Upload
lengoc
View
212
Download
0
Embed Size (px)
Citation preview
1
Universität Duisburg-Essen 4. Semester Fak. 5 , IVG Fachgebiet Thermodynamik Laborleiter: Dr. Siddiqi
Thermodynamics Lab Humidity Measurement
Humidity Measurement
Introduction Air comprises a mixture of gases of which nitrogen makes up more than 78%, oxygen 21%
and carbon dioxide and the inert gases (argon, neon, krypton, helium etc.) the remainder.
These are known as the dry gases of the atmosphere. In addition to these dry gases, the air
also contains varying amounts of water vapor. At normal temperatures and pressures water is
able to exist in both a liquid and a gaseous (or vapor) form, but it cannot be treated in the
same way as the other gases of the atmosphere because its quantity, and hence proportion are
continually varying. Water vapor is said to be associated with dry air and the more vapor
associated with the dry gases, the more humid the air. Humidity measurements are important
not only in meteorology but also in pulp and paper industry, drug and food storage and
delivery, film industry etc. Hygroscopic materials draw water from their surroundings and
can form solution (e.g., salt, sugar). Strongly hygroscopic materials are used directly for the
drying of substances (e.g., gases). The humidity can have dramatic effect on our perception of
comfort. It can affect the well-being of a person or even cause diseases.
A measure of the water vapor content in the air is the absolute humidity (Feuchte). It shows
how many kg of water vapor are present in unit volume (1 m3) of the gas mixture. Depending
on the state of the gas mixture (dry or wet) taken as reference one defines:
a) Related to the unit volume of the wet gas mixture
air wetmOH kg
Vm = f 3
2W
′′
(1a)
mw = Mass of water vapor
= Volume of wet gas mixture V ′
(slash means here that it is the wet gas mixture)
is also called the density of the water vapor. f ′
b) Related to the unit volume of the dry gas mixture
2
airdry mOH kg
Vm = f 3
2w
(1b)
These both humidity values depend on temperature via the temperature dependence of V
under a given atmospheric pressure. If one does not relate the water vapor content to the unit
volume but to unit mass of dry air, one gets temperature independent value, known as
moisture content or humidity ratio.
airdry kgOH kg
mm = x 2
L
w
(2)
Dry air can absorb water vapor in a certain state (at certain pressure and temperature) up to a
certain amount. The absorption capacity grows with a rising temperature. If the maximum
absorption capacity is reached at a certain temperature, it is then „saturated air“.
Another way to express the absorption capacity of wet air in a certain state (1) (p1, t1, x1), is
the relative humidity ϕ. It is defined as the ratio of the available absolute humidity in state (1)
humidity , to the maximum possible humidity at the same temperature t1, i.e. the
absolute humidity in the saturation state:
´f ´sf
)t(’ f)/t(f = 1s1′ϕ (3)
1
S
x
h 1+x
t1 t =const.
1
ϕ =con
st.
ϕ=1
3
Another possibility to describe the degree of the saturation of moist air exists in the
comparison of the actual air temperature t1 and the so-called dew point temperature
(Taupunkt) tTpkt. The dew point temperature is the temperature at which the first drop of
water condenses when we cool the air in state (1) at constant water content x1. If the air is
cooled below the dew point temperature to t2, the air loses water vapor. The mass of the water
lost (due to condensation) is . The information (expression) of the dew point
temperature is well suited in the climate technology because it gives the temperature up to
which a test atmosphere (room) can be cooled without any condensation takes place.
)x - x( m = m 21Lw
1
2
x
h 1+x
tTpkt
ϕ=1
t2
x2 x1
Taupunkt
The third possibility to describe the state of moist air at a certain temperature t1, makes use of
the cooling limit temperature (Kühlgrenztemperatur) tF. The cooling limit (Kühlgrenze) lies in
h, x- diagram on the intersection of the extended nebulous isotherms of the state point (1) of
the air to and the line ϕ = 1. The temperature at this point is the temperature which the water,
initially at any temperature, attains if it is brought in contact with unsaturated air of state (1).
