5
Tips & Tricks GPC/SEC: Strategies to Save Solvent Tetrahydrofuran (THF) — a widely used solvent in gel permeation chromatography/size-exclusion chromatography (GPC/SEC) — is set to be classified as carcinogenic as proposed by the Committee for Risk Assessment (RAC). Analytical laboratories therefore need to find ways to reduce THF solvent use and waste, far beyond the demands of green chemistry. This instalment of Tips & Tricks presents different strategies to meet this goal. Jasmin Preis and Daniela Held, PSS Polymer Standards Service GmbH, Mainz, Germany. Resolution in gel permeation chromatography/ size-exclusion chromatography (GPC/SEC) depends on two parameters — the available pore volume and the dispersion in the system as indicated by the plate count. It is common practice in GPC/SEC to combine several columns in a column bank to increase resolution and molar mass separation range; however, this increases analysis time, solvent consumption, and waste. Here, we present a number of ways to perform environmentally friendly GPC/SEC. Overlaid Injection Overlaid injection introduces a sample onto the column while the previous sample is still eluting from the column. The maximum amount that the samples can be overlaid is equivalent to the interstitial volume of the column. Using overlaid injection can reduce solvent consumption by 25–35% and is an elegant way to save time. The big advantage of this is that it does not sacrifice resolution and can be performed with columns from all manufacturers. Figure 1 shows an example of an overlaid injection with UV and light scattering data for two samples. The first sample elutes with the internal flow marker around 12.5 mL (min); however, the next sample PS 1124a is already injected after approximately 8.25 mL (min). If software is used correctly, overlaid injection does not affect the data evaluation of results. Column Concepts Reducing particle size can increase resolution in GPC/SEC as in Photo Credit: Tuomas Kujansuu/Getty Images 12 Q&A: Ramos 2 News 8 Tips and Tricks 12 The Essentials 17 8 8 12 2 HPLC Beijing Event Preview 19 CHROMacademy 21 Training & Events 22 Staff 23 21 2 22 22

Tips & Tricks GPC/SEC: Strategies to Save Solvent · 2015. 9. 1. · chemistry. This instalment of Tips & Tricks presents different strategies to meet this goal. Jasmin Preis and

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Tips & Tricks GPC/SEC: Strategies to Save Solvent · 2015. 9. 1. · chemistry. This instalment of Tips & Tricks presents different strategies to meet this goal. Jasmin Preis and

Tips & Tricks GPC/SEC: Strategies to Save Solvent

Tetrahydrofuran (THF) — a widely used solvent in gel permeation

chromatography/size-exclusion chromatography (GPC/SEC) — is set to

be classifi ed as carcinogenic as proposed by the Committee for Risk

Assessment (RAC). Analytical laboratories therefore need to fi nd ways

to reduce THF solvent use and waste, far beyond the demands of green

chemistry. This instalment of Tips & Tricks presents different strategies

to meet this goal. 

Jasmin Preis and Daniela Held, PSS Polymer Standards Service GmbH, Mainz, Germany.

Resolution in gel permeation chromatography/

size-exclusion chromatography (GPC/SEC)

depends on two parameters — the

available pore volume and the dispersion

in the system as indicated by the plate

count. It is common practice in GPC/SEC

to combine several columns in a column

bank to increase resolution and molar

mass separation range; however,

this increases analysis time, solvent

consumption, and waste. Here, we

present a number of ways to perform

environmentally friendly GPC/SEC.

Overlaid Injection

Overlaid injection introduces a sample

onto the column while the previous

sample is still eluting from the column.

The maximum amount that the samples

can be overlaid is equivalent to the

interstitial volume of the column. Using

overlaid injection can reduce solvent

consumption by 25–35% and is an elegant

way to save time. The big advantage of

this is that it does not sacrifice resolution

and can be performed with columns from

all manufacturers.

Figure 1 shows an example of an overlaid

injection with UV and light scattering

data for two samples. The first sample

elutes with the internal flow marker

around 12.5 mL (min); however, the next

sample PS 1124a is already injected after

approximately 8.25 mL (min). If software is

used correctly, overlaid injection does not

affect the data evaluation of results.

