1
AAPM 2005
The Ultrasound Research Interface: A New Tool for Biomedical Investigations
The Ultrasound Research Interface: A New Tool for Biomedical Investigations
Shelby Brunke1, Laurent Pelissier2, Kris Dickie2, Jim Zagzebski3, Tim Hall3, Thaddeus Wilson4
1Siemens Medical Systems, Issaquah WA2Ultrasonix Medical Corporation, Vancouver, BC
3Department of Med. Physics, University of Wisconsin4Department of Radiology, University of Tennessee
AAPM 2005
IntroductionIntroductionIntroductionDigitally controlled ultrasound scanners offer extensive levels of programmability, which enable manufacturers to explore and to readily incorporate alternative beam formation, signal and image processing, networking, and interfacing capabilities.
Recent efforts have led manufacturers to share these tools for innovation with academic and clinical researchers. This discussion will present capabilities of two such machines and present examples of research uses.
AAPM 2005
The Ultrasound Research Interface: A New Tool for Biomedical Investigations
Examples of Use
The Ultrasound Research Interface: A New Tool for Biomedical Investigations
Examples of Use
Jim Zagzebski, Tim Hall, Thaddeus Wilson*
Department of Med. Physics, University of Wisconsin*Department of Radiology, University of Tennessee
AAPM 2005
Parametric ImagingParametricParametric ImagingImaging
• Except for Doppler, ultrasound imaging is based entirely on echo amplitudes
• Parametric images•Acoustic attenuation•Scatterer size •Speed of sound•Elasticity (many forms)
2
AAPM 2005
Ultrasound Attenuation Ultrasound Attenuation Ultrasound Attenuation
• Attenuation is used diagnostically in the liver, breast, etc.
• However, only qualitative estimates are made– “mass exhibits shadowing”– “mass exhibits good through transmission”
• Goals: incorporate methods for determining attenuation locally, and in the form of images.
AAPM 2005
Measuring AttenuationMeasuring AttenuationMeasuring Attenuation
• Record RF echo data from ROI• Filter, measure reduction of rf signal
with depth• In a clinical machine, signal amplitude
changes with depth are also affected by:– Beam focusing, beam shapes– TGC settings set by operator– Internal TGC set by the manufacturer– Nonlinear processing in scanners
• Reference phantom techniques have been developed that effectively account for instrumentation effects.
Depth
( )( ) 2
2
logω
ω
r
s
S
S
AAPM 2005
Human liverHuman liverHuman liver
• ROI outlined from B-mode image (blue line)
• Areas of inhomogeneityeliminated (red line)
• Algorithm retrieves RF echo data from ROI, computes attenuation
• Results in normal liver agree with many previous reports (0.5 dB/cm-MHz)
Attenuation in liver vs. frequency
00.5
11.5
22.5
33.5
4
0 1 2 3 4 5 6
Frequency (MHz)
Att
enu
atio
n (d
B/c
m)
Tu et al
Lu et al
AAPM 2005
Attenuation ImagingSiemens SONOLINE Antares
Attenuation ImagingAttenuation ImagingSiemens SONOLINE AntaresSiemens SONOLINE Antares0.3 dB/cm/MHz contrast, 1 cm diameter
Spatial and Frequency Compounding
•Acquire RF data from multiple angles
•Compute α from ROI’s at each angle
3
AAPM 2005
Attenuation ImagingSiemens SONOLINE Antares
Attenuation ImagingAttenuation ImagingSiemens SONOLINE AntaresSiemens SONOLINE Antares0.3 dB/cm/MHz contrast, 1 cm diameter
Spatial and Frequency Compounding
•Acquire RF data from multiple angles
•Compute α from ROI’s at each angleAAPM 2005
“Scatterer Size” Imaging““ScattererScatterer SizeSize”” ImagingImaging
• RF data can be processed to yield the “backscatter coefficient” at frequencies throughout the signal bandwidth
• Values of backscatter vs. frequency reflect the size of scatterers contributing to the signal.
• By applying scatterer size dependent correlation models to the backscatter vs. data, possible to estimate size.
AAPM 2005
Mouse tumor model (Oelze and O’Brien, IEEE UFFC, 2004)Mouse tumor model Mouse tumor model ((OelzeOelze and Oand O’’Brien, IEEE UFFC, 2004)Brien, IEEE UFFC, 2004)
Carcinoma
Fibroadenoma
Scatterer size
Use single element transducer, 20 MHz
Reflects histological structure
AAPM 2005
Overlaid B-mode and Scatterer sizeOverlaid BOverlaid B--mode and Scatterer sizemode and Scatterer size
• Acquire RF echo data from normal human thyroid
• Siemens Antares, 6-13 MHz• Histology book: 100-200 µm
lobules• Scatterer size image data
appears to correlate with anatomy.
4
AAPM 2005
Patient with thyroid nodule (Wilson et al)Patient with thyroid nodule Patient with thyroid nodule (Wilson et al)(Wilson et al)
Near real-time scatterer size imaging mode on UltrasonixRP500
AAPM 2005
Elasticity ImagingElasticity Imaging
• Improve on manual palpation • Use a clinical ultrasound imaging system as
a sensor of anatomic deformation• Relative deformation quantifies the bulk
mechanical properties of tissue• Provides new diagnostic information
Estimation of Strain (Uses RF data frames)Estimation of Strain (Uses RF data frames)
Pre-compression RF line
Post-compression RF line
2τ
1τ T∆
TStrain
∆−= 12
ττ(Gradient of the axial displacement)
Array Transducer
AAPM 2005
Implementation on Machine with URIImplementation on Machine with URIImplementation on Machine with URI
5
AAPM 2005
Relative Size of LesionsRelative Size of LesionsRelative Size of Lesions
Invasive Ductal Carcinoma
Fibroadenoma
AAPM 2005
169 Breast Lesions; 1 Observer169 Breast Lesions; 1 Observer169 Breast Lesions; 1 Observer
AAPM 2005
0
1
2
3
4
5
6
7
8x 10
-3
0 10 20 30 40
0
5
10
15
20
25
30
35
40
0
1
2
3
4
5
6
7
8x 10
-3
0 10 20 30 40
0
5
10
15
20
25
30
35
400
1
2
3
4
5
6
7
8x 10
-3
0 10 20 30 40
0
5
10
15
20
25
30
35
40
0º
12º
10º
15º
0
1
2
3
4
5
6
7
8x 10
-3
0 10 20 30 40
0
5
10
15
20
25
30
35
40
(Ultrasonix, gel phantom, inclusion is 3x stiffer)
Spatial Angular Compounded Elastograms(Ultrasonix 500 RP)
AAPM 2005Compounded ElastogramsCompounded ElastogramsCompounded Elastograms
0
1
2
3
4
5
6
7
8x 10
-3
0 10 20 30 40
0
5
10
15
20
25
30
35
40
0
1
2
3
4
5
6
7
8x 10
-3
0 10 20 30 40
0
5
10
15
20
25
30
35
40
-15 º ~ 15 º
Elastogram Without Compounding
Spatial Angular Compounded Elastograms
0 5 10 15 2025
26
27
28
29
30
31
SN
Re
(dB
)
Maximum Angle ( o )
0.5o
1o
2o
3o
6
AAPM 2005
Speed of soundSpeed of soundSpeed of sound
Hayashi et al, A new method for measuring in vivo speed of sound speed in the reflection mode, J. Clin Ultrasound 16: 87-93, 1988.
Can you measure SOS in pulse-echo mode?
AAPM 2005
Speed of soundSpeed of soundSpeed of sound
• Beam former adds time delays to echo data picked up from elements in an array. Assumes SOS = 1540 m/s.
• Some URI’s allow assumed speed of sound to be programmed.
• Proper focusing is obtained only when the assumed speed of sound matches the speed of sound in the subject. Affects image intensity, for example.
AAPM 2005
Speed of soundSpeed of soundSpeed of sound• Generate images of phantoms for different assumed
SOS in the beamformer
• Measure image brightness vs assumed SOS
• Peak occurs near SOS of phantom
RMI 403C=1540 m/s
ATS 539C=1460m/s
AAPM 2005
ConclusionsConclusionsConclusions• Parametric imaging adds new information to
improve diagnostic accuracy• The ‘research interface’ on high-end systems are
somewhat limiting in what the user can control Extended control and system programming available through close working relationship with manufacturer
• The ‘research interface’ on lower-end systems provide greater access to system resources and system parameters
• These research interfaces provide an opportunity to investigate imaging algorithms that were impractical with laboratory data acquisition systems