4
1
x
h 1+x
t1ϕ=1
tF
x1
Kühlgrenze
For the representation of the concentration of water vapor in the gas mixture the mass fraction
mmw , the mole fraction
nnw and the partial pressure pD = p
nnW are commonly used. In general
p = pG + pD is valid for every vapor gas mixture. In the special case of vapor gas mixture if
we apply the ideal gas law, then
, VT Rm = p WWDand
VT Rm = p GGG (4)
and it means that every component behaves in such a way as if it were alone in the volume V
(Dalton´s law). RG and RL are the specific gas constants for dry gas mixture and water vapor.
The partial pressure of the water vapor at saturation is . It can be shown that for total
pressure p <10 bar the saturated partial pressure (t) can be approximated (better than 1%)
with the vapor pressure of pure water at temperature “t” which is given by the vapor
pressure curve as a temperature function (Table 1). The water vapor pressure in
unsaturated air can be calculated from pD = pWS(tTpkt), as
pDS
pDS
pWS
Dp
(t) f p n
n = pDW ≠ remains constant
at dew point temperature and at dew point it is saturated.
With equation (4) it follows for equation (3)
(t)p)t(p
= (t)p
p = W
TpktW
D
D
S
S
S
ϕ
The difference of both pressure pD and is called saturation deficit Δp. pDS
5
6
Lab Experiment
Comparative measurements of the humidity in a moist chamber should be made with different
measuring methods and the values will be represented in different humidity units. Two
different climates are to be examined.
1.Object of the Experiment The humidity of air in chamber should be measured with different measuring instruments.
The following instruments will be used to measure the humidity. The working principle, and
the advantages and disadvantages will be described.
- Humidity indicator
- Hair hygrometer as a polymeter
- Assmann Aspirations-Psychrometer
- LiCl dew point measuring instrument
- Mirror dew point measuring instrument.
The different humidity values are determined by suitable conversions, tables and diagrams.
2. Working principles
2.1 Humidity indicator
The humidity indicator is a chemically preserved paper strip which reacts like litmus paper
(used for pH) by changing its color with a change of the relative humidity. With increasing
humidity the color segments will become pink (rosa), with increasing dryness they becomes
blue (blau). This process can be repeated many times and the humidity value can be read in %
of relative humidity (relative Feuchtigkeit) on a printed scale (kept beside). The both strips
used for the experiment are shown in Fig. 1(Abb. 1).
The advantage of these measuring stripes is primarily, the low price and lower place
requirement. For a continuous view control with low demand for accuracy (5 - 10%) the
indicators can be used even in very small places (self sticking strips, basic implementation: 50
x 19 mm). Their use is not possible in air containing ammonia and with the appearance of
liquid water on it’s surface.
30 40 50 60 70 80 90 100
trocken normal feucht
20
Feuchtigkeit in %Ablesung im Grenzfeld rosa-blau
Abb. 1 Zwei Typen von Feuchte-Indikatoren mittels Farbumschlag
Ablesung der relativen Feuchte in %am Übergang rosa-blau, oder wo Strei-fen der Untergrundfarbe entspricht
Feuchte-Indikator
20
30
40
50
60
70
8075
65
55
45
35
25
normal
trocken
feucht
10
0
10
20
30
4060
40
20
10
5
2
SkalaTemperatur
in °C
SkalaSättigungsdruck
in Torr
Haarbündel
3015 10
5
0
10080
503010
0
°C
%rel. Feuchte
Abb. 2 Haarhygrometer in Polymeterform
2.2 Hair polymeter
The hair-polymeter represents the easiest and most inexpensive instrument for hygrometric
measurements. In hygrometric procedures the physical properties of substances which are
dependent on humidity are used for the measurement. These can be mechanical (length +
twist) and electrical properties (surface resistance, conductivity) as well as heat evolution,
diffusion, infrared absorption and the electrolysis.
