2
Data Sheet IMS-Analyzer The environmental and industrial IMS multi-gas analyzers The IMS-Analyzer is a multi-gas analyzer as Ion Mobility Spectrometer (IMS) with high sensitivity in the lower ppb-range and high resolution. With this trace-gas analyzer toxic gases can be detected and identified without any enrichment directly in situ already at a very low concentration level. For the further enhancement of the selectivity and further reduction of cross sensitivities the IMS- Analyzer is also available as GC-IMS approach. There the IMS is coupled with a gas chromatographic column (GC) for pre-separation. price / performance ratio which makes it quite interesting for replacements of expensive process gas chromatographs. This unique combination of GC column with IMS to a GC-IMS is also offered as mobile analyzer and is the smallest GC-IMS analyzer available on the market. With the IMS-Analyzer, IUT offers a high sensitive, high selective, reliable, and easy to operate analyzer. The analyzer offers an excellent Analysis Application • Ambient Air Monitoring • Process Gas Monitoring • Continuously Emission Control (Stack) • Semiconductor Fabrication • Environmental Monitoring • Medical / Pharmaceutical Monitoring Features • Real-time analysis of Toxic Industrial Compounds (TICs) • High sensitivity in lower ppb-range • Superior selectivity and cross sensitivity rejection • Enhancement with GC-column coupling (GC-IMS) • Analysis directly in situ (No Tedlar bag sampling / no lab analysis) • Operation under ambient air pressure • Low costs of operation, no consumables, no carrier gas required Ion Mobility Spectrometry Photo Ionization Detector Gas Chromatography

The environmental and industrial IMS multi-gas analyzers · The environmental and industrial IMS multi-gas analyzers ... The environmental and industrial IMS multi-gas analyzers

  • Upload
    ngoque

  • View
    223

  • Download
    0

Embed Size (px)

Citation preview

Data Sheet

IMS-Analyzer

The environmental and industrial IMS multi-gas analyzers

The IMS-Analyzer is a multi-gas analyzer as Ion Mobility Spectrometer (IMS) with high sensitivity in the lower ppb-range and high resolution. With this trace-gas analyzer toxic gases can be detected and identified without any enrichment directly in situ already at a very low concentration level.

For the further enhancement of the selectivity and further reduction of cross sensitivities the IMS-Analyzer is also available as GC-IMS approach. There the IMS is coupled with a gas chromatographic column (GC) for pre-separation.

price / performance ratio which makes it quite interesting for replacements of expensive process gas chromatographs.

This unique combination of GC column with IMS to a GC-IMS is also offered as mobile analyzer and is the smallest GC-IMS analyzer available on the market.

With the IMS-Analyzer, IUT offers a high sensitive, high selective, reliable, and easy to operate analyzer. The analyzer offers an excellent

Analysis Application

• Ambient Air Monitoring

• Process Gas Monitoring

• Continuously Emission Control

(Stack)

• Semiconductor Fabrication

• Environmental Monitoring

• Medical / Pharmaceutical

Monitoring

Features

• Real-time analysis of

Toxic Industrial Compounds (TICs)

• High sensitivity in lower ppb-range

• Superior selectivity and cross sensitivity

rejection

• Enhancement with GC-column

coupling (GC-IMS)

• Analysis directly in situ

(No Tedlar bag sampling / no lab analysis)

• Operation under ambient air pressure

• Low costs of operation, no consumables,

no carrier gas required

Ion Mobility Spectrometry

Photo Ionization Detector

Gas Chromatography

Data Sheet

Technical DataIMS-AnalyzerThe environmental and industrial IMS multi-gas analyzers

With this trace-gas analyzer toxic gases can be detected and identified without any enrichment directly in situ already at a very low concentration level. The detection including sampling and identification process takes few seconds, depending on the possible GC-Setup.

Compared with other systems the IMS-Analyzer has the advantage to operate under atmospheric pressure. This makes it best suitable for mobile applications for on-site measurements. The powerful rechargeable battery offers an optimal mobile use. Due to the special approach the sensitivity is extremely high. Unlike other IMS systems the IMS-Analyzer does not use dopants and does not use a membrane.

For the enhancement of the selectivity and further reduction of cross sensitivities the IMS-Analyzer is also available as GC-IMS approach. There the IMS is coupled with a gas chromatographic column (GC) for pre-separation. The GC-IMS version operates without carrier gas which keeps the operation costs low and makes it best suitable for the mobile use.

EX-protection (purge system) is available.

Calibration functions with an additional calibrator for SPAN/ZERO calibration is offered as well as Multi-Point-Sample (MPS) systems, dilution- or even gas probe chiller systems.

The IMS-Analyzer enclosures can range from Class 1 Division 1 to general purpose. Also

Detection principle:

Ion Mobility Spectrometry (IMS), or Gas-Chromatography-IMS (GC-IMS)

Ionization: Tritium (3H). No license or wipe test required

Measurablecompounds(only as example;other compounds possible):

Acetone, Ammonia, Acrylonitrile, Benzyl chloride, Bromine, Chlorine, Hydrogen chloride, Hydrogen bromide, Hydrogen fluoride, Dimethyl sulfate (DMS), Chlorocyane, CMME, Dibutyl phthalate (DBP), Hydrazine, Isocyanates, NMP, PFIB, Phosgene, Sulfur dioxide, Vinyl chloride

Sensitivity: Lower ppb range

Moisture: 0% to 90% RH

Data output:(depending on version)

• Integrated GUI• 4 - 20 mA (analog)• RS-232 (optional)• RS-485 (optional)• USB (optional)• LAN / WLAN (optional)• Digital in/out

Warm-up time: 20 minutes

Measuring time: Seconds to minutes (depending on GC setup)

Power: 100 V - 240 V / 50Hz - 60 Hzmobile: typically 8 hours with Li-ion battery

Carrier gas: No carrier gas required

Enclosure: 19'' rack mounted

Industrial enclosure (NEMA 4)

Mobile

Dimension: W:H:D:

483 mm135 mm420 mm

W:H:D:

500 mm500 mm210 mm

W:H:D:

280 mm100 mm280 mm

Weight: ~10 kg (~22 lbs)

18 kg (40 lbs.)

~7 kg (~15.5 lbs)

Operationtemperature:

0°C - 50°C(+32°F - 122°F)

-40°C - 50°C(-40°F - 122°F)

-10°C - 50°C(+14°F - 122°F)

Storage temperature:

-10°C - 50°C (+14°F - 122°F)

-10°C - 50°C (+14°F - 122°F)

-10°C - 50°C (+14°F - 122°F)

IUT Techmation B.V.

phone: +31 (183) 500 331fax: +31 (183) 500 342e-mail: [email protected]

Van Tienhoven v.d. Boogaardstraat 14aPO BOX 195; 4250 DD WerkendamNetherlands