Software – Batch Molecular Formula Generation and Database Searching
• METLIN Personal Metabolite Database B.01.00 $3,759 – Accurate Mass Retention Time Database (AMRT) – Convert *.CSV to MTL Formats – User Generated Databases (Plant Extract, Pesticides, AA’s)
Agilent Users Meeting
September 17th, 2008
Questions and Answers Thank You