Upload
vuthu
View
240
Download
2
Embed Size (px)
Citation preview
THE NPO FATIGUE TESTER: THE DESIGN & DEVELOPMENT OF A
NEW DEVICE FOR TESTING PROSTHETIC FEET
Tara Ziolo, B.Sc.E
Rad Zdero, Ph.D
Tim Bryant, Ph.D, P.Eng
THE NPO FATIGUE TESTERTM: THE DESIGN AND DEVELOPMENT OF
A NEW DEVICE FOR TESTING PROSTHETIC FEET
Tara Ziolo, B.Sc.E
Radovan Zdero, Ph.D
J. Timothy Bryant, Ph.D, P.Eng
Human Mobility Research Centre (HMRC) Apps Medical Research Centre, Angada 1
Kingston General Hospital, Kingston, ON, Canada and
Mechanical Engineering Department Queen’s University, Kingston, ON, Canada
(i)
The NPO Fatigue TesterTM Ziolo, Zdero, and Bryant (2001)
EXECUTIVE SUMMARY
The ongoing problem of amputations resulting from land mines in developing and post-
conflict countries such as El Salvador, Nicaragua, Cambodia, Croatia, and Angola provides
the impetus for the current study. Although the need for community based rehabilitation in
these areas is recognized as advantageous, it is carried out often without the use of
appropriate context-sensitive prosthetic technology.
To this end, the overall purpose of the present ongoing investigation is the development
of a lower limb prosthetic system - starting with the foot - that is affordable for the target
population, highly functional, esthetically and culturally acceptable to users, and
scientifically validated through a series of clinical, field, and laboratory studies.
Briefly, this document outlines the scientific work done to date in the development of the
Niagara Foot, provides a literature survey of previous mechanical cyclic fatigue testers for
assessing prosthetic feet, describes the design and development of the current NPO Fatigue
Tester, discusses the preliminary test results for Niagara and SACH feet using the this tester,
and gives consideration to future work.
This work is being done by the Human Mobility Research Centre (Kingston General
Hospital and Queen’s University, Kingston, ON) in conjunction with Niagara Prosthetics and
Orthotics (St. Catherines, ON).
(ii)
The NPO Fatigue TesterTM Ziolo, Zdero, and Bryant (2001)
TABLE OF CONTENTS
(iii)
EXECUTIVE SUMMARY (ii) 1 BACKGROUND 1 1.1 The Need 1 1.2 The Rehabilitation Goal 1 1.3 The Technical Goal 1 2 WORK TO DATE SUMMARY 2 2.1 Niagara Foot Shape 2 2.2 Material Selection 2 2.3 Stress Analysis 2 2.4 Action Tests 2 2.5 Clinical Trials 5 2.6 Final Design Optimization 5 3 REVIEW OF FATIGUE TESTER LITERATURE 6 4 DESIGN OF THE NPO FATIGUE TESTER 11 4.1 Mechanical Components 11 4.2 Pneumatics 15 4.3 Electrical/Control System 15 4.4 Electrical/Computer System 19 5 OPERATING PROCEDURES FOR THE NPO FATIGUE TESTER 20 5.1 Preparation of the Prosthesis for Testing 20 5.2 Test Startup Procedure 20 5.3 Test Conditions 22 6 A PILOT STUDY USING THE NPO FATIGUE TESTER 23 6.1 Materials and Methods 23 6.2 Results and Discussion 23 7 INITIAL MATERIALS TEST RESULTS 25 7.1 Summary 25 7.2 Proof Testing 25 7.3 Cyclic Testing 26 8 FUTURE WORK 28 8.1 Validation and Calibration of the NPO Fatigue Tester 28 8.2 Waveform Modifications 30 8.3 Environmental Durability of the Niagara Foot 30 8.4 Complete Sized Modular System 31 8.5 Cosmetic Cover for Foot 32 8.6 Modification of Test Stands 32 9 FINAL REMARKS 33 REFERENCES 34
APPENDIX 1 – ISO Testing Standards for Prostheses
APPENDIX 2 – U.S. Patent for the Niagara Foot APPENDIX 3 – Material Properties APPENDIX 4 – PPC Calibration Tables APPENDIX 5 – Control System (Technical Report)
The NPO Fatigue TesterTM Ziolo, Zdero, and Bryant (2001)
APPENDIX 6 – Test Stand Modifications APPENDIX 7 – Results of Proof Testing APPENDIX 8 – Inspection Sheets for Cyclic Testing APPENDIX 9 – Cyclic Fatigue Testing Report APPENDIX 10 – Deformation of Feet During Cyclic Testing
(iii)
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
1. BACKGROUND 1.1 The Need The current project’s motivation is the ongoing problem of amputations resulting from land
mines, especially in developing and post-conflict countries such as El Salvador, Nicaragua,
Cambodia, Croatia, and Angola. About 70 people per day are killed or injured due to
landmines, of which 110 million still lay active in the ground on every continent
(www.oneworld.org/guides/landmines/stats.html). The Canadian government, universities,
service organizations, and the private sector have been leaders in addressing this problem.
1.2 The Rehabilitation Goal
The uniqueness of the current project is two-fold. First, most governments have invested in
programs promoting dependence on costly solutions and centralized institutional approaches
to rehabilitation, even though the majority of landmine survivors are from rural areas. As an
alternative, the internationally recognized concept of Community Based Rehabilitation
(CBR) develops innovative programs that help disabled people rely less on centralized social
and health systems, live more independently, and be re-integrated into their communities.
This project uses CBR as its philosophy to address equipment needs for disabled people.
Second, prostheses typically provided have been inappropriate for developing and post-
conflict countries, in that they have been far too expensive for poor clients, especially since
they must be replaced frequently throughout one’s life, and were designed to fit urban
Western environmental conditions. On the other hand, the Niagara FootTM (Niagara
Prosthetics and Orthotics, St. Catherines, ON, Canada), was designed to be highly functional
in these environments, but at only 1/10TH the price of other commercial prosthetic feet.
1.3 The Technical Goal
This report describes the design and development of a mechanical cyclic fatigue tester,
namely the NPO Fatigue TesterTM, to assess prosthetic feet as a predictor of field service life.
Specifically, using ISO (International Organization for Standardization) testing standards
(Appendix 1), the Niagara FootTM is to be compared to and tested in parallel with a popular
commercially available foot, namely the SACH (Solid Ankle Cushion Heel), which is often
exported from the West to places of need.
- 1 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
2. WORK TO DATE SUMMARY The Niagara FootTM was designed by the Human Mobility Research Centre (HMRC, Queen's
University and Kingston General Hospital, Kingston, Canada) and the Ergonomics Research
Group (ERG, Queen's University, Kingston, Canada) under contract with Niagara Prosthetics
and Orthotics (NPO).
2.1 Niagara FootTM Shape
The first step was the design and construction of an inexpensive and high-function foot
component, a lengthy iterative process, modular with existing prosthesis shafts being used in
the host nation. The modular foot component is unique because of its single unit construction
and its ability to be mass-produced in a few standard sizes (Fig. 1).
