Upload
others
View
3
Download
0
Embed Size (px)
Citation preview
The Extraction of Perfluorinated Alkyl Acids
(PFAAs) in Drinking Water using Automated
Cartridge SPE
Key Words
PFOS, PFOA, perfluorinated alkyl acids, PFAA, drinking water, EPA 537,
ISO 25101:2009
Russ Wolff and Craig Caselton, Northern Lake Service, Inc., Crandon, WI, USA; Alicia Cannon and Michael Ebitson, Horizon Technology, Inc., Salem, NH, USA
Introduction
Perfluorinated chemicals (PFCs), also known as PFAAs, can be found in commercial and industrial uses such as firefighting foams, package
material coatings, nonstick cookware, waterproofing and stain proof fabrics. PFCs are all manmade and have unique properties such as
repelling water, oil and heat. They are very difficult to break down and persist in the environment. Small amounts of PFCs can dissolve in
water and is a cause for concern for human exposure, especially in drinking water sources such as large public water systems and private
wells.
PFC contamination poses risks to the developmental, immune, metabolic, and endocrine health of consumers.1 The largest
potential source of human exposure is through drinking water. To better understand PFC occurrence levels throughout the US
the US EPA included six of these chemicals into the third Unregulated Contaminant Monitoring Rule (UCMR 3) in 2012. UCMR 3
required monitoring for 30 contaminants (28 chemicals and two viruses) between 2013 and 2015 using analytical methods
developed by EPA, consensus organizations or both. This monitoring provides a basis for future regulatory actions to protect
public health. The data summary from the extensive study found 0.9% of the public water supplies studied showed
concentrations of PFOS greater than the reference level (0.07 µg/L).2 Of the public water supplies studied 0.3% showed levels
of PFOA above the reference level (0.07 µg/L). As a result of this study, a Health Advisory has been issued by US EPA. Water
utilities should notify customers if greater than 70 ppt (0.07 µg/L) PFOS or PFOA or a total for the two combined are detected in
the water supply.3
The US is not the only country concerned about exposure to PFCs.4 In Europe, the twelfth meeting of the United Nations
Persistent Organic Pollutants Review Committee was held in Rome in September 2016 to move the consideration of PFCs
forward for further regulatory consideration.5 Rules developed under this framework will have global impact.
PFCs have become ubiquitous in the environment and in the laboratory. As very nonreactive compounds they are included in
equipment parts, tubing and other commonly found laboratory supplies. Therefore, it is a challenge to prepare samples and
measure low concentrations without contamination and interferences. This work will demonstrate the capability of the
SmartPrep® Cartridge Extraction II System for the extraction of six PFCs in compliance with US EPA 537 or ISO 25101:2009.6,7
Low concentrations will be measured with strict contamination control and reproducible performance demonstrated through
the entire analytical process.
Page 2
The perfluorinated compounds include:
perfluorooctanesulfonic acid (PFOS)
perfluorooctanoic acid (PFOA)
perfluorononanoic acid (PFNA)
perfluorohexanesulfonic acid (PFHxS)
perfluoroheptanoic acid (PFHpA)
perfluorobutanesulfonic acid (PFBS)
These six compounds were selected in the third UCMR although other similar
compounds will likely be captured with the same methodology.
Experimental
Solid phase extraction (SPE) is used to remove and concentrate the PFCs from
water samples. In US EPA method 537, a 0.5 g, 6-mL SPE cartridge containing styrene
divinylbenzene (SDVB) sorbent phase is specified. The analysis is done using HPLC/
MS/MS for a sensitive detection step.
Automated conditioning, application of sample and elution was done using the SmartPrep® Cartridge Extractor II (Horizon
Technology, Inc.) configured to handle 6-mL cartridges with 20-mL vessel collection tray. The original sample bottle was
rinsed automatically with the following procedure:
The configuration for the rinse kit was enabled in the software.
Sample lines 1-6 (how sample is loaded) were configured normally to load the samples onto the cartridge.
The sample lines that rinse the container (samples lines 7-12) were connected directly to the shower head, omitting the extra tubing from the rinse kit.
