TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

Embed Size (px)

Citation preview

Page 1: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

TIMERS

Page 2: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

BASIC X-RAY SCHEMATIC

Page 3: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

TIMER CIRCUIT:

• IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM.

TIMER

Page 4: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

TIMER

• A MECHANICAL OR ELECTRONIC DEVICE WHOSE ACTION IS TO MAKE OR BRAKE THE HIGH VOLTAGE ACROSS THE X-RAY TUBE

Page 5: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

TYPES OF TIMING CIRCUITSTIMERS:

• MECHANICAL

• SYNCHRONOUS

• ELECTRONIC

• mAs

• BACKUP

• AUTOMATIC EXPOSURE CONTROL

Page 6: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

MACHANICAL TIMER

• OPERATE BY CLOCKWORK.• SPRING MECHANISMS IS WOUND UP.• DURING EXPOSURE THE SPRING UNWINDS

AND TERMINATES EXPOSURE.

Page 7: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

SYNCHRONOUS TIMER• IN SYNCH WITH FREQUENCY OF THE A/C

CURRENT (60 Hz in the US).• MINIMUM EXPOSURE TIME IS 1/60 s.• TIMING INTERVALS INCREASE BY 1/60 s.• MUST BE RESET AFTER EACH EXPOSURE

Page 8: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

ELECTRONIC TIMER• THE MOST SOPHISTICATED.• THE MOST ACCURATE• IT’S CIRCUITRY CONSISTS OF

THYRATRON TUBE OR THYRISTOR.

• ALLOW WIDE RANGE OF TIME INTERVALS.

• ALLOW SERIAL EXPOSURES.• NOWADAYS CONTROLLED BY

MICROPROCESSOR

Page 9: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

THYRATRON CIRCUIT

Page 10: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

mAs TIMER

• MONITORS THE PRODUCT OF mA & EXPOSURE TIME AND TERMINATES EXPOSURE WHEN THE DESIRED mAs VALUE IS ATTAINED.

Page 11: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

BACKUP

• SET TO TERMINATE EXPOSURE IN THE EVENT OF FAILURE IN THE AEC CIRCUIT.

• MAX. mAs ALLOWABLE – 600 mAs

UNLESS TUBE POTENTIAL IS LESS THAN 50 kVp – 2,000 mAs IS THE LIMIT.

Page 12: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

AUTOMATIC EXPOSURE CONTROL

Automatic exposure control devices can assist the radiographer in producing consistent radiographic images from patient to patient, regardless of size or presence of pathology. The advantages of this consistency are numerous and include: decreased repeat rate; decreased patient exposure; and increased department efficiency. The most important benefit being decreased repeat rate

Page 13: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

When an AEC device is used to terminate an exposure, the technologist sets the kVp and mA, but the time of the exposure is automatically determined by the machine. The AEC device differs from a manual timer because the AEC does not stop the exposure until the film has reached an appropriate density. Unlike manual timers, which simply stop the exposure after the preset time has elapsed.

Page 14: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

A major benefit of the AEC device is its ability to consistently obtain accurately exposed radiographs, even in the presence of pathology. While manual timers terminate the exposure at the preset time, regardless of pathology or achievement of proper film density. The following example demonstrates the difference

Page 15: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

AUTOMATIC EXPOSURE CONTROL (AEC)

• FLAT IONIZATION CHAMBER

• PHOTOTIME

Page 16: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

IONIZATION CHAMBER AEC

X-RAY TABLE

X-RAY TUBE

BUCKY

Page 17: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

IONIZATION CHAMBERS ARE SITUATED

• BETWEEN THE GRID MECHANISM & THE CASSETTE

Page 18: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

CHAMBERS

Page 19: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

CHAMBER SELECTION ON THE CONSOLE

DETECTOR SELECTION

    Along with proper

Page 20: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

AEC

Page 21: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

IONIZATION CHAMBER AEC

X-RAY TABLE

X-RAY TUBE

CASSETTE

SCREEN

Page 22: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

PM TUBELIGHT

ELECTRONS

Page 23: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

PHOTOTIME MECHANISM IS SITUATED

• BEHIND THE CASSETTE

• CASSETTE NEEDS SPECIAL DESIGN

IT HAS OPENING IN THE BACK

Page 24: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

PHOTOTIME CASSETTE

Page 26: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

Even though the ionization chamber and the phototimer operate differently, they both have the same function: convert radiation into an electrical signal which will be used to automatically stop the exposure when the film has reached the proper density

Page 27: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

FACTORS INFLUENCING AEC

• PATIENT POSITIONING• DETECTOR SELECTION

Page 28: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

POSITIONING!!!!

• AEC USES DETECTORS TO RECEIVE THE RADIATION THAT PASSED THROUGH THE PATIENT WE MUST POSITION THE ANAOMY OF INTEREST DIRECTLY ABOVE THE DETECTORS.

Page 29: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER
Page 30: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

Without a technologist who is very knowledgeable about anatomy and positioning, automatic timers are worthless. In fact, they may actually decrease department efficiency because of the increased amount of repeat radiographs that will result if used improperly.

Page 31: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

DETECTOR SELECTION

Page 32: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

LAT. L-SPINE

Page 33: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

C-SPINE

Page 34: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

C-SPINE -AP

Page 35: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

SHOULDER

Page 36: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

KNEE

Page 37: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

HIP

Page 38: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

PELVIS

Page 39: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

ABDOMEN

Page 40: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

CHEST PA

Page 41: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

LAT. CHEST

Page 42: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

DENSITY SETTING

• +4

• +3

• +2

• +1

• 0

• -1

• -2

• -3

• -4

• +100

• +75

• +50

• +25

• 0

• -25

• -50

• -75

• -100

D. SETTING % DENSITY CHANGE

Page 43: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

MINIMUM RESPONSE TIME

• SHORTEST EXPOSURE TIME POSSIBLE

Page 44: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

EFFECTS OF mA, kVp, SID ON AEC

• mA AEC TIME

• kVp AEC TIME

• SID AEC TIME

Page 45: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

PROBLEMS WITH AEC

• PATIENT PATHOLOGY

• PROSTHETIC DEVICE

• COLLIMATION

• STROBOSCOPIC EFFECT

• WRONG IMAGE RECEPTOR

Page 46: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

CERTAIN PATHOLOGIES ALTER AEC READING

Page 47: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

PROSTHESIS GIVES FALSE AEC READING

Page 48: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

NOT ENOUGH COLLIMATION ALLOWS TOO MUCH

SCATTER REACHING THE DETECTOR AND PREMATURE

TERMINATION OF THE EXPOSURE TAKES PLACE

Page 49: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

STROBOSCOPIC EFFECT

• TOO SHORT OF THE EXPOSURE TIME IS SELECTED BY THE AE, SHORTER THAN THE SPEED OF THE OSCILLATION OF THE GRID AND IT FREEZES THE GRID MOVEMENT.

Page 50: TIMERS. BASIC X-RAY SCHEMATIC TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER

WRONG SPEED OF IMAGE RECEPTOR!!!