2.2 Material Selection
The material used to manufacture the Niagara foot is another critical factor in its success.
After considering the stress analysis effects and load-deflection characteristics of a wide
variety of potential materials, a preliminary shape of the Niagara foot [Series 1] was
developed (Fig. 1a) composed of a medium modulus injection-moldable thermoplastic (Tsai,
1999). The selection was done in tandem with the stress analysis described below.
2.3 Stress Analysis
The Niagara Series 1 foot shape was assessed using computer Finite Element Analysis (FEA)
by our group for a variety of materials (Ziolo, 1999). The model predicts the mechanical
stress levels the foot would experience under extreme conditions of normal use, i.e. at heel
strike and toe off. Heel strike load (700 N) was applied at 15° whereas toe off load (700 N)
was applied at 45°. From Fig. 2, it was evident that for a given set of material properties, the
inner-C portion was the most susceptible to experiencing highest stress.
2.4 Action Tests
The FEA model was verified with a load-deflection test using an Instron tester (Tsai, 1999).
The Niagara [Series 1] foot was instrumented with IREDs (infra-red emitting diodes) and
mounted onto the Instron (Fig. 3). A 5-70 kg cyclic force was applied separately at the toe
and heel at 45° and 15°, respectively, and the resulting angular deflections videotaped and
- 2 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
(b)
(a)
(c)
Fig. 1. Design Iterations of the Niagara Foot. (a) Niagara Foot [Series 1] (b) Niagara Foot [Series 2] (c) Niagara Foot [Series 3] final design shape being tested currently.
Figure 2. FEA Stress Analysis on the Niagara Foot [Series 1] prosthesis. The inner-C section experiences the highest von Mises stress levels.
- 3 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
(a) (b)
Fig. 3. Load-Deflection Tests on the Niagara Foot [Series 1]. Forces were applied at (a) toe and (b) heel and the deflections were determined.
Computer
Infrared EmittingDiodes (IREDS)
(a) (b) Fig. 4. Clinical Gait Trials on the Niagara Foot [Series 1] and SACH. (a) clinical gait test setup (from Noce-Kirkwood, 1997). (b) photo of a below knee amputee subject walking with the Niagara Foot [Series 1]. Results showed that the Niagara Foot [Series 1] is a highly functional foot with positive feedback from subjects, equal to or better than the SACH.
- 4 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
later analyzed. The action tests yielded a toe stiffness result of 24 kN/m compared to the FEA
model result of 14.8 kN/m.
2.5 Clinical Trials
The Niagara Series 1 and a SACH type foot were clinically tested with 4 amputees (Potter
2000; Potter et al., 1999). These quantitative tests involved videotape footage and computer
controlled gait analysis equipment (Fig. 4). Results indicated no statistical difference (p >
0.05) in the velocity (Niagara Series 1 - 67.01 m/min, SACH - 64.54 m/min) and cadence
(Niagara Series 1 - 100.38 steps/min, SACH - 101.38 steps/min) between the prostheses.
However, the Niagara Series 1 statistically outperformed (p < 0.05) the SACH with respect to
stride length (Niagara Series 1 - 1.33 m, SACH - 1.27 m) and % stance phase (Niagara Series
1 - 61.24 %, SACH - 64.95 %). Qualitative feedback from the subjects, using a standardized
questionnaire and personal interview, showed amputee satisfaction with the Niagara foot in
terms of stability, effectiveness, and comfort. Subjects indicated no significant difference
between the Niagara Series 1 and SACH.
2.6 Final Design Optimization
Because the foot stiffness target was 50 kN/m after the ISPO (1978) standard, the Niagara
Series 1 shape was further optimized to arrive at the Niagara Series 2 by Ziolo (1999) (Fig.
1b). Subsequent patient considerations and stress analysis led to further design modifications,
which concluded with the Niagara Foot [Series 3] shape currently being used (Fig. 1c), full
technical for which can be found in Appendix 2. The final material chosen was the acetal
resin Delrin 150E, the material and mechanical properties for which can be found in
Appendix 3.
- 5 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
3. REVIEW OF FATIGUE TESTER LITERATURE In addition to the numerous clinical gait studies of prosthetic lower limbs and feet have been
conducted (Barth et al., 1991; Macfarlane et al., 1990; Menard and Murray, 1989; Pitkin,
1995; Potter et al., 1999; Valenti, 1990; Yamamoto et al., 1992), a survey of the literature has
revealed several studies involving mechanical fatigue testing of prosthetic feet or other
orthotic devices.
The earliest fatigue tester the current authors are aware of was developed by Daher
(1975). It consisted of a Scotch-Yoke mechanism simulating the kinematics of the knee and
foot, providing both the swing and stance phases of gait (Fig. 5). The tester had a mobile flat
ground plate with which foot contact was made as the mechanism cycled through, with heel-
strike and toe-off occurring at 20° and 35°, respectively. Load levels were controlled by
timed cams that cyclically pressurized the air contained in pneumatic cylinders. Stance load
levels were maintained at a relatively constant level until just prior to toe-off. Daher’s study
involved evaluating 9 types of SACH feet, applying socks and shoes to two simultaneously
tested prosthetic feet, which were initially aligned at 5° toe-out. The feet were cycled up to
500,000 cycles and maximum loads of 100 kg (981 N), with several load-vs-deflection
checks during the entire test regimen. Most of the permanent deformation, due to compacting
and breakdown of foam, occurred within the first 5000 cycles as evidenced from X-rays of
the feet and hysteresis in the load-deflection curves. The major advantages of this system
were that it was force controlled in order to reproduce general loading pattern and had the
capability of testing two feet simultaneously. However, the axial load levels were far below
the peak levels normally experienced by prosthetic feet during in-service use (approx. 1350
N) for a 150 lb person, thereby artificially extending the longevity of the prostheses tested.
A displacement control device previously developed by our group consisted of a cam-
driven rocker ground plate that created 17-20° and 57-61° angles, respectively, with
prosthetic test feet at heel-strike (HS) and toe-off (TO) positions, as shown in Fig. 6
(Singleton, 1983; Wevers and Durance, 1987). The tester was capable of durability testing of
an entire foot-shank-socket complex. The ground plate had adjustable inversion/eversion
settings in order to test the effect of improper prosthetic fitting or use when desired; the effect
of this would normally be felt at the socket or contralateral hip. It was driven by a 110 V
- 6 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
Heel Strike Toe OffRotating Pin
Fixed Pin
Mobile Ground Plate
Disk Disk
GroundPlate
A B
Fig. 5. Fatigue Tester as proposed by Daher (1975) (A) schematic (B) photo of tester.
b
c
c
a
d
e f
g
Fig. 6. Fatigue Tester designed and constructed by Singleton (1983) and Wevers and Durance (1987): [a] - shaper crank, [b] - cam, [c] - sliders, [d] - connecting rod, [e] - upper crank, [f] - rocker plate, [g] - spring loaded stop.
