16
S cribne r Associates i n c o r p o r a t e d Through-thickness Membrane Conductivity Measurement for HTM Program: Issues and Approach Kevin Cooper Scribner Associates, Inc. DOE HTMWG Meeting 14 September 2006 Subcontractor to FSEC/UCF DOE Award No. DE-FC36-06GO16028

Through-thickness Membrane Conductivity Measurement for ... · Through-thickness Membrane Conductivity Measurement for HTM Program: ... MFC MFC TC MFC MFC TC Top ... Through-thickness

  • Upload
    lyquynh

  • View
    219

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Through-thickness Membrane Conductivity Measurement for ... · Through-thickness Membrane Conductivity Measurement for HTM Program: ... MFC MFC TC MFC MFC TC Top ... Through-thickness

ScribnerAssociates i n c o r p o r a t e d

Through-thickness Membrane Conductivity Measurement

for HTM Program: Issues and Approach

Kevin CooperScribner Associates, Inc.

DOE HTMWG Meeting

14 September 2006

Subcontractor to FSEC/UCFDOE Award No. DE-FC36-06GO16028

Page 2: Through-thickness Membrane Conductivity Measurement for ... · Through-thickness Membrane Conductivity Measurement for HTM Program: ... MFC MFC TC MFC MFC TC Top ... Through-thickness

ScribnerAssociates i n c o r p o r a t e d

Through-thickness membrane conductivity: Motivation• Development of new ionomers and novel proton conductors is

focused on improving membrane conductivity for challenging fuel cell operating needs– i.e., High conductivity at high T and low RH– Allows for system simplification to meet commercial applications

• There is a need for the ability to rapidly and accurately measure key membrane properties as a function of T and RH

• Through-thickness membrane resistance (or conductance) is a key property of interest for fuel cells

• Traditionally measured with catalyzed membrane (MEA) in 2-probe configuration (i.e., a single-cell fuel cell)– Must account for

• Rcontact (current collector/GDL, GDL/MEA)• Relectronic (CC, GDL, leads)• Rinterface (electrolyte/catalyst)

– Time consuming to make MEA, build FC and test

Page 3: Through-thickness Membrane Conductivity Measurement for ... · Through-thickness Membrane Conductivity Measurement for HTM Program: ... MFC MFC TC MFC MFC TC Top ... Through-thickness

ScribnerAssociates i n c o r p o r a t e d

Through-thickness membrane conductivity: Objective

• Develop hardware and protocols to measure through-plane conductivity of PEMFC membranes

• Ideally, test device will – Use bare (uncatalyzed) membranes - 10 to 200 μm thick– Employ a 4-probe technique to eliminate confounding Ri– Operate over a wide range of atmospheric conditions

• T, specific humidity / RH, pressure– Measurement would be fast ~ 15 min / test condition– Hardware and protocols would be robust

• accurate, repeatable, reliable• user-friendly

• Versatile R&D tool to support development of HTMs

• HTM Program Year 1 Milestone: Prototype measurement system

Page 4: Through-thickness Membrane Conductivity Measurement for ... · Through-thickness Membrane Conductivity Measurement for HTM Program: ... MFC MFC TC MFC MFC TC Top ... Through-thickness

ScribnerAssociates i n c o r p o r a t e d

Nomenclature

• Conductivity = σ (S/m)• Conductance = κ (S)• Resistivity = ρ (Ω-m)• Resistance = R (Ω)

• Resistance and Conductance are extrinsic properties dependent on specimen geometry

• Resistivity and conductivity are intrinsic material properties– e.g., density, elastic modulus, yield strength– independent of specimen geometry– orientation dependence for anisotropic

materials– f (microstructure, T, humidity, …)

• To obtain σ, we need to know V, I, L and A at the time of the measurement

RIV =

κ/1ALρR ==

A

I

L

V

ρ/1σ =

AL

VIσ =

S = siemens = 1 / Ω = mho

Page 5: Through-thickness Membrane Conductivity Measurement for ... · Through-thickness Membrane Conductivity Measurement for HTM Program: ... MFC MFC TC MFC MFC TC Top ... Through-thickness

