Upload
polly-gaines
View
218
Download
2
Embed Size (px)
Citation preview
The Importance of Commutability of Reference Materials Used as Calibrators: The Example of Ceruloplasmin
I. Zegers, R. Beetham, T. Keller, J. Sheldon, D. Bullock, F. MacKenzie, S. Trapmann, H. Emons, and H. Schimmel
September 2013
www.clinchem.org/content/59/9/1322.full
© Copyright 2013 by the American Association for Clinical Chemistry
© Copyright 2009 by the American Association for Clinical Chemistry
Ceruloplasmin measurementsCeruloplasmin measurements
Ceruloplasmin a major serum protein, about 150 kDa binds 6 copper ions
Clinical use decreased concentrations of ceruloplasmin may
indicate aceruloplasminemia, Menkes disease, or Wilson disease
Normal concentrations about 0.2 – 0.5 g/L
© Copyright 2009 by the American Association for Clinical Chemistry© Copyright 2009 by the American Association for Clinical Chemistry
Figure 1. The standardisation of ceruloplasmin measurements was based on the use of the common reference material ERM-DA470 (certified for 15 major serum proteins). The traceability chain involves also the manufacturer’s working calibrators and kit calibrators.
Traceability Traceability chain for chain for ceruloplasmin ceruloplasmin measurementsmeasurements
Routine sample result
Immunoassay
US National Reference Preparation
ERM-DA470 (matrix reference material)
Reference procedure, immunoassay ring trial
Value transfer 1: LC-MS, immunoassay
Value transfer 2: immunoassay
Immunoassay
Pure ceruloplasmin
Manufacturer’s working calibrator
Manufacturer’s kit calibrator
© Copyright 2009 by the American Association for Clinical Chemistry© Copyright 2009 by the American Association for Clinical Chemistry
Figure 2. In 2002, Beetham et al. wrote a Letter to the Editor of Clinical Chemistry. Observations from the UK National External Quality Assessment Service (NEQAS) showed that for ceruloplasmin discrepancies remained between results from different methods. This was the case despite the fact that all measurement results were traceable to value of the reference material ERM-DA470 (formerly called CRM 470). For other proteins certified in ERM-DA470 discrepancies were reduced after its generalised use.
R. Beetham, P. White, P. Riches,D. Bullock, F. MacKenzie, and for UK NEQAS Protein Assays Specialist Advisory Group. Use of CRM 470/RPPHS Has Not Achieved True Consensus for Ceruloplasmin Measurement. Clin Chem 2002; 48: 2293-2294.
UK NEQAS observes discrepanciesUK NEQAS observes discrepancies
© Copyright 2009 by the American Association for Clinical Chemistry
CommutabilityCommutability
Commutability of reference materials (RM) is an essential requirement to achieve comparability of patient results
It is required when a RM is used as calibrator or as EQAS sample
Commutability is a property of a reference materialsuch that values measured for that reference materialand for representative clinical samples have the samerelationship between two, or more, measurementprocedures for the same measurand.
© Copyright 2009 by the American Association for Clinical Chemistry
Materials and MethodsMaterials and Methods Sets of samples containing 3 reference materials (RMs)
and 30 serum samples from healthy people were measured
The following RMs were evaluated: ERM-DA470 (freeze-dried)
certified for 15 serum proteins, distributed from 1994 to 2008 the use of this material led to a significant reduction of
between-method variability for proteins like IgG, albumin, etc.
ERM-DA470k/IFCC (freeze-dried) certified for 12 serum proteins, distributed since 2008 not certified for ceruloplasmin concentration
ERM-DA472/IFCC (liquid frozen) certified for CRP, replaced by ERM-DA474/IFCC in 2011
© Copyright 2009 by the American Association for Clinical Chemistry© Copyright 2009 by the American Association for Clinical Chemistry
Table 1. Methods used to evaluate the commutability of reference materials for ceruloplasmin measurements.
Methods usedMethods usedPlatform Reagents Laboratory code
Abbott Architect c16000 Abbott Ceruloplasmin REF 6K9101 Lot 70185
Sentinel Diagnostics, Milano, IT Abbott
BN ProSpec Siemens reagents (N Antiserum to Human Ceruloplasmin)
Siemens Healthcare Diagnostic Products GmbH, Marburg, DE
ProSpec I
BN ProSpec Dade Behring anti-ceruloplasmin reagent lot: 155289, Dade Behring standard SL lot:083678
St. Georges Hospital, London, UK ProSpec II
Beckman Immage Beckman calibrant 2 lot:704575, Beckman anti-ceruloplasmin reagent lot:706567
St. Georges Hospital, London, UK Immage
Hitachi 917 Tina-quant C4-2, Id. 11 875 051, Lot 696 008: C4, ACN 032
Roche Professional Diagnostics, Penzberg, DE
Roche Hitachi
Roche Integra Ceruloplasmin, Id. 20 764 663, Lot 696825: CERU3, Test-Id. 0-666
Roche Professional Diagnostics, Penzberg, DE
Roche Integra
Hitachi 917 DAKO (Antibody: Q0121 Lot 00012259 Ceruloplasmin)
DAKO, Glostrup, DK DAKO Hitachi
Olympus systems analyser AU2700 OA28
Olympus OSR6164 # 5576 Olympus Clare (now Beckman Coulter Biomedical Ltd.), Clare, IE
AU2700
© Copyright 2009 by the American Association for Clinical Chemistry© Copyright 2009 by the American Association for Clinical Chemistry
Figure 3. Example of a pair-wise commutability assessment for ceruloplasmin (CER) measurements in sera from 30 healthy individuals (CS), ERM-DA470, ERM-DA470k, and ERM-DA472. The plotted values are the means per sample. The continuous black line is the results of a Deming regression, the dotted black lines correspond to the limits of the prediction interval at the 95% confidence level.
Results: example Results: example commutability commutability assessment assessment
The recovery of ceruloplasmin in ERM-DA470 is reasonable with both the DAKO Hitachi and Beckman Immage methods (certified concentration 205 ± 11 mg/L)
ERM-DA470 is not commutable for this combination of methods
© Copyright 2009 by the American Association for Clinical Chemistry© Copyright 2009 by the American Association for Clinical Chemistry
Table 1. Comparison of results for serum samples and commutability of reference materials. The upper-right part of the table contains results from the evaluation of the commutability of ERM-DA470. The lower-left half of the table contains the slope of Passing-Bablok regression. a1 indicates commutability, 0 indicates a lack of commutability.
Commutability of ERM-DA470Commutability of ERM-DA470
© Copyright 2009 by the American Association for Clinical Chemistry
The slopes of the correlation lines vary from 0.56 to 1.42, showing that there are large discrepancies
The largest deviation from unity is between results from DAKO Hitachi and Beckmann Immage
ERM-DA470 is not commutable for many combinations of methods
ERM-DA470 is commutable for clusters of methods: - Abbott Architect and BN ProSpec- AU270, DAKO Hitachi and Roche Integra
Main findings:Main findings:
© Copyright 2009 by the American Association for Clinical Chemistry
Reference materials are stabilised materials stored for long times. The stabilisation of ERM-DA470 included the addition of stabilising agents like sodium azide.
Ceruloplasmin is a complex molecule that requires the binding of 6 copper ions. It can undergo structural changes.
Crossed immunoelectrophoresis shows that in ERM-DA470 a second form of ceruloplasmin is present that is not found in fresh serum. At present biochemical data on are not conclusive about the nature of differences between the forms.
What causes the lack of commutability?What causes the lack of commutability?
© Copyright 2009 by the American Association for Clinical Chemistry© Copyright 2009 by the American Association for Clinical Chemistry
Figure 4. Results from measurements of samples distributed by UK NEQAS. Fresh samples: Results from different methods were very discrepant. Aged Samples (distributed after 6 months storage in the presence of sodium azide): Consistent results. ALTM stands for all-laboratory trimmed mean, Roche stands for Roche Integra, Dade for BN ProSpec and BN II, Array for the Beckmann Array and Immage for the Beckman Immage.
Effect of prolonged storage of EQAS samplesEffect of prolonged storage of EQAS samples
© Copyright 2009 by the American Association for Clinical Chemistry
The reference material ERM-DA470 behaved like an aged serum sample
All methods gave results that were traceable to the ERM-DA470 value
Methods gave equivalent results for aged serum samples, like UK NEQAS samples stored in the presence of sodium azide
The methods gave discrepant results for fresh samples
Use of a non-commutable RM causes discrepanciesUse of a non-commutable RM causes discrepancies
US National Reference Preparation
ERM-DA470 (matrix reference material
Pure ceruloplasmin
Manufacturer’s working calibrator
Manufacturer’s kit calibrator
External quality assessment sample
© Copyright 2009 by the American Association for Clinical Chemistry
ConclusionsConclusions The reference material ERM-DA470 was not commutable for
ceruloplasmin. Therefore, its generalised use did not lead to equivalent results for ceruloplasmin (in contrast to most of the other parameters certified in ERM-DA470).
A reference material format may be appropriate for certain analytes and not for others.
UK-NEQAS has developed procedures for preparing commutable EQAS samples. These might also be used for preparing a commutable reference material for ceruloplasin.
Commutability of reference materials is an essential requirement to achieve comparability of patient results.
© Copyright 2009 by the American Association for Clinical Chemistry
Thank you for participating in this month’sClinical Chemistry Journal Club.
Additional Journal Clubs are available atwww.clinchem.org
Download the free Clinical Chemistry app on iTunes for additional content!
Follow us