For the mechanical properties the substances which change their length with the humidity
seem to be a simple and good choice. It is also required that the change in length should be
reproducible, i.e. it should come to the initial value when the humidity changes back to its
initial value. Hair fulfills this requirement and is also inexpensive. It has a thermal coefficient
of expansion of about 3.4 10-5 per degree which is almost temperature independent. It is,
therefore, used for the purpose of making a device to measure the humidity. The working
principle of hair hygrometer is that the length of hair changes with the increase or decrease in
humidity. The hair hygrometer (Fig. /Abb. 2) uses hair under tension to measure humidity.
When relative humidity increases, hair becomes longer, and when it drops, hair becomes
shorter. This instrument uses strands of human or horse hair attached to levers that magnify 7
8
small changes in length. Hair hygrometers are not as accurate as their counterparts and
register significant errors at very high and very low relative humidity.
The disadvantages of the method are:
- The accuracy of the measurement is approx. 3 - 5% in the relative humidity value. It shows
an additional hysteresis effect of 8%.
- It can be used in the relative humidity range of 30 – 90% and below 70 °C (at this
temperature the hair are affected by the heat). Also the use below - 30 °C is not recommended
as the reading is not shown promptly (it is sluggish).
- To avoid false measurements, the hair must be "regenerated" in 2-weeks cycle. It is done by
keeping the instrument in saturated atmosphere for ½ hour.
The hair hygrometer used by us is in the form of a Polymeter. So besides hair hygrometer it is
also provided with a thermometer which shows the surrounding temperature as well as the
saturation pressure in Torr at this temperature. Other directly readable quantities are:
- the relative humidity (Feuchte) on the lower scale of the round indicator
- the difference Δt between ambient temperature tU and the dew point temperature tTpkt on the
upper scale of the round indicator. In this case the left point of the triangle gives the value at
+20 °C surroundings temperature, the middle point the value at +10 °C surroundings and the
right point at 0 °C surroundings (at other surrounding temperatures the values must be
estimated accordingly).
- the saturation water vapor pressure (in Torr) is shown on the right side of the thermometer
besides the ambient temperature.
From it one can find immediately.
- the dew point temperature tTpkt = tU - Δt
- the partial pressure of water vapor pD = φ pDs [equation (5)]
- the absolute humidity ~ partial pressure of water vapor pD in Torr. This is a coarse
approximation in the range of -30 °C to + 45 °C (see Fig./Abb. 3). The exact values must be
taken from tables (Table 1) or diagrammes (Fig. 4a and Fig./Abb. 4b).
1000
400
200
1006040
20
1064
2
1-20 0 20 40 60 80 100
Dam
pfdr
uck
mbar
Temperatur
°C
Abb. 3 Dampfdruckkurve des Wassers
üb. unterkühltem Wasser
50
40
30
20
10
00-10 10 20 30 40°C
Gas-Temperatur
abso
lute
Feu
chte
f'g
/
m
Luft
H O 2
3
100
90
80
70
60
50
40
30
20
Abb. 4a Abhängigkeit von relativer, absoluter Feuchte un Temperatur im Bereich von -10 °C bis +40 °C
φ
9
500
°CGas-Temperatur
abso
lute
Feu
chte
f'g
/ m
Luft
H O 2
3
450
400
350
300
250
200
150
100
50
030 40 50 60 70 80 90 100
80
70
90100
60
50
40
30
20
10
φ
Abb. 4b Abhängigkeit von relativer, absolute Feuc und Temperatur im Bereich von 30 - 100 °
10
2.3 Aspiration-Psychrometer (Assmann)
The psychrometric measuring procedure, also called “Dry and wet thermometer procedure”,
uses the cooling effect which is observed when an unsaturated gas stream flows over a
moistened thermometer thereby evaporation. The process which takes place is a combined
heat and mass transfer process.
Figure 5 (Abb. 5) shows the basic construction of an Aspiration-Psychrometer.