- 7 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
synchronous electric motor through a reduction gear box, with displacement control cams
ensuring proper kinematics during 1 Hz cycling was simulated. Due to softening of prosthetic
feet during repetitive motion and in order to meet the ISPO standard 1350 N maximum axial
load, a force sensor was mounted on the shank support bar to provided force feedback. In
between TO and HS, no load was applied to simulate the swing phase. This 0-1350 N loading
pattern was cycled a minimum of 3 million times or until prosthesis failure, evidenced by
either sudden catastrophic failure of the metal Otto Bock foot-shank connector, the
appearance of cracks in the plastic socket, or cracking in the soles of the feet. In their study,
Wevers and Durance (1987) tested 4 different SACH feet and found that they showed cracks
in the soles at between 43,800 and 105,000, reaching only 1.5-3.5 % of the 3 million cycles
required by ISPO and ISO standards. Plastic residual limb sockets tested separately,
however, reached an average of 415,000 cycles before failure. The appeal of this tester is that
it both simulated the load transfer from heel to toe and provided material recovery during
swing phase, mimicking actual use more closely and was able to test durability of an entire
foot-shank-socket complex. However, because loading patterns and levels were dependent on
accurate cam design, the authors recommended future modifications to include the use of
hydraulic semi-controlled machine.
Carlson et al. (1990) developed a durability tester designed to test ankle-foot orthoses
(AFOs), specifically those constructed from polyurethane and polypropylene. It consisted of
a central mounting shaft onto which four AFOs were attached at the foot segment and
weighted at the shank end (Fig. 7). As the entire system cycled at 14 cycles/min (0.23 Hz),
the AFOs were rotated backwards causing maximum dorsiflexion of up to 250°. As the
AFOs were rotated over the top of the cycle, the weighted shank would suddenly fall,
‘banging’ the orthosis against its plantarflexion stop. The inertia of the shank weights (1 and
1.95 kg-m) was such that they created significant sheer, torsion, and tension ‘shock’
transmitted to the attachment interface during the sudden ‘bang’. The failure criterion was the
creation of fatigue cracks at the flexural interface. Although the apparatus did not yield
results capable of predicting AFO service life, it was a good relative measure in comparing
durability of AFOs of varying designs and material compositions.
- 8 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
Fig. 7. Cyclic Fatigue Tester for ankle-foot orthoses (AFOs) developed by Carlson et al. (1990).
E
Fig. 8. Simplified Fatigue Tester proposed by Toh et al. (1993): A - plate, B - angle block, C - cam, D - load cell, E - variable speed induction motor.
- 9 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
A more recent approach by Toh et al. (1993) has been to avoid simulating the complex
loading patterns experienced during normal walking and move to a simpler apparatus in
which the toe and heel are tested separately up to an appropriate peak axial load level (Fig.
8). The toe and heel were tested with sinusoidal cyclic axial loads peaking at 1.5 x body
weight (approx. 600 N) at 2 Hz up to 500 000 cycles. Blocks wedges of 30° and 15° were
used to create inclines for testing the prosthesis forefoot and heel, respectively. Static load
deflection tests were conducted between cyclic runs to detect mechanical property changes in
the foot. No impact loads simulating external events or protective footwear was used. Their
results for the Lambda foot and the Kingsley and Proteor SACH feet were consistent with
other studies using more complex fatigue tester systems. Although the appeal of this system
lies in its simplicity and is capable of testing individual prosthetic feet, it would need to be
significantly modified to mimic more closely actual ground reaction loading patterns and to
test foot-shank-socket complexes.
A commercially available system, the Universal Testing Machine (UTM) from Endolite-
Blatchford (Hampshire, UK, www.endolite.com) is a structural testing device capable of
testing one or two prosthetic feet simultaneously. Typical technical specifications include: 1
Hz cycling frequency, 5000 N maximum axial force, 60 mm stroke length at toe off, 30 mm
stroke length at heel strike, 400-500 kg weight, 1.3x1.6 m size, and 110/240 VAC and 4 amp
power requirements. The tester is available at a cost of $ 7000 U.S. (£ 3500), is computer
controlled with pneumatic actuators requiring 325-650 L/min of air, and capable of up to
5000 N of axial load. Although available commercially, the manufacturer’s product
catalogue does not disclose criteria used to indicate prosthesis failure, calibration procedures,
or hysteresis effects. The current authors are unaware of any validation tests or applications
of this unit described in the scientific literature.
- 10 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
4. DESIGN OF THE NPO FATIGUE TESTERTM The following description of the current tester is that of a working model which will include
several design modifications, as described below in the Section 7, Future Work.4.
4.1 Mechanical Test Stands and Control Panels
Test Stations
Each cyclic test station (Fig. 9a) consists of a pair of pneumatic cylinders (TRD cylinders, 2.5”
bore, 3” stroke, bumper piston, over-sized rods; Cowper Inc., Kingston, ON, Canada), a prosthetic
foot-shank assembly, support brackets for the shank, a non-resetable mechanical stroke
counter (McMaster Carr, www.mcmaster.com) that records the number of test cycles completed,
and a magnetic limit switch (Model MRS-0.087-PXBL, 24 VDC reed switch, 2 wires with LED;
Cowper Inc., Kingston, ON, Canada) that detects excessive cylinder stroke length and, hence,
excessive deflection at the toe and heel. These components are mounted onto a 24”x30”x3/8”
aluminum plate (6061-T6), which is in turn screw mounted onto a test stand consisting of a
2”x4” frame secured onto a ¾” reinforced base. In total, there are two test stands housing
four testing stations (Fig. 9b).
Pneumatic and Electrical/Control Panels
Atop the test stands sit two 1”x6” boards onto which are secured the pneumatic and control
system components (Fig. 10a). The pneumatic components consist of 8 solenoid valves
(MAC series 55B; 24 VDC, 3/8” NPT ports, internal pilot, normally closed; Cowper Inc.) and 3
rapid control manifolds (Model RC303754; 0.5” inlet, 3/8” outlet, Cowper Inc.). The control
system is comprised of the main control box (Fig. 10b), DPDT relays (Daltco, Kingston, ON,
Canada), and a 24 VDC-2 Amp power supplies (Entrelec type, www.entrelec.com, Daltco,
Kingston, ON, Canada).
Electrical/Computer Panel
This panel (Fig. 11) is identical to the test station frames above, except that the aluminum
plate has been replaced by a ¾” plywood plate, onto which are mounted the 120 VAC supply
box, a 24 VDC-2.1 Amp power supply (IDEC switching type, A-Tech Instruments Ltd.,
Scarborough, ON, Canada), two signal conditioning modules (DSCA 41, DIN rail mount,
analogue, +/- 10 VDC out; purchased from A-Tech Instruments Ltd.), a back pressure regulator
- 11 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
(a) (b) Fig. 9. NPO Fatigue Tester Station. (a) Layout of components, including Niagara Foot [Series 3], shank and support brackets, 2 pneumatic cylinders, 2 magnetic limit switches, and a mechanical stroke counter. (b) Typical test frame, each of which houses 2 test stations.