The extraction procedure implemented with the SmartPrep is summarized in Figure 1. The details of the method necessary to enter into the software are included in the Appendix.
A Strata® SDBL 100 µm Styrene
divinylbenzene, 6-mL cartridge was used
(Phenomenex).
The extracts were evaporated using an N
-Evap 112 (Organomation) to dryness
and then reconstituted in 1 mL of 96:4
methanol/water.
The HPLC system used was a Prominence
System (Shimadzu) with Atlantis® dc18, 5
µm, 2.1 x 150 mm HPLC column
(Waters). The mass spectrometer used
was an API4000 LC/MS/MS (SCIEX) for
the analysis step.
SmartPrep Automated Extraction
System II
Figure 1. Method Summary for the SmartPrep Extractor II.
Reagents
Reagent Water
Methanol
Method Summary
1. Prior to extraction the 250 mL samples must be preserved with a dechlorinating reagent.
2. Do not transfer the sample to another bottle as some of the PFAAs may adsorb to surfaces.
3. Add all appropriate Standards and Surrogates, then cap and invert sample bottle to mix.
4. Load each sample into a position on the SmartPrep extractor with sample sip tube and rinse cap.
5. Load the sample method and sample name in the sequence for each sample to be extracted.
6. Once the sequence is initiated, all the appropriate conditioning, air-dry, rinsing and elution steps started will be fully
automated.
7. Once the extracts are collected, they must be concentrated to dryness under a gentle stream of nitrogen to remove the
water/methanol mixture.
8. Once the extracts are evaporated to dryness,
9. Then add Internal Standard and the appropriate amount of methanol: water solution (96:4%) to bring up to a final volume of
1.0 mL.
10. Follow appropriate techniques (EPA method 537, in this case) for the storage of extracts.
11. Analyze by LC/MS/MS.
Results and Discussions
Prior to running field samples, an Initial
Demonstration of Capability (IDC) is required. Within
the results, five of the minimum acceptable QC criteria
are demonstrated with the SmartPrep Cartridge
Extractor System.
Table 1 shows a Laboratory Fortified Blank (LFB)
extracted with the SmartPrep Extractor System. The
SmartPrep II has FEP tubing as standard and this will
minimize any contamination. All of the sample, rinse
and solvent lines pass through the same valves and
tubing within the instrument. This table demonstrates
one instrument valve position (as per method 537)
and all of the SmartPrep’s internal components,
solvents and the SPE cartridge to demonstrate low
background contribution. In addition the surrogate compounds are within the range specified of 70-130 % recovery.
Figure 2, on the following page, shows a chromatogram of the LFB. It contains the Surrogates and Internal standards as well as a
non-detectable PFOS peak. The Final Concentration of PFOS is well below the DL (0.0037 µg/L, determined by DL in Table 3) and is
not reportable.
Page 3
Table 1. Initial Demonstration of Low System Background
Compound Area INST ( ng/L) Final Conc. (ng/L)
PFBS 0.00 0 0
PFHpA 0.00 0 0
PFHxS 0.00 0 0
PFOA 0.00 0 0
PFNA 0.00 0 0
PFOS 1177 71.6 0.290
SUR C13-PFHxA 176007 10.1 µg/L 101%
SUR C13-PFDA 404135 9.42 µg/L 94.2%
C13-PFOS-(ISTD) 166900 10.0 µg/L 0.0400 µg/L
C13-PFOA-(ISTD) 561125 10.0 µg/L 0.0400 µg/L
Table 2 confirms that the upper and lower limits for the prediction interval of result (PIR) met the recovery criteria for the six PFCs.
Each compound is within the limits and shows consistency throughout the seven runs. The upper criteria’s are set at <= 150% and
the lower PIR is >= 50%.
Table 3 shows the Method Detection Limit (MDL) determined from seven samples run on the SmartPrep over a three day period.
