xxggeennuuss aacc//ddcc 2

The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.” reserves itself the right to modify the production and the manual without notice on their designated distribution. “de Götzen S.r.l.” shall not be liable for the incorrect use of the information contained in this manual. Copies, even excerpts, of this manual are allowed only for personal use and with written consent of the manufacturer.

Edition: 01/2004

Drawn up by A. Colombo Approved by M. de Götzen

Manual Revision Date Description

1.0 01/2004 English Version

Page 3: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

3

INTRODUCTION page 2 SUMMARY page 3 PRELIMINARY INFORMATION page 4 INFORMATION FOR THE INSTALLER page 5 1. “xgenus ac/dc” RADIOGRAPHIC SYSTEMS page 6 2. PICTOGRAMS USED page 7 3. OVERALL DIMENSIONS page 9 4. INSTALLATION SPECIFICATIONS page 13 5. INSTALLATION page 15 5.1 UNPACKING page 15 5.2 ASSEMBLING THE WALL PLATE page 16 5.3 ASSEMBLING THE BRACKET/HORIZONTAL ARM page 18 5.4 ASSEMBLING THE POSITIONER ARM page 20 5.5 CONNECTION TO THE TERMINAL BOARD page 21 5.6 ASSEMBLING THE RADIOGRAPHIC UNIT page 23 5.7 BALANCING THE POSITIONER ARM page 26 5.8 ASSEMBLING THE “xgenus® ac/dc” X-RAY CONTROL page 29 5.9 X-RAY CONTROL ELECTRICAL CONNECTION page 31 6. X-RAY CONTROL PANEL page 33 7. SYSTEM CONFIGURATION page 35 8. CHANGING THE CONFIGURATION page 38 9. START UP page 40 10. CHECKING THE INSTALLATION page 41 11. CHECKING THE EXPOSURE FACTORS page 43 12. DIAGNOSTIC page 46 13. TUBE HEAD CALIBRATION page 47 14. ERROR MESSAGES page 48 15. REPLACEMENT OF FUSES page 49 16. MAINTENANCE page 50 17. CLEANING THE OUTER SURFACES page 51 18. IF A REPAIR SHOULD BECOME NECESSARY page 51 19. ATTACHMENTS page 52

SUMMARY

Page 4: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 4

Before beginning to use the “xgenus ac/dc” radiographic systems, it is advisable to carefully read and follow the instructions contained herein, so as to obtain the best possible performance. Always pay close attention to the CAUTION, WARNING, and PLEASE NOTE messages when operating the system. LEGEND

CAUTION The word CAUTION identifies those situations that might compromise the operator's personal safety or cause injuries to others.

WARNING The word WARNING identifies those occurrences which might compromise the xgenus® ac/dc x-ray systems’ performance.

PLEASE NOTE The words PLEASE NOTE serves to draw your attention to special indications to facilitate maintenance or make important information more clear.

PRELIMINARY INFORMATION

Page 5: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

5

CAUTION The installer is responsible for the installation, calibration and testing with regard to system safety and efficient operation.

For safe and reliable installation of the “xgenus ac/dc” radiographic systems, it is advisable to: Check that the voltage mentioned on the rating plate matches the line voltage.

Install the “xgenus ac/dc” radiographic systems according to the procedures described in this manual.

Provide the user with any information regarding the use of the “xgenus ac/dc” radiographic systems according to that which is provided in the operators manual.

Certify the work done by completing the “Declaration of Conformity” form.

Return the completed warranty certificate to: TG Group. Only when the completed warranty certificate is received is the warranty valid.

INFORMATION FOR THE INSTALLER

Page 6: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 6

The “xgenus ac/dc” radiographic systems (Fig. 1) consists of:

TUBE HEAD – XGG / TUBEHEAD – XGDC∗∗∗∗ CONE

POSITIONER ARM (Scissor)

CONTROL – XGT

WALL BOX

HORIZONTAL ARM OPTIONAL ACCESSORIES : − long 12” (31cm) cone

− short 8” (20cm) cone

− long cone 12” (31cm) with a rectangular section sized 44x35mm

- remote control button: xgenus ECB

∗∗∗∗ xgenus® dc only

1. “xgenus ac/dc” RADIOGRAPHIC SYSTEMS

Fig. 1

Page 7: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

7

The identification tags on the tube heads XGG/XGDC∗∗∗∗ , on the control XGT and on the cone indicate the model name, the serial number, the manufacturing date and the symbols of the main technical characteristics.

ID TAG OF THE “xgenus ac” TUBE HEAD XGG

ID TAG OF THE “xgenus dc”

TUBE HEAD XGDC

ID TAG OF THE “xgenus”

CONTROL XGT

ID TAG OF THE SHORT 8” CONE: BLD20 - Ø 60mm (20 cm SSD)

ID TAG OF THE LONG 12” CONE: BLD31 - Ø 60mm (31 cm SSD)

ID TAG OF THE RECTANGULAR CONE: BLD31SQ - 44x35mm (31 cm SSD)

GRADUATED SCALE TAG ON TUBE HEADS

∗∗∗∗ “xgenus® dc” only

2. PICTOGRAMS USED

Page 8: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 8

SYMBOLS:

This symbol guarantees that the radiographic system complies with the regulations contained in the European Directive EEC 93/42 regarding Medical Devices

The degree of protection against direct and indirect electric contacts is B type

Symbol indicating danger due to “ionizing radiation”

Refer to Manual's instructions

N “NEUTRAL" power supply wire

L “LIVE" power supply wire

Ground cable

Page 9: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

9

Fig. 1A, 1B, 1C are the overall dimensions to accommodate varying installation situations: SHORT ARM (optional) length 41cm - 16.2”

STANDARD ARM length 82,5cm - 32.5”

LONG ARM (optional) length 110cm - 43.5”

SHORT ARM 41cm (16.2”) Fig. 1A

STANDARD ARM 82.5cm (32.5”) Fig. 1B

LONG ARM 110cm (43.5”) Fig. 1C

3. OVERALL DIMENSIONS

Page 10: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 10

Fig. 2, 3A, 3B, 3C e 4 show the typical dimensions of the radiographic system:

STANDARD ARM 82.5cm (32.5”) ONE INCH = 2.54 cm Fig. 2

Page 11: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

11

STANDARD ARM 82.5cm (32.5”) Fig. 3A

Fig. 3B Fig. 3C SHORT ARM 41cm (16.2”)

ONE INCH = 2.54cm LONG ARM 110cm (43.5”)

Page 12: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 12

ONE INCH = 2.54 cm Fig. 4

Page 13: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

13

WARNING Prior to installing the radiographic system the Office Owner must be sure that the operatory and its’ electrical services comply with all governing regulations and required installation specifications. Otherwise one must bring them all to compliance.

1. OPERATORY REQUIREMENTS a. The operatory must be of suitable width to accommodate the sizes and overall

dimensions provided (refer to section 3 “OVERALL DIMENSIONS”). Check that no obstacles are present while positioning the radiographic system.

b. The operatory must not be pressurized or constitute an explosion hazard.

c. While operating, the ambient temperature must be within +5°C and +40°C

d. The storage temperature must be within -15°C and +50°C

e. The relative humidity must be within 25% and 75% 2. REQUIREMENTS OF THE SUPPORTING WALL a. The supporting wall for the radiographic system must be able to withstand 200Kg sheer

stress at every bolt. b. The subject wall material and construction must be checked. If required, consult a contractor to determine strength and stability. c. If there is any doubt as to the strength or stability of the wall to support the x-ray, reinforce by providing either a steel or double plywood plate. Refer to the Pre-installation drawings.

3. REQUIREMENTS OF THE ELECTRIC SYSTEM a. The electric system must comply with all applicable regulatory bodies.

b. The electric system must be able to supply the power-voltage and current required by the Manufacturer's rating plate of the radiographic system as per Chart 1

Model xgenus® ac xgenus® dc

Input voltage of the radiographic systems 115V -5% +10%

115V ±±±±15%

NOMINAL LINE VOLTAGE 115V 115V MINIMUM LINE VOLTAGE 109 Vac 98 Vac MAXIMUM LINE VOLTAGE 126.5 Vac 132 Vac FREQUENCY 50/60 Hz 50/60 Hz

CH

AR

T 1

POWER DRAWN DURING EXPOSURE (MAX. TIME = 3.9 SEC) 800 VA 1.4 kVA

PLEASE NOTE∗∗∗∗ the communication cables (C11, C12 – C21, C22) between the timer and the tube head must be two-pole, twisted and shielded with ≥0,25mm2 (ex.: Belden 9501 type).