- 12 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
(a) (b)
Fig. 10. Pneumatics and Electrical/Control System (a) panel at top and bottom of figure, respectively. (b) the control box.
- 13 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
Fig. 11. Electrical/Computer Panel. Mounted are a power supply box, a 24 VDC-2.1 Amp power supply, two signal conditioning modules, a back pressure regulator, and two proportional pressure controllers (PPCs).
- 14 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
(Wilkerson Model R21-04-L00, 0.5” NPT; Cowper Inc., Kingston, ON, Canada), and two
proportional pressure controllers (PPCs) (MAC valve, 0-10 VDC control voltage, 0-100 psi;
Cowper Inc., Kingston, ON, Canada).
4.2 Pneumatics
The pneumatic network (Fig. 12) is powered by an air compressor (6.5 hp, 60 gal, vertical),
which acts as the main air supply for all test stations. The air pressure is regulated before
passing through an air filter (Wilkerson model VF18-04, Cowper Inc., Kingston, ON, Canada)
used to remove any moisture from the supply. The main supply line then branches from a
back pressure regulator and the two PPCs, one activating heel cylinders at all 4 stations and
the other the toe cylinders at all 4 stations. Thus, the cylinders apply load to the heel and toe.
The heel and toe PPCs each have their own manifold, which branch one air-flow inlet into 4
outlets. Each outlet is connected to a dedicated solenoid before being attached to the cylinder
inlet port. The solenoids are normally closed and must be powered to open and allow air to
flow into the cylinder inlet port.
In order to return the cylinders to an ‘off’ or ‘non-load’ position a small amount of back
pressure, controlled by the back pressure regulator, is required. Various combinations of
compressor pressure and back pressure, determined from a calibration table (Appendix 4),
can yield identical forces desired for a test.
The force and, hence, the speed at which the cylinders are retracted depend on the
application of a minimum back pressure level. The back pressure regulator is connected into
a manifold which has 1 inlet and 8 outlets that are connected into the exhaust ports of the
respective cylinders.
4.3 Electrical / Control System
Control Box
The operation of the entire fatigue tester is dictated by the control box (Figs. 10b and 13),
details for which are found in Appendix 5. The key switch controls power on/off, which
activates the green ‘Power On’ and ‘Ready’ lights. The red ‘Off’ switch is a momentary
contact switch that shuts down the control and pneumatics systems by cutting all electrical
power, causing solenoid valves to close and, hence, removing air supply to the cylinders.
- 15 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
1
357
0
123
4
567
024
6
SOL1
SOL3
SOL7
SOL5
SOL0
SOL2
SOL6
SOL4
1 2
1 2
1 2
1 2
1 2
1 2
1 2
1 2
To Cylinder Exhaust 0To Cylinder Exhaust 1To Cylinder Exhaust 2To Cylinder Exhaust 3
To Cylinder Exhaust 4
To Cylinder Exhaust 5To Cylinder Exhaust 6
To Cylinder Exhaust 7
To Cylinder Inlet 0
To Cylinder Inlet 2
To Cylinder Inlet 4
To Cylinder Inlet 6
To Cylinder Inlet 1
To Cylinder Inlet 3
To Cylinder Inlet 5
To Cylinder Inlet 7
Heel Manifold
ToeManifold
ExhaustManifold
Regulator
Back PressureRegulator
Air Filter
Compressor(Main Supply)
ProportionalPressure
Controller
ProportionalPressure
Controller
Solenoids
PNEUMATICS
PPC 0
PPC 1
1
357
1
357
0
123
4
567
0
123
0
123
4
567
4
567
024
6
024
6
SOL1
SOL3
SOL7
SOL5
SOL1
SOL3
SOL7
SOL5
SOL0
SOL2
SOL6
SOL4
SOL0
SOL2
SOL6
SOL4
1 2
1 2
1 2
1 2
1 2
1 2
1 2
1 2
To Cylinder Exhaust 0To Cylinder Exhaust 1To Cylinder Exhaust 2To Cylinder Exhaust 3
To Cylinder Exhaust 4
To Cylinder Exhaust 5To Cylinder Exhaust 6
To Cylinder Exhaust 7
To Cylinder Inlet 0
To Cylinder Inlet 2
To Cylinder Inlet 4
To Cylinder Inlet 6
To Cylinder Inlet 1
To Cylinder Inlet 3
To Cylinder Inlet 5
To Cylinder Inlet 7
Heel Manifold
ToeManifold
ExhaustManifold
Regulator
Back PressureRegulator
Air Filter
Compressor(Main Supply)
ProportionalPressure
Controller
ProportionalPressure
Controller
Solenoids
PNEUMATICS
PPC 0
PPC 1
Fig. 12. Schematic Layout for the Pneumatics system.
- 16 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
Fig. 13. Schematic Layout for Electrical/Control box.
- 17 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
5 6 7 8
1 2 3 4
SignalConditioner
0
5 6 7 8
1 2 3 4
SignalConditioner
1
xNC
xNC
NCx
- V+ V
24 VDC PowerSupply
N L
fromPPCs0 & 1
fromPPCs0 & 1
from PPC 1
Wires to Termination Board
from PPC 0
+ -
From 120 VPower Supply
1 – White +CMD2 – Red LMS23 – Green Common 6 Wires4 – Yellow LMS15 – Black Power 24 V6 – Blue -CMD
2 and 4 are Capped
PPC 0
1 – White +CMD2 – Red LMS23 – Green Common 6 Wires4 – Yellow LMS15 – Black Power 24 V6 – Blue -CMD
2 and 4 are Capped
PPC 1
ProportionalPressureControllers(PPCs)
x
ClearWhiteGreen
BlackRed
Analog Output - Green & WhiteGround - ClearAnalog Input - Black (in #1)
- Red (in #2)
TerminationPanel
ELECTRICAL / COMPUTER PANEL
Fig. 14. Schematic Layout for the Electrical/Computer system.
- 18 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
Simultaneously, the red ‘Warning’ light is activated while the green ‘Ready’ light is turned
off, signifying there is an error or problem in the system. In order to reset the system, the
momentary contact ‘System Reset’ switch must be pressed, thus opening the solenoid valves.
The ‘Limit Over-ride’ toggle switch either sets or over-rides the magnetic limit switches,
described above. The control box is powered by one of two 24 VDC power supplies.
Relays
Each test station is equipped with a double-pole double-throw (DPDT) relay that controls the
switches and lights of the control box, as well as the solenoid valves and limit switches.
4.4 Electrical / Computer System
Signal Conditioners
Two signal conditioners (Fig. 14), one for each of the two PPCs, reduce signal noise levels
and remove potential disturbances caused by the 60 Hz frequency prior to the signal being
sent to the computer. The PPCs and signal conditioners are powered by a 24 VDC power
supply.