6 0
05
04
03
02
01
0
01
6 0
05
04
03
02
01
0
01
Uhrwerkmit Ventilator
trockenesThermometer
feuchtesThermometer
Luft
Abb. 5 Prinzip-Aufbauskizze eines Aspirations- Psychrometers nach Assmann
11
It consist of two thermometers. The bulb of one thermometer is kept wet (by means of a thin,
wet cloth wick) so that the cooling that results from evaporation makes it register a lower
temperature than the other (called dry bulb) thermometer. The flowing air (at the inlet of the
measuring device) has the water vapor partial pressure pD and the temperature tu , which is
shown by the dry thermometer. If pD is smaller than the saturation pressure pS at this
temperature, the water evaporates from the wet bulb wick and saturates the passing gas
stream. The needed energy for the vaporization of water is supplied by the internal energy of
the gas(unsaturated air), the thermometer bulb and the wet wick. Through this, the
temperature of the wet thermometer drops below that of the gas to a typical value, the cooling
limit temperature tF. This depends on the saturation deficit of the gas. After the equilibrium
temperature is reached the necessary heat energy for the continuous vaporization of water is
taken only from the gas stream.
We equate the heat amount which is taken away from the gas by cooling to the heat amount
necessary for the evaporation and get the expression for the water vapor partial pressure pD:
h Rt c p
- p = pV
*p
FD ⋅Δ⋅⋅
(6)
where p = Barometer pressure
Δt = (tu - tF) psychrometric temperature difference
tu = Temperature of the measuring air
pF = Water vapor saturation pressure at tF
cp = spec. heat capacity of the gas at constant pressure
R* = RL/RW , RL is the specific gas constant for dry air and RW for water vapor
hv = specific enthalpy of vaporization for water
For technical measurements we mostly use (using the empirical approximation formula pD =
pF - p ⋅ A ⋅ Δt, with A = Psychrometer constant) the formula from SPRUNG (1888) for the
measurements in air, which simplifies equation (6) to:
,t 755
p C150 - p = p FD Δ⋅°. (7)
12
13
where the constant 0.5 refers to Assmann aspirations psychrometer with an air speed > 2.0
m/s. For practical evaluation mostly the tables, graphical representations or special slide
rules or discs are used. Thus the absolute humidity can be determined by means of Table 1
with the help of the water vapor partial pressure, or one can get the relative humidity
directly from the psychrometric temperature difference (Δt) using Fig./Abb. 6 ("Graphical
Psychrometeric Table") or with the help of the humidity slide rule. Very accurate results are
read from the Psychrometer Table provided by the manufacturer, because it uses the
Psychrometer constant very precisely which depends on the respective construction details,
the wind speed and the gas kind.
The Aspiration psychrometer comes from the meteorology where it is also used intensively
even today because of its mobility. The accuracy in the humidity measurement is limited by
the accuracy of the temperature readings on the thermometers. With the calibrated accuracy
of the thermometers available in our experiment, we reach an accuracy better than 0.2 K in
the psychrometric temperature difference; this corresponds at an ambient temperature of 25
°C the exactness in humidity measurements of 1% of relative humidity. For accurate
measurement the side effects that have an influence on the psychrometric effect should be
taken into account. Primarily, we have to keep the wind speed to a defined value because the
saturation of air strongly depends on its value.
As many experiments show the speeds over 2 m/s also give acceptable values. Also, for lower
speeds calibrations are possible, nevertheless then the variations in wind speed are possible
which enhance the measuring errors. It is favorable anyway to provide artificial ventilation
(e.g. by fan) for the necessary gas speed. However it is to be assured that the ventilation does
not raise the temperature of the gas before the measurement.
Beside the necessary least gas speed of 2 m/s, the application of the Psychrometers is limited
furthermore by the necessary moistening of the wet thermometer bulb which should remain
the same during the measurement and the extreme sensitivity against impurities (e.g. dust
particles) in the air which can lower the vapor pressure of the water and thus affect the
psychrometric effect. For these reasons continuous measurements are only very hardly
possible with this device. There are of course some devices which can continuously humidify
the wet bulb thermometer but due to the impurities in air the wet wick has to be changed
regularly in short periods.