∗∗∗∗ xgenus® dc only

4. INSTALLATION SPECIFICATIONS

Page 14: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 14

PLEASE NOTE For longer lines, the conductor gauge must be increased in proportion.

Model xgenus® ac xgenus® dc

Line voltage 115V -5% +10%

115V ±±±±15%

MAXIMUM LINE VOLTAGE DROP 1.2% 3% MINIMUM LINE APPARENT RESISTANCE 0.2Ω 0.2Ω

5. ELECTRICAL CONNECTIONS

WARNING Prior to installing the radiographic system, it is advisable that the electrical supply complies with the recommended specifications in this manual.

X-ray control electrical connections

CAUTION According to relevant and practical standards, the x-ray control must be installed in a position allowing the operator to see and have full control of the radiographic exposure.

On the x-ray control installation wall, suitable runs for the following electrical cables must be provided, according to the electrical installation diagram (refer to section 18 “ATTACHMENTS”):

a. X-ray control electrical cables

(refer to section 5.9 “CONTROL ELECTRIC CONNECTIONS”) b. Cables connecting the control and the radiographic unit

(refer to section 5.9 “ CONTROL ELECTRIC CONNECTIONS ”) Radiographic unit electric connections

To the wall plate location on the installation wall, a suitable run for the cable connecting the

x-ray control and the radiographic unit must be provided (refer to section 5.5 “CONNECTION TO THE POWER SUPPLY TERMINAL BOARD”)

4. REQUIREMENTS OF THE ELECTRIC LINE a. The electric line must be "single-phase" type.

b. Separate circuit of 15 amps to be supplied.

c. Wire size according to AWG.

Page 15: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

15

CAUTION The “xgenus ac/dc” radiographic systems must be installed by professionally trained technicians, who must be able to certify their work.

CAUTION Prior to proceeding with the installation, check that all requirements are met.

The components of the “xgenus ac/dc” radiographic systems are delivered by the factory securely packaged within a specifically designed box, as shown in the following sketch (Fig. 5):

D DOCUMENTS: INSTRUCTIONS MANUAL, INSTALLATION & MAINTENANCE MANUAL, ANNUAL MAINTENANCE INSTRUCTIONS AND WARRANTY

RADIOGRAPHIC UNIT PACKAGING

X-RAY CONTROL PACKAGING + EXTERNAL LAMP (IF PROVIDED)

POSITIONER ARM PACKAGING

HORIZONTAL ARM AND WALL BOX PACKAGING

PLEASE NOTE Prior to installation, please check all components for shipping damage. Advise the carrier.

PLEASE NOTE The carton materials and the polystyrene foam can be completely recycled or disposed of by authorized recycling companies.

PLEASE NOTE It is advisable to store the original packaging to return the goods for repairs if required.

5. INSTALLATION

5.1 UNPACKING

Page 16: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 16

ASSEMBLY INSTRUCTIONS (Fig. 6)

1. Take the wall plate out of the packaging (refer to Fig. 5) Use the drilling template provided

2. Position the drilling template on the radiographic system installation wall, at the required height (130 cm from the base is the suggested height)

3. Fix the template to the wall with adhesive tape

4. Check the holes for vertical and horizontal alignment with the floor, using a plumb line

5. Mark the plate fixing holes

6. Mark the holes for the electric cables connecting the x-ray control to the radiographic unit PLEASE NOTE To prevent any flaking of the white paint coat and to control the center distances between the holes, it is advisable to start drilling with a tip Ø7, increasing this measure gradually.

7. Drill the plate fixing holes

8. If required, drill the holes for the electric cables connecting the x-ray control to the radiographic unit

9. Remove the template and insert the suitable anchor screws, according to the wall characteristics

10. Unscrew the screw and remove the plug from the wall plate

11. Withdraw the sliding cover

12. Attach the plate to the wall and insert the screws with the relevant washers, then tighten alternately

13. Check that the plate is securely fixed to the wall and is plumb and level. PLEASE NOTE If the wall is not perfectly leveled, put a suitable shim between the wall and the plate, so as to prevent any possible warping of the mounting plate.

5.2 ASSEMBLING THE WALL PLATE

Page 17: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

17

Fig. 6

Page 18: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 18

PLEASE NOTE The brackets come in the following lengths: 41cm (16.2”) - 82.5cm (32.5”) - 110cm (43.5”) The 82.5 cm and 110 cm brackets are provided with a stop key (Fig. 7A and 7B) to prevent the electric cable from twisting.

PLEASE NOTE Generally, the stop key is installed so that the equipment position at rest is on the right side of a possible viewer when standing in front of the wall plate (Fig. 7A). Should the position at rest be on the left side, the stop key must be rotated by 180° (Fig. 7B).

5.3 ASSEMBLING THE BRACKET/HORIZONTAL ARM

Fig. 7A Fig. 7B

WALL

Page 19: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

19

After fixing the plate to the wall, assemble the bracket according to the following procedure (Fig. 8):

1. Take the bracket out of the packaging (refer to Fig. 5)

2. Insert the bracket pin into the wall plate (upwards)

3. Insert the supporting rest

Fig. 8 PLEASE NOTE Prevent any foreign material (drywall, dust, cement, etc. ) from settling on the pin seat. The pin must slide freely in its seat. If required, thoroughly clean and lubricate with grease “Molikote D".

PLEASE NOTE Carefully check the bracket is parallel to the floor by using a water level.

Page 20: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 20

ASSEMBLY INSTRUCTIONS (Fig. 9)

1. Take out the positioner arm from the packaging (refer to Fig. 5)

2. Remove the arm and cap by unscrewing the fixing screw

3. Slide the arm guard slat

4. Insert the arm cable through the washer and then the positioner pin

Fig. 9

5. If required, clean the pin and the washer and lubricate with grease “Molikote D".

6. Insert the electric cable into the bracket housing

7. Assemble the guard slat

8. Insert the cable into the arm and push it until it reaches the pin outlet near the supply terminal board

5.4 ASSEMBLING THE POSITIONER ARM

Page 21: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

21

CAUTION For electrical safety, it is essential that the ground conductors be correctly connected.

WARNING The electric cable length is suitable for 82.5cm long arms. DOT NOT tamper with the terminals arrangement when 41cm long arms are used. The excess cable must be housed in the housing provided. The 110cm long arms are equipped with a special extension cable.

OPERATING INSTRUCTIONS (Fig. 10)

1. Remove the terminal board cover by removing the screw

Fig. 10

5.5 CONNECTION TO THE TERMINAL BOARD

Page 22: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 22

2. Proceed to the electric connection as shown in Fig. 11

PLEASE NOTE The “xgenus® dc” unit, differently from “xgenus® ac”, is equipped with 2 communication cables.

Fig. 11

TERMINAL BOARD CONNECTION DIAGRAM

BROWN L (line)

YELLOW GREEN T (ground)

BLUE N (neutral)

BLACK∗∗∗∗ COMMUNICATION∗∗∗∗

RED∗∗∗∗ COMMUNICATION∗∗∗∗

YELLOW GREEN T (ground)

YELLOW GREEN T (ground)

3. Connect the positioner arm shielded cable to the grounding potential

4. Connect the wall plate to the grounding potential

5. Clamp the cables with the stress relief clamps provided

6. Reassemble the terminal board insulating cover.

∗∗∗∗ xgenus® dc only

Page 23: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

23

ASSEMBLY INSTRUCTIONS

1. Take out the radiographic unit from the packaging (refer to Fig. 5)

2. Check that all the rating data match the power supply voltage

3. Remove both guards from the positioner arm by loosening the relevant screws (Fig. 12)

Fig. 12

4. Using a screw driver tip, separate the front coupling device (Fig. 13A)

5. Remove both guards (Fig. 13B)

Fig. 13A Fig. 13B

5.6 ASSEMBLING THE RADIOGRAPHIC UNIT

Page 24: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 24

6. Insert the arm unit pin into the x-ray head (Fig. 14A) and insert the supporting retainer “C” clip (Fig. 14B)

Fig. 14A Fig. 14B 7. Check that during insertion, the pin of the rotation prevention device correctly fits the seat located

on the x-ray head (Fig. 15)

Fig. 15

Page 25: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

25

8. Couple the arm and x-ray head connectors and make sure that they are seated (Fig. 16)

Fig. 16

Page 26: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 26

CAUTION The arm must be adjusted only when the radiographic unit is fully assembled and the x-ray head is mounted.