Computer System and Interface
All electrical connections to the computer occur via the termination panel (Model PCI-20429T,
Intelligent Instrumentation), which is plugged into the internal A/D data acquisition multi-
function board (Model PCI-20428W, Intelligent Instrumentation). The ‘Visual Designer’
software (Version 4 for Windows from Intelligent Instrumentation; purchased from A-Tech
Instruments Ltd., Scarborough, ON, Canada) creates sine waves used by the PPCs for cycling the
cylinders on and off in a sinusoidal wave pattern, and is used to collect and process raw data.
- 19 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
5. OPERATING PROCEDURES FOR THE NPO FATIGUE TESTERTM The NPO fatigue tester experimental protocol can be described in terms of preparation of the
prosthesis for testing, test startup procedure, and test conditions.
5.1 Preparation of the Prosthesis for Testing Step 1. Foot Assembly (Fig. 15)
attach Otto Bock connector to foot
attach Otto Bock sleeve to shank
attach foot and shank
Step 2. Foot Alignment
all testing to be done on foot/shank with respect to ‘normal’ alignment:
o foot - 6° heel from horizontal (sagittal plane), no inversion/eversion
o shank - perpendicular to ground in frontal and sagittal planes
Step 3. Foot Mounting
remove nuts and washers from pylon connecting brackets
remove top piece of bracket and insert shank loosely
push cylinders to rest position (minimum stroke)
position foot such that acorn nut on top of heel cylinder (when cylinder is down) is approx. 10 mm from heel contact point; toe is about 1.5 inches from toe cylinder at this position
5.2 Test Startup Procedure Step 1. Initial Startup
open valve on air filter that is attached to air compressor
turn the compressor regulator on
plug in air compressor and power supply switch
turn on computer
Step 2. PPCs (Proportional Pressure Controllers)
set compressor regulator (which controls cylinder ‘on’ action) to desired pressure
set back stream pressure regulator (which controls cylinder ‘off’ action) to desired pressure
Note: pressures are different for each test depending on desired load (Appendix 4)
Step 3. Computer Control
execute ‘Visual Designer/Diagram’ and ‘Visual Designer/Run’ software programs which initiate sine wave form for actuator
- 20 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
Fig. 15. Exploded view of the preparation of the Niagara [Series 3] foot-shank assembly for fatigue testing.
Time (s)
Load (N) Load Waveform for Fatigue Testing 970 N
50 N load ramp up
load ramp down
Fig. 16. Representative Load Waveform during Cycling Fatigue Testing.
- 21 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
5.3 Test Conditions The following test conditions will be applied in the current study and, where cited, comply
directly with standards recommended by the International Organization for Standardization
(ISO, 1996), the full document being given in Appendix 1:
Temperature: approx. 23 °C or room temperature
Cycling Frequency: 1 Hz (since using plastic; ISO 2-6.4)
No. Cycles: 3 million cycles or until failure (ISO 4, Table 6)
Loads: cycle between 50-970 N (ISO 4, Table 6; ISO 3-4.4)
Cycling Waveform: sine wave (ISO 3-7.1, Fig. 3) as in Fig. 16
Niagara FootTM Size
o prosthesis rated for 60 kg person (A60 person; ISO 3-4.4), being the weight range of average amputee in the current target country (Cambodia)
o 24 cm length available (ISO 2-5.2)
Force-Foot Angles:
o heel-strike (15°) and toe-off (20°) (ISO 5-7.5, Fig 1; ISO 6-4, Table 7)
o angles cycled relative to ‘normal’ shank alignment described above
Failure Criteria for Cycling Testing:
o Test Completion: a test is considered complete when ‘failure’ occurs prior to reaching the prescribed 3 million cycles or when the 3 million cycles has been achieved with no failure of the test sample; ‘failure’ is defined and quantified in terms of deflection and force levels reached during (ISO 3-4.2, Table 6; ISO 3-7.3)
o Deflections: monitored by magnetic ‘reed’ limit switch, which is a band attached to the piston and a sensor placed on the cylinder surface; it is adjustable and shuts down solenoid valves at that station (ISO 3-5.4)
o Forces: axial forces on tibia/shank to be monitored with strain gages and not exceed critical values (ISO 4-4.2, Table 6)
- 22 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
6. A PILOT STUDY USING THE NPO FATIGUE TESTERTM This section summarizes briefly the results of preliminary fatigue tests performed on the
Niagara and SACH feet using the NPO Fatigue Tester.
6.1 Materials and Methods
Prosthetic Feet: Two variants of the Niagara Foot [Series 3] were tested. Niagara Foot A was
cut from a block of Delrin 150E using a water jet technique and was non-annealed. Niagara
Foot A* was also water jet cut from a block of the same material, but was also air annealed at
160°C (320°F) for 1.25 hours. Both feet were cut to the basic Niagara Foot shape and size
described earlier (Fig. 17a), being rated for a 60 kg individual and having a 24 cm length.
Full technical details of the U.S. patent for the foot can be found in Appendix 2. For
comparison, a Solid Ankle Cushion Heel (SACH) prosthetic foot (Fig. 17b) was also tested.
Test Protocol: The general test protocol and conditions were those described above in
Section 5, with the specific compressor and back stream pressures set at 80 and 25 psi,
respectively, to achieve a sinusoidal load range of 50-970 N for all runs. The SACH feet are
tested using a flat plate indenter, instead of the spherically ended indenters used for the
Niagara foot, to recreate desired load patterns because the SACH surface material is much
softer.
6.2 Results and Discussion
Results are summarized in Table 1. The tests reveal no catastrophic failure or signs of crack
initiation up to 63,000 cycles for the Niagara Foot. The current SACH foot is still undergoing
testing and has currently reached the 20,000 cycle mark, being too early to draw any
conclusions. However, as a point of comparison, the results of Wevers and Durance (1987)
on 4 different SACH feet betrayed cracks and early failure in the soles between 43,800 and
105,000 cycles. It is premature to perceive these results as an indication of the potential long-
term performance of the Niagara Foot relative to the SACH. Varying test conditions between
the studies may have played a role in the apparent differences: the current peak load levels
(50-970 N) were slightly below those of Wevers and Durance (1000-1350 N) for a given
loading cycle; the waveform of the present regime does not include a zero-load swing phase
but is rather continuous cycling of 50-970 N, unlike the SACH study which incorporates
- 23 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
swing; current test angles (heel-strike 15°, toe-off 20°) were slightly different from the
SACH study (heel-strike 17-20°, toe-off 57-61°). True comparison between prostheses
requires testing under exactly the same test conditions. Only further testing toward the 3
million loading cycles recommended by ISO and ISPO will reveal the actual mode(s) of
failure or wear of the Niagara and SACH feet, the extent of material softening, and the
effects of prosthesis construction and heat treatment method on performance.
