63

Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Embed Size (px)

Citation preview

Page 1: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide
Page 2: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide
Page 3: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Timing

Page 4: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

How is proper timing achieved?

• Always performed using the arterial pressure waveform as the guide

Page 5: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Arterial Pressure Waveform

PSP PSP

75% SV 25% SV

DN DN

AVO AVOIVCXAEDP AEDP

X

Page 6: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Arterial Pressure Waveform

Without IABP

With IABPAssist 1:2

Page 7: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

APSP

AUG

PSP

PAEDP

BAEDP

DN

PatientAorticEndDiastolicPressure

DicroticNotch

AssistedPeakSystolicPressure

PeakSystolicPressure

Augmentation

Balloon Aortic End Diastolic Pressure

Page 8: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Assist Ratios

1:1

1:2

1:4

Page 9: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

DN DN

Correct Inflation

Just prior to DN

Page 10: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

PSP AUG

AUG should be higher than PSP

Unless:1. Patient’s SV significantly greater

than balloon volume

2. Balloon is positioned too low

3. Hypovolemia

4. Balloon is too small

5. Low SVR

6. Improper timing

7. Partial obstruction of gas flow

Page 11: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Correct Deflation

BAEDP < PAEDPAPSP < PSP PSP

APSP

BAEDP

PAEDP

Page 12: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

May be caused by:

1. Balloon not large enough or not filled to full volume

2. Compliant aortic wall

3. Improper placement

4. Partial obstruction of gas flow

Poor afterload reduction

PSP APSP

BAEDPPAEDP

Page 13: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Timing Errors

• Early Inflation• Late Inflation• Early Deflation• Late Deflation

Page 14: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Early Inflation

Page 15: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Early Inflation

Correct Timing

AUG

DN

move inflation

Page 16: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Late Inflation

Page 17: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Late Inflation

Correct Timing

AUG

DN

move inflation

Page 18: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Early Deflation

Page 19: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

move deflation

PSP APSP

Early Deflation

Correct Timing

Page 20: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Late Deflation

Page 21: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

move deflation

BAEDPPAEDP

Late Deflation

Correct Timing

Page 22: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide
Page 23: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

NYHA class III patient, EF 25%, during off-pump CABG. HR varied between 73-139 bpm with significant changes in Pulse Pressure on a beat to beat basis.

Page 24: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

AutoPilot™ Mode

• Automatically selects the best available ECG source / lead

• Automatically selects the AP source• Automatically selects the appropriate

trigger mode• Automatically selects the optimal timing

method and settings

Page 25: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Operator Mode

Clinician selects:• ECG source / lead• AP source• Trigger mode• Timing settings

Page 26: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Both Modes

• Automatically adjust ECG gain– Unless manual gain function selected

• Automatically selects AP scale– Unless manual scale function selected

• User can select ECG source / lead and AP source

Page 27: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Triggering

Page 28: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Definition

The computer in the IAB console needs a stimulus to cycle the pneumatic system which inflates and deflates the balloon.

The trigger signal tells the computer that another cardiac cycle has begun.

Page 29: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Options

In most cases it is preferable to use the R wave of the ECG as the trigger signal.

However, the operator also has the option of using the arterial pressure waveform or pacing spikes as the trigger event.

Page 30: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Patient Signal Connections

Direct connections are always best

Page 31: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

HR 100

BP 115/50

ECG

AP

ECGDirect

“Slaved”Monitor

SpaceLabs

MarquetteTram

HP Merlon

Use 4 or 5Lead cable

Page 32: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

HR 100

BP 115/50

ECG

AP

Direct“Slaved”

Monitor

SpaceLabs

MarquetteTram

HP Merlon

AP

one of these

Fiber Optic AP connection if LightWAVE™ catheter is used

Page 33: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

FOR GOOD, CONSISTENT TRIGGERING IT IS IMPORTANTTO PROVIDE THE PUMP WITH A GOOD ECG SIGNAL

Good Choices –

Unidirectional QRS with minimal artifact

Poor Choices –

Biphasic QRS, tall T or P waves, wandering baseline, artifact present

Page 34: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

This lead will give you both triggeringand timing problems

Page 35: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

ECG Pattern

This is the preset trigger mode.

The computer analyzes the height, width (25-135 msec), and slope of a positively or negatively deflected QRS complex.

Rejection of pacer spikes is automatic.

AutoPilot™’s choice when the QRS complex is normal and the HR < 130

Page 36: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

ECG Peak

The computer analyzes the height and slope of a positively or negatively deflected QRS complex.