WARNING To prevent damage to the internal mechanism while performing adjustments and the balancing tests, the adjustment wrench/key must not be left in place.

WARNING The wrench/key provided must remain with the unit for future adjustments.

PLEASE NOTE To reach the adjustment screw X the arm “A” must be put in vertical position. To reach the adjustment screw Y the arm “B” must be put in horizontal position. The adjustment key provided can be inserted only under the above conditions (Fig. 17)

Fig. 17

Fig. 17

5.7 BALANCING THE POSITIONING ARM

arm “A”

arm “B”

X

Y

Page 27: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

27

Procedure to be followed to correctly balance the positioner arm (refer to Fig. 17): 1. BALANCING THE ARM “A”

PLEASE NOTE The positioner type arm is supplied with arm “A” already factory adjusted. The Arm “B” is supplied unloaded (no tension) for safety reasons.

2. BALANCING THE ARM “B” For correct balancing, proceed as follows: a. ARM “A” VERTICAL

b. ARM “B” HORIZONTAL

c. INSERT THE ADJUSTMENT WRENCH/KEY IN “Y”

d. LOAD THE SPRING BY APPLYING 22 TURNS WITH THE WRENCH IN A CLOCKWISE DIRECTION.

e. WITHDRAW THE WRENCH/KEY 3. CHECKING THE BALANCE To check the balance proceed as follows:

BRING THE ARM “B” TO SEVERAL DIFFERENT POSITIONS

if the head and arm do not hold the position you have selected:

a. BRING THE ARM “B” BACK TO THE HORIZONTAL POSITION

b. INSERT THE ADJUSTMENTWRENCH IN “Y”

c. ROTATE THE ADJUSTMENT WRENCH BY HALF TURN: − CLOCKWISE IF IT TENDS TO DRIFT DOWN − COUNTERCLOCKWISE IF IT DRIFTS UP

d. WITHDRAW THE WRENCH

PLEASE NOTE Repeat the test and adjustment until the arm ”A” is steady and stable in all positions, even when the arm “B” is completely extended.

4. READJUSTMENT OF ARM “A” If the head and arm do not hold the position you have selected: a. BRING THE ARM “A” TO THE VERTICAL POSITION AGAIN

b. INSERT THE ADJUSTMENT WRENCH IN “X”

c. TURN THE ADJUSTMENT WRENCH BY HALF TURN: − CLOCKWISE IF IT TENDS TO COME DOWN − COUNTERCLOCKWISE IF IT DRIFTS UP

d. WITHDRAW THE WRENCH

Page 28: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 28

PLEASE NOTE Repeat the test and adjustment until the arm “A” is steady and stable in all positions, even when the arm “B” is completely extended.

5. UPON COMPLETION OF BALANCING:

a. Insert the movable finish cover (A) between the arm guard and metal frame (Fig. 18)

Fig. 18

b. Insert the pins of one guard into the relevant seats; then position it and check that the finish piece is coupled to the guards (Fig. 19)

Fig.19

A

Page 29: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

29

CAUTION Check that the cable runs are arranged in the x-ray control installation wall; check the compliance of the power supply with the installation specifications (refer to § 4 “INSTALLATION SPECIFICATIONS”).

CAUTION Check that the rating data match the power supply voltage.

ASSEMBLY INSTRUCTIONS (Fig. 20)

1. Take the x-ray control out of the packaging (refer to Fig. 5)

2. Take out the drilling template

3. Confirm the vertical and horizontal alignment. Mark the mounting holes on the wall using the drilling template.

4. Drill using a Ø3mm tip, then drill again with a Ø6mm tip to prevent any flaking of the white coat

5. Insert the three screw anchors provided into the holes

6. Open the x-ray control by unscrewing and removing the three screws

5.8 ASSEMBLING “xgenus® ac/dc” X-RAY CONTROL

Fig. 20

Page 30: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 30

7. Remove the 26-pole connector from its seat to release both control guards (Fig. 21)

Fig. 21

8. Insert the electric feed cables into the hole (Fig. 22)

Fig. 22

9. Insert the connection cables coming from the radiographic units into the rectangular slot

10. Insert the cables of the external signal lamps (IF INSTALLED) (Fig. 22)

11. Insert the cables of the remote control button (IF INSTALLED) (Fig. 22)

12. Attach the x-ray control base to the wall, matching the three anchor screws with the holes

13. Tighten the screws (refer to Fig. 20 ) with the relevant washers (refer to Fig. 20 ) into the screw anchors

PLEASE NOTE

If the wall is uneven, use appropriate shims to prevent the x-ray control from warping.

PLEASE NOTE Keep the x-ray control clean. Protect the board from dust, dirt or wire particles.

CONNECTOR 26 pole

POWER SUPPLY CABLE INLET HOLE

INLET HOLE FOR THE RADIOGRAPHIC UNIT(S) AND WARNING LAMP(S) CABLE

Page 31: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

31

CAUTION Before proceeding to connections, the power supply must be turned off.

CAUTION If the equipment is mounted on metal wall or studs, these must be connected to the grounding circuit of the system.

WARNING While carrying out electrical connection procedures, always respect the polarity PHASE – NEUTRAL.

WARNING While stripping the cable, pay attention to the small copper wires that may fall on the printed circuit board and cause short circuits or malfunctions.

OPERATING INSTRUCTIONS (Fig. 23) (refer to §19 “ATTACHMENTS”)

1. Connect the power supply cable to the terminal board

2. Insert the three mains cables into the rack

3. Fix them with the cable clamp

4. Connect the cables coming from the operatory unit# 1 to the terminals XRAY1

4a.∗∗∗∗ Connect the communication cables from the tube head 1 to the terminals C11 and C12 5. Connect the YELLOW-GREEN grounding cable to the metal "Grounding plate”

6. Connect the cables coming from the operatory unit# 2 to the terminals XRAY2

6a.∗∗∗∗ Connect the communication cables from the tube head 2 to the terminals C21 and C22 7. Connect the YELLOW-GREEN grounding cable to the metal “Grounding plate” 8. Fasten the cables in the cable clamp

9. Connect the remote control button (if installed) (refer to § 19 “ATTACHMENTS”) 10. Check the configuration on the dip-switches (refer to section 7 “SYSTEM CONFIGURATION”) 11. Reconnect the 26-pole connector 12. Close the x-ray control with the three screws

13. Mount the plate guard (refer to Fig. 6 ) 14. Reconnect the power supply

∗∗∗∗ xgenus® dc only

5.9 X-RAY CONTROL ELECTRICAL CONNECTION

Page 32: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 32

Fig. 23

The electronic PC card terminal boards include the follow:

1 KEY1 RADIOGRAPHIC UNIT N° 1 CONTROL BUTTON WITH LED 2 KEY2 RADIOGRAPHIC UNIT N° 2 CONTROL BUTTON (OPTIONAL) WITH LED 3A S12/S11 RADIOGRAPHIC UNIT N° 1 SIGNAL LAMP OUTSIDE THE OPERATORY

3B S22/S21 RADIOGRAPHIC UNIT N° 2 SIGNAL LAMP OUTSIDE THE OPERATORY

4A∗∗∗∗ C12/C11 RS232 COMMUNICATION FOR DC TUBE HEAD 1 4B∗∗∗∗ C22/C21 RS232 COMMUNICATION FOR DC TUBE HEAD 2 5 XRAY1 RADIOGRAPHIC UNIT N° 1 POWER SUPPLY 6 XRAY2 RADIOGRAPHIC UNIT N° 2 POWER SUPPLY 7 LINE X-RAY CONTROL POWER SUPPLY

WARNING Check the value of the fuses F1-F2 (see schemes on § 19) which must be in compliance with the type of tube head connected to the XRAY1 and XRAY2 terminal boards:

TUBE HEAD TYPE FUSE xgenus® ac xgenus® dc

F1 (XRAY1) 8AF - 250V 12.5A - 250V F2 (XRAY2) 8AF - 250V 12.5A - 250V

∗∗∗∗ xgenus dc only

1 2 3A-3B 4A-4B 5 6 7

Page 33: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

33

The control panel is the user's interface, where the operator can: with the keys select the radiographic unit operating parameters

change the exposure times displayed

store any new customized exposure times

with the light indicators “green Led” check the x-ray control operating conditions

check the active function key

check the selected radiographic unit parameter

6. X-RAY CONTROL PANEL

Page 34: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 34

control panel

MAIN ON/OFF SWITCH

LED INDICATOR

DISPLAY

X-RAY EXPOSURE

BUTTON

RADIOGRAPHIC DISTANCE INDICATOR

(LONG CONE RECOMMENDED-

MANDATORY OF DC OPERATION)

KEY TO DECREASE EXPOSURE TIME

KEY TO INCREASE EXPOSURE TIME

TUBEHEAD TYPE INDICATOR

TUBEHEAD SELECTION

RADIOGRAPHIC KV/VOLTAGE SELECTOR

SELECTION OF PATIENT’S SIZE

MEMORY

MAXILLARY UPPER TEETH

RADIOGRAPHIC CURRENT SELECTOR

MANDIBULARY LOWER TEETH

OCCLUSAL EXAM

X-RAY EMISSION SIGNAL

BITE-WING EXAM

DUTY CICLE PAUSE INDICATOR

DIGITAL RADIOGRAPHIC TECHNIQUE (HOLD FOR 3 SEC.-ALWAYS USE LONG

CONE TECHNIQUE) MALFUNCTIONING

INDICATOR

FILM TYPE SELECTION

KEY SWITCH LOCK

Page 35: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

35

The “xgenus ac/dc” radiographic systems is factory configured for the following operative “standard mode” configuration

Installation of 2 “xgenus” tube heads

on the control panel by pressing the RX button

Led 1 lights up, by pressing again the button, Led 2 lights up

IF THE LONG 12” (31cm) CONE IS USED A radiographic distance SSD = 31cm with long 12” cone

on the control panel Led 12” is lit

IF THE SHORT 8” (20cm) CONE IS USED A radiographic distance SSD = 20cm with short 8” cone

on the control panel Led 8” is lit

with film type “D” on the control panel Led “D” is lit

CONTROL BUTTON to perform the exposure

The control includes a control button with long flexible coil cord.

The installed radiographic units include the following features:

tube head type Led “AC” lit: “xgenus® ac” Led “DC” lit: “xgenus® dc”

radiographic voltage

the Led 70 kV or 60 kV∗∗∗∗ on the control panel must be lit

radiographic current

the Led 4mA∗∗∗∗ or 8mA on the control panel must be lit (ONLY AVAILABLE IN DC MODE)

The above configuration depends on the position of n° 8 mini-switches (dip-switch ) on the x-ray control board.

DIP SWITCH PARAMETER ON OFF

1 RADIOGRAPHIC UNIT N° 1 INSTALLED NOT INSTALLED 2 TYPE OF RADIOGRAPHIC UNIT N° 1 DC AC 3 RADIOGRAPHIC UNIT N° 2 INSTALLED NOT INSTALLED 4 TYPE OF RADIOGRAPHIC UNIT N° 2 DC AC 5 CONTROL BUTTON N° 2 INSTALLED NOT INSTALLED

6 CONE LONG (12”) INSTALLED

SHORT (8”) INSTALLED

7 NOT AVAILABLE NOT AVAILABLE 8 NOT AVAILABLE NOT AVAILABLE

∗∗∗∗ xgenus® dc only

7. SYSTEM CONFIGURATION

Page 36: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 36

IF THE LONG 12” (31cm) CONE IS USED

ON OFF 1 radiographic unit n° 1 2

∗∗∗∗ AC (=off) / DC (=on) 3 radiographic unit n° 2 4

∗∗∗∗ AC (=off) / DC (=on) 5 control button n° 2 6 cone 12” (31cm) 7 not available 8 not available

IF THE SHORT 8” (20cm) CONE IS USED

ON OFF 1 radiographic unit n° 1 2

∗∗∗∗ AC (=off) / DC (=on) 3 radiographic unit n° 2 4

∗∗∗∗ AC (=off) / DC (=on) 5 control button n° 2 6 cone 8” (20cm) 7 not available 8 not available

∗ xgenus® dc only

Page 37: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

37

The configuration may be changed if:

POSSIBLE MODIFICATIONS HOW TO CARRY OUT THE MODIFICATIONS

− type “E” and “F” films are used

− a digital system is used − radiographic current (4mA∗∗∗∗ /8mA) − radiographic voltage (60kV∗∗∗∗ /70kV)

By pressing the keys of the control panel

refer to USER’S MANUAL sect. 6 “ USE INSTRUCTIONS”

− the short 8” (20cm) cone is used − the long 12” (31cm) cone is used − one single radiographic unit is used − n° 2 CONTROL BUTTONS are used

by changing the dip-switch position

THIS OPERATION MUST BE CARRIED OUT BY THE AUTHORIZED SERVICE PERSONAL ONLY

PLEASE NOTE “xgenus® dc” tube head allows the following combinations for the radiographic voltage and current:

60kV/4mA – 60kV/8mA – 70kV/4mA – 70kV/8mA

“xgenus® ac” tube head allows only:

70kV/8mA

∗∗∗∗ xgenus® dc only

Page 38: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 38

CAUTION This operation must be carried out by a qualified installer/technician.

WARNING To install the changes you have made, turn off the x-ray control and then turn it on again.

To change the configuration, the dip-switch positions in the x-ray control must be changed:

DIP

SWITCH PARAMETER ON OFF

1 RADIOGRAPHIC UNIT N° 1 INSTALLED NOT INSTALLED 2 TYPE OF RADIOGRAPHIC UNIT N° 1 DC AC 3 RADIOGRAPHIC UNIT N° 2 INSTALLED NOT INSTALLED 4 TYPE OF RADIOGRAPHIC UNIT N° 2 DC AC 5 CONTROL BUTTON N° 2 INSTALLED NOT INSTALLED

6 CONE LONG (12”) INSTALLED

SHORT (8”) INSTALLED

7 NOT AVAILABLE NOT AVAILABLE 8 NOT AVAILABLE NOT AVAILABLE

1. CHANGING THE NUMBER OF RADIOGRAPHIC UNITS INSTALLED To change the number of radiographic units installed, move the dip-switch n° 1 or n° 3 a. If the radiographic unit is connected to the RX1 terminal board, put the dip-switch n° 1

to the ON position, otherwise to the OFF position. b. If the radiographic unit is connected to the RX2 terminal board, put the dip-switch n° 3

to the ON position, otherwise to the OFF position. 2. CHANGE THE TYPE OF TUBE HEAD INSTALLED a. if the tube head connected to the XRAY1 is “ac” type, move the dip-switch n° 2 to the

OFF position; otherwise to the ON∗∗∗∗ position. b. if the tube head connected to the XRAY2 is “ac” type, move the dip-switch n° 4 to the

OFF position; otherwise to the ON∗∗∗∗ position. 3. CHANGING THE CONE To replace the 12” cone with the 8” cone, move the dip-switch n° 6 to the ON position.

Check that the led “8” lights up on the control panel.

PLEASE NOTE After modification, the set exposure times are automatically changed.

∗∗∗∗ xgenus® dc only

8. CHANGING THE CONFIGURATION

Page 39: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

39

To replace the 8” cone with the 12” cone, move the dip-switch n° 6 to the ON position. Check that the led “12” lights up on the control panel.

PLEASE NOTE

After modification, the set exposure times are automatically changed. 4. CHANGING THE AMOUNT OF THE REMOTE BUTTONS To change the number of control buttons, move the dip-switch n° 5 to the ON position.

PLEASE NOTE After modification, each radiographic unit is controlled by the relevant button.

PLEASE NOTE It’s possible to connect to the timer: n° 2 “xgenus® ac” n° 2 “xgenus® dc” n° 1 “xgenus® ac” + n°1 “xgenus® dc”

Page 40: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 40

CAUTION For further information, see the USER'S MANUAL.