Column Concepts 

Reducing particle size can increase

resolution in GPC/SEC as in Ph

oto

Cre

dit

: Tu

om

as

Ku

jan

suu

/Ge

tty I

ma

ge

s

12

Q&A: Ramos2 News8 Tips and Tricks12 The Essentials1788 122HPLC Beijing Event Preview19 CHROMacademy21 Training & Events22 Staff23212 2222

Page 2: Tips & Tricks GPC/SEC: Strategies to Save Solvent · 2015. 9. 1. · chemistry. This instalment of Tips & Tricks presents different strategies to meet this goal. Jasmin Preis and

The Column www.chromatographyonline.com

ultrahigh-pressure liquid chromatography

(UHPLC), saving time and solvent, but it

is still a concept under investigation. In

the case of macromolecules, for example,

forcing high molar mass or stiff polymer

chains through an LC system at very high

pressure can result in chain degradation

and will generate results only for the

fragments.1 Despite this, many applications

for oligomers or spherical and stable

macromolecules can be transferred to newer

solutions. Depending on the GPC/SEC

equipment two approaches can be applied: 

a) Continue to use traditional analytical

Figure 1: Overlaid injection example. Before the system peaks and the internal standard (light green triangle) of “PS 280 br” are eluted, the next sample “PS 1124 a” is already injected at 8.5 mL (blue triangle). The analysis and data evaluation of sample “PS 280 br” is not affected by that; baseline and integration limits can be set as required by national and international GPC/SEC guidelines, for example, ISO 13885/DIN 55672.

Tips and Tricks

13

Q&A: Ramos2 News8 Tips and Tricks12 The Essentials1788 122HPLC Beijing Event Preview19 CHROMacademy21 Training & Events22 Staff23212 2222

Ascentis Express UHPLC and HPLC ColumnsFaster HPLC on Any System

©2015 Sigma-Aldrich Co. LLC. All rights reserved. SIGMA-ALDRICH and SUPELCO are trademarks of Sigma-Aldrich Co. LLC, registered in the US and other countries. Ascentis is a registered

trademark of Sigma-Aldrich Co. LLC. Solutions within and BIOshell are trademarks of Sigma-Aldrich Co. LLC. Fused-Core is a registered trademark of Advanced Materials Technology, Inc.

Page 3: Tips & Tricks GPC/SEC: Strategies to Save Solvent · 2015. 9. 1. · chemistry. This instalment of Tips & Tricks presents different strategies to meet this goal. Jasmin Preis and

The Column www.chromatographyonline.com

columns with an internal diameter of

approximately 8 mm and fi lled with

particles of the same chemistry and pore

size, but a different particle size. Figure 2

compares the resolution of a polystyrene

oligomer on styrene-divinylbenzene

material with different particle sizes.

Using 3-µm particles allows separations

with the same or better resolution on just

one or two analytical columns instead

of three. This decreases the amount of

solvent and waste signifi cantly. 

b) Even more solvent can be saved when

the internal diameter of the columns is

Figure 2: Resulting molar mass distribution for a polystyrene mixture of four different molar mass calibrants including an oligomer mixture. For one styrene-divinylbenzene column (8 × 300 mm, 3-µm) a better resolution can be achieved in less time and using less solvent than when using a combination of three 5-µm styrene-divinylbenzene columns with a comparable pore size distribution.

Tips and Tricks

14

Q&A: Ramos2 News8 Tips and Tricks12 The Essentials1788 122HPLC Beijing Event Preview19 CHROMacademy21 Training & Events22 Staff23212 2222

Page 4: Tips & Tricks GPC/SEC: Strategies to Save Solvent · 2015. 9. 1. · chemistry. This instalment of Tips & Tricks presents different strategies to meet this goal. Jasmin Preis and

The Column www.chromatographyonline.com

reduced from 8 mm to around 4.6 mm;

however, the cell volume of a lot of

traditional GPC/SEC detector cells is

too large to profit from the enhanced

separation using semi-microcolumns.

While UV–vis detectors are, in general,

ready for these applications, either

upgrade of existing refractive index

detectors (RI) or purchase of new,

optimized µRIs, light scatterings

detectors, or viscometers is required.

Figure 3 shows how the excellent

resolution of microcolumns gets lost

when analytical RI detectors are used

for detection. Optimized µRIs provide

much better results.  

Intelligent Calibration 

Another way that solvent can be saved

is by using an intelligent approach to

calibration. GPC/SEC calibration involves

measuring the elution volume of several

molar mass calibration standards. It is

normally possible to inject a combination

of three to four standards in the same

injection if the standards differ significantly

Figure 3: Comparison of the chromatogram of a polystyrene oligomer separated in 5.5 mL using micro columns and a standard RI detector (blue trace) versus an optimized µRI detector with smaller cell volume (green trace).