(a) (b)
Fig. 17. Prosthetic Feet Tested (a) Niagara Foot [Series 3] (b) SACH Foot.
Table 1. Preliminary Test Results for Niagara and SACH Feet
Foot Type
Cycling Frequency
(Hz)
No. Cycles
Load Waveform
(N) Comments
Niagara Foot A 1 30,000 50-970 no signs of failure cracks; superficial
polishing tracks at toe and heel contact Niagara Foot A* 1 63,000 50-970 no signs of failure cracks; superficial
polishing tracks at toe and heel contact
SACH 1 20,000
(testing in progress)
50-970 at present, no signs of failure cracks or polishing tracks
- 24 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
7. INITIAL MATERIALS TEST RESULTS
7.1 Summary
There are two different types of testing that the ISO Standards require. One being the static
proof test or failure test and only once test samples have passed that test can they be cyclic
tested, which is the second type of testing required. (ISO 4-4.2)
After preliminary analysis at Queen’s University on the possible family of materials to use in
for the Niagara Foot, experts from Dupont were contacted and have been involved with the
final material selection procedures for the Niagara Foot.
7.2 Proof Testing
Protocol
The foot to be tested is inverted and mounted to a metal plate using the same mounting hole
as the connector for the completed assembly. The feet are held in place using a tool vice set
to the required angle for heel strike and toe off (15o and 200) respectively as stated in the ISO
Standards.
Fig. 18. Set up for Proof Testing: Exploded view (left) and assembled view (right) of Niagara
Foot and mounting plate
- 25 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
Fig. 19. Setup of the Instron for Proof Testing: Loads and deflections upon failure of the feet
at both heel strike (left) and toe-off (right) were determined
Testing was terminated when either the feet broke (catastrophic failure) or the deflection was
so great as to cause the heel to touch the back of the C-section or to cause the top of the toe to
touch the tool vice.
Prior to Dupont’s work on this project, failures were occurring with the feet at approximately
300 kg load. Dupont has since fine tuned the materials and suggested other variations of
Delrin® (Appendix 3) which have the necessary strength properties while allowing enough
deflection in the feet to cause a termination of testing rather than catastrophic failures.
Results for the last 15 Delrin® injection molded feet are included in Appendix 7.
7.3 Cyclic Fatigue Testing
Once feet have passed the proof testing, they go onto cyclic fatigue testing. Feet are
currently tested to ISO Standards however in the future, the ISO tests may be modified to
allow for a better representation of the actual biomechanical forces generated throughout a
gait cycle.
At the present time, spot inspections on the cyclic testers are performed every half hour and
complete inspections are performed every four hours. Details of these inspections can be
found in Appendix 8.
- 26 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
After the first 100,000 cycles on the tester, Zytel® ST801 (Nylon) was discarded as a
potential candidate as it appeared that permanent plastic deformation occurred. Failure
occurred on the SACH foot during this time (the first 100,000 cycles) as well. All of the
Delrin® versions of the feet that have been tested to date have made it past this crucial point.
(Appendix 9)
Photos are taken daily of the feet being tested and compared to a profile of the feet as well as
compared to previous days photos. Reports are generated on a regular basis, several of
which are found in Appendix 10.
- 27 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
8. FUTURE WORK
8.1 Validation and Calibration of the NPO Fatigue Tester
Calibration of Axial Force Measurements
To ensure that the fatigue tester axial force measurements are accurate and repeatable, the
tibial shank used for cycling testing will be instrumented with strain gages arranged in the
Wheat Stone bridge configuration used during actual test runs (Fig. 18). The shank will be
placed in an Instron machine (model 1122), known axial compressive forces will be applied,
and the corresponding strain gage voltages recorded. This calibration is required annually
and must be accurate within +/- 1% (ISO 3-9).
Calibration of Cycling Frequency
To ensure that the cycling frequency of choice is maintained during testing to within +/- 10%
(ISO 3-9), annual calibration will be done using a mechanical stroke counter (model A,
McMaster-Carr, Dayton, NJ, www.mcmaster.com) and stop watch for several trials of 1
minute duration.
Comparison with Gait Analysis
To ensure that the NPO Fatigue Tester mimics adequately the axial tibial force levels and
waveforms experienced by the prosthesis in the field, measurements of axial forces will be
measured during cycling testing of the Niagara Foot and compared to data from two separate
gait studies, described below, one for normals and the other for below knee amputee subjects.
1. Fatigue Tester Force Measurements (to be done)
Test Procedure - using Wheat Stone bridge strain gage circuit, axial forces acting along the tibial shank of the prosthesis during cyclic fatigue testing of the Niagara Foot will be measured
Test Conditions - as described in Section 5.3 above
2. Gait Tests for Normals (completed)
Test Procedure - IRED (infra-red emitting diode) markers placed on lateral aspect of their test leg at bony landmarks; Watsmart (Northern Digital, ON) video system used to record motion / kinematic data; embedded force plate in raised walkway recorded force / kinetic data; motion and force data sampled at 50 Hz
Test Conditions - n = 34 subjects, average age (24.5 yrs), average weight (65.4 kg), average height (171 cm), level walking frequency = naturally chosen by subject
- 28 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
1
24
3
XY
Z
Note: 350 Ω Strain Gages
Shank
StrainGage
ForceFX
MomentMY
MomentMX
1 + + 02 - 0 -3 - 0 +4 + - 0
Fig. 18. Strain gage arrangement on the tibial shank for measuring axial tibial force (FX) and moments (MX and MY).
- 29 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
3. Gait Tests for Below Knee Amputees (to be done)
Test Procedure - the tibial shank will be instrumented with a Wheat Stone bridge strain gage circuit to measure tibial axial forces
Test Conditions - gait study performed on adult below knee amputees in level walking, Niagara Foot prosthesis used, number of subjects and trials per subject yet to be determined
8.2 Waveform Modifications
In the field, it is expected that the prosthesis will be subject to load levels typical of non-gait
conditions and events (e.g. kicking, climbing, rough terrain). In addition, the mechanical
durability of the Niagara Foot may be ‘successful’ in that it exceeds the 3 million cycles
recommended by ISO standards. Thus, one of the methods used to simulate non-gait events
and assess the failure mode of ‘successful’ prostheses, is to modify the load waveform (Fig.
16) either by gradually increasing peak-to-peak loading or the load baseline level imposed on
the foot until failure eventually occurs. Although, this does not accurately predict the fatigue
strength of the foot, it allows manufacturers to make inter-prosthesis comparisons by
observing how and at what load levels failures occur.