Abb. 6 Graphische Psychrometertafel für stark bewegte Luft Luftgeschwindigkeit 2 m/s>=
°CTemperatur des trockenen Thermometers
020 30 40 50
5
10
15
20
25
30
35
10
Psyc
hrom
etris
che
Diff
eren
z
°C
0 %
10 %
20 %
30 %
40 %
50 %
60 %70 %
80 %90 %
14
15
A consideration of the Psychrometer formula (6) or (7) shows that at low temperatures and
water contents even small errors in temperature readings can result in very high errors in the
humidity values because of the small value of the psychrometric temperature difference (Δt).
This fixes the lower temperature limit for the accuracy in psychrometeric measurements. The
upper limit is where the cooling limit temperature reaches the boiling temperature of water,
because then water does not exist anymore in liquid aggregate state.
The relative humidity ϕ can be read in h-x-Diagram [Fig.(Abb.) 7] comfortably in the
intersection of the cooling border fog isotherm with the ambient temperature.
16
9000
8000
1000
0
7000
6000
5000
4500
4250
4000 3900
38003700
36003500
3400
3300
3200
3100
3000
2900
2800
2700
2600
2500
2400
2300
22002100200019001750
16501500
12501000
5000
h =50 kJ/kg
1+x
Δh /Δx [kJ/kg]1+x
t=60°C
t=30°C
t=50°C
t=0 °C(Eis)
t=0 °C(W
asser)t=-10°C
p* =0,09 bard
p* =0,08 bard
p* =0,07 bard
p* =0,06 bard
p* =0,05 bard
p* =0,04 bardp* =0,03 bard
p* =0,02 bard
0 10 20 30 40 50 70
h
[kJ/
kg]
1+x
Δh /Δx [kJ/kg]1+x
t=10°C
t=20°C
t=30°C
t=35°C
t=25°C
t=45°C
p* =0,1 bard
p=1,01325 bar
ϕ=0,5
-100
1020
3040
5060
Abb. 7 Mollier-h,x-Diagramm für feuchte Luft
h =200 kJ/kg
1+x
h =175 kJ/kg
1+x
t=40°C
h =150 kJ/kg
1+x
h =125 kJ/kg
1+x
h =100 kJ/kg
1+x
ϕ =0,4
ϕ=0,6
ϕ=0,8
ϕ =0,2
ϕ =0,1
t=15°C
h =75 kJ/kg
1+x
h =25 kJ/kg
1+x
t=10°C
t=20°C
h =0 kJ/kg
1+x
t=40°C
ϕ=1,0
x [g/kg] 60
2.4 LiCl-dew point measuring instrument
The measurement principle is based on the hygroscopic nature (ability of a substance to
attract water molecules) of lithium chloride. The hygroscopic property of LiCl to draw
moisture from the environment and the strong dependency of the electrical conductivity of
LiCl solution on its water content is used to measure the dew-point of a gas mixture (air).
Figure 8 shows the basic construction of a LiCl- dew point cell. The sensor (e.g. a resistance
thermometer/Widerstandsthermometer) consists of a reel covered with an absorbent fabric
(glass fibre/ Glasgewebe) and a bifilar winding (two insulated wires, with current traveling
through them in opposite directions) of inert electrodes. The reel is coated with lithium
chloride. An alternating current is passed through the winding and the lithium chloride
solution, causing resistive heating.
Widerstandsthermometer
Elektroden Glasgewebe mit LiCl
~
Abb. 8 Ausführungsprinzip einer LiCl-Taupunktzelle
If an alternating current is supplied to the precious metal wires of the hot probe (low water
content in LiCl solution) then first of all only a low current flows which is proportional (via
electrical resistance) to the water content in the LiCl solution. As the time passes the water
content rises slowly due to the hygroscopic effect of the LiCl, which leads to a reduction of
17
18
the electric resistance and an increase of current flow. The more and more Joule´s energy
released thereby causes, through the evaporation of water and a growing counter hygroscopic
effect, the flow of water from the LiCl solution to the surroundings. When the sensor has
been heated up so much that the mass flow of the evaporated water becomes larger than the
diffusion flow which results from the hygroscopic effect of the LiCl, then the water content in
the solution decreases again, the resistance increases and the electric current decreases. The
sensor cools to the extent that the proportion of the mass flow changes, i.e. the hygroscopic
mass flow is again just higher than the flow of evaporated water.
