Upload
clinton-lee-hudson
View
220
Download
0
Embed Size (px)
Citation preview
The gas system for the CMS-RPC detector
Roberto GuidaCollaborazione CMS-RPC
Riunione CSN1 Roma, 2 Aprile 2007
•Introduction to the CMS-RPC gas system
•R&D studies on the closed loop gas system: •Development of the purifier module
•Gas monitoring systems:•Gas quality monitoring system •Gas gain monitoring system
•Conclusions
Outlook
Characteristics and requirements
Total volume of the CMS-RPC detector: 18 m3
Gas mixture: 96.2 % C2H2F4, 3.5 % iC4H10, 0.3% SF6
Mixture relative humidity: 40%
Number of channels: 360
Nominal gas flow: 8 m3/h (0.5 volume change/h)
Gas recirculation factor (CLOSED LOOP MODE): ~95%
Fresh gas replenishing rate: 1 m3/h (~5%)
Humidity
SuppliersMixerHumidifier
Purifier module
Pre-distribution and pump
Final distribution
Layout of the system
Schedule
P um pM o d u le
P urifi erM o d u le
M ix er C ircu latio nM o d u le.
E x h aust
SG - B u ild in g
B uff er
P rim arySu p p ly
U SC - area
M ain P ip es / len gth = 70 mSu p p ly : 1 1 x Ø = 14/ 1 6 m mR etu rn : 11 x Ø = 20 / 22 m m
U X - C av ern
Pre-
Dis
trib
utio
nPr
essu
re R
egul
.
U X - D istrib u tio nR ack s
H u m id ityC o n d itio n in g
Apr-May 07(installation)
May 07(CL commissioning)
almost ready
• All gas racks are installed except purifiers and analysis
• All racks are leak tested and functionally tested in the lab
• Control software 95% ready• PVSS user interface ready
Gas racks layout
RPC Modules: Pre-distribution; Pump; Analysis; Control rack
RPC Modules:Mixer; Exhaust; Humidifier; Purifiers; Analysis; Control rack
SG5 (surface) USC55 (service cavern)
…at the end the RPC system will be the biggest
CMS UX cavern
The last step is the final distribution of the gas mixture on the experiment
11 gas distribution racks:•50 channels each •Gas flow read-out by means of flowcells•Manual adjustment between channels
A careful equalization of the gas flow is needed in order to exploit the full capacity of the system
-Between stations (i.e. wheel level): •Results independent on the number of channels connected•Optimized at the minimum pressure needed •The key parameter is the number of volume change/h
-Station level:When needed a needle valve has been inserted in order to equalize the flow between the two chambers
Final gas distribution
Present specificationsThe system described is the result of an intense R&D program. In fact,the results from the GIF ageing test carried out by the CMS-RPC and ATLAS-RPC collaborations (1999-2004) as well as the test of many RPC with cosmic have implied an important evolution in the requirement for the RPC gas system and mixture composition in CMS:
Today CMSin 1999/018
Gas volume 18 m3 18 m3
Gas mixture R134a/iC4H10/SF6
96.2/3.5/0.3
R134a/iC4H10
96.5/3.5
Tollerable contamination: H2O
Mixture RH 40%
(7,000-10,000 ppm)
<1000 ppm
Chamber pressure <3 mbar <3 mbar
Maximum flow rate 18 m3/h 2.2 m3/h
Flow rate at operating conditions
8 m3/h 1.2 m3/h
Fresh gas replenishing ≤1 m3/h <0.12 m3/h
• The CMS-RPC collaboration started an intense R&D program since 2003 using a closed loop gas system prototype, provided by the CERN gas group, connected with two final RB1 detectors under irradiation at the CERN-GIF.
• After, in a second phase (started in 2005) the system was moved to the ISR area and used for the mass production chamber test. The aim of this phase was a further optimization of the purifier effectiveness (Preprint LNF 06/27).
Phase 1: test at the GIF facility Phase 2: test at ISR
• Further studies of the materials properties: Bakelite, purifiers, gas mixture
R&D on the closed loop system
• The purifier module is a fundamental part of the closed loop gas system.
• 1999: Standard purifier using a twin column (6 l) filled with e.g. Molecular sieve and activated copper can be used to remove water and oxygen. Since the expected contaminant were H2O and O2
• 2003: first hints from the GIF first results. However the idea of an upgrade was not considered in detail.
• 2005: The final results from the GIF ageing test have confirmed that:– RPCs need a humidified gas mixture (RH about 40%)– The gas flow should be 0.5 – 1 volume change/hour– New component in the gas mixture is needed (0.3% SF6)– The outlet gas from a reference chamber (not irradiated) has a similar gas
chromatogram as the inlet gas (except air and H2O)– In the outlet gas from a chamber inside GIF 5-7 extra components are
visible in the gas chromatogram – The best configuration found for the purifier is:
1. Molecular Sieve 5A 2. Cu-Zn / Cu3. Ni – Al2O3
Purifier R&D on the closed loop system
Gas chromatogram: GIF results
Fresh mixture
purifiers
Closed loop recirculation (95%) after about one month: many small extra signals are visible especially at low retention time
Results obtained with a MicroGC Varian CP2002P (PoraPLOT Q and TCD detector).
Fresh mixture
Mixture in the closed loop circuit
Final RPCs in closed loop mode with cosmic ray
System stable from November 2005, but just before the purifier regeneration …
slight increase in the currents. Reversible after regeneration.
Log scale
Gas chromatogram: ISR results
Usually no trace of impurities and pollutants in GC analysis.Seldom something is visible only in log scale.
Upgrade of the Purifier Module
Fill
Empty
Ø=16
Mol.Sieve
5Å
Mol.Sieve
5Å
Fill
Empty
Ø=16
Cu-ZnBASFR3 12
CuBASF
R311-G
Cu-ZnBASFR3 12
CuBASF
R311-G
fromBuffer
to ExhaustModule
Buffer Volume
Free Rack placefor future upgrade of
third purifier.
First possibility: 2 purifiers, the second one including the two metallic filters In case of necessity a further upgrade including a third module will be considered for the LHC high luminosity phase
An upgrade of the purifier module is mandatory for both capacity and purifiers material reasons
• Purifier #1: Molecular Sieve 5, 24 l cartridge. Expected lifetime at operating conditions 1.5 day of running time
• Purifier #2: Cu-Zn/Cu and Ni-Al2O3 Adsorption capacity x 50 with respect to the GIF setup.
Expected lifetime at operating conditions 15 day
P #1 P #2 P #3
?
• 8 double gaps are now under test in ISR (they are equivalent to 4 RB1 detectors)
• We started with a open mode gas system and we will move soon in closed loop mode
• Results expected in 1-2 months
• New more gaps in 2 months (?)
ISR test: second phase, a dedicated test
Development of the gas system
Extra signals in the closed loop gas
Increase of the gas flow
New component (SF6)
Humidified mixtureHumidity
Gas monitoring for CMS-RPC
We foreseen two systems able to monitor the gas mixture for the RPC detector:
• Gas quality and composition monitoring system based on some chemical analyses of the gas: • F- electrode • pH electrode• Pick-up point for a possible use of a Gas Chromatograph (GC)
• Gas gain monitoring system based on three sets of small RPCs supplied with fresh mixture, input mixture to CMS, return mixture from CMS respectively. It will be also possible to scan all the gas lines.
Results from both the monitoring systems need be recorded in the DCS system.
Gas Monitoring Systems
USC55UXC55SG5
RPCs #1Fresh gas
Purifiers
Mixer
Humidifier
GCF-
pHsample
RPCs #2Before purifiers
RPCs #3After purifiers
Gas quality monitoring
Gas gain monitoringGC+others(half wheel pick-up points)
CMS
Specific for fluoride/HF concentration monitoring.
Analysis station with two independent input channels (possibility to monitor simultaneously different sampling points).
F- electrode
GIF results
Low-cost and very simple system, with PC readout, already used during tests at GIF, in our labs and in ISR … nevertheless quite sensitive
y = 5,29E-09x - 5,91E-06
R2 = 9,75E-01
0,E+00
5,E-06
1,E-05
2,E-05
2,E-05
3,E-05
0 1000 2000 3000 4000 5000 6000
time (min)
[HF
]
Serie1
Lineare (Serie1)
pH meter
Gas Chromatograph
purifiers
GC Results:Signals from impurities produced by RPCs working in Closed loop mode (95%) after about one month.
GC is the most flexible system it has already demonstrated at GIF its ability to detect produced pollutants in the gas mixture
A monitoring system of the RPC working point able to provide much faster and sensitive response than the CMS RPC system
Monitor of efficiency and charge continuously Monitoring of charge and efficiencies with cosmic rays in SG5 (surface
gas building) Three sub-sets of 45x45 cm2 Single Gaps
1 Reference RPCs, i.e. fresh mixture
2 Monitor “IN” RPCs, i.e. mixture for CMS-RPC
3 Monitor “OUT” RPCs, i.e. mixture downstream of CMS-RPC
In case work point changes– A warning goes on action needed
– The Gas quality monitoring system (i.e. GC,probes, etc) will understand the reason
Gas gain monitoring system
The R&D program carried out by the RPCs community since 1999 has allowed to found the proper solution in order to guarantee the detector functionality for a long term period (increase of the gas flow, humidified mixture, a more complex purifier module, ….)
However the results of this R&D program have produced many important changes on the CMS-RPC gas system
Now the system is much more complex, but the changes are effective
Further studies are needed especially for the purifier module (most of this work is already going on or well defined)
Conclusions