TURN ON THE X-RAY CONTROL to provide power to the system a. Turn THE “KEY SWITCH” to position “I” (ON). b. Set the “MAIN SWITCH” located on the upper part of the control to the position “I” (ON). The green light turns on indicating that the system is powered

The LED’s of the set radiographic parameters automatically light up

The exposure time is shown on the display

c. THE RADIOGRAPHIC SYSTEM IS NOW READY FOR USE

ATTENTION If an error is detected when the system is turned on, the anomaly is indicated as follows: • an intermittent beep sounds off

• The “MALFUNCTIONING INDICATOR” Led turns on intermittently

The error code (E ….) appears on the display

(refer to §14 “ERROR MESSAGES”)

• All CONTROL PANEL functions are inhibited In this case, turn off the timer and then turn it back on again. If the error should repeat itself, call for “Technical Assistance”.

PLEASE NOTE The exposure time and radiographic parameters which appear on the display are the last that were set before the x-ray control was turned off.

PLEASE NOTE If installed outside the operatory, the RX signaling lamp corresponding to the selected tube head turns on.

PLEASE NOTE If the x-ray control remains unused and inactive for a few minutes, it switches to the stand-by mode. Press any key of the CONTROL PANEL to restore it to the operative mode. The previous settings will be retained.

9. START UP

Page 41: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

41

CAUTION When assembly and all connections are completed, the installer must check the electrical safety and function of the system for compliance. OPERATING INSTRUCTIONS: 1. Checking the configuration Check on the control panel that all LEDs corresponding to the required configuration are lit;

otherwise, change them. (Refer to section. 7 “SYSTEM CONFIGURATION” and section.8 “CHANGING THE CONFIGURATION”.)

2. Checking the x-ray control operation a. Verify the correct operation of the control panel by selecting different exposure times

b. Verify the time on the display

c. When changing the selected radiographic unit, check that the corresponding external RX signal lamps (if installed) turn on.

3. Checking the exposure: CAUTION! Radiation will be emitted. a. Set an exposure time of 1 sec

b. Lift the "CONTROL BUTTON" on the x-ray control

c. Using the extensible cable of the “CONTROL BUTTON”, keep a safe distance (length of the cable) or at least 2 meters (7 feet) from the radiographic unit, to check the radiographic exposure.

d. Press the “X-RAY KEY” and keep it pressed until the acoustic signal (beep)

stops and the yellow “X-RAY OUTPUT SIGNAL” Led turns off. PLEASE NOTE If the “EXPOSURE BUTTON” is released early, the exposure is immediately interrupted and the E12 error message appears on the display.

e. At the end of the exposure, verify the operation of the Led “PAUSE INDICATOR”

f. Verify on the display the indication of the actual exposure time (Refer to section 11.4 “EXPOSURE TIME (sec)”)

10. CHECKING THE INSTALLATION

Page 42: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 42

4. Checking the operation of the radiographic unit: CAUTION! Radiation will be

emitted Carry out several exposures on the installed radiographic units installed and check for the

following: a. There are no errors.

b. The LED (AC or DC) of the selected radiographic unit is lit.

c. The "CONTROL BUTTON" LED is lit over the duration of the acoustic signal. 5. Checking the power drawn by the radiographic system: CAUTION! Radiation will

be emitted. To check the power drawn by the radiographic system, a meter must be used, with the

Ammeter in “AC” a. Connect the instrument to the power supply line

(refer to section 5.9 “X-RAY CONTROL ELECTRICAL CONNECTION”)

b. Set an exposure time of approx. 3 sec. on the x-ray control. b.∗∗∗∗ If using an xgenus® dc, set 70kV-8mA-3.2s on the x-ray control.

c. Carry out an exposure and read the current value on the instrument

PLEASE NOTE The radiographic system complies with the requirements when:

the current drawn is ≤ 8A for line voltage of 115V

the current drawn is ≤ 12.5A∗∗∗∗ for line voltage of 115V Otherwise, check the electrical system or call for “Technical Assistance”.

6. Checking the electrical system To check the electrical system, a meter must be used, with the Voltmeter in “AC” mode. a. Connect the instrument to the terminals L and N on the x-ray control

b. Measure the line voltage (load voltage V0)

c. Connect the instrument to the terminals L and N of the wall plate terminal board (refer to section 5.5 – fig.11 - “CONNECTION TO THE TERMINAL BOARD”)

d. Set an exposure time of approx. 3 sec on the x-ray control d.∗∗∗∗ If using an xgenus® dc, set 70kV-8mA-3.2s on the x-ray control. e. Release the exposure

f. Measure the line voltage during exposure (load voltage V1)

PLEASE NOTE The electric system complies with the requirements when the result of the formula is as follows:

(V0 – V1 ) / V0 ≤ 0.03∗∗∗∗ or 0.012∗∗∗∗∗∗∗∗ Otherwise, the electric system must be adjusted (refer to section 4: “INSTALLATION SPECIFICATIONS”)

∗∗∗∗ xgenus® dc only

Page 43: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

43

1. Radiographic voltage (kVp)

The radiographic high voltage is measured using a calibrated "non invasive" instrument with an initial delay set of 50ms∗∗∗∗∗∗∗∗ . The kVp value is factory checked with a PMX-I D.

xgenus® ac xgenus® dc Line Voltage Vn -5% / +10% Vn ±15% Max Voltage drop 1.2% 3% Nominal High Voltage 70 kV 60-70 kV Nominal current 8mA 4mA – 8 mA Set exposure time 3.2 s 3.2 s

SET TECHNICAL FACTORS

The radiographic VOLTAGE is: 70kVp ±15% for “xgenus® ac” 60kVp – 70kVp ±10% for “xgenus® dc”

2. Radiographic current (mA)

The radiographic current is measured by connecting a milliampmeter. The mA value is factory checked with a digital multimeter.

xgenus® ac xgenus® dc Line Voltage Vn -5% / +10% Vn ±15% Max Voltage drop 1.2% 3% Nominal High Voltage 70 kV 60-70 kV Nominal current 8mA 4mA – 8 mA Set exposure time 3.2 s 3.2 s

SET TECHNICAL FACTORS

The radiographic CURRENT is: 8mA ±15% for “xgenus® ac” 4mA – 8mA ±10% for “xgenus® dc”

3. Dose (mGy)

The dose in air is measured with a "non invasive" instrument, by positioning the detector at a spot-skin distance = 310mm (12”). The mGy are factory checked with the instrument SOLIDOSE.

xgenus® ac xgenus® dc Line Voltage Vn -5% / +10% Vn ±15% Max Voltage drop 1.2% 3% Nominal High Voltage 70 kV 60-70 kV Nominal current 8mA 4mA – 8 mA Set exposure time 1 s 1 s

SET TECHNICAL FACTORS

Dose in air∗∗∗∗∗∗∗∗ 3.3 mGy/s ±30% (with 12” cone) 6.5 mGy/s ±30% (with 8” cone) Dose in air∗∗∗∗ (60kVp – 4mA)=2.2mGy/s±30% (with 12” cone) (70kVp – 4mA)=3mGy/s±30% (with 12” cone) (60kVp – 8mA)=4.5mGy/s±30% (with 12” cone) (70kVp – 8mA)=5.5mGy/s±30% (with 12” cone)

∗∗∗∗∗∗∗∗ xgenus® ac only ∗∗∗∗∗∗∗∗ xgenus® ac only ∗∗∗∗ xgenus® dc only equipped with a 12” cone. See §13 of the “USER’S MANUAL” for the dose in air with a 8” cone.

11. CHECKING THE EXPOSURE FACTORS

Page 44: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 44

4. Exposure time (s) The exposure time is measured with a “non invasive” instrument.

PLEASE NOTE∗∗∗∗∗∗∗∗ The “xgenus” x-ray control features the "self-compensating" technology. The automatic correction or compensation of the exposure time offsets the effects of the line voltage fluctuations on x-ray film quality. Radiographic over/under exposure is eliminated. The automatic correction system works as follows:

• A voltmeter located inside the timer constantly reads the mains voltage while the operator selects the required exposure time.

• The last voltage measured (before the exposure) is sent to the

microprocessor. The selected exposure time is corrected to voltage by applying a special algorithm.

• At the end of the exposure, during the pause time, the time-corrected value

is shown on the display of the x-ray control. THE PAUSE TIME IS APPROX. 32 TIMES THE SET TIME.