The MDL values for this method with this sample size and equipment are extremely low. They will be able to measure water
samples at the US Health Advisory level of 0.07 µg/L for PFOS or PFOA with MDLs of 0.004 and 0.002 respectively. The limit of
quantitation (LOQ) is 0.013 µg/L for PFOS and 0.0076 µg/L for PFOA, which is acceptable for very low level determinations.
Page 4
Table 2. Minimum Reporting Limit (MRL) Confirmation (ng/L)
Table 3. MDL Study
Compound 1 2 3 4 5 6 7 AVG SD HR for PIR
Upper PIR Limit (%)
Lower PIR Limit (%) CONC
PFBS 64.7 73.6 64.9 70.5 70.9 72.9 68.3 69.4 3.58 14.2 92.9 61.4 90.0
PFHpA 7.94 9.14 7.99 9.16 9.09 8.76 8.44 8.65 0.532 2.11 107 65.4 10.0
PFHxS 21.8 25.2 22.0 25.6 26.3 25.6 25.0 24.5 1.80 7.15 116 63.4 27.3
PFOA 16.1 18.2 16.3 18.3 18.5 17.9 16.9 17.5 0.987 3.91 107 67.7 20.0
PFNA 16.8 18.2 17.0 19.8 19.7 19.0 18.7 18.5 1.19 4.70 116 68.8 20.0
PFOS 27.9 30.6 28.2 33.2 32.1 32.6 32.5 31.0 2.19 8.67 107 60.2 37.2
Compound MDL 1 MDL 2 MDL 3 MDL 4 MDL 5 MDL 6 MDL 7 MDL 8 AVG STD CONC. MDL
(µg/L) LOQ
(µg/L) CONC/MDL
PFBS 0.036 0.039 0.029 0.039 0.041 0.037 0.041 0.039 0.0375 0.0037 0.0450 0.0111 0.0371 4.05
PFHpA 0.004 0.004 0.003 0.004 0.004 0.004 0.004 0.004 0.0041 0.0003 0.0050 0.00099 0.00331 5.04
PFHxS 0.013 0.013 0.010 0.013 0.013 0.012 0.012 0.013 0.0124 0.0013 0.0137 0.00385 0.0128 3.55
PFOA 0.008 0.009 0.007 0.009 0.009 0.008 0.009 0.008 0.0084 0.0008 0.0100 0.00229 0.00764 4.37
PFNA 0.009 0.010 0.007 0.009 0.010 0.009 0.009 0.009 0.0090 0.0008 0.0100 0.00231 0.00772 4.32
PFOS 0.016 0.016 0.013 0.016 0.017 0.016 0.017 0.016 0.0157 0.0013 0.0186 0.00376 0.0126 4.94
Figure 2. Chromatogram of Low System Background on the SmartPrep Extractor II
Table 4. IDC Study, Measure of Precision and Accuracy (Spikes are in ng/L)
Table 5. Laboratory Control Spike
Page 5
Compound Result (ng/L)
Spike Amount (ng/L) Spike Recovery (%)
Acceptable Range (%)
PFBS 99.6 90.0 111 (50-150)
PFHpA 10.6 10.0 106 (50-150)
PFHxS 30.9 27.3 113 (50-150)
PFOA 20.5 20.0 103 (50-150)
PFNA 22.6 20.0 113 (50-150)
PFOS 39.3 37.2 106 (50-150)
Table 4 demonstrates four samples that were extracted using the SmartPrep Extractor II. This data can determine the Initial
Demonstration of Precision (IDP) and Accuracy (IDA). Each need a minimum of four mid-level range replicates. The IDP must meet a
RSD of less than 20% and the IDA must meet an average recovery of ± 30% of the compounds true value.
Each of the compounds pass the criteria for the IDP by demonstrating a range from 0.94%-3.41% RSD. The IDA is within the method
requirement of ± 30% for determining accuracy of the spiked amounts for each of the six PFCs.