The apparatus comes to an equilibrium state where the temperature and the water
content of the LiCl solution reach a state where the mass flow of the evaporated water is
equal to the diffusion flow as a result of the hygroscopic effect of the LiCl solution. In this
equilibrium state the water vapor pressure over the LiCl solution is equal to the partial
pressure of the water in the surrounding air.
Starting the measuring process with a saturated probe (very high water content)
would lead, due to the low electric resistance of the solution, to the damage of the sensor by
overheating.
The temperature of the LiCl layer can be determined by the thermometer arranged
inside and with this the water vapor partial pressure from the water vapor partial pressure
curve of saturated LiCl solution. This will be equal to the water vapor partial pressure of
surrounding air at room temperature (surrounding temperature) due to the material
equilibrium. Also this is equal to the saturation pressure at the dew point and hence with the
help of the vapor pressure curve the relation between the sensor temperature and the dew
point is established. The display can be calibrated directly in such a way that it indicates the
dew point temperature, which as mentioned already in the introduction is suitable for
humidity characterization. Moreover, one can determine the relative humidity from the vapor
pressure curve (Fig. 3).
The main advantage of this method of humidity measurement lies in its wide range
of application (extremely low and high temperatures, -30 °C to +100 °C) and in its
insensitivity against dirt or electrolytically reactive impurities (by the strong hygroscopic
properties of LiCl are already dry at the equilibrium temperature and do not contribute to
electrical conductivity). For this reason this measuring method is well suited for continuous
measurements and for automatic check arrangements (the electrical signal can be used for
19
control), also in industries where the humidity is to be controlled. The measuring inaccuracy
is in general ± 1°C in dew point temperature, so that for the relative humidity an error of
approx. 3% can arise. The disadvantage is the long time required for the setting time of the
device which is about 1 min for every 1 °C dew point temperature.
2.5 Mirror dew point measuring instrument
This utilizes the occurrence of dew at dew point temperature when air containing water vapor
is cooled. A mirror is cooled until it reaches the dew point of the gas in question. As dew
condensation forms, it changes the light reflected from the mirror. When the mirror surface
reaches an equilibrium state whereby evaporation and condensation are occurring at the same
rate, the temperature of the mirror is equal to the dew point temperature of the tested gas.
With the determined dew point temperature the relative humidity can be determined with the
help of h- x diagram.
3. Experimental procedure
The aim of the experiment is the measurement of the humidity in a humidity chamber with
the help of the methods introduced in chapter 2. The experiment results are to be noted in the
enclosed test protocols. A comparison of the results using different methods reveals the
advantages and disadvantages as already given in chapter 2.
The experiment devices are accommodated in a chamber. The following eleven points
of the experimental procedure are to be done under two different conditions (2 test protocols).
The first measurement is done under the condition that the window of the chamber is open,
i.e. the measurement is done under ambient condition. These are recorded in Protocol 1. The
second measurement is done after changing the humidity of the chamber and the values are
recorded in Protocol 2.
Before the second measurement is carried out the following things are to be done:
a) Carefully close the window.
b) Turn on the humidifier: For this the three switches on the right side of the humidifier
are pressed. (do not change the thermostat setting!)
20
c) Wait, until the humidifier switches off itself automatically (approx. 10 min.). Then
Switch off the humidifier through the three switches on the right side of the
humidifier. (During the waiting period, the calculations of the first test protocol can be
carried out.)
d) Wait further 30 minutes before you start the second measurement.
The experiment should be carried out in the following steps:
1. Check whether the switch on the "Dew-Point" measuring device (LiCl-dew point
measuring device) on position "Standby" or "Dew-Point" stays. If not, immediately
inform the instructor, he will switch on the dew point apparatus! (Then an additional
half-hour will be required for the heating of the sensor)
2. When the switch is on the position "Dew-Point" note the current time as “start
measuring time" τe under point 4 in the protocol.
3. Read the relative humidity according to the printed instruction from both color
indicators hanging on the working place and note it under point 1 on the protocol.
4. Read (or evaluate) the following properties from the Hair Polymeter according to the
methods described in chapter 2.2 and note under point 2 in the protocol.
-The relative humidity in %
-The room temperature in °C
- The dew point temperature in °C
- The saturation pressure in Torr on the right side of the thermometer beside the room
temperature value
- The water vapor partial pressure in Torr
- The absolute humidity in g/m3
5. Preparation for the measurement with Psychrometer: The thermometer provided with
the "wick" (green tape) is to be made wet with the help of water. For this small
amount of water filled in the syringe (or a dropper) is sprayed over the wick.
6. Psychrometer measurements: After moistening the "wick" the ventilating fan is pulled
completely with the help of the key provided. During the running time of the
ventilating fan ( 3 to 4 minutes) both thermometers (especially the wet thermometer)
are observed whether they reach a constant final value . These values are noted under
point 3 on the protocol. If no constant value is reached during the run time, the
ventilating fan must be drawn up again. During the temperature reading it is to assured
that the temperature values are not influenced by the breathing.
7. Read the pressure on the mercury barometer in the Lab.
8. Determine the relative humidity from Fig. 6 (Abb. 6), calculate the water vapor
partial pressure from equation (7) (pF from Table 1) and write down the values. With
pD and you can determine the relative humidity from equation (5) and then from
Fig. 4 the absolute humidity and the dew point temperature can be read. The saturation
deficit is to be calculated according to the definition given in the introduction of the
experiment.
pDS
9. Before reading the dew point temperature on the LiCl-dew point measuring device
the time is to be noted under number 3 on the protocol. It is to be assured that the
instrument is on for more than 30 minutes. The value of the temperature is
noted.
T Tpkt
10. With the help of humidity slide rule (disc) the relative humidity and the absolute
humidity (in g/m3) can be calculated from the ambient temperature and the dew point
temperature. With the saturation pressure from table 1 and the read water vapor partial
pressure the saturation deficit can be calculated.
The above mentioned humidity variables can also be evaluated with the figures and
equations given above instead of using the humidity slide rule.
11. For the "Reflecting mirror dew point instrument" the air pump is started. Then the
depth of the reflecting stick into the ice water is so adjusted with the help of the
spindle that the mirror is just fogged. The reflecting mirror temperature TTpkt is read
on the thermometer. The relative humidity is to be read from Fig 7. As a check the
relative humidity can also be calculated with the help of equation (5).
21
Table 1 Saturated pressure of the water vapor (ps), water content of the saturated air (xs) and the absolute humidity of the saturated air, with reference to moist (humidified) state as a function of temperature between 0 to 30 °C. )’sf(
t
ps
xs ’sf
°C
mbar
Torr g/kg dry g/m3 moist
0 1 2 3 4 5 6 7 8 9 10
11 12 13 14 15
16 17 18 19 20
21 22 23 24 25
26 27 28 29 30
6.11 6.5 7.1 7.6 8.1 8.7
9.3
10.0 10.7 11.5 12.3
13.1 14.0 14.9 16.0 17.1
18.1 19.3 20.6 22.0 23.4
24.9 26.4 28.1 29.9 31.7
33.6 35.6 37.7 40.0 42.4
4.58 4.9 5.3 5.7 6.1 6.5
7.0 7.5 8.0 8.6 9.2
9.8
10.5 11.2 12.0 12.8
13.6 14.5 15.5 16.5 17.5
18.7 19.8 21.1 22.4 23.8
25.2 26.7 28.3 30.0 31.8
3.78 4.07 4.37 4.70 5.03 5.40
5.79 6.21 6.65 7.13 7.63
8.15 8.75 9.35 9.97 10.6
11.4 12.1 12.9 13.8 14.7
15.6 16.6 17.7 18.8 20.0
21.4 22.6 24.0 25.6 27.2
4.84 5.2 5.6 6.0 6.4 6.8
7.3 7.8 8.3 8.8 9.4
10.0 10.7 11.4 12.1 12.8
13.6 14.5 15.4 16.3 17.3
18.3 19.4 20.6 21.8 23.0
24.4 25.8 27.2 28.7 30.3
22
Table 1 Saturated pressure of the water vapor (ps), water content of the saturated air
(xs) and the absolute humidity of the saturated air, with reference to moist (humidified) state as a function of temperature between 31 to 70 °C. )’sf(
t
ps
xs ’sf
°C
mbar
Torr g/kg dry g/m3 moist
31 32 33 34 35
36 37 38 39 40
41 42 43 44 45
46 47 48 49 50
52 54 56 58 60
62 64 66 68 70
44.9 47.6 50.3 53.2 56.3
59.5 62.8 66.3 69.9 73.7
77.7 82.0 86.4 91.0 95.0
100.9 106.1 111.6 117.3 123.3
136.1
150 165
181.5 199.2
218.4 239.4 261.4 285.6 311.6
33.7 35.7 37.7 39.9 42.2
44.6 47.1 49.7 52.4 55.3
58.3 61.5 64.8 68.3 71.9
75.7 79.6 83.7 88.0 92.5
102.1 112.5 123.8 136.1 149.4
163.8 179.3 196.1 214.2 233.7
28.8 30.6 32.5 34.4 36.6
38.8 41.1 43.5 46.0 48.8
51.7 54.8 58.0 61.3 65.0
68.9 72.8 77.0 81.5 86.2
96.6 108 121 136 152
171 192 216 244 276
32 34 35 37 39
41 44 46 48 51
53 56 59 62 65
68 72 75 79 83
90 99 108 119 130
142 154 168 182 198
23
Table 1 Saturated pressure of the water vapor (ps), water content of the saturated air
(xs) and the absolute humidity of the saturated air, with reference to moist (humidified) state as a function of temperature between 71 to 100 °C. )’sf(
t
ps
xs ’sf
°C
mbar
Torr g/kg dry g/m3 wet
72 74 76 78 80
82 84 86 88 90
92 94 96 98
100
339.4 369.6 401.8 436.4 473.4
513.2 555.7 601.2 649.4 701.1
755.9 814.5 876.7 943.0
1013.3
254.6 277.2 301.4 327.4 355.1
384.9 416.8 450.9 487.1 525.8
567.0 610.9 657.6 707.3 760.0
313 357 409 471 546
638 755 907
1110 1396
1827 2549 3994 8348
-
214 232 251 271 293
316 340 367 394 423
454 487 522 559 597
24
Protocol 1 tu =
p = pDS
=
1. Color indicator
Color indicator 1 Color indicator 2
ϕ
2. Polymeter ϕ =
Δt = tTpkt =
f ′ =
pD = Δp =
3. Aspirations-Psychrometer tF =
Δt = ϕ = [Fig. (Abb.) 6]
pF =
pD = [Eq.(7)] ϕ = [Eq.( 5)]
f ′ = Δp =
4. LiCl-Dew point measuring device τe =
τm = > τe + 30 min
tTpkt =
pD = Δ p =
ϕ =
f ′ =
5. Reflecting mirror dew point measuring device tTpkt =
pD = Δp =
ϕ =
f ′ =
25
Protocol 2 tu =
p = pDS
=
1. Color indicator
Color indicator 1 Color indicator 2
ϕ
2. Polymeter ϕ =
Δt = tTpkt =
f ′ =
pD = Δp =
3. Aspirations-Psychrometer tF =
Δt = ϕ = [Fig.(Abb.) 6]
pF =
pD = [Eq.(7)] ϕ = [Eq.( 5)]
f ′ = Δp =
4. LiCl-Dew point measuring device τe =
τm = > τe + 30 min
tTpkt =
pD = Δ p =
ϕ =
f ′ =
5. Reflecting mirror dew point measuring device tTpkt =
pD = Δp =
ϕ =
f ′ =
26