The correct time is the actual duration of the exposure

The accuracy of the “actual time” is factory checked with instrument PMX-I.

xgenus® ac xgenus® dc Line Voltage Vn -5% / +10% Vn ±15% Max Voltage drop 1.2% 3% Nominal High Voltage 70 kV 60-70 kV Nominal current 8mA 4mA – 8 mA Set exposure time 0.2 s 1.6 s

SET TECHNICAL FACTORS

The EXPOSURE TIME measured is: 0.2s ±10% for “xgenus® ac” 1.6s ±10% for “xgenus® dc”

5. Checking the exposure time∗∗∗∗∗∗∗∗ (continued on p. 45: “Operating instructions”)

WARNING While performing periodic quality checks, some differences between the set exposure time and the measured exposure time may be detected. The differences are due to the automatic time compensation that intervenes when the power voltage differs from the nominal one.

The x-ray control is provided with two types of accuracy measurements for the exposure time:

A. Set exposure (is the exposure time set by the operator on the x-ray control).

The max error is +100% / - 100% when the voltage fluctuation is -5% / +10%.

B. After exposure: it’s the exposure duration (automatically adjusted). It’s calculated by the microprocessor according to the line voltage, and it’s shown on the display.

∗∗∗∗∗∗∗∗ xgenus® ac only

Page 45: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

45

OPERATING INSTRUCTIONS:

CAUTION Radiations will be emitted.

To check the exposure times (sec) it is advisable to use calibrated “non invasive” instruments; i.e.:

NERO by Victoreen PMX by RTI TRIAD QA by Keithley

a. Put the “X-ray detector” at the cone outlet; center the primary beam. b. Turn on the x-ray control c. Set one exposure time on the control. d. Release the exposure. e. At the end of the exposure, when the functional pause period starts, read the actual time of

exposure on the display. f. To check the accuracy of the exposure time, use the chart below.

CONFORMITY CHECK CHART At the end of the exposure

read the time value on the display = Actual exposure time

read the measure on the instrument = Emission time measure

subtract 1° from 2°

divide 3° by 2° = Absolute error

multiply 4° by 100 = Percentage error (%)

The accuracy is checked when the % of error is

±10% o ±20msec by adopting the most favorable condition.

PLEASE NOTE The exposure time complies with the requirements when:

((Measured time – Actual time) / Measured time)) x 100 ≤ ±10% or ±20 ms

Otherwise, please contact “Technical Support”.

Page 46: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 46

The “xgenus ac/dc” radiographic systems allows the installer to set or display some functional parameters of the x-ray control.

1. To set the minimum exposure time: a. Select the tube head b. Turn off the x-ray control

c. Turn on the x-ray control, by keeping the key pressed d. The message “SEC” is displayed for approx 1sec e. The present value of the lower set limit is displayed

f. To change the value, press the keys and

g. To confirm, press the key

h. To exit this mode, turn the x-ray control off and then on again

WARNING Both in digital and in conventional mode, the minimum exposure time MUST NOT BE LESS than 0,08s∗∗∗∗∗∗∗∗

2. To visualize the parameters, the installer must:

a. Press simultaneously and keep pressed the following keys (2) UPPER JAW MOLAR + (31) LOWER JAW MOLAR

b. Press the key associated to the parameter one wishes to see

KEY DISPLAYED PARAMETER Example U.M.

BITE-WING ANTERIOR

RADIOGRAPHIC SYSTEM VOLTAGE 110 Volt

BITE-WING POSTERIOR LINE VOLTAGE 105 Volt

UPPER INCISOR

MAXIMUM LINE VOLTAGE VALUE DETECTED∗∗∗∗∗∗∗∗

115

Volt

LOWER INCISOR

MINIMUM LINE VOLTAGE VALUE DETECTED∗∗∗∗∗∗∗∗

118 Volt

OCCLUSAL SOFTWARE VERSION 1.5 ----

∗∗∗∗∗∗∗∗

xgenus® ac only

12. DIAGNOSTIC

Page 47: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

47

CAUTION During this operation radiations will be emitted. It is mandatory to carry out all the safety measures relevant to radioprotection. INSTRUCTIONS 1. turn off the timer 2. turn on the timer by keeping the key pressed. 3. the message TUBE is displayed. 4. select the appropriate tube head to calibrate.

5. take an exposure by pressing the EXPOSURE BUTTON key and keep it pressed until the acoustic signal (beeping) stops and the yellow led turns off. 6. Once the exposure has been taken, if the display does not show errors, the calibration has been

successfully done. ∗∗∗∗xgenus® dc only

13. TUBE HEAD CALIBRATION∗∗∗∗

Page 48: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 48

The following chart gives a list of error messages that may appear while the “xgenus ac/dc” radiographic systems is working. The chart also includes the reason for the error messages and what to do to solve them. ERROR

MESSAGES CAUSE SOLUTION

RX1 TUBEHEAD

IS NOT CONNECTED OR IS OUT OF ORDER CALL TECHNICAL SUPPORT

RX2 TUBEHEAD

IS NOT CONNECTED OR IS OUT OF ORDER CALL TECHNICAL SUPPORT

CORRUPTED EEPROM DATA CALL TECHNICAL SUPPORT

EEPROM DATA NOT SAVED PROPERLY CALL TECHNICAL SUPPORT

LINE VOLTAGE VALUE NOT

WITHIN RANGE CALL TECHNICAL SUPPORT

THE X-RAY KEY

ALWAYS SEEMS TO BE PRESSED MAKE SURE IT IS NOT JAMMED OR DEFECTIVE

ANOMALY IN THE CONTROL PANEL CALL TECHNICAL SUPPORT

THE EXPOSURE

HAS BEEN PREMATURELY INTERRUPTED KEEP THE X-RAY KEY

PRESSED TILL THE END OF THE EXPOSURE

ANOMALY IN THE TRIAC/RELAY CALL TECHNICAL SUPPORT

ANOMALY IN THE ELECTRONIC CIRCUIT CALL TECHNICAL SUPPORT

ANOMALY IN THE CONTROL CIRCUIT CALL TECHNICAL SUPPORT

INCORRECT DIP-SWITCH CONFIGURATION CALL TECHNICAL SUPPORT

THE CONTROL BUTTON DOES NOT CORRESPOND

TO THE SELECTED TUBEHEAD CALL TECHNICAL SUPPORT

∗∗∗∗ THE TUBEHEAD

DOES NOT WORK PROPERLY CALL TECHNICAL SUPPORT

∗∗∗∗ THE TUBEHEAD

IS NOT IN THE CORRECT MODE CALL TECHNICAL SUPPORT

∗∗∗∗ THE TUBEHEAD

HAS NOT COMPLETED THE EXPOSURE REPEAT THE EXPOSURE

CALL TECHNICAL SUPPORT ∗∗∗∗

THE TUBEHEAD IS NOT CALIBRATED

CALL TECHNICAL SUPPORT

∗∗∗∗ EEPROM DATA NOT SAVED PROPERLY CALL TECHNICAL SUPPORT

∗∗∗∗ CORRUPTED EEPROM DATA CALL TECHNICAL SUPPORT

∗∗∗∗ OVERVOLTAGE ERROR CALL TECHNICAL SUPPORT

∗∗∗∗ ANODE VOLTAGE OUT OF TOLERANCE CALL TECHNICAL SUPPORT

∗∗∗∗ ANODE CURRENT OUT OF TOLERANCE CALL TECHNICAL SUPPORT

∗∗∗∗ CONTROL CONNECTOR CALL TECHNICAL SUPPORT

∗∗∗∗ PROBLEM IN THE REFERENCE VOLTAGE CALL TECHNICAL SUPPORT

MAJOR ERROR ALL FUNCTIONS ARE DISABLED CALL TECHNICAL SUPPORT

∗∗∗∗ xgenus® dc only

14. ERROR MESSAGES

Page 49: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

49

The x-ray control electronic equipment is protected by 4 fuses located on the electronic P.C.B. To replace them, proceed as follows (Fig. 25):

1. Turn off the power supply

2. Remove the front cover from the x-ray control by unscrewing the fixing screws

3. Identify the fuse to be replaced (Fig. 26)

4. Remove the plastic protection (Fig. 27)

5. Remove the fuse

6. Replace it with one of same type‡

TYPE OF TUBE HEAD CONNECTED xgenus® ac xgenus® dc POWER VOLTAGE

OF THE RADIOGRAPHIC SYSTEM 115V (-5% /+10%) 115V (±15%)

F1 (XRAY1) 8AF – 250V 12.5A – 250V F2 (XRAY2) 8AF – 250V 12.5A – 250V

Internal protection fuses (5x20) quick

(see schemes on § 19) F3 – F4 12.5A – 250V 12.5A – 250V

7. Put the fuse protection back in place.

Fig. 26 Fig. 27 8. Replace the x-ray control cover.