8.3 Environmental Durability of the Niagara Foot
Environmental Conditions
Because of the various environments the foot will potentially be exposed to, the danger of
severe material degradation of the Niagara Foot exists. To this end, the resistance of the foot
material to anticipated conditions such as thermal and UV sunlight exposure, humidity,
saltwater, micro-organisms and mechanical abrasion, should be tested. During previous
clinical studies (Potter 2000; Potter et al., 1999), a layer of polyurethane rubber was attached
to the sole of the Niagara 1 version of the Niagara Foot to provide some cushioning for the
client and for protection of the Hivalloy G7062 (20% glass-filled polypropylene) material
from abrasion. In order to assess the effect of more adverse conditions during gait, the bond
between the rubber sole and the foot should also be examined. To these ends, a series of
standard tests are available from the American Society for Testing and Materials (ASTM):
aerobic biodegradability of degradable plastics by micro-organisms (ASTM D5247) outdoor accelerated weathering of plastics using concentrated sunlight (ASTM
D4364)
- 30 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
response of rigid cellular plastics to thermal and humid aging (ASTM D2126) weathering of plastics under marine floating exposure (ASTM D5437) resistance of plastics to abrasion (ASTM D1242) peel or stripping strength of adhesive bonds (ASTM D903)
Related Materials Issues
The importance of examining material microstructure of the material before and after
mechanical fatigue and environmental resistance tests has been suggested. The detection of
structural change and weakness may be especially significant in areas where high stresses are
expected. One available method developed by our group (Dwyer, 1996) detects localized
regions of structural weakness by the level of oxidation present using an SO2 staining
technique. Finally, in order to minimize waste, further reduce foot cost, and catalyze
technology transfer to the host community, the possibility of using local waste plastics for re-
processing can be investigated (Bunton et al., 1999).
8.4 Complete Sized Modular System
The next project will be to design residual limb sockets in a range of sizes to fit an initial
target population. The majority of amputees are affected in the lower limb, below the knee
(BK). The most common approach to provide prostheses for these patients is to use a
modular system composed of a foot, pylon and socket. Feet are generally available in a
series of sizes and styles, while sockets require custom fitting to ensure proper load transfer
to the residual limb. All components are expensive by world standards, since the majority of
amputees are in developing countries. It is proposed that the use of modern design and
fabrication methods could greatly reduce these costs, increasing accessibility. Previous work
has produced a low cost foot component fabricated by injection molding; the specific
objective of this project is to design and test a sized modular socket system produced using
mass production methods.
The socket project has five phases as described below: socket sizing using statistical
methods, internal shape and size optimization to ensure fit, stress analysis and material
selection, initial prototype fabrication, and durability testing. The components will be
manufactured in Ontario for a global market of amputees that are currently under serviced by
expensive custom-made devices.
- 31 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
1. Collection of Anthropometric Data (NPO) Development of sampling method Statistical analysis of data set
2. Sizing Using CAD Interference Method (Queen’s University) Importing of data and data reduction in CAD program Determination of characteristics of fit Determination of suitable size range for sample population
3. Stress Analysis of Socket-Pylon Interface (Queen’s University) Simulation of loading at socket-pylon interface Material characterization
4. Prototyping (NPO) Creation of forms for various sizes for the chosen manufacturing method Verification of fit using blown liners
5. Mechanical Testing (Queen’s University) Modification of tester to accept residual limb form Verification of instrumentation Determination of tibial forces using gait analysis Test of initial set of sockets
8.5 Cosmetic Cover for Foot
In some cultures, the appearance of a prosthetic device is of almost equal importance as its
functionality. To this end, Boyer (2001) developed a cosmetic cover for the Niagara Foot,
which has a strictly cosmetic purpose and is not necessary for the proper function of the
prosthesis. The cover is a leather boot based on a positive mold of a natural foot the same
length as that of the prosthesis. The design is unisex, ambidextrous, simple, and low in cost.
The proposed manufacturing scheme involves local shoemakers, thereby minimizing
specialized training, and local materials, thereby minimizing cost. Distribution of a cosmetic
cover pattern with each prosthesis to local shoemakers is envisioned. A prototype cover has
been constructed and is available for focus group evaluation, from which modifications can
be made. The full report by Boyer (2001) is available upon request.
8.6 Modification of Test Stands
The current tester is a first working model made to ensure that all components functioned as
expected. As such, the stands for the stations and control panels are temporary. New stands
and panels will be constructed that are ergonomic, functional, and versatile (Appendix 6).
The proposed design concept is the result of an unpublished study (Fisher et al., 2001).
- 32 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
9. FINAL REMARKS
In many post-conflict and developing countries, prostheses are used on rough physical
terrain. The current mechanical fatigue tester uses a standardized cyclic loading regime that
mimics normal walking conditions and potential damage events (kicking, tripping,
negotiating rough terrain, etc.) Testing will be consistent with the voluntary international
standards developed by ISO (1996) and ISPO (1978), which have been used by few
researchers (Daher, 1975; Toh et al., 1993), two of which were undertaken by our research
group (Singleton, 1983; Wevers and Durance, 1987). The success of the NPO Fatigue Tester
could make Canada an international centre for durability testing of prosthetic lower limbs
and feet. Also, because the SACH prosthesis is commonly used in developing countries,
benchmark tests will be done for comparison with the Niagara Foot. Finally, since
preliminary tests show that the Niagara Foot’s durability and performance exceeds that of the
SACH, the Niagara Foot's cost - 10% of a low end $ 75 (U.S.) SACH - and manufacturing
ease, would make it the foot of choice.
- 33 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
REFERENCES Barth DG, Schumacher L, Thomas SS. “Gait Analysis and Energy Cost of Below-Knee Amputees Wearing Six Different Prosthetic Feet.” J. Prosthetics and Orthotics, 4(2):63-75, 1991.
Boyer K. A Cosmetic Cover for the Niagara Prosthetic Foot. MECH 461 Course Report, Mechanical Engineering Dept., Queen’s University, Kingston, ON, Canada, April 2001. Bunton, E, Kenney, S, Meilenner, C, and S Wilson, Low Cost Wheels, MECH 212 Design Project, Mechanical Engineering Dept., Queen’s University, Kingston, ON, Canada, 1999. Carlson JM, Day B, Berglund G. “Double Short Flexure Type Orthotic Ankle Joints.” J. Prosthetics and Orthotics, 2(4):289-300, 1990.
Daher RL, “Physical Response of SACH Feet under Laboratory Testing”, Bulletin of Prosthetics Research, 10(23):4-50, 1975. Dwyer, KA, Effect of Heterogeneous Structure on Ultra High Molecular Weight Polyethylene Failure Mechanisms in Total Joint Arthroplasty, PhD Thesis, Mechanical Engineering Dept., Queen’s University, Kingston, ON, Canada, 1996. Fisher J, Burnier E, Johansen C, Corley J. Test Stand for Fatigue Tester. MECH 212 Course Report, Mechanical Engineering Dept., Queen’s University, Kingston, ON, Canada, April 2001. ISO (International Organization for Standardization), Prosthetics – Structural Testing of Lower-Limb Prostheses, Reference No. ISO 10328:1-8 / 1996E, 1996.
ISPO (International Society for Prosthetics and Orthotics), Standards for Lower-Limb Prostheses – Report of a Conference 1977, 1978.