Tips and Tricks

15

Q&A: Ramos2 News8 Tips and Tricks12 The Essentials1788 122HPLC Beijing Event Preview19 CHROMacademy21 Training & Events22 Staff23212 2222

www.gerstel.com

Sample Prep AutomationOn Your Mind? Talk to us - we are experts in automation for GC/MS and LC/MS. We listen and adapt our modular solutions to your requirements. Some examples: s�Food contaminants and allergenss�Pesticide residues (QuEChERS)s�Flavors, fragrances and off-fl avorss�Migration from packaging and E&Ls�Emissions from polymerss�Forensic analysis and metabolomics To automate your analysis, please contact us to determine the best solution for your lab. What can we do for you?

Derivatization,Addition of Standards,Dilution, Calibration

Extraction, Filtration,SPE, Evaporation (mVAP)

Thermal Desorption,Dynamic Headspace(DHS) and PYRO

MAESTRO PrepAheadProductivity

Page 5: Tips & Tricks GPC/SEC: Strategies to Save Solvent · 2015. 9. 1. · chemistry. This instalment of Tips & Tricks presents different strategies to meet this goal. Jasmin Preis and

The Column www.chromatographyonline.com

in molar mass. If the separation is

sufficient, even more standards can be

added and analyzed within the same

run. A 10- to 12-point calibration curve

can therefore be created with three

injections or less, but care is required in

choosing the correct concentration for the

different molar masses.2 Mixtures of high

quality reference materials with precise

concentrations and molar masses are

available as polymer cocktails pre-weighed

into autosampler vials, eliminating the

need to weigh in calibrants.

Detector Concepts

Multi-detection is one of the strengths

of GPC/SEC — molar mass sensitive

detectors, such as on-line viscometers

(stand-alone or as part of a triple

detection system), provide valuable

information for many applications. All

commercially available detectors apply the

same major principle: they split the sample

eluting from the column in a bridge

design and compare the viscosity (or, more

precisely, the pressure in the tubing) of

the pure mobile phase with that of the

sample. 

The viscosity of the pure solvent is

measured using either hold-up columns

or a reservoir filled with solvent. If a

detector has a reservoir, the split part of

the sample is diluted to zero concentration

in that reservoir. If a detector uses hold-up

columns, the split part of the sample is

delayed while travelling through these

hold-up columns. The delayed sample

parts then elute after the system peaks of

the sample. They are visible as negative

peaks in the delta pressure signals of

the viscosity detector. Therefore, the

amount of solvent required per sample

is much higher for viscometers with

hold-up columns than for viscometers with

reservoirs. 

Summary

There are several ways to save solvent and

to avoid waste in GPC/SEC.

t��0WFSMBJE�JOKFDUJPO�SFEVDFT�TPMWFOU�

consumption by 25–35%, without

sacrificing the resolution; however,

dedicated software is required.

t��3FEVDJOH�QBSUJDMF�TJ[F�UP�SFEVDF�UIF�

number of columns required for a

good resolution is applicable for many

applications of compact and low molar

mass macromolecules. For example,

applications in THF or other organic

solvents can be transferred from

5 µm styrene-divinylbenzene to 3 µm.

Aqueous separations can be run on

5 µm for typical water-soluble polymers

or biopolymers, or 2–3 µm for proteins. 

t��5IF�VTF�PG�NJDSPDPMVNOT�XJUI�B�

reduced internal diameter requires an

adjustment of the cell dimensions of

typical GPC/SEC detectors. For some

RI detectors a detector modification is

available. 

t��'PS�PO�MJOF�WJTDPNFUSZ�USJQMF�EFUFDUJPO�

the use of viscometers with a

reservoir instead of hold-up columns

halves the amount of solvent required

because there is no need to wait for a

negative peak.

References

1. A. Striegel, W.W.Yau, J.J.Kirkland, and

D.D. Bly, Modern Size-Exclusion Liquid

Chromatography: Practice of Gel Permeation

and Gel Filtration Chromatography

(John Wiley & Sons, New York,

USA, 2009).

2. D. Held, The Column 9(6), 9–12 (2013).

Daniela Held studied polymer chemistry

in Mainz, Germany, and works in the PSS

software and instrument department.

She is also responsible for education and

customer training.

Jasmin Preis studied polymer chemistry

in Mainz, Germany, and works in the PSS

production department. She is responsible

for custom synthesis of specialty polymers

and particles.

E-mail: [email protected]: www.pss-polymer.com

Tips and Tricks

16

Q&A: Ramos2 News8 Tips and Tricks12 The Essentials1788 122HPLC Beijing Event Preview19 CHROMacademy21 Training & Events22 Staff23212 2222