9. Restore the power supply. ‡ n° 2 spare fuses are supplied in a plastic bag inside the XGT control.

15. REPLACEMENT OF FUSES

Page 50: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 50

To guarantee the safety of the radiographic system, the following checks must be carried out:

CAUTION The owner is responsible for arranging the maintenance schedule. The radiographic system must undergo maintenance to check correct operation every 12 months. 1. Turn off the power 2. Release the spring tension of the arm “B” of the positioner using the wrench provided.

(Please see page 27) 3. Remove the radiographic generator.

(Refer to section 5.6) 4. Remove the wall plate cover.

(Refer to section 5.2) 5. Remove the terminal board cover and disconnect the cable 6. Remove the bracket plug and withdraw the guard slat.

(Refer to section 5.4) 7. Remove the positioner and the relevant cable from the bracket 8. Remove the bracket from the wall plate

(Refer to section 5.3) 9. Check the vertical alignment of the wall plate – adjust if required 10. Check the six screws on the wall plate – tighten if required 11. Clean off the old lubricating grease from the bracket shaft - should the bracket shaft be

damaged, install a new bracket 12. Clean the old lubricating grease off the bracket bushing - should the bracket bushing be

damaged, install a new bracket 13. Grease the bracket shaft (use grease MOLIKOTE D) 14. Lubricate the bracket bushing with lubrication grease (use grease MOLIKOTE D) 15. Install the bracket in the wall plate

(Refer to section 5.3) 16. Check the arm cable – should it be damaged, send the arm to the manufacturer for repairs 17. Check the arm covers 18. Replace if the guards are damaged 19. Clean the old grease off the shaft - should the shaft be damaged, send the arm to the

manufacturer/dealer for repairs. (Refer to section 5.4)

20. Lubricate the arm shaft with lubricating grease (use grease MOLIKOTE D) and reposition it in the bracket

21. Put again the arm cable in the bracket and the wall plate. Connect it to the terminal board and re-attach the terminal board cover (Refer to section 5.5)

22. Position the guard slat in the bracket 23. Position the bracket plug 24. Position the plate guard 25. Check the electric contact of the radiographic unit. If damaged, send the radiographic unit to

the manufacturer/dealer for repairs 26. Clean the old grease from the radiographic unit assembly shaft 27. Grease the assembly shaft of the radiographic unit with a thin layer of lubricating grease

(use grease MOLIKOTE D) 28. Position the radiographic unit again 29. Load the spring of the positioner arm “B” using the wrench provided

(Refer to 5.7) 30. Turn on the power and check the correct operation of the radiographic system.

16. MAINTENANCE

Page 51: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

51

Use a soft cloth dampened with water and mild soap to clean the outer surfaces of the x-ray. DO NOT permit water or any liquid to penetrate the machine or severe electrical component damage may occur. Damage of this nature is not covered under warranty. For protection against cross-infection, it is recommended that barrier protection be used and replaced for each patient. Germicides must be selected carefully, realizing that many are not suitable for use with such plastics that are used in manufacturing. Damage of this nature is not covered under warranty. In case of a malfunction, contact your local Technical Service Representative. They will call TG Group for an RA Number (Return Authorization). This will authorize one the return of the defective part to TG Group for replacement or repair. USING THE ORIGINAL PACKAGING Send all materials to: TG Group, Inc. 3505 Laird Road, Unit # 7 Mississauga, Ontario L5L 5Y7 Canada Tel. 1-877-557-4888 Fax 1-905-832-0732 Manufacturer: de Götzen S.r.l. Via Roma, 45 21057 OLGIATE OLONA VA - ITALY Tel. +39 0331 376760 r.a. Fax +39 0331 376763 E-mail: [email protected]

18. IF A REPAIR SHOULD BECOME NECESSARY

17. CLEANING THE OUTER SURFACES

Page 52: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

The manufacturer undertakes to supply, upon request, drawings, circuit diagrams, component partslists, instructions or other information needed by qualified technical personnel to perform repairs onthose parts of the “xgenus ac/dc” radiographic systems which may be repaired.

5252 xgenusxgenus ac/dc ac/dc

19. ATTACHMENTS

Page 53: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xgenusxgenus ac/dc ac/dc 5353

Page 54: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 54

A B C

REF SPARE PARTS TUBEHEAD ‘xgenus® ac’ CODE Q.TY PCS 1 AC XG (1) FORK PLASTIC COVER 29700983 7 2 AC/DC XG (2) RING NUT SET FOR THE

FORK 29700980 5

3 AC/DC XG (3) FORK 29700777 3 4 AC/DC XG (4) ROTATION LIMITING

DEVICE 29700750 1

5 AC XG (5) CABLE + 3 POLES CONNECTOR 39700302 1 A LONG CONE 31cm (12”) Ø60 29700397 1 B SHORT CONE 20cm (8”) Ø60 29700396 1 C RECTANGULAR CONE 31cm (12”) 44x35 29700615 1

Page 55: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

55

REF SPARE PARTS TUBEHEAD ‘xgenus® dc’ CODE Q.TY PCS 1 DC XG (1) FORK PLASTIC COVER 29700979 8 2 AC/DC XG(2) RING NUT SET FOR THE

FORK 29700980 5

3 AC/DC XG (3) FORK 29700777 3 4 AC/DC XG (4) ROTATION LIMITING

DEVICE 29700750 1

5 DC XG (5) WASHER 27500959 1 6 DC XG (6) AIR CARD 29700981 9 7 DC XG (7) AIR CARD COVER 29700982 11 8 DC XG (8) HEAD CABLE 39700385 1

Page 56: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 56

REF SPARE PARTS ‘xgenus® ECB’

(OPTIONAL) CODE Q.TY PCS

1 XG ECB (1) SUP & INF SHELL 29700978 4 2 MAIN BOARD WITH SWITCH AND LED 39700303 1 3 XG ECB (3) KEYLOCK SWITCH 35100298 1

Page 57: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

57

REF SPARE PARTS xgenus® LIGHT

(OPTIONAL) CODE Q.TY PCS

1 XG LIGHT (1) SUP & INF SHELL 29700977 4 2 XG LIGHT (2) LAMP HOLDER 33400372 1 3 XG LIGHT (3) LAMP 230V 15W 33300352 1 3 XG LIGHT (3) LAMP 115V 15W 33300390 1 4 XG LIGHT (4) 230V BOARD 39200354 3 4 XG LIGHT (4) 115V BOARD 39200353 3

Page 58: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 58

Ref SPARE PARTS TIMER ‘xgenus®’ ac/dc CODE Q.TY PCS 1 AC/DC XG TIMER (1) TEMPLATE 24200883 1 2 AC/DC XG TIMER (2) CONTROL BOARD 39200306 7 2 AC/DC XG TIMER (2) CONTROL BOARD US 39200307 7 3 AC/DC XG TIMER (3) XM COILED CABLE 32000325 1 4 AC/DC XG TIMER (4) XM DOUBLE COILED

CABLE 32000359 1

5 AC/DC XG TIMER (5) EXPOSURE BUTTON HOUSING

29700974 4

6 AC/DC XG TIMER (6) EXPOSURE BUTTON BOARD

39700303 1

7 AC/DC XG TIMER (7) LED BOARD 39200313 7 8 AC/DC XG TIMER (8) DEG KEYBOARD 39200308 1 8 AC/DC XG TIMER (8) SAT KEYBOARD 39200309 1 9 AC/DC XG TIMER (9) REAR COVER 29700975 1 10 AC/DC XG TIMER (10) DEG FRONT COVER 29700304 4 10 AC/DC XG TIMER (10) SAT FRONT COVER 29700976 4 11 AC/DC XG TIMER (11) KEYSWITCH LOCK 39700305 2 12 AC/DC XG TIMER (12) FILTER 36200382 3 13 GREEN BIPOLAR SWITCH 35000044 1 14 BIPOLAR SWITCH PROTECTION 33500228 1

Page 59: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

59

REF SPARE PARTS POSITINER ARM ‘xgenus®’ ac/dc

CODE QTY PCS

1 AC XG B (1) CABLE + 3 POLES CONNECTOR 39700384 1 2 DC XG B (2) CABLE + 5 POLES CONNECTOR 39700381 1 3 AC/DC XG B (3) HEAD SIDE SHELLS 29700990 9 4 POSITIONER ARM CONNECTING ROD ANT 29700460 4 5 POSITIONER ARM TIE ROD HOLDER 29700458 6 6 AC/DC XG B (6) CAPS (set) 29700991 7 7 AC/DC XG B (7) EXT COVER 27500757 1 8 AC/DC XG B (8) INT COVER 27500758 1 9 AC/DC XG B (9) CENTRAL SHELLS 29700993 8 10 POSITIONER ARM LOADING WRENCH 22000543 1 11 AC/DC XG B (11) INT COVER 27500758 1 12 AC/DC B (12) EXT COVER 27500757 1 13 POSITIONER ARM CONNECTING ROD POST 29700992 7 14 AC/DC XG B SAFETY KEY 22000176 1 15 AC/DC B (15) BRACKET SIDE SHELLS 29700994 8

Page 60: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 60

REF SPARE PARTS WALL PLATE 350 CODE Q.TY PCS 1 WALL BOX COVER (1) DEG 29700985 5 1 AC/ XG WALL BOX (1) COVER SAT 29700986 5 2 AC/DC XG B & WALL PLATE SAFETY KEY 22000176 1 3 WALL PLATE (3) TEMPLATE 24200984 1 4 WALL PLATE (4) INSERT + CAPS 29701036 3 WALL PLATE 350 DEG 29700383 COMPLETO AC/ XG WALL PLATE 350 SAT 29700987 COMPLETO

Page 61: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

61

REF SPARE PARTS WALL PLATE 350 TOP CODE

Q.TY PCS

1 WALL PLATE (1) COVER DEG 29700985 5 1 AC/ XG WALL PLATE (1) COVER SAT 29700986 5 2 AC/DC XG B & WALL PLATE SAFETY KEY 22000176 1 3 WALL PLATE (3) TEMPLATE TOP MOUNT 24201002 1 4 WALL PLATE (4) INSERT + CAPS 29701036 3 5 WALL PLATE (5) WASHER TOP MOUNT 22000597 1 WALL PLATE 350 TOP DEG 29700988 COMPLETO AC/ XG WALL PLATE 350 TOP SAT 29700989 COMPLETO

Page 62: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 62

Ref ARM SPARE PARTS CODE Q.TY PCS

BRACKET 400 29700380 COMPLETO BRACKET 800 29700378 COMPLETO BRACKET 1100 29700379 COMPLETO BRACKET 1100 DC XG 29701019 CPMPLETO

Ref ARM SPARE PARTS CODE Q.TY PCS

CEILING BRACKET 29700382 COMPLETO

Ref ARM SPARE PARTS CODE Q.TY PCS

UNIT BRACKET 29700381 COMPLETO

Ref ARM SPARE PARTS CODE Q.TY PCS

BRACKET 400 TOP MOUNT 29700620 COMPLETO BRACKET 800 TOP MOUNT 29700621 COMPLETO BRACKET 1100 TOP MOUNT 29700622 COMPLETO BRACKET 1100 TOP MOUNT DC XG 29701035 COMPLETO

Page 63: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc

63

D i c h i a r a z i o n e

d i C o n f o r m i t à

Declaration of Conformity Statement - Déclaration de Conformité Konformitäts-Erklärung - Déclaration de Conformidad de la

Impresa Company Società Firma Empresa

dichiara che il dispositivo medico sistema radiografico per diagnosi intra-orale declares that the medical device intraoral X-ray system declare que le dispositif medical appareil pour radioagnostic dentaire erklärt, daß das medizinisches Gerät Röntgendiagnose-Einheit declara que el dispositivo medico rayos-X infraorales modello type of machine modèle de machine bauart der maschine modelo de la máquina

xgenus®

è conforme ai requisiti delle direttive is in compliance with the directives

est conforme aux directives ist den Richtlinien gemäß

comple las directives

dispositivi medici medical devices dispositives medicaux medizinische geräte dispositivos medicos

93/42/CEE

organismo notificato notified body organisme notizie verantwortliche gestaltung organismo responsible

0434

certificato n° certificate No. certificat n° zertifikat Nr. certificato n°

2002 – OSL – MDD – 0044

compatibilità elettromagnetica e successive modifiche electromagnetic compatibility and subsequent amendments compatibilità electromagnétique et modifications suivantes elektromagnetische verordnung und folgenden Änderungen contabilidad electromagnetica

89/336/CEE

y modificaciones singuentes

bassa tensione low voltage basse tension niederspannung baja tension

73/23/CEE

Il dispositivo definito di classe IIb è conforme ai requisiti dell’allegato II della direttiva 93/42/CEE This product, definited as a class IIb device, is conformed to the requirements of annex II of the council directive 93/42/ECC

Le produit ci-mentionné, defini comme dispositif de classe IIb, est conforme aux requises de l’annex II de la directive 93/42/EEC Dieses Gerät, bestimmt als Klasse IIb, ist den richtilinien 93/42/EWG, annex II gemäß

El producto aqui mencionado esta definido como dispositivo de clase IIb y esta conformado a los requisitos de l’annex II de la directiva 93/42/CEE

data di emissione issuing date date de èmission ausstelldatum fecha de emisio

18.03.2002

Page 64: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

xxggeennuuss aacc//ddcc 64

D i c h i a r a z i o n e

d i C o n f o r m i t à

Declaration of Conformity Statement - Déclaration de Conformité Konformitäts-Erklärung - Déclaration de Conformidad de la

Impresa Company Società Firma Empresa

dichiara che il dispositivo medico sistema radiografico per diagnosi intra-orale declares that the medical device intraoral X-ray system declare que le dispositif medical appareil pour radioagnostic dentaire erklärt, daß das medizinisches Gerät Röntgendiagnose-Einheit declara que el dispositivo medico rayos-X infraorales modello type of machine modèle de machine bauart der maschine modelo de la máquina

xgenus® dc

è conforme ai requisiti delle direttive is in compliance with the directives

est conforme aux directives ist den Richtlinien gemäß

comple las directives

dispositivi medici medical devices dispositives medicaux medizinische geräte dispositivos medicos

93/42/CEE

organismo notificato notified body organisme notizie verantwortliche gestaltung organismo responsible

0434

certificato n° certificate No. certificat n° zertifikat Nr. certificato n°

2002 – OSL – MDD – 0044

compatibilità elettromagnetica e successive modifiche electromagnetic compatibility and subsequent amendments compatibilità electromagnétique et modifications suivantes elektromagnetische verordnung und folgenden Änderungen contabilidad electromagnetica

89/336/CEE

y modificaciones singuentes

bassa tensione low voltage basse tension Niederspannung baja tension

73/23/CEE

Il dispositivo definito di classe IIb è conforme ai requisiti dell’allegato II della direttiva 93/42/CEE This product, definited as a class IIb device, is conformed to the requirements of annex II of the council directive 93/42/ECC

Le produit ci-mentionné, defini comme dispositif de classe IIb, est conforme aux requises de l’annex II de la directive 93/42/EEC Dieses Gerät, bestimmt als Klasse IIb, ist den richtilinien 93/42/EWG, annex II gemäß

El producto aqui mencionado esta definido como dispositivo de clase IIb y esta conformado a los requisitos de l’annex II de la directiva 93/42/CEE

data di emissione issuing date date de èmission ausstelldatum fecha de emisio

25.06.2002

Page 65: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”
Page 66: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”
Page 67: The radiographic system described in this manual refers to a...xgenus ac/dcc 2 The radiographic system described in this manual refers to a “wall installation”. “de Götzen S.r.l.”

NOTES

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………

……………………………………………………………………………………………………………………………………………………