Compound 1 2 3 4 AVG CONC. % Rec STD DEV -30% 30.0% RSD
PFBS 640 671 661 629 650 720 90.3 19.1 504 936 2.94
PFHpA 79.4 81.7 78.6 76.5 79.1 80.0 98.8 2.15 56.0 104 2.72
PFHxS 208 218 210 206 210 219 96.2 5.11 153 284 2.43
PFOA 152 155 153 153 153 160 95.9 1.44 112 208 0.94
PFNA 160 170 164 158 163 160 102 5.56 112 208 3.41
PFOS 278 287 274 269 277 297 93.3 7.68 208 386 2.77
SUR C13-PFHxA 34.8 36.3 36.8 34.6 35.6 40.0 89.0 1.09 28.0 52.0 3.06
SUR C13-PFDA 41.6 42.3 43.1 40.9 42.0 40.0 105 0.920 28.0 52.0 2.19
Table 5 and Figure 3 show the results for the Laboratory Control Sample (LCS). The sample is representative of a drinking water
matrix and spiked at a mid-range concentration. The LCS spike recoveries are well within the acceptable range specified by the
method of 50-150%.
Figure 3. Laboratory Control Spike
Table 6. Laboratory Fortified Sample Matrix and Duplicate (LFSM and LFSMD)
Compound Spike Amount
(ng/L) Field Sample Result LFSM (ng/L)
LFSMD (ng/L)
LFSM Recovery (%)
LFSMD Recovery (%) RPD
PFBS 90.0 ND 102 116 113 128 12.7
PFHpA 10.0 ND 8.98 10.5 89.8 105 15.5
PFHxS 27.3 ND 28.5 33.7 104 123 16.8
PFOA 20.0 ND 18.6 20.9 93.3 105 11.8
PFNA 20.0 ND 20.5 21.8 102 109 6.21
PFOS 37.2 ND 37.1 38.9 99.9 105 4.77
Page 6
Table 6 shows the results for a sample matrix and duplicate spiked with the analytes of interest. The agreement of the duplicates
is well within the < 30% criteria for concentrations spiked near native concentrations.
References:
1. Kyle Steenland, Tony Fletcher, and David A. Savitz, Epidemiologic Evidence on the Health Effects of Perfluorooctanoic Acid (PFOA), Environ Health Perspect. 2010 Aug; 118(8): 1100–1108. https://www.ncbi.nlm.nih.gov/pmc/articles/PMC2920088/.
2. US UCMR-3 Data Summary, https://www.epa.gov/sites/production/files/2016-05/documents/ucmr3-data-summary-april-2016.pdf, (accessed October 14, 2016).
3. Jennifer Morrison, Chronic exposure limit set for PFOA in drinking water http://cen.acs.org/articles/94/web/2016/05/Chronic-exposure-limit-set-PFOA-drinking-water.html (accessed October 14, 2016).
4. Cheryl Hogue, PFOA moves closer to global restrictions, C&EN, October 3, 2016 (http://cen.acs.org/articles/94/i39/PFOA-moves-closer-global-restrictions.html )
5. Twelfth meeting of the Persistent Organic Pollutants Review Committee (POPRC.12), http://chm.pops.int/TheConvention/POPsReviewCommittee/Meetings/POPRC12/Overview/tabid/5171/Default.aspx (accessed October 14, 2016).
6. USEPA Method 537, https://cfpub.epa.gov/si/si_public_record_report.cfm?dirEntryId=198984&simpleSearch=1&searchAll=EPA%252F600%252FR-08%252F092+ (accessed October 14, 2016).
7. ISO 25101:2009, http://www.iso.org/iso/iso_catalogue/catalogue_tc/catalogue_detail.htm?csnumber=42742
www.horizontechinc.com 16 Northwestern Drive, Salem, NH 03079 USA ▪ Tel: (603) 893 -3663 ▪ Email: [email protected]
AN1131610_01
Conclusion
PFC compounds are becoming an increasingly important concern as research into health effects uncovers other related illnesses.
As countries around the world move to regulate exposure through drinking water, reliable and accurate measurement methods
become critical.
Methods US EPA 537 and ISO 25101:2009 specify concentration with SPE and analysis using HPLC/MS/MS. Automation of the
cartridge SPE step is important to ensure the best reproducibility with the least chance of contamination. It also ensures the
efficient use of technician time through walk-away operation.