Macfarlane PA, Nielsen DH, Shurr DG, Meier K. “Gait Comparisons for Below-Knee Amputees using a Flex-Foot versus a Conventional Prosthetic Foot.” J. Prosthetics and Orthotics, 3(4):150-161, 1990.
Menard MR, Murray DD. “Subjective and Objective Analysis of an Energy-Storing Prosthetic Foot”. J. Prosthetic and Orthotics, 1(4):220-230, 1989.
Noce-Kirkwood R. Kinematic and kinetic analysis of the hip joint during level walking, stair climbing and exercise protocols, PhD Thesis, School of Physical and Health Education, Queen’s University, Kingston, ON, Canada, 1997.
Pitkin MR. “Mechanical Outcomes of a Rolling-Joint Prosthetic Foot and Its Performance in the Dorsiflexion Phase of Transtibial Amputee Gait.” J. Prosthetics and Orthotics, 7(4):114-123, 1995.
Potter DW. Gait Analysis of a New Low Cost Foot Prosthetic for use in Developing Countries. M.Sc Thesis, School of Physical and Health Education, Queen’s University, Kingston, ON, Canada, 2000.
Potter D, Costigan P, Bryant JT, and R Gabourie, "Clinical Gait Trial of a New Prosthetic Foot Design for Developing Countries", International Society of Biomechanics, 17th Congress, Calgary, Canada, Aug. 8-13, 1999.
Singleton JD, A Fatigue Tester for Below-Knee Prostheses, M.Sc. Thesis, Mechanical Engineering Dept., Queen’s University, Kingston, ON, Canada, 1983.
Toh SL, Goh JCH, Tan PH, and TE Tay, “Fatigue Testing of Energy Storing Prosthetic Feet”, Prosthetics and Orthotics International, 17:180-188, 1993.
Tsai A, “A Summary of the Initial Phase of the NPO Low Cost Modular Prosthetic Foot Project”, Technical Report, Clinical Mechanics Group (Kingston General Hospital), Queen’s University, Kingston, ON, Canada, Apr. 25, 1999.
Valenti TG. “Experience with Endoflex: A Monolithic Thermoplastic Prosthesis for Below Knee Amputees.” J. Prosthetics and Orthotics, 3(1):43-50, 1990.
Wevers HW and JP Durance, “Dynamic Testing of Below-Knee Prosthesis: Assembly and Components”, Prosthetics and Orthotics International, 11:117-123, 1987.
- 34 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
Wong P, Addison G, Austin M. Control System for Prosthetic Foot Cyclical Testing Machine. MECH 212 Course Report, Mechanical Engineering Dept., Queen’s University, Kingston, ON, Canada, April 2001.
Yamamoto S, Ebina M, Kubo S, et al. “Quantification of the Effect of Dorsi-Plantarflexibility of Ankle-Foot Orthoses on Hemiplegic Gait: A Preliminary Report.” J. Prosthetics and Orthotics, 5(3):88-94, 1992.
Ziolo T, Stress Analysis and Design Optimization of an Injection Molded Prosthetic Foot, Undergraduate Thesis, Mechanical Engineering Dept., Queen’s University, Kingston, ON, Canada, 1999.
- 35 -
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
APPENDIX 1 – ISO TESTING STANDARDS
FOR PROSTHESES
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
APPENDIX 2 – U.S. PATENT FILE
FOR THE NIAGARA FOOT
The NPO Fatigue Tester Ziolo, Zdero, and Bryant (2001)
Inlet Pressurepsi 0 psi 5 psi 10 psi 15 psi 20 psi 25 psi 30 psi 35 psi 40 psi
5 109 17 -74 -166 -258 -349 -441 -533 -610 218 127 35 -57 -148 -240 -332 -424 -515 328 236 144 52 -39 -131 -223 -314 -40620 437 345 253 162 70 -22 -114 -205 -29725 546 454 362 271 179 87 -4 -96 -18830 655 563 472 380 288 197 105 13 -7935 764 673 581 489 397 306 214 122 3140 873 782 690 598 507 415 323 231 14045 983 891 799 707 616 524 432 341 24950 1092 1000 908 817 725 633 542 450 35855 1201 1109 1018 926 834 742 651 559 46760 1310 1218 1127 1035 943 852 760 668 57665 1419 1328 1236 1144 1052 961 869 777 68670 1528 1437 1345 1253 1162 1070 978 887 79575 1638 1546 1454 1363 1271 1179 1087 996 90480 1747 1655 1563 1472 1380 1288 1197 1105 101385 1856 1764 1673 1581 1489 1397 1306 1214 112290 1965 1873 1782 1690 1598 1507 1415 1323 123295 2074 1983 1891 1799 1708 1616 1524 1432 1341
100 2184 2092 2000 1908 1817 1725 1633 1542 1450105 2293 2201 2109 2018 1926 1834 1742 1651 1559110 2402 2310 2218 2127 2035 1943 1852 1760 1668115 2511 2419 2328 2236 2144 2053 1961 1869 1777120 2620 2529 2437 2345 2253 2162 2070 1978 1887
The Net Force Exerted on Foot (N) Using 2.5" Bore Cylinders (CYCLIC TESTING)
Back Pressure
2415
Inlet Pressurepsi 0 psi 5 psi 10 psi 15 psi 20 psi 25 psi 30 psi 35 psi 40 psi
5 185 33 -118 -270 -421 -573 -724 -876 -102710 369 218 66 -85 -237 -388 -540 -691 -84315 554 402 251 99 -52 -204 -355 -507 -65820 738 587 435 284 132 -19 -171 -322 -47425 923 771 620 468 317 165 14 -138 -28930 1107 956 804 653 501 350 198 47 -10535 1292 1140 989 837 686 534 383 231 8040 1476 1325 1173 1022 870 719 567 416 26445 1661 1509 1358 1206 1055 903 752 600 44950 1845 1694 1542 1391 1239 1088 936 785 63355 2030 1878 1727 1575 1424 1272 1121 969 81860 2214 2063 1911 1760 1608 1457 1305 1154 100265 2399 2247 2096 1944 1793 1641 1490 1338 118770 2583 2432 2280 2129 1977 1826 1674 1523 137175 2768 2616 2465 2313 2162 2010 1859 1707 155680 2952 2801 2649 2498 2346 2195 2043 1892 174085 3137 2985 2834 2682 2531 2379 2228 2076 192590 3321 3170 3018 2867 2715 2564 2412 2261 210995 3506 3354 3203 3051 2900 2748 2597 2445 2294
100 3690 3539 3387 3236 3084 2933 2781 2630 2478105 3875 3723 3572 3420 3269 3117 2966 2814 2663110 4059 3908 3756 3605 3453 3302 3150 2999 2847115 4244 4092 3941 3789 3638 3486 3335 3183 3032120 4428 4277 4125 3974 3822 3671 3519 3368 3216
Back Pressure
Net Force Exerted on Foot (N) Using 3.25" Bore Cylinders (PROOF TESTING)