This may be the trigger mode of choice for wide complex or rapid rhythms.

Rejection of pacer spikes is automatic.

AutoPilot™’s choice when the QRS complex is wide, the HR > 130, or during arrhythmia when Arrhythmia Timing is OFF.

Page 37: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

AFIB

The computer analyzes the QRS complex in the same manner as Peak mode an initiates “Real-Time” timing.

Deflation is automatic when the next trigger event is identified, allowing for more consistent deflation timing when R to R intervals are irregular. Rejection of pacer spikes is automatic.

AutoPilot™’s choice the rhythm is irregular and Arrhythmia Timing is ON.

Page 38: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Timing with Irregular Rhythms

Conventional Timing

“Real-Time” Timing

Page 39: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Arterial Pressure

The computer uses the systolic upstroke of the arterial pressure waveform as the trigger signal.

This mode is an option when an ECG is unavailable or distorted.

AutoPilot™’s choice when there are no R-waves available.

HR

Page 40: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

V Pace

The computer uses the ventricular spike as the trigger signal. This mode can be used with ventricular or AV paced rhythms.

Must be 100% paced.

AutoPilot™’s uses this mode when there are no R-waves or AP waveforms present, however, there are V or AV pacer spikes.

Page 41: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

A Pace

The computer uses the atrial pacing spike as the trigger signal. This mode can be used with atrially paced rhythms only.

Must be 100% paced.

AutoPilot™’s choice when the ECG is intermittent and pacer spike to R wave is > 100ms.

Page 42: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Internal

The balloon inflates and deflates at a preset rate regardless of the patient’s cardiac activity.

This mode is only to be used when there is no cardiac output and no ECG.

Preset rate is 80 bpm; can be varied between 40 to 120.

Page 43: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Cardiac Arrest

What do you do with the IABP?

Page 44: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Helium Delivery

Page 45: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Pneumatic Systems

2 types of gas delivery systems:

Vacuum / compressor system

Bellows / stepper motor system

Page 46: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

12

3

4

5

6

7

8

9

1. Zero Baseline

2. Balloon Pressure Baseline

3. Rapid Inflation

4. Peak Inflation Artifact

5. Plateau Pressure

6. Rapid Deflation

7. Deflation Artifact

8. Return to Baseline

9. Duration of Balloon Cycle

Balloon Pressure Waveform

Page 47: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

0

transducer

pump helium

BPW

Stepper motor / bellows system

Helium Fill Pressure

Page 48: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

transducer

pump helium

Stepper motor / bellows system

Balloon Inflation

0

BPW

Page 49: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

transducer

pump helium

Stepper motor / bellows system

Balloon Deflation

0

BPW

Page 50: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Heart Rate Variations

250

0

250

0

Tachycardia Bradycardia

Page 51: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

BPW in Irregular Diastole (Afib)

250

0

Page 52: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Pressure Variations

250

0

250

0

Hypertension Hypotension

Page 53: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Comparison of Augmented AP waveform and Plateau of Balloon Pressure waveform

Page 54: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

0

250

50

150

IAB

fu

lly

infl

ated

aort

a

aort

a

transmembranepressure

Note that the endpoint of the BPW plateau should be equal or up to 20 mmHg higher than the Augmented Pressure.

Intra-aortic Relationship of Inflated Balloon Catheter and Vascular Pressure

Page 55: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Troubleshooting

Gas Surveillance Alarms

Page 56: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Purge Failure

Pump did not fill adequately with helium to establish the balloon pressure waveform baseline

250

0

Verify:

Helium tank not empty

Catheter connections intact

Trigger present

Page 57: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Check for:

1. Partially wrapped balloon

2. Kinked catheter

High Baseline

250

0

Page 58: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Helium Loss

250

0

Check for:

1. Leak in tubing and connections

2. Blood in catheter tubing

3. Kinked catheter

4. Ectopic beats

Page 59: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Classic BPW appearance of aKinked Catheter

Page 60: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Internal Leak Test

BALLOON

0

0

pass

fail

Observe BPW baselinefor 1 - 2 minutes

•Turn alarms OFF

•Select INTERNAL trigger

Page 61: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

0

BPW

IAB too Large for Aortic Environment

transducer

pump helium

Page 62: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide

Hands on review of the AutoCAT® 2

WAVETM

Page 63: Timing How is proper timing achieved? Always performed using the arterial pressure waveform as the guide