ScribnerAssociates i n c o r p o r a t e d

aqueouselectrolyte

membrane clampedbetween flanges

Ionic conductivity of membrane is always determined by measuring its resistance

V

I

Rmembrane = R′cell - Rcell

R′cell

(w/ membrane)

Rcell(w/o membrane)

• 2-probe measurement includes contributions from

• leads, electrodes• contact• aqueous electrolyte• interfacial

• bringing electrodes in contact with membrane eliminates ionic resistance of electrolyte (but not others)• concept behind Mercury cell

V

PS I

Mercury Mercury

= Rcell

Page 6: Through-thickness Membrane Conductivity Measurement for ... · Through-thickness Membrane Conductivity Measurement for HTM Program: ... MFC MFC TC MFC MFC TC Top ... Through-thickness

ScribnerAssociates i n c o r p o r a t e d

4-probe• sense electrodes in Luggin probes monitor potential drop across membrane• minimizes Rcell• better

But aqueous electrolyte and mercury are not relevant test environments for PEM FC because these are not the conditions under which the membrane operates

Measurements must be made in humid gas at relevant temperatures and pressures

V

PS

Ionic conductivity of membrane is always determined by measuring its resistance

Page 7: Through-thickness Membrane Conductivity Measurement for ... · Through-thickness Membrane Conductivity Measurement for HTM Program: ... MFC MFC TC MFC MFC TC Top ... Through-thickness

ScribnerAssociates i n c o r p o r a t e d

In-plane• most readily adopted method because dimensions are reasonable to work with

• small A + large L large VIR for small I good• relatively easy measurement

• 4-probe technique easy to implement

• in-plane σ not through-thickness σ• anisotropic materials• suitable for rapid evaluation of new materials, trending

• surface vs. bulk conductivity

Electrodes must be in contact with membrane when measuring conductivity in gaseous environment

PS

A

V

L

AL

σIVIR =

Page 8: Through-thickness Membrane Conductivity Measurement for ... · Through-thickness Membrane Conductivity Measurement for HTM Program: ... MFC MFC TC MFC MFC TC Top ... Through-thickness

ScribnerAssociates i n c o r p o r a t e d

Measurement challenge: Through-thickness membrane resistance in gaseous environments2-probe method• suffers from very high contact resistance which confounds measurement of very small membrane resistance• Alberti et. al. used up to 5-layer stacks of membrane in order to determine conductivity of low-resistance membranes

V

PS

G. Alberti, et. al., J. Membrane Science 185 73 (2001)

Page 9: Through-thickness Membrane Conductivity Measurement for ... · Through-thickness Membrane Conductivity Measurement for HTM Program: ... MFC MFC TC MFC MFC TC Top ... Through-thickness

ScribnerAssociates i n c o r p o r a t e d

Measurement challenge: Through-thickness membrane resistance in gaseous environments

Making a 4-probe measurement in thin electrolyte presents a challenge

PS

How to place the V-sense electrodes “in” the electrolyte without disturbing the current distribution?

V

?

Page 10: Through-thickness Membrane Conductivity Measurement for ... · Through-thickness Membrane Conductivity Measurement for HTM Program: ... MFC MFC TC MFC MFC TC Top ... Through-thickness

ScribnerAssociates i n c o r p o r a t e d

Measurement challenge: Through-thickness membrane resistance in gaseous environments

• sandwich and hot-press V-sense electrodes between layers of membrane

• not feasible for this project

V

PS

S. Ma, A. Kuse, Z. Siroma and K. Yasuda, "Measuring Conductivity of Proton Conductive Membranes in the Direction of Thickness," Espec Technology Report No. 20 (ESPEC) 12-20 (Year?).

Page 11: Through-thickness Membrane Conductivity Measurement for ... · Through-thickness Membrane Conductivity Measurement for HTM Program: ... MFC MFC TC MFC MFC TC Top ... Through-thickness

ScribnerAssociates i n c o r p o r a t e d

Source electrode (ring)V1 sense (disk)

Membrane

V2 senseSource electrode

Ring electrode

Insulator

V2

V1

Pt wire

Plan View

Side View

Side View

4-probe configuration #1: Concentric disk-ring

Page 12: Through-thickness Membrane Conductivity Measurement for ... · Through-thickness Membrane Conductivity Measurement for HTM Program: ... MFC MFC TC MFC MFC TC Top ... Through-thickness

ScribnerAssociates i n c o r p o r a t e d

Conclusion: concentric disk-ring concept not feasible - electrode geometry required are not possible with traditional or micro-machining, or lithography

Electrode configuration #1: Modeling Results

r, mm

0.0 0.2 0.4 0.6 0.8 1.0

z, μ

m

0

20

40

60

80

100

-0.016-0.012-0.008-0.004 00.0040.0080.0120.016

E, V

σ = 0.1 S/cmL = 100 μmI ~ 10 mA (1 A/cm2)

V1

V2

ΔE = V1 – V2 ~ 0 V

Page 13: Through-thickness Membrane Conductivity Measurement for ... · Through-thickness Membrane Conductivity Measurement for HTM Program: ... MFC MFC TC MFC MFC TC Top ... Through-thickness

ScribnerAssociates i n c o r p o r a t e d

For fuel cell, we want small ohmic voltage loss through the membrane

• i.e., small R– small L (thin)– big A– big σ

• What is “good” for fuel cell can make it difficult for property measurement– Large I required for

reasonable ΔVIR

AL

σIRIV ==

Electrolyte thickness, μm

0 25 50 75 100

Cur

rent

, A

0

2

4

6

8

10

σ = 0.001 S/cmσ = 0.01 S/cmσ = 0.1 S/cm

Current vs. Membrane Thickness and σfor ΔVIR = 0.1 V

Increasing σ

A L

Page 14: Through-thickness Membrane Conductivity Measurement for ... · Through-thickness Membrane Conductivity Measurement for HTM Program: ... MFC MFC TC MFC MFC TC Top ... Through-thickness

ScribnerAssociates i n c o r p o r a t e d

Measurement challenge: Wide range of conductivity and membrane thickness

• To accommodate membranes of various thicknesses and broad range in σ, the measurement system must be able to accommodate large changes in V and/or I

Electrolyte thickness, μm

0 25 50 75 100

Cur

rent

, A

0

2

4

6

8

10

σ = 0.001 S/cmσ = 0.01 S/cmσ = 0.1 S/cm

Current vs. Membrane Thickness and σ for ΔVIR = 0.1 V

Increasing σ

σ = f (RH, T)

Page 15: Through-thickness Membrane Conductivity Measurement for ... · Through-thickness Membrane Conductivity Measurement for HTM Program: ... MFC MFC TC MFC MFC TC Top ... Through-thickness

ScribnerAssociates i n c o r p o r a t e d M. Murthy, Proton Conducting Membrane Fuel Cells III, ECS PV02-31, 257 (2005)

Measurement challenge: Rapid relative humidity cycling in both directions

Approach: Wet + dry gas mixingprecise control of RHrapid ↑↓ in RH Scribner has experience with this

systemmore complex and $ than single

MFC

RH ~ 20% to 100 %T ambient to > 120 oC

PoImA

MFC

MFC

TC

MFC

MFC

TC

TopGas In

TopGas Out

BottomGas OutBottom

Gas In

F

Membrane

Sealing Gasket

PoImA

MFC

MFC

TCTC

MFC

MFC

TCTC

TopGas In

TopGas Out

BottomGas OutBottom

Gas In

F

Membrane

Sealing Gasket

Page 16: Through-thickness Membrane Conductivity Measurement for ... · Through-thickness Membrane Conductivity Measurement for HTM Program: ... MFC MFC TC MFC MFC TC Top ... Through-thickness

ScribnerAssociates i n c o r p o r a t e d

Path forward …

• 4-terminal electrode concept developed• approach validated with 2-D model• electrodes designed, fabricated and under proof-of-concept testing

• Cell hardware with electrode / membrane clamping mechanism being fabricated

• pressurized operation feasible for >100 oC, high RH test capability

• Gas humidification system being evaluated• accuracy • stability – within a day• repeatability – day to day