Demonstration of compliance with all requirements of the method was shown. Method detection limits are excellent and
sufficiently low to provide adequate results at the level of interest in the current Health Advisory and in future regulations. The
matrix spike and matrix spike duplicate show excellent spike recoveries at low concentrations. The difference between duplicate
spikes was less than 20% in all cases. The combination of automated SPE for sample concentration and LC/MS/MS for
determination was shown to be effective and contamination free for the low-level determination of PFCs.
Appendix
RA
TE 1
R
ATE
2
LIQ
UID
PU
RG
E TI
ME
Sip
or
Sip
or
SEN
SE
MIN
. EX
PEC
TED
SO
AK
[D
RY
, Rin
se]
V
OLU
ME
De
live
ry
De
live
ry
Tru
e (
ON
) V
OLU
ME
VO
LUM
E TI
ME
N2
Pu
rge
STEP
O
PER
ATI
ON
R
eag
en
t, V
en
t,
or
Sam
ple
(m
L)
(mL/
min
) (m
L/m
in)
Fals
e (
OFF
) (m
L)
(mL)
(s
ec)
(s
ec)
1
Co
nd
itio
n C
artr
idge
M
eth
ano
l 1
5
10
0
FA
LSE
0
0
1
0
0
2
Co
nd
itio
n C
artr
idge
R
eage
nt
Wat
er
18
1
0
0
FALS
E
0
0
0
0
3
Load
Lar
ge V
olu
me
Sam
ple
Sa
mp
le b
ott
le
0
75
1
0
FALS
E
30
0
10
00
0
0
4
Sam
ple
Bo
ttle
Rin
se
Rea
gen
t W
ater
2
.5
30
0
FA
LSE
0
0
0
3
5
Sam
ple
Bo
ttle
Rin
se
Rea
gen
t W
ater
2
.5
30
0
FA
LSE
0
0
0
3
6
Sam
ple
Bo
ttle
Rin
se
Rea
gen
t W
ater
2
.5
30
0
FA
LSE
0
0
0
3
7
Sam
ple
Bo
ttle
Rin
se
Rea
gen
t W
ater
2
.5
30
0
FA
LSE
0
0
0
3
8
Sam
ple
Bo
ttle
Rin
se
Rea
gen
t W
ater
2
.5
30
0
FA
LSE
0
0
0
3
9
Sam
ple
Bo
ttle
Rin
se
Rea
gen
t W
ater
2
.5
30
0
FA
LSE
0
0
0
2
.5
10
Lo
ad L
arge
Vo
lum
e Sa
mp
le
Sam
ple
bo
ttle
0
1
5
10
FA
LSE
1
5
20
0
0
11
N
2 P
urg
e Ti
mer
Ste
p
0
0
0
0
FALS
E
0
0
0
5
12
Sa
mp
le B
ott
le R
inse
M
eth
ano
l 2
.5
30
0
FA
LSE
0
0
0
3
13
Sa
mp
le B
ott
le R
inse
M
eth
ano
l 2
.5
30
0
FA
LSE
0
0
0
3
14
Sa
mp
le B
ott
le R
inse
M
eth
ano
l 2
.5
30
0
FA
LSE
0
0
0
3
15
Sa
mp
le B
ott
le R
inse
M
eth
ano
l 2
.5
30
0
FA
LSE
0
0
0
3
16
Lo
ad S
amp
le F
or
Cle
anu
p
Sam
ple
bo
ttle
2
0
20
8
FA
LSE
0
2
0
1
5
17
C
lean
Sam
ple
Lin
e
Met
han
ol
6
10
0
FA
LSE
0
0
0
0
18
C
lean
Sam
ple
Lin
e
Rea
gen
t W
ater
6
1
0
0
FALS
E
0
0
0
10
19
C
lean
Plu
nge
r M
eth
ano
l 6
1
0
0
FALS
E
0
0
0
0
20
C
lean
Plu
nge
r R
eage
nt
Wat
er
6
10
0
FA
LSE
0
0
0
1
0
Detailed SmartPrep Extractor II Method for PFCs.