Upload
isadora-figueroa
View
36
Download
0
Embed Size (px)
DESCRIPTION
AirFree Platform. Closed-Loop Process Control. The SmartDispenser™ comes ready to integrate with a AND AD4212B-102 Analytical Weighing Sensor. - PowerPoint PPT Presentation
Citation preview
• The SmartDispenser™ comes ready to integrate with a AND AD4212B-102 Analytical Weighing Sensor.
AirFree Platform. Closed-Loop Process Control
• Perform a capabilities study with the weighting sensor before production and use the data to find the target weight and +/- tolerances of the SmartDispenser™ with your assembly fluid.
• Dispense a full syringe allowing the SmartDispenser™ to automatically record weights.
• The statistical mean of the data is your target weight.
• The statistical standard deviation is your +/- tolerance.
• The SmartDispenser™ authorized administrator activates the scale feature and inputs the target weight and +/- tolerance percentage into the dispenser program.
• Dispense onto the scale allowing the SmartDispenser™ to read the dispense weight after each dispense. If dispense is out of tolerance a warning message appears stopping the SD until cleared.
• Actual weight dispensed and number of stops are recorded in an Excel spreadsheet for post production statistical analysis and costing.
• Reports can be accessed by any authorized SD administrator worldwide.
AirFree Platform. Closed-Loop Process Control
Production Data in Excel• 24, 7 day and 30 day machine cycles (one fluid dispense)
• All programs stored in the SmartDispenser™
• The program used for the last 500 dispenses